Search Filters

Search Results

Found 13 results

510(k) Data Aggregation

    K Number
    K131515
    Manufacturer
    Date Cleared
    2013-08-14

    (78 days)

    Product Code
    Regulation Number
    862.1700
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    . § 862.1360)
    Product Code JPZ |
    | | Trade Name:

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACE y-GT Reagent is intended for the quantitative determination of gamma-glutamyltransferase activity in serum and lithium heparin plasma using the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. Gamma-glutamyltransferase measurements are used in the diagnosis and treatment of liver diseases such as alcoholic cirrhosis and primary and secondary liver tumors. This test is intended for use in clinical laboratories and physician office laboratories. For in vitro diagnostic use only.

    The ACE Lipase Reagent is intended for the quantitative determination of lipase activity in serum and lithium heparin plasma using the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. Lipase measurements are used in diagnosis and treatment of diseases of the pancreas such as acute pancreatitis and obstruction of the pancreatic duct. This test is intended for use in clinical laboratories and physician office laboratories. For in vitro diagnostic use only.

    The ACE T4 Reagent is intended for the quantitative determination of total thyroxine (T4) in serum and lithium heparin plasma using the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. Total thyroxine measurements are used in the diagnosis and treatment of thyroid diseases. This test is intended for use in clinical laboratories and physician office laboratories. For in vitro diagnostic use only.

    Device Description

    In the ACE γ-GT Reagent assay, γ-GT in serum or heparin plasma catalyzes the transfer of the γ-glutamyl group from L-γ-glutamyl-3-carboxy-4-nitroanilide to glycylglycine in the reagent. The product, 5-amino-2-nitrobenzoate, absorbs strongly at 408 nm. The rate of increase in absorbance, monitored bichromatically at 408 nm/486 nm, is directly proportional to the γ-GT activity in the sample.

    In the ACE Lipase Reagent Assay, lipase in serum or heparin plasma acts on a natural substrate, 1,2-diglyceride, to liberate 2-monoglyceride. This is hydrolyzed by monoglyceride lipase (a highly specific enzyme for monoglyceride) into glycerol and free fatty acid. Glycerol kinase acts on glycerol to form glycerol-3-phosphate, which is in turn acted on by glycerol-3-phosphate oxidase to generate hydrogen peroxide. Peroxidase converts the hydrogen peroxide, 4-Aminoantipyrine and TOOS (N-ethyl-N-(2-hydroxy-3-sulfopropyl)-m-toluidine) into a quinine dye. The rate of formation of the dye, determined bichromatically at an absorbance of 573 nm/692 nm, is proportional to the lipase activity in the sample.

    The ACE T4 Assay is a homogeneous enzyme immunoassay using ready-to-use liquid ACE T4 Reagent. The assay uses 8-anilino-1-naphthalene sulfonic acid (ANS) to dissociate thyroxine from the plasma binding proteins. Using specific antibodies to thyroxine, this assay is based on the competition of glucose-6-phosphate dehydrogenase (G6PD) labeled thyroxine and the dissociated thyroxine in the sample for a fixed amount of specific antibody binding sites. In the absence of thyroxine from the sample, the thyroxine labeled G6PD in the second reagent is bound by the specific antibody in the first reagent, inhibiting the enzyme's activity. The enzyme G6PD catalyzes the oxidation of glucose-6-phosphate (G6P) with nicotinamide adenine dinucleotide (NADT) to form 6-phosphogluconate and reduced nicotinamide adenine dinucleotide (NADH). NADH strongly absorbs at 340 nm whereas NAD does not. The rate of conversion, determined by measuring the increase in absorbance bichromatically at 340 nm/505 nm during a fixed time interval, is directly proportional to the amount of thyroxine in the sample. The concentration of thyroxine is determined automatically by the ACE Clinical Chemistry Systems using a logarithmic calibration curve established with calibrators, which are provided separately.

    AI/ML Overview

    The information provided describes the performance of the ACE γ-GT, ACE Lipase, and ACE T4 Reagents on the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. This is not an AI/ML device, however, I will address the other requested points to the best of my ability with the provided text.

    Here's a breakdown of the acceptance criteria and study information, where applicable:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in a separate table. However, it provides performance data for precision, matrix comparison (serum vs. plasma), detection limits, linearity, and interference. Based on the "Conclusions" section, the goal was to demonstrate "substantial equivalence" of the reagents for lithium heparin plasma samples (compared to serum) and the ACE Alera System (compared to the predicate ACE Clinical Chemistry System). The performance data presented are implicitly intended to support this substantial equivalence.

    Implied Acceptance Criteria (based on predicate comparison and performance data) and Reported Performance:

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance (Summary)
    PrecisionFor In-House Precision (Serum vs. Plasma): Low, Mid, and High analyte concentrations should demonstrate acceptable within-run and total %CV on ACE, ACE Alera, and ACE Axcel systems, comparable to expected values for clinical chemistry assays.

    For POL Precision: Similar acceptable %CV values for low, mid, and high samples across different POL sites and in-house, on ACE and ACE Alera systems. | In-House Precision (Serum vs. Plasma):

    • γ-GT: Total %CV generally 0.997, Slope 0.960-0.987, Intercept 1.5-4.0 across systems.
      Lipase: Correlation > 0.994, Slope 0.980-1.024, Intercept -2.5 to -0.9 across systems (for ACE and ACE Alera, Axcel missing intercept CI).
      T4: Correlation > 0.984, Slope 0.963-1.007, Intercept 0.01-0.35 across systems. |
      | Method Comparison (POL) | When comparing results from POL sites to in-house results on the same instrument, correlation coefficients should be high (close to 1), slopes close to 1, and with small intercepts, indicating consistency across testing locations. | ACE System:
    • γ-GT: Correlation > 0.9997, Slope 0.964-0.976, Intercept -2.7 to 0.7.
    • Lipase: Correlation > 0.9966, Slope 0.994-1.031, Intercept -5.3 to 0.0.
    • T4: Correlation > 0.9908, Slope 1.010-1.019, Intercept -0.09 to -0.04.

    ACE Alera System:

    • γ-GT: Correlation > 0.9996, Slope 0.950-1.028, Intercept 1.9 to 2.9.
    • Lipase: Correlation > 0.9960, Slope 0.992-1.028, Intercept -3.5 to 3.3.
    • T4: Correlation > 0.9868, Slope 1.022-1.048, Intercept -0.31 to -0.10. |
      | Detection Limits (ACE Alera) | Limits of Blank (LOB), Detection (LOD), and Quantitation (LOQ) should be clinically acceptable. | γ-GT: LOB 3 U/L, LOD 5 U/L, LOQ 7 U/L.
      Lipase: LOB 7 U/L, LOD 11 U/L, LOQ 13 U/L.
      T4: LOB 0.3 µg/dL, LOD 0.8 µg/dL, LOQ 1.3 µg/dL. |
      | Linearity (ACE Alera) | The assay should be linear up to the stated measuring range, with a linear regression equation demonstrating good fit. | γ-GT: Linear to 950 U/L ($y = 1.036x + 0.8$).
      Lipase: Linear to 700 U/L ($y = 0.971x + 0.2$).
      T4: Linear to 19.6 µg/dL ($y = 1.057x - 0.09$). |
      | Interferences (ACE Alera) | No significant interference from common exogenous or endogenous substances at physiologically relevant or elevated concentrations. | γ-GT: No significant interference at or below Icterus 14.2 mg/dL, Hemolysis 125 mg/dL, Lipemia 500 mg/dL, Ascorbic Acid 6 mg/dL.
      Lipase: No significant interference below Icterus 12.5 mg/dL, Hemolysis 1000 mg/dL, Lipemia 803 mg/dL, Ascorbic Acid 6 mg/dL.
      T4: No significant interference below Icterus 47.2 mg/dL, Hemolysis 1000 mg/dL, Lipemia 1000 mg/dL, Ascorbic Acid 6 mg/dL.
      Heterophile (T4): HAMA 800 ng/mL, RF 516 IU/mL.
      Cross-Reactivity (T4): 3,3',5,5'- Tetraiodothyroacetic Acid (18.4%), L-Thyroxine (91.6%), D-Thyroxine (68.0%) at 5 µg/dL. |

    2. Sample Sizes Used for the Test Set and Data Provenance

    The document does not explicitly use the term "test set" in the context of AI/ML, but rather describes clinical performance studies. The sample sizes for these studies are as follows:

    • In-House Matrix Comparison (Serum vs. Plasma):
      • ACE γ-GT Reagent: 100 pairs (ACE), 97 pairs (ACE Alera), 53 pairs (ACE Axcel)
      • ACE Lipase Reagent: 42 pairs (ACE), 43 pairs (ACE Alera), 62 pairs (ACE Axcel)
      • ACE T4 Reagent: 55 pairs (ACE), 55 pairs (ACE Alera), 55 pairs (ACE Axcel)
    • Method Comparison (POL vs. In-House):
      • ACE System: 50-54 samples per reagent per POL site (3 POL sites)
      • ACE Alera System: 48-51 samples per reagent per POL site (3 POL sites)
    • Precision (In-House and POL): The number of replicates per sample level (Low, Mid, High) is not explicitly stated, but precision studies typically involve multiple runs over several days.
    • Detection Limits, Linearity, Interferences, Cross-Reactivity: Sample sizes for these specific experiments are not detailed but are generally conducted with a sufficient number of replicates and concentrations to statistically establish the parameters.

    Data Provenance: The studies are described as "In-House" and "POL" (Physician Office Laboratory) studies. This indicates that the data was collected at the manufacturer's facility ("In-House") and potentially at various POL sites. The country of origin is not explicitly stated, but given the 510(k) submission to the FDA, it is likely the studies align with US regulatory requirements and are potentially from US-based labs. The studies are prospective in nature, as they involve newly generated data to demonstrate the performance of the devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This section is not applicable as the device is a clinical chemistry reagent and not an AI/ML device that generates interpretations requiring expert ground truth for image or diagnostic data. The "ground truth" in this context refers to the measured analyte concentrations obtained from established laboratory methods, calibrators, and reference materials.

    4. Adjudication Method for the Test Set

    This section is not applicable as the device is a clinical chemistry reagent. Adjudication methods like 2+1 or 3+1 are used in contexts like human reader studies for diagnostic imaging, where discordant interpretations need resolution by additional experts.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    This section is not applicable as the device is a clinical chemistry reagent. MRMC studies are designed to assess the performance of diagnostic devices or AI algorithms by multiple human readers across multiple cases, especially in imaging.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This section is not applicable as the device is a clinical chemistry reagent. This term is relevant for AI/ML diagnostic tools. The "performance" of this device is inherently standalone in that the instrument processes samples and generates quantitative results without human intervention in the measurement process itself, beyond sample loading and general operation.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance studies presented is based on quantitative chemical measurements of the specific analytes (gamma-glutamyltransferase, lipase, total thyroxine) in control materials, patient samples, and comparison with established reference methods or predicate devices. This includes:

    • Known concentrations: For precision, linearity, detection limits, and interference studies, samples with known or spiked concentrations are used.
    • Comparison to predicate device: For method comparison studies, the results from the new device/system are compared against the results from the legally marketed predicate device/system.
    • Reference materials/calibrators: The accuracy and calibration of the assays depend on traceable reference materials and calibrators.

    8. The Sample Size for the Training Set

    This section is not applicable as the device is a clinical chemistry reagent and not an AI/ML device. There is no "training set" in the context of machine learning model development.

    9. How the Ground Truth for the Training Set Was Established

    This section is not applicable for the same reasons as #8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K113382
    Device Name
    ACE ALT REAGENT
    Manufacturer
    Date Cleared
    2012-07-19

    (246 days)

    Product Code
    Regulation Number
    862.1030
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    . § 862.1360)
    Product Code JPZ |
    | Predicate Devices: | Manufacturer for analyzer/reagent system

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACE ALT Reagent is intended for the quantitative determination of alanine aminotransferase activity in serum using the ACE Axcel Clinical Chemistry System. Alanine aminotransferase measurements are used in the diagnosis and treatment of certain liver diseases (e.g., viral hepatitis and cirrhosis) and heart diseases. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE AST Reagent is intended for the quantitative determination of aspartate aminotransferase activity in serum using the ACE Axcel Clinical Chemistry System. Measurements of aspartate aminotransferase are used in the diagnosis and treatment of certain types of liver and heart disease. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE γ-GT Reagent is intended for the quantitative determination of gamma-glutamyltransferase activity in serum using the ACE Axcel Clinical Chemistry System. Gamma-glutamyltransferase measurements are used in the diagnosis and treatment of liver diseases such as alcoholic cirrhosis and primary and secondary liver tumors. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    Device Description

    In the ACE ALT Reagent assay, alanine aminotransferase in serum converts the L-alanine and α-ketoglutarate substrates in the reagent to L-glutamate and pyruvate. Lactate dehydrogenase (LDH) catalyzes the oxidation of the reduced cofactor to the cofactor. The rate of conversion of the reduced cofactor to the cofactor can be determined by monitoring the decrease in absorbance bichromatically at 340 nm/647 nm. This rate of conversion from the reduced cofactor to the cofactor is a function of the activity of ALT in the sample.

    In the ACE AST Reagent assay, aspartate aminotransferase in serum converts the L-aspartate and α-ketoglutarate in the reagent to oxalacetate and L-glutamate. The oxalacetate undergoes reduction with simultaneous oxidation of NADH to NAD+ in the malate dehydrogenase catalyzed indicator reaction. NADH absorbs strongly at 340 nm, whereas NAD+ does not. Therefore, the rate of conversion of NADH to NAD+ can be determined by monitoring the decrease in absorbance bichromatically at 340 nm/647 nm. This rate of conversion from NADH to NAD+ is a function of the activity of AST in the sample. Lactate dehydrogenase is added to prevent interference from endogenous pyruvate, which is normally present in serum.

    In the ACE γ-GT Reagent assay, γ-GT in serum catalyzes the transfer of the γ-glutamyl group from L-γ-glutamyl-3-carboxy-4-nitroanilide to glycylglycine in the reagent. The product, 5-amino-2-nitrobenzoate, absorbs strongly at 408 nm. The rate of increase in absorbance, monitored bichromatically at 408 nm/486 nm, is directly proportional to the γ-GT activity in the sample.

    AI/ML Overview

    This document describes the performance of the ACE ALT, AST, and γ-GT Reagents on the ACE Axcel Clinical Chemistry System. These reagents are intended for the quantitative determination of alanine aminotransferase (ALT), aspartate aminotransferase (AST), and gamma-glutamyl transferase (γ-GT) activity in serum, used in the diagnosis and treatment of certain liver and heart diseases.

    Here's an analysis of the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly state "acceptance criteria" as distinct from the reported performance. Instead, it presents the results of precision and accuracy studies. The implication is that these results demonstrate acceptable performance for the device. For the purpose of this table, the reported performance values are presented with the understanding that they met the internal requirements of the manufacturer for regulatory submission.

    MetricAcceptance Criteria (Implied)ACE ALT ReagentACE AST ReagentACE γ-GT Reagent
    PrecisionLow CV% for within-run and total CVLab Setting (4 levels, 22 days): Within-run CV: 0.8% - 6.9%; Total CV: 1.1% - 6.9%.
    POL Sites (3 sites, 5 days): Within-run CV: 0.8% - 8.7%; Total CV: 1.3% - 8.7%.Lab Setting (4 levels, 22 days): Within-run CV: 0.9% - 7.1%; Total CV: 1.4% - 8.5%.
    POL Sites (3 sites, 5 days): Within-run CV: 1.1% - 11.3%; Total CV: 1.2% - 11.3%.Lab Setting (4 levels, 22 days): Within-run CV: 1.0% - 3.0%; Total CV: 1.1% - 6.1%.
    POL Sites (3 sites, 5 days): Within-run CV: 0.7% - 12.4%; Total CV: 1.3% - 13.0%.
    Accuracy (Correlation to Predicate)High correlation coefficient (near 1), low standard error, slope near 1, intercept near 0Correlation Study (102 samples, 4-472 U/L): Correlation coefficient: 0.9996; Standard error: 2.4; CI slope: 1.035 to 1.047; CI intercept: -0.3 to 0.9.
    POL Sites: Correlation coefficients: 0.9997 to 0.9999; Standard error: 2.4 to 3.1; CI slopes: 1.009 to 1.035; CI intercepts: -1.1 to 2.4.Correlation Study (117 samples, 8-440 U/L): Correlation coefficient: 0.9996; Standard error: 2.2; CI slope: 1.002 to 1.012; CI intercept: 1.9 to 2.8.
    POL Sites: Correlation coefficients: 0.9996 to 0.9998; Standard error: 2.5 to 2.9; CI slopes: 1.005 to 1.038; CI intercepts: -1.4 to 2.1.Correlation Study (128 samples, 7-902 U/L): Correlation coefficient: 0.9998; Standard error: 3.4; CI slope: 0.981 to 0.988; CI intercept: -0.6 to 0.8.
    POL Sites: Correlation coefficients: 0.9992 to 0.9999; Standard error: 3.6 to 8.8; CI slopes: 0.967 to 1.053; CI intercepts: -1.7 to 4.8.
    Detection LimitLow U/L value3.1 U/L1.5 U/L2.7 U/L

    2. Sample Sizes Used for the Test Set and Data Provenance:

    • ACE ALT Reagent:

      • Precision: Not explicitly stated as a "test set" sample size for patient samples. Precision was assessed at four ALT levels over 22 days in a lab setting and at three Physician Office Laboratory (POL) sites over 5 days. These studies would typically involve repeated measurements of control materials or pooled patient samples.
      • Accuracy: 102 samples for the main correlation study (ALT values from 4 to 472 U/L) and patient correlation studies conducted at three separate POL sites.
      • Data Provenance: Not explicitly stated, but the mention of Physician Office Laboratory (POL) sites suggests data from clinical settings. It is implied to be prospective data collected for the study, rather than retrospective.
    • ACE AST Reagent:

      • Precision: Similar to ALT, assessed at four AST levels over 22 days in a lab and at three POL sites over 5 days.
      • Accuracy: 117 samples for the main correlation study (AST values from 8 to 440 U/L) and patient correlation studies conducted at three separate POL sites.
      • Data Provenance: Not explicitly stated, but implies clinical settings and prospective data.
    • ACE γ-GT Reagent:

      • Precision: Similar to ALT and AST, assessed at four γ-GT levels over 22 days in a lab and at three POL sites over 5 days.
      • Accuracy: 128 samples for the main correlation study (γ-GT values from 7 to 902 U/L) and patient correlation studies conducted at three separate POL sites.
      • Data Provenance: Not explicitly stated, but implies clinical settings and prospective data.

    The country of origin for the data is not specified, but the manufacturer is based in the US.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not applicable as the device is an in vitro diagnostic (IVD) chemistry system that provides quantitative measurements of enzyme activity. The "ground truth" for IVD devices like this is typically established by comparing performance against a reference method or a legally marketed predicate device, rather than expert interpretation of images or clinical assessments.

    For accuracy, the device's performance was compared to the Alfa Wassermann ACE Clinical Chemistry System (predicate device K931786).

    4. Adjudication Method for the Test Set:

    This information is not applicable. Adjudication methods (like 2+1 or 3+1) are typically used in studies where human experts interpret results or clinical cases, and a consensus needs to be reached. For quantitative IVD tests, the "ground truth" is the result obtained from a reference method or predicate device.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable. MRMC studies are relevant for medical imaging AI devices where human readers interpret images. This device is an in vitro diagnostic system for chemical analysis and does not involve human readers interpreting cases in the same way.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Yes, the performance studies described (precision, accuracy, and detection limit) represent the standalone performance of the ACE ALT, AST, and γ-GT Reagents on the ACE Axcel Clinical Chemistry System. These are objective, quantitative measurements generated directly by the analytical system, without human interpretation in the "loop" of the measurement process itself.

    7. The Type of Ground Truth Used:

    The ground truth used for the accuracy studies was the measurements obtained from the predicate device, the Alfa Wassermann ACE Clinical Chemistry System (K931786). The new reagents on the ACE Axcel system (y) were compared against the existing reagents on the ACE system (x) using patient samples. This is a common method for demonstrating substantial equivalence for new IVD devices by showing good correlation with an already legally marketed device.

    8. The Sample Size for the Training Set:

    This information is not applicable. The device is an in vitro diagnostic reagent and system, not an AI/ML algorithm that requires a "training set" in the conventional sense of machine learning. The reagents and system are developed and optimized through traditional chemical and engineering processes, followed by validation studies as described.

    9. How the Ground Truth for the Training Set was Established:

    This information is not applicable for the same reason as point 8. There is no training set in the machine learning context for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K060677
    Device Name
    N LATEX CDT
    Manufacturer
    Date Cleared
    2006-08-25

    (164 days)

    Product Code
    Regulation Number
    862.1360
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Device Name

    Classification:

    Panel: Product Code:

    Class I 21 CFR 862.1360 Clinical Chemistry (75) NAO
    19714

    , ,

    AUG 25 2006

    K060677 Re:

    Trade/Device Name: N Latex CDT Kit Regulation Number: 21 CFR 862.1360

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    In vitro diagnostic for the quantitative determination of carbohydrate-deficient transferrin (CDT) in human serum by means of particle-enhanced immunonephelometry using the BN™ II and BN ProSpec® System. The N Latex CDT assay must be run concurrently with the N Antisera to Human Transferrin assay so that the result can be expressed as a relative ratio, i.e., %CDT of the total transferrin. The calculation of %CDT can be used as a tool to identify possible chronic heavy alcohol consumption.

    Device Description

    The CDT in the sample competes with CDT-coated polystyrene particles for the bond to specific monoclonal antibodies against human CDT, which are likewise bound to polystyrene particles. In the presence of CDT in the sample, there is no or little aggregation of the polystyrene particles. In the absence of CDT in the sample, the polystyrene particles aggregate. The higher the CDT content in the assay, the lower the scattered light signal. The evaluation is performed by comparison with a standard of known concentration.

    AI/ML Overview

    The provided text describes the 510(k) summary for the N Latex CDT Kit, which is an in vitro diagnostic device. The study presented focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than defining and meeting new acceptance criteria for standalone performance or comparative effectiveness.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly state "acceptance criteria" in the traditional sense of pre-defined thresholds for performance metrics. Instead, it demonstrates performance by comparing the new device (N Latex CDT assay) against a predicate device (Axis-Shield %CDT assay). The criteria for "acceptance" appear to be based on a high correlation and agreement with the predicate.

    Performance MetricAcceptance Criteria (Implicit)Reported Device Performance (N Latex CDT vs. Predicate)
    CorrelationClose to 1.0Correlation coefficient: 0.99
    Agreement (Linear Regression)Slope close to 1, Intercept close to 0y = 0.720x + 0.75 %CDT

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size for Test Set: 116 serum samples
    • Data Provenance: The document does not specify the country of origin of the data or whether the study was retrospective or prospective. It only states that 116 serum samples were "evaluated."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    Not applicable. The study does not establish "ground truth" through expert consensus. Instead, it uses a legally marketed predicate device's results as the reference for comparison.

    4. Adjudication Method for the Test Set:

    Not applicable. There was no expert adjudication process described, as the comparison was against a predicate device's results.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is an in vitro diagnostic assay, not an AI-assisted diagnostic tool involving human readers. Therefore, an MRMC study or evaluation of human reader improvement with AI assistance is not relevant to this submission.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    The device's performance, as described by the regression analysis (y = 0.720x + 0.75 %CDT, correlation coefficient 0.99), is essentially its "standalone" performance when compared against the predicate device. It's a chemical assay, so its performance is inherent to the assay itself.

    7. The Type of Ground Truth Used:

    The "ground truth" in this context is the results obtained from the predicate device (Axis-Shield %CDT Assay) for the same serum samples. This is a form of reference standard comparison rather than ground truth established by pathology or clinical outcomes.

    8. The Sample Size for the Training Set:

    Not applicable. This is an in vitro diagnostic assay, not a machine learning or AI-based device that typically requires a distinct training set in the same manner. The "training" of such a device generally refers to its chemical formulation and calibration, which is not detailed in terms of a "training set sample size" in this document.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no mention or implication of a "training set" in the context of machine learning or AI. The development of an in vitro diagnostic largely relies on chemical and biological principles, calibration, and validation, rather than ground truth established for a training dataset in the AI sense.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    510k Summary Text (Full-text Search) :

    test system |
    | 862.1175 Cholesterol (total) test system | Cholesterol test system |
    | 862.1360
    cartridge | Dimension® GGT
    Flex® reagent
    cartridge | K861700 | I* | 862.1360

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Dimension Vista™ Acetaminophen (ACTM) Flex® reagent cartridge is a device intended to measure acetaminophen, an analgesic and antipyretic (fever reducing) drug, in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of acetaminophen overdose.

    The Dimension Vista™ Amylase (AMY) Flex® reagent cartridge is a device intended to measure the activity of the enzyme amylase in serum, plasma and urine. Amylase measurements are used primarily for the diagnosis and treatment of pancreatitis (inflammation of the pancreas).

    The Dimension Vista™ Creatine Kinase (CK) Flex® reagent cartridge is a device intended to measure the activity of the enzyme creatine kinase in serum and plasma. Measurements of creatine kinase are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive Duchenne-type muscular dystrophy.

    The Dimension Vista™ Cholesterol (CHOL) Flex® reagent cartridge is a device intended to measure cholesterol in serum and plasma. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.

    The Dimension Vista™ Gamma-glutamyl transferase (GGT) Flex® reagent cartridge is a device intended to measure gamma-glutamyl transferase in human serum and plasma. Gamma-glutamyl transferase measurements are used in the diagnosis and treatment of liver diseases such as alcoholic cirrhosis and primary and secondary liver tumors.

    The Dimension Vista™ Glucose (GLU) Flex® reagent cartridge is a device intended to measure glucose in human serum, plasma, urine and cerebrospinal fluid. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal and idiopathic hypoglycemia, and pancreatic islet cell carcinoma.

    The Dimension Vista™ High-Density Lipoprotein Cholesterol (HDLC) Flex® reagent cartridge is intended to measure high-density lipoprotein cholesterol in serum and plasma. Measurements of high-density lipoprotein cholesterol are used in the diagnosis of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.

    The Dimension Vista™ Low-Density Lipoprotein Cholesterol (LDLC) Flex® reagent cartridge is intended to measure low-density lipoprotein cholesterol in serum and plasma. Measurements of low-density lipoprotein cholesterol are used in the diagnosis of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.

    The Dimension Vista™ Lidocaine (LIDO) Flex® reagent cartridge is a device intended to measure lidocaine, an antiarrythmic and anticonvulsant drug, in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of lidocaine overdose or in monitoring levels of lidocaine to ensure appropriate therapy.

    The Dimension Vista™ Magnesium (MG) Flex® reagent cartridge is intended for the measurement of magnesium levels in serum and plasma. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low plasma levels of magnesium) and hypermagnesemia (abnormally high plasma levels of magnesium).

    The Dimension Vista™ Pseudocholinesterase (PCHE) Flex® reagent cartridge is a device intended to measure pseudocholinesterase activity in human serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of cholinesterase inhibition disorders (e.g., insecticide poisoning and succinylcholine poisoning).

    The Dimension Vista™ Phosphorus (PHOS) Flex® reagent cartridge is a device intended to measure inorganic phosphorus in serum, plasma, and urine. Measurements of phosphorus (inorganic) are used in the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases, and vitamin D imbalance.

    The Dimension Vista™ Procainamide (PROC) Flex® reagent cartridge is a device intended to measure procainamide in serum and plasma. Measurements obtained may be used in the diagnosis and treatment of procainamide overdose and in monitoring levels of procainamide to ensure appropriate therapy.

    The Dimension Vista™ Salicylate (SAL) Flex® reagent cartridge is a device intended to measure salicylates, a class of analgesic, antipyretic and anti-inflammatory drugs that includes aspirin, in human serum. Measurements obtained by this device are used in the diagnosis and treatment of salicylate overdose and in monitoring salicylate levels to ensure appropriate therapy.

    The Dimension Vista™ Thyroxine (T4) Flex® reagent cartridge is a device intended to measure total (free and protein bound) thyroxine (thyroid hormone) in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of thyroid diseases.

    The Dimension Vista™ Tobramycin (TOBR) Flex® reagent cartridge is a device intended to measure tobramycin, an aminoglycoside antibiotic drug, in palsma and serum. Measurements obtained by this device are used in the diagnosis and treatment of tobramycin overdose and in monitoring levels of tobramycin to ensure appropriate therapy.

    The Dimension Vista™ Triglyceride (TRIG) Flex® reagent cartridge is a device intended to measure triglyceride (neutral fat) in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders.

    The Dimension Vista™ Uric Acid (URCA) Flex® reagent cartridge is a device intended to measure uric acid in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions, and of patients receiving cytotoxic drugs.

    The Dimension Vista™ Valproic Acid (VALP) Flex® reagent cartridge is a device intended to measure valproic acid, an anti-convulsant drug in serum and plasma. Measurements obtained may be used in the diagnosis and treatment of valproic acid overdose and in monitoring levels of valproic acid to ensure appropriate therapy.

    The Dimension Vista™ Vancomycin (VANC) Flex® reagent cartridge is a device intended to measure vancomycin, an antibiotic drug, in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of vancomycin overdose and in monitoring the level of vancomycin to ensure appropriate therapy.

    Device Description

    Dade Behring Dimension Vista™ Flex® reagent cartridges are prepackaged in-vitro diagnostic test methods (assays) that are specifically designed to be used on the Vade Behring Dimension Vista™ Integrated system, a floor model, fully automated, microprocessor-controlled, integrated instrument system. The Dimension Vista™ system was previously cleared with seven associated test methods (K 051087). This Special 510(k) is submitted for a packaging modification to in-vitro diagnostic devices that have been cleared under the 510(k) process for use on Dimension® clinical chemistry systems. The packaging change is to allow use on the Dimension Vista™ system.

    The reagents contained in the Dimension Vista™ Flex® reagent cartridges are the same as those contained in the Flex® reagent cartridges manufactured for the Dimension® clinical chemistry systems, another family of Dade Behring analyzers. The packaging modification, does not affect the intended use of the devices, nor does it alter the fundamental scientific technology of the devices.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Dade Behring Dimension Vista™ Flex® reagent cartridges, based on the provided 510(k) summary:

    This device submission is a Special 510(k) for a packaging modification, meaning the core technology and reagents are the same as previously cleared devices. Therefore, the primary goal of the study is to demonstrate substantially equivalent performance after the packaging change, rather than to establish initial performance claims for a novel device.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of numerical acceptance criteria or specific performance metrics (e.g., accuracy, precision values) for each analyte. Instead, it relies on a comparative equivalency approach to a predicate device.

    The overarching acceptance criterion is "substantially equivalent performance" to the predicate Dimension® Flex® reagent cartridges.

    Acceptance CriterionReported Device Performance (Summary)
    Substantial Equivalence to Predicate Device"Comparative testing described in the protocol included in this submission demonstrates substantially equivalent performance."
    Same Intended Use and Indications for UseConfirmed; the packaging modification does not affect intended use or indications.
    Same Reagents and Fundamental Scientific TechnologyConfirmed; reagents are the same, and the fundamental scientific technology is unaltered.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states: "Comparative testing described in the protocol included in this submission demonstrates substantially equivalent performance."

    • Sample Size for Test Set: This information is not explicitly stated in the provided summary. The summary refers to a "protocol included in this submission," which would contain these details.
    • Data Provenance: This information is not explicitly stated in the provided summary.
    • Retrospective or Prospective: This information is not explicitly stated. However, given the nature of in-vitro diagnostic testing for performance comparison, it would typically involve prospective testing on patient samples or spiked samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This is an in-vitro diagnostic device for quantitative measurement of analytes in human samples (serum, plasma, urine, CSF). The ground truth for such devices is established by:

    • Reference Methods: Highly accurate and precise laboratory methods, often gold standards like GC-MS, HPLC, or other well-validated enzymatic or spectrophotometric methods.
    • Certified Reference Materials (CRMs): Samples with known, certified concentrations of the analytes.

    Therefore, the concept of "experts" in the clinical imaging or diagnostic interpretation sense (e.g., radiologists) is not applicable here. The ground truth is laboratory-based and instrumental.


    4. Adjudication Method for the Test Set

    Not applicable for this type of in-vitro diagnostic device. Ground truth is established by reference methods or certified materials, not by expert consensus or adjudication.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • No, an MRMC comparative effectiveness study was not done.
    • This device is an in-vitro diagnostic reagent cartridge, not an AI-powered diagnostic imaging tool or a system designed for human interpretation with or without AI assistance. The performance is measured instrumentally.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Yes, the performance evaluated is inherently "standalone" in the context of an automated analytical instrument. The Flex® reagent cartridges are designed to be used on the Dimension Vista™ Integrated system, a "fully automated, microprocessor-controlled, integrated instrument system." The performance of the reagent (device) is measured by its output on this automated system.
    • There is no "human-in-the-loop" decision-making component for the measurement process itself, although clinical interpretation of the results by a healthcare professional is expected.

    7. The Type of Ground Truth Used

    The ground truth for this type of in-vitro diagnostic device would typically involve:

    • Reference Method Assays: Using established, highly accurate, and precise laboratory methods (e.g., a recognized primary reference measurement procedure or a well-characterized predicate device itself) to determine the true concentration of the analytes in the test samples.
    • Certified Reference Materials: Commercial or internal standards with known, traceable concentrations of the analytes.
    • Sample Matrix: Patient samples (serum, plasma, urine, CSF) with concentrations spanning the analytical range.

    The summary states "Comparative testing... demonstrates substantially equivalent performance." This strongly implies that the new device's measurements were compared against the measurements obtained by the predicate device on the same samples, which serves as the "reference" or "ground truth" for the equivalence claim.


    8. The Sample Size for the Training Set

    This device is a reagent cartridge for an in-vitro diagnostic test, not a machine learning or AI algorithm in the contemporary sense that requires a "training set" to learn. The reagents and their chemical reactions are based on established scientific principles.

    Therefore, the concept of a "training set" as understood in machine learning is not applicable to this device.


    9. How the Ground Truth for the Training Set Was Established

    As noted above, the concept of a "training set" is not applicable to this device. The ground truth for the performance evaluation (test set) would be established by reference methods or comparison to the predicate device, as described in point 7.

    Ask a Question

    Ask a specific question about this device

    K Number
    K024182
    Date Cleared
    2003-03-03

    (74 days)

    Regulation Number
    862.1770
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    862.1100 |
    | Gamma Glutamyl Transferase | 931958 | I | 941315 | 862.1360
    Glutamyl Transferase | 862.1360

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Precision Systems™ ANALETTE™ Chemistry Analyzer is intended for the quantitative determination of Calcium, Creatinine, Phosphorus, Albumin, Total Protein, Glucose, Urea Nitrogen, Magnesium, Creatine Kinase, Alkaline Phosphatase, Cholesterol(includes HDL), Triglycerides, Total Bilirubin, Direct Bilirubin, Uric Acid, Lactate Dehydrogenase L, Alanine Aminotransferase, Aspartate Aminotransferase, Gamma Glutamyl Transferase, Chloride, and etc. analytes in solution such as serum, plasma, or urine. It is an "open" System, which can use a variety of commercially manufactured reagents such as but not limited to Synermeds® Reagents, Medical Analysis Systems Reagents and STANBIO Laboratory Reagents. It is used to monitor various physiological diseases or conditions. Precision Systems Inc will distribute, recommend and sales STANBIO Reagents without any modification of STANBIO packaging using PSI Applications sheets.

    Device Description

    The ANALETTE™ Chemistry Analyzer is an in vitro diagnostic automated clinical chemistry analyzer for the analysis of analytes in solution. It is an "open" System, which can use a variety of commercially manufactured reagents.

    AI/ML Overview

    The document describes the acceptance criteria and the study conducted to demonstrate the substantial equivalence of the Precision Systems™ ANALETTE™ Chemistry Analyzer using STANBIO Laboratory Reagents to its predicate devices (ANALETTE™ using Synermed® Reagents and ANALETTE™ using Medical Analysis Systems Inc® Reagents).

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document explicitly states: "Performance specifications: None established under Section 514." Instead, it refers to "Acceptance Criteria" (Exhibit E and F) but does not detail the specific numerical acceptance criteria within the provided text.

    However, the "Results" section (G.) provides the reported performance relative to "acceptable/equivalent results" or "Manufacturers' claim."

    Performance MetricAcceptance Criteria (Implied/Referenced)Reported Device Performance
    ImprecisionAcceptable/equivalent results (Implied)Serum controls give acceptable/equivalent results using the described procedure for within run and total imprecision with each of the representative test methods (Synermed, Medical Analysis Systems, and STANBIO Laboratory Reagents, as shown in Table 1 and Table 2 vs insert values).
    CorrelationAcceptable results (Implied)Slopes, Intercepts and Correlation Coefficients show acceptable results. The regression (slope and intercept) and correlation coefficients are shown in Table 3 and Graphs 1-21. Acceptable results are shown between both methods (STANBIO Laboratory Reagents vs. Synermeds® or Medical Analysis Systems Reagents).
    LinearityNot exceeding Manufacturers' claimLinearity did not exceed the Manufacturers' claim (shown in Table 4 vs insert values). A comparison is made between STANBIO Laboratory Reagents and the Least Square line to establish linearity.
    RecoveryAcceptable results for assigned rangesAcceptable results are shown between both methods (using assigned control serums ranges, shown in Table 5).
    Normal RangeRemains as recommended by manufactureParameters were not tested, assumed to remain as recommended by manufacture as no modifications to STANBIO Laboratory Reagents or packaging.
    SensitivityRemains as recommended by manufactureParameters were not tested, assumed to remain as recommended by manufacture as no modifications to STANBIO Laboratory Reagents or packaging.
    StabilityRemains as recommended by manufactureParameters were not tested, assumed to remain as recommended by manufacture as no modifications to STANBIO Laboratory Reagents or packaging.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Imprecision: Two control serums were used for both within-run and total precision.
      • Within-run: Up to 20 repeats.
      • Total precision: Duplicates for up to 20 days.
    • Correlation: "about 100 serums" were used.
    • Linearity: "Commercially available linearity material" was assayed.
    • Recovery: "Commercial available Controls with assigned values" were used.

    Data Provenance: The document does not specify the country of origin for the data or explicitly state if it was retrospective or prospective. However, the nature of the tests (using control serums, commercial linearity material, and patient serums for correlation by assaying them) suggests it was a prospective study conducted for the purpose of this 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not provided in the document. The "ground truth" for the test set appears to be established by comparing the performance of the STANBIO Laboratory Reagents on the ANALETTE™ to the performance of predicate reagents (Synermed® and Medical Analysis Systems Inc® Reagents) on the same ANALETTE™ or to manufacturer's claims for linearity and recovery. This is a comparison study, not a ground truthing exercise with independent experts reviewing clinical cases.

    4. Adjudication Method for the Test Set:

    This information is not applicable as the study described is a laboratory performance study comparing reagent efficacy, not a human reader or image-based diagnostic study requiring adjudication. The performance is assessed against established laboratory methods or manufacturer claims for the predicate reagents.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If So, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable as the device is a chemistry analyzer and reagents, not an AI-assisted diagnostic tool for human readers. No MRMC study was conducted.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done:

    This is an algorithm-only (standalone) performance study in the sense that it evaluates the analytical performance of the ANALETTE™ Chemistry Analyzer with STANBIO reagents. The purpose is to demonstrate that the device produces accurate measurement results independently. Human intervention is limited to operating the analyzer and interpreting the numerical output.

    7. The Type of Ground Truth Used:

    The "ground truth" in this context is established by:

    • Comparison to Predicate Devices/Reagents: For imprecision and correlation, the performance of the STANBIO reagents is compared to the performance of legally marketed Synermed® and Medical Analysis Systems Inc® reagents on the ANALETTE™ device (which effectively act as the reference standard).
    • Manufacturer's Claims/Expected Values: For linearity and recovery, the results are compared against the manufacturer's claims for the reagents or assigned values for commercial controls.

    This is therefore a form of comparative analytical performance against established and accepted methods/claims, rather than clinical outcomes or pathology reports.

    8. The Sample Size for the Training Set:

    This information is not applicable as the ANALETTE™ is a chemistry analyzer, not a machine learning or AI-based device that requires a "training set" in the conventional sense. The "training" for such a system would involve instrument calibration and quality control procedures, which are standard for laboratory devices.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable for the reasons stated above (not an AI/ML device requiring a training set with established ground truth).

    Ask a Question

    Ask a specific question about this device

    K Number
    K022072
    Date Cleared
    2002-08-23

    (58 days)

    Regulation Number
    862.1035
    Reference & Predicate Devices
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    | 862.1360

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Precision Systems™ ANALETTE™ Chemistry Analyzer is intended for the quantitative determination of Calcium, Creatinine, Phosphorus, Albumin, Total Protein, Glucose, Urea Nitrogen, Magnesium, Creatine Kinase, Alkaline Phosphatase, Carbon Dioxide, Amylase, Cholesterol(includes HDL), Triglycerides, Total Bilirubin, Direct Bilirubin, Uric Acid, Lactate Dehydrogenase L, Lactate Dehydrogenase P, Alanine Aminotransferase. Aspartate Aminotransferase; Gamma Glutamyl Transferase, Lipase, Chloride, and etc. analytes in solution such as serum, plasma, or urine. It is an "open" System, which can use a variety of commercially manufactured reagents such as but not limited to Synermeds® Reagents and Medical Analysis Systems Reagents. It is used to monitor various physiological diseases or conditions. Precision Systems Inc will distribute, recommend and sales MAS Reagents without any modification of MAS packaging using PSI Applications sheets.

    Device Description

    An in vitro diagnostic automated clinical chemistry analyzer for the analysis of analytes in solution.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the ANALETTE™ clinical chemistry analyzer and Medical Analysis Systems Reagents:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text describes a submission for substantial equivalence (510(k)) for the ANALETTE™ clinical chemistry analyzer using Medical Analysis Systems (MAS) Reagents, comparing it to the same ANALETTE™ analyzer using Synermeds® 072 reagents (the predicate device). The core of the acceptance criteria here is the demonstration of "substantial equivalence" of the new reagent system to the predicate. Specific quantitative acceptance criteria are not explicitly detailed in the provided text in the form of numerical thresholds for accuracy, precision, or comparison studies. Instead, the performance section broadly states:

    Acceptance CriteriaReported Device Performance
    Substantial equivalence to the predicate device (ANALYETTE™ with Synermeds® 072 reagents)"Substantially equivalence was established in comparative studies. It was concluded from these results that this product is safe and effective."
    Effective performance for the quantitative determination of various analytes in solution (serum, plasma, or urine).The device is intended for the quantitative determination of a comprehensive list of analytes (Calcium, Creatinine, Phosphorus, Albumin, Total Protein, Glucose, Urea Nitrogen, Magnesium, Creatine Kinase, Alkaline Phosphatase, Carbon Dioxide, Amylase, Cholesterol (includes HDL), Triglycerides, Total Bilirubin, Direct Bilirubin, Uric Acid, Lactate Dehydrogenase L, Lactate Dehydrogenase P, Alanine Aminotransferase, Aspartate Aminotransferase, Gamma Glutamyl Transferase, Lipase, Chloride). The statement of substantial equivalence implies effective performance.
    Safety of the device."It was concluded from these results that this product is safe and effective."

    2. Sample Size Used for the Test Set and Data Provenance

    The document states that "comparative studies" were conducted. However, it does not provide any details regarding the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective nature).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. For a clinical chemistry analyzer, the ground truth is typically established by reference methods or highly accurate laboratory instruments rather than expert adjudication in the way it would be for image-based diagnostics.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    This information is not applicable in the context of a clinical chemistry analyzer's performance evaluation as described. Ground truth is established through analytical measurements, not through human adjudication of diagnostic findings. Therefore, no adjudication method like 2+1 or 3+1 would be used.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done for this device. MRMC studies are typically used for diagnostic imaging devices where human interpretation is a critical component, often comparing human readers with and without AI assistance. This device is a clinical chemistry analyzer, which provides quantitative measurements directly.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The performance described for the ANALETTE™ clinical chemistry analyzer with MAS reagents is inherently a standalone performance in the context of the instrument measuring analyte concentrations. There is no "human-in-the-loop" performance component in the direct measurement by the analyzer. The comparison is between two reagent systems on the same analyzer, assessing the analytical performance.

    7. The Type of Ground Truth Used

    The ground truth for this type of device (a clinical chemistry analyzer) would typically be established by:

    • Reference standard methods: Highly accurate and precise laboratory methods, often more complex or expensive than routine clinical tests.
    • Certified reference materials: Samples with known, validated concentrations of the analytes.
    • Comparison to the predicate device: For a 510(k) submission seeking substantial equivalence, the performance of the new device (or reagent system) is directly compared to the legally marketed predicate device using patient samples and/or quality control materials. The predicate device's results serve as the pragmatic "ground truth" for demonstrating equivalence in a clinical setting.

    The document implies the latter, stating "Substantially equivalence was established in comparative studies," meaning performance was compared against the predicate system.

    8. The Sample Size for the Training Set

    This information is not provided in the document. Clinical chemistry analyzers and their associated reagents are developed through analytical validation, which involves extensive testing, but the term "training set" is more commonly associated with machine learning algorithms. If there were any computational models or algorithms within the analyzer's software that required training (which is not explicitly indicated as relevant here beyond basic instrument calibration), the details of such a training set are absent.

    9. How the Ground Truth for the Training Set Was Established

    Since no "training set" in the machine learning sense is explicitly mentioned or detailed, and the focus is on analytical performance comparison (substantial equivalence), the method for establishing ground truth for a training set is not applicable or provided. The development of a clinical chemistry reagent kit involves rigorous analytical validation, where performance characteristics like accuracy, precision, linearity, and interference are established using known standards and patient samples, rather than a "ground truth for training" in the way an AI model would be trained.

    Ask a Question

    Ask a specific question about this device

    K Number
    K020486
    Device Name
    CARESIDE GGT
    Manufacturer
    Date Cleared
    2002-07-01

    (138 days)

    Product Code
    Regulation Number
    862.1360
    Reference & Predicate Devices
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    CARESIDE® GGT

    GGT test system Clinical chemistry panel GGT test system Regulation Number: 21 CFR 862.1360
    Parkway Culver City, CA 90230

    Re: K020486 Trade/Device Name: Careside® GGT Regulation Number: 21 CFR 862.1360

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For in vitro diagnostic use with the CARESIDE Analyzer to quantitatively measure GGT from anti-coagulated whole blood, plasma, or serum specimens to aid in the diagnosis and treatment of liver diseases such as alcoholic cirrhosis and primary and secondary liver tumors.

    Device Description

    CARESIDE® GGT cartridges are used with the CARESIDE Analyze® to measure GGT activity in anti-coagulated whole blood, plasma or serum specimens. The CARESIDE GGT cartridge, a single use disposable in vitro diagnostic test cartridge, aids in specimen separation and delivers a measured volume of plasma or serum to a dry film to initiate the measurement of GGT activity. The patented film cartridge contains all reagents necessary to measure GGT activity.

    Each CARESIDE® GGT cartridge consists of a GGT-specific multi-layer reagent film mounted in a plastic base with a hinged lid. The user introduces the anti-coagulated whole blood, serum. or plasma specimen into the cartridge sample well, closes the lid and inserts the cartridge into the CARESIDE Analyzer.

    Once loaded, the CARESIDE Analyzer scans the cartridge barcode, brings the cartridge and the contained specimen to 37℃, and spins the cartridge to move the sample from the sample deposition well into the cartridge channels and chambers. As the cartridge continues to spin, the blood cells are separated from the plasma/serum and the cells accumulate in the separation well. Approximately 8.5 microliters of plasma (or serum, as applicable) remain in the metering passage. Any excess sample flows into an overflow well.

    The plasma (or serum, as applicable) is automatically dispensed onto the multi-layer reagent film. The spreading and substrate layer distributes the GGT containing specimen uniformly. The GGT in the specimen reacts with the substrate L-y-glutamyl-p-nitroanilide to release p-nitroaniline resulting in a change in film color. The rate of change of color intensity, as measured by the amount of reflected light at 425 nanometers, directly relates to the specimen GGT activity.

    As the cartridges spin, photodiodes measure reflectance of light emitted by wavelengthspecific light emitting diodes (LEDs) over a fixed time period. The instrument uses the reflectance measurements and the lot-specific standard curve to calculate GGT activity.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information for the CARESIDE® GGT device, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance (CARESIDE® GGT)Predicate Device Performance (Vitros GGT DT Slides)
    Detection limit20 U/L5 U/L
    Reportable Range20 to 1000 U/L5 to 1400 U/L
    AccuracyMean recovery 101%Not available (for predicate)
    PrecisionTotal CV, 344 U/L, 2.6%Total CV, 166 U/L, 2.4%
    Method ComparisonCARESIDE = 1.00 (BM/Hitachi 902 GGT) + 0.91 U/L, r= 1.00Not provided (for predicate)
    LinearityLinearity by dilution yielded slope and correlation coefficient within acceptable limits.Not available (for predicate)
    InterferenceNo significant interference observed at tested concentrations of Ascorbic Acid 10 mg/dL, Bilirubin 10 mg/dL, Triglycerides 3000 mg/dLNone stated (for predicate)
    Specimen Types & AnticoagulantsNo clinically significant difference between sodium heparinized whole blood, serum, and sodium heparin plasma.No clinically significant difference between serum, heparin plasma, or EDTA plasma. Whole blood is unsuitable.

    Note: The document explicitly states that the CARESIDE® GGT is "substantially equivalent in principle, intended use, and clinical performance to the currently marketed Vitros slides for the quantitative measurement of GGT on the Vitros DTSC 60 II." This broad statement acts as the overarching acceptance criterion, with the detailed performance characteristics providing the evidence for this claim. The differences (e.g., direct whole blood specimen, no accurate pipetting required, no reagent pre-warming required for CARESIDE® GGT) are presented as differences rather than failures of acceptance, implying they are either improvements or clinically insignificant for the intended use.

    Study Details

    1. Sample size used for the test set:

      • The document does not explicitly state the sample size (number of patient samples) used for the comparison studies (accuracy, precision, method comparison, interference, specimen type evaluation).
      • Data provenance: Not explicitly stated, but clinical data is implied for "Method Comparison" and "Specimen Types & Anticoagulants" by comparing against other analyzers and evaluating different sample types. Given the context of a 510(k) summary, it's highly likely to be prospective testing for the CARESIDE device.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not provided in the document. For an in vitro diagnostic device measuring an analyte (GGT), the "ground truth" is typically established by reference laboratory methods or a predicate device.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • This is not applicable for this type of in vitro diagnostic device study. Adjudication methods like 2+1 or 3+1 are typically used in imaging or clinical trials involving subjective expert interpretation to resolve discrepancies. For quantitative laboratory tests, the "ground truth" is typically a quantitative value from a reference method or predicate device, not subject to subjective expert interpretation requiring adjudication.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • This is not applicable. The CARESIDE® GGT is an in vitro diagnostic device to measure GGT activity, not an AI-assisted diagnostic tool requiring human interpretation. Therefore, an MRMC study or assessment of human reader improvement with AI assistance is not relevant.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Yes, the performance characteristics (accuracy, precision, linearity, interference, method comparison) are inherently "standalone" in that they describe the performance of the CARESIDE® GGT device and analyzer system directly, without requiring human-in-the-loop performance assessment to determine the GGT value. The device provides a quantitative result.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for the CARESIDE® GGT performance evaluation is established by:
        • Reference laboratory methods: For the "Method Comparison," it states "CARESIDE = 1.00 (BM/Hitachi 902 GGT) + 0.91 U/L, r= 1.00", indicating a comparison against a commercially available clinical chemistry analyzer (BM/Hitachi 902 GGT), which serves as a reference or predicate method.
        • Known concentrations: For "Accuracy" (mean recovery 101%) and "Linearity" (by dilution), the ground truth would be known concentrations of GGT in control materials or diluted samples.
        • Predicate device comparison: The overall claim of substantial equivalence to the Vitros GGT DT Slides implies that the Vitros system values served as a de facto "ground truth" or comparator for aspects of clinical performance.
    7. The sample size for the training set:

      • The document does not specify a separate "training set" or its sample size. For an IVD device like this, the calibration data (bar-coded on each cartridge) effectively serves as the "training" for the device to interpret raw signals into GGT activity. However, this is distinct from statistical machine learning model training sets.
    8. How the ground truth for the training set was established:

      • As above, explicitly defined "training set" in a machine learning sense is not described. However, for the calibration of the device, the ground truth would have been established using calibrators with known GGT concentrations, typically verified against reference methods. The document mentions "Calibration information bar-coded on each cartridge. Calibration information may change with each lot." This indicates that each lot is calibrated, likely against reference standards, to ensure accurate GGT measurement.
    Ask a Question

    Ask a specific question about this device

    K Number
    K011720
    Date Cleared
    2001-06-25

    (21 days)

    Regulation Number
    862.1035
    Reference & Predicate Devices
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    862.1145 862.1160 862.1170 862.1175 866.5240 866.5240 866.5270 862.1215 862.1215 862.1225 866.5340 862.1360

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Olympus AU5400 Clinical Chemistry Analyzer is a fully automated photometric analyzer intended for clinical laboratory use. Applications include colorimetric, turbidimetric, latex agglutination, and homogeneous enzyme immunoassay.

    Device Description

    The Olympus AU5400 Clinical Chemistry Analyzer is a fully automated photometric analyzer.

    AI/ML Overview

    While the provided document is a 510(k) clearance letter for the Olympus AU5400 Clinical Chemistry Analyzer, it does not contain the detailed performance study results, acceptance criteria, or ground truth information typically found in the actual 510(k) submission or a scientific publication.

    The letter confirms that the device has been found substantially equivalent to predicate devices, meaning it is considered safe and effective for its indicated use. However, it does not explicitly state the specific performance metrics (like sensitivity, specificity, accuracy), the thresholds for acceptance of those metrics, or the specifics of the validation study.

    Therefore, I cannot populate all the requested fields from the given text. I can only infer some information based on the nature of a 510(k) submission for a clinical chemistry analyzer.

    Here's what I can convey based on the provided document and general understanding of 510(k) submissions for similar devices:

    1. Table of Acceptance Criteria and Reported Device Performance

    • Acceptance Criteria: Not explicitly stated in the provided letter. For a clinical chemistry analyzer, acceptance criteria would typically involve demonstrating analytical performance similar to or better than a predicate device across various parameters, including:

      • Accuracy: Agreement with a reference method.
      • Precision (Reproducibility & Repeatability): Consistency of results.
      • Linearity: Accuracy across the analytical measurement range.
      • Detection Limits: Lowest concentration that can be reliably measured.
      • Interference: Lack of significant impact from common interfering substances.
      • Carry-over: Minimal contamination between samples.
      • Stability: Reagent and calibration stability.
      • Correlation: Strong correlation with predicate device or reference method.
    • Reported Device Performance: Not explicitly stated in the provided letter. The 510(k) submission would have contained data supporting these performance characteristics, demonstrating that the device meets the established acceptance criteria. The FDA's clearance implies that this evidence was found satisfactory.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not specified in the provided letter. For a clinical chemistry analyzer, test sets would include a variety of patient samples (normal, abnormal) and spiked samples to assess different analytical aspects.
    • Data Provenance: Not specified in the provided letter. Typically, clinical chemistry analyzer validation involves prospective collection of patient samples, often from multiple sites to ensure representativeness, as well as characterization of control materials.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Experts and Qualifications: Not specified in the provided letter. For clinical chemistry analyzers, "ground truth" for analytical performance is typically established through:
      • Reference interval studies: Involving a statistically significant number of healthy individuals.
      • Comparison studies: Against a recognized reference method or a legally marketed predicate device, where the predicate device's results serve as the comparison standard.
      • Control materials and calibrators: With known, certified values.
      • Analytical experts (e.g., clinical chemists, laboratory directors) would be involved in designing and overseeing these studies, and interpreting the results.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable in the traditional sense for analytical performance of a clinical chemistry analyzer. Adjudication methods (like 2+1, 3+1) are typically used for subjective interpretations, such as image analysis or pathology review, where expert opinion is directly establishing "ground truth." For an automated analyzer, the output is quantitative, and performance is assessed against established analytical standards or comparison methods.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • MRMC Study: Not applicable. MRMC studies are used to evaluate human reader performance, often with AI assistance, for tasks involving interpretation (e.g., radiology). The Olympus AU5400 is an automated clinical chemistry analyzer that produces quantitative results, not an AI-assisted diagnostic imaging tool with human interpretation.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • Standalone Performance: As an automated analyzer, the device's performance is inherently "standalone" in generating the quantitative results. The entire 510(k) submission would be focused on demonstrating this standalone analytical performance. However, there's no "algorithm only without human-in-the-loop" contrast needed, as the device's function is to perform the chemical analysis automatically.

    7. The Type of Ground Truth Used

    • Ground Truth Type: For a clinical chemistry analyzer, the "ground truth" is typically established through:
      • Reference methods: Highly accurate and validated analytical methods.
      • Certified reference materials/calibrators: Materials with known, traceable analyte concentrations.
      • Comparison to a legally marketed predicate device: Demonstrating equivalent performance to a device already on the market.
      • Pathology/Outcomes data: Would generally not be the primary "ground truth" for the analytical performance of the analyzer itself, though the results generated by the analyzer would be used in conjunction with such data for clinical decision-making.

    8. The Sample Size for the Training Set

    • Training Set Sample Size: Not applicable in the conventional machine learning sense. This device is a traditional analytical instrument, not a machine learning or AI model that requires a "training set" to learn its function. Its operational parameters are determined by its design, engineering tolerances, and chemical principles, not by training on a dataset.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set: Not applicable, as there is no "training set" for a traditional clinical chemistry analyzer. The device's calibration involves using calibrator materials with known concentrations, but this is part of routine operation and quality control, not "training" in the ML sense.
    Ask a Question

    Ask a specific question about this device

    K Number
    K980757
    Date Cleared
    1998-04-21

    (53 days)

    Regulation Number
    862.1170
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    | II |
    | 862.1345 | Glucose test system | I |
    | 862.1360

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ILab 600 is an automated, random access clinical chemistry analyzer which uses analytical techniques (photometry and potentiometry) for the in vitro quantitation of analytes found in physiological fluids such as serum, plasma, urine and cerebrospinal fluid. The results of the measurements are used as medical diagnostic tools.

    Device Description

    The ILab 600 is an automated, random access clinical chemistry analyzer which uses analytical techniques (photometry and potentiometry) for the in virro quantitation of analytes found in physiological fluids such as serum, plasma, urine and cerebrospinal fluid. The results of the measurements are used as medical diagnostic tools.

    AI/ML Overview

    The ILab 600 Clinical Chemistry System is an automated, random access clinical chemistry analyzer that quantifies analytes in physiological fluids using photometry and potentiometry. The device was found substantially equivalent to the ILab 900/1800 Clinical Chemistry System (K932467, K943595) and IL Test assays (K943366, K952646, K943367, K952647).

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the ILab 600 are implicit in its claim of substantial equivalence to the predicate device, the ILab 900. This means the performance of the ILab 600 must be "statistically similar" to that of the ILab 900 across various analytes and sample types (serum, urine, cerebrospinal fluid).

    The reported device performance is demonstrated through method comparison studies and precision studies.

    Method Comparison Studies (ILab 600 vs. ILab 900):

    IL TestUnitsnRangeSlope (IL600 vs. IL900)Interceptr (Correlation Coefficient)
    Serum Samples
    Acid Phosp., Non-ProstaticU/L970.3-27.60.8670.380.986
    Acid Phosp., TotalU/L940.4-33.40.9570.1970.996
    Albuming/dL982.9-5.51.047-0.1830.985
    Alkaline PhosphataseU/L10824-6231.0525.940.999
    ALT/GPTU/L1092-25571.006-1.40.999
    AmylaseU/L11025-3771.0400.10.997
    AST/GOTU/L11514-23771.0650.10.998
    Bilirubin, Directmg/dL1000.03-15.190.989-0.0290.999
    Bilirubin, Totalmg/dL990.02-26.520.9550.0340.999
    Calciummg/dL986.5-15.61.040-0.0280.990
    Cholesterolmg/dL11740-9441.0052.3340.997
    CholinesteraseU/L1072166-126921.002195.40.990
    CK/CPKU/L10318-37590.9336.720.998
    CK-MBU/L1100.6-237.21.003-1.40.997
    Creatininemg/dL990.8-7.11.0290.1470.998
    Glucose Hexokinasemg/dL11360-4571.0170.3580.997
    Glucose Oxidasemg/dL13751-3930.9447.920.997
    γ-GTU/L1224-4971.0521.20.999
    Ironµg/dL9710-2531.0402.370.998
    LD-L/LDH-LU/L9545-4040.9973.560.992
    LipaseU/L648-27190.972-1.70.999
    Magnesiummg/dL1031.60-8.240.9880.010.994
    Phosphorusmg/dL1002.5-11.30.9760.060.998
    TCO2mmol/L10210-361.079-1.430.987
    Total Proteing/dL984.6-8.80.9660.160.992
    Triglyceridesmg/dL9637-10390.9781.4140.999
    Urea Nitrogenmg/dL1197.0-68.01.007-0.0150.998
    Uric Acidmg/dL991.9-15.90.9630.110.995
    ISE Chloridemmol/L9036.8-143.21.028-1.390.998
    ISE Potassiummmol/L792.0-7.31.013-0.040.999
    ISE Sodiummmol/L9062.4-157.41.0111.390.999
    Urine Samples
    AmylaseU/L6526-60680.953-210.999
    Calciummg/dL7020-920.923-0.070.995
    Creatininemg/dL5949-2631.0654.20.992
    Glucose Hexokinasemg/dL954-6901.0272.550.996
    Glucose Oxidasemg/dL800-8010.9494.290.997
    Phosphorusmg/dL6036-1610.953-2.30.980
    Urea Nitrogenmg/dL58200-16491.06013.70.992
    Uric Acidmg/dL709-910.9891.840.997
    ISE Chloridemmol/L5073-2491.042-5.740.998
    ISE Potassiummmol/L4919-851.083-1.30.999
    ISE Sodiummmol/L4973-1941.0005.360.999
    Cerebrospinal Fluid Samples
    Glucose Oxidasemg/dL2040-2260.932-0.1171.000

    Precision Studies (ILab 600):

    • Serum Samples: Two levels of serum samples (three for Cholesterol) were tested in triplicate twice a day for 10 days (n=60 total). The Total %CV for most analytes was generally low, indicating good precision. For example:
      • Albumin: Level 1 (1.79%), Level 2 (1.08%)
      • ALT/GPT: Level 1 (1.26%), Level 2 (0.99%)
      • Cholesterol: Level 1 (2.11%), Level 2 (1.35%), Level 3 (1.37%)
      • ISE Sodium: Level 1 (0.94%), Level 2 (0.67%)
    • Urine Samples: Two levels of urine samples were tested in triplicate twice a day for 10 days (n=60 total). Similar to serum, Total %CV remained low:
      • Amylase: Level 1 (2.56%), Level 2 (2.02%)
      • Creatinine: Level 1 (2.11%), Level 2 (1.73%)
      • ISE Sodium: Level 1 (1.13%), Level 2 (0.65%)
    • Cerebrospinal Fluid Samples: Two levels of CSF samples were tested using IL Test Glucose Oxidase in triplicate twice a day for 5 days (n=30 total).
      • Glucose Oxidase: Level 1 (1.49%), Level 2 (0.98%)

    The studies conclude that the ILab 600 and ILab 900 are "statistically similar" for the tests evaluated. The precision studies demonstrate acceptable levels of reproducibility based on the reported Coefficient of Variation (%CV) values for within-run, among-run, among-day, and total precision. No specific numerical thresholds for acceptance criteria were explicitly stated, but the robust statistical similarity and low %CV values imply the device meets the necessary performance standards for clinical use.

    2. Sample Sizes and Data Provenance

    • Test Set (Method Comparison):

      • Serum Samples: Sample sizes ranged from a minimum of 64 (Lipase) to a maximum of 137 (Glucose Oxidase).
      • Urine Samples: Sample sizes ranged from 49 (ISE Potassium and Sodium) to 95 (Glucose Hexokinase).
      • Cerebrospinal Fluid Samples: Sample size was 20 (Glucose Oxidase).
      • Data Provenance: The document does not specify the country of origin of the data or whether it was retrospective or prospective.
    • Test Set (Precision Studies):

      • Serum Samples: n=60 for most analytes (triplicate measurements, twice a day, for 10 days). Cholesterol used n=60 across three levels.
      • Urine Samples: n=60 for all analytes (triplicate measurements, twice a day, for 10 days).
      • Cerebrospinal Fluid Samples: n=30 for Glucose Oxidase (triplicate measurements, twice a day, for 5 days).
      • Data Provenance: The document does not specify the country of origin of the data or whether it was retrospective or prospective.

    3. Number of Experts and Qualifications for Ground Truth

    The document pertains to the performance characteristics of a clinical chemistry analyzer, which measures quantitative values of analytes. The "ground truth" in this context refers to the actual concentration of the analytes in the samples. Clinical chemistry analyzer performance is typically evaluated by comparing results to a reference method (in this case, the predicate device ILab 900) or by using certified reference materials with known concentrations. Therefore, expert interpretation or consensus, as might be used in image-based diagnostic AI, is not applicable here. No mention of human experts defining "ground truth" is provided or expected.

    4. Adjudication Method

    Not applicable. As described in point 3, this is a quantitative measurement device, not an interpretation task requiring adjudication of expert opinions.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. The device is a clinical chemistry analyzer, not an AI system assisting human readers in diagnostic interpretation. The study evaluates the analyzer's performance directly against a predicate device and for precision, not the human reader's effectiveness with or without AI.

    6. Standalone Performance Study

    Yes, standalone performance was done.

    • Method Comparison Studies: The performance of the ILab 600 was compared directly to that of the ILab 900 (predicate device). This is a standalone comparison of the new device against an established one.
    • Precision Studies: The precision of the ILab 600 was evaluated independently, measuring its reproducibility across different runs and days. This is also a standalone performance evaluation of the device itself.

    7. Type of Ground Truth Used

    The ground truth used for these studies is the quantitative analytical result obtained from the predicate device (ILab 900) for method comparison studies, and the inherent, measured concentration within the biological samples for precision studies. This is a form of reference method comparison or analytical accuracy assessment, rather than pathology, expert consensus, or outcomes data, which are typically associated with qualitative or interpretative diagnostics. For precision, the ground truth is implicitly the true, stable concentration in the control/patient samples being repeatedly measured.

    8. Sample Size for the Training Set

    Not applicable. The ILab 600 is a clinical chemistry analyzer, which operates based on established chemical and photometric/potentiometric principles and internal calibration curves. It is not an AI/ML device that requires a "training set" in the sense of supervised learning. Its analytical methods are pre-programmed and validated, not learned from data in the way a machine learning algorithm would be.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no "training set" for this type of device. The device's operational parameters and calibration are established through engineering design and standard laboratory calibration procedures, not through a data-driven training process.

    Ask a Question

    Ask a specific question about this device

    K Number
    K973869
    Date Cleared
    1998-01-07

    (105 days)

    Regulation Number
    862.1030
    Reference & Predicate Devices
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    mellitus,neonatal hypoglycemia , idiopathic hypoglycemia and pancreatic islet cell carcinoma. " CFR 862.1360

    Intended Use :

    mellitus,neonatal hypoglycemia , idiopathic hypoglycemia and pancreatic islet cell carcinoma. " CFR 862.1360

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for the the In Vitro, quantitative determination of ALT (L-Alanine;2-Oxoglutarate Aminotransferase EC2.6.1.2) in human serum on the Cobas MIRA® clinical chemistry system. This alanine aminotransferase (ALT/SGPT) test system is a device intended to measure the activity of the enzyme alanine aminotransferase (ALT) (also known as a serum glutamic pyruvic transaminase or SGPT) in serum and plasma. Alanine aminotransferase measurements are used in the diagnosis and treatment of certain liver diseases (e.g. viral hepatitus and cirrhosis) and heart diseases. CFR 862.1030

    Intended for the the In Vitro, quantitative determination of KAMYLASE (1,4-%- D glucan glucanohydrolase EC3.2.1.1) in human serum and urine on the Cobas MIRA ® clinical chemistry system. This Amylase test system is a device intended to measure the activity of the enzyme Amylase in serum and urine . Amylase measurements are used primarily for the diagnosis and treatment of pancreatitis (inflammation of the pancreas) CFR 862.1070

    Intended for the the In Vitro, quantitative determination of AST (Aspartate aminotransferase EC2.6.1.1) in human serum on the Cobas MIRA ® clinical chemistry system. This "aspartate aminotransferase (AST/SGOT) test system is a device intended to measure the activity of the enzyme aspartate aminotransferase (AST) (also known as a serum glutamic oxaloacetic transferase or SGOT) in serum and plasma. Aspartate aminotransferase measurements are used in the diagnosis and treatment of certain types of liverand heart disease." CFR 862.1100

    Intended for the the In Vitro, quantitative determination of Direct Bilirubin in human serum on the Cobas MIRA ® clinical chemistry system. This "Direct Bilirubin test system is a device intended to measure the levels of bilirubin (direct) in plasma or serum. Measurements of levels of bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, if used in the diagnosis and treatment of the liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block." CFR 862.1110

    Intended for the the In Vitro, quantitative determination of Total Bilirubin in human serum on the Cobas MIRA ® clinical chemistry system. This "Total Bilirubin test system is a device intended to measure the levels of bilirubin (Total) in plasma or serum. Measurements of levels of bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, if used in the diagnosis and treatment of the liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block." CFR 862.1110

    Intended for the the In Vitro, quantitative determination of Calcium in human serum or urine on the Cobas MIRA ® clinical chemistry system. This "Calcium test system is a device intended to measure the total calcium level in serum. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms" CFR 862.1145

    Intended for the the In Vitro, quantitative determination of Total CO2 in human serum on the Cobas MIRA ® clinical chemistry system. This " bicarbonatel carbon dioxide test system is a device intended to measure the total bicarbonate/carbon dioxide plasma, serum and whole blood. Bicarbonate/carbon dioxide measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid base balance. " CFR 862.1160

    Intended for the the In Vitro, quantitative determination of Chloride in human serum on the Cobas MIRA ® clinical chemistry system. This " Chloride test system is a device intended to measure the Chloride in plasma, serum, sweat and urine. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis. " CFR 862.1170

    Intended for the the In Vitro, quantitative determination of Cholesterol in human serum on the Cobas MIRA ® clinical chemistry system. This " Cholesterol test system is a device intended to measure the Cholesterol in plasma and serum. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. " CFR 862.1175

    Intended for the In Vitro, quantitative determination of Creatinine in human serum and urine on the Cobas MIRA ® clinical chemistry system. This " Creatinine test system is a device intended to measure the Creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases,in monitoring renal dialysis, and a calculation basis for measuring other urine analytes. " CFR 862.1225

    Intended for the the In Vitro, quantitative determination of Glucose in human serum and urine on the Cobas MIRA ® clinical chemistry system. This " Glucose test system is a device intended to measure the Glucose levels in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus,neonatal hypoglycemia , idiopathic hypoglycemia and pancreatic islet cell carcinoma. " CFR 862.1360

    Intended for the the In Vitro, quantitative determination of Iron in human serum on the Cobas MIRA ® clinical chemistry system. This " Iron (non-heme) test system is a device intended to measure Iron (non-heme)in serum and plasma. Iron (non-heme) measurements are used in the diagnosis and treatment of diseases such as iron deficiency anemia , hemochromatosis ( a disease associated with widespread deposit in the tissues of two iron containing pigments, hemosiderin and hemofuscin, and characterized by pigmentation of the skin) and chronic renal disease. " CFR 862.1410

    Intended for the In Vitro, quantitative determination of Magnesium in human serum and urine on the Cobas MIRA ® clinical chemistry system. This " Magnesium test system is a device intended to measure magnesium levels in serum and plasma. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low plasma levels of magnesium) and hypermagnasemia (abnormally high levels of magnesium). " CFR 862.1495

    Intended for the the In Vitro, quantitative determination of Inorganic Phosphorus in human serum or urine on the Cobas MIRA ® clinical chemistry system. This " Phosphorus (inorganic) test system is a device intended to measure inorganic phosphorus in serum, plasma and urine. Measurements of phosphorus (inorganic) are used in the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases , and vitamin D imbalance. " CFR 862.1580

    Intended for the the In Vitro, quantitative determination of Total Protein in human serum on the Cobas MIRA ® clinical chemistry system. This " Total Protein test system is a device intended to measure Total protein(s) in serum or plasma. Measurements obtained by this device are used in the diagnosis and treatment of a variety of diseases involving the liver, kidney or bone marrow as well as other metabolic or nutritional disorders " CFR 862. 1635

    Intended for the the In Vitro, quantitative determination of Triglycerides in human serum on the Cobas MIRA ® clinical chemistry system. This " Triglyceride test system is a device intended to measure Triglyceride (neutral fat) in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders. " CFR 862.1705

    Intended for the the In Vitro, quantitative determination of Urea Nitrogen in human serum or urine on the Cobas MIRA ® clinical chemistry system. This " urea nitrogen test system is a device intended to measure urea nitrogen (an end product of nitrogen metabolism) in whole blood, serum or plasma. Measurements obtained by this device are used in the diagnosis and treatment of certain renal and metabolic diseases. " CFR 862.1770

    Intended for the the In Vitro, quantitative determination of Unbound Iron-Binding Capacity (UIBC) in human serum on the Cobas MIRA ® clinical chemistry system. This " Iron-binding capacity test system is a device intended to measure Iron-binding capacity in serum. Iron-binding capacity measurements are used in the diagnosis and treatment of anemia. " CFR 862.1415

    Intended for the the In Vitro, quantitative determination of Uric Acid in human serum or urine on the Cobas MIRA ® clinical chemistry system. This " Uric Acid test system is a device intended to measure uric acid in serum, plasma or urine. Measurements obtained by this device are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions, and of patients receiving cytotoxic drugs. " CFR 862.1775

    Device Description

    Not Found

    AI/ML Overview

    The provided documents are Indications for Use (IFU) statements and an FDA 510(k) clearance letter for the TRACE Reagent Line for the Cobas MIRA. These documents state the intended use and regulatory classification of the device but do not contain any information regarding acceptance criteria, study design, reported device performance, sample sizes, ground truth establishment, or expert qualifications.

    Therefore, I cannot provide the requested information based on the provided text. The documents merely confirm that the TRACE Reagent Line for the Cobas MIRA is cleared for marketing as substantially equivalent to existing devices for the in vitro quantitative determination of various analytes (ALT, Amylase, AST, Direct Bilirubin, Total Bilirubin, Calcium, Total CO2, Chloride, Cholesterol, Creatinine, Glucose, Iron, Magnesium, Inorganic Phosphorus, Total Protein, Triglycerides, Urea Nitrogen, Unbound Iron-Binding Capacity, and Uric Acid) on the Cobas MIRA clinical chemistry system.

    To answer your request, detailed performance study reports, often found in the 510(k) submission itself (which is typically not publicly available in its entirety without a FOIA request), or in peer-reviewed publications, would be necessary.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2