Search Results
Found 190 results
510(k) Data Aggregation
(51 days)
NFV | Oxazepam (BZO) | 862.3170, Benzodiazepine Test System | Toxicology |
| NFY | Cocaine (COC) | 862.3250
The CLUNGENE Multi-Drug Test Easy Cup is a lateral flow immunoassay for the qualitative detection of Morphine, Methamphetamine, Cocaine, Marijuana, Methylenedioxymethamphetamine, Buprenorphine, Propoxyphene, Amphetamine, Phencyclidine, EDDP (Methadone metabolite), Oxycodone, Oxazepam, Nortriptyline, Secobarbital, Methadone, 6-Monoacetylmorphine and Fentanyl in human urine at the following cut off concentrations:
Drug | Calibrator | Cut-off (ng/mL) |
---|---|---|
Morphine (MOP/OPI300) | Morphine | 300 |
Morphine (MOP/OPI2000) | Morphine | 2,000 |
Methamphetamine (mAMP/MET1000) | D-Methamphetamine | 1,000 |
Methamphetamine (mAMP/MET500) | D-Methamphetamine | 500 |
Cocaine (COC300) | Benzoylecgonine | 300 |
Cocaine (COC150) | Benzoylecgonine | 150 |
Marijuana (THC) | 11-nor-9-THC-9-COOH | 50 |
Methylenedioxymethamphetamine (MDMA) | D,L-Methylenedioxymethamphetamine | 500 |
Buprenorphine (BUP) | Buprenorphine | 10 |
Propoxyphene (PPX) | D-Propoxyphene | 300 |
Amphetamine (AMP1000) | D-Amphetamine | 1,000 |
Amphetamine (AMP500) | D-Amphetamine | 500 |
Phencyclidine (PCP) | Phencyclidine | 25 |
Methadone metabolite (EDDP) | 2-Ethylidine-1,5-dimethyl-3,3-diphenylpyrrolidine | 300 |
Oxycodone (OXY) | Oxycodone | 100 |
Oxazepam (BZO) | Oxazepam | 300 |
Nortriptyline (TCA) | Nortriptyline | 1,000 |
Secobarbital (BAR) | Secobarbital | 300 |
Methadone (MTD) | Methadone | 300 |
6-Monoacetylmorphine (6-MAM) | 6-Monoacetylmorphine | 10 |
Fentanyl (FYL) | Fentanyl | 1 |
The single or multi-test cups can consist of any combination of the above listed drug analytes, but only one cut off concentration under same drug condition will be included per device.
This test provides only preliminary result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. Evaluate preliminary positive results carefully. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
The CLUNGENE Multi-Drug Home Test Easy Cup is a lateral flow immunoassay for the qualitative detection of Morphine, Methamphetamine, Cocaine, Marijuana, Methylenedioxymethamphetamine, Buprenorphine, Propoxyphene, Amphetamine, Phencyclidine, EDDP (Methadone metabolite), Oxycodone, Oxazepam, Nortriptyline, Secobarbital, Methadone, 6-Monoacetylmorphine and Fentanyl in human urine at the following cut off concentrations:
Drug (Identifier) | Cut-off (ng/mL) |
---|---|
Morphine (MOP/OPI2000) | 300 or 2000 |
Methamphetamine (mAMP/MET) | 500 or 1,000 |
Cocaine (COC) | 150 or 300 |
Marijuana (THC) | 50 |
Methylenedioxymethamphetamine (MDMA) | 500 |
Buprenorphine (BUP) | 10 |
Propoxyphene (PPX) | 300 |
Amphetamine (AMP) | 500 or 1,000 |
Phencyclidine (PCP) | 25 |
Methadone metabolite (EDDP) | 300 |
Oxycodone (OXY) | 100 |
Oxazepam (BZO) | 300 |
Nortriptyline (TCA) | 1,000 |
Secobarbital (BAR) | 300 |
Methadone (MTD) | 300 |
6-Monoacetylmorphine (6-MAM) | 10 |
Fentanyl (FYL) | 1 |
The single or multi-test cup offers any combination from above 1 to 17 drugs, but only one cut off concentration under same drug condition will be included per device. |
The test provides only preliminary test results. To obtain a confirmed analytical result, a more specific alternative chemical method must be used. GC/MS or LC-MS/MS is the preferred confirmatory method.
It is intended for over-the-counter (OTC) use. For in vitro diagnostic use only.
CLUNGENE Multi-Drug Test Easy Cup and CLUNGENE Multi-Drug Home Test Easy Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine.
The device is a cup format. Each test device is sealed with two sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format and no longer requires assembly before use.
This document provides details on the performance characteristics of the CLUNGENE Multi-Drug Test Easy Cup and CLUNGENE Multi-Drug Home Test Easy Cup. Since this is an in vitro diagnostic device (specifically, a drug screening test), the acceptance criteria and study design are typically focused on analytical performance (accuracy, precision, analytical specificity) rather than a multi-reader multi-case (MRMC) comparative effectiveness study, which is more common for imaging AI. Similarly, "human readers improving with AI vs without AI" is not applicable here as the device is the test, not an aid to human interpretation of another modality.
Here's the breakdown based on the provided text:
Acceptance Criteria and Reported Device Performance
The core acceptance criterion for qualitative drug screening tests like this is accurate detection around a specific cutoff concentration. The reported performance demonstrates the device's ability to correctly classify samples as positive or negative relative to these cutoffs.
Table of Acceptance Criteria and Reported Device Performance (Analytical Precision/Reproducibility)
The "Acceptance Criteria" column represents the desired performance for a qualitative assay around its cutoff. For positive results, this means detecting drug concentrations above the cutoff, and for negative results, it means not detecting concentrations below the cutoff. The provided precision data shows the number of positive (+) and negative (-) results out of 50 tests for various concentrations relative to the cutoff. An ideal performance would show 100% positive for concentrations above cutoff and 100% negative for concentrations below, with roughly 50/50 split at the cutoff itself (due to inherent variability).
Drug (Cut-off ng/mL) | Acceptance Criteria (Implicit for qualitative assay) | Reported Device Performance (Accuracy as evidenced by reproducibility at various concentrations) Number of negative/positive results out of 50 tests. Values are aggregated across 3 lots where available. |
---|---|---|
MOP300 | All samples >cutoff should test positive; all samples |
Ask a specific question about this device
(25 days)
II | 21 CFR § 862.3870, Cannabinoids Test System | Toxicology (91) |
| NFY Cocaine | II | 21 CFR § 862.3250
The AssureTech Quick Cup Tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Secobarbital, Buprenorphine, Oxazepam, Cocaine, EDDP, Fentanyl, Ecstasy, Propoxyphene, Morphine, Methadone, Phencyclidine, Oxycodone, Norfentanyl, Methamphetamine, Nortriptyline, 6-Monoacetylmorphine, Tramadol and Marijuana in human urine at the cutoff concentrations of:
Drug (Identifier) | Cut-off level |
---|---|
Amphetamine (AMP) | 500 ng/mL or 300 ng/mL |
Secobarbital (BAR) | 300 ng/mL |
Buprenorphine (BUP) | 10 ng/mL |
Oxazepam (BZO) | 300 ng/mL |
Cocaine (COC) | 150 ng/mL |
EDDP | 300ng/ml |
Fentanyl (FYL) | 1 ng/mL |
Ecstasy (MDMA) | 500 ng/mL |
Propoxyphene (PPX) | 300 ng/mL |
Morphine (MOR) | 2000 ng/mL or 300 ng/mL |
Methadone (MTD) | 300 ng/mL |
Phencyclidine (PCP) | 25 ng/mL |
Oxycodone (OXY) | 100 ng/mL |
Norfentanyl (NFYL) | 5 ng/mL |
Methamphetamine (MET) | 500 ng/mL or 300 ng/mL |
Nortriptyline (TCA) | 1000 ng/mL |
6-Monoacetylmorphine (6-MAM) | 10 ng/mL |
Tramadol (TML) | 100 ng/mL |
Marijuana (THC) | 50 ng/mL or 20 ng/mL |
Configuration of the AssureTech Quick Cup Tests can consist of any combination of the above listed drug analytes. The test may yield positive results for the prescription drugs above when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.
For in vitro diagnostic use only.
The AssureTech Multi-drug Urine Test Cup are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Secobarbital, Buprenorphine, Oxazepam, Cocaine, EDDP, Fentanyl, Ecstasy, Propoxyphene, Morphine, Methadone, Phencyclidine, Oxycodone, Norfentanyl, Methamphetamine, Nortriptyline, 6-Monoacetylmorphine, Tramadol and Marijuana in human urine at the cutoff concentrations of:
Drug (Identifier) | Calibrator | Cut-off level |
---|---|---|
Amphetamine (AMP) | d-Amphetamine | 500 ng/mL or 300 ng/mL |
Secobarbital (BAR) | Secobarbital | 300 ng/mL |
Buprenorphine (BUP) | BUP-3-D-Glucuronide | 10 ng/mL |
Oxazepam (BZO) | Oxazepam | 300 ng/mL |
Cocaine (COC) | Benzoylecgonine | 150 ng/mL |
EDDP | 2-Ethylidine-1,5-dimethyl-3,3-diphenylpyrrolidine | 300ng/ml |
Fentanyl (FYL) | Fentanyl | 1 ng/mL |
Ecstasy (MDMA) | 3,4-Methylenediioxy-MET | 500 ng/mL |
Propoxyphene (PPX) | D-Propoxyphene | 300 ng/mL |
Morphine (MOR) | Morphine | 2000 ng/mL or 300 ng/mL |
Methadone (MTD) | Methadone | 300 ng/mL |
Phencyclidine (PCP) | Phencyclidine | 25 ng/mL |
Oxycodone (OXY) | Oxycodone | 100 ng/mL |
Norfentanyl (NFYL) | Norfentanyl | 5 ng/mL |
Methamphetamine (MET) | Methamphetamine | 500 ng/mL or 300 ng/mL |
Nortriptyline (TCA) | Nortriptyline | 1000 ng/mL |
6-Monoacetylmorphine (6-MAM) | 6-Monoacetylmorphine | 10 ng/mL |
Tramadol (TML) | Cis-Tramadol | 100 ng/mL |
Marijuana (THC) | 11-nor-9-THC-9-COOH | 50 ng/mL or 20 ng/mL |
Configuration of the AssureTech Multi-drug Urine Test Cup can consist of any combination of the above listed drug analytes. It is for in vitro diagnostic use only.
The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
The AssureTech Quick Cup Tests are immunochromatographic assays that use a lateral flow system for the qualitative detection of Amphetamine, Secobarbital, Buprenorphine, Oxazepam, Cocaine, EDDP, Fentanyl, Ecstasy, Propoxyphene, Morphine, Methadone, Phencyclidine, Oxycodone, Norfentanyl, Methamphetamine, Nortriptyline, 6-Monoacetylmorphine, Tramadol and Marijuana (target analytes) in human urine. The products are single-use in vitro diagnostic devices. Each test kit contains a Test Device and a package insert. Each test device is sealed with a desiccant in an aluminum pouch.
The provided FDA 510(k) Clearance Letter details the performance of the AssureTech Quick Cup Tests and AssureTech Multi-drug Urine Test Cup for qualitative and simultaneous detection of various drugs in human urine.
Here's an analysis of the acceptance criteria and the study proving the device meets those criteria:
1. Acceptance Criteria and Reported Device Performance
For in vitro diagnostic devices like these, acceptance criteria are typically related to the accuracy of the qualitative detection (positive vs. negative) compared to a gold standard, particularly around the established cutoff concentrations. The performance is assessed through analytical studies (precision, specificity, interference) and comparison studies with a confirmatory method.
Here's a table summarizing the implicit acceptance criteria based on the precision and lay-user studies, and the reported device performance. The acceptance criterion is inferred as the ideal performance for these types of tests, where results near or above the cutoff should be positive, and results significantly below should be negative. The performance data below is extracted from the "Precision" and "Lay-user study" sections.
Table of Acceptance Criteria and Reported Device Performance
Drug (Identifier) | Cut-off Level | Implicit Acceptance Criterion (Qualitative) | Reported Performance (Precision Study - Total of 3 Lots, 50 observations per concentration) | Reported Performance (Lay-user study - Agreement %) |
---|---|---|---|---|
AMP300 | 300 ng/mL | -100% to -25% Cutoff: All Negative | -100%/-75%/-50%/-25% Cutoff: 100% Negative (50-/0+) | -100% to -25% Cutoff: 95%-100% agreement |
+25% to +100% Cutoff: All Positive | +25%/+50%/%75%/+100% Cutoff: 100% Positive (50+/0-) | +25% to +75% Cutoff: 95%-100% agreement | ||
Cutoff: Majority Positive, some Negative acceptable (reflecting assay variability) | Cutoff: 64-68% Positive (30-34+/16-20-) | Not explicitly reported for cutoff in lay-user study | ||
MET300 | 300 ng/mL | -100% to -25% Cutoff: All Negative | -100%/-75%/-50%/-25% Cutoff: 100% Negative (50-/0+) | -100% to -25% Cutoff: 100% agreement |
+25% to +100% Cutoff: All Positive | +25%/+50%/%75%/+100% Cutoff: 100% Positive (50+/0-) | +25% to +75% Cutoff: 95%-100% agreement | ||
Cutoff: Majority Positive, some Negative acceptable (reflecting assay variability) | Cutoff: 66-72% Positive (33-36+/14-17-) | Not explicitly reported for cutoff in lay-user study | ||
TML100 | 100 ng/mL | -100% to -25% Cutoff: All Negative | -100%/-75%/-50%/-25% Cutoff: 100% Negative (50-/0+) | -100% to -25% Cutoff: 100% agreement |
+25% to +100% Cutoff: All Positive | +25%/+50%/%75%/+100% Cutoff: 100% Positive (50+/0-) | +25% to +75% Cutoff: 95% agreement | ||
Cutoff: Majority Positive, some Negative acceptable (reflecting assay variability) | Cutoff: 64-72% Positive (32-36+/14-18-) | Not explicitly reported for cutoff in lay-user study | ||
All other listed drugs (Configuration 1 & 2) | Various | -100% to -25% Cutoff: All Negative | Data for these drugs were reported in previous 510(k)s (K243996, K201630, K181768, K180349, K170049, K161044, K153465, K152025, K151211). The precision study for AMP300, MET300, TML100 suggests similar performance. | -100% to -25% Cutoff: 90%-100% agreement |
+25% to +100% Cutoff: All Positive | +25% to +75% Cutoff: 90%-100% agreement |
Note: The precision study for AMP300, MET300, and TML100 used 3 lots, with "50-/0+" meaning 50 negative results and 0 positive results, and "50+/0-" meaning 50 positive results and 0 negative results. For the 'Cutoff' concentration, it shows a mix of positive and negative results, which is expected due to the nature of qualitative assays around the threshold.
2. Sample Sizes and Data Provenance
-
Precision Study:
- For AMP300, MET300, and TML100: Each drug had 8 concentrations (spanning -100% to +100% of cutoff, plus the cutoff). For each concentration, tests were performed two runs per day for 25 days, for 3 different lots.
- This means 50 observations per concentration per lot (2 runs/day * 25 days/run).
- Total observations per drug for all 3 lots: 8 concentrations * 50 observations/concentration * 3 lots = 1200 observations per drug.
- Data for other drugs refer to previous 510(k) clearances (K243996, K201630, K181768, K180349, K170049, K161044, K153465, K152025, K151211).
- Data Provenance: Retrospective, as samples were "prepared by spiking drug in negative samples" and confirmed by LC/MS. No specific country of origin is mentioned, but typically for FDA submissions, studies are conducted under GLP (Good Laboratory Practice) guidelines, often in the US or by international labs adhering to comparable standards.
- For AMP300, MET300, and TML100: Each drug had 8 concentrations (spanning -100% to +100% of cutoff, plus the cutoff). For each concentration, tests were performed two runs per day for 25 days, for 3 different lots.
-
Comparison Studies (Clinical Samples):
- For AMP300, MET300, and TML100: The tables show data broken down by "Negative" (presumably drug-free), "Low Negative" ( +50%). The sum of the positive and negative results across these categories for each operator represents the number of clinical samples tested for that drug.
- AMP300: 5 (Negative) + 15 (LN) + 19 (NCN) + 24 (NCP) + 16 (HP) = 79 samples per operator. (Operator 1)
- MET300: 4 (Negative) + 13 (LN) + 23 (NCN) + 20 (NCP) + 20 (HP) = 80 samples per operator. (Operator 1)
- TML100: 2 (Negative) + 18 (LN) + 18 (NCN) + 19 (NCP) + 20 (HP) = 77 samples per operator. (Operator 1)
- Total (approximate, as numbers vary slightly between operators): ~79+80+77 = ~236 clinical samples for AMP300, MET300, TML100 combined.
- Data for other drugs refer to previous 510(k) clearances.
- Data Provenance: Retrospective, using "unaltered clinical samples." No specific country of origin is mentioned.
- For AMP300, MET300, and TML100: The tables show data broken down by "Negative" (presumably drug-free), "Low Negative" ( +50%). The sum of the positive and negative results across these categories for each operator represents the number of clinical samples tested for that drug.
-
Lay-User Study:
- Sample Size: 280 lay persons tested the device.
- Configuration 1: 66 male + 74 female = 140 lay persons.
- Configuration 2: 87 male + 53 female = 140 lay persons.
- Data Provenance: Retrospective, samples were "prepared at the following concentrations; negative, +/-75%, +/-50%, +/-25% of the cutoff by spiking drugs into drug free-pooled urine specimens." Confirmed by LC/MS. Conducted "at three intended user sites." No specific country of origin is mentioned.
- Sample Size: 280 lay persons tested the device.
3. Number of Experts and Qualifications for Ground Truth (Clinical Samples)
- Ground Truth Establishment for Clinical Samples: LC/MS (Liquid Chromatography/Mass Spectrometry) is stated as the preferred confirmatory method and was used to confirm the concentrations of the samples. This is an objective chemical method, considered the gold standard for drug detection and quantification in urine.
- Experts: The comparison studies were performed "in-house with three laboratory assistants." While these individuals are performing the rapid tests, the ultimate ground truth is established by the LC/MS results. The "laboratory assistants" are not explicitly designated as "experts" in establishing ground truth, but rather as trained users of the device whose results are compared to the LC/MS gold standard.
4. Adjudication Method for the Test Set (Clinical Samples)
- The document states that "Operators ran unaltered clinical samples for each drug. The samples were blind labeled and compared to LC/MS results."
- There were three operators. The "Discordant Results" tables show discrepancies between the rapid test results and the LC/MS results, sometimes across multiple operators for the same sample.
- No explicit adjudication method (e.g., 2+1, 3+1) for the rapid test results themselves is described. The comparison seems to be a direct comparison of each operator's rapid test result against the LC/MS ground truth, and then discrepancies are noted. The LC/MS data serves as the final, objective ground truth.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- This document describes performance characteristics of an in-vitro diagnostic device (a qualitative urine drug test cup).
- No MRMC comparative effectiveness study was performed in the context of comparing human readers (e.g., radiologists interpreting images) with and without AI assistance. This type of study design is specific to AI/CADe (Computer-Assisted Detection) or CADx (Computer-Assisted Diagnosis) devices in imaging, which is not applicable to a lateral flow immunoassay like the AssureTech Quick Cup Tests.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
- Given this is a physical immunoassay test cup, the concept of a "standalone (algorithm only without human-in-the-loop performance)" study does not directly apply in the same way it would for a software-based AI device.
- The "Comparison Studies" with laboratory assistants and the "Lay-user study" assess the device's performance when interpreted by human users. The device itself, by producing a visual result (line/no line), is the "algorithm." Its performance is inherently tied to human interpretation of that visual output. The precision and specificity studies represent the analytical performance of the device itself.
7. Type of Ground Truth Used
- Analytical Performance Studies (Precision, Specificity, pH/SG Effect): The ground truth was established by spiking known concentrations of drugs into negative urine samples, with concentrations confirmed by LC/MS.
- Comparison Studies (Clinical Samples): The ground truth was established by LC/MS results on unaltered clinical urine samples. LC/MS is a highly accurate chemical analytical method.
- Lay-User Study: Ground truth was established by spiking known concentrations of drugs into drug-free pooled urine specimens, confirmed by LC/MS.
8. Sample Size for the Training Set
- This document describes a 510(k) submission for a traditional in-vitro diagnostic device (immunoassay). It does not mention any artificial intelligence (AI) or machine learning (ML) components that would typically require a "training set" in the computational sense.
- The terms "training set" and "test set" are common in AI/ML validation. For a traditional medical device, the studies described (precision, interference, specificity, comparison, lay-user) serve as the "validation set" against pre-defined performance criteria.
- Therefore, N/A for "training set" in the context of AI/ML.
9. How the Ground Truth for the Training Set Was Established
- N/A (as above, no "training set" in the AI/ML context).
- However, if we consider how the device itself was developed, the ground truth for optimizing its performance (e.g., antibody binding, membrane characteristics) would have relied on highly controlled experiments with known concentrations of analytes, likely confirmed by advanced analytical chemistry methods like LC/MS. This process is part of the extensive R&D and quality control that precedes a 510(k) submission.
Ask a specific question about this device
(29 days)
NFV | Oxazepam (BZO) | 862.3170, Benzodiazepine Test System | Toxicology |
| NFY | Cocaine (COC) | 862.3250
The Deepblue Multi-Drug Urine Test Cup is a rapid lateral flow immunoassay for the qualitative detection of 6-Monoacetylmorphine, d-Amphetamine, Benzoylecgonine, Buprenorphine, EDDP, fentanyl, Methadone, d-Methamphetamine, d/l-Methylenedioxymethamphetamine, Morphine, Nortriptyline, Oxazepam, Oxycodone, Phencyclidine, d-Propoxyphene, Secobarbital, and THC-COOH in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:
Test | Calibrator | Cut-off(ng/mL) |
---|---|---|
6-MAM | 6-Monoacetylmorphine | 10 |
AMP | d-Amphetamine | 500 or 1000 |
BAR | Secobarbital | 300 |
BUP | Buprenorphine | 10 |
BZO | Oxazepam | 300 |
COC | Benzoylecgonine | 150 or 300 |
EDDP | 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrolidine | 300 |
FYL | Fentanyl | 1 |
MDMA | Methylenedioxymethamphetamine | 500 |
MET | d-Methamphetamine | 500 or 1000 |
MTD | d/l-Methadone | 300 |
MOP/OPI | Morphine | 300 or 2000 |
OXY | Oxycodone | 100 |
PCP | Phencyclidine | 25 |
PPX | d-Propoxyphene | 300 |
TCA | Nortriptyline | 1000 |
THC | 11-nor-Δ9-THC-COOH | 50 |
The single or multi-test cups can consist of up to seventeen (17) of the above listed analytes in any combination with or without on-board adulteration/specimen validity tests (SVT).
The tests provide only a preliminary result. A more specific alternative chemical method must be used to obtain a confirmed presumptive positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
The Deepblue Home Multi-Drug Urine Test Cup is a rapid qualitative immunoassay. The device provides preliminary results for the detection of one or more of the following drugs.
Code | Substance | Cut-off(ng/mL) |
---|---|---|
AMP | Amphetamine | 1000 or 500 |
BUP | Buprenorphine | 10 |
BAR | Secobarbital | 300 |
BZO | Oxazepam | 300 |
COC | Cocaine | 300 or 150 |
EDDP | EDDP | 300 |
FYL | Fentanyl | 1 |
MET | Methamphetamine | 1000 or 500 |
MDMA | Ecstasy | 500 |
OPI | Morphine | 2000 or 300 |
MTD | Methadone | 300 |
OXY | Oxycodone | 100 |
PCP | Phencyclidine | 25 |
PPX | Propoxyphene | 300 |
TCA | Nortriptyline | 1000 |
THC | Marijuana | 50 |
6-MAM | 6-Monoacetylmorphine | 10 |
This drug test cup may contain any combination of the drug tests listed in the table above.
This test provides only preliminary result. A more specific alternative chemical method must be used to obtain a confirmed presumptive positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
Deepblue Home Muti-Drug Urine Test Cup and Deepblue Muti-Drug Urine Test Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine.
The device is a cup format. Each test device is sealed with two sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format and no longer requires assembly before use.
The provided FDA 510(k) clearance letter details the performance characteristics of the Deepblue Multi-Drug Urine Test Cup and Deepblue Home Multi-Drug Urine Test Cup. This information allows us to describe the acceptance criteria and the study that proves the device meets these criteria.
It's important to note that this device is a qualitative lateral flow immunoassay for initial drug screening, not a diagnostic imaging AI, so the criteria and study methodology differ significantly from those for an AI-powered diagnostic tool. Specifically, there are no references to AI assistance, human readers, or image adjudication, as these are not relevant to this type of device.
Acceptance Criteria and Device Performance for Deepblue Multi-Drug Urine Test Cups
The acceptance criteria for this type of qualitative immunoassay are primarily based on its analytical performance, specifically its ability to correctly identify the presence or absence of target drugs at specified cutoff concentrations. This is demonstrated through precision/reproducibility studies, analytical specificity (cross-reactivity and interference), and method comparison studies against a gold standard (LC-MS/MS).
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly derived from the successful demonstration of performance in the analytical and method comparison studies. The goal is to achieve high agreement rates with the confirmatory method (LC-MS/MS) and consistent results at and around the cutoff concentrations.
Performance Metric Category | Specific Acceptance Criteria (Implicit) | Reported Device Performance |
---|---|---|
Precision/Reproducibility | Consistent results across multiple lots and runs, especially at and near cutoff concentrations. High percentage of correct results for spiked samples at various concentrations. | Across three lots and 25 days of testing (50 runs per concentration), the device showed excellent performance. For concentrations at +50%, +75%, and +100% of cutoff, all results were positive (0-/50+). For concentrations at -50%, -75%, and -100% of cutoff, all results were negative (50-/0+). At the cutoff concentration, there was a mix of positive and negative results, indicating the assay's sensitivity at the threshold (e.g., AMP 1000: Lot 1 had 8-/42+, meaning 8 negatives and 42 positives out of 50 runs). Near the cutoff (e.g., -25% and +25%), expected mixed results were observed, demonstrating the assay's ability to differentiate. |
Analytical Specificity (Cross-Reactivity) | Minimal cross-reactivity with non-target compounds and sufficient cross-reactivity with known metabolites/analogs to ensure broad detection. | Detailed tables provided showing specific cross-reactivity percentages for various compounds. For example, for AMP 1000, D,L-Amphetamine and D-Amphetamine showed 100% cross-reactivity as expected. For BAR 300, Alphenal showed 200% cross-reactivity, and Phenobarbital showed 150%. For FYL 1, Acetyl fentanyl showed 100%, and Butyryl Fentanyl showed 50%. Norfentanyl showed |
Ask a specific question about this device
(24 days)
NFV | Oxazepam (BZO) | 862.3170, Benzodiazepine Test System | Toxicology |
| NFY | Cocaine (COC) | 862.3250
AllTest Multi-Drug Urine Test Cup tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Benzodiazepine, Cocaine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline, Marijuana, Fentanyl, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Tramadol, Propoxyphene and 6-monoacetylmorphine in human urine at the cutoff concentrations of:
Drug (Identifier) | Cut-off level |
---|---|
Amphetamine (AMP) | 500 ng/mL or 1000 ng/mL |
Buprenorphine (BUP) | 10 ng/mL |
Secobarbital (BAR) | 300 ng/mL |
Benzodiazepine (BZO) | 300 ng/mL |
Cocaine (COC) | 150 ng/mL or 300 ng/mL |
Methamphetamine (MET) | 500 ng/mL or 1000 ng/mL |
Methylenedioxymethamphetamine (MDMA) | 500 ng/mL |
Morphine (MOP/OPI) | 300 ng/mL or 2000 ng/mL |
Methadone (MTD) | 300 ng/mL |
Oxycodone (OXY) | 100 ng/mL |
Phencyclidine (PCP) | 25 ng/mL |
Nortriptyline (TCA) | 1000 ng/mL |
Marijuana (THC) | 50 ng/mL |
Fentanyl(FYL) | 1 ng/mL |
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) | 300 ng/mL |
Tramadol (TRA) | 100 ng/mL |
Propoxyphene (PPX) | 300 ng/mL |
6-monoacetylmorphine (6-MAM) | 10 ng/mL |
AllTest Multi-Drug Urine Test Cup can be a single drug test cup or used for any combinations of the above listed analytes. It is for in vitro diagnostic use only. It is intended for OTC use.
The tests may yield positive results for the prescription drugs when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.
The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS or LC/MS is the recommended confirmatory method.
AllTest Multi-Drug Rapid Urine Test Cup tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Benzodiazepine, Cocaine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline, Marijuana, Fentanyl, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Tramadol, Propoxyphene and 6-monoacetylmorphine in human urine at the cutoff concentrations of:
Drug (Identifier) | Calibrator | Cut-off (ng/mL) |
---|---|---|
Amphetamine (AMP) | d-Amphetamine | 500 or 1000 |
Buprenorphine (BUP) | Buprenorphine | 10 |
Secobarbital (BAR) | Secobarbital | 300 |
Benzodiazepine (BZO) | Oxazepam | 300 |
Cocaine (COC) | Benzoylecgonine | 150 or 300 |
Methamphetamine (MET) | d-Methamphetamine | 500 or 1000 |
Methylenedioxymethamphetamine (MDMA) | d,l-Methylenedioxymethamphetamine | 500 |
Morphine (MOP/OPI) | Morphine | 300 or 2000 |
Methadone (MTD) | Methadone | 300 |
Oxycodone (OXY) | Oxycodone | 100 |
Phencyclidine (PCP) | Phencyclidine | 25 |
Nortriptyline (TCA) | Nortriptyline | 1000 |
Marijuana (THC) | 11-nor-Δ9-THC-9 COOH | 50 |
Fentanyl (FYL) | Fentanyl | 1 |
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) | 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine | 300 |
Tramadol (TRA) | Tramadol | 100 |
Propoxyphene (PPX) | Propoxyphene | 300 |
6-monoacetylmorphine (6-MAM) | 6-monoacetylmorphine | 10 |
AllTest Multi-Drug Rapid Urine Test Cup can be a single drug test cup or used for any combination of the above listed analytes. It is for in vitro diagnostic use only.
The tests may yield positive results for the prescription drugs when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.
The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS or LC/MS is the recommended confirmatory method.
AllTest Multi-Drug Urine Test Cup and AllTest Multi-Drug Rapid Urine Test Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine.
The devices are a cup format. Each test device is sealed with sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format.
The provided document describes the analytical and user performance of the "AllTest Multi-Drug Rapid Urine Test Cup" and "AllTest Multi-Drug Urine Test Cup" for detecting various drugs in human urine.
Here's a breakdown of the acceptance criteria and study information:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state pre-defined acceptance criteria in terms of specific percentages for sensitivity, specificity, or agreement. However, the performance studies demonstrate that the device is designed to correctly identify drug presence/absence relative to a defined cutoff concentration. For qualitative drug tests, a common expectation is high agreement rates for samples significantly above or below the cutoff, with some variability allowed for samples near the cutoff.
The reported device performance can be summarized from the precision and lay person studies. The precision studies show ideal performance at concentrations far from the cutoff (100% agreement), and expected variability (around 50% positive/negative calls) at the cutoff concentration. The lay person study similarly shows very high agreement (typically 90-100%) for samples adequately far from the cutoff concentration.
Since no explicit quantitative acceptance criteria are given in the provided text, the reported performance is presented in relation to the ideal behavior of a qualitative assay around its cutoff.
Performance Metric | Acceptance Criteria (Implied) | Reported Device Performance (Summary from studies) |
---|---|---|
Precision (at Cutoff) | Expected to be approximately 50% positive and 50% negative calls at the exact cutoff concentration, with 100% agreement for concentrations significantly above or below the cutoff. | Achieved: For all tested drugs (MOP 2000, EDDP, COC 300, TRA, PPX, AMP 1000, MET 1000, 6-MAM), results at the cutoff concentration generally show a mix of positive and negative calls (e.g., 20-30 positive / 30-20 negative out of 50 tests per lot), while concentrations +/- 50-100% from cutoff show 100% agreement. |
Accuracy (Method Comparison) | High agreement with LC-MS/MS, especially for samples sufficiently above or below the cutoff. Limited discordant results expected primarily near the cutoff. | Achieved: Very high agreement for drug-free, low negative ( +50% cutoff) samples (typically 100% agreement). Limited discordant results ( |
Ask a specific question about this device
(411 days)
. | Toxicology |
| DIO | 862.3250/Cocaine system.
The LYHER® Oral fluid Multi-Drug Test Kit (Cube) is a rapid lateral flow immunoassay for the qualitative detection of d-Amphetamine, d-Methamphetamine, Benzoylecgonine, Morphine, Phencyclidine and Delta-9-Tetrahydrocanabinol in human oral fluid. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:
Test | Calibrator | Cut-off (ng/mL) |
---|---|---|
Opiates(OPI) | Morphine | 40 |
Cocaine (COC) | Benzoylecgonine | 20 |
Amphetamine (AMP) | d-Amphetamine | 50 |
Marijuana (THC) | Delta-9-Tetrahydrocannabinol | 40 |
Methamphetamine (MET) | d-Methamphetamine | 50 |
Phencyclidine (PCP) | Phencyclidine | 10 |
The single or multi-test panels can consist of the above insted analytes in anycombination, up to a maximum of 6 analytes. The tests provide only a preliminary result. A more specific alternate chemical must be used in order to obtain a confirmed analytical test result. Gas Chromatography/Mass Spectrometry (GC/MS), Liquid Chromatography / Mass Spectrometry (LC/MS) and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
The LYHER® Oral fluid Multi-Drug Test Kit (Cube) is an immunochromatographic assay that uses a lateral flow system for the qualitative detection of d-Amphetamine, d-Methamphetamine, Benzoylecgonine, Morphine, Phencyclidine and Delta-9-Tetrahydrocannabinol in human oral fluid. The LYHER® Oral fluid Multi-Drug Test Kit (Cube) device consists of a cube device, an oral fluid collection swab and a package insert.
The document describes the analytical performance studies for the LYHER® Oral fluid Multi-Drug Test Kit (Cube), a rapid lateral flow immunoassay. The device is designed to detect d-Amphetamine, d-Methamphetamine, Benzoylecgonine, Morphine, Phencyclidine, and Delta-9-Tetrahydrocannabinol in human oral fluid.
Here's a breakdown of the requested information based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state "acceptance criteria" as a pass/fail threshold for performance, but rather presents the results of precision-reproducibility studies across various concentrations relative to the cut-off. The performance is summarized by the number of positive and negative results at each concentration.
To construct a table of "acceptance criteria" (inferred as target performance, though not explicitly defined as such) and reported performance, we can focus on the precision-reproducibility studies around the cut-off and the comparison study data. For the comparison study, the implicit acceptance criterion is that the device accurately identifies samples as positive or negative relative to the LC/MS confirmation.
Inferred Acceptance Criteria (Conceptual) and Reported Device Performance
Parameter / Concentration Level | Expected Outcome (Inferred) | Reported Performance (Counts from all operators and lots, combined) |
---|---|---|
D-Amphetamine | ||
-100% cut off (Negative) | All Negative | 0+/180- (All Negative) |
-75% cut off (Negative) | All Negative | 0+/180- (All Negative) |
-50% cut off (Negative) | All Negative | 0+/180- (All Negative) |
-25% cut off (Negative) | Majority Negative, some Positive | Approx. 8+/532- (Some false positives) |
Cut off (Threshold) | Mix of Positive/Negative (close to 50/50 split implies good performance at cutoff) | Approx. 142+/348- (for combined precision study across drugs and operators), For the comparison study, at/around cutoff, you see 47-49+/11-13- for precision (ideal is balanced +/-), and 1+/74- for comparison at -50 cutoff for Amphetamines, and 180+/9- for comparison at cutoff for Amphetamines |
+25% cut off (Positive) | Majority Positive, some Negative | Approx. 165+/15- (Some false negatives) |
+50% cut off (Positive) | All Positive | 60+/0- (All Positive) |
+75% cut off (Positive) | All Positive | 60+/0- (All Positive) |
+100% cut off (Positive) | All Positive | 60+/0- (All Positive) |
Cocaine | ||
-100% cut off (Negative) | All Negative | 0+/180- (All Negative) |
-75% cut off (Negative) | All Negative | 0+/180- (All Negative) |
-50% cut off (Negative) | All Negative | 0+/180- (All Negative) |
-25% cut off (Negative) | Majority Negative, some Positive | ~17+/523- |
Cut off (Threshold) | Mix of Positive/Negative | ~145+/245- |
+25% cut off (Positive) | Majority Positive, some Negative | ~168+/12- |
+50% cut off (Positive) | All Positive | 60+/0- (All Positive) |
+75% cut off (Positive) | All Positive | 60+/0- (All Positive) |
+100% cut off (Positive) | All Positive | 60+/0- (All Positive) |
All other Drugs (similarly to above) | ||
At or below -50% Cut-off | Generally all Negative | Very high negative rate (e.g., 0+/60- for most negative concentrations in precision study) |
At or above +50% Cut-off | Generally all Positive | Very high positive rate (e.g., 60+/0- for most positive concentrations in precision study) |
Near Cut-off (e.g. +/-25%) | Mixed results, reflecting the nature of a qualitative test around cutoff | Variable, as expected (e.g., for Amphetamine at -25% cut-off, ~2+/58- across operators/lots for precision; 47-49+/11-13- at cut-off) |
Comparison Study vs. LC/MS: | ||
True Negative (Analyte absent) | Device Negative | E.g., Amphetamine: 360 Negative (device) out of 360 Negative (LC/MS) |
False Positive (Analyte negative, Device positive) | Very low / zero | E.g., Amphetamine: 0 Positive (device) where LC/MS was Negative |
True Positive (Analyte positive, Device positive) | Device Positive | E.g., Amphetamine: 180 Positive (device) at Cut off to +50% cut off vs. 180 identified by LC/MS |
False Negative (Analyte positive, Device negative) | Very low / zero | E.g., Amphetamine: 9 Negative (device) where LC/MS was Cut off to +50% cut off (indicating 9 False Negatives in that range) |
Note on Acceptance Criteria: The document provides raw performance data. For a qualitative immunoassay, the "acceptance criteria" are usually demonstrated by a high rate of negativity well below the cutoff, a high rate of positivity well above the cutoff, and a predictable transition zone around the cutoff concentration. The data presented supports this.
2. Sample size used for the test set and the data provenance
-
Test Set Sample Size:
- Precision-Reproducibility-Cut-Off Study: For each drug, 9 concentration levels were tested. For each concentration, there were 2 runs per day for 30 days, across 3 device lots, and by 3 operators.
- This means 9 concentrations * 2 runs/day * 30 days * 3 lots * 3 operators = 4860 total tests per drug analyte for the precision study.
- Each concentration within each operator/lot combination was tested 60 times (2 runs/day * 30 days). So for each specific concentration, lot, and operator, n=60.
- Comparison Studies (Method Comparison):
- Negative oral fluid: 360 samples (across all operators and sites for each drug).
- Positive oral fluid at various ranges:
- +50% cut off: For D-Amphetamine, 183 samples. For Cocaine, 186 samples. For d-Methamphetamine, 189 samples. For Morphine, 180 samples. For Phencyclidine, 192 samples. For Delta-9-Tetrahydrocannabinol, 195 samples.
- The total number of samples for the comparison study for each drug is 360 (negatives) + (sum of the positive ranges for that drug). For example, for D-Amphetamine: 360 + 93 + 74 + 180 + 183 = 890 samples. This applies similarly to other drugs.
- Precision-Reproducibility-Cut-Off Study: For each drug, 9 concentration levels were tested. For each concentration, there were 2 runs per day for 30 days, across 3 device lots, and by 3 operators.
-
Data Provenance: The document states "Method comparison studies for the LYHER Oral fluid Multi-Drug Test Kit(Cube) were performed at three testing sites with three operators at each site." The location of these sites is not explicitly mentioned (e.g., country of origin), but the submitter is based in China. The data origin is prospective as samples were prepared by spiking known concentrations into negative oral fluid for analytical studies.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Ground Truth Establishment: The ground truth for the test set for both the precision and comparison studies was established by Laboratory Confirmed Concentrations, specifically using LC/MS (Liquid Chromatography/Mass Spectrometry). For the precision study, it states: "Each drug concentration was confirmed by LC/MS." For the method comparison studies, it states: "Operators tested the samples using the candidate device and the results were compared to LC/MS results."
- Number of Experts/Qualifications: LC/MS is a laboratory analytical method, not reliant on subjective expert interpretation like radiological imaging. Therefore, there were no "experts" in the sense of human annotators (e.g., radiologists) involved in establishing the ground truth via consensus or adjudication. The "experts" would be the qualified laboratory personnel performing and interpreting the LC/MS results, whose qualifications are implicit given the professional standards for such testing.
4. Adjudication method for the test set
- Since LC/MS is used to establish quantitative drug concentrations, and the device provides a qualitative "positive" or "negative" result based on a defined cutoff, there is no expert adjudication method (like 2+1, 3+1) mentioned or necessary. The device's result is compared directly to the LC/MS result relative to the established cut-off.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- This question is not applicable to this device. This is a qualitative diagnostic test (immunoassay) for drug detection in oral fluid, not an AI-powered image analysis device that assists human readers. The "operators" mentioned are performing the test, not interpreting complex medical images.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- This device is a standalone test (a rapid lateral flow immunoassay), essentially an "algorithm only" in the sense that its chemical reactions produce a visual result (line presence/absence). Human interaction involves collecting the sample, applying it to the device, and visually interpreting the presence or absence of test lines. There isn't a separate "algorithm" that operates outside of the device itself to interpret the results. The performance data presented measures the device's inherent analytical accuracy against known concentrations and LC/MS.
7. The type of ground truth used
- The ground truth used for both precision and comparison studies was laboratory-confirmed drug concentrations via LC/MS (Liquid Chromatography/Mass Spectrometry). This is a highly accurate and quantitative analytical method, considered a gold standard for drug detection and quantification.
8. The sample size for the training set
- This device is a lateral flow immunoassay, not a machine learning or AI-based device that requires a "training set" in the computational sense. Its performance is based on the inherent chemical and biological properties of the reagents and test strip design. Therefore, the concept of a "training set" for model development is not applicable. The studies described are analytical validation studies, not AI model training.
9. How the ground truth for the training set was established
- As explained in point 8, there is no "training set" for an AI model. The ground truth for the analytical validation (which is analogous to testing the device's inherent design performance) was established by LC/MS confirmation of spiked drug concentrations in oral fluid.
Ask a specific question about this device
(59 days)
| 862.3250
AllTest Multi-Drug Rapid Test Cup tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Secobarbital, Benzodiazepines, Cocaine, 2ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencycline, Nortriptyline, Marijuana, Tramadol, Propoxyphene, Fentanyl and 6monoacetylmorphine in human urine at the cutoff concentrations of:
Drug (Identifier) | Cut-off level |
---|---|
Amphetamine (AMP) | 500 or 1000 ng/mL |
Buprenorphine (BUP) | 10 ng/mL |
Secobarbital (BAR) | 300 ng/mL |
Benzodiazepines (BZO) | 300 ng/mL |
Cocaine (COC) | 150 or 300 ng/mL |
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) | 300 ng/mL |
Methamphetamine (MET) | 500 or 1000 ng/mL |
Methylenedioxymethamphetamine (MDMA) | 500 ng/mL |
Morphine (MOP/OPI) | 300 or 2000 ng/mL |
Methadone (MTD) | 300 ng/mL |
Oxycodone (OXY) | 100 ng/mL |
Phencyclidine (PCP) | 25 ng/mL |
Nortriptyline (TCA) | 1000 ng/mL |
Marijuana (THC) | 50 ng/mL |
Tramadol (TRA) | 100 ng/mL |
Propoxyphene (PPX) | 300 ng/mL |
Fentanyl(FYL) | 1 ng/mL |
6-monoacetylmorphine (6-MAM) | 10 ng/mL |
AllTest Multi-Drug Rapid Test Cup can be a single drug test cup or used for any combination |
est Multi-Drug Rapid Test Cup can be a single drug test cup or used for any combination of the above listed analytes. It is for in vitro diagnostic use only. It is intended for OTC use.
The tests may yield positive results for the prescription drugs when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these crugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS or LC/MS is the recommended confirmatory method.
AllTest Multi-Drug Test Cup tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secodiazepines, Cocaine, 2- ethylidene-1.5dimethyl-3,3- diphenylpyrrolidine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline, Marijuana, Tramadol, Propoxyphene Fentanyl and 6-monoacetylmorphine in human urine at the cutoff concentrations of:
Drug (Identifier) | Calibrator | Cut-off (ng/mL) |
---|---|---|
Amphetamine (AMP) | d-Amphetamine | 500 or 1000 |
Buprenorphine (BUP) | Buprenorphine | 10 |
Secobarbital (BAR) | Secobarbital | 300 |
Benzodiazepines (BZO) | Oxazepam | 300 |
Cocaine (COC) | Benzoylecgonine | 150 or 300 |
2-ethylidene-1,5-dimethyl-3,3- | ||
diphenylpyrrolidine (EDDP) | 2-ethylidene-1,5-dimethyl-3,3- | |
diphenylpyrrolidine | 300 | |
Methamphetamine (MET) | d-Methamphetamine | 500 or 1000 |
Methylenedioxymethamphetamine (MDMA) | d,l-Methylenedioxymethamphetamine | 500 |
Morphine (MOP/OPI) | Morphine | 300 or 2000 |
Methadone (MTD) | Methadone | 300 |
Oxycodone (OXY) | Oxycodone | 100 |
Phencyclidine (PCP) | Phencyclidine | 25 |
Nortriptyline (TCA) | Nortriptyline | 1000 |
Marijuana (THC) | 1-nor-Δ9-THC-9 COOH | 50 |
Tramadol (TRA) | Tramadol | 100 |
Propoxyphene (PPX) | Propoxyphene | 300 |
Fentanyl(FYL) | Fentanyl | 1 |
6-monoacetylmorphine (6-MAM) | 6-monoacetylmorphine | 10 |
AllTest Multi-Drug Test Cup can be a single drug test cup or used for any combination of the above listed analytes. It is for in vitro diagnostic use only.
The tests may yield positive results for the prescription drugs when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.
The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS or LC/MS is the recommended confirmatory method.
AllTest Multi-Drug Rapid Test Cup and AllTest Multi-Drug Test Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine.
The devices are a cup format. Each test device is sealed with sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format and no longer requires assembly before use.
Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:
Device Name: AllTest Multi-Drug Rapid Test Cup; AllTest Multi-Drug Test Cup
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are generally established by the performance characteristics demonstrated in the studies, particularly the agreement/precision at and around the cutoff concentrations, and the specificity. Since this is an in vitro diagnostic (IVD) device, the "performance" here refers to its accuracy in detecting the target analytes and its overall reliability in a diagnostic context. This is fundamentally about how well the device matches a known, established ground truth.
Acceptance Criterion | Reported Device Performance (Summary) |
---|---|
Precision/Reproducibility | For Tramadol (TRA), Propoxyphene (PPX), Fentanyl (FYL), and 6-monoacetylmorphine (6-MAM), results at -100%, -75%, -50% cut-off were 100% negative calls. For +100%, +75%, +50% cut-off, results were 100% positive calls (with a few exceptions for PPX Lot 3 at +25% cutoff, and 6-MAM Lot 2 at +25% cutoff). At the exact cutoff, performance ranged from 22-28 positive/negative calls out of 50 total tests, showing expected variability around the cutoff. This demonstrates the device's ability to consistently provide the expected result across multiple runs and lots at various concentrations relative to the cutoff. |
Linearity/Assay Reportable Range | Not applicable, as the device is intended for qualitative use only. |
Stability | Stable at 2-30°C for 24 months based on real-time stability study. |
Analytical Specificity/Interference | Cross-Reactivity: Demonstrated varying degrees of cross-reactivity for structurally related compounds (e.g., N-Desmethyl-cis-tramadol for TRA, Norpropoxyphene for PPX, Acetyl fentanyl, Acrylfentanyl for FYL, Diacetylmorphine for 6-MAM). Specificity was shown by very low or no cross-reactivity (+50%):** The device typically reported high positive counts and 0 negative counts, indicating good agreement for positives. |
Small numbers of discordant results (e.g., a sample near the cutoff called positive by LC/MS/MS but negative by the device, or vice-versa) were observed consistently with the nature of qualitative cutoff tests. | |
Lay Person Study Agreement | For all tested drugs (AMP, BAR, BZO, BUP, COC, EDDP, MDMA, MET, MOP, MTD, OXY, PCP, TCA, THC, TRA, PPX, FYL, 6-MAM) across two configurations: |
- Agreement at -100%, -75%, -50% Cutoff: Generally 100% negative agreement.
- Agreement at +50%, +75% Cutoff: Generally 100% positive agreement.
- Agreement at -25% and +25% Cutoff: Ranged from 85.0% to 95.0% negative/positive agreement, respectively, demonstrating expected variability around the cutoff.
All participants found the instructions easy to understand and follow (Flesch-Kincaid Grade Level 7). |
2. Sample Size Used for the Test Set and Data Provenance
-
Precision/Reproducibility Study:
- Sample Size: For each drug and each concentration (-100%, -75%, -50%, -25%, Cutoff, +25%, +50%, +75%, +100% of cutoff), tests were performed in 2 runs per day for 25 days, using 3 lots of test cups. This means for each concentration and drug, there were 50 tests per lot (2 runs/day * 25 days) and 150 tests across 3 lots.
- Data Provenance: Urine samples spiked with target drugs into drug-free urine. The drug concentrations were confirmed by LC-MS/MS. This is an analytical/laboratory-controlled study, not from actual patients.
-
Analytical Specificity/Interference Study:
- Sample Size: Not explicitly stated as a total number, but drugs and interfering compounds were spiked into drug-free urine samples and tested using 3 lots of devices for each interference condition.
- Data Provenance: Laboratory prepared urine samples (drug-free urine spiked with various compounds).
-
Method Comparison Study:
- Sample Size: 80 unaltered urine clinical samples (40 negative and 40 positive) for each drug.
- Data Provenance: Unaltered clinical urine samples. The country of origin is not specified but it is an in-house study.
-
Lay Person Study:
- Sample Size: 280 lay persons.
- Data Provenance: Urine samples prepared by spiking drug(s) into drug-free pooled urine specimens. The concentrations were confirmed by LC-MS/MS. This is an analytical/laboratory-controlled study designed to evaluate user comprehension and ease of use, not true clinical performance with patient samples.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
-
Precision/Reproducibility, Analytical Specificity/Interference, Lay Person Study: Ground truth was established by LC-MS/MS (Liquid Chromatography-Mass Spectrometry/Mass Spectrometry). This is a highly accurate and standardized analytical method for drug concentration determination. No human experts were explicitly stated to establish this ground truth, as it's an objective chemical analysis.
-
Method Comparison Study: Ground truth was established by LC-MS/MS. Similar to the above, this is an objective analytical method.
4. Adjudication Method for the Test Set
- For the Method Comparison Study, the device results were compared directly to the LC-MS/MS results. There is no mention of an adjudication process (e.g., 2+1, 3+1 expert review) for the ground truth itself, as LC-MS/MS is considered the definitive truth.
- For the Lay Person Study, the "ground truth" for the expected positive/negative call was based on the known spiked concentrations confirmed by LC-MS/MS. The participants' interpretations of the device results were then compared to this established truth. There was no adjudication of the lay person's reading, but rather an assessment of their accuracy against the known sample content.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, an MRMC comparative effectiveness study was not done.
- This device is a qualitative lateral flow immunochromatographic assay ("rapid test cup") for in vitro diagnostic use, intended for direct reading by a user (either professionals or lay persons). It does not involve AI or human readers interpreting complex images or data assisted by AI.
- The closest analogy is the "Lay Person Study," which assesses how well lay users can interpret the results without assistance on complex outputs, but there is no AI component.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Yes, in spirit, the analytical performance studies (precision, specificity, stability) and the LC-MS/MS ground truth in the method comparison study represent "standalone" performance. The device itself (the immunochromatographic strip) is the "algorithm," and its chemical reactions are assessed according to established scientific principles, independent of human interpretation prior to result line formation.
- However, the final reading of the lines on the rapid test cup is still a human-in-the-loop step, even for laboratory personnel in the precision and method comparison studies, and especially for lay users in their study. The intent of these "rapid test" types of devices is precisely for direct human interpretation.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- The primary ground truth used across all analytical and method comparison studies was LC-MS/MS (Liquid Chromatography-Mass Spectrometry/Mass Spectrometry). This is a highly accurate and quantitative laboratory method used for definitive drug detection and concentration measurement.
8. The Sample Size for the Training Set
- This submission describes a premarket notification (510(k)) for an in vitro diagnostic device. Unlike AI/ML devices, these types of products typically do not involve "training sets" in the machine learning sense.
- The manufacturing process, antibody selection, and assay design for these immunochromatographic tests are developed and refined through internal R&D, often using a range of samples and iterations. The document does not describe a formal 'training set' analogous to those used for AI algorithms. The data presented are for validation and verification of the finished device.
9. How the Ground Truth for the Training Set was Established
- As noted above, a "training set" in the AI/ML context is not applicable here. The development of the reagents and test design would have relied on standard laboratory practices for developing highly specific and sensitive immunoassays, with positive and negative controls and calibration against known concentrations, usually verified by gold-standard analytical methods like LC-MS/MS.
Ask a specific question about this device
(41 days)
| 862.3250
Dochek® Multi-Drug Urine Test Cup is an immunoassay for the qualitative determination of single or multiple drugs in human urine at the following cutoff concentrations.
Dochek® Multi-Drug Urine Test Cup offers any combinations from 1 to 17 drugs but only one cutoff concentration under same drug condition will be included per device.
It is intended for over-the-counter (OTC) use. For in vitro diagnostic use only.
The test provides only preliminary results. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a postive result. To obtain a confirmed analytical result, a more specific alternate chemical method is needed. GC/MS or LC/MS is the recommended confirmatory method.
Dochek® Multi-Drug Urine Test Cup Pro is an immunoassay for the qualitative determination of single or multiple drugs in human urine at the following cutoff concentrations.
Dochek® Multi-Drug Urine Test Cup Pro offers any combinations from 1 to 17 drugs but only one cutoff concentration under same drug condition will be included per device.
For in vitro diagnostic use only.
The test provides only preliminary results. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. To obtain a confirmed analytical result, a more specific alternate chemical method is needed. GC/MS or LC/MS is the recommended confirmatory method.
Dochek® Multi-Drug Urine Test Cup Pro and Dochek® Multi-Drug Urine Test Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine at or above the cut-off levels as indicated. The products are single use in vitro diagnostic medical devices.
This device is a cup format, with the test strips integrated into the plastic cup provided, and the urine sample is collected directly into the cup containing the strips. Each cup device is sealed in an aluminum foil pouch with two sachets of desiccant. The device is in a ready-to-use format and no longer requires assembly before use.
The provided text describes the Dochek® Multi-Drug Urine Test Cup and Dochek® Multi-Drug Urine Test Cup Pro, which are immunoassays for the qualitative determination of single or multiple drugs in human urine. The acceptance criteria and the studies performed to demonstrate performance are detailed below. It is important to note that the acceptance criteria are implied by the reported performance, as explicit criteria are not stated in terms of thresholds for sensitivity/specificity. Instead, the studies demonstrate accuracy and agreement against a reference method and other concentrations.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the goal of demonstrating substantial equivalence to a predicate device and achieving certain performance levels in precision, accuracy against a reference method, and lay user comprehension.
Notes on the tables below:
- "Cutoff" refers to the specified ng/mL for each drug.
- "+" indicates a preliminary positive result (drug detected).
- "-" indicates a negative result (drug not detected).
- LC-MS/MS is the reference method for confirming drug concentrations.
- The "Percentage of correct results (%)" in the Lay User Study is derived from the reported counts of negative and positive results compared to the expected outcome given the drug concentration relative to the cutoff. For example, a sample at -100% cutoff should be negative, and a sample at +100% cutoff should be positive.
1.1. Precision/Reproducibility Study (Fentanyl (FTY) Example)
Drug | Lot Number | Drug Concentration Categories (relative to Cutoff = 1 ng/mL) | Reported Performance (Negative/Positive) | Implied Acceptance Criteria |
---|---|---|---|---|
FTY | Lot I | +100%, +75%, +50%, +25% | 0-/50+ (100% Positive) | 100% Positive |
Cutoff (1 ng/mL) | 12-/38+ (76% Positive) | High % Positive | ||
-25%, -50%, -75% | 50-/0+ (100% Negative) | 100% Negative | ||
-100% | 50-/0+ (100% Negative) | 100% Negative | ||
FTY | Lot II | +100%, +75%, +50%, +25% | 0-/50+ (100% Positive) | 100% Positive |
Cutoff (1 ng/mL) | 11-/39+ (78% Positive) | High % Positive | ||
-25%, -50%, -75% | 50-/0+ (100% Negative) | 100% Negative | ||
-100% | 50-/0+ (100% Negative) | 100% Negative | ||
FTY | Lot III | +100%, +75%, +50%, +25% | 0-/50+ (100% Positive) | 100% Positive |
Cutoff (1 ng/mL) | 11-/39+ (78% Positive) | High % Positive | ||
-25%, -50%, -75% | 50-/0+ (100% Negative) | 100% Negative | ||
-100% | 50-/0+ (100% Negative) | 100% Negative |
1.2. Method Comparison Study (Fentanyl (FTY) Example)
This study compares the device's results to LC-MS/MS, a highly accurate confirmatory method. The "Discordant results" highlight where the device deviated from the LC-MS/MS findings.
Drug | LC-MS/MS Result Category | Device Result | Viewer A | Viewer B | Viewer C | Implied Acceptance Criteria |
---|---|---|---|---|---|---|
FTY | Drug-Free ( +50% Cutoff) | + | 13 | 13 | 13 | |
- | 0 | 0 | 0 |
Summary of Discordant Fentanyl Results (FTY 1 ng/mL):
Drug | Operator | Sample ID | LC-MS/MS Result (ng/mL) | Dochek Result (Viewer) | Expected Result (based on Cutoff=1 ng/mL) | Discordance type |
---|---|---|---|---|---|---|
FTY | Viewer C | F046 | 0.945 | + | - | False Positive |
FTY | Viewer A, B, C | F062 | 1.012 | - | + | False Negative |
FTY | Viewer A, C | F083 | 1.020 | - | + | False Negative |
FTY | Viewer B | F049 | 1.044 | - | + | False Negative |
1.3. Lay Person Study (Configuration 1 & 2 - Example for AMP, BAR, BUP, BZO, COC, EDDP, MDMA, MET, OPI, MTD, OXY, PCP, PPX, TCA, THC, 6-MAM, FTY)
This study evaluates the device's performance when used by non-professionals. Results are generally expected to be 100% correct for samples at -100% and +100% of the cutoff, and high percentages for other concentrations (e.g., ≥90-95% for +/-25% of cutoff).
Drug (Cutoff shown) | Results Category | Drug Concentration Categories (relative to Cutoff) | Reported Performance (% Correct Results) | Implied Acceptance Criteria (Typically ≥95% at +/-25% cutoff, 100% elsewhere) |
---|---|---|---|---|
AMP (1000 ng/mL) | Correct | -100%, -75%, -50%, -25% | 100% | 100% |
+25%, +50%, +75% | 100% | 100% | ||
BAR (300 ng/mL) | Correct | -100%, -75%, -50%, -25% | 100% | 100% |
+25% | 95% | high % (e.g., ≥90%) | ||
+50%, +75% | 100% | 100% | ||
... similar data for many drugs ... | ||||
FTY (1 ng/mL) | Correct | -100%, -75%, -50% | 100% | 100% |
-25% | 95% | high % (e.g., ≥90%) | ||
+25%, +50%, +75% | 100% | 100% |
2. Sample Size and Data Provenance
Precision Study:
- Sample Size (Test Set): For Fentanyl, 50 tests were performed at each of the 9 concentration levels (+/-100%, +/-75%, +/-50%, +/-25% of cutoff, and cutoff) across 3 lots, for a total of 9 concentrations * 50 measurements * 3 lots = 1350 tests.
- Data Provenance: Samples were prepared by spiking target drug in drug-free urine samples. The source of the drug-free urine or spiked drugs is not explicitly stated in terms of country of origin. This was a prospective study, with samples specifically prepared for the testing.
Method Comparison Study:
- Sample Size (Test Set): 80 unaltered clinical urine samples were used for each drug (40 negative and 40 positive). For Fentanyl, this means 80 samples were tested.
- Data Provenance: Unaltered clinical urine samples. The country of origin of these clinical samples is not specified. This appears to be a retrospective study using existing clinical samples.
Lay Person Study:
- Sample Size (Test Set): 280 lay users participated.
- Configuration 1: 140 users (68 male, 72 female).
- Configuration 2: 138 users (74 male, 64 female).
- Across 7 concentration levels (+/-100%, +/-75%, +/-50%, +/-25% of cutoff, and cutoff), with 20 samples per concentration level for each drug. This means for each drug, 7 * 20 = 140 results were generated by lay users.
- Data Provenance: Urine samples were prepared by spiking drug(s) into drug-free pooled urine specimens. The source of the drug-free pooled urine or spiked drugs is not explicitly stated in terms of country of origin. This was a prospective study.
3. Number of Experts and Qualifications for Ground Truth
- Precision Study: Ground truth for sample concentrations was confirmed by LC-MS/MS. This method is a highly qualified and generally accepted gold standard for drug concentration determination, not relying on human expert interpretation of the test result itself.
- Method Comparison Study: Ground truth was established by LC-MS/MS results. The operators in this study were "three operators" (presumably laboratory personnel or technicians, but their specific qualifications are not detailed). These operators read the device results, which were then compared to the LC-MS/MS ground truth.
- Lay Person Study: Ground truth for sample concentrations was confirmed by LC-MS/MS. The lay users themselves provided the device readings, and the percentage of correct results was calculated against the LC-MS/MS confirmed concentrations.
4. Adjudication Method for the Test Set
- Precision Study: The results are quantitative (counts of positive/negative) based on pre-defined concentrations. No adjudication method is explicitly described for subjective interpretation as the test is qualitative and the results are directly read as positive or negative by trained personnel (implied).
- Method Comparison Study: "Three operators" read the device results. The individual results for each viewer (A, B, C) are presented. There is no explicit adjudication method (e.g., 2-out-of-3 consensus) mentioned to derive a single device result per sample if the operators disagreed. The discordant results table shows instances where operators disagreed, or where the device result from an individual operator disagreed with LC-MS/MS.
- Lay Person Study: Lay users performed the tests independently. There is no mention of an adjudication process among lay users for their readings. Each participant provided a single result for their assigned sample/device.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was mentioned to quantify the improvement of human readers with AI assistance versus without AI assistance. The device is a lateral flow immunoassay, not an AI-powered diagnostic for image interpretation or similar tasks often associated with MRMC studies.
6. Standalone Performance Study
Yes, standalone performance was conducted for the device.
- Precision Study: The device's inherent precision was evaluated across different drug concentrations and lots, independent of human interpretation variability (though human reading is still involved for the qualitative result).
- Method Comparison Study: The device's performance against the gold standard (LC-MS/MS) was evaluated by three operators independently, representing a standalone assessment of the device's accuracy in a laboratory setting.
- Lay Person Study: This study specifically assessed the standalone performance of the device when used by the intended lay users, including their ability to follow instructions and interpret results correctly.
7. Type of Ground Truth Used
The primary ground truth used for evaluating the device's accuracy in all relevant studies (Precision, Method Comparison, Lay Person) was:
- LC-MS/MS (Liquid Chromatography-Mass Spectrometry/Mass Spectrometry): This is a highly sensitive and specific analytical chemistry technique used to precisely confirm the presence and concentration of drugs and their metabolites in urine samples. This serves as the objective, quantitative ground truth for drug concentrations.
8. Sample Size for the Training Set
The provided document describes performance studies (precision, method comparison, lay person study) for the Dochek® Multi-Drug Urine Test Cup devices. These are immunoassay devices, not machine learning or AI-based devices that typically have "training sets" in the computational sense. The document does not describe any such training set for an algorithm. The development of the immunoassay itself relies on chemical and biological principles rather than algorithm training.
9. How the Ground Truth for the Training Set Was Established
Since there is no "training set" in the context of a machine learning algorithm for this immunoassay device, this question is not applicable. The device's performance characteristics are inherent to its biochemical design.
Ask a specific question about this device
(43 days)
| Toxicology (91) |
| NFY
Cocaine | II | 21 CFR § 862.3250
The AssureTech Panel Dip Tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Methamphetamine, Fentanyl, Norfentany], Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, EDDP, Nortriptyline and d-Propoxyphene in human urine at the cutoff concentrations listed. The single or multi-test panels can consist of up to seventeen (17) of the above listed analytes in any combination. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. For in vitro diagnostic use only.
The AssureTech Quick Cup Tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Methamphetamine, Fentanyl, Norfentany], Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, EDDP, Nortriptyline and d-Propoxyphene in human urine at the cutoff concentrations listed. The single or multi-test panels can consist of up to seventeen (17) of the above listed analytes in any combination. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. For in vitro diagnostic use only.
The AssureTech Multi-drug Urine Test Panel are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Cocaine, Marijuana, Methamphetamine, Morphine, Fentanyl, Norfentanyl, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, EDDP, Nortriptyline, d-Propoxyphene and adulterants in human urine at the cutoff concentrations listed. The single or multi-test panel can consist of up to seventeen (17) of the above listed analytes in any combination. It is for in vitro diagnostic use only. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
The AssureTech Multi-drug Urine Test Cup are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Marijuana, Methamphetamine, Morphine, Fentanyl, Norfentanyl, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, EDDP, Nortriptyline, d-Propoxyphene and adulterants in human urine at the cutoff concentrations listed. The single or multi-test cups can consist of up to seventeen (17) of the above listed analytes in any combination. It is for in vitro diagnostic use only. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
The AssureTech Panel Dip Tests and AssureTech Quick Cup Tests are immunochromatographic assays that use a lateral flow system for the qualitative detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Fentanyl, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, EDDP. Nortriptyline and Propoxyphene (target analytes) in human urine. The products are single use in vitro diagnostic devices, which come in the formats of Panel Dip Cards or Cups. Each test kit contains a Test Device (in one of the two formats), a package insert and a urine cup for sample collection. Each test device is sealed with a desiccant in an aluminum pouch.
The provided document describes the FDA 510(k) premarket notification for AssureTech Panel Dip Tests and Quick Cup Tests, which are in vitro diagnostic devices for qualitative and simultaneous detection of various drugs of abuse in human urine. The document focuses on demonstrating substantial equivalence to a predicate device (K181768).
Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:
Acceptance Criteria and Reported Device Performance
The acceptance criteria for this type of device are primarily related to its analytical performance, specifically its ability to accurately detect the presence or absence of target drugs at specified cutoff concentrations. The device is a qualitative test, meaning it provides a "positive" or "negative" result, rather than a quantitative measurement.
The study demonstrates performance through:
- Precision: Consistency of results across multiple runs and lots, especially near the cutoff concentrations.
- Specificity: Ability to react only with the target drug/metabolite and not with other substances or structurally similar compounds.
- Interference: Lack of false positives/negatives due to common interfering substances in urine, or variations in urine specific gravity and pH.
- Method Comparison: Agreement of device results with a known, more precise reference method (LC/MS).
- Lay-user study: Evaluation of the device's performance when used by non-professionals, assessing ease of use and accuracy of interpretation.
Here is a table summarizing the reported device performance for Fentanyl (FYL), Norfentanyl (NFYL), and as an example for another drug, Amphetamine (AMP) from the "Lay-user study" data. The document does not explicitly state numerical "acceptance criteria" for each performance metric, but rather presents the results of the studies conducted to show sufficient performance for regulatory clearance. The implicit acceptance criterion for a qualitative test like this is generally very high accuracy, especially around the cutoff, and a low rate of false positives/negatives.
Table of Performance for Key Drugs (Fentanyl, Norfentanyl, Amphetamine)
Precision Study (Fentanyl - Panel Dip/Quick Cup, Norfentanyl - Panel Dip/Quick Cup):
The precision data is presented for three lots and various concentrations relative to the cutoff. The data shows very high consistency. For instance, for Fentanyl:
- At -100%, -75%, -50% cut off (negative range), all 50 tests across 3 lots consistently yielded negative results (50-/0+).
- At +25%, +50%, +75%, +100% cut off (positive range), all 50 tests consistently yielded positive results (50+/0-).
- At the cutoff concentration, the device shows variability, as expected for tests near the decision threshold. For Fentanyl Panel Dip, results were 28+/22-, 29+/23-, 28+/22- for Lot 1, 2, 3 respectively (meaning some tests were positive and some negative at the cutoff). This variability is inherent for qualitative tests around the cutoff and implies that some samples at the cutoff may read positive and others negative, which is acceptable performance for a qualitative test. Similar patterns are observed for Quick Cup Fentanyl, Panel Dip Norfentanyl, and Quick Cup Norfentanyl.
Method Comparison Study (Fentanyl - Panel Dip/Quick Cup, Norfentanyl - Panel Dip/Quick Cup):
This study compared the device results against LC/MS, the preferred confirmatory method. The results are presented in tables showing agreement across different concentration ranges (Negative, Low Negative, Near Cutoff Negative, Near Cutoff Positive, High Positive).
Example for FYL (Fentanyl) - Panel Dip, Operator 1:
- Negative (LC/MS 0): Device: 0 Positive, 1 Negative (1 discordant result here, sample 1484, LC/MS 0.78 ng/mL, Device: +)
- Low Negative (LC/MS +50%): Device: 20 Positive, 0 Negative
Lay-User Study (Selected data for AMP, FYL, NFYL):
This study evaluates the percentage of correct results when used by lay persons at various concentrations relative to the cutoff.
Example for AMP (Amphetamine):
- Negative (100% below cutoff): 100% correct (0 positive, 20 negative)
- Low Negative (-75% to -25% Cutoff): 100% correct negative for -75% and -50%, but 0 positive/20 negative for -25% cutoff.
- Positive (+25% to +75% Cutoff): Generally high correctness (95%-100%). For +25% cutoff, 95% correctness (19 positive, 1 negative).
Drug (Identifier) | Cut-off Level | Reported Device Performance (Summary) |
---|---|---|
Fentanyl (FYL) | 1 ng/mL | Precision: At -100% to -50% cutoff, 100% negative calls (50-/0+ over 3 lots for Panel Dip & Quick Cup). At +25% to +100% cutoff, 100% positive calls (50+/0- over 3 lots for Panel Dip & Quick Cup). At cutoff, performance is mixed (e.g., Panel Dip Lot 1: 28+/22-). |
Method Comparison: High concordance with LC/MS, especially for samples well above or below cutoff. Some discordant results near cutoff for both negative (e.g., sample 1484, LC/MS 0.78 ng/mL, device +) and positive (e.g., sample 5419, LC/MS 1.05 ng/mL, device -) as expected for qualitative tests. | ||
Lay-User Study: All 20 negative samples at -100%, -75%, -50% cutoff were correctly identified as negative (100% correct). At -25% cutoff, 95% correct (19 negative, 1 positive). All 20 positive samples at +50%, +75% cutoff were correctly identified as positive (100% correct). At +25% cutoff, 100% correct (20 positive). | ||
Norfentanyl (NFYL) | 5 ng/mL | Precision: At -100% to -50% cutoff, 100% negative calls (50-/0+ over 3 lots for Panel Dip & Quick Cup). At +25% to +100% cutoff, 100% positive calls (50+/0- over 3 lots for Panel Dip & Quick Cup). At cutoff, performance is mixed (e.g., Panel Dip Lot 1: 27+/23-). |
Method Comparison: High concordance with LC/MS, with some discordance near cutoff (e.g., sample 4074, LC/MS 4.39 ng/mL, device +; sample 0687, LC/MS 5.05 ng/mL, device -). | ||
Lay-User Study: All 20 negative samples at -100%, -75%, -50% cutoff were correctly identified as negative (100% correct). At -25% cutoff, 95% correct (19 negative, 1 positive). All 20 positive samples at +25%, +50%, +75% cutoff were correctly identified as positive (100% correct). | ||
Amphetamine (AMP) | 500 ng/mL | Lay-User Study: All 20 negative samples at -100%, -75%, -50%, -25% cutoff were correctly identified as negative (100% correct). All 20 positive samples at +50%, +75% cutoff were correctly identified as positive (100% correct). At +25% cutoff, 95% correctness (19 positive, 1 negative). |
Detailed Study Information:
-
Sample sizes used for the test set and the data provenance:
- Precision Study: For Fentanyl and Norfentanyl, the reported data is for 3 lots, with 2 runs per day for 25 days, for 9 concentrations (e.g., -100% cutoff, -75% cutoff, etc.). This implies 50 tests per concentration per lot (2 runs * 25 days), leading to 450 tests per drug type per lot (9 concentrations * 50 tests), and 1350 tests per drug type across all 3 lots. The data provenance implies these samples were prepared by spiking known concentrations of drug into negative samples, indicating a controlled laboratory environment. Data for other analytes was "reported in K181768" (the predicate device documentation), so the exact sample sizes are not explicitly stated in this document but are assumed to be similar.
- Method Comparison Study: For Fentanyl and Norfentanyl, 80 unaltered clinical samples (40 negative and 40 positive based on LC/MS results) were used per drug. Each sample was tested by three laboratory assistants for each device type (Panel Dip and Quick Cup). This means 80 samples * 3 operators = 240 tests per drug for each device type. The data provenance is "in-house" and "unaltered clinical samples." The document does not specify the country of origin, but given the FDA submission, it's likely US-based or compliant with US standards. The study appears to be retrospective, using already collected clinical samples for comparison.
- Lay-user Study: 280 lay persons were used for each device format (Panel Dip and Quick Cup, though the results summarized apply to the overall device type). Urine samples were prepared at 7 different concentrations (negative, +/-25%, +/-50%, +/-75%, +/-100% of cutoff). Each participant received one blind-labeled sample and one device. Assuming each person tested one sample, this implies 280 samples were tested for each specific drug evaluated by lay-users on each format. The data provenance: "samples were prepared by spiking drugs into drug-free pooled urine specimens" and confirmed by LC/MS. This is a controlled experimental set-up rather than real-world patient samples.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Precision Study: Ground truth was established by spiking known concentrations of drugs into negative samples and confirmed by LC/MS. No human experts were involved in establishing the ground truth directly for this part.
- Method Comparison Study: The ground truth was established by LC/MS (Liquid Chromatography-Mass Spectrometry), which is explicitly stated as the "preferred confirmatory method" and is considered a gold standard for drug detection and quantification in urine. No human expert readers established the ground truth; it was a laboratory instrument measurement. The study used three laboratory assistants to read the device results, but they were comparing their readings against the LC/MS truth, not establishing the truth themselves. Their qualifications are not specified beyond "laboratory assistants."
- Lay-user Study: The ground truth was established by spiking known concentrations of drugs confirmed by LC/MS. No human experts established the ground truth of the samples. The study assessed the lay-users' ability to interpret the device results against this known truth.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Precision Study: No adjudication method mentioned as samples were prepared with known concentrations.
- Method Comparison Study: No adjudication method was explicitly mentioned for the device results. Each of the three operators performed their own reads, and their individual results were compared to the LC/MS. Discordant results are noted for each operator.
- Lay-user Study: No adjudication method was mentioned. Each lay user tested one sample against a pre-defined truth.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted in this report. This device is a rapid diagnostic test (lateral flow immunoassay), not an AI-assisted diagnostic tool for interpretation of medical images or other complex data. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable here.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- This is a lateral flow immunoassay, which is a physical diagnostic device producing a visual result (colored lines). It does not involve an "algorithm" in the sense of a software-based AI or computational algorithm. The device itself is the "standalone" diagnostic. Its performance characteristics (precision, specificity, interference) are essentially its "algorithm only" performance. The method comparison study is akin to assessing the device's standalone performance against a gold standard. The lay-user study assesses human interpretation of the device's standalone output.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The primary ground truth for the analytical and method comparison studies was Liquid Chromatography-Mass Spectrometry (LC/MS), which is a highly accurate chemical method for detecting and quantifying substances.
- For the precision and lay-user studies, the ground truth was based on spiked urine samples with known drug concentrations, which were then confirmed by LC/MS.
-
The sample size for the training set:
- This document describes a 510(k) premarket notification for an in vitro diagnostic device (lateral flow immunoassay). Unlike AI/ML-driven devices that require extensive training data, such chemical-based devices are developed and optimized through chemical engineering and biological principles, not by "training" on datasets in the AI sense. Therefore, the concept of a "training set" with a statistical sample size as understood in machine learning is not applicable to this type of device. The development process involves chemical formulation and validation, not data training.
-
How the ground truth for the training set was established:
- Since the concept of a "training set" as it pertains to AI/ML devices is not applicable, the establishment of ground truth for a training set is also not relevant in this context. The "ground truth" for the performance evaluation of the device relied on LC/MS results and carefully prepared spiked samples with known drug concentrations.
Ask a specific question about this device
(49 days)
| 862.3250
The Healgen® Accurate Multi-Drug Urine Screening Cup is a rapid lateral flow immunoassay for the qualitative detection of 6-Monoacetylmorphine, d-Amphetamine, Benzoylecgonine, Buprenorphine, EDDP, Norfentanyl, Methadone, d-Methamphetamine, d/l-Methylenedioxymethamine, Mortriptyline, Oxazepam, Oxycodone, Phencyclidine, d-Propoxyphene, Secobarbital, THC-COOH and Tramadol in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:
Test | Calibrator | Cut-off (ng/mL) |
---|---|---|
6-MAM | 6-Monoacetylmorphine | 10 |
AMP | d-Amphetamine | 500 or 1000 |
BAR | Secobarbital | 300 |
BUP | Buprenorphine | 10 |
BZO | Oxazepam | 300 |
COC | Benzoylecgonine | 150 or 300 |
EDDP | 2-ethylidene-1,5-dimethyl-3,3- | |
diphenylpyrolidine | 300 | |
FEN or FYL | Norfentanyl | 5 |
MDMA | Methylenedioxymethamphetamine | 500 |
MET | d-Methamphetamine | 500 or 1000 |
MTD | Methadone | 300 |
OPI | Morphine | 300 or 2000 |
OXY | Oxycodone | 100 |
PCP | Phencyclidine | 25 |
PPX | d-Propoxyphene | 300 |
TCA | Nortriptyline | 1000 |
THC | 11-nor-Δ9-THC-COOH | 50 |
TRA or TMI | Tramadol | 100 |
The single or multi-test cups can consist of up to eighteen (18) of the above listed analytes in any combination with or without on-board adulteration/specimen validity tests (SVT).
The tests provide only a preliminary result. A more specific alternative chemical must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
The Healgen® Accurate Home Multi-Drug Urine Test Cup is a rapid qualitative immunoassay. The device provides preliminary results for the detection of one or more of the following drugs
CODE | SUBSTANCE | Cut-off (ng/mL) |
---|---|---|
AMP | Amphetamine | 1000 or 500 |
BUP | Buprenorphine | 10 |
BAR | Secobarbital | 300 |
BZO | Oxazepam | 300 |
COC | Cocaine | 300 or 150 |
EDDP | EDDP | 300 |
FYL | Norfentanyl | 5 |
MET/mAMP | Methamphetamine | 1000 or 500 |
MDMA | Ecstasy | 500 |
OPI | Morphine | 2000 or 300 |
MTD | Methadone | 300 |
OXY | Oxycodone | 100 |
PCP | Phencyclidine | 25 |
PPX | Propoxyphene | 300 |
TCA | Nortriptyline | 1000 |
THC | Marijuana | 50 |
TML | Tramadol | 100 |
6-MAM | 6-Monoacetylmorphine | 10 |
This drug test cup may contain any combination of the drug tests listed in the table above.
This test provides only preliminary result. A more specific alternative chemical must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
Healgen® Accurate Home Muti-Drug Urine Test Cup and Healgen® Accurate Muti-Drug Urine Drug Screen Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine.
The device is a cup format. Each test device is sealed with two sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format and no longer requires assembly before use.
Acceptance Criteria and Device Performance for Healgen® Accurate Multi-Drug Urine Drug Screen Cup
This document outlines the acceptance criteria and the evidence provided to demonstrate that the Healgen® Accurate Multi-Drug Urine Drug Screen Cup meets these criteria. The information is extracted from the provided FDA 510(k) summary (K243365).
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for this qualitative immunoassay are primarily demonstrated through a combination of precision, analytical specificity/interference, and a method comparison study. The precision study evaluates the device's ability to consistently produce correct results at various concentrations relative to the cutoff, while the method comparison study assesses its agreement with a gold standard (LC-MS/MS) using clinical samples. The layperson study demonstrates ease of use for over-the-counter applications.
Acceptance Criteria Category | Specific Criteria | Reported Device Performance |
---|---|---|
Precision/Reproducibility | Consistent qualitative results (+/- at least 95% at the cutoff concentration, and 100% agreement for samples far from the cutoff) across multiple lots and runs, at various concentrations relative to the cutoff (-100%, -75%, -50%, -25%, Cutoff, +25%, +50%, +75%, +100%). | For all drugs tested (AMP 1000/500, BAR 300, BUP 10, BZO 300, COC 300/150, EDDP 300, FYL 5, MDMA 500, MET 1000/500, OPI 300/2000, OXY 100, PCP 25, PPX 300, TCA 1000, THC 50, TML 100, 6-MAM 10) across 3 lots: |
- 100% agreement for concentrations at -100%, -75%, -50%, +50%, +75%, and +100% cutoff (e.g., 0-/50+ or 50-/0+ indicating no false negatives or positives respectively).
- At the cutoff concentration, results consistently showed a mix of positive and negative interpretations, typically near 50%/50% (e.g., 23-/27+, 25-/25+), demonstrating sensitivity around the cutoff.
- At -25% cutoff, the device predominantly showed negative results (e.g., 50-/0+ or 49-/1+), with a few false positives.
- At +25% cutoff, the device predominantly showed positive results (e.g., 0-/50+ or 1-/49+), with a few false negatives.
This indicates good precision around the cutoff with expected variability at borderline concentrations. |
| Analytical Specificity | No significant cross-reactivity with commonly encountered substances (drug metabolites, prescription medications, endogenous compounds) at relevant concentrations, and no interference from variations in pH and specific gravity within a physiological range. | Diverse lists of compounds were tested for cross-reactivity. The results show that: - Many cross-reacting compounds are identified with their minimum concentration needed to yield a positive result and their % Cross-Reactivity (e.g., Hydroxyamphetamine with AMP 1000, Amobarbital with BAR 300).
- Numerous compounds showed no detection or were negative at high concentrations (e.g., 100,000 ng/mL), indicating no cross-reactivity for those substances.
- pH levels of 4 to 9 and specific gravity levels of 1.000 to 1.035 were demonstrated to not affect the assay results. |
| Method Comparison | A high degree of concordance between the device's qualitative results and confirmatory Gas Chromatography-Mass Spectrometry (GC-MS) or Liquid Chromatography-Mass Spectrometry (LC-MS) results for clinical urine samples, particularly for samples at or near the cutoff. | For each drug and its cutoff configuration: - 100% agreement was observed for drug-free samples and "low negative" samples (less than -50% of cutoff).
- 100% agreement was primarily observed for "high positive" samples (greater than +50% of cutoff).
- For "near cutoff negative" (between -50% and cutoff) and "near cutoff positive" (between cutoff and +50%) samples, there was expected variability, with some samples interpreted differently by the device compared to LC-MS/MS, indicating the device's expected performance around the cutoff. The discordant results table details specific samples where device results differed from LC-MS/MS. Generally, the majority of samples near the cutoff were correctly identified by the device. |
| Lay Person Usability | The device should be easy to understand and use by laypersons as indicated by a high agreement in understanding instructions and performance on prepared samples across different drug classes. A reading level analysis of the instructions for use should demonstrate appropriate readability. | A study with 280 laypersons (mix of male/female, aged 21 to >50, diverse backgrounds) demonstrated: - 100% agreement for interpretation of drug-free and concentrations at -100%, -75%, -50% below cutoff.
- 100% agreement for concentrations at +50% and +75% above cutoff.
- Expected variability in agreement around the -25% and +25% cutoff concentrations, mostly ranging from 90% to 95%.
- All participants indicated the instructions were easy to understand and follow.
- The Flesch-Kincaid reading analysis yielded a Grade Level of 7, which is suitable for layperson use. |
| Stability | The device should maintain its performance characteristics for a specified duration under defined storage conditions. | The device is stable at 2-30°C for 36 months based on real-time stability studies. |
2. Sample Size Used for the Test Set and Data Provenance
Precision/Reproducibility Study:
- Sample Size: For each drug and each concentration level (total 9 levels per drug), 50 individual tests were performed (2 runs per day for 25 days). Given there are 17 distinct drug analytes, this amounts to 17 drugs * 9 concentrations * 50 tests/concentration = 7,650 tests.
- Data Provenance: Samples were prepared by spiking target drugs into drug-free urine samples. These samples were likely prepared in a laboratory setting, making the data provenance prospective and controlled. The document does not specify a country of origin for the samples/urine, but the study was conducted "in-house."
Method Comparison Study:
- Sample Size: For each drug, 80 unaltered urine clinical samples were used. This consisted of 40 negative and 40 positive samples. Given there are 17 distinct drug analytes, this amounts to 17 drugs * 80 samples/drug = 1,360 total clinical samples tested across all drugs.
- Data Provenance: Unaltered clinical urine samples. The document does not specify the country of origin but states the study was performed "in-house." The nature of "unaltered clinical samples" suggests these were collected from patients, and their retrospective or prospective nature is not explicitly stated, though being "blind labeled" for comparison suggests they were handled specifically for this study.
Lay Person Study:
- Sample Size: 280 lay persons participated. Urine samples were prepared at 7 concentration levels (-100%, +/-75%, +/-50%, +/-25% of cutoff). Each participant tested 1 blind labeled sample. While it states urine samples were prepared at 7 concentrations, the agreement tables show 7 concentrations and 20 tests per concentration for each drug. If we assume each of the 280 participants tested one sample for one drug, the sample sizes per concentration shown in the table (20 tests) indicate that multiple participants contributed data for each drug and concentration, but a single participant did not test all drugs or all concentrations.
- Data Provenance: Urine samples were prepared by spiking drugs into drug-free pooled urine specimens. The samples were prepared in a laboratory environment, indicating prospective and controlled data provenance. The lay persons were recruited for the study, making their involvement prospective. The study was performed at "three intended user sites."
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Precision/Reproducibility Study:
- Ground Truth: The "ground truth" for the spiked samples was established by the known concentrations of the spiked drugs, which were confirmed by LC-MS/MS.
- Experts/Qualifications: The document states that "Each drug concentration was confirmed by LC-MS/MS." LC-MS/MS is a highly accurate analytical method, and while specific "experts" (e.g., toxicologists, clinical chemists) are not explicitly named, the use of this method implies analysis by qualified laboratory personnel specializing in mass spectrometry. "One operator per lot" (for 3 lots) performed the tests on the device, but the LC-MS/MS confirmation implies separate, expert analysis for ground truth.
Method Comparison Study:
- Ground Truth: The ground truth for the clinical samples was established by LC-MS/MS results.
- Experts/Qualifications: Similar to the precision study, the ground truth was determined by LC-MS/MS, implying analysis by qualified laboratory professionals. No specific number or qualification of "experts" is provided beyond the analytical method itself.
Lay Person Study:
- Ground Truth: The "ground truth" for the spiked samples was established by the known concentrations of the spiked drugs, which were confirmed by LC-MS/MS.
- Experts/Qualifications: As above, the ground truth was derived from LC-MS/MS analysis, indicating reliance on qualified laboratory personnel.
4. Adjudication Method for the Test Set
Precision/Reproducibility Study:
- Adjudication Method: Not applicable in the traditional sense of expert consensus. The ground truth was based on the quantitative LC-MS/MS results of the spiked samples. The device's qualitative results were compared against these known concentrations relative to the cutoff. Discrepancies were noted directly.
Method Comparison Study:
- Adjudication Method: Not applicable/Not explicitly stated as an adjudication process involving multiple human readers. The device results were directly compared to the LC-MS/MS results, which served as the definitive "true" value (ground truth). The tables show the device's output (positive/negative) versus the LC-MS/MS quantification, which then categorizes samples as drug-free, low negative, near cutoff negative, near cutoff positive, or high positive.
Lay Person Study:
- Adjudication Method: Not applicable in terms of expert consensus. The laypersons' interpretations of the device results were compared against the established ground truth (LC-MS/MS confirmed concentrations of spiked samples). The study aimed to assess if laypersons could correctly interpret the device results against this objective ground truth, not for them to establish ground truth or adjudicate.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done in the context of comparing human readers with and without AI assistance, as this is an in-vitro diagnostic device (a test cup) not an AI-powered diagnostic imaging or interpretation system. The "operators" in the method comparison study refer to individuals performing the device test, not medical interpreters assisted by AI. The "lay person study" evaluated if laypersons could correctly interpret the device, not an AI system.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This is an in-vitro diagnostic (IVD) device (a lateral flow immunoassay test cup), not an AI algorithm. Therefore, the concept of "standalone (algorithm only)" performance without human-in-the-loop is not applicable in the typical sense for AI/software devices. The device itself provides a visual qualitative result (lines appearing or not appearing), which then requires human observation and interpretation. The performance characteristics described (precision, specificity, method comparison) are inherently "standalone" in that they assess the device's ability to produce correct visual results, which a human then reads. The layperson study specifically evaluates the human (layperson) interpretation of these visual results.
7. The Type of Ground Truth Used
The primary type of ground truth used across all analytical performance studies (Precision, Analytical Specificity, Method Comparison, and Lay Person Study where applicable) is Gas Chromatography-Mass Spectrometry (GC-MS) and Liquid Chromatography-Mass Spectrometry (LC-MS) and their tandem mass-spectrometer versions (LC-MS/MS).
These are considered confirmatory analytical methods and are widely accepted as the gold standard for quantitative drug detection in toxicology.
8. The Sample Size for the Training Set
As this device is a lateral flow immunoassay (a chemical/biological test kit), and not an AI/machine learning algorithm, there is no "training set" in the computational sense. The device's performance is based on its inherent physical and chemical properties and reagent formulations, which are developed and optimized through laboratory research and development, rather than machine learning training.
9. How the Ground Truth for the Training Set Was Established
Since there is no "training set" for this type of device, this question is not applicable. The development and optimization of the immunoassay undoubtedly involved extensive experimentation with samples of known drug concentrations, likely confirmed by GC-MS/LC-MS, but this is part of the product's research and development process, not analogous to establishing ground truth for a machine learning training set.
Ask a specific question about this device
(40 days)
Benzoylecognine (COC) | NFY | Class II | 21 CFR § 862.3250
Pocguide Multi-Drug Test Panel OTC is competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5dimethyl-3,3-diphenylpyrrolidine, Methylenedioxy-methamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline and Marijuana in human urine at the cutoff concentrations of:
Drug ( Identifier) | Cut-off level |
---|---|
Amphetamine (AMP) | 1000 ng/mL or 500 ng/mL |
Buprenorphine (BUP) | 10 ng/mL |
Secobarbital (BAR) | 300 ng/mL |
Oxazepam (BZO) | 300 ng/mL |
Benzoylecognine (COC) | 300 ng/mL or 150 ng/mL |
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) | 300 ng/mL |
Methamphetamine (MET) | 1000 ng/mL or 500 ng/mL |
Methylenedioxymethamphetamine (MDMA) | 500 ng/mL |
Morphine (OPI2000/MOP300) | 2000 ng/mL or 300 ng/mL |
Methadone (MTD) | 300 ng/mL |
Oxycodone (OXY) | 100 ng/mL |
Phencyclidine (PCP) | 25 ng/mL |
Nortriptyline (TCA) | 1000 ng/mL |
Marijuana (THC) | 50 ng/mL |
The single or multi-test panels can consist of up to the above listed analytes in any combination. The tests provide only a preliminary result. A more specific alternative chemical must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
For over-the-counter use. For in vitro diagnostic use only
Pocguide™ Multi-Drug Test Panel and Pocguide™ Multi-Drug Test Panel OTC are immunochromatographic assays that use a lateral flow system for the qualitative detection of single drugs in human urine at or above the cut-off levels as indicated. The products are single use in vitro diagnostic devices.
This device is a dipcard format in which the test strips are integrated into the plastic dipcard. After removing the cap of the dipcard, the absorbent end of the test strips is exposed and can be in direct contact with the urine sample. The device is in a ready-to-use format and no longer requires assembly before use.
The provided document describes the Pocguide Multi-Drug Test Panel and Pocguide Multi-Drug Test Panel OTC, which are in vitro diagnostic devices for qualitative and simultaneous detection of various drugs in human urine.
Here's an analysis of the acceptance criteria and the study proving the device meets those criteria, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for this type of qualitative diagnostic device are typically related to its analytical performance, specifically precision (reproducibility) around the cutoff concentration, and its ability to correctly identify positive and negative samples when compared to a confirmed method (method comparison).
Acceptance Criteria (Implied based on study design and regulatory context for qualitative drug tests):
- Precision/Reproducibility: A high percentage of agreement (e.g., typically >80%) for samples near the cutoff (e.g., +/- 25% cutoff, cutoff itself) over multiple lots and runs, and 100% agreement for samples far from the cutoff (e.g., +/- 100% cutoff).
- Method Comparison: High overall agreement (sensitivity and specificity) with a gold standard confirmatory method (LC/MS or GC/MS) for clinical samples, especially for true positive and true negative samples. Acceptable performance for samples near the cutoff where some discordance is expected due to the nature of qualitative assays.
- Analytical Specificity (Cross-Reactivity): No significant cross-reactivity with common substances or structurally similar compounds to avoid false positives.
- Interference: No interference from common physiological substances found in urine.
- Lay-User Study (for OTC devices): High agreement with trained professionals and ease of use for the intended lay user.
Reported Device Performance (from "Precision/Reproducibility" and "Method Comparison" sections):
Test Characteristic | Drug (Cutoff) Examples | Reported Device Performance |
---|---|---|
Precision/Reproducibility | AMP 500 ng/mL, BUP 10 ng/mL, etc. | +100%, +75%, +50%, +25% Cutoff: 100.0% Positive (Across all tested drugs and cutoffs) |
-100%, -75%, -50%, -25% Cutoff: 100.0% Negative (Across all tested drugs and cutoffs) | ||
Cutoff: |
- AMP 500: 82.0% Positive, 18.0% Negative
- BUP 10: 84.0% Positive, 16.0% Negative
- BAR 300: 82.7% Positive, 17.3% Negative
- And similar ranges for other drugs listed in Table 2. Each drug showed similar performance around the cutoff. |
| Method Comparison | AMP 1000 ng/mL, AMP 500 ng/mL, BAR 300 ng/mL, etc. | Excellent agreement for Drug-Free, Low Negative, and High Positive Samples: Typically 100% correct classification by the device for these categories (e.g., "Viewer A Positive" for High Positive by LC/MS and "Viewer A Negative" for Drug-Free by LC/MS are 100% for almost all drugs).
Expected Discordance Near Cutoff: As anticipated for qualitative tests, some samples near the cutoff (especially -25% and +25%) show mixed results (discordance) between the device and LC/MS, as detailed in Table 5 and Table 6 (Discordant results). These are typically within acceptable ranges for qualitative tests, acknowledging the inherent variation around a precise cutoff. |
| Analytical Specificity | AMP, BUP, BAR, etc. | Tested numerous substances. Most showed no cross-reactivity or very low percentages at very high concentrations, indicating good specificity. Specific cross-reactivity percentages are provided in Table 3. |
| Interference | N/A - broadly tested | No interference observed for a wide range of common substances and physiological conditions (urine specific gravity 1.000-1.035, pH 4-9) as listed in Tables 4. |
| Lay-User Study | All Configuration 1 & 2 Drugs | Agreement (%): - -100%, -75%, -50% Cutoff: 100% negative calls.
- +25%, +50%, +75% Cutoff: Mostly 100% positive calls, some 95% for +25% cutoff.
- -25% Cutoff: 95% negative calls for most drugs.
- Raw numbers show 19/20 or 20/20 correct calls for most categories.
Ease of Use: All participants indicated instructions were easy to understand and follow (Flesch-Kincaid Grade Level 7). |
2. Sample Sizes and Data Provenance
- Test Set Sample Sizes:
- Precision/Reproducibility: For each drug and each cutoff, 50 samples were tested at each concentration level (-100%, -75%, -50%, -25%, Cutoff, +25%, +50%, +75%, +100%). This was done across 3 lots, so $50 \text{ samples/level} \times 9 \text{ levels} \times 3 \text{ lots} = 1350$ tests per drug. (For AMP alone, this would be $1350 \times 2 \text{ cutoffs} = 2700$ tests).
- Method Comparison: 100 unaltered clinical samples were used for each target drug (40 negative, 40 positive, and an additional 20 samples around the cutoff as seen in the breakdown of results). So, for 13 drug analytes, this would be $13 \times 100 = 1300$ clinical samples.
- Cross-Reactivity / Interference: Specific numbers for each substance are not given, but samples were spiked at various concentrations and tested using three lots of each device.
- Lay-User Study: For Configuration 1, 140 participants (58 male, 82 female). For Configuration 2, 140 participants (56 male, 84 female). Each participant tested 1 blind-labeled sample. For each drug within each configuration, 20 samples were prepared per concentration level (-100%, -75%, -50%, -25%, +25%, +50%, +75%).
- Data Provenance: The document does not explicitly state the country of origin for the data. The consulting firm is in Shanghai, China, and the applicant's address is Irvine, CA, USA. Given the FDA 510(k) submission, it's implied that the data is intended to be representative and valid for the US market. The studies are described as retrospective as they involve samples prepared at specific concentrations or existing clinical samples compared to a gold standard.
3. Number of Experts and their Qualifications
- For Precision/Reproducibility, Cross-Reactivity, Interference, and Method Comparison: The document does not explicitly state the number of "experts" used to establish ground truth or interpret results. These are quantitative/analytical laboratory tests where the ground truth (concentration by LC/MS or GC/MS) is established by analytical instrumentation. The "Viewers" (A, B, C) mentioned in the Method Comparison section appear to be individuals performing the visual interpretation of the device results, not necessarily independent experts establishing ground truth. Their qualifications are not specified but are implied to be trained laboratory personnel.
- For Lay-User Study: No "experts" were used to establish ground truth for the lay-user study. The ground truth for the samples used in this study was established by LC-MS/MS confirming the spiked drug concentrations.
4. Adjudication Method for the Test Set
- For Precision/Reproducibility, Cross-Reactivity, Interference: No adjudication method is described. The results are reported as counts of positive/negative readings against a known (spiked) concentration.
- For Method Comparison: No explicit "adjudication" among multiple readers is described. Results for the candidate device were observed by "Viewer A, B, C." The comparison is directly between the "Candidate Device Result" (presumably individual Viewer results, though aggregated in Table 5) and the LC/MS reference method. The discordant results in Table 6 specify which viewers made the discordant call (e.g., "Viewer A, B," "Viewer C"). This suggests independent readings by three viewers, but no formal adjudication process to resolve disagreements among them is mentioned; the individual viewer calls are presented relative to the ground truth.
- For Lay-User Study: No adjudication method is mentioned. The tables report the number of positive/negative results per concentration.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No MRMC comparative effectiveness study was done. The study design is an analytical performance study and a comparison study against a laboratory reference method, along with a lay-user study for OTC claims. It assesses device performance in a standalone or simulated user setting, not direct human reader improvement with AI assistance. The device is a lateral flow immunoassay, not an AI-powered diagnostic.
6. Standalone (Algorithm Only) Performance
- This question is not applicable as the device is a lateral flow immunochromatographic assay, not an algorithm or software-based diagnostic. Its performance is inherent to the chemical reactions on the test strip and visual interpretation, not an algorithm.
7. Type of Ground Truth Used
- The primary ground truth used for performance evaluation (Precision/Reproducibility, Method Comparison, Lay-User Study) is analytical confirmation by Gas Chromatography-Mass Spectrometry (GC-MS) or Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions (LC-MS/MS), which are stated as the "preferred confirmatory methods." This is a highly accurate and quantitative method for determining drug concentrations.
8. Sample Size for the Training Set
- This question is not applicable. The device is a qualitative diagnostic test based on immunoassay principles, not a machine learning or AI-based device that requires a "training set" in the computational sense. The development of the immunoassay itself relies on antigen-antibody binding characteristics and optimization, not a deep learning model.
9. How the Ground Truth for the Training Set Was Established
- This question is not applicable for the same reasons as #8. The "ground truth" for developing the test and optimizing its performance would be established through standard immunoassay R&D processes, involving controlled experiments with known concentrations of analytes and cross-reactants, guided by established analytical chemistry principles.
Ask a specific question about this device
Page 1 of 19