K Number
K240287

Validate with FDA (Live)

Date Cleared
2025-03-18

(411 days)

Product Code
Regulation Number
862.3610
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LYHER® Oral fluid Multi-Drug Test Kit (Cube) is a rapid lateral flow immunoassay for the qualitative detection of d-Amphetamine, d-Methamphetamine, Benzoylecgonine, Morphine, Phencyclidine and Delta-9-Tetrahydrocanabinol in human oral fluid. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

TestCalibratorCut-off (ng/mL)
Opiates(OPI)Morphine40
Cocaine (COC)Benzoylecgonine20
Amphetamine (AMP)d-Amphetamine50
Marijuana (THC)Delta-9-Tetrahydrocannabinol40
Methamphetamine (MET)d-Methamphetamine50
Phencyclidine (PCP)Phencyclidine10

The single or multi-test panels can consist of the above insted analytes in anycombination, up to a maximum of 6 analytes. The tests provide only a preliminary result. A more specific alternate chemical must be used in order to obtain a confirmed analytical test result. Gas Chromatography/Mass Spectrometry (GC/MS), Liquid Chromatography / Mass Spectrometry (LC/MS) and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.

Device Description

The LYHER® Oral fluid Multi-Drug Test Kit (Cube) is an immunochromatographic assay that uses a lateral flow system for the qualitative detection of d-Amphetamine, d-Methamphetamine, Benzoylecgonine, Morphine, Phencyclidine and Delta-9-Tetrahydrocannabinol in human oral fluid. The LYHER® Oral fluid Multi-Drug Test Kit (Cube) device consists of a cube device, an oral fluid collection swab and a package insert.

AI/ML Overview

The document describes the analytical performance studies for the LYHER® Oral fluid Multi-Drug Test Kit (Cube), a rapid lateral flow immunoassay. The device is designed to detect d-Amphetamine, d-Methamphetamine, Benzoylecgonine, Morphine, Phencyclidine, and Delta-9-Tetrahydrocannabinol in human oral fluid.

Here's a breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state "acceptance criteria" as a pass/fail threshold for performance, but rather presents the results of precision-reproducibility studies across various concentrations relative to the cut-off. The performance is summarized by the number of positive and negative results at each concentration.

To construct a table of "acceptance criteria" (inferred as target performance, though not explicitly defined as such) and reported performance, we can focus on the precision-reproducibility studies around the cut-off and the comparison study data. For the comparison study, the implicit acceptance criterion is that the device accurately identifies samples as positive or negative relative to the LC/MS confirmation.

Inferred Acceptance Criteria (Conceptual) and Reported Device Performance

Parameter / Concentration LevelExpected Outcome (Inferred)Reported Performance (Counts from all operators and lots, combined)
D-Amphetamine
-100% cut off (Negative)All Negative0+/180- (All Negative)
-75% cut off (Negative)All Negative0+/180- (All Negative)
-50% cut off (Negative)All Negative0+/180- (All Negative)
-25% cut off (Negative)Majority Negative, some PositiveApprox. 8+/532- (Some false positives)
Cut off (Threshold)Mix of Positive/Negative (close to 50/50 split implies good performance at cutoff)Approx. 142+/348- (for combined precision study across drugs and operators), For the comparison study, at/around cutoff, you see 47-49+/11-13- for precision (ideal is balanced +/-), and 1+/74- for comparison at -50 cutoff for Amphetamines, and 180+/9- for comparison at cutoff for Amphetamines
+25% cut off (Positive)Majority Positive, some NegativeApprox. 165+/15- (Some false negatives)
+50% cut off (Positive)All Positive60+/0- (All Positive)
+75% cut off (Positive)All Positive60+/0- (All Positive)
+100% cut off (Positive)All Positive60+/0- (All Positive)
Cocaine
-100% cut off (Negative)All Negative0+/180- (All Negative)
-75% cut off (Negative)All Negative0+/180- (All Negative)
-50% cut off (Negative)All Negative0+/180- (All Negative)
-25% cut off (Negative)Majority Negative, some Positive~17+/523-
Cut off (Threshold)Mix of Positive/Negative~145+/245-
+25% cut off (Positive)Majority Positive, some Negative~168+/12-
+50% cut off (Positive)All Positive60+/0- (All Positive)
+75% cut off (Positive)All Positive60+/0- (All Positive)
+100% cut off (Positive)All Positive60+/0- (All Positive)
All other Drugs (similarly to above)
At or below -50% Cut-offGenerally all NegativeVery high negative rate (e.g., 0+/60- for most negative concentrations in precision study)
At or above +50% Cut-offGenerally all PositiveVery high positive rate (e.g., 60+/0- for most positive concentrations in precision study)
Near Cut-off (e.g. +/-25%)Mixed results, reflecting the nature of a qualitative test around cutoffVariable, as expected (e.g., for Amphetamine at -25% cut-off, ~2+/58- across operators/lots for precision; 47-49+/11-13- at cut-off)
Comparison Study vs. LC/MS:
True Negative (Analyte absent)Device NegativeE.g., Amphetamine: 360 Negative (device) out of 360 Negative (LC/MS)
False Positive (Analyte negative, Device positive)Very low / zeroE.g., Amphetamine: 0 Positive (device) where LC/MS was Negative
True Positive (Analyte positive, Device positive)Device PositiveE.g., Amphetamine: 180 Positive (device) at Cut off to +50% cut off vs. 180 identified by LC/MS
False Negative (Analyte positive, Device negative)Very low / zeroE.g., Amphetamine: 9 Negative (device) where LC/MS was Cut off to +50% cut off (indicating 9 False Negatives in that range)

Note on Acceptance Criteria: The document provides raw performance data. For a qualitative immunoassay, the "acceptance criteria" are usually demonstrated by a high rate of negativity well below the cutoff, a high rate of positivity well above the cutoff, and a predictable transition zone around the cutoff concentration. The data presented supports this.

2. Sample size used for the test set and the data provenance

  • Test Set Sample Size:

    • Precision-Reproducibility-Cut-Off Study: For each drug, 9 concentration levels were tested. For each concentration, there were 2 runs per day for 30 days, across 3 device lots, and by 3 operators.
      • This means 9 concentrations * 2 runs/day * 30 days * 3 lots * 3 operators = 4860 total tests per drug analyte for the precision study.
      • Each concentration within each operator/lot combination was tested 60 times (2 runs/day * 30 days). So for each specific concentration, lot, and operator, n=60.
    • Comparison Studies (Method Comparison):
      • Negative oral fluid: 360 samples (across all operators and sites for each drug).
      • Positive oral fluid at various ranges:
        • <-50% cut off: For D-Amphetamine, 93 samples. For Cocaine, 69 samples. For d-Methamphetamine, 63 samples. For Morphine, 57 samples. For Phencyclidine, 51 samples. For Delta-9-Tetrahydrocannabinol, 93 samples.
        • -50% cut off - cut off: For D-Amphetamine, 74 samples. For Cocaine, 83 samples. For d-Methamphetamine, 70 samples. For Morphine, 84 samples. For Phencyclidine, 109 samples. For Delta-9-Tetrahydrocannabinol, 96 samples.
        • Cut off - +50% cut off: For D-Amphetamine, 180 samples. For Cocaine, 172 samples. For d-Methamphetamine, 164 samples. For Morphine, 183 samples. For Phencyclidine, 170 samples. For Delta-9-Tetrahydrocannabinol, 172 samples.
        • >+50% cut off: For D-Amphetamine, 183 samples. For Cocaine, 186 samples. For d-Methamphetamine, 189 samples. For Morphine, 180 samples. For Phencyclidine, 192 samples. For Delta-9-Tetrahydrocannabinol, 195 samples.
      • The total number of samples for the comparison study for each drug is 360 (negatives) + (sum of the positive ranges for that drug). For example, for D-Amphetamine: 360 + 93 + 74 + 180 + 183 = 890 samples. This applies similarly to other drugs.
  • Data Provenance: The document states "Method comparison studies for the LYHER Oral fluid Multi-Drug Test Kit(Cube) were performed at three testing sites with three operators at each site." The location of these sites is not explicitly mentioned (e.g., country of origin), but the submitter is based in China. The data origin is prospective as samples were prepared by spiking known concentrations into negative oral fluid for analytical studies.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Ground Truth Establishment: The ground truth for the test set for both the precision and comparison studies was established by Laboratory Confirmed Concentrations, specifically using LC/MS (Liquid Chromatography/Mass Spectrometry). For the precision study, it states: "Each drug concentration was confirmed by LC/MS." For the method comparison studies, it states: "Operators tested the samples using the candidate device and the results were compared to LC/MS results."
  • Number of Experts/Qualifications: LC/MS is a laboratory analytical method, not reliant on subjective expert interpretation like radiological imaging. Therefore, there were no "experts" in the sense of human annotators (e.g., radiologists) involved in establishing the ground truth via consensus or adjudication. The "experts" would be the qualified laboratory personnel performing and interpreting the LC/MS results, whose qualifications are implicit given the professional standards for such testing.

4. Adjudication method for the test set

  • Since LC/MS is used to establish quantitative drug concentrations, and the device provides a qualitative "positive" or "negative" result based on a defined cutoff, there is no expert adjudication method (like 2+1, 3+1) mentioned or necessary. The device's result is compared directly to the LC/MS result relative to the established cut-off.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This question is not applicable to this device. This is a qualitative diagnostic test (immunoassay) for drug detection in oral fluid, not an AI-powered image analysis device that assists human readers. The "operators" mentioned are performing the test, not interpreting complex medical images.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • This device is a standalone test (a rapid lateral flow immunoassay), essentially an "algorithm only" in the sense that its chemical reactions produce a visual result (line presence/absence). Human interaction involves collecting the sample, applying it to the device, and visually interpreting the presence or absence of test lines. There isn't a separate "algorithm" that operates outside of the device itself to interpret the results. The performance data presented measures the device's inherent analytical accuracy against known concentrations and LC/MS.

7. The type of ground truth used

  • The ground truth used for both precision and comparison studies was laboratory-confirmed drug concentrations via LC/MS (Liquid Chromatography/Mass Spectrometry). This is a highly accurate and quantitative analytical method, considered a gold standard for drug detection and quantification.

8. The sample size for the training set

  • This device is a lateral flow immunoassay, not a machine learning or AI-based device that requires a "training set" in the computational sense. Its performance is based on the inherent chemical and biological properties of the reagents and test strip design. Therefore, the concept of a "training set" for model development is not applicable. The studies described are analytical validation studies, not AI model training.

9. How the ground truth for the training set was established

  • As explained in point 8, there is no "training set" for an AI model. The ground truth for the analytical validation (which is analogous to testing the device's inherent design performance) was established by LC/MS confirmation of spiked drug concentrations in oral fluid.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 18, 2025

Hangzhou Laihe Biotech Co., Ltd. % Ethan Liu RA Specialist Shanghai Thinkwell Consulting Co., Ltd. Xinling Rd., 211/6F Shanghai. 201100 China

Re: K240287

Trade/Device Name: LYHER® Oral fluid Multi-Drug Test Kit (Cube) Regulation Number: 21 CFR 862.3610 Regulation Name: Methamphetamine Test System Regulatory Class: Class II Product Code: DJC, DIO, DJG, DKZ, LCM, LDJ Dated: January 24, 2025 Received: February 6, 2025

Dear Ethan Liu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportinemdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Marianela Perez-torres -S

Marianela Perez-Torres, Ph.D. Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K240287

Device Name

LYHER® Oral fluid Multi-Drug Test Kit (Cube)

Indications for Use (Describe)

The LYHER® Oral fluid Multi-Drug Test Kit (Cube) is a rapid lateral flow immunoassay for the qualitative detection of d-Amphetamine, d-Methamphetamine, Benzoylecgonine, Morphine, Phencyclidine and Delta-9-Tetrahydrocanabinol in human oral fluid. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

TestCalibratorCut-off (ng/mL)
Opiates(OPI)Morphine40
Cocaine (COC)Benzoylecgonine20
Amphetamine (AMP)d-Amphetamine50
Marijuana (THC)Delta-9-Tetrahydrocannabinol40
Methamphetamine (MET)d-Methamphetamine50
Phencyclidine (PCP)Phencyclidine10

The single or multi-test panels can consist of the above insted analytes in anycombination, up to a maximum of 6 analytes. The tests provide only a preliminary result. A more specific alternate chemical must be used in order to obtain a confirmed analytical test result. Gas Chromatography/Mass Spectrometry (GC/MS), Liquid Chromatography / Mass Spectrometry (LC/MS) and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the

time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary K240287

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

1. Submitter

Submitted by:Hangzhou Laihe Biotech Co., Ltd.
Address: Room 401-406, F1-3, Building 1, No.425Miaohouwang Road, Xixing Street, BinjiangDistrict, Hangzhou, 310051 Zhejiang P.R. ChinaTel.: +86 571 8765 3090Fax: +86 571 8665 8000www.lyher.com
Contact Person:Ethan Liu
Shanghai Thinkwell Consulting Co., LtdAddress:Xinling Rd., 211/6F Shanghai, 201100,P.R.ChinaPhone:0086-15216699240Email:xtdeepwater@126.com
Date Prepared:March 6, 2025

2. Device

LYHER® Oral fluid Multi-Drug Test Kit (Cube)

Classification:

Product CodeCFR #Panel
DJC862.3610/Methamphetaminetest system.Toxicology
DIO862.3250/Cocaine system.Toxicology
DJG862.3650/Opiate test system.Toxicology
DKZ862.3100/Amphetamine TestSystemToxicology
LCMUnclassifiedToxicology
LDJ862.3870/Cannabinoid testsystem.Toxicology

3. Predicate Device:

OralTox™ Oral Fluid Drug Test, K171403

{5}------------------------------------------------

4. Device Description

The LYHER® Oral fluid Multi-Drug Test Kit (Cube) is an immunochromatographic assay that uses a lateral flow system for the qualitative detection of d-Amphetamine, d-Methamphetamine, Benzoylecgonine, Morphine, Phencyclidine and Delta-9-Tetrahydrocannabinol in human oral fluid. The LYHER® Oral fluid Multi-Drug Test Kit (Cube) device consists of a cube device, an oral fluid collection swab and a package insert.

5. Intended Use / Indications for Use:

The LYHER® Oral fluid Multi-Drug Test Kit (Cube) is a rapid lateral flow immunoassay for the qualitative detection of d-Amphetamine, d-Methamphetamine, Benzoylecgonine, Morphine, Phencyclidine and Delta-9-Tetrahydrocannabinol in human oral fluid. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

TestCalibratorCut-off (ng/mL)
Opiates(OPI)Morphine40
Cocaine (COC)Benzoylecgonine20
Amphetamine (AMP)d-Amphetamine50
Marijuana (THC)Delta-9-Tetrahydrocannabinol40
Methamphetamine (MET)d-Methamphetamine50
Phencyclidine (PCP)Phencyclidine10

The single or multi-test panels can consist of the above listed analytes in any combination, up to a maximum of 6 analytes.

The tests provide only a preliminary result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical test result. Gas Chromatography/Mass Spectrometry (GC/MS), Liquid Chromatography/Mass Spectrometry (LC/MS) and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.

6. Substantial Equivalence

A summary comparison of features of the LYHER® Oral fluid Multi-Drug Test Kit (Cube) and the predicate devices is provided in following tables.

{6}------------------------------------------------

ItemProposed DevicePredicateDevice-K171403
Similarities
Indication(s) for UseFor the qualitative determination of drugs of abuse in human oral fluid.Same
MethodologyCompetitive binding, Lateral flow immunochromatographic assay based on principle of antigen antibody immunochemistrySame
Type of TestQualitativeSame
Specimen TypeHuman oral fluidSame
Cut-off ValuesTest Cut-off (ng/mL) OPI 40 COC 20 AMP 50 THC 40 MET 50 PCP 10Same
Intended UseFor prescription use.Same
Differences
ConfigurationsCubeCups

7. Test Principle

Each device employs lateral flow immunochromatographic technology and is based on the principle of competitive binding. Drugs, if present in concentrations below the cutoff level, will not saturate the binding sites of the antibody coated particles on the drug specific test strips. The goat-anti-rabbit IgG antibody-coated particles will then be captured by immobilized drug-specific conjugate. If the level of drug in the oral fluid specimen is below the cutoff concentration, the T line appears as a visible burgundy line. If the level of drug in the oral fluid specimen is above the cutoff, no T line develops. The control line (C line) serves as an internal quality control. The control line should always appear as a burgundy-colored band regardless of the presence of the drug, if enough sample volume has been added to the test and if the sample has correctly migrated up the test strip.

Testing is based on the principle of a competitive immunochemical reaction between a chemically labeled drug (drug-protein conjugate) and the drug or drug metabolites which may be present in the oral fluid sample competing for the limited antibody binding sites.

8. Performance Characteristics

8.1 Analytical Performance

  • a. Precision-Reproducibility-Cut-Off

{7}------------------------------------------------

Precision-Reproducibility-Cut-Off studies were carried out for samples with concentrations of -100% cut off, -75% cut off, -50% cut off, -25% cut off, cut off, +25% cut off, +50% cut off , +75% cut off and +100% cut off. These samples were prepared by spiking drug in negative oral fluid samples. Each drug concentration was confirmed by LC/MS. All sample aliquots were blindly labeled by the person who prepared the samples and didn't take part in the sample testing. For each concentration, tests were performed two runs per day for 30 days per device lot in a randomized order. The data is summarized below.

Operator 1
-100%cut off-75%cut off-50%cut off-25%cut offCut off+25%cut off+50%Cutoff+75%cut off+100%cut off
Lot10+/60-0+/60-0+/60-૨+/૨૨-48+/12-54+/6-60+/0-60+/0-60+/0-
Lot20+/60-0+/60-0+/60-૨+/૨૨-47+/13-ર્ડન/2-60+/0-60+/0-60+/0-
Lot30+/60-0+/60-0+/60-૨+/૨૨-47+/13-ર્સન/ર-60+/0-60+/0-60+/0-
Operator 2
LotDrug-100%cut off-75%cut off-50%cut off-25%cut offCut off+25%cut off+50%Cutoff+75%cut off+100%cut off
Lot10+/60-0+/60-0+/60-૨+/૨૨-47+/13-ર્સન/ર-60+/0-60+/0-60+/0-
Lot20+/60-0+/60-0+/60-૨+/૨૨-48+/12-54+/6-60+/0-60+/0-60+/0-
Lot30+/60-0+/60-0+/60-૨+/૨૨-47+/13-ર્રન/ર-60+/0-60+/0-60+/0-
Operator 3
LotDrug-100%cut off-75%cut off-50%cut off-25%cut offCut off+25%cut off+50%Cutoff+75%cut off+100%cut off
Lot10+/60-0+/60-0+/60-૨+/૨૨-47+/13-56+/4-60+/0-60+/0-60+/0-
Lot20+/60-0+/60-0+/60-૨+/૨૨-47+/13-56+/4-60+/0-60+/0-60+/0-
Lot30+/60-0+/60-0+/60-6+/54-48+/12-54+/6-60+/0-60+/0-60+/0-

Amphetamine

{8}------------------------------------------------

Cocaine
Operator 1
LotDrug-100%cut off-75%cutoff-50%cutoff-25%cutoffCut off+25%cutoff+50%Cutoff+75%cutoff+100%cut off
Lot10+/60-0+/60-0+/60-6+/54-48+/12-56+/4-60+/0-60+/0-60+/0-
Lot20+/60-0+/60-0+/60-5+/55-48+/12-56+/4-60+/0-60+/0-60+/0-
Lot30+/60-0+/60-0+/60-6+/54-49+/11-55+/5-60+/0-60+/0-60+/0-
Operator 2
LotDrug-100%cut off-75%cutoff-50%cutoff-25%cutoffCut off+25%cutoff+50%Cutoff+75%cutoff+100%cut off
Lot10+/60-0+/60-0+/60-6+/54-48+/12-56+/4-60+/0-60+/0-60+/0-
Lot20+/60-0+/60-0+/60-5+/55-48+/12-56+/4-60+/0-60+/0-60+/0-
Lot30+/60-0+/60-0+/60-5+/55-49+/11-56+/4-60+/0-60+/0-60+/0-
Operator 3
LotDrug-100%cut off-75%cutoff-50%cutoff-25%cutoffCut off+25%cutoff+50%Cutoff+75%cutoff+100%cut off
Lot10+/60-0+/60-0+/60-5+/55-49+/11-56+/4-60+/0-60+/0-60+/0-
Lot20+/60-0+/60-0+/60-6+/54-48+/12-56+/4-60+/0-60+/0-60+/0-
Lot30+/60-0+/60-0+/60-5+/55-48+/12-56+/4-60+/0-60+/0-60+/0-

{9}------------------------------------------------

Methamphetamine
Operator 1
LotDrug-100%cut off-75%cut off-50%cut off-25%cut offCut off+25%cut off+50%Cut off+75%cut off+100%cut off
Lot 10+/60-0+/60-0+/60-6+/54-47+/13-56+/4-60+/0-60+/0-60+/0-
Lot 20+/60-0+/60-0+/60-5+/55-48+/12-55+/5-60+/0-60+/0-60+/0-
Lot 30+/60-0+/60-0+/60-6+/54-48+/12-54+/6-60+/0-60+/0-60+/0-
Operator 2
LotDrug-100%cut off-75%cut off-50%cut off-25%cut offCut off+25%cut off+50%Cut off+75%cut off+100%cut off
Lot 10+/60-0+/60-0+/60-6+/54-48+/12-56+/4-60+/0-60+/0-60+/0-
Lot 20+/60-0+/60-0+/60-5+/55-47+/13-56+/4-60+/0-60+/0-60+/0-
Lot 30+/60-0+/60-0+/60-6+/54-48+/12-55+/5-60+/0-60+/0-60+/0-
Operator 3
LotDrug-100%cut off-75%cut off-50%cut off-25%cut offCut off+25%cut off+50%Cut off+75%cut off+100%cut off
Lot 10+/60-0+/60-0+/60-5+/55-48+/12-55+/5-60+/0-60+/0-60+/0-
Lot 20+/60-0+/60-0+/60-5+/55-48+/12-56+/4-60+/0-60+/0-60+/0-
Lot 30+/60-0+/60-0+/60-5+/55-47+/13-56+/4-60+/0-60+/0-60+/0-

Methamphetamine

{10}------------------------------------------------

Morphine
Operator 1
LotDrug-100%cut off-75%cut off-50%cut off-25%cut offCut off+25%cut off+50%Cutoff+75%cut off+100%cut off
Lot10+/60-0+/60-0+/60-5+/55-48+/12-56+/4-60+/0-60+/0-60+/0-
Lot20+/60-0+/60-0+/60-6+/54-49+/11-56+/4-60+/0-60+/0-60+/0-
Lot30+/60-0+/60-0+/60-6+/54-49+/11-55+/5-60+/0-60+/0-60+/0-
Operator 2
LotDrug-100%cut off-75%cut off-50%cut off-25%cut offCut off+25%cut off+50%Cutoff+75%cut off+100%cut off
Lot10+/60-0+/60-0+/60-5+/55-49+/11-55+/5-60+/0-60+/0-60+/0-
Lot20+/60-0+/60-0+/60-6+/54-48+/12-55+/5-60+/0-60+/0-60+/0-
Lot30+/60-0+/60-0+/60-5+/55-49+/11-56+/4-60+/0-60+/0-60+/0-
Operator 3
LotDrug-100%cut off-75%cut off-50%cut off-25%cut offCut off+25%cut off+50%Cutoff+75%cut off+100%cut off
Lot10+/60-0+/60-0+/60-6+/54-49+/11-56+/4-60+/0-60+/0-60+/0-
Lot20+/60-0+/60-0+/60-5+/55-49+/11-55+/5-60+/0-60+/0-60+/0-
Lot30+/60-0+/60-0+/60-5+/55-48+/12-55+/5-60+/0-60+/0-60+/0-

Morphine

{11}------------------------------------------------

Operator 1
LotDrug-100%cut off-75%cut off-50%cut off-25%cut offCut off+25%cut off+50%Cutoff+75%cut off+100%cut off
Lot10+/60-0+/60-0+/60-6+/54-48+/12-55+/5-60+/0-60+/0-60+/0-
Lot20+/60-0+/60-0+/60-6+/54-48+/12-54+/6-60+/0-60+/0-60+/0-
Lot30+/60-0+/60-0+/60-5+/55-48+/12-56+/4-60+/0-60+/0-60+/0-
Operator 2
LotDrug-100%cut off-75%cut off-50%cut off-25%cut offCut off+25%cut off+50%Cutoff+75%cut off+100%cut off
Lot10+/60-0+/60-0+/60-5+/55-48+/12-54+/6-60+/0-60+/0-60+/0-
Lot20+/60-0+/60-0+/60-5+/55-48+/12-55+/5-60+/0-60+/0-60+/0-
Lot30+/60-0+/60-0+/60-5+/55-48+/12-54+/6-60+/0-60+/0-60+/0-
Operator 3
LotDrug-100%cut off-75%cut off-50%cut off-25%cut offCut off+25%cut off+50%Cutoff+75%cut off+100%cut off
Lot10+/60-0+/60-0+/60-6+/54-48+/12-55+/5-60+/0-60+/0-60+/0-
Lot20+/60-0+/60-0+/60-6+/54-48+/12-54+/6-60+/0-60+/0-60+/0-
Lot30+/60-0+/60-0+/60-6+/54-48+/12-55+/5-60+/0-60+/0-60+/0-

Phencyclidine

{12}------------------------------------------------

Operator 1
LotDrug-100%cut off-75%cut off-50%cut off-25%cut offCut off+25%cut off+50%Cutoff+75%cut off+100%cut off
Lot10+/60-0+/60-0+/60-6+/54-49+/11-56+/4-60+/0-60+/0-60+/0-
Lot20+/60-0+/60-0+/60-6+/54-48+/12-55+/5-60+/0-60+/0-60+/0-
Lot30+/60-0+/60-0+/60-5+/55-48+/12-56+/4-60+/0-60+/0-60+/0-
Operator 2
LotDrug100%cutoff-75%cut off-50%cut off-25%cut offCut off+25%cut off+50%Cutoff+75%cut off+100%cut off
Lot10+/60-0+/60-0+/60-5+/55-48+/12-55+/5-60+/0-60+/0-60+/0-
Lot20+/60-0+/60-0+/60-6+/54-49+/11-55+/5-60+/0-60+/0-60+/0-
Lot30+/60-0+/60-0+/60-6+/54-48+/12-55+/5-60+/0-60+/0-60+/0-
Operator 3
LotDrug100%cutoff-75%cut off-50%cut off-25%cut offCut off+25%cut off+50%Cutoff+75%cut off+100%cut off
Lot10+/60-0+/60-0+/60-6+/54-48+/12-55+/5-60+/0-60+/0-60+/0-
Lot20+/60-0+/60-0+/60-6+/54-48+/12-55+/5-60+/0-60+/0-60+/0-
Lot30+/60-0+/60-0+/60-6+/54-49+/11-56+/4-60+/0-60+/0-60+/0-

Marijuana (THC)

The following cut-off values for the candidate devices have been verified.

CalibratorCut-off (ng/mL)
d-Amphetamine50
Benzoylecgonine20
d-Methamphetamine50
Morphine40
Phencyclidine10
Delta-9-Tetrahydrocannabinol40

{13}------------------------------------------------

b. Linearity Not applicable.

c. Stability

Device stability has been evaluated through accelerated and real-time studies. The devices are stable at 2-30°C for 24 months based on the real time stability study at room temperature.

d. Interference

Potential interfering substances were added to drug-free oral fluid and target drugs oral fluid with concentrations at 50% below and 50% above cut-off levels. These oral fluid samples were tested using three batches of the device. Compounds that showed no interference for all the six drugs at a concentration of 10 µg/mL are summarized in the following table.

AcetaminophenDextromethorphanNaltrexone
AcetylcodeineDigoxinNaproxen
AllobarbitalDihydrocodeineNicotinamide
Alprazolamdiltiazem HClNicotine
AmobarbitalDiphenhydramineHClNoscapine
ApomorphineDL-PropranololOmeprazole
AtenololDoxylaminePapaverine
AtropineEcgonine methylesterPentazocine
BaclofenEstradiolPhentermine
BenzocaineEstronePhenylpropanolamine
ButabarbitalFluconazolePhenytoin
CaffeineFurosemidePioglitazone HCl
CannabidiolHexobarbitalPrednisolone
CarbamazepineHydrochlorothiazidePrednisone
ChlordiazepoxideIbuprofenProcainamide HCl
ChlorpromazineImipramineProcaine HC1
CimetidineLamotriginePromethazine
Citalopram HBrLevetiracetamQuinine HCl
ClobazamLidocaineR,R(-)-Pseudoephedrine
ClomipramineLormetazepamSalicylic Acid
ClonazepamL-ThyroxineSertraline HCl
ClonidineMetformin HClSimvastin
Clopidogrel bisulfateMethylphenidate HClTheophylline
CortisolMetoprololThiamine
CotinineMetronidazoleTopiramate
d,l-SalbutamolMontelukast sodiumsaltValproic Acid
DeoxycorticosteroneNaloxoneVerapamil
Zonisamide

{14}------------------------------------------------

Food items such as methanol cough drops, cough syrup, cola, mouthwash, coffee, tea, milk, sugar, chewing gum, alcohol, baking soda, salt, cranberry juice, orange juice, food coloring (red, blue, green), toothpaste, tomatoes and MSG were added in either drug-free oral fluid or oral fluid containing the target drug with concentrations of 50% below and 50% above cutoff levels to a concentration of 5%. None of the substances showed interference. Hemoglobin showed no interference at 100 ug/mL. Cigarette smoking showed no interference.

e. Specificity

To test specificity, drug metabolites and other components that are likely to interfere in oral fluid samples were tested using three batches of the device. The following are summaries.

Amphetamine (AMP)(Cut-off=50 ng/mL)Result Positive at(ng/mL)% CrossReactivity
D-Amphetamine50100.00%
L-Amphetamine40001.25%
D,L-Amphetamine12540.00%
Methoxyamphetamine50010%
Methylenedioxyamphetamine(MDA)15033%
Benzodioxolylbutanamine (BDB)100000.5%
3-Hydroxy Tyramine50001%
d,l-p-Chloramphetamine50010%
Phenethylamine40001.25%
Hydroxyamphetamine800< 6.25%
d,1-PhenylpropanolamineNegative at 10000< 0.50%
PhentermineNegative at 10000< 0.50%
Methylenedioxyethylamphetamine(MDEA)Negative at 10000< 0.50%
Methylenedioxy-methamphetamine(MDMA)Negative at 10000< 0.50%
d-MethamphetamineNegative at 10000< 0.50%
l-MethamphetamineNegative at 10000< 0.50%
Dimethylamylamine (DMAA)Negative at 10000< 0.50%
MethylbenzodioxolylbutanamineNegative at 10000< 0.50%
para-MethoxymethamphetamineNegative at 10000< 0.50%
PhendimetrazineNegative at 10000< 0.50%
PhenmetrazineNegative at 10000< 0.50%
D-EphedrineNegative at 10000< 0.50%
L-EphedrineNegative at 10000< 0.50%
D,L-EphedrineNegative at 10000< 0.50%
diphenhydramineNegative at 10000< 0.50%
d-PseudoephedrineNegative at 10000< 0.50%
FenfluramineNegative at 10000< 0.50%
IsoxsuprineNegative at 10000<0.50%

{15}------------------------------------------------

1-PseudoephedrineNegative at 10000< 0.50%
MephentermineNegative at 10000< 0.50%
COCAINE (COC)(Cut-off=20 ng/mL)Result Positive at(ng/mL)% CrossReactivity
Cocaine20100.00%
Benzoylecgonine20100.00%
Cocaethylene2580.00%
Ecgonine15001.33%
Ecgonine methyl ester125000.16%
ProcaineNegative at 20000< 0.10%
NorcocaineNegative at 20000<0.10%
METHAMPHETAMINE (MET)Result Positive% Cross
(Cut-off=50 ng/mL)at (ng/mL)Reactivity
D-Methamphetamine50100.00%
L-Methamphetamine30001.67%
Methoxymethamphetamine50100.00%
Ephedrine40012.50%
Phenylephrine40001.25%
Procaine20002.50%
Methylephedrine40012.50%
Methylenedioxyethylamphetamine(MDEA)40012.50%
3,4-methylenedioxy-methamphetamine(MDMA)50100.00%
AmphetamineNegative at100000< 0.05%
D-AmphetamineNegative at100000< 0.05%
L-AmphetamineNegative at100000< 0.05%
3,4-methylenedioxyamphetamineNegative at100000< 0.05%
OPIATES (OPI)(Cut-off=40 ng/mL)Result Positive at(ng/mL)% CrossReactivity
Morphine40100.00%
Acetylmorphine30133.33%
Codeine10400.00%
Ethylmorphine30133.33%
Heroin5080.00%
Hydromorphone10040.00%
Thebaine15002.67%
Norcodeine15002.67%
Morphine 6-β-glucuronide5080.00%

{16}------------------------------------------------

OPIATES (OPI)Result Positive at% Cross
(Cut-off=40 ng/mL)(ng/mL)Reactivity
Oxycodone250000.16%
Oxymorphone250000.16%
Nalorphine100000.40%
Hydrocodone10040.00%
6-Monoacetylmorphine30133.33%
Morphine 3-β-glucuronide5080.00%
Phencyclidine (PCP)Result Positive at% Cross
(Cut-off=10 ng/mL)(ng/mL)Reactivity
Phencyclidine10100.00%
HydrocodoneNegative at 30000< 0.03%
HydromorphoneNegative at 30000< 0.03%
NalorphineNegative at 30000< 0.03%
Tenocyclidine (TCP)20000.50%
1-(1-phenylcyclohexyl)morpholine(PCM)2050.00%
4-hydroxyphencyclidine2050.00%
EDDPNegative at 100000< 0.10%
KetamineNegative at 100000< 0.10%
PrazepamNegative at 100000< 0.10%
AmitriptylineNegative at 100000< 0.10%
(+) BrompheniramineNegative at 100000< 0.10%
(+) ChlorphenamineNegative at 100000< 0.10%
DesmethylvenlafaxineNegative at 100000< 0.10%
ChlorpromazineNegative at 100000< 0.10%
ClomipramineNegative at 100000< 0.10%
CyclizineNegative at 100000< 0.10%
CyclobenzaprineNegative at 100000< 0.10%
DexbrompheniramineNegative at 100000< 0.10%
DextromethorphanNegative at 100000< 0.10%
DiphenhydramineNegative at 100000< 0.10%
DoxepinNegative at 100000< 0.10%
DoxylamineNegative at 100000< 0.10%
ImipramineNegative at 100000< 0.10%
TrioridazineNegative at 100000< 0.10%
VenlafaxineNegative at 100000< 0.10%
MARIJUANA (THC)(Cut-off=40 ng/mL)Result Positive at(ng/mL)% CrossReactivity
Delta-9-Tetrahydrocannabinol40100.00%
11-nor-Δ9-THC-9 COOH12333.33%
Δ8-Tetrahydrocannabinol125000.32%

{17}------------------------------------------------

MARIJUANA (THC)Result Positive at% Cross
(Cut-off=40 ng/mL)(ng/mL)Reactivity
11-hydroxy-Δ9-THC40010.00%
Cannabinol125000.32%
CannabidolNegative at 12500< 0.32%
11-Nor-Δ9-THC-carboxy-glucuronide8050.00%
(-)-11-nor-9-carboxy-Δ9-THC6066.67%
11-nor-Δ8-THC-9-COOH5800.00%
8-beta- 11-dihydroxy-Δ9-THC40010.00%
8-beta-hydroxy-Δ9-THC30013.33%
Exo-THC8050.00%
1-11-Nor-Δ9-THC-9-CarboxylicAcyl-Glucuronide15266.67%
Δ8-THC60000.67%
Δ8-THC Carboxylic Acid5800.00%
Δ9-THC Carboxylic Acid12333.33%

f. Effect of Oral fluid pH

To investigate the effect of oral fluid pH, oral fluid samples with pH 3 to 9 were spiked with target drugs at 50% below and 50% above cut-off levels. These samples were tested using three lots of the device. Results were all positive for samples at and above +50% cut-off and all negative for samples at and below -50% cut-off.

g. Drug Recovery Study

Negative oral fluid samples in glass bottles were spiked with the drug to concentrations of -50% and +50% of the cutoff. The samples were transferred to the devices and store at room temperature, at -20℃ and at 40℃. Over 90% recoveries were observed for all drugs in the devices. Oral fluid samples can be stored in the device at -20℃ for at least 3 months. Oral fluid samples can be shipped overnight in the device for LC/MS confirmation.

8.2 Comparison Studies

Method comparison studies for the LYHER Oral fluid Multi-Drug Test Kit(Cube) were performed at three testing sites with three operators at each site. Operators tested the samples using the candidate device and the results were compared to LC/MS results. The results are presented in the tables below.

{18}------------------------------------------------

D-Amphetamine

OperatorTestresultNegativeoral fluid<-50%cut off-50% cutoff-cut offCut off-+50% cutoff>+50%cut off
Total (alloperatorsand sites)Negative360937490
Positive001180183

Cocaine

OperatorTestresultNegativeoral fluid<-50%cut off-50% cutoff-cut offCut off-+50% cutoff>+50%cut off
Total (alloperatorsand sites)Negative360698380
Total (alloperatorsand sites)Positive004172186

d-Methamphetamine

OperatorTestresultNegativeoral fluid<-50%cut off-50% cutoff-cut offCut off-+50% cutoff>+50%cut off
Total (alloperatorsand sites)Negative360637070
Total (alloperatorsand sites)Positive002164189

Morphine

OperatorTestresultNegativeoral fluid<-50%cut off-50% cutoff-cut offCut off+50% cutoff>+50%cut off
Total (alloperatorsand sites)Negative360578460
Total (alloperatorsand sites)Positive003183180

Phencyclidine

OperatorTestresultNegativeoral fluid<-50%cut off-50% cutoff-cut offCut off-+50% cutoff>+50%cut off
Total (alloperatorsand sites)Negative36051109130
Total (alloperatorsand sites)Positive005170192

{19}------------------------------------------------

OperatorTestresultNegativeoral fluid<-50%cut off-50% cutoff-cut offCut off-+50% cutoff>+50%cut off
Total (all operatorsand sites)Negative360939680
Total (all operatorsand sites)Positive003172195

Delta-9-Tetrahydrocannabinol

8.3 Clinical Studies Not Applicable.

9. Conclusion

Based on the test principle and acceptable performance characteristics including precision, interference, specificity, and method comparison studies of the devices, it's concluded that the LYHER Oral fluid Multi-Drug Test Kit(Cube) is substantially equivalent to the predicate.

§ 862.3610 Methamphetamine test system.

(a)
Identification. A methamphetamine test system is a device intended to measure methamphetamine, a central nervous system stimulating drug, in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of methamphetamine use or overdose.(b)
Classification. Class II (special controls). A methamphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).