K Number
K020486
Device Name
CARESIDE GGT
Manufacturer
Date Cleared
2002-07-01

(138 days)

Product Code
Regulation Number
862.1360
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For in vitro diagnostic use with the CARESIDE Analyzer to quantitatively measure GGT from anti-coagulated whole blood, plasma, or serum specimens to aid in the diagnosis and treatment of liver diseases such as alcoholic cirrhosis and primary and secondary liver tumors.

Device Description

CARESIDE® GGT cartridges are used with the CARESIDE Analyze® to measure GGT activity in anti-coagulated whole blood, plasma or serum specimens. The CARESIDE GGT cartridge, a single use disposable in vitro diagnostic test cartridge, aids in specimen separation and delivers a measured volume of plasma or serum to a dry film to initiate the measurement of GGT activity. The patented film cartridge contains all reagents necessary to measure GGT activity.

Each CARESIDE® GGT cartridge consists of a GGT-specific multi-layer reagent film mounted in a plastic base with a hinged lid. The user introduces the anti-coagulated whole blood, serum. or plasma specimen into the cartridge sample well, closes the lid and inserts the cartridge into the CARESIDE Analyzer.

Once loaded, the CARESIDE Analyzer scans the cartridge barcode, brings the cartridge and the contained specimen to 37℃, and spins the cartridge to move the sample from the sample deposition well into the cartridge channels and chambers. As the cartridge continues to spin, the blood cells are separated from the plasma/serum and the cells accumulate in the separation well. Approximately 8.5 microliters of plasma (or serum, as applicable) remain in the metering passage. Any excess sample flows into an overflow well.

The plasma (or serum, as applicable) is automatically dispensed onto the multi-layer reagent film. The spreading and substrate layer distributes the GGT containing specimen uniformly. The GGT in the specimen reacts with the substrate L-y-glutamyl-p-nitroanilide to release p-nitroaniline resulting in a change in film color. The rate of change of color intensity, as measured by the amount of reflected light at 425 nanometers, directly relates to the specimen GGT activity.

As the cartridges spin, photodiodes measure reflectance of light emitted by wavelengthspecific light emitting diodes (LEDs) over a fixed time period. The instrument uses the reflectance measurements and the lot-specific standard curve to calculate GGT activity.

AI/ML Overview

Here's an analysis of the acceptance criteria and study information for the CARESIDE® GGT device, based on the provided text:

Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance (CARESIDE® GGT)Predicate Device Performance (Vitros GGT DT Slides)
Detection limit20 U/L5 U/L
Reportable Range20 to 1000 U/L5 to 1400 U/L
AccuracyMean recovery 101%Not available (for predicate)
PrecisionTotal CV, 344 U/L, 2.6%Total CV, 166 U/L, 2.4%
Method ComparisonCARESIDE = 1.00 (BM/Hitachi 902 GGT) + 0.91 U/L, r= 1.00Not provided (for predicate)
LinearityLinearity by dilution yielded slope and correlation coefficient within acceptable limits.Not available (for predicate)
InterferenceNo significant interference observed at tested concentrations of Ascorbic Acid 10 mg/dL, Bilirubin 10 mg/dL, Triglycerides 3000 mg/dLNone stated (for predicate)
Specimen Types & AnticoagulantsNo clinically significant difference between sodium heparinized whole blood, serum, and sodium heparin plasma.No clinically significant difference between serum, heparin plasma, or EDTA plasma. Whole blood is unsuitable.

Note: The document explicitly states that the CARESIDE® GGT is "substantially equivalent in principle, intended use, and clinical performance to the currently marketed Vitros slides for the quantitative measurement of GGT on the Vitros DTSC 60 II." This broad statement acts as the overarching acceptance criterion, with the detailed performance characteristics providing the evidence for this claim. The differences (e.g., direct whole blood specimen, no accurate pipetting required, no reagent pre-warming required for CARESIDE® GGT) are presented as differences rather than failures of acceptance, implying they are either improvements or clinically insignificant for the intended use.

Study Details

  1. Sample size used for the test set:

    • The document does not explicitly state the sample size (number of patient samples) used for the comparison studies (accuracy, precision, method comparison, interference, specimen type evaluation).
    • Data provenance: Not explicitly stated, but clinical data is implied for "Method Comparison" and "Specimen Types & Anticoagulants" by comparing against other analyzers and evaluating different sample types. Given the context of a 510(k) summary, it's highly likely to be prospective testing for the CARESIDE device.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not provided in the document. For an in vitro diagnostic device measuring an analyte (GGT), the "ground truth" is typically established by reference laboratory methods or a predicate device.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • This is not applicable for this type of in vitro diagnostic device study. Adjudication methods like 2+1 or 3+1 are typically used in imaging or clinical trials involving subjective expert interpretation to resolve discrepancies. For quantitative laboratory tests, the "ground truth" is typically a quantitative value from a reference method or predicate device, not subject to subjective expert interpretation requiring adjudication.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This is not applicable. The CARESIDE® GGT is an in vitro diagnostic device to measure GGT activity, not an AI-assisted diagnostic tool requiring human interpretation. Therefore, an MRMC study or assessment of human reader improvement with AI assistance is not relevant.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, the performance characteristics (accuracy, precision, linearity, interference, method comparison) are inherently "standalone" in that they describe the performance of the CARESIDE® GGT device and analyzer system directly, without requiring human-in-the-loop performance assessment to determine the GGT value. The device provides a quantitative result.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for the CARESIDE® GGT performance evaluation is established by:
      • Reference laboratory methods: For the "Method Comparison," it states "CARESIDE = 1.00 (BM/Hitachi 902 GGT) + 0.91 U/L, r= 1.00", indicating a comparison against a commercially available clinical chemistry analyzer (BM/Hitachi 902 GGT), which serves as a reference or predicate method.
      • Known concentrations: For "Accuracy" (mean recovery 101%) and "Linearity" (by dilution), the ground truth would be known concentrations of GGT in control materials or diluted samples.
      • Predicate device comparison: The overall claim of substantial equivalence to the Vitros GGT DT Slides implies that the Vitros system values served as a de facto "ground truth" or comparator for aspects of clinical performance.
  7. The sample size for the training set:

    • The document does not specify a separate "training set" or its sample size. For an IVD device like this, the calibration data (bar-coded on each cartridge) effectively serves as the "training" for the device to interpret raw signals into GGT activity. However, this is distinct from statistical machine learning model training sets.
  8. How the ground truth for the training set was established:

    • As above, explicitly defined "training set" in a machine learning sense is not described. However, for the calibration of the device, the ground truth would have been established using calibrators with known GGT concentrations, typically verified against reference methods. The document mentions "Calibration information bar-coded on each cartridge. Calibration information may change with each lot." This indicates that each lot is calibrated, likely against reference standards, to ensure accurate GGT measurement.

{0}------------------------------------------------

CARESIDE, Inc. Page 1

CARESIDE GGT Premarket Notification 1 2002 JUL

May 10, 2002

K020486

510(K) SUMMARY: CARESIDE® GGT SAFETY AND EFFECTIVENESS

I. Applicant Information

  • A. Applicant Name
  • B. Applicant/Manufacturer Address
  • C. Telephone Number
  • D. Contact Person
  • E. FAX Number
  • F. e-Mail Address
  • Date 510(k) Summary prepared G.

II. Device Information

  • A. Device Name (Trade)
  • B. Device Name (Classification)
  • C. Device Classification

CARESIDE, Inc. 6100 Bristol Parkway Culver City, CA 90230 310-338-6767 Renate A. MacLaren, Ph.D. 310-670-6986 rmaclaren@CARESIDE.com May 10, 2002

CARESIDE® GGT

GGT test system Clinical chemistry panel GGT test system Regulation Number: 21 CFR 862.1360 Regulatory Class 1 (non-exempt for point of care use) Classification Number: 75JPZ None applicable

D. Special controls and performance standards

III. Substantial Equivalence Claim

A. General equivalency claim

The ability to monitor analyte-specific biochemical reactions in dry film and other formats is widely recognized and has gained widespread acceptance for use in chemistry assays.

GGT in vitro diagnostic products, in both dry film and other formats, are currently marketed in the U.S., including GGT products which utilize GGT catalyzed generation of p-nitroaniline from the artificial substrate L-y-glutamyl-p-nitroanilide.

B. Specific equivalency claim

The CARESIDE® GGT test is substantially equivalent in principle, intended use, and clinical performance to the currently marketed Vitros slides for the quantitative measurement of GGT on the Vitros DTSC 60 II.

Name of Predicate Device: Johnson and Johnson's (formerly Eastman Kodak, Inc.) Vitros GGT Slides for Johnson and Johnson's Vitros DTSC 60 (formerly Eastman Kodak's DTSC 60 II).

Predicate Device 510K number: K912844/A Product Code: 75JPZ

{1}------------------------------------------------

IV. Device Description

CARESIDE® GGT cartridges are used with the CARESIDE Analyze® to measure GGT activity in anti-coagulated whole blood, plasma or serum specimens. The CARESIDE GGT cartridge, a single use disposable in vitro diagnostic test cartridge, aids in specimen separation and delivers a measured volume of plasma or serum to a dry film to initiate the measurement of GGT activity. The patented film cartridge contains all reagents necessary to measure GGT activity.

A. Explanation of Device Function

Each CARESIDE® GGT cartridge consists of a GGT-specific multi-layer reagent film mounted in a plastic base with a hinged lid. The user introduces the anti-coagulated whole blood, serum. or plasma specimen into the cartridge sample well, closes the lid and inserts the cartridge into the CARESIDE Analyzer.

Once loaded, the CARESIDE Analyzer scans the cartridge barcode, brings the cartridge and the contained specimen to 37℃, and spins the cartridge to move the sample from the sample deposition well into the cartridge channels and chambers. As the cartridge continues to spin, the blood cells are separated from the plasma/serum and the cells accumulate in the separation well. Approximately 8.5 microliters of plasma (or serum, as applicable) remain in the metering passage. Any excess sample flows into an overflow well.

The plasma (or serum, as applicable) is automatically dispensed onto the multi-layer reagent film. The spreading and substrate layer distributes the GGT containing specimen The GGT in the specimen reacts with the substrate L-y-glutamyl-puniformly. nitroanilide to release p-nitroaniline resulting in a change in film color. The rate of change of color intensity, as measured by the amount of reflected light at 425 nanometers, directly relates to the specimen GGT activity.

Test Reaction Sequence:

L-Y-glutamyl-p-nitroanilide + glycylglycine

GGT ↓

L-glutamylglycylglycine + p-nitroaniline

As the cartridges spin, photodiodes measure reflectance of light emitted by wavelengthspecific light emitting diodes (LEDs) over a fixed time period. The instrument uses the reflectance measurements and the lot-specific standard curve to calculate GGT activity.

B. Test Summary

y-Glutamyltransferase measurements are used in the diagnosis and treatment of liver diseases including alcoholic cirrhosis and primary and secondary liver tumors. GGT in the serum originates primarily from the hepatobiliary system even though renal tissue contains the highest levels of GGT. GGT is elevated in all forms of liver disease. GGT elevations are observed earlier and are more pronounced than those of other liver enzymes in cases of obstructive jaundice and metastatic neoplasms. In intra- or posthepatic biliary obstruction, GGT levels may reach levels 5 to 30 times higher than in intra- or post-hepatic biliary obstruction

{2}------------------------------------------------

V. Intended Use

A. Intended Use

The CARESIDE® GGT cartridge is intended for in vitro diagnostic use in conjunction with the CARESIDE® Analyzer to quantitatively measure GGT activity in anticoagulated whole blood, plasma or serum.

  • Indications for Use B.
    For in vitro diagnostic use with the CARESIDE Analyzer to quantitatively measure GGT from anti-coagulated whole blood, plasma, or serum specimens to aid in the diagnosis and treatment of liver diseases such as alcoholic cirrhosis and primary and secondary liver tumors.

Technological Characteristics VI.

  • Similarities A.
CARESIDE® GGTVitros GGT DT Slides
Intended UseFor in vitro Diagnostic useSame
IndicationsPrimarily to aid in thediagnosis and treatment ofpatients with liver diseasessuch as alcoholic cirrhosis andprimary and secondary livertumors.Same
MeasurementQuantitativeSame
MethodDry film based quantitation ofSame
Principleenzymatic activity byreflectance photometry usingL-g-glutamyl transferase andglycylglycine substrate.
SpecimenNot requiredSame
Dilution
MaterialsL-y-glutamyl-p-nitroanilideand glycylglycineL-y-glutamyl-p-nitroanilideand glycylglycine
DetectorReflectance (425 nm)Reflectance (400 nm)
Test TimeApprox. 4 min. warm-up (on-board) plus 5 minute test time.15 minutes slide warm-up (off-line) plus 5 minutes test time.
Sample TypeAnti-coagulated whole blood,serum or plasma [whole bloodapplied sample, plasma testsample]Serum, plasma
Specimen8.5 µl test volume10 µl
Volume(90 ± 10 µl applied volume)
CalibrationCalibration information bar-coded on each cartridge.Calibration information maychange with each lot.Run Vitros DTSC IIcalibrators whenever a newslide lot is used or whennecessary.
Quality Control2 levelsSame
Reporting UnitsU/LSame
Reaction Temp.37 °CSame

{3}------------------------------------------------

B. Differences

CARESIDE® GGTVitros GGT DT Slides
Direct BloodSpecimenYes, whole bloodNo, requires separation ofwhole blood prior tosample application
Reportablerange20 to 1000 U/L5 to 1400 U/L
AccuratePipettingNot requiredRequired
Reagent Pre-warmingNot requiredRequired

C. Comparative Performance Characteristics

CARESIDE® GGTVitros GGT DT Slides
Detection limit20 U/L5 U/L
ReportableRange20 to 1000 U/L5 to 1400 U/L
AccuracyMean recovery 101%Not available
PrecisionTotal CV, 344 U/L, 2.6%Total CV, 166 U/L, 2.4%
MethodComparisonCARESIDE = 1.00 (BM/Hitachi 902 GGT) + 0.91 U/L,r= 1.00
LinearityLinearity by dilutionyielded slope andcorrelation coefficientwithin acceptable limits.Not available
InterferenceNo significant interferenceobserved at testedconcentration ofinterferent:Ascorbic Acid 10 mg/dLBilirubin 10 mg/dLTriglycerides 3000 mg/dLNone stated
Specimen Types&AnticoagulantsNo clinically significantdifference between sodiumheparinized whole blood,serum, and sodium heparinplasma.No clinically significantdifference between serum,heparin plasma, or EDTAplasma. Whole blood isunsuitable.

D. Conclusion

The nonclinical and clinical data provided demonstrate that the CARESIDE® GGT product is as safe, effective, and performs as well as or better than the legally marketed predicate device

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around it. Inside the circle is an abstract symbol that resembles three stylized human profiles or figures, possibly representing people or services provided by the department.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Renate A. MacLauren, Ph.D. Clinical Affairs Manager Careside. Inc. 6100 Bristol Parkway Culver City, CA 90230

Re: K020486 Trade/Device Name: Careside® GGT Regulation Number: 21 CFR 862.1360 Regulation Name: Gamma-glutamyl transpeptidase and isoenzymes test system Regulatory Class: Class I, reserved Product Code: JPZ Dated: May10, 2002 Received: May13, 2002

:

'JUL -

1 2002

Dear Dr. MacLauren:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

Page 2 =

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and ' additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device. please contact.the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html",

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

K0204:86

INDICATIONS FOR USE

510(k). Number:

Device Name:

· CARESIDE® GGT

For in vitro diagnostic use with the CARESIDE Analyzer to quantitatively Indications for use: measure GGT from anti-coagulated whole blood, plasma, or serum specimens to aid in the diagnosis and treatment of liver diseases such as alcoholic cirrhosis and primary and secondary liver tumors.

Jean Cooper
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K020486

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use X (Per 21 CFR 801.109)

OR

Over-The-Counter Use (Optional Format 1-2-96)

§ 862.1360 Gamma-glutamyl transpeptidase and isoenzymes test system.

(a)
Identification. A gamma-glutamyl transpeptidase and isoenzymes test system is a device intended to measure the activity of the enzyme gamma-glutamyl transpeptidase (GGTP) in plasma and serum. Gamma-glutamyl transpeptidase and isoenzymes measurements are used in the diagnosis and treatment of liver diseases such as alcoholic cirrhosis and primary and secondary liver tumors.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 862.9.