Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K242988

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-06-27

    (274 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended Use: Plate and screw Implants are intended for temporary fixation, correction or stabilization of bones in various anatomical regions.

    One-Quarter Tubular Plates 2.4/2.7 are indicated for fixation of fractures of the foot, hand, wrist, and forearm (including radius). The plates should only be used for load-sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.

    Straight Plates 2.4/2.7 are indicated for fixation of fractures, osteotomies, malunions and nonunions of the foot, hand, wrist, and forearm (including radius) and arthrodesis of small bones of the foot and hand.

    Reconstruction Plates 2.4/2.7 are indicated for fixation of fractures of the foot, hand, wrist, and forearm (including radius)

    Radius Distal Styloid Plate 2.4/2.7 is indicated for the fixation for fractures of the radial styloid that have metaphyseal comminution or as an adjunct to either volar of dorsal plate fixation when additional support of the radial styloid is required.

    The Radius Proximal head plates 2.4/2.7 are indicated for extra-articular and intra-articular fractures of the proximal radius and multifragmented radial neck fractures.

    The Radius Distal Volar Plates 2.4/2.7 are indicated for the fixation for fractures, fusions, or osteotomies of the distal radius.

    One-Third Tubular Plates 3.5 are indicated for fixation of fractures of the elbow (distal humerus and ulna), and fibula. The plates should only be used for load-sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.

    Reconstruction Plates 3.5 are indicated for fixation of fractures of the elbow (distal humerus and ulna), wrist, and forearm (including radius), fibula, and calcaneus

    Straight Plates 3.5 are indicated for fixation of fractures, osteotomies, malunions and nonunions of the elbow (distal humerus and ulna), wrist, forearm (including radius), and fibula.

    Tibia T-Plates 3.5 are indicated for buttressing of partial articular fractures and bone fragments of the distal tibia.

    The Clavicle Superior Plates 3.5 are indicated for:

    • Fractures of the clavicle shaft
    • Fractures of the lateral clavicle
    • Malunions of the clavicle
    • Non-unions of the clavicle

    The Clavicle Hook Plate 3.5 is indicated for fixation of lateral clavicle fractures and dislocations of the acromioclavicular joint.

    The Humerus Proximal Plates 3.5 are indicated for the fixation of fractures, fracture dislocations, non-unions and osteotomies of the proximal humerus.

    The Tibia Proximal Lateral Plates 3.5 are indicated for:

    • Split-type fractures of the lateral tibial plateau
    • Lateral split fractures with associated depressions
    • Pure central depression fractures
    • Split or depression fractures of the medial plateau

    The Tibia Distal Medial Plates 3.5 are indicated for the fixation of fractures, osteotomies, and non-unions of the distal tibia.

    The Tibia Distal Anterolateral Plates 3.5 are indicated for the fixation of fractures, osteotomies, and non-unions of the distal tibia.

    The Humerus Distal Dorsolateral Plates 2.7/3.5 are indicated for

    • Intraarticular fractures of the distal humerus
    • Supracondylar fractures of the distal humerus
    • Non-unions of the distal humerus

    The Humerus Distal Medial Plates 2.7/3.5 are indicated for

    • Intraarticular fractures of the distal humerus
    • Supracondylar fractures of the distal humerus
    • Non-unions of the distal humerus

    The Ulna Proximal Olecranon Plates 3.5 are indicated for

    • Complex extra- and intra-articular olecranon fractures
    • Non-unions of the proximal ulna
    • Osteotomies
    • Simple olecranon fractures

    Broad Curved Plates 4.5/5.0 are indicated for fixation of fractures, osteotomies, malunions and nonunions of humerus, femur, and tibia.

    Broad Plates 4.5/5.0 are indicated for fixation of fractures, osteotomies, malunions and nonunions of humerus, femur, and tibia.

    Narrow Plates 4.5/5.0 are indicated for fixation of fractures, osteotomies, malunions and nonunions of humerus, and distal tibia.

    T Plate 4.5/5.0 is intended to buttress metaphyseal fractures of the proximal humerus, medial tibial plateau and distal tibia. The plates should only be used for load-sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.

    T-Buttress Plates 4.5/5.0 are intended to buttress metaphyseal fractures of the proximal humerus, medial tibial plateau and distal tibia. The plates should only be used for load sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.

    LL-Buttress Plates 4.5/5.0 are intended to buttress metaphyseal fractures of the proximal humerus, medial tibial plateau and distal tibia. The plates should only be used for load sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.

    Femur Distal Plates 4.5/5.0 are indicated for distal shaft fractures, supracondylar fractures, intraarticular fractures, and periprosthetic fractures of the distal femur

    Tibia Proximal Lateral Plates 4.5/5.0 are indicated for proximal shaft fractures, metaphyseal fractures, intra-articular fractures and periprosthetic fractures of the proximal tibia.

    Device Description

    The Genostis Osteosynthesis System comprises of fracture fixation plates, corresponding screws and required surgical instruments. Different shapes and sizes are available to accommodate for different anatomical regions and locations of fractures. The plate screw interface is either locking, non-locking or, for some plates, realized by a combination hole offering both options (locking and non-locking).

    All plates and screws are made of titanium alloy (Ti 6Al 4V ELI) according to ASTM F136 and ISO 5832-3.

    The implants of the Genostis Osteosynthesis System are sterilized by radiation and are delivered sterile.

    AI/ML Overview

    The provided FDA 510(k) clearance letter for the Genostis Osteosynthesis System does not describe a study involving an AI/software device or human readers. Instead, it describes a traditional medical device (bone fixation plates and screws) and primarily references non-clinical (biomechanical and biocompatibility) testing to demonstrate substantial equivalence to predicate devices.

    Therefore, many of the requested criteria related to AI/software performance, human reader studies, and ground truth establishment in a diagnostic context are not applicable to this document.

    However, I can extract and structure the information that is relevant to the device's acceptance criteria based on the provided text, particularly concerning non-clinical testing.

    Here's the breakdown based on the information available:


    Acceptance Criteria and Study for Genostis Osteosynthesis System

    The Genostis Osteosynthesis System is a traditional medical device (bone fixation plates and screws), not an AI/software diagnostic device. As such, the "acceptance criteria" and "study" proving it meets them are primarily based on non-clinical engineering and biological tests, not AI performance metrics or human-in-the-loop diagnostic studies.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Test/StandardAcceptance CriteriaReported Device Performance
    Material BiocompatibilityCytotoxicity (ISO 10993-5)Satisfactory bio-compatibility endpointsPerformed; results satisfactory.
    Sensitization (ISO 10993-10)Satisfactory bio-compatibility endpointsPerformed; results satisfactory.
    Irritation/Intracutaneous Reactivity (ISO 10993-23)Satisfactory bio-compatibility endpointsPerformed; results satisfactory.
    Chemical Characterization (ISO 10993-18)Satisfactory bio-compatibility endpointsPerformed; results satisfactory.
    Toxicological Risk AssessmentSatisfactory bio-compatibility endpointsPerformed; results satisfactory.
    Bacterial Endotoxin TestingAcceptable endotoxin levelsPerformed; results satisfactory.
    Mechanical PerformanceScrew Performance (ASTM F543)Satisfactory mechanical performancePerformed; results satisfactory.
    Plate Performance (ASTM F382)Satisfactory mechanical performancePerformed; results satisfactory.
    Packaging & SterilizationSterility (Radiation)Device must be sterile at deliveryImplants are sterilized by radiation and delivered sterile. Test conducted on final sterile product.
    GeneralSubstantial Equivalence to PredicatesAll non-clinical testing demonstrates substantial equivalence.All testing indicates substantial equivalence to predicates.

    Note: The document states "All results were satisfactory" for all tests, indicating the device met the pre-defined acceptance criteria for each test.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify the exact number of units or samples used for each biomechanical or biocompatibility test. It generally states that testing was "performed" and "conducted on final sterile product."
    • Data Provenance: Not explicitly stated as "country of origin," but testing would be performed in a controlled laboratory environment. The tests are non-clinical (biomechanical and biocompatibility), not data from patients. The study is by nature prospective in that new manufactured devices were tested to meet the standards, not a retrospective analysis of existing data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Not Applicable. This is a medical device (hardware) clearance, not an AI/software diagnostic device that requires expert ground truth for interpretation of medical images or other diagnostic data. The "ground truth" here is the adherence to established engineering and biocompatibility standards.

    4. Adjudication Method for the Test Set

    • Not Applicable. As this is non-clinical mechanical and biocompatibility testing, there is no "adjudication" in the sense of reconciling expert opinions on diagnostic findings. The results are quantitative and objective measurements against established standard thresholds.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • Not Applicable. MRMC studies are for evaluating clinical interpretation of diagnostic results, typically with AI assistance. This device is a physical implant, not a diagnostic tool or AI.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Not Applicable. This pertains to AI algorithm performance. The device is a bone fixation system and does not rely on an algorithm for its primary function.

    7. The Type of Ground Truth Used

    • Pre-defined Standards and Controls: The "ground truth" for this device's performance is established by well-defined, internationally recognized (ISO) and American (ASTM) standards for material properties, mechanical strength, and biocompatibility. The device must perform within the specified parameters of these standards. For example, screws must meet certain torque resistance (ASTM F543), and plates must meet specific load-sharing characteristics (ASTM F382).

    8. The Sample Size for the Training Set

    • Not Applicable. This device does not involve a "training set" in the context of machine learning or AI.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. There is no "training set" for this type of medical device clearance. The design and manufacturing processes are established based on engineering principles and existing predicate device designs, with performance validated against the standards mentioned above.
    Ask a Question

    Ask a specific question about this device

    K Number
    K233665

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-03-21

    (127 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DePuy Synthes VOLT™ Mini Fragment Plating System:

    The VOLT™ Mini Fragment Plating System is indicated for internal fracture fixation of bones and bone fragments of the appendicular skeleton appropriate for the implant size.

    The VOLT™ Mini Fragment Plating System is intended for adults and both children (2-12 years) and adolescents (12-21 years) in which growth plates (physes) have fused or in which unfused growth plates will not be compromised by fixation.

    If used in the femur, tibia, humerus, patella, or pelvis the VOLT™ Mini Fragment Plating System can only be used for non-load bearing stabilization and reduction.

    DePuy Synthes VOLT™ Small Fragment Plating System:

    The VOLT Small Fragment Plating System is indicated for internal fracture fixation of bones and bone fragments of the appendicular skeleton appropriate for the implant size.

    The VOLT Small Fragment Plating System is intended for adults and both children (2-12 years) and adolescents (12-21 years) in which growth plates (physes) have fused or in which unfused growth plates will not be compromised by fixation.

    If used in the femur the VOLT Small Fragment Plating System can only be used for non-load bearing stabilization and reduction.

    Device Description

    DePuy Synthes VOLT™ Mini Fragment Plating System

    The DePuy Synthes VOLT™ Mini Fragment Plating System is a family of implantable devices, consisting of non-contoured, non-anatomic straight and shaped plates (Adaption, Adaption Combi, Compact Straight, Straight Condylar, Hook, T-Plate, Tine, Y-Plate and Triangle), with variable angle screw holes, and two screw types (cortex and locking screws). The system is available in three sizes; 2.0mm, 2.4mm and 2.7mm. The Triangle Plate is available in 2.4mm and 2.7mm sizes only.

    The plates of this system are available in Stainless Steel and Commercially Pure Titanium, whilst the corresponding screws are available in Stainless Steel and Titanium Alloy (TAV) respectively. Plates and screws within the VOLT™ Mini Fragment Plating System are available either sterile or non-sterile and are single-use only.

    DePuy Synthes VOLT™ Small Fragment Plating System

    The DePuy Synthes VOLT™ Small Fragment Plating System is a family of implantable devices consisting of 3.5mm non-contoured, non-anatomic plates with variable angle screw holes. and 3.5mm locking, 3.5mm cortex, and 4.0mm cancellous screws.

    The plates of this system are available in Stainless Steel and Titanium, whilst the corresponding screws are available in Stainless Steel and Titanium Alloy (TAV). Plates and screws within the VOLT™ Small Fragment Plating System are available either sterile or non-sterile and are single-use only

    AI/ML Overview

    The provided text is a 510(k) summary for the DePuy Synthes VOLT Mini Fragment Plating System and VOLT Small Fragment Plating System. This document describes the device, its indications for use, and a comparison to predicate devices, focusing on non-clinical performance testing (mechanical and engineering analyses) to demonstrate substantial equivalence.

    Crucially, this document explicitly states: "Clinical testing was not necessary for the determination of substantial equivalence."

    Therefore, the requested information regarding acceptance criteria, study details (sample size, data provenance, expert ground truth, adjudication, MRMC, standalone performance, training set details), which are typically found in clinical performance studies validating AI/ML-based medical devices, is not applicable to this submission.

    This 510(k) clearance is based on the devices being "substantially equivalent" to existing predicate devices, primarily demonstrated through non-clinical performance testing (mechanical simulations and analyses), rather than clinical trials or studies involving human data and AI algorithm performance.

    Therefore, I cannot provide the requested information as it is not present in the provided text. The device in question is a physical orthopedic implant, not an AI/ML software device that would require such performance evaluations.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191412

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-08-23

    (87 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K063049

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VariAx 2 Mini Fragment System is indicated for fracture fixation, replantation, stabilization, reduction, fusions, osteotomies, mal-unions of small bones and small bone fragments including normal and osteopenic bones in adult and adolescent (12 - 21 years) patients. The system is also indicated for non-load bearing stabilization and reduction of bone fragments in long bones.

    The VariAx 2 System is indicated for adult and pediatric patients, where the implant would not cross open growth plates, for the treatment of normal or osteopenic bone for the following conditions or procedures:

    • · Fracture fixation, including single, segmental, and comminuted fractures
    • · Revision, including nonunion and malunion
    • Intra- and extra-articular fractures
    • · Compression fracture
    • Displaced fracture
    • · Reconstruction
    • · Replantation
    • · Arthrodesis
    • Osteotomy
    Device Description

    The VariAx 2 Mini Fragment System is used for internal fixations and is composed of sterile and nonsterile plates. Additionally, the plates are provided in locking and non-locking options, as well as various lengths, thicknesses, and configurations.

    The VariAx 2 System is used for internal fixation applications and is composed of sterile and nonsterile screws, washers, and instruments. In addition to independent use, screws of this system are also used with compatible, internal fixation systems. These devices are made of commercially pure titanium and titanium alloy with Type III anodization, and are available in a variety of sizes and diameters, as well as locking and non-locking types.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the Stryker VariAx 2 System and VariAx 2 Mini Fragment System. This type of submission relies on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through clinical trials with specific acceptance criteria as might be seen for novel devices.

    Therefore, the document does not contain the detailed information requested regarding acceptance criteria and a study proving the device meets them in the way this request envisions for a software or AI-based medical device.

    However, I can extract the information provided about the non-clinical testing performed to demonstrate substantial equivalence, which serves a similar purpose in the context of a 510(k) submission.

    Here's the breakdown of what is available based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not specify quantitative acceptance criteria. Instead, it states that "The aforementioned testing demonstrated that the VariAx 2 Mini Fragment System and the VariAx 2 System are substantially equivalent to the predicate devices." This implies that the performance of the new devices in these tests was comparable to or better than the predicate devices, satisfying the FDA's requirement for substantial equivalence for mechanical properties.

    Acceptance Criteria (Implied)Reported Device Performance
    Substantial equivalence to predicate devices in:Torsion: Demonstrated substantial equivalence to predicate devices.
    - TorsionInsertion Torque: Demonstrated substantial equivalence to predicate devices.
    - Insertion TorquePull-out: Demonstrated substantial equivalence to predicate devices.
    - Pull-outFour-point bending: Demonstrated substantial equivalence to predicate devices.
    - Four-point bendingMRI Safety (Magnetically-induced displacement force, Magnetically-induced torque, RF-induced heating, and Image artifacts): Demonstrated to be MR conditional.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the sample sizes (e.g., number of screws, plates tested) for the non-clinical mechanical tests. It also does not provide information on data provenance as these are physical device tests, not data-driven studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This was a non-clinical, mechanical testing study of physical medical devices, not a study involving expert interpretation of data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This was a non-clinical, mechanical testing study.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This was a non-clinical, mechanical testing study, not a human reader study involving AI.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This was a non-clinical, mechanical testing study.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the mechanical tests, the "ground truth" or reference for comparison was the performance of the identified predicate devices (Smith & Nephew Variable-Angle Locking Mini-Fragment Plating System (K132886) and Synthes (USA) Modular Mini Fragment LCP System (K063049) for the Mini Fragment System, and VariAx 2 System (K180500) for the VariAx 2 System). The tests measured physical properties and compared them against known standards or the predicate's performance. For MRI safety, the ground truth would be established by industry standards or FDA guidance for MR compatibility.

    8. The sample size for the training set

    Not applicable. This was a non-clinical, mechanical testing study, not an AI or machine learning study requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable. This was a non-clinical, mechanical testing study.


    Summary of the Study:

    The study referenced in the 510(k) submission is a non-clinical mechanical testing study. Its purpose was to demonstrate substantial equivalence of the VariAx 2 Mini Fragment System and the VariAx 2 System to their respective predicate devices.

    Tests Performed:

    • Torsion
    • Insertion Torque
    • Pull-out
    • Four-point bending
    • MRI assessments (magnetically-induced displacement force, magnetically-induced torque, RF-induced heating, and image artifacts)

    Conclusion of the Study:

    The non-clinical testing "demonstrated that the VariAx 2 Mini Fragment System and the VariAx 2 System are substantially equivalent to the predicate devices." Additionally, the devices were found to be MR conditional. Clinical testing was not required for this submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K163361

    Validate with FDA (Live)

    Date Cleared
    2017-04-07

    (128 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ANTHEM™ Fracture System is indicated for fixation of fractures, arthrodesis and reconstruction of bones for the appropriate size of the device to be used in adult patients, including the clavicle, scapula, humerus, radius, ulpa, small bones (metacarpals, metatarsals, phalanges), wrist, pelvis, fibula, ankle, and foot. Small fragment and distal fibula plates may be used in all pediatric subgroups (except neonates) and small stature adults. Distal radius plates may be used in adolescents (12-21 years of age).

    Device Description

    The ANTHEM™ Fracture System is a family of plates and screws designed to be used for internal bone fixation. The implants are available in various sizes and shapes to accommodate patient anatomy, and may be contoured or straight, sterile and non-sterile, with locking and non-locking screws. ANTHEM™ implants are manufactured from medical grade titanium alloy, cobalt chromium molybdenum alloy, or stainless steel. All implants are for single use only.

    AI/ML Overview

    This document is a 510(k) Summary for the ANTHEM™ Fracture System, a medical device. Based on the provided text, there is no study described that proves the device meets specific acceptance criteria based on AI/algorithm performance, nor is there any mention of an AI device. The document explicitly describes the ANTHEM™ Fracture System as a system of "plates and screws designed to be used for internal bone fixation."

    Therefore, I cannot provide the requested information regarding:

    • A table of acceptance criteria and reported device performance related to AI/algorithmic accuracy.
    • Sample size and data provenance for an AI test set.
    • Number of experts and qualifications for AI ground truthing.
    • Adjudication method for an AI test set.
    • MRMC comparative effectiveness study for AI assistance.
    • Standalone AI performance.
    • Type of ground truth used for AI.
    • Sample size for training set for AI.
    • How ground truth for a training set was established for AI.

    The document focuses on the mechanical and material performance of bone fixation devices, not on diagnostic or AI-driven performance.

    Here's a summary of the performance data that is mentioned in the document, which pertains to the physical device itself:

    1. Acceptance Criteria and Reported Device Performance (for the physical device):

    • Acceptance Criteria (Implicit Standard Performance): The device's performance was evaluated in accordance with recognized ASTM standards for bone fixation devices. This implies that the device had to meet performance benchmarks defined by these standards to demonstrate substantial equivalence to predicate devices. While specific numerical acceptance criteria are not explicitly stated (e.g., "minimum bending strength of X N-m"), the adherence to these standards serves as the implicit acceptance criteria for mechanical performance.
    • Reported Device Performance:
      • Tests Conducted: Engineering analysis, bending strength tests (for plates and screws), pullout strength tests (for screws), and insertion/removal torque tests (for screws).
      • Outcome: "Performance data demonstrates substantial equivalence to the predicate devices."
      • Bacterial Endotoxin Testing (BET): Conducted in accordance with ANSI/AAMI ST-72:2011 to ensure sterility or low endotoxin levels.

    2. Sample Size and Data Provenance (for the physical device testing):

    • The document does not specify the sample sizes used for the mechanical performance tests (e.g., how many plates or screws were tested for bending strength).
    • The data provenance is not mentioned (e.g., where the tests were conducted, whether the materials were from specific batches, etc.).

    3. Number of experts used to establish the ground truth... and qualifications of those experts:

    • Not applicable. This document is about a physical medical device (bone plates and screws), not an AI algorithm requiring expert review for ground truth.

    4. Adjudication method for the test set:

    • Not applicable. This document is about a physical medical device (bone plates and screws), not an AI algorithm.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done...:

    • Not applicable. This document is about a physical medical device (bone plates and screws), not an AI algorithm.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This document describes a physical medical device, not an algorithm.

    7. The type of ground truth used:

    • For the physical device, "ground truth" would be the objective measurements obtained from standardized mechanical tests (e.g., actual force at failure, actual torque values) against the requirements of the ASTM standards.

    8. The sample size for the training set:

    • Not applicable, as this is not an AI/machine learning device.

    9. How the ground truth for the training set was established:

    • Not applicable, as this is not an AI/machine learning device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K163046

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-02-09

    (100 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K063049

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DePuy Synthes 2.4/2.7 mm VA-LCP Two-Column Volar Distal Radius Plate, Extra-Long is intended for intra- and extra-articular fractures, osteotomies, nonunions of the distal radius, with or without extension into the radial diaphysis in adults and adolescents (12-21) where the growth plates have fused or will not be crossed, and the following adolescent distal radius fractures:

    • Intra-articular fractures exiting the epiphysis
    • Intra-articular fractures exiting the metaphysis
    • Physeal crush injuries
    • Any injuries which cause growth arrest to the distal radius.
    Device Description

    The DePuy Synthes 2.4/2.7mm VA-LCP Two-Column Volar Distal Radius Plate, Extra-Long will be available in stainless steel and titanium. The head of the plate will remain the same as the existing 2.4mm VA-LCP Two-Column Volar Distal Radius Plate (K102694) and will therefore allow the use of the existing guide block for these plates. The DePuy Synthes 2.4/2.7mm VA-LCP Two-Column Volar Distal Radius Plate, Extra-Long will come in 3 different lengths (7 shaft holes, 10 shaft holes and 13 shaft holes). To accommodate different patient anatomy, the plate with 7 shaft holes will be available with three different plate head widths (Standard, Narrow and Wide). The 10 shaft hole and the 13 shaft hole plate will only be released with the Standard plate head width and additionally they feature a curvature in the shaft. All plates will feature left- and right-specific designs and will be offered in sterile packed only.
    The DePuy Synthes 2.4/2.7mm VA-LCP Two-Column Volar Distal Radius Plates, Extra-Long are designed to accept existing 2.4mm Variable Angle Locking Screws, 2.4mm Cortex Screws, 2.4mm Locking Screws, 1.8mm VA Buttress Pins and 1.8mm Locking Buttress Pins in the plate head holes. In the plate shaft, existing 2.4mm Cortex Screws, 2.7mm Cortex Screws, 2.4mm Variable Angle Locking Screws, 2.7mm Variable Angle Locking Screws, 2.4mm Locking Screws, and 2.7mm Locking Screws can be inserted.

    AI/ML Overview

    This 510(k) premarket notification is for a medical device called the "DePuy Synthes 2.4/2.7mm VA-LCP Two-Column Volar Distal Radius Plate, Extra-Long". It is a metallic bone fixation appliance and accessories. This document does not describe the acceptance criteria or studies for an AI/ML device. It focuses on the substantial equivalence of a physical medical implant to existing predicate devices through mechanical and biocompatibility testing. Therefore, I cannot provide the requested information.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121601

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-07-06

    (35 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Predicate For
    Why did this record match?
    Reference Devices :

    K100776, K063049, K111230, K120070, K110592, K043185

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Synthes 2.7/3.5mm Variable Angle LCP Ankle Trauma System is intended for fixation of the ankle in adults and adolescents (12-21) in which the growth plates have fused, and particularly in osteopenic bone. Specifically,

    • Anterolateral Distal Tibia Plates are intended for fixation of osteotomies, fractures, nonunions, malunions, and replantations of bones and bone fragments of the diaphyseal and metaphyseal regions of the distal tibia.
    • Medial and Anteromedial Distal Tibia Plates are intended for fixation of osteotomies, fractures, nonunions, malunions, and replantations of bones and bone fragments of the diaphyseal and metaphyseal regions of the distal tibia,
    • Distal Tibia T Plates and Distal Tibia L Plates are intended to buttress partial articular fractures and bone fragments of the distal tibia, and
    • Lateral Distal Fibula Plates are intended for fixation of osteotomies, fractures, nonunions, malunions, and replantations of bones and bone fragments of the diaphyseal and metaphyseal regions of the distal fibula.
    Device Description

    The Synthes Anterolateral Distal Tibia Plates are intended to treat fractures of the distal tibia. These plates can be used with a new screw configuration, 3.5mm VA Locking Screws. The Anterolateral Distal Tibia Plates and 3.5mm VA Locking Screws will be offered in both stainless steel and titanium alloy (TAN), and in both sterile and non-sterile configurations. The Anterolateral Distal Tibia Plates will be offered in left and right configurations. The system accepts existing cortical screws, locking screws, dynamic locking screws, and metaphyseal screws (K100776, K063049, K112583, K000684, K111230, K120070, K110592, and K043185) as well as new 3.5mm VA Locking Screws, and allows for both dynamic compression and locking through Combi holes.

    AI/ML Overview

    This is a 510(k) premarket notification for a medical device, specifically the Synthes 2.7/3.5mm Variable Angle LCP Ankle Trauma System - Anterolateral Distal Tibia Plates.

    This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than providing clinical safety and efficacy data. Therefore, the information you're requesting regarding acceptance criteria, study details, expert review, and ground truth establishment, which are typical for studies validating AI/ML diagnostic devices, are not present in this document. This submission focuses on engineering analysis and comparison to existing, already approved devices.

    Here's why your specific questions cannot be answered from this document:

    • This is not an AI/ML device. The device is a physical bone fixation system (plates and screws). Therefore, concepts like "device performance" in terms of classification metrics (sensitivity, specificity), "test sets," "training sets," "ground truth" (in the context of expert consensus or pathology for diagnostic accuracy), or "MRMC studies" are not applicable.
    • 510(k) is about substantial equivalence. The primary way this device met its acceptance criteria was by demonstrating that it is "at least as strong as" and has similar indications, design, materials, and performance characteristics as previously cleared predicate devices.

    However, I can extract information related to the device and its "proof" of meeting requirements based on the document provided:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Demonstration of Substantial Equivalence)Reported Device Performance / Evidence
    Similar Indications for Use: Intended for fixation of the ankle in adults and adolescents (12-21) where growth plates have fused, particularly in osteopenic bone, for various distal tibia and fibula fractures.The subject device (Synthes 2.7/3.5mm VA-LCP Ankle Trauma System) has similar indications to the predicate Synthes 2.7/3.5 VA-LCP Ankle Trauma System (K120854), Synthes 4.5/3.5 LCP Metaphyseal Plate (K033805), and Zimmer Periarticular and ECT Plate Systems (various K numbers).
    Similar Design Characteristics: Refers to the physical shape, features, and configurations of the plates and screws.Both the subject and predicate devices have similar design characteristics. The Anterolateral Distal Tibia Plates are offered in left and right configurations, and the system accepts existing cortical, locking, dynamic locking, and metaphyseal screws, as well as new 3.5mm VA Locking Screws, allowing for dynamic compression and locking through Combi holes.
    Similar Materials: Stainless steel and titanium alloy (TAN) for the plates and screws.Both the subject and predicate devices use similar materials (stainless steel and titanium alloy (TAN)).
    Similar Performance Characteristics (Mechanical Strength): The device (plates and screws) must be at least as strong as predicate devices to ensure mechanical integrity for bone fixation.The subject system (plates) has been shown to be at least as strong as the predicate devices through engineering analysis and fatigue strength testing. The new 3.5mm VA Locking Screws have been shown to be substantially equivalent to existing screws through analysis and dimensional comparison to predicate screws (3.5mm Cortex Screws K112583 and 3.5mm Locking Screws K000684).

    Regarding the other points, as explained, they are not applicable or not provided in this 510(k) submission for a physical implantable device:

    1. Sample size used for the test set and the data provenance: Not applicable. Performance was demonstrated through engineering analysis and fatigue testing, not a "test set" of clinical data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No "ground truth" was established by experts for a diagnostic assessment.
    3. Adjudication method for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted device.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This is not an algorithm.
    6. The type of ground truth used: Not applicable. The "ground truth" equivalent here would be established engineering standards for strength and fatigue, against which the device was tested.
    7. The sample size for the training set: Not applicable. No "training set" for an algorithm.
    8. How the ground truth for the training set was established: Not applicable.

    In summary, this 510(k) relied on demonstrating substantial equivalence through a comparison of indications, design, materials, and engineering performance (mechanical strength) to already approved predicate devices. It did not involve clinical studies or expert evaluations of diagnostic accuracy as would be the case for AI/ML diagnostic software.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1