Search Filters

Search Results

Found 11 results

510(k) Data Aggregation

    K Number
    K251547

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-10-31

    (163 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.

    Device Description

    The purpose of this submission is to expand the DESS Dental Smart Solutions for the DESSLoc Attachment system cleared under K170588, K191986, K212628, K222288, K240208, and K242340 and to:

    • include OEM platform compatibilities to the DESSLoc Abutment design that have been previously cleared in other DESS Abutment designs,
    • include new OEM platform compatibility for MIS C1 Dental Implant System,
    • include attachment components (retention inserts and housing) including reprocessing information in labeling.

    The DESSLoc Attachment System consists of abutments and device-specific accessories (retention inserts and denture housings) for resilient attachment of prostheses to endosseous dental implants. There have been no changes to the design of the DESSLoc abutments, the design is the same that has been cleared in the above submissions. The abutments are made of titanium alloy and coated with zirconium nitride (ZrN). The nylon retention insert is manufactured from Polynil® (polyamide 6.6) or Vestamid® Care ML GB30 (polyamide 12). The denture housing is made of titanium alloy with a machined surface or anodized surface. The DESSLoc abutment is compatible with OEM implants, as listed below.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252384

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-10-23

    (84 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DESS® Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.

    All digitally designed custom abutments for use with DESS® Pre-Milled Blank abutments are to be sent to a Terrats Medical validated milling center for manufacture.

    Device Description

    The purpose of this submission is to add abutments to the DESS Dental Smart Solutions system, which includes abutments cleared previously in K242340 and K240208. The subject device abutment designs include Pre-Milled Blank Abutments, Multi-Unit Abutments (straight and angled), and abutment screws.

    Pre-Milled Blank Abutments are designed for custom abutment fabrication by a CAD-CAM process. All patient-specific custom abutment fabrication is by prescription on the order of the clinician. The Pre-Milled Blank Abutments have a maximum (before milling) diameter of 10 mm or 14 mm and a solid cylindrical design with an engaging or non-engaging implant connection. They are available in versions with a straight screw channel and with an angled screw channel for esthetic considerations. The Pre-Milled Blank Abutments are manufactured from titanium alloy (Ti 6Al-4V).

    The design parameters for the CAD-CAM fabrication of a custom abutment from the Pre-Milled Blank Abutment are:

    • Minimum wall thickness – 0.45 mm
    • Minimum cementable post height for single-unit restoration – 4.0 mm (minimum cementable post height for single-unit restoration is defined as the height above the restorative margin)
    • Minimum gingival height – 0.5 mm
    • Maximum gingival height – 6.0 mm
    • Pre-Milled Blanks are for straight abutments only

    Multi-Unit Abutments: Straight and Angled are designed for attachment of multi-unit screw-retained restorations and are provided in three (3) designs, straight (0°), angled 17°, and angled 30°. The designs of the subject Multi-Unit Abutments are similar to the designs of Multi-Unit Abutments cleared in K242340 and K240208. All Multi-Unit Abutments are manufactured from titanium alloy (Ti-6Al-4V).

    The subject device straight Multi-Unit Abutments have a non-engaging, threaded design that attaches directly to the implant. The subject device straight (0°) Multi-Unit Abutments are included in this submission are similar in design to straight Multi-Unit Abutments cleared in K242340, K230143, and K222288. Subject device straight Multi-Unit Abutments are provided with a prosthetic platform diameter of 4.8 mm, and with a gingival height ranging from 1 mm to 5.5 mm.

    The subject device angled Multi-Unit Abutments are provided only in an engaging design that requires an abutment screw. The subject device Multi-Unit Abutments angled 17° and 30° are provided with a prosthetic platform diameter of 4.8 mm, and with a gingival height ranging from 2.5 to 4.5 mm.

    Select straight and angled Multi-Unit Abutments have an added coating of zirconium nitride (ZrN). This coating is identical to the ZrN coating used on abutments with ZrN cleared in K242340.

    DESS® Dental Smart Solutions Screws are designed to attach the abutment to the implant or the prosthesis to the abutment. There are a total of seventeen (17) subject device screws compatible with the subject device components or previously cleared components. The new screws have designs that are similar to those of screws cleared in K242340 and K240208. Screws are made of titanium alloy (Ti-6Al-4V). Select subject device screws are available with DLC (Diamond-like Carbon) coating and are identical to that used on screws cleared in K240208.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251280

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-07-23

    (90 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DESS® dental implants are indicated for surgical placement in the upper or lower jaw in edentulous or partially edentulous patients for anchoring or supporting tooth replacements to restore patient esthetics and chewing function. They are designed to support single or multi-unit restorations in splinted or non-splinted applications, as well as to support overdenture attachment systems. DESS® dental implants may be used for immediate or early implantation following extraction or loss of natural teeth, and may be used for immediate or delayed loading techniques. Implants may be loaded immediately when good primary stability is achieved and occlusal loading is appropriate.

    DESS® NEO GM Dental Implants are compatible with DESS® Dental Smart Solutions abutments having the identical NEO GM connection manufactured by Terrats Medical SL.

    DESS® NEO GM Dental Implants with a diameter of 3.5 mm are indicated for use in reduced interdental spaces, where there is not enough alveolar bone for a larger diameter implant. The use of 3.5 mm implants is intended only for rehabilitation of the anterior region of the mouth.

    DESS® Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.

    All digitally designed custom abutments for use with DESS® Pre-Milled Blank abutments are to be sent to a Terrats Medical validated milling center for manufacture.

    Device Description

    This submission includes dental implants that are compatible with DESS® Dental Smart Solutions abutments having the identical NEO GM connection manufactured by the sponsor of this submission, Terrats Medical SL, and cleared under K212628, K222288, K233316, K240208, and K242340. No claims of compatibility between the subject device implants and abutments from any OEM other than DESS® Dental Smart Solutions will be made. This submission also includes DESSLoc abutments that are compatible with eight (8) dental implant lines from five (5) OEM manufacturers. Also included in this submission is one (1) Pre-Milled Blank abutment that is compatible with three (3) implant lines manufactured by Alpha Dent Implants GmbH.

    This submission includes one implant line, the Dental Implant NEO GM, a series of self-tapping, threaded, root-form dental implants to be placed at bone level. The subject device implants are provided in body diameters of 3.55 mm, 3.75 mm, 4.0 mm, 4.3 mm, 5.0 mm, 6.0 mm, and 7.0 mm. The subject device implant body diameters will be labeled as 3.5 mm, 3.75 mm, 4.0 mm, 4.3 mm, 5.0 mm, 6.0 mm, and 7.0 mm. Implant with body diameters ranging from 3.55 mm to 5.0 mm are provided in overall lengths of 7.9 mm, 9.9 mm, 11.4 mm, 12.9 mm, 15.9 mm, and 17.9 mm. Implants with body diameters 6.0 mm and 7.0 mm are provided in overall lengths of 7.9 mm, 9.9 mm, and 11.4 mm. The subject device implant lengths will be labeled as 8 mm, 10 mm, 11.5 mm, 13 mm, 16 mm, and 18 mm. All subject device implants, regardless of body diameter, have an internal Morse taper connection with a 16° included angle and 2.99 mm diameter opening at the top of the implant. This NEO GM connection is identical to the connection for abutments cleared previously in K242340.

    All subject device implants are made of unalloyed titanium conforming ASTM F67 and ISO 5832-2. The entire endosseous surface, except for a small coronal bevel, features a grit blasted and double acid etched (SLA) surface, which is identical to the surface treatment for DESS® implants that were cleared in K212538.

    This submission also includes DESSLoc Abutments designed for overdenture retention. The subject device DESSLoc Abutments are straight, non-engaging abutments that attach directly to the implant and are compatible with eight (8) dental implant lines from five (5) OEM manufacturers.

    The subject device DESSLoc Abutments are manufactured from titanium alloy (Ti-6Al-4V) conforming to ASTM F136. The subject device DESSLoc abutments have a zirconium nitride (ZrN) coating produced by a physical vapor deposition (PVD) process. The ZrN coating is applied to increase the surface hardness and reduce wear of the abutment surface. The ZrN coating for the subject device DESSLoc Abutments is identical to the ZrN coating applied to DESSLoc Abutments cleared in K242340, K240208, K222288, K191986, and K170588.

    This submission also includes one (1) Pre-Milled Blank Abutment that is compatible with three (3) implant lines manufactured by Alpha Dent Implants GmbH, including Implant Active Konus, Implant Classic Konus, and Implant Active Bio, cleared in cleared in K210499. Reverse engineering compatibility analysis of the Alpha Dent implants, abutments, and abutment screws and Terrats Medical SL abutments and abutment screws was provided in the prior Terrats Medical SL submission K243212.

    The Pre-Milled Blank Abutment has a maximum (before milling) diameter of 10 mm and a solid cylindrical design and an engaging implant connection. The Pre-Milled Blank Abutment is manufactured from titanium alloy (Ti-6Al-4V) conforming to ASTM F136. All patient-specific custom abutment fabrication for the Pre-Milled Blank Abutment is by prescription on the order of the clinician and will be done at a Terrats Medical validated milling center under FDA quality system regulations. The design parameters for the CAD-CAM fabrication of a custom abutment from the Pre-Milled Blank Abutment are:

    • Minimum wall thickness – 0.45 mm
    • Minimum post height for single-unit restoration – 4.0 mm (post height measured above the gingival height of the final patient-matched design)
    • Minimum gingival height – 0.5 mm
    • Maximum gingival height – 6.0 mm
    • Pre-Milled Blanks are for straight abutments only

    The abutment screw for use with the Pre-Milled Blank Abutment and the Alpha Dent implants was cleared previously in K243212.

    All subject device implants are provided sterile, and all subject device abutments are provided non-sterile. The Pre-Milled Blank Abutment is supplied with the previously-cleared non-sterile abutment screw for attachment to the corresponding compatible implant. All subject device components are provided in single-unit packages for single-patient, single-use only.

    AI/ML Overview

    The provided FDA 510(k) Clearance Letter for DESS® Dental Implants (K251280) does not contain information about acceptance criteria or a specific study proving the device meets those criteria in the context of AI/ML device performance.

    The document is a clearance letter for a dental implant system, which is a physical medical device, not an AI/ML-driven diagnostic or treatment planning software. Therefore, the questions related to AI/ML device performance (such as sample size for test/training sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, and ground truth) are not applicable to the content provided.

    The document primarily focuses on demonstrating substantial equivalence to predicate devices based on:

    • Intended Use and Indications for Use
    • Technological Characteristics (design, materials, surface treatment, sterilization methods, dimensions)
    • Performance Data (non-clinical data such as MRI review, sterilization validation, bacterial endotoxin testing, shelf life, biocompatibility, and mechanical testing according to ISO 14801).

    Since the request specifically asks for AI/ML-related performance evaluation, and the provided document is not for an AI/ML device, I cannot fulfill the request as stated with the given input.

    However, I can extract the non-clinical performance data and the comparison to predicate devices, which serve as the "study" and "acceptance criteria" for this type of physical medical device in the context of a 510(k) submission.

    Non-Clinical Performance Data (Serving as "Study" for Substantial Equivalence):

    1. Worst-Case MRI Review:

      • Method: Scientific rationale and published literature (T.O. Woods, J.G. Delfino, and S. Rajan, "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices," Journal of Testing and Evaluation Volume 49, No. 2 (March/April 2021): 783–795).
      • Scope: Evaluated all compatible implant bodies, abutments, and fixation screws and their material composition.
      • Parameters: Magnetically induced displacement force and torque, per FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment."
      • Outcome: Rationale addressed parameters; device components found suitable for the MR environment.
    2. Gamma Irradiation Sterilization Validation (for implants):

      • Method: Selecting and substantiating a 25 kGy dose using method VDmax25, according to ISO 11137-1 and ISO 11137-2.
      • Acceptance Criteria: Sterility Assurance Level (SAL) of $10^{-6}$.
      • Outcome: Analysis showed the subject device implants met the SAL and did not create a new worst case compared to predicate device K212538.
    3. Bacterial Endotoxin Testing:

      • Method: Limulus amebocyte lysate (LAL) testing according to ANSI/AAMI ST72.
      • Acceptance Criteria: Limit of < 20 EU/device.
      • Outcome: Demonstrated sterile product met the limit.
    4. Sterile Barrier Shelf Life:

    5. Moist Heat Sterilization (for non-sterile abutments, validated for end-user sterilization):

      • Method: Overkill method according to ANSI/AAMI/ISO 17665-1 and ANSI/AAMI/ISO TIR 17665-2.
      • Acceptance Criteria: Sterility Assurance Level (SAL) of $10^{-6}$.
      • Outcome: Analysis showed the subject device abutments met the SAL and did not create a new worst case compared to predicate device K240208.
    6. Biocompatibility:

    7. Compatibility Information:

    8. Mechanical Testing:

      • Method: Conducted according to ISO 14801.
      • Purpose: To support the performance of the subject device implants with previously cleared NEO GM abutments with angulation (up to 30° Multi-Unit Abutments and up to 17° Cement-retained Abutments).
      • Outcome: Supports performance for indicated uses.

    Regarding AI/ML specific questions:

    1. AI/ML Acceptance Criteria & Reported Device Performance: Not applicable, as this is not an AI/ML device.
    2. Sample size for test set and data provenance: Not applicable.
    3. Number of experts and qualifications for ground truth: Not applicable.
    4. Adjudication method: Not applicable.
    5. Multi Reader Multi Case (MRMC) comparative effectiveness study: Not applicable.
    6. Standalone (algorithm only) performance: Not applicable.
    7. Type of ground truth used: Not applicable.
    8. Sample size for training set: Not applicable.
    9. How ground truth for training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.

    All digitally designed custom abutments for use with DESS Bases or Pre-milled Blanks are to be sent to a Terrats Medical validated milling center for manufacture, or to be designed and manufactured according to the digital dentistry workflow. The digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine, and associated tooling and accessories.

    Device Description

    The purpose of this submission is to expand the DESS Dental Smart Solutions abutment system cleared under K221301 and K240982 to allow additional options of zirconia material, scanners, CAM software, and milling machines to the digital dentistry workflow. The subject devices are to be sent to Terrats Medical validated milling centers for manufacture, or to be designed and manufactured via a digital dentistry workflow. The digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, titanium and ceramic material, milling machine, and associated tooling and accessories. There are no changes to the abutment design, implant compatibilities, or design parameters. All part numbers have been cleared for manufacturing via a validated milling center and digital dentistry workflows (also referred to as point of care) under K221301 and K240982.

    The subject device DESS Dental Smart Solutions abutments provide a range of prosthetic solutions for dental implant restoration. DESS abutments are offered in a variety of connection types to enable compatibility with currently marketed dental implants. All abutments are provided non-sterile, and each abutment is supplied with the appropriate abutment screw (if applicable) for attachment to the corresponding implant.

    Subject device Base Abutments are designed for fabrication of a patient-specific CAD/CAM zirconia superstructure on which a crown may be placed. They are two-piece abutments for which the second part (or top half) is the ceramic superstructure. They also may be used for support of a crown directly on the abutment.

    All patient-specific custom abutment fabrication for Base Abutments and Pre-milled (Blank) Abutments is by prescription on the order of the clinician. The subject device Pre-milled (Blank) Abutments and all zirconia superstructures for use with the subject device Ti Base Interface, DESS Aurum Base, ELLIPTIBase, and DESS C-Base will be manufactured using a validated milling center or a digital dentistry workflow. A validated milling center will be under FDA quality system regulations. The digital dentistry workflow scans files from intra-oral and lab (desktop) scanners, CAD software, CAM software, titanium and ceramic material, milling machine and associated tooling and accessories.

    The digital dentistry workflow uses scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories.

    AI/ML Overview

    The provided 510(k) summary for DESS Dental Smart Solutions focuses on demonstrating substantial equivalence to predicate devices for dental implant abutments. It primarily addresses the expansion of compatible materials, scanners, CAM software, and milling machines within an existing digital dentistry workflow. The document does not describe an AI/ML-based device that would typically have acceptance criteria related to diagnostic performance.

    Therefore, many of the requested items related to AI/ML device performance (like acceptance criteria for diagnostic metrics, sample size for test sets, data provenance, expert qualifications, adjudication methods, MRMC studies, standalone performance, and training set details) are not applicable to this submission.

    The acceptance criteria and supporting "study" (non-clinical data) for this device are related to its mechanical performance, biocompatibility, and integration within the digital workflow, demonstrating that the expanded components maintain the safety and effectiveness of the previously cleared predicate devices.

    Here's a breakdown based on the information provided and the non-applicability of AI/ML-specific questions:

    1. A table of acceptance criteria and the reported device performance

    Since this is not an AI/ML diagnostic device, the acceptance criteria are not in terms of traditional diagnostic metrics (sensitivity, specificity, AUC). Instead, they are related to material properties, mechanical integrity, and the digital workflow's accuracy.

    Acceptance Criteria CategoryReported Device Performance (Summary from Submission)
    Mechanical Integrity / Compatibility- Fatigue testing of OEM implant bodies with patient-specific abutments made at worst-case angled conditions. (Implies successful completion to similar or better standards than predicate.)
    Biocompatibility- Biocompatibility testing according to ISO 10993-5 and ISO 10993-12. (Implies successful demonstration of biocompatibility.)
    Sterilization- Sterilization validation according to ISO 17665-1, ISO 17665-2, and ISO 14937. (Implies successful validation for sterilization.)
    CAD Design Restrictions- Software verification included testing of restrictions that prevent design of components outside of the stated design parameters. - Abutment design library validated to demonstrate established design limitations are locked and cannot be modified by the user. (Implies successful implementation and verification of design constraints.)
    CAM Restriction Zones / Manufacturing Accuracy- Validation testing of CAM restriction zones conducted, including verification to show avoidance of damage or modifications of the connection geometry, and locking of restriction zones from user editing in CAM software. (Implies successful validation to ensure manufacturing precision and prevent damage.)
    Material Conformance- Zirconia materials conform to ISO 6872. - Titanium alloy conforms to ASTM F136. - Co-Cr-Mo alloy conforms to ASTM F15337. (Implies materials meet standards.)
    Physical Dimensions- Device encompasses the same range of physical dimensions as the predicate device. (Implies dimensional equivalence.)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not explicitly stated in terms of a "test set" for diagnostic performance. The validation involved physical testing of components (e.g., fatigue testing) and software verification. The specific number of abutments or digital design instances used for these non-clinical tests is not detailed in this summary.
    • Data Provenance: Not applicable in the context of patient data for an AI/ML device. The "data" here refers to engineering and material testing results, likely conducted in controlled lab environments (implied to be in accordance with international standards like ISO and ASTM). The manufacturer is Terrats Medical SL, in Spain, so testing would likely originate from their facilities or contracted labs.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This is not a diagnostic device requiring expert interpretation for ground truth. The "ground truth" for this device relates to engineering specifications and material science.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not a diagnostic device involving expert review adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For this device, the "ground truth" is based on:

    • Engineering Specifications: Defined design parameters (e.g., minimum wall thickness, post height, angulation limits).
    • Material Standards: Conformance to international standards such as ASTM F136, ISO 6872.
    • Benchmarking/Predicate Equivalence: Performance is assessed against established performance of the predicate devices and OEM implant systems.
    • Software Validation Logic: Verification that software correctly enforces design rules and CAD/CAM restrictions.

    8. The sample size for the training set

    Not applicable. This device does not involve a machine learning training set.

    9. How the ground truth for the training set was established

    Not applicable. This device does not involve a machine learning training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243212

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-02-14

    (135 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K111287, K120414, K102034, K060657, K051507, K210499

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.

    All digitally designed custom abutments for use with DESS Ti Base abutments or Pre-Milled Blank abutments are to be sent to a Terrats Medical validated milling center for manufacture.

    Device Description

    The purpose of this submission is to add components to the DESS Dental Smart Solutions system, which includes abutments, and prosthetic components cleared previously in various submissions. The previously cleared abutments and prosthetic components are compatible with a variety of original equipment manufacturer (OEM) dental implants as well as DESS Dental Smart Solutions dental implants.

    This submission adds various abutments to the OEM implant lines as summarized on the following pages in Table 1 Summary of Subject Device Abutment Designs, and Table 2 Summary of Subject Device Abutment Sizes.

    The subject device abutment designs include Healing Abutments, Temporary Abutments, Straight Abutments, Ti Base Abutments, AURUM Base Abutments with a screw channel design that allows easier instrument access to the abutment screw), C-Base Abutments, ELLIPTIBase CoCr Base Abutments, Pre-Milled Blank Abutments, Multi-Unit Abutments (straight), DESSLoc Abutments (Locator-type abutments), and abutment screws.

    AI/ML Overview

    The provided document is an FDA 510(k) clearance letter for the DESS Dental Smart Solutions, which are endosseous dental implant abutments. This document primarily focuses on demonstrating substantial equivalence to previously cleared predicate devices rather than providing detailed performance data from studies against specific acceptance criteria.

    Therefore, the document does not contain the information required to populate a table of acceptance criteria and proven device performance as typically expected for studies proving device efficacy or safety against performance metrics. It does not present specific quantitative performance data for the device.

    Based on the content, here's what can be inferred or explicitly stated regarding the study that proves the device meets acceptance criteria:

    1. Table of Acceptance Criteria and Reported Device Performance:

    As noted above, the document does not present a table of specific performance acceptance criteria (e.g., accuracy, sensitivity, specificity, or other quantitative measures) and the device's reported performance against these criteria. The "Performance Data" section primarily describes non-clinical tests (MR compatibility, dimensional analysis, sterilization validation, shelf-life testing) that demonstrate that the device meets the design specifications and regulatory requirements for substantial equivalence, rather than clinical efficacy or diagnostic accuracy.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Sample Size: Not applicable in the context of clinical performance data. The non-clinical tests would have their own sample sizes for their specific assessments (e.g., a certain number of abutments for sterilization validation, a certain number of parts for dimensional analysis). These specific numbers are not provided in this summary.
    • Data Provenance: The "reverse engineering dimensional analysis" was done for BlueSkyBio implant connections and for Strawmann bone level connections (referenced from previous submissions). For Alpha dent, a contractual agreement provided measurements. No specific country of origin is mentioned for the data itself, but Terrats Medical SL is based in Barcelona, Spain. The document states "No clinical data were included in this submission," indicating a reliance on non-clinical testing and substantial equivalence. The studies referenced are non-clinical in nature.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts:

    This information is typically relevant for studies involving human interpretation or subjective assessment (e.g., imaging studies). Since "No clinical data were included in this submission," and the evaluation relies on non-clinical testing and engineering analyses, this concept of "experts establishing ground truth" as it pertains to medical image interpretation or clinical outcomes is not applicable here. Ground truth in this context would be engineering specifications and validated test methods.

    4. Adjudication Method for the Test Set:

    Not applicable, as there is no human interpretation or subjective assessment being adjudicated in the described non-clinical studies.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    No. The document explicitly states: "No clinical data were included in this submission." Therefore, an MRMC study comparing human readers with or without AI assistance was not performed or submitted.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device is a physical dental implant abutment, not an AI algorithm or software that performs standalone analysis. The "DESS Dental Smart Solutions" in the product name refers to the brand of these physical components, not an intelligent software.

    7. The Type of Ground Truth Used:

    For the non-clinical tests described:

    • Dimensional Analysis: The "ground truth" was established by reverse engineering dimensional analysis of OEM implant bodies, OEM abutments, and OEM abutment screws, or by contractual agreement providing measurements from the OEM manufacturer. This means the OEM specifications and existing components served as the reference.
    • Sterilization: The ground truth for sterilization was meeting the sterility assurance level of 10-6 according to ANSI/AAMI/ISO 17665-1 and ANSI/AAMI/ISO TIR 17665-2 for moist heat, and ISO 11137-1 and ISO 11137-2 for gamma irradiation, and bacterial endotoxin testing according to ANSI/AAMI ST72.
    • Material properties: Conformance to ASTM F136 for Ti-6Al-4V alloy and ASTM F1537 for Co-Cr-Mo alloy, and ISO 13356 for zirconia.

    8. The Sample Size for the Training Set:

    Not applicable. This is a physical medical device, not a machine learning or AI algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as no training set was used.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240982

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-08-26

    (138 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.

    Device Description

    The purpose of this submission is to expand the DESS Dental Smart Solutions abutment system cleared under K221301 to add the ability for the subject device Base Abutments and Pre-milled (Blank) Abutments to be designed using AbutmentCAD software in the digital dentistry workflow, as well as add angulation to some of the Pre-Milled (Blank) Abutments. The subject devices are to Terrats Medical validated milling centers for manufacture, or to be designed and manufactured via a digital dentistry workflow. The digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, titanium and ceramic material, milling machine, and associated tooling and accessories. The proposed change is to allow the subject devices to be designed using AbutmentCAD by exocad GmbH, the current workflow allows only the use of 3Shape by 3Shape A/S for the design software. Another purpose of this submission is to expand the design parameters to allow angulation (up to 30°) on Pre-milled (Blank) Abutments that are compatible with Neodent Grand Morse, Nobel Active/Nobel Parallel Conical 3.0 mm, and Straumann BLX implants. There are no changes to the abutment design or implant compatibilities. All part numbers have been cleared for manufacturing via a validated milling center and digital dentistry workflows in under K221301.

    The subject device DESS Dental Solutions abutments provide a range of prosthetic solutions for dental implant restoration. DESS abutments are offered in a variety of connection types to enable compatibility with currently marketed dental implants. All abutments are provided non-sterile, and each abutment is supplied with the appropriate abutment screw (if applicable) for attachment to the corresponding implant.

    Subject device Base Abutments are designed for fabrication of a patient-specific CAD/CAM zirconia superstructure on which a crown may be placed. They are two-piece abutments for which the second part (or top half) is the ceramic superstructure. They also may be used for support of a crown directly on the abutment.

    All patient-specific custom abutment fabrication for Base Abutments and Pre-milled (Blank) Abutments is by prescription on the order of the clinician. The subject device Pre-milled (Blank) Abutments and all zirconia superstructures for use with the subject device Ti Base Interface, DESS Aurum Base, ELLIPTIBase, and DESS C-Base will be manufactured using a validated milling center or a digital dentistry workflow. A validated milling center will be under FDA quality system regulations. The digital dentistry workflow scans files from intra-oral and lab (desktop) scanners, CAD software, titanium and ceramic material, milling machine and associated tooling and accessories.

    The digital dentistry workflow uses scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories. The digital workflow includes the following products (not subject devices of this submission):

    • Ceramic material: VITA YZ ST and VITA YZ XT (K180703)
    • . Cement: Ivoclar Vivadent Multilink Hybrid Abutment Cement (K130436)
    • . Intraoral Scanner: 3Shape TRIOS A/S Series Intraoral Scanner (510(k) exempt under 21 CFR 872.3661)
    • Desktop scanner: 3Shape D900 Dental Lab Scanner (510(k) exempt under 21 CFR 872.3661)
    • Abutment design software: 3Shape Abutment Designer Software (K151455) and AbutmentCAD ● (K193352)
    • . Milling machine: VHF R5 by vhf camfacture AG with DentalCAM and DentalCNC 7 software
    AI/ML Overview

    The provided text describes a 510(k) premarket notification for DESS Dental Smart Solutions, which are dental implant abutments. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving a device's effectiveness through clinical performance studies with specific statistical acceptance criteria for accuracy, sensitivity, or specificity.

    Therefore, the document does not contain the information requested regarding:

    • A table of acceptance criteria and reported device performance (in terms of clinical metrics like accuracy, sensitivity, specificity).
    • Sample size used for the test set or its provenance.
    • Number of experts used to establish ground truth or their qualifications.
    • Adjudication method for the test set.
    • Multi-Reader Multi-Case (MRMC) comparative effectiveness study or its effect size.
    • Standalone (algorithm only) performance.
    • Type of ground truth used (expert consensus, pathology, outcomes data).
    • Sample size for the training set.
    • How the ground truth for the training set was established.

    The study described in this document focuses on non-clinical performance data to demonstrate substantial equivalence, specifically:

    1. Sterilization validation: According to ISO 17665-1, ISO 17665-2, and ISO 14937.
    2. Biocompatibility testing: According to ISO 10993-5 and ISO 10993-12.
    3. Fatigue testing and reverse engineering analysis: Of OEM implant bodies, OEM abutments, and OEM abutment screws to confirm compatibility. This includes fatigue testing of OEM implant bodies with patient-specific abutments made at worst-case angled conditions.
    4. MR Conditional labeling.
    5. Validation testing of CAM restriction zones: Including verification to show avoidance of damage or modifications of the connection geometry, and locking of restriction zones from user editing in the CAM software.
    6. Software verification: Included testing of restrictions that prevent design of components outside of the stated design parameters. The abutment design library was validated to demonstrate that established design limitations and specifications are locked and cannot be modified by the user.

    The acceptance criteria and reported "performance" for this submission are based on these engineering and design validations, ensuring the device meets safety and performance standards equivalent to the predicate device, K221301. The key "performance" metrics are about maintaining physical and material integrity and compatibility.

    The core of the submission is to expand the DESS Dental Smart Solutions abutment system to:

    • Allow design using AbutmentCAD software (in addition to 3Shape software).
    • Add angulation (up to 30°) to some Pre-milled (Blank) Abutments for specific implant systems.

    The document explicitly states: "No clinical data were included in this submission." and "The subject device, the predicate device, and reference devices have the same intended use, technological characteristics, and are materials. The subject device, the predicate device, and reference devices encompass the same range of physical dimensions, manufactured by similar methods, are packaged in similar materials, and are to be sterilized using similar methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above."

    Therefore, this FDA submission is for a physical medical device (dental implant abutment) and its manufacturing/design software modifications, not an AI or diagnostic device that would involve clinical performance metrics like sensitivity or specificity.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.

    All digitally designed custom abutments for use with DESS Ti Base abutments or Pre-Milled Blank abutments are to be sent to a Terrats Medical validated milling center for manufacture.

    Device Description

    The purpose of this submission is to add components to the DESS Dental Smart Solutions system, which includes dental implants, abutments, and prosthetic components cleared previously in various submissions. The previously cleared abutments and prosthetic components are compatible with a variety of original equipment manufacturer (OEM) dental implants as well as DESS Dental Smart Solutions dental implants.

    This submission adds various abutments to the DESS and OEM implant lines as summarized on the following pages in Table 1 Summary of Subject Device Abutment Designs, and Table 2 Summary of Subject Device Abutment Sizes.

    The subject device abutment designs include Cover Screws, Healing Abutments, Temporary Abutments, Ti Base Abutments, AURUM Base Abutments (Ti Base abutments with a screw channel design that allows easier instrument access to the abutment screw), CoCr Base Abutments, Pre-Milled Blank Abutments, Multi-Unit Abutments (straight and angled), DESSLoc Abutments (Locator-type abutments), and abutment screws.

    AI/ML Overview

    This document is a 510(k) summary for the DESS Dental Smart Solutions, a line of endosseous dental implant abutments. It details the device's characteristics, intended use, and a comparison to predicate devices to establish substantial equivalence.

    Based on the provided text, the device itself is a physical medical device (dental implant abutments and associated components), not a software or AI-driven system. Therefore, the questions related to AI acceptance criteria, training/test sets, expert adjudication, MRMC studies, and ground truth establishment for AI would not be directly applicable to this product as described.

    The document focuses on establishing substantial equivalence to previously cleared predicate devices through:

    • Identical Intended Use: The device is intended for "functional and esthetic rehabilitation of the edentulous mandible or maxilla" by providing support for prosthetic restorations, which is the same as the predicate devices.
    • Similar Technological Characteristics: The device utilizes similar designs, materials (Ti-6Al-4V, Co-Cr-Mo alloy), manufacturing processes, and sterilization methods as its predicates.
    • Performance Data: Non-clinical testing (mechanical testing per ISO 14801, MR environment assessment) and comparison to existing data from predicate devices are used to demonstrate safety and effectiveness.

    Therefore, many of the requested elements for describing AI acceptance criteria and studies are not present or applicable in this document.

    However, I can extract information relevant to the device's performance assessment and criteria for its type of submission.

    Here's an attempt to answer the questions based solely on the provided text, acknowledging that the nature of the device (a physical implant component) means many AI-specific questions will be answered as "Not Applicable" (N/A):

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state quantifiable "acceptance criteria" in a table format with corresponding "reported device performance" metrics for this specific 510(k) submission. Instead, it relies on demonstrating substantial equivalence to existing predicate devices.

    The underlying "acceptance criteria" for demonstrating substantial equivalence for this device type are primarily through:

    • Mechanical Testing (ISO 14801): This is a critical performance standard for dental implants and abutments. The document states that mechanical testing was "conducted according to ISO 14801 to support the performance." The acceptance criteria would be successful completion of these tests, demonstrating the device's mechanical strength and fatigue resistance are comparable to or better than predicate devices. The reported performance is simply that the tests supported the performance.
    • Material Conformance: Materials must conform to specific ASTM standards (e.g., ASTM F136 for Ti-6Al-4V, ASTM F1537 for Co-Cr-Mo). The reported performance is that the materials conform to these standards.
    • Biocompatibility: While not detailed in this excerpt, the mention of "biocompatibility" in relation to predicates implies conformance to relevant biocompatibility standards (e.g., ISO 10993 series). The reported performance is that it is compatible.
    • Sterilization Validation: Demonstrated sterility assurance level (SAL) of 10⁻⁶ via validated methods (moist heat or gamma irradiation). The reported performance is that validation was performed and met this SAL.
    • Dimensional Compatibility: The abutments must fit the corresponding OEM implants correctly. The reported performance is that reverse engineering dimensional analysis confirmed compatibility.

    Due to the nature of the document being a 510(k) summary focusing on substantial equivalence rather than a full study report, specific numerical performance results for the device tests are not provided in this text.

    2. Sample sizes used for the test set and the data provenance

    • Sample Size for Mechanical Testing: The document states "mechanical testing conducted according to ISO 14801." For such tests, ISO 14801 typically specifies minimum sample sizes (e.g., 10-11 samples for static strength, typically more for fatigue). The exact number of samples used for this specific submission is not explicitly stated, but it would have followed the standard's requirements.
    • Data Provenance: The mechanical testing and material analyses are assumed to be "non-clinical data submitted or referenced" by the manufacturer, Terrats Medical SL, based in Barcelona, Spain. The "reverse engineering dimensional analysis" was done by Terrats Medical SL or through contractual agreement. This is prospective testing performed to support the 510(k). The document itself does not specify the country of origin for the underlying OEM implant data used for reverse engineering, although the OEM companies are listed (e.g., Astra Tech AB, BioHorizons).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    N/A. This is a physical device. Ground truth, in the context of AI, refers to validated labels for data used to train and test an algorithm. For a physical device, performance is evaluated through engineering and biocompatibility testing against defined standards. There are no "experts" establishing ground truth in the AI sense. Testing would be performed by qualified engineers and technicians.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    N/A. Adjudication methods are typically used in clinical studies involving interpretation (e.g., by radiologists) to resolve discrepancies. This document describes non-clinical performance testing of a physical device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    N/A. This product is a dental implant abutment, not an AI software intended to assist human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    N/A. This is a physical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    N/A. For engineering tests of physical devices, the "ground truth" is typically derived from established engineering principles, international standards (e.g., ISO 14801 for mechanical properties, ASTM for materials), and the physical properties of the materials and designs themselves. There isn't "expert consensus" or "pathology" in the AI or clinical trials sense.

    8. The sample size for the training set

    N/A. This is a physical device; there's no "training set" in the machine learning sense. The device is manufactured based on established engineering designs and material specifications.

    9. How the ground truth for the training set was established

    N/A. No training set for AI. For device manufacturing, the "ground truth" for design and production parameters comes from established engineering best practices, prior successful device designs (predicate devices), and adherence to quality systems regulations (21 CFR Part 820).

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.

    Device Description

    The purpose of this submission is to expand the DESS Dental Smart Solutions abutment system by a change in sterilization status to provide products sterile to the end user that were previously cleared to be provided non-sterile. The subject device abutments and abutment screws were cleared previously to be provided non-sterile to the end user in K170588, K191986, K212628, and K22288. All subject device components will now be provided sterile.

    The subject device components include Healing Abutments, Multi-Unit Abutments (0, 17°, and 30°), and abutment screws.

    AI/ML Overview

    The provided text is a 510(k) summary for the DESS Dental Smart Solutions, an endosseous dental implant abutment. It details the device, its intended use, and its substantial equivalence to previously cleared predicate and reference devices. However, this document does not contain the acceptance criteria or a study proving the device meets those criteria in the context of an AI/ML medical device.

    The 510(k) submission for this dental abutment focuses on establishing substantial equivalence based on:

    • Design and Material: The subject device components are identical in design, material (Ti-6Al-4V alloy, DLC coating), and technological characteristics to previously cleared devices.
    • Manufacturing: The manufacturing process is consistent with previously cleared devices.
    • Biocompatibility: Referenced from previous K-clearances.
    • Sterilization: The main change in this submission is expanding the system to provide products sterile to the end-user via gamma irradiation, which was validated by referencing a previous K-clearance (K212538).
    • MR Environment Testing: Non-clinical analysis was performed to evaluate the subject devices in the MR environment, referencing published literature and FDA guidance.
    • Shelf Life Testing: Referenced from a previous K-clearance (K212538) for samples after accelerated aging.

    Therefore, I cannot fulfill your request to describe the acceptance criteria and a study proving an AI/ML device meets them based on the provided text, as this document is not about an AI/ML medical device. It's about dental implant abutments and establishes substantial equivalence through non-clinical performance data and comparison to predicate devices, not through AI/ML performance metrics.

    To provide the information you requested, I would need a document detailing the validation of an AI/ML medical device, which would include definitions of acceptance criteria (e.g., accuracy, sensitivity, specificity), details of training and test datasets, ground truth establishment, and potentially clinical effectiveness studies if applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210610

    Validate with FDA (Live)

    Date Cleared
    2021-08-27

    (179 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K120414, K163350, K183079, K111287

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The implants are intended for both one- and two-stage surgical procedures in the following situations and with the following clinical protocols:

    • Replacing missing teeth in single or multiple unit applications in the mandible or maxilla.
    • Immediate placement in extraction sites and in situations with a partially or completely healed alveolar ridge.
    • Especially indicated for use in soft bone applications with other implant surface treatments may be less effective.
    • Immediate and early loading for all indications, except in single tooth situations on implant shorter than 8 mm or in soft bone (type IV) where implant stability may be difficult to obtain and immediate loading may not be appropriate.
    Device Description

    The proposed PrimeTaper EV Dental Implants are endosseous dental implants, which are manufactured from commercially pure titanium (Grade 4). The implant surface is blasted, and acid etched. The dental implants are intended to be surgically placed in the bone of the mandible and maxilla to replace missing teeth in order to restore chewing function. The proposed PrimeTaper EV Dental Implants have a root form shape with a self-tapping thread design. The coronal portion of the implant has a minute threading (MicroThread™). The residual implant body has macro threads. The proposed implants are available in the diameters 3.6, 4.8, 5.4 mm and in the lengths 6.5, 8, 9, 11, 13, 15, 17 mm. The proposed PrimeTaper EV Dental Implants incorporate a conical implant - abutment connection interface with indexing option (EV Connection), which is identical to the internal connection of the predicate OsseoSpeed Plus (K120414) implants. The PrimeTaper EV Dental Implants are intended to be used in combination with prosthetic components with the conical EV connection (cover screws, healing abutments, cement-retained abutments and screw-retained abutments), which were previously cleared for use with the OsseoSpeed Plus Dental Implant System (K120414, K163350, K183079 and K111287). Dental implant abutments are intended to be used in the upper or lower jaw used for supporting tooth replacements to restore chewing function. No new Abutments are being proposed as part of this submission. The PrimeTaper EV Dental Implants are provided sterile by electron-beam irradiation.

    AI/ML Overview

    I am sorry, but the provided text does not contain information about acceptance criteria and a study proving a device meets these criteria in the context you specified (AI/ML device performance).

    The text is a 510(k) summary for a PrimeTaper EV Dental Implant, which is a physical medical device (an endosseous dental implant). The document details the device description, indications for use, comparison to predicate devices, and non-clinical as well as clinical performance data to demonstrate substantial equivalence to previously cleared devices.

    Therefore, I cannot extract the following information that you requested:

    • A table of acceptance criteria and the reported device performance.
    • Sample size used for the test set and data provenance.
    • Number of experts used to establish ground truth and their qualifications.
    • Adjudication method for the test set.
    • MRMC comparative effectiveness study details.
    • Standalone (algorithm only) performance.
    • Type of ground truth used.
    • Sample size for the training set.
    • How ground truth for the training set was established.

    These specific points are relevant to AI/ML-enabled devices and their evaluation, not to the traditional medical device (dental implant) described in the provided document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191890

    Validate with FDA (Live)

    Date Cleared
    2020-02-06

    (206 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K091239, K111287, K171799, K120414, K183518

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Elos Accurate® Customized Abutments are intended for attaching to dental implants in order to provide basis for single or multiple tooth prosthetic restorations. The Elos Accurate® Customized Abutment will be attached to a dental implant using the included Elos Prosthetic screw.

    The Elos Accurate® Customized Abutments are compatible with the implant systems listed in table 1:

    Table 1.

    Elos Accurate Customized Abutment – Model TypePlatform compatibilityPlatform diameter [mm]Implant Body diameter [mm]
    AB-ATO35Astra Tech 3.5/4.03.5 & 43.5 & 4
    AB-ATO45Astra Tech 4.5/5.04.5 & 54.5 & 5
    AB-ATE36Astra Tech EV 3.63.63.6
    AB-ATE42Astra Tech EV 4.24.23.6 & 4.2
    AB-ATE48Astra Tech EV 4.84.84.2 & 4.8
    AB-ATE54Astra Tech EV 5.45.45.4

    All digitally designed CAD/CAM customizations for the Elos Accurate® Customized Abutments are only intended to be sent and manufactured at a FDA registered Elos Medtech approved milling facility.

    Device Description

    The Elos Accurate® Customized Abutment is a patient specific abutment intended for attaching to dental implants in order to provide basis for single- or multiple tooth prosthetic restorations. The Elos Accurate® Customized Abutment will be attached to the implant using the included Elos Prosthetic Screw and attached to the crown/coping manually by cementation. The Elos Accurate® Customized Abutment consists of an Abutment Blank used in fabricating of a full patient-specific abutment in Titanium alloy per ASTM F136. The Abutment Blank used in creation of the Elos Accurate® Customized Abutment has a pre-manufactured connection interface that fits directly to a pre-specified dental implant. The customized shape of the abutment is intended to be manufactured by an Elos Medtech approved milling facility. The Elos Accurate® Customized Abutment is delivered nonsterile and the final restoration including corresponding Elos Prosthetic Screw is intended to be sterilized at the dental clinic before it is placed in the patient. The Elos Accurate® Customized Abutment provides clinicians and laboratories with a prosthetic device that can be used in definitive (permanent) single- or multi restorations.

    AI/ML Overview

    The provided document, a 510(k) Premarket Notification summary for the Elos Accurate® Customized Abutment, outlines the device's indications for use, technological characteristics, and non-clinical testing conducted to demonstrate substantial equivalence to a predicate device. However, it does not contain information typically found in an acceptance criteria and performance study for a medical device that uses an algorithm or AI. This document is for a physical dental abutment, not a device that relies on an algorithm or AI for its function.

    Therefore, I cannot extract the requested information regarding acceptance criteria and a study proving device meets acceptance criteria related to an AI/algorithm-based device. The original document does not offer data points for:

    • A table of acceptance criteria and reported device performance (in the context of algorithm performance)
    • Sample size for a test set or data provenance
    • Number of experts or their qualifications for ground truth
    • Adjudication method
    • MRMC comparative effectiveness study
    • Standalone algorithm performance
    • Type of ground truth (e.g., pathology, outcomes data)
    • Sample size for a training set
    • How ground truth for a training set was established

    The "testing" mentioned in the document (fatigue testing, biocompatibility testing, dimensional analysis) is for the physical material and design of the abutment, not for an algorithm's performance.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2