Search Filters

Search Results

Found 53 results

510(k) Data Aggregation

    K Number
    K232404
    Date Cleared
    2024-08-09

    (365 days)

    Product Code
    Regulation Number
    862.1175
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CHOLESTEROL: Reagent kit intended for the quantitative determination of Cholesterol in human serum. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood, of lipid and lipoprotein metabolism disorders.

    HDL-Cholesterol: Reagent kit intended for the quantitative determination of high-density lipoprotein in human serum. Measurements are used in the diagnosis and treatment of lipid disorders mellitus), atherosclerosis, and various liver and renal diseases.

    LDL-Cholesterol: Reagent kit intended for the quantitative determination of low-density lipoprotein in human serum. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.

    TRIGLYCERIDES: Reagent kit intended for the quantitative determination of triglycerides (neutral fat) in human serum. Measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders.

    Device Description

    CHOLESTEROL: The Cholesterol Oxidase peroxidase (CHOD-PAP) enzymatic method is used. The cholesterol esterase enzyme catalyzes the hydrolysis of cholesterol and free fatty and free fatty acids. Free cholesterol, including that originally present in the sample, is then oxidized by the enzyme cholesterol oxidase (CHOD) to cholest-4-en-3-one, by using molecular oxygen as the electron acceptor and concurrently producing hydrogen peroxide (H2O2). The H2O2 produced is then used in a subsequent chromogenic oxidative coupling reaction, catalyzed by the enzyme peroxidase, in the presence of a redox indicator system, which leads to the formation of a colored compound, absorbing light at 550 nm. The increase in absorbance is directly proportional to the cholesterol concentration in the sample.

    HDL-Cholesterol: The Accelerator Selective Detergent method is applied. The determination of HDL-Cholesterol is based on the following reactions: LDL, VLDL, and chylomicrons are neutralized by the combined action of the enzymes Cholesterol Oxidase, Peroxidase, accelerators and N,N-bis-(4-sulfobutyl)-m-toluidine-disodium (DSBmT). HDL remaining in the sample is disrupted by the action of a selective detergent and cholesterol is converted to △4 Cholestenone by the enzymatic action of Cholesterol Esterase and Cholesterol Oxidase, with the subsequent production of H2O2, which reacts with DSBmT and 4-aminoantipyrine in the presence of Peroxidase to a colored complex that absorbs light at 590 nm. The absorbance measured is proportional to the concentration of HDL-Cholesterol in the sample.

    LDL-Cholesterol: The Selective Detergent method is applied. The method is in a two-reagent format and depends on the properties of a unique detergent. The first detergent solubilizes only the non-LDL lipoprotein particles. The cholesterol released is consumed by cholesterol esterase and cholesterol oxidase in a non-color forming reaction. The second detergent solubilizes the remaining LDL particles, and a chromogenic coupler allows for color formation. The enzyme reaction with LDL-Cholesterol in the presence of the coupler at 590 nm produces color that is proportional to the amount of LDL cholesterol present in the sample.

    TRIGLYCERIDES: The enzymatic glycerol-3-phosphate-peroxidase (GPO-POD) method is used. The method enzymatically hydrolyzes by lipase to free fatty acids and glycerol is phosphorylated by adenosine triphosphate (ATP) with glycerokinase (GK) to produce glycerol-3-phosphate and adenosine diphosphate (ADP). Glycerol-3-phosphate-oxidase oxidizes glycerol-3-phosphate to dihydroxyacetone phosphate and H2O2. The catalytic action of peroxidase (POD) forms quinoneimine from H202, aminoantipyrine, and Dihydrate (N-Ethyl-N-(2hydroxy-3-sulfopropyl)-m-toluidine (TOOS). The absorption change at 550 nm is proportional to the triglycerides concentration in the sample.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the Medicon Hellas Cholesterol, HDL-Cholesterol, LDL-Cholesterol, and Triglycerides test systems, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally established by comparison to legally marketed predicate devices and alignment with clinical laboratory guidelines (CLSI). The document presents a clear comparison in the "Device Comparison Table" sections. For this summary, I'll focus on the key performance indicators for each analyte.

    CHOLESTEROL

    Acceptance Criteria (Predicate: OLYMPUS CHOLESTEROL REAGENT (K925603))Reported Device Performance (Medicon Hellas CHOLESTEROL)
    Method comparison (correlation to comparator): 1.000Method comparison (correlation to comparator): 0.9980
    Reportable range: 20 to 700 mg/dLReportable range: 20 to 700 mg/dL
    Sensitivity LoD: 1 mg/dL (Predicate LoQ not defined)Sensitivity LoD / LoQ: LoD 4.4 / LoQ 4.6 (mg/dL)
    Precision (within run & total for all LVs): <= 3%Precision (within run & total for all LVs): <= 4%
    Endogenous Interferences: Hemoglobin: up to 500 mg/dLEndogenous Interferences: Hemoglobin: up to 500 mg/dL
    Endogenous Interferences: Triglycerides: up to 1,000 mg/dLEndogenous Interferences: Triglycerides: up to 1,500 mg/dL
    Calibration frequency: 30 daysCalibration frequency: 14 days
    On-board stability: Not definedOn-board stability: 28 days
    Specimen Type: Human serum, plasma and urineSpecimen Type: Human serum

    HDL-Cholesterol

    Acceptance Criteria (Predicate: DIRECT HDL (K981224))Reported Device Performance (Medicon Hellas HDL-Cholesterol)
    Method comparison (correlation to comparator): 1.999 (Typo in document, likely 0.999)Method comparison (correlation to comparator): 0.997
    Reportable range: 5.0 to 221 mg/dLReportable range: 6.0 to 200 mg/dL
    Sensitivity LoD / LoQ: LoD 2.5 / LoQ 5.0 (mg/dL)Sensitivity LoD / LoQ: LoD 3.0 / LoQ 5.8 (mg/dL)
    Precision (within run & total for all LVs): <= 4.0%Precision (within run & total for all LVs): <=4.0%
    Endogenous Interferences: Hemoglobin: up to 2,000 mg/dLEndogenous Interferences: Hemoglobin: up to 1,000 mg/dL
    Endogenous Interferences: Triglycerides: MDL1 1,000mg/dL & MDL2 2,000mg/dLEndogenous Interferences: Triglycerides: up to 1,500 mg/dL
    Calibration frequency: 28 daysCalibration frequency: 28 days
    On-board stability: Not definedOn-board stability: 28 days
    Specimen Type: Human serum & plasmaSpecimen Type: Human serum

    LDL-Cholesterol

    Acceptance Criteria (Predicate: DIRECT LDL (K981303))Reported Device Performance (Medicon Hellas LDL-Cholesterol)
    Method comparison (correlation to comparator): 0.960Method comparison (correlation to comparator): 0.999
    Reportable range: 1 to 800 mg/dLReportable range: 3 to 800mg/dL
    Sensitivity LoD / LoQ: < 10mg/dLSensitivity LoD / LoQ: LoD 2 / LoQ 3 mg/dL
    Precision (within run & total for all LVs): < 4.0%Precision (within run & total for all LVs): < 4.0%
    Endogenous Interferences: Hemoglobin: up to 500mg/dLEndogenous Interferences: Hemoglobin: up to 1,000mg/dL
    Endogenous Interferences: Triglycerides: up to 1,293 mg/dLEndogenous Interferences: Triglycerides: up to 1,500 mg/dL
    Calibration frequency: 28 daysCalibration frequency: At new lot
    On-board stability: Not definedOn-board stability: 28 days
    Specimen Type: Human serum & plasmaSpecimen Type: Human serum

    TRIGLYCERIDES

    Acceptance Criteria (Predicate: OLYMPUS TRIGLYCERIDE TEST SYSTEM (K063804))Reported Device Performance (Medicon Hellas TRIGLYCERIDES)
    Method comparison (correlation to comparator): 0.999Method comparison (correlation to comparator): 0.999
    Reportable range: 10 to 1,000mg/dLReportable range: 10 to 1,000mg/dL
    Sensitivity LoD / LoQ: < 0.31 / 5.0 mg/dLSensitivity LoD / LoQ: LoD 5.5 / LoQ 9.7 mg/dL
    Precision (within run & total for all LVs): < 5.0%Precision (within run & total for all LVs): < 4.0%
    Endogenous Interferences: Hemoglobin: up to 500mg/dLEndogenous Interferences: Hemoglobin: up to 400mg/dL
    Calibration frequency: 30 daysCalibration frequency: 28 days
    On-board stability: 30 daysOn-board stability: 28 days
    Specimen Type: Human serum, plasma & urineSpecimen Type: Human serum

    2. Sample size used for the test set and the data provenance

    • Accuracy (Method Comparisons):

      • A minimum of 75 leftover specimens.
      • For the specific analytes:
        • CHOLESTEROL: 93 human serum samples
        • HDL-Cholesterol: 141 human serum samples
        • LDL-Cholesterol: 107 human serum samples
        • TRIGLYCERIDES: 163 human serum samples
      • Data Provenance: The document states "left-over specimens," implying retrospective use of clinical samples. The country of origin is not explicitly stated, but the company is Medicon Hellas, S.A. based in Greece, and testing was performed at the company premises.
    • Precision/Reproducibility:

      • Three human serum pools for Cholesterol and Triglycerides.
      • Two pools for HDL-Cholesterol.
      • Four pools for LDL-Cholesterol.
      • Each sample was tested for 20 testing days, two different runs, and two replicate measurements per run (morning and afternoon run), for a total of 80 results per level concentration (e.g., for Cholesterol, 3 pools * 80 results/pool = 240 results).
      • Data Provenance: Human serum pools, likely prepared in-house or acquired for the study.
    • Linearity:

      • 11 to 13 levels per analyte, prepared by dilution of a human serum pool.
      • Each level was tested in 4 replicates.
      • Data Provenance: Human serum pool.
    • Analytical Specificity / Interference:

      • Serum pools at low and high levels of each analyte.
      • Each measurement of the blank and the sample containing the interferent was repeated at least 5 times.
      • Data Provenance: Serum pools.
    • Detection Limit:

      • LoB: 5 blank serum samples measured in 4 replicates for 3 days (total of 60 measurements).
      • LoD: 5 low-level samples measured in 4 replicates for 3 days (total of 60 measurements).
      • LoQ: 10 samples that span the low end of linearity, measured 5 times each day for 3 days (total of 150 measurements).
      • Data Provenance: Serum samples.
    • Stability and Calibration Frequency:

      • Two fresh serum pools and two serum-based commercial controls.
      • Measurements were repeated in triplicate at regular time points.
      • Data Provenance: Serum pools and commercial controls.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The ground truth for performance studies like those described (method comparison, precision, linearity, interference, detection limits) for in vitro diagnostic (IVD) devices like these typically involves established reference methods or highly accurate comparative analyzers, rather than expert human interpretation of results. The document states that the performance of the Medicon Hellas reagents was compared with "comparator methods" (Beckman Coulter reagents on AU400, Abbott Diagnostics reagents on Architect c8000), which serve as the reference for ground truth in these types of analytical performance studies. The qualifications of the operators performing these studies are not specified.

    4. Adjudication method for the test set

    This information is not applicable and therefore not provided. Adjudication methods (e.g., 2+1, 3+1) are typically used in studies where human interpretation of complex data (like medical images) is involved and a consensus is needed to establish ground truth. For quantitative chemical assays, the ground truth is established by the highly precise and accurate measurement of reference methods or predicate devices.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable and therefore not provided. MRMC studies are specific to evaluating diagnostic devices where human readers interpret medical cases, often with and without AI assistance (e.g., radiology AI). The Medicon Hellas devices are in vitro diagnostic reagents for quantitative chemical measurements in serum, not image-based diagnostic tools that require human "readers" in the context of an MRMC study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This concept is not applicable in the traditional sense for these in vitro diagnostic reagents. These devices are chemical assays that produce a quantitative numerical output (e.g., cholesterol level in mg/dL). There isn't an "algorithm" making a diagnostic interpretation independently in the way AI software would. The device is the standalone measurement system. Its performance is evaluated independently through analytical studies (precision, linearity, accuracy against reference methods, etc.). The results are then read and interpreted by a human clinician.

    7. The type of ground truth used

    The ground truth for the analytical performance studies (precision, linearity, interference, detection limits, and method comparison) was established against:

    • Reference Methods/Predicate Devices: For method comparison, the device's performance was compared against established comparator methods (Beckman Coulter reagents on AU400 and Abbott Diagnostics reagents on ABBOTT Architect c8000). The document explicitly states these as the comparators.
    • A Priori Values/Established Standards: For linearity, precision, and detection limits, the ground truth is based on the known concentrations of prepared samples (e.g., serially diluted pools, spiked samples, blank serum) and statistical analysis according to CLSI guidelines.
    • Traceability to Reference Methods/Materials: For Cholesterol and Triglycerides, traceability is to Gas-chromatography-isotope dilution mass spectrometry (GC-IDMS). For HDL-Cholesterol and LDL-Cholesterol, traceability is to the Abell-Kendall (AK) reference method.

    8. The sample size for the training set

    This information is not applicable and therefore not provided. These are chemical reagent kits, not machine learning (AI/ML) models that require a "training set" in the computational sense. The development of such reagents involves chemical and enzymatic research and optimization, often tested with various batches and concentrations of samples during R&D. The studies described in this document are for validation and verification of the final device, not for "training" an algorithm.

    9. How the ground truth for the training set was established

    As noted above, the concept of a "training set" with ground truth established in the AI/ML sense is not applicable to these chemical reagent devices. The "ground truth" for evaluating the analytical performance of the developed reagent kits is established through the reference methods and standardized protocols described in section 7.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203597
    Device Name
    Cholesterol2
    Date Cleared
    2022-06-30

    (568 days)

    Product Code
    Regulation Number
    862.1175
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cholesterol2 assay is used for the quantitation of cholesterol in human serum or plasma on the ARCHITECT c System. The Cholesterol2 assay is to be used as an aid in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.

    Device Description

    The Cholesterol2 assay is an automated clinical chemistry assay for the quantitation of cholesterol in human serum or plasma on the ARCHITECT c System. Cholesterol esters are enzymatically hydrolyzed by cholesterol esterase to cholesterol and free fatty acids. Free cholesterol, including that originally present, is then oxidized by cholesterol oxidase to cholest-4-ene-3-one and hydrogen peroxide. The hydrogen peroxide oxidatively couples with N,N-Bis(4-sulfobutyl)-3-methylaniline (TODB) and 4-aminoantipyrine to form a chromophore (quinoneimine dye) which is quantitated at 604 nm.

    AI/ML Overview

    The provided text describes the Abbott Cholesterol2 assay, an in vitro diagnostic device for quantifying cholesterol in human serum or plasma.

    Here's an analysis of the acceptance criteria and study data:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria for each performance characteristic. Instead, it presents the results of various studies as proof of device performance. The table below summarizes the reported performance, which implicitly serves as the "met" criteria.

    Performance CharacteristicReported Device Performance (Cholesterol2)
    Analytical Measuring Interval (AMI)5-748 mg/dL
    Extended Measuring Interval (EMI)748-2992 mg/dL
    Reportable Interval2-2992 mg/dL
    Precision
    Control Level 1 (251 mg/dL)SD: 1.9 mg/dL (Within-Run), 2.6-3.1 mg/dL (Within-Laboratory); %CV: 0.7% (Within-Run), 1.0-1.2% (Within-Laboratory)
    Control Level 2 (106 mg/dL)SD: 1.0 mg/dL (Within-Run), 1.3-1.7 mg/dL (Within-Laboratory); %CV: 1.0% (Within-Run), 1.2-1.6% (Within-Laboratory)
    Panel A (21 mg/dL)SD: 0.6 mg/dL (Within-Run), 0.7-0.8 mg/dL (Within-Laboratory); %CV: 3.0% (Within-Run), 3.2-4.1% (Within-Laboratory)
    Panel B (237 mg/dL)SD: 2.8 mg/dL (Within-Run), 3.7-4.9 mg/dL (Within-Laboratory); %CV: 1.2% (Within-Run), 1.5-2.0% (Within-Laboratory)
    Panel C (718 mg/dL)SD: 6.4 mg/dL (Within-Run), 4.6-6.9 mg/dL (Within-Laboratory); %CV: 0.9% (Within-Run), 0.7-1.0% (Within-Laboratory)
    Limit of Blank (LoB)0 mg/dL
    Limit of Detection (LoD)2 mg/dL
    Limit of Quantitation (LoQ)5 mg/dL (at 20% CV maximum allowable precision)
    LinearityLinear across the analytical measuring interval of 5 to 748 mg/dL
    Interference (Endogenous)
    Conjugated Bilirubin (7 mg/dL)No significant interference (within ± 10%)
    Unconjugated Bilirubin (11 mg/dL)No significant interference (within ± 10%)
    Hemoglobin (1000 mg/dL)No significant interference (within ± 10%)
    Total Protein (15 g/dL)No significant interference (within ± 10%)
    Conjugated Bilirubin (40 mg/dL)Interference: -39% at 150 mg/dL analyte, -31% at 220 mg/dL analyte
    Unconjugated Bilirubin (16 mg/dL)Interference: -11% at 150 mg/dL analyte
    Interference (Exogenous)
    Acetaminophen (160 mg/L)No significant interference
    Acetylcysteine (150 mg/L)No significant interference
    Acetylsalicylic acid (30 mg/L)No significant interference
    Aminoantipyrine (40 mg/L)No significant interference
    Ampicillin-Na (80 mg/L)No significant interference
    Biotin (4250 ng/mL)No significant interference
    Ca-dobesilate (60 mg/L)No significant interference
    Cefotaxime (53 mg/dL)No significant interference
    Cefoxitin (6600 mg/L)No significant interference
    Cyclosporine (2 mg/L)No significant interference
    Desacetylcefotaxime (6 mg/dL)No significant interference
    Dipyrone (100 mg/L)No significant interference
    Dobutamine (0.2 mg/dL)No significant interference
    Doxycycline (20 mg/L)No significant interference
    Ibuprofen (220 mg/L)No significant interference
    Intralipid (1050 mg/dL)No significant interference
    Levodopa (8 mg/L)No significant interference
    Methotrexate (140 mg/dL)No significant interference
    Metronidazole (130 mg/L)No significant interference
    Methylaminoantipyrine (40 mg/L)No significant interference
    Methyldopa (20 mg/L)No significant interference
    N-Acetyl-p-benzoquinone (NAPQI) (20 mg/L)No significant interference
    Phenylbutazone (330 mg/L)No significant interference
    Phenytoin (6 mg/dL)No significant interference
    Rifampicin (50 mg/L)No significant interference
    Sodium heparin (4 U/mL)No significant interference
    Sulpiride (15 mg/L)No significant interference
    Theophylline (60 mg/L)No significant interference
    Ascorbic acid (60 mg/L)Interference: -10% at 150 mg/dL analyte
    Intralipid (2000 mg/dL)Interference: -27% at 150 mg/dL analyte, -22% at 220 mg/dL analyte
    Methyldopa (30 mg/L)Interference: -14% at 150 mg/dL analyte
    Method Comparison (vs. Predicate)
    Serum (n=138)Correlation Coefficient: 1.00; Intercept: 0.41; Slope: 0.98 (Range: 7-684 mg/dL)
    Tube Type SuitabilityAcceptable for Serum, Serum separator, Lithium heparin, Lithium heparin separator, Sodium heparin tubes.
    Dilution VerificationAutomated dilution protocol (1:5.97) and manual dilution procedure (1:4) evaluated. (Performance details not provided in summary).
    TraceabilityTraceable to National Reference System for Cholesterol (Abell-Kendall reference method in a CDC-Certified CRMLN).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Precision Study: 2 controls and 3 human serum panels were tested. Each sample was tested in duplicate, twice per day for 20 days. This means 80 measurements per sample (2 duplicates x 2 times/day x 20 days).
    • Lower Limits of Measurement Study: n ≥ 60 replicates for LoB, LoD, and LoQ determinations. They used low-analyte level samples and zero-analyte samples.
    • Linearity Study: The number of samples for the linearity study is not explicitly stated, but it covered the range of 5 to 748 mg/dL.
    • Interference Studies: Each endogenous substance was tested at 2 analyte levels (approximately 150 mg/dL and 220 mg/dL). Exogenous substances were tested at various specified interferent levels. The number of samples for each interferent is not provided.
    • Method Comparison Study: 138 serum samples were used.
    • Tube Type Study: Samples were collected from a minimum of 40 donors.
    • Dilution Verification: 8 human serum samples.

    Data Provenance: The document does not specify the country of origin of the data or whether the studies were retrospective or prospective. Given the context of medical device regulatory submission, these are typically prospective studies conducted in a controlled laboratory environment. The "human serum panels" and "human serum samples" imply human-derived biological samples.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This device is an in vitro diagnostic (IVD) chemistry assay. The concept of "experts establishing ground truth" as it applies to image interpretation or clinical diagnosis by medical professionals (like radiologists) does not directly apply here in the same way.

    For IVDs like this, the "ground truth" or reference values are established through:

    • Reference methods: The Cholesterol2 reagent is certified to be traceable to the National Reference System for cholesterol, against the Abell-Kendall reference method in a CDC-Certified Cholesterol Reference Method Laboratory Network (CRMLN). The Abell-Kendall method is considered the gold standard for cholesterol measurement.
    • Analytically Validated Methods: For values outside the AMI but within the EMI, samples were "value assigned using the analytically validated method."
    • Known concentrations: For studies like linearity, spiked samples with known concentrations are used.

    Therefore, the "experts" are the methodologists and laboratory professionals overseeing and validating the reference methods and the analytical validation processes. No specific number or qualification of clinical experts (e.g., radiologists) is relevant for establishing the ground truth for a quantitative chemistry assay.

    This is a standalone performance evaluation of the assay itself, demonstrating its analytical accuracy, precision, and robustness.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The concept of "adjudication" (e.g., 2+1, 3+1 where multiple human readers agree or have a tie-breaker by an expert) is not applicable to this type of quantitative diagnostic assay. The results are numerical values generated by the automated instrument and reagents. Deviations or discrepancies would be resolved through re-testing, quality control, or investigation into instrument or reagent issues, rather than human expert adjudication of a "diagnosis."


    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This is an in vitro diagnostic assay, not an AI-powered diagnostic imaging device or an AI assistant for human readers. Its output is a quantitative measurement of cholesterol, not an interpretation that requires human "reading" or decision support.


    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Yes, a standalone performance evaluation of the device (Cholesterol2 assay on the ARCHITECT c8000 System) was done. The studies described (reportable interval, precision, lower limits of measurement, linearity, interference, method comparison, tube type, dilution verification) all evaluate the analytical performance of the assay and instrument directly, without human interpretation as part of the primary outcome measure. The output is a numerical concentration of cholesterol.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth used for this quantitative assay primarily relies on:

    • Reference Methods: Specifically, the Abell-Kendall reference method, which is considered the gold standard for cholesterol measurement and is used in CDC-Certified Cholesterol Reference Method Laboratory Networks (CRMLN). The device's traceability to this method is explicitly stated.
    • Analytically Validated Methods: For verifying values in the extended measuring interval.
    • Known Spiked Concentrations: For studies such as linearity and dilution verification, where samples are prepared with precisely known concentrations.

    This is an analytical ground truth, not a clinical ground truth derived from pathology or patient outcomes.


    8. The sample size for the training set

    This document does not describe a typical "training set" in the context of machine learning or AI. This is a chemistry assay that uses reagents and enzymatic reactions, not an algorithm that is "trained" on data. Therefore, the concept of a training set as used in AI development is not applicable here. The assay's analytical characteristics are determined through standard laboratory validation studies.


    9. How the ground truth for the training set was established

    As explained above, there is no "training set" in the AI sense for this device. The analytical accuracy and reliability are established through comparisons to certified reference methods and known standard concentrations, as described in point 7.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190490
    Date Cleared
    2019-03-27

    (27 days)

    Product Code
    Regulation Number
    862.1175
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Rx Only For in vitro diagnostic use only

    The TRIG test within the VITROS XT Chemistry Products TRIG-CHOL Slides quantitatively measure triglyceride (TRIG) concentration in serum and plasma using VITROS XT 7600 Integrated Systems. Triglyceride measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver diseases involving lipid metabolism, or various endocrine disorders.

    The CHOL test within the VITROS XT Chemistry Products TRIG-CHOL Slides quantitatively measure cholesterol (CHOL) concentration in serum and plasma using VITROS XT 7600 Integrated Systems. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders mellitus), atherosclerosis, and various liver and renal diseases.

    Device Description

    The new device, the VITROS XT Chemistry Products TRIG-CHOL Slide is a single device that contains both a TRIG test and a CHOL test multilayered, analytical elements coated on a polyester support separated by a plastic barrier sealed within a single slide frame. In this format, individual reactions occur and test results are generated for each analyte independently of the other analyte.

    To perform the TRIG test, a drop of patient sample is deposited on the slide and is evenly distributed by the spreading layer to the underlying layers. The Triton X-100 surfactant in the spreading layer aids in dissociating the triglycerides from lipoprotein complexes present in the sample. The triglyceride molecules are then hydrolyzed by lipase to yield glycerol and fatty acids. Glycerol diffuses to the reagent layer, where it is phosphorylated by glycerol kinase in the presence of adenosine triphosphate (ATP). In the presence of L-a-glycerolphosphate oxidase, L-α-glycerophosphate is then oxidized to dihydrox vacetone phosphate and hydrogen peroxide. The final reaction involves the oxidation of a leuco dye by hydrogen peroxide, catalyzed by peroxidase, to produce a dye. The density of the dye formed is proportional to the triglyceride concentration present in the sample and is measured by reflectance spectrophotometry.

    To perform the CHOL test, a drop of patient sample is deposited on the slide and is evenly distributed by the spreading layer to the underlying layers. The Triton X-100 (TX100) surfactant in the spreading layer aids in dissociating the cholesterol and cholesterol esters from lipoprotein complexes present in the sample. Hydrolysis of the cholesterol esters to cholesterol is catalyzed by cholesterol ester hydrolase. Free cholesterol is then oxidized in the presence of cholesterol oxidase to form cholestenone and hydrogen peroxide. Finally, hydrogen peroxide oxidizes a leuco dye in the presence of peroxidase to generate a colored dye. The density of dye formed is proportional to the cholesterol concentration present in the sample and is measured by reflectance spectrophotometry.

    AI/ML Overview

    The provided text describes the analytical performance of the VITROS XT Chemistry Products TRIG-CHOL Slides, an in vitro diagnostic device for quantitatively measuring triglyceride (TRIG) and cholesterol (CHOL) concentrations in serum and plasma.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of "acceptance criteria" but rather reports the analytical performance of the new device and shows its comparison to the predicate devices. The implicit acceptance criteria are that the new device performs at least as well as, and is substantially equivalent to, the predicate devices. The reported performance metrics are detailed in the tables for method comparison, precision, detection capability, and linearity.

    TRIG Test

    Performance MetricAcceptance Criteria (Implicit)Reported Device Performance (VITROS XT TRIG-CHOL Slides)
    Method ComparisonSubstantially equivalent to predicate device (VITROS TRIG Slides)Slope: 0.99, Intercept: 1.49, Correlation Coefficient: 1.000
    Precision (CV%)Acceptable variability for clinical usePool 1: 1.3%, Native Pool: 1.5%, Control 1: 0.7%, Control 2: 0.8%, Pool 2: 0.9%, Pool 3: 0.9%
    Limit of Quantitation (LoQ)Clinically relevant LoQ for TRIG10 mg/dL (Criteria: %CV < 20%)
    LinearityLinear within claimed measuring rangeLinear Range: 8.0 - 542.8 mg/dL (Claimed: 10 - 525 mg/dL)

    CHOL Test

    Performance MetricAcceptance Criteria (Implicit)Reported Device Performance (VITROS XT TRIG-CHOL Slides)
    Method ComparisonSubstantially equivalent to predicate device (VITROS CHOL Slides)Slope: 0.97, Intercept: 0.09, Correlation Coefficient: 0.999
    Precision (CV%)Acceptable variability for clinical usePool 1: 2.1%, Control 1: 1.9%, Native Pool: 1.3%, Pool 2: 1.5%, Control 2: 1.6%, Pool 3: 1.5%
    Limit of Quantitation (LoQ)Clinically relevant LoQ for CHOL50 mg/dL (Criteria: %CV < 9%)
    LinearityLinear within claimed measuring rangeLinear Range: 27 - 358 mg/dL (Claimed: 50 - 325 mg/dL)

    2. Sample sizes used for the test set and the data provenance

    • Method Comparison: 148 serum samples were used for both TRIG and CHOL tests.
    • Precision: 80 observations (2 replicates per run, 2 runs per day over 20 days) for serum samples (patient pools and quality control materials).
    • Detection Capability (LoQ): 180 determinations for both TRIG and CHOL.
    • Linearity: Eighteen proportionally related admixtures of low and high test fluids, each tested in quadruplicate.

    The data provenance is not explicitly stated in terms of country of origin or whether the samples were retrospective or prospective. It implies the use of patient samples, but specific details are absent.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This document describes the analytical performance of an in vitro diagnostic device, not a device that relies on expert human interpretation for its output (like an AI imaging device). Therefore, there is no mention of "experts used to establish the ground truth" in the way it would apply to a clinical imaging study or a study validating human performance. The "ground truth" for this type of device is established by its quantitative measurements against established analytical standards and reference methods/predicate devices.

    4. Adjudication method for the test set

    Not applicable. As noted above, this is an analytical performance study for an in vitro diagnostic device, not a human reader study requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is an analytical performance study of an in vitro diagnostic device, not an AI-assisted diagnostic tool that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, this entire study represents a "standalone" performance evaluation of the device (VITROS XT Chemistry Products TRIG-CHOL Slides on the VITROS XT 7600 Integrated System) without human intervention in the result generation. The device quantitatively measures the analytes.

    7. The type of ground truth used

    The ground truth for this in vitro diagnostic device is established through:

    • Comparison to a legally marketed predicate device: The VITROS Chemistry Products TRIG Slides and VITROS Chemistry Products CHOL Slides. This implies that the predicate devices serve as the established reference standard for performance.
    • Established analytical methods and materials: CLSI protocols (EP09c, EP05-A3, EP17-A2, EP06-A, EP07-03) are referenced, which dictate the methodology for evaluating analytical performance parameters.
    • Quality Control Materials and reference pools: Used in precision and linearity studies.
    • Clinically established guidelines: NCEP guidelines are referenced for expected values/classification of TRIG and CHOL, indicating alignment with clinical understanding of these analytes.

    8. The sample size for the training set

    This document describes pre-market validation studies for a diagnostic test kit and instrument system. It does not refer to a "training set" in the context of machine learning or AI. The development of such a device involves internal optimization and development work, but the data presented in this 510(k) summary are for the validation of the finalized product.

    9. How the ground truth for the training set was established

    Not applicable, as this refers to a diagnostic test kit and instrument's analytical validation, not an AI or machine learning model that would involve a "training set" with ground truth established through expert annotation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181373
    Date Cleared
    2018-10-18

    (147 days)

    Product Code
    Regulation Number
    862.1175
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Extended Lipid Panel Assay is an in vitro diagnostic test for quantitative determination of Total Cholesterol, High Density Lipoprotein Cholesterol, and Triglycerides in human serum and Apolipoprotein B in human serum. Values for Total Cholesterol, High Density Lipoprotein Cholesterol, Triglycerides and Apolipoprotein B are calculated by the Vantera® Clinical Analyzer.

    · Total Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood, lipid and lipoprotein metabolism disorders.

    · High Density Lipoprotein Cholesterol measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.

    · Triglyceride measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders.

    · Apolipoprotein B measurements are used in the diagnosis and treatment of lipid disorders and atherosclerosis.

    Device Description

    The Extended Lipid Panel Assay involves the acquisition of a 400 MHz proton NMR spectrum of serum or plasma, passing the spectral information through a Partial Least Squares (PLS) regression model, and deriving analyte concentrations from the spectrum based on the trained PLS model. The proton NMR spectrum of serum and plasma is replete with information from the lipids packaged in lipoproteins. The spectrum consists of multiple proton signals emanating from the TG, cholesteryl esters and free cholesterol present in chylomicrons, VLDL, LDL and HDL, out of which the methylene and methyl proton signals are the most abundant. NMR spectra were recorded for several hundred to several thousand representative serum specimens for which the TG, TC, HDL-C and ApoB were chemically measured. Using a PLS regression routine, the spectral information in the combined methylene and methyl region (0.56 - 1.40 ppm) was trained against the chemical measurements where the information is connected through latent variables. Cross-validation was performed with PRESS statistics to optimize the regression model with an appropriate number of latent variables. Once trained with sufficient number of specimens, for any test specimen spectrum, the spectral information is then converted into lipid or ApoB concentrations through the optimum number of 24 to 27 latent variables for which the regression coefficients were known from the predictor matrix.

    AI/ML Overview

    The Extended Lipid Panel Assay is an in vitro diagnostic test for the quantitative determination of Total Cholesterol (TC), High Density Lipoprotein Cholesterol (HDL-C), Triglycerides (TG), and Apolipoprotein B (ApoB) in human serum and plasma. The device uses Nuclear Magnetic Resonance (NMR) technology to derive analyte concentrations.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this device are demonstrated through various analytical performance studies. The document does not explicitly state pre-defined acceptance criteria in a dedicated table with pass/fail results against specific thresholds for each performance metric, but rather presents the results of these studies. Based on the provided performance data, the implicit acceptance criteria would be for the device to show good analytical sensitivity, precision, linearity, and minimal interference, and to demonstrate substantial equivalence to predicate devices through method comparison studies.

    Below is a table summarizing the reported device performance, which implicitly demonstrates it meets the necessary criteria for "substantial equivalence" based on the FDA's assessment in the 510(k) clearance.

    Performance MetricAnalyteReported Device PerformanceImplicit Acceptance Standard (Inferred from predicates/regulatory guidance for clinical chemistry tests)
    Analytical Sensitivity (LoQ)TC24 mg/dLLowest concentration measurable with acceptable precision and accuracy (implicitly deemed acceptable by FDA)
    TG15 mg/dLLowest concentration measurable with acceptable precision and accuracy
    HDL-C13 mg/dLLowest concentration measurable with acceptable precision and accuracy
    ApoB18 mg/dLLowest concentration measurable with acceptable precision and accuracy
    Within-run Precision (%CV) (n=20)TC0.9 - 1.6%Low %CV indicating high within-run precision (typically < 5% or < 10% depending on analyte/concentration)
    TG1.0 - 1.0%Low %CV indicating high within-run precision
    HDL-C1.3 - 2.4%Low %CV indicating high within-run precision
    ApoB1.1 - 1.2%Low %CV indicating high within-run precision
    Within-Laboratory Precision (%CV) (n=80)TC1.1 - 1.6%Low %CV indicating high within-laboratory precision (typically < 5% or < 10%)
    TG1.0 - 1.4%Low %CV indicating high within-laboratory precision
    HDL-C1.4 - 2.8%Low %CV indicating high within-laboratory precision
    ApoB1.9 - 2.4%Low %CV indicating high within-laboratory precision
    Reproducibility (%CV)TC2.1 - 3.6%Low %CV across multiple sites/runs (typically < 10%)
    TG2.1 - 3.4%Low %CV across multiple sites/runs
    HDL-C2.1 - 4.6%Low %CV across multiple sites/runs
    ApoB3.7 - 6.7%Low %CV across multiple sites/runs
    Linearity (R²)TC1.000R² close to 1 indicates excellent linearity over the measuring range
    TG1.000R² close to 1 indicates excellent linearity over the measuring range
    HDL-C1.000R² close to 1 indicates excellent linearity over the measuring range
    ApoB0.999R² close to 1 indicates excellent linearity over the measuring range
    Linearity (Measuring Range)TC66-868 mg/dLBroad medically relevant measuring range
    TG35-950 mg/dLBroad medically relevant measuring range
    HDL-C14-152 mg/dLBroad medically relevant measuring range
    ApoB35-366 mg/dLBroad medically relevant measuring range
    Interfering SubstancesEndogenous/ExogenousNo significant interference observed for numerous agents at specified concentrations.No clinically significant bias from common interferents.
    Method Comparison (Correlation with Predicate)TCr = 0.994High correlation (r > 0.975 typically sought) and Deming regression slope/intercept close to 1 and 0, respectively.
    TGr = 0.9805 (implied from y= -3.858 + 0.9805x and visual)High correlation
    HDL-Cr = 0.985High correlation
    ApoBr = 0.980High correlation

    2. Sample sizes used for the test set and the data provenance

    • Analytical Sensitivity (LoQ): Not specified how "LoQ" was determined by sample size, but indicates "lowest concentration measurable with acceptable precision and accuracy."
    • Assay Precision (Within-run and Within-laboratory): 20 replicates for within-run and 80 replicates (for n=80, it means 40 total runs over 20 days with 2 replicates per run on 1 instrument) for within-laboratory precision, using three patient serum pools (low, medium, high). The provenance is "patient serum pools."
    • Reproducibility: Five levels of serum panels were tested for 5 days, 6 runs per day, 2 replicates per run at 3 sites. This results in 5 (levels) * 5 (days) * 6 (runs/day) * 2 (replicates/run) * 3 (sites) = 900 measurements per analyte. The provenance is from "serum panels," presumably patient-derived.
    • Linearity: "Reference serum pools were prepared from patient specimens with low to high values" and "mean values from analysis of four replicates of each pool" were used. The exact number of 'pools' created for each analyte is not explicitly stated but implies multiple points across the range. The provenance is "patient specimens."
    • Interfering Substances: "samples with spiked concentrations of interferent" were used. The number of samples is not specified, but it covered "Eight endogenous agents and thirty drugs."
    • Method Comparison:
      • TC: n=281 pooled serum samples
      • TG: n=270 pooled serum samples
      • HDL-C: n=15575 (This number seems exceptionally high compared to other n values; it might be a typo or represent cumulative data points over various experiments) pooled serum samples
      • ApoB: n=266 pooled serum samples
        The data provenance for method comparison is "pooled serum samples across the reportable range" of the device. All studies appear to be retrospective using banked or prepared samples, as typical for analytical performance studies of this nature. The country of origin is not specified but is presumably the US given the FDA submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This device is a clinical chemistry assay for quantifying specific analytes (TC, TG, HDL-C, ApoB) rather than an imaging or diagnostic AI device that interprets qualitative information and requires human expert consensus for ground truth.

    • Ground Truth for Method Comparison: The "ground truth" for the method comparison studies was established by predicate clinical chemistry devices (biochemical assays like enzymatic colorimetric assay or nephelometric immunoassay). These are established methods that serve as the comparative standard. No human experts were involved in establishing the ground truth for the values themselves, as it's a quantitative measurement.
    • Qualifications of Experts: Not applicable in the context of establishing ground truth for quantitative chemical measurements. However, the development, validation, and regulatory submission would have been overseen by qualified scientists and regulatory affairs professionals.

    4. Adjudication method for the test set

    Not applicable. Adjudication methods (like 2+1, 3+1) are typically used for qualitative or semi-quantitative diagnostic devices where human interpretation might differ, and an expert panel resolves discrepancies to establish ground truth. For quantitative clinical chemistry assays, the reference method (predicate device in this case) provides the "adjudicated" ground truth.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. MRMC studies are designed for evaluating diagnostic devices that involve human interpretation of images or other qualitative data, often with AI assistance, to assess the impact of AI on reader performance. This device is a standalone clinical chemistry analyzer that provides quantitative measurements; it does not involve human "readers" or AI assistance in a diagnostic interpretation workflow.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, this device is inherently a standalone algorithm. The Extended Lipid Panel Assay is described as utilizing "passing the spectral information through a Partial Least Squares (PLS) regression model, and deriving analyte concentrations from the spectrum based on the trained PLS model." This process is entirely automated and does not involve human intervention in the result generation once the sample is run on the Vantera® Clinical Analyzer. The performance studies (precision, linearity, method comparison) directly evaluate this standalone algorithmic performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for evaluating the Extended Lipid Panel Assay was based on measurements obtained from legally marketed predicate clinical chemistry devices. These predicate devices are established methods for quantifying the specific lipid and lipoprotein analytes (TC, TG, HDL-C, ApoB).

    8. The sample size for the training set

    For the Partial Least Squares (PLS) regression model: "NMR spectra were recorded for several hundred to several thousand representative serum specimens for which the TG, TC, HDL-C and ApoB were chemically measured." The exact number is not precisely stated but indicates a substantial training dataset.

    9. How the ground truth for the training set was established

    The ground truth for the training set was established by chemical measurements (presumably using standard, often wet-chemistry or enzymatic reference methods) for TG, TC, HDL-C, and ApoB on the "several hundred to several thousand representative serum specimens." These chemical measurements served as the target values for the PLS regression model to learn the relationship between the NMR spectra and the analyte concentrations.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180504
    Date Cleared
    2018-03-28

    (30 days)

    Product Code
    Regulation Number
    862.1175
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mission® Lipid Panel Monitoring System is intended for the quantitative determination of Total Cholesterol. High Density Lipoprotein Cholesterol, and Triglycerides in human capillary whole blood from the fingerin venous whole blood. The Mission® Lipid Panel Monitoring System consisting of the Mission® Lipid Panel Meter, Mission® Lipid Panel Test Cartridges, Mission® Cholesterol Optical Verifier, and Mission® Cholesterol Control Solution, and is intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto disabling lancing devices. This system is for in vitro diagnostic use only.

    Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. HDL (High Density Lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.

    Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, and other diseases involving lipid metabolism or various endocrine disorders.

    An estimated value for Low Density Lipoprotein Cholesterol is calculated by the Mission Cholesterol Pro Meter and is reported only when Triglycerides are ≤400 mg/dL.

    Device Description

    Not Found

    AI/ML Overview

    The document provided (K180504) is a 510(k) premarket notification for the Mission Lipid Panel Monitoring System. Based on the content, this device is a quantitative assay system for measuring Total Cholesterol, HDL-C, and Triglycerides, primarily used in professional healthcare settings. It is NOT an AI/ML powered device, nor does it involve image analysis by experts, or require MRMC studies.

    Therefore, many of the requested criteria (like number of experts, adjudication methods, multi-reader multi-case studies, effect size of human improvement with AI, training set details, etc.) are not applicable to this type of medical device and the information is not present in the provided document.

    However, I can extract the acceptance criteria and performance data relevant to a clinical chemistry device, which typically involves analytical accuracy and precision.

    Here's an attempt to answer the request based only on the provided document, acknowledging the limitations for an AI/ML context:


    The Mission Lipid Panel Monitoring System is a quantitative in vitro diagnostic device. The documentation provided focuses on clinical chemistry performance, not AI/ML algorithm performance. As such, many of the acceptance criteria and study details commonly associated with AI/ML devices (e.g., number of experts, MRMC studies, training set details, ground truth for image analysis) are not relevant or provided for this device type.

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document (K180504) is a 510(k) clearance letter and an Indications for Use statement. It does not contain the detailed performance data or the specific acceptance criteria tables that would typically be part of the predicate comparison or detailed validation studies submitted to the FDA. The letter only states that the device has been reviewed and determined to be substantially equivalent to legally marketed predicate devices.

    To illustrate what would typically be found for a quantitative diagnostic like this, and based on common regulatory expectations for lipid panels, performance would involve metrics like:

    Performance MetricAcceptance Criteria (Illustrative - Not explicitly stated in the provided document, but typical for lipid panels)Reported Device Performance (Illustrative - Not explicitly stated in the provided document)
    Accuracy (Bias)For each analyte (Total Cholesterol, HDL-C, Triglycerides), often assessed against a reference method (e.g., CDC-certified laboratory correlation). Criteria might be expressed as a percentage of bias or absolute bias within clinically acceptable limits for different concentration ranges.e.g., Total Cholesterol: Bias within ±5% or ±X mg/dL compared to reference method.
    Precision (Reproducibility)Within-run, between-run, and total precision (CV% or SD) for each analyte at different concentration levels (low, medium, high). Criteria based on CLIA or professional guidelines.e.g., HDL-C: CV% < 5% at clinically relevant concentrations.
    Linearity/Reportable RangeThe device should demonstrate linear response across the entire claimed measurement range for each analyte.Demonstrated linearity across the stated measurement ranges for each analyte.
    InterferenceNo significant interference from common endogenous substances (e.g., hemoglobin, bilirubin, triglycerides for cholesterol measurements) or exogenous substances (common medications).Showed no significant interference from tested substances within specified limits.
    Method ComparisonStrong correlation (e.g., R-squared > 0.95) and acceptable agreement (e.g., Bland-Altman plots) with a predicate device or established reference method.Strong correlation and agreement with predicate device/reference method.
    User Performance/Lay User Study (if applicable, though this is professional use)For professional use, demonstrates acceptable performance across different users or sites.Performance consistent across different operators in a professional setting.

    Important Note: The provided document does not contain any of these specific performance values or acceptance criteria. It is a regulatory clearance letter, not a summary of the validation study.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample size for any test set or the data provenance (e.g., country of origin, retrospective/prospective). Such details would be found in the more detailed 510(k) submission or a summary of safety and effectiveness, neither of which is part of this provided extract.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not applicable. This device measures chemical analytes (Total Cholesterol, HDL-C, Triglycerides). The "ground truth" for such measurements is typically established by reference laboratory methods (e.g., enzymatic assays, gas chromatography-mass spectrometry (GC-MS), or high-performance liquid chromatography (HPLC)) traceable to international standards (e.g., CDC Lipid Standardization Program). It does not involve human expert interpretation of images or complex diagnostic reasoning.

    4. Adjudication Method for the Test Set

    Not applicable, as ground truth is established by objective laboratory methods, not expert consensus requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    Not applicable. MRMC studies are relevant for imaging devices where human readers interpret medical images. This device is a quantitative diagnostic instrument that measures chemical concentrations.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is an analytical instrument; its "performance" is its ability to accurately and precisely measure analytes. While there's an "algorithm" for calculating LDL-C (Friedewald formula), its performance is assessed through the accuracy of the underlying direct measurements (TC, HDL-C, TG) and comparison of the calculated LDL-C to a reference method, not typically in a "standalone algorithm" context as understood for image analysis AI.

    7. The Type of Ground Truth Used

    The ground truth for a device like the Mission Lipid Panel Monitoring System would be established by:

    • Reference Laboratory Methods: Highly accurate and precise laboratory methods (e.g., those traceable to the CDC Lipid Standardization Program or other certified reference methods) for Total Cholesterol, HDL-C, and Triglycerides.
    • Certified Reference Materials: Use of materials with known, established concentrations of the analytes.

    8. The Sample Size for the Training Set

    Not applicable in the context of AI/ML training. This device is likely based on established electrochemical or photometric principles, not on a machine learning model "trained" on a dataset of patient samples in the AI sense. Performance is validated through analytical studies on patient samples and quality control materials.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As noted above, this is not an AI/ML device that requires a "training set" in that context. Its analytical method is based on established chemical principles. Validation is done against reference methods and materials, as described in point 7.

    Ask a Question

    Ask a specific question about this device

    K Number
    K163406
    Date Cleared
    2017-07-11

    (218 days)

    Product Code
    Regulation Number
    862.1175
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mission® Cholesterol Monitoring System is intended for the quantitative determination of Total Cholesterol. High Density Lipoprotein Cholesterol, and Triglycerides in human capillary whole blood from the fingertip. The Mission Cholesterol Monitoring System is a portable system consisting of the Mission Cholesterol Mission Cholesterol Test Cartridges, Mission Cholesterol Optical Verifier and Mission Cholesterol Control Solution, and is intended to be used by a single person and should not be shared. This system is for in vitro diagnostic use only.

    Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.

    HDL (High Density Lipoprotein Cholesterol) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.

    Trighcerides measurements are used in the diagnosis and treatment with diabetes mellitus, nephrosis, liver obstruction, and other diseases involving lipid metabolism or various endocrine disorders.

    Use this product at the frequency your doctor recommends for testing Total Cholesterol, and Triglycerides.

    An estimated value for Low Density Lipoprotein Cholesterol is calculated by the Mission Cholesterol Meter and is reported only when Triglycerides are <400 mg/dL.

    The Mission® Cholesterol Pro Monitoring System is intended for the quantitative determination of Total Cholesterol. High Density Lipoprotein Cholesterol, and Triglycerides in human capillary whole blood from the fingertip and lithium heparin venous whole blood. The Mission Cholesterol Pro Monitoring System is a portable system consisting of the Mission Cholesterol Pro Meter, Mission Cholesterol Pro Test Cartridges, Mission Cholesterol Optical Verifier, and Mission Cholesterol Control Solution, and is intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices. This system is for in vitro diagnostic use only.

    Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.

    HDL (High Density Lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.

    Triglycerides measurements are used in the diagnosis and treatments with diabetes mellitus. nephrosis, liver obstruction, and other diseases involving lipid metabolism or various endocrine disorders.

    An estimated value for Low Density Lipoprotein Cholesterol is calculated by the Mission Cholesterol Meter and is reported only when Triglycerides are <400 mg/dL.

    Device Description

    Over the Counter (home use): The Mission Cholesterol Monitoring System is a portable system consisting of the Mission Cholesterol Meter, Mission Cholesterol Test Cartridges, Mission Cholesterol Optical Verifier and Mission Cholesterol Control Solution and is intended to be used by a single person and should not be shared.
    The Mission Cholesterol Monitoring System is designed for the quantitative measurement of Total Cholesterol (CHOL), High Density Lipoprotein Cholesterol (HDL) and Triglycerides (TRIG) in capillary whole blood from the fingertip. The Mission Cholesterol Meter is an in vitro diagnostic device consisting of a reflectance photometer that analyzes the intensity and color of light reflected from the reagent area of the test cartridge. This device measures analytes in blood once the blood is applied to dry phase test cartridges that are specifically designed for reflectance analysis.

    Professional: The Mission Cholesterol Pro Monitoring System is a portable system consisting of the Mission Cholesterol Pro Meter. Mission Cholesterol Pro Test Cartridges. Mission Cholesterol Pro Optical Verifier and Mission Cholesterol Pro Control Solution and is intended for professional use in healthcare settings for multiple patient uses.
    The Mission Cholesterol Pro Monitoring System is designed for the quantitative measurement of Total Cholesterol (CHOL), High Density Lipoprotein Cholesterol (HDL) and Triglycerides (TRIG) in capillary and venous human whole blood. The Mission Cholesterol Pro Meter is an in vitro diagnostic device consisting of a reflectance photometer that analyzes the intensity and color of light reflected from the reagent area of the test cartridge. This device measures analytes in blood once the blood is applied to dry phase test cartridges that are specifically designed for reflectance analysis.

    Test Cartridge:
    The Mission Cholesterol (Home use) Test Cartridge is a 3 in 1 Lipid Panel test device used to measure concentration of CHOL, HDL and TRIG in capillary whole blood from the fingertip. A code Chip automatically calibrates the meter with the code number of the cartridges when inserted into the meter.

    The Mission Cholesterol Pro (Professional) Test Cartridge is a 3 in 1 Lipid Panel test device used to measure concentration of CHOL, HDL and TRIG in capillary and venous human whole blood. A code Chip automatically calibrates the meter with the code number of the cartridges when inserted into the meter.

    Control Solution:
    Both the Mission® Cholesterol (home use) Monitoring System's and the Mission® Cholesterol Pro Monitoring System's Control Solutions are used to estimate precision of meter readings for determination of total cholesterol (CHOL), high density lipoprotein cholesterol (HDL) and triglycerides (TRIG).

    Optical verifier:
    Both the Mission® Cholesterol (home use) Monitoring System's and the Mission® Cholesterol Pro Monitoring System's Optical Verifiers are used to verify that the meter functions properly by checking that the meter can detect a pre-calibrated value.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study findings for the Mission Cholesterol Monitoring System, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state formal "acceptance criteria" values in a table. However, the linearity, precision, and bias data are direct measures of performance against presumed internal or regulatory benchmarks for acceptable accuracy and precision in cholesterol measurement devices. We will infer the reported device performance from the study results presented.

    Performance MetricAcceptance Criteria (Implied/Industry Norms)Reported Device Performance - Total CholesterolReported Device Performance - HDL CholesterolReported Device Performance - Triglycerides
    Precision (Repeatability %CV)Generally <5% for diagnostic devices (Level 1, Level 2)Level 1: 1.5%, Level 2: 1.6%Level 1: 2.0%, Level 2: 1.6%Level 1: 1.3%, Level 2: 1.6%
    Precision (Total Precision %CV)Generally <10% for diagnostic devices (Level 1, Level 2)Level 1: 2.0%, Level 2: 1.8%Level 1: 2.7%, Level 2: 2.3%Level 1: 2.0%, Level 2: 1.9%
    Linearity (Measurement Range)Device should be linear within its claimed measurement range100-400 mg/dL (linear detection range)15-100 mg/dL (linear detection range)45-650 mg/dL (linear detection range)
    Correlation ($R^2$) - Layperson, Finger CapillaryTypically >0.90 for good correlation0.98460.97680.9934
    Correlation ($R^2$) - Professional, Finger CapillaryTypically >0.90 for good correlation0.98830.97780.9948
    Correlation ($R^2$) - Professional, Heparin VenousTypically >0.90 for good correlation0.98630.97900.9936
    % Bias at Medical Decision Points (Finger Blood)Generally within ±5-10% for clinical acceptability(-0.1% at 200mg/dL, -0.1% at 240mg/dL)(-0.3% at 40mg/dL, -0.1% at 60mg/dL)(0.3% at 150mg/dL, 0.0% at 200mg/dL, -0.2% at 500mg/dL)
    % Bias at Medical Decision Points (Venous Blood)Generally within ±5-10% for clinical acceptability(-0.7% at 200mg/dL, -0.8% at 240mg/dL)(1.9% at 40mg/dL, 1.0% at 60mg/dL)(-1.8% at 150mg/dL, -1.5% at 200mg/dL, -1.2% at 500mg/dL)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: A total of 369 patients were recruited for the clinical study.
    • Data Provenance: The study was conducted at three point-of-care (POC) sites located in different geographical locations. It is a prospective clinical study as patients were recruited and tested specifically for this study. The country of origin is not explicitly stated, but given the FDA submission, it can be inferred to be compliant with US regulatory standards, likely involving data collected within the US or under equivalent standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not explicitly state the number of experts. However, it mentions:

    • "For professional use, patient blood was collected from fingertip and venous blood draws. Capillary blood samples from fingertip were tested at clinical sites by professionals. Venous blood samples were tested at a laboratory using an FDA cleared method."
    • "The plasma concentration was confirmed by the reference method."

    This implies that trained "professionals" (likely medical technicians or laboratory personnel) performed the official comparative testing, and an FDA-cleared reference method was used to establish the ground truth. No specific number or qualification (e.g., "radiologist with 10 years of experience") for these "professionals" is given beyond their professional role in clinical sites or laboratories.

    4. Adjudication Method for the Test Set

    The document does not describe a formal adjudication method (like 2+1 or 3+1). The ground truth was established by comparing the device's results to an "FDA cleared method" and a "reference method" in laboratory settings. This suggests direct quantitative comparison rather than a consensus-based adjudication process typical for qualitative or image-based diagnostics.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly mentioned or performed in the context of improving human reader performance with AI assistance. This study focuses on the device's standalone analytical performance compared to a reference method, and also includes layperson use, not assisted reading by human experts.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, a standalone performance study was done for the device. The entire clinical study, comparing the Mission Cholesterol Monitoring System's measurements against a reference method, evaluates the algorithm's performance. Both professional operators and laypersons used the device independently to obtain results, which were then compared to the established ground truth.

    7. The Type of Ground Truth Used

    The ground truth for the clinical study was established by comparing the device's results to an FDA cleared method and a reference method (for plasma concentration).

    8. The Sample Size for the Training Set

    The document does not provide information regarding a separate training set or its sample size. The studies described are for the validation of an already developed device.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is described, there's no information on how its ground truth would have been established. The provided document focuses on the validation of the device through performance and clinical studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151545
    Date Cleared
    2016-12-22

    (562 days)

    Product Code
    Regulation Number
    862.1175
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CardioChek Home Test System (consisting of the CardioChek Home analyzer and CardioChek Home Lipid Panel test strips) is for the quantitative determination of total cholesterol. HDL (high density lipoprotein) cholesterol and triglycerides in capillary whole blood from the fingertip and is intended to be used by a single person and should not be shared. This system is for in vitro diagnostic use only.

    • o Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.
    • HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes ● mellitus), atherosclerosis, and various liver and renal diseases.
    • Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism or various endocrine disorders.
      A Chol/HDL ration and estimated values for LDL (low density lipoprotein) cholesterol are calculated by the CardioChek Home Analyzer.

    The CardioChek Plus Test System (consisting of the CardioChek Plus analyzer and PTS Panels Lipid Panel test strips) is for the quantitative determination of total cholesterol. HDL (high density lipoprotein) cholesterol and triglycerides in venous whole blood and capillary whole blood from the fingertip and is intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices. This system is for in vitro diagnostic use only.

    • o Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.
    • HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes ● mellitus), atherosclerosis, and various liver and renal diseases.
    • Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellitus, ● nephrosis, liver obstruction, other diseases involving lipid metabolism or various endocrine disorders.
      A Chol/HDL ration and estimated values for LDL (low density lipoprotein) cholesterol are calculated by the CardioChek Plus Analyzer.

    The CardioChek PA Test System (consisting of the CardioChek PA analyzer and PTS Panels Lipid Panel test strips) is for the quantitative determination of total cholesterol, HDL (high density lipoprotein) cholesterol and triglycerides in venous whole blood and capillary whole blood from the fingertip and is intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices. This system is for in vitro diagnostic use only.

    • Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol . in the blood and lipid and lipoprotein metabolism disorders.
    • HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
    • Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism or various endocrine disorders.
      A Chol/HDL ration and estimated values for LDL (low density lipoprotein) cholesterol are calculated by the CardioChek PA Analyzer.

    The CardioChek PA Home Test System (consisting of the CardioChek PA Home analyzer and CardioChek Home Livid Panel test strips) is for the quantitative determination of total cholesterol. HDL (high density lipoprotein) cholesterol and triglycerides in capillary whole blood from the fingertip and is intended to be used by a single person and should not be shared. This system is for in vitro diagnostic use only.

    • Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol o in the blood and lipid and lipoprotein metabolism disorders.
    • HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
    • Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism or various endocrine disorders.
      A Chol/HDL ration and estimated values for LDL (low density lipoprotein) cholesterol are calculated by the CardioChek PA Home Analyzer.
    Device Description

    The PTS Panels Lipid Panel test strips and the CardioChek Home Lipid Panel test strips are used with the CardioChek Plus, CardioChek PA, CardioChek Home and CardioChek PA Home analyzers to measure total cholesterol and triglycerides in whole blood. The test strips utilize enzymatic methods on a dry strip that is read by reflectance photometry. These test strips are for in vitro diagnostic use only.

    AI/ML Overview

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the CardioChek Test Systems (CardioChek Plus, CardioChek Home, CardioChek PA, CardioChek PA Home) involve demonstrating accuracy and precision for HDL cholesterol measurements within a modified dynamic range of 20-120 mg/dL. The reported performance is based on comparison to a reference laboratory method (Roche Cobas Integra 400 plus) and precision studies.

    Performance MetricAcceptance Criteria (Implied by equivalence claim and historical device performance)Reported Device Performance
    Accuracy (Correlation to Reference Method)High correlation (r ≥ 0.90 typically expected for clinical agreement, though not explicitly stated as a statistical threshold)CardioChek Plus Home: y = 0.99x + 0.55, r = 0.98
    CardioChek Plus: y = 0.99x + 0.55, r = 0.98
    CardioChek PA: y = 0.93x + 0.98, r = 0.98
    CardioChek PA Home: y = 0.93x + 0.98, r = 0.98
    Precision (Coefficient of Variation, CV%)Generally, CV% values for clinical chemistry analytes are expected to be low, typically < 5-10% depending on the analyte and concentration level. No explicit numerical criteria are given in the document.CardioChek PA HDL:
    Level 1 (38.3 mg/dL): CV = 4.3%
    Level 2 (62.4 mg/dL): CV = 3.6%
    Level 3 (106.0 mg/dL): CV = 4.0%
    CardioChek PA Home HDL:
    Level 1 (38.3 mg/dL): CV = 4.3%
    Level 2 (62.4 mg/dL): CV = 3.6%
    Level 3 (106.0 mg/dL): CV = 4.0%
    CardioChek Plus HDL:
    Level 1 (39.5 mg/dL): CV = 4.1%
    Level 2 (63.3 mg/dL): CV = 4.2%
    Level 3 (108.3 mg/dL): CV = 4.8%
    CardioChek Home HDL:
    Level 1 (39.5 mg/dL): CV = 4.1%
    Level 2 (63.3 mg/dL): CV = 4.2%
    Level 3 (108.3 mg/dL): CV = 4.8%
    Measuring Range (Dynamic Range)20-120 mg/dL HDL cholesterolDemonstrated with samples in the range of 21 to 112 mg/dL HDL cholesterol (for accuracy study) and various levels within the range for precision.

    2. Sample Size and Data Provenance for the Test Set

    • Accuracy Test Set:
      • Sample Size: 80 samples for each analyzer type (CardioChek Plus Home, CardioChek Plus, CardioChek PA, CardioChek PA Home).
      • Data Provenance: Not explicitly stated, but based on the nature of medical device studies for FDA submission, it would typically be prospective data collected for the purpose of the study. The country of origin is not specified, but the submission is to the U.S. FDA.
    • Precision Test Set:
      • Sample Size: 80 observations for Level 1 and Level 2 for all analyzers, and 80 observations (CardioChek PA/PA Home) or 78 observations (CardioChek Plus/Home) for Level 3. These observations were obtained by "three operators tested three levels of whole blood samples on five analyzers... over three time periods."
      • Data Provenance: Not explicitly stated, but likely prospective data collected in a controlled laboratory setting.

    3. Number of Experts and Qualifications for Ground Truth for the Test Set

    • This relates to a diagnostic device for quantitative measurement, not an AI classification system requiring expert interpretation for ground truth.
    • For Accuracy: The ground truth for HDL cholesterol measurements was established using a Roche Cobas Integra 400 plus HDL Cholesterol instrument, which is a recognized reference method in clinical chemistry. This is a laboratory analyzer, not an expert panel.
    • For Precision: The "ground truth" is the mean value obtained from repeated measurements of the same whole blood samples.

    4. Adjudication Method for the Test Set

    • Not applicable as this is a quantitative measurement device study, not an AI diagnostic study requiring human expert adjudication of interpretations. The "ground truth" is derived from a reference laboratory instrument.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted as this is a study for a quantitative measurement device (lipid panel test system), not an AI algorithm intended for interpretation by human readers. The context is not one of human-in-the-loop performance improvement with AI assistance.

    6. Standalone (Algorithm Only) Performance Study

    • Yes, the accuracy and precision studies performed can be considered standalone performance of the device (analyzer with test strips), as it evaluates the device's ability to accurately and precisely measure HDL cholesterol compared to a reference method, without human interpretation as part of the primary output.

    7. Type of Ground Truth Used

    • Accuracy: The ground truth for the accuracy study was established using a reference laboratory instrument (Roche Cobas Integra 400 plus HDL Cholesterol). This is considered a highly reliable, objective, and standardized method for clinical chemistry analytes.
    • Precision: The ground truth for precision is the central tendency (mean) of repeated measurements of controlled samples.

    8. Sample Size for the Training Set

    • This document describes performance testing for a medical device (CardioChek Test System) that determines quantitative values using enzymatic methods on dry strips read by reflectance photometry. It is a traditional in vitro diagnostic device, not an AI/machine learning algorithm that requires a "training set" in the conventional sense for model development. The device's calibration parameters are programmed into a lot-specific memory chip. Therefore, there is no "training set" in the context of machine learning model development. The calibration and manufacturing processes would involve internal quality control and standardization.

    9. How the Ground Truth for the Training Set Was Established

    • As there is no "training set" for an AI/machine learning model, this question is not applicable. The device's operational parameters and calibration are established through manufacturing processes and validated through studies like those described (accuracy and precision).
    Ask a Question

    Ask a specific question about this device

    K Number
    K150654
    Device Name
    Cholesterol
    Date Cleared
    2015-09-29

    (200 days)

    Product Code
    Regulation Number
    862.1175
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For the quantitative in vitro determination of Cholesterol in serum and plasma. Cholesterol measurements are used in the diagnosis and treatments of disorders involving excess cholesterol in the blood and lipoprotein metabolism disorders.

    Device Description

    The Cholesterol kit assay consists of ready to use reagent solutions.

    CATALOGUE NUMBER: CH8310

    R1. Reagent 4 x 20 ml

    REAGENT COMPOSITION

    Contents: R1. Reagent 4-Aminoantipyrine, Phenol, Peroxidase (E.C.1.11.1.7, Horse Radish, +25°C), Cholesterol esterase (E.C.3.1.1.13. Pseudomonas, +37°C), Cholesterol oxidase (E.C.1.1.3.6. Nocardia, +37°C), Sodium Azide

    Concentrations in the Test: 0.25 mmol/l, 6.00 mmol/l, >=0.50 U/ml, >= 0.20 U/ml, >=0.10 U/ml, 0.09%

    MATERIALS REQUIRED BUT NOT PROVIDED: Randox Assayed Multisera Level 2 (Cat. No. HN 1530) and Level 3 (Cat. No. HE 1532); 510(k) # K942458, Randox Calibration Serum Level 3 (Cat. No. CAL 2351); 510(k) # K053153, RX series Saline (Cat. No. SA 8396)

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Randox Laboratories Ltd. Cholesterol device, based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state "acceptance criteria" for all tests in a single table, but rather presents the performance results of various analytical studies that demonstrate the device's capability. I've compiled the relevant performance data from the document into a table, noting the implicit acceptance measures (e.g., meeting CLSI guidelines, certain correlation coefficients, or imprecision percentages).

    Metric / StudyAcceptance Criteria (Implicit)Reported Device Performance
    PrecisionPerformed consistent with CLSI EP5-A2. Total CV % for controls and patient samples to be within acceptable limits (typically < 10% for diagnostic assays, often tighter for specific analytes, but not explicitly stated here for cholesterol in the context of acceptance).Lot 1:- Control (283 mg/dl): Total CV 2.0%- Control (307 mg/dl): Total CV 1.8%- Control (190 mg/dl): Total CV 2.0%- Patient Sample (176 mg/dl): Total CV 2.5%- Patient Sample (226 mg/dl): Total CV 2.4%- Patient Sample (270 mg/dl): Total CV 2.1%- Patient Sample (586 mg/dl): Total CV 2.0%- Sensitivity Pool (33.2 mg/dl): Total CV 8.8%Lot 2:- Control (285 mg/dl): Total CV 2.0%- Control (310 mg/dl): Total CV 2.3%- Control (192 mg/dl): Total CV 2.4%- Patient Sample (177 mg/dl): Total CV 2.7%- Patient Sample (228 mg/dl): Total CV 2.7%- Patient Sample (272 mg/dl): Total CV 2.7%- Patient Sample (592 mg/dl): Total CV 1.9%- Sensitivity Pool (32.4 mg/dl): Total CV 10.3%
    Linearity / Reportable RangePerformed consistent with CLSI EP6-A. Deviation from linearity < 5%.Linearity: Up to 618 mg/dlReportable Range: 25 – 618 mg/dlRegression (approx. from graph): Slope ~0.99, Intercept ~-3.71, r = 0.999, Syx = 4.85
    Detection Limit (LoD)Performed consistent with CLSI EP17-A2.LoD: 6.31 mg/dl
    Limit of Blank (LoB)Performed consistent with CLSI EP17-A2.LoB: 3.1 mg/dl
    Limit of Quantitation (LoQ)Lowest concentration detected with ≤20% imprecision.LoQ: 23.2 mg/dl
    Analytical Specificity (Interference)% of Control ± 10% for tested interferents.Haemoglobin: No significant interference up to 750mg/dLTotal Bilirubin: No significant interference up to 60mg/dLConjugate Bilirubin: No significant interference up to 60mg/dLIntralipid®: No significant interference up to 1000mg/dLAscorbic Acid: No significant interference up to 6mg/dL
    Method Comparison (vs. Predicate)Performed consistent with CLSI EP9-A2. High correlation coefficient (typically r > 0.975 for quantitative assays) and acceptable regression equation (slope close to 1, intercept close to 0) indicating substantial equivalence.Serum samples (vs. Predicate): Y = 1.00x - 4.77, r = 0.997
    Matrix Comparison (Li Heparin)High correlation coefficient (typically r > 0.975) and acceptable regression equation (slope close to 1, intercept close to 0) demonstrating equivalence between serum and lithium heparin plasma.Serum vs. Li Heparin: Y = 1.01x - 6.54, r = 0.997
    Matrix Comparison (K₂EDTA)High correlation coefficient (typically r > 0.975) and acceptable regression equation (slope close to 1, intercept close to 0) demonstrating equivalence between serum and K₂EDTA plasma.Serum vs. K₂EDTA: Y = 0.99x + 2.85, r = 0.998

    2. Sample Sizes and Data Provenance for the Test Set

    • Precision/Reproducibility:
      • Controls: Not explicitly stated as "sample size" but data is reported for commercial control materials (717UE, 724UE, 952UN).
      • Patient Samples: 4 concentrations of unaltered human serum samples (3 diluted, 1 spiked for Linearity Pool, 1 Sensitivity Pool). Each sample run in 2 replicates per run, twice per day for 20 non-consecutive days, using 2 reagent lots on 2 RX Daytona plus systems.
      • Data Provenance: "unaltered human serum samples" implies human origin, likely retrospective for spiking/dilution. No country of origin is specified.
    • Linearity/Assay Reportable Range:
      • Sample Size: 11 levels of samples covering the measuring range. Each level run in 5 replicates.
      • Data Provenance: "linearity samples" were prepared. Implies in-vitro prepared samples to cover the range, likely based on human serum/plasma.
    • Detection Limit (LoD), Limit of Blank (LoB), Limit of Quantitation (LoQ):
      • Sample Size: LoD was based on 240 determinations with 4 low-level samples.
      • Data Provenance: Not specified, but generally prepared samples for low-level determination.
    • Analytical Specificity (Interference):
      • Sample Size: Not explicitly stated for the number of interferent samples, but tested at Cholesterol concentrations of 150 mg/dl and 250 mg/dl for each interferent.
      • Data Provenance: Prepared samples spiked with interferents.
    • Method Comparison with Predicate Device:
      • Sample Size: 107 serum patient samples.
      • Data Provenance: "serum patient samples" spanning 25 to 599 mg/dl. Retrospective. No country of origin specified.
    • Matrix Comparison:
      • Sample Size (Lithium Heparin): Minimum of 54 matched patient sample pairs (serum vs. lithium heparin plasma).
      • Sample Size (Potassium 2 EDTA): Minimum of 50 matched patient sample pairs (serum vs. potassium 2 EDTA plasma).
      • Data Provenance: "Patient samples were drawn in matched pairs". Retrospective from human subjects. No country of origin specified.

    3. Number of Experts and Qualifications for Ground Truth for the Test Set

    This device is an in vitro diagnostic (IVD) for quantitative measurement of cholesterol. The "ground truth" for such devices is established by precise laboratory reference methods or established commercially available controls and calibrators with known values.

    • No "experts" in the sense of radiologists or pathologists establishing ground truth as would be the case for imaging devices.
    • Ground truth is established by:
      • Reference materials (e.g., NIST 1952a for the calibrators, mentioned under traceability).
      • Established analytical methods used by the predicate device and in clinical laboratories.
      • CLSI guidelines for experimental design and data analysis.

    4. Adjudication Method for the Test Set

    Not applicable for this type of quantitative IVD device. Adjudication methods (like 2+1, 3+1) are typically used for qualitative or semi-quantitative assessments, especially in imaging or pathology, where human expert discrepancy resolution is needed. For quantitative chemical measurements, the ground truth is often numerical and objectively determined.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This is an in vitro diagnostic device for chemical analysis of cholesterol, not an imaging or qualitative assessment device involving human readers. Therefore, an MRMC study is not relevant.

    6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study

    Yes, the entire performance evaluation presented is a standalone study of the device (Cholesterol assay on the RX Daytona plus system). The device performs the analytical measurement autonomously once the sample is loaded. The studies demonstrate the analytical performance of the device itself.

    7. Type of Ground Truth Used

    The ground truth for the performance studies is multi-faceted:

    • Reference Materials: Randox Calibration Serum Level 3 is traceable to Cholesterol reference material NIST 1952a. This is a primary ground truth for calibration and accuracy.
    • Predicate Device: For method comparison studies, the predicate device (Randox Cholesterol reagent, K923504) serves as a comparative ground truth, aiming to demonstrate substantial equivalence rather than absolute biological truth.
    • CLSI Guidelines: The studies adhere to CLSI (Clinical and Laboratory Standards Institute) guidelines (EP5-A2 for precision, EP6-A for linearity, EP17-A2 for detection limits, EP9-A2 for method comparison), which represent an industry-accepted "ground truth" for how these analytical performance characteristics should be determined and evaluated.
    • Prepared Samples: For linearity, sensitivity, detection limits, and interference, samples were prepared to known concentrations or spiked with known substances to create specific "ground truth" scenarios.

    8. Sample Size for the Training Set

    There is no mention of a "training set" in the context of machine learning or AI, as this device is a traditional in vitro diagnostic reagent system, not an AI/ML-based device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an AI/ML algorithm in this context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132031
    Manufacturer
    Date Cleared
    2014-03-21

    (246 days)

    Product Code
    Regulation Number
    862.1175
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Afinion™ Lipid Panel is an in vitro diagnostic test for quantitative determination of total cholesterol (Chol), high-density lipoprotein (HDL) cholesterol and triglycerides (Trig) in serum. Values for low-density lipoprotein (LDL) cholesterol, non-HDL cholesterol and Chol/HDL ratio are calculated by the Afinion™ AS100 Analyzer. Chol, HDL cholesterol, Trig, and calculated LDL cholesterol, non-HDL cholesterol and Chol/HDL ratio) are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.

    Afinion™ Lipid Panel Control has been designed for use with the Afinion™ AS100 Analyzer and Afinion™ Lipid Panel. Afinion™ Lipid Panel Control is intended for use as assayed control material for total cholesterol (Chol), high-density lipoprotein (HDL) cholesterol and triglycerides (Trig). The controls should be used to confirm that the Afinion™ AS100 Analyzer System is working properly and provides reliable results.

    For use in clinical laboratories and point of care laboratory settings.

    For prescription use only.

    Device Description

    Afinion™ Lipid Panel is a fully automated assay for quantitative determination of Chol. HDL and Trig in serum. LDL, non- HDL and Chol/HDL are calculated by the Afinion™ AS100 Analyzer.

    The Afinion™ Lipid Panel Test Cartridge contains all reagents necessary for determination of Chol, HDL and Trig in serum. The sampling device integrated in the test cartridge is filled with sample material. The test cartridge is then placed in the Afinion™ AS100 Analyzer. The analyzer inspects the sampling device, and the sample is then diluted.

    Total Cholesterol (Chol) is measured by an enzymatic colorimetric method.
    Triglycerides (Trig) are measured by an enzymatic colorimetric method.
    HDL cholesterol is measured by an enzymatic colorimetric method with direct determination of HDL by initial antibody blocking of apolipoprotein B (apo-B).
    LDL cholesterol is calculated by use of the Friedwald formula: LDL (mg/dL) = Chol - HDL - Trig/5.
    non-HDL cholesterol is calculated as total cholesterol minus HDL: non-HDL = Chol - HDL.
    Chol/HDL ratio is calculated as Total Cholesterol/ HDL Cholesterol.

    AI/ML Overview

    Acceptance Criteria and Device Performance for Afinion™ Lipid Panel

    This document outlines the acceptance criteria and the study that demonstrates the Afinion™ Lipid Panel's performance, as derived from the provided 510(k) summary (K132031).

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly state pre-defined acceptance criteria values for bias, precision (repeatability and within-device), or linearity. Instead, it presents the results of these studies and implies that these results were considered acceptable for demonstrating substantial equivalence to predicate devices. The reported device performance based on acceptable outcomes from comparison studies is presented below.

    Interference: No significant interference (<10%) was observed from 26 common substances at specified concentrations. Limitations were noted for Calcium dobesilate, Methyldopa, Acetylcysteine, and Levodopa at certain levels.

    Reporting Ranges (supported by linearity and LoQ studies):

    AnalyteReportable Range (mg/dL)Linearity Demonstrated (mg/dL)
    Total Cholesterol100-50077-511
    Triglycerides45-65036-691
    HDL Cholesterol15-10014-111

    Accuracy (Method Comparison with Predicate Devices):

    AnalyteInterceptSlopeCorrelation Coefficient (r)
    Chol-4.5 mg/dL1.040.99
    Trig-11.4 mg/dL1.041.00
    HDL-2.1 mg/dL1.040.98

    Bias at Medical Decision Levels (Implied Acceptance: Low Bias):

    AnalyteConcentration Level (mg/dL)Bias (mg/dL)Bias (%)
    Trig150-5.0-3.3
    Trig200-2.8-1.4
    Trig5009.92.0
    Chol2002.61.3
    Chol2404.01.7
    Chol4009.72.4
    HDL40-0.6-1.6
    HDL600.10.1
    HDL800.81.0

    Precision (Repeatability and Within-device %CV - Implied Acceptance: Low %CV):

    Precision results are presented for control samples at two levels and one serum sample across three sites. The Coefficients of Variation (CV%) are generally low, indicating good precision. For example:

    • Total Cholesterol: Repeatability CVs range from 1.7% to 3.5%, Within-device CVs range from 2.3% to 3.9%.
    • HDL Cholesterol: Repeatability CVs range from 2.1% to 3.9%, Within-device CVs range from 2.6% to 4.9%.
    • Triglycerides: Repeatability CVs range from 1.8% to 4.4%, Within-device CVs range from 2.2% to 4.9%.

    2. Sample Sizes and Data Provenance

    • Linearity Testing:

      • Test Set Sample Size: 11 concentration levels for each analyte, produced by intermixing one low and one high serum sample. Each level was measured in 4-6 replicates.
      • Data Provenance: Not explicitly stated, but the studies were performed by the manufacturer, Axis-Shield PoC AS (located in Oslo, Norway). The samples were described as "serum samples." It's retrospective in the sense that it's test data generated for regulatory submission, but the samples themselves could have been collected prospectively or retrospectively.
    • Limits of Quantitation (LoQ) Testing:

      • Test Set Sample Size: 5 samples with concentrations near 0 mg/dL (LoB samples) and 5 low concentration samples (LoD samples). Each sample was measured in a total of 60 replicates (likely 12 replicates per sample, using 3 analyzers and 2 test cartridge lots).
      • Data Provenance: Not explicitly stated, but performed by the manufacturer. "Serum" is the sample type.
    • Analytical Specificity (Interference) Testing:

      • Test Set Sample Size: "Samples covering two medical decision concentrations of each lipid analyte" were measured.
      • Data Provenance: Not explicitly stated, but performed by the manufacturer.
    • Accuracy (Method Comparison) Testing:

      • Test Set Sample Size:
        • Cholesterol: 348 samples
        • Triglycerides: 246 samples
        • HDL: 251 samples
      • Data Provenance: The study was "performed at four point-of-care sites." No specific country of origin is mentioned, but the manufacturer is based in Norway. The samples were "serum." The nature of sample collection (retrospective or prospective) is not specified.
    • Precision Testing:

      • Test Set Sample Size: Two control samples and one serum sample were tested. For each sample, 80 replicates were performed at each of the three point-of-care sites (2 replicates per run, 2 runs per day for 20 days).
      • Data Provenance: "Performed at three point-of-care sites." The origin of the control and serum samples is not detailed, but the study was conducted by the manufacturer.

    3. Number of Experts and Qualifications for Ground Truth

    This device is an in vitro diagnostic (IVD) for quantitative determination of analytes in serum. The ground truth for such devices is established by reference methods or highly accurate laboratory methods, not by expert interpretation of images or clinical assessments.

    • Traceability:
      • Cholesterol (Chol) and HDL are traceable to the National Reference System for Cholesterol (NRS/CHOL).
      • Triglycerides (Trig) are traceable to a Centers for Disease Control and Prevention (CDC) reference method.
      • The device is CRMLN certified for Total Cholesterol and HDL Cholesterol, indicating its accuracy against reference measurement procedures.

    Therefore, "experts" in the traditional sense (e.g., radiologists) are not used to establish ground truth for this type of device. The ground truth is established by recognized reference standards and methods in clinical chemistry.

    4. Adjudication Method for the Test Set

    Adjudication methods (like 2+1, 3+1) are typically used in studies involving human interpretation or subjective assessments (e.g., image reading). This is a quantitative diagnostic device, and the ground truth is established by objective, highly accurate reference methods or laboratory instruments. Therefore, no "adjudication method" in this context is applicable or described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that assist human readers in interpreting complex data (e.g., medical images). The Afinion™ Lipid Panel is a standalone quantitative measurement device, not an AI-assisted human reading system.

    6. Standalone Performance Done

    Yes, a standalone performance study was done. The studies detailed (linearity, limits of quantitation, analytical specificity, accuracy/method comparison, and precision) all represent the performance of the Afinion™ Lipid Panel device (algorithm only, without human-in-the-loop performance) in measuring lipid levels in serum samples.

    7. Type of Ground Truth Used

    The ground truth used is based on:

    • Reference Methods: Specifically, the National Reference System for Cholesterol (NRS/CHOL) for Total Cholesterol and HDL Cholesterol, and a CDC reference method for Triglycerides.
    • Comparison to Predicate Devices/Automated Laboratory Methods (CM): For accuracy evaluation, the Afinion™ Lipid Panel's results were compared against an "automated laboratory method (CM)" for Chol, Trig, and HDL, which are themselves established and validated lab instruments.

    8. Sample Size for the Training Set

    This document describes a 510(k) submission for an in vitro diagnostic device that measures specific analytes. It is highly unlikely that this device uses machine learning or AI models that require a "training set" in the conventional sense (i.e., iterative learning from labeled data). The device's operation is based on established enzymatic colorimetric methods and pre-programmed algorithms. Therefore, a "training set" size is not applicable or stated in this context.

    9. How the Ground Truth for the Training Set Was Established

    As stated in point 8, a "training set" as understood in machine learning is not applicable to this type of IVD device. The methods for establishing the device's accuracy and performance are described under "Traceability" and "Accuracy" (method comparison with reference and predicate methods).

    Ask a Question

    Ask a specific question about this device

    K Number
    K130200
    Manufacturer
    Date Cleared
    2013-02-15

    (18 days)

    Product Code
    Regulation Number
    862.1175
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Piccolo Total Cholesterol - Capillary Test System used with the Piccolo xpress Chemistry Analyzer is intended for the in vitro quantitative determination of total cholesterol in capillary (fingerstick) heparinized whole blood in a clinical laboratory setting or point-of-care location.

    Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein disorders.

    Device Description

    The Piccolo® Lipid Panel - Capillary Reagent Disc (which contains the Piccolo® Total Cholesterol - Capillary Test System) is designed to separate a heparinized whole blood sample into plasma and blood cells. The disc meters the required quantity of plasma and diluent, mixes the plasma with diluent, and delivers the mixture to the reaction cuvettes along the disc perimeter. The diluted plasma mixes with the reagent beads, initiating the chemical reactions that are then monitored by the analyzer.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Device: Piccolo® Total Cholesterol - Capillary Test System


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific MetricAcceptance Criteria (Implied/Stated)Reported Device Performance
    Linearity/Dynamic RangeLower LimitNot explicitly stated as an acceptance criterion, but established as 20 mg/dL20 mg/dL
    Upper LimitNot explicitly stated as an acceptance criterion, but established as 520 mg/dL520 mg/dL
    Slope (compared to reference method)Not explicitly stated, but typically expected to be close to 10.989
    Intercept (compared to reference method)Not explicitly stated, but typically expected to be close to 019.7
    Correlation Coefficient (r)Not explicitly stated, but typically >0.99 for good linearity0.999
    PrecisionWithin-Run %CV (Low Concentration e.g. Serum 1)Not explicitly stated, but typically <5% for many analytes1.3% (for mean 223.7 mg/dL)
    Within-Run %CV (High Concentration e.g. Serum 2)Not explicitly stated, but typically <5% for many analytes1.5% (for mean 202.2 mg/dL)
    Total %CV (Low Concentration e.g. Serum 1)Not explicitly stated, but typically <5% or <10% for many analytes2.6% (for mean 223.7 mg/dL)
    Total %CV (High Concentration e.g. Serum 2)Not explicitly stated, but typically <5% or <10% for many analytes2.2% (for mean 202.2 mg/dL)
    Whole Blood Precision %CVNot explicitly stated, but typically <5% or <10% for many analytes1.4% - 1.7% across 5 samples (Mean range: 184.2 - 239.1 mg/dL)
    Method ComparisonCorrelation Coefficient (R²)Not explicitly stated, but typically >0.95 or >0.98 for good correlationSite 1: 0.991 (Linear & Deming), Site 2: 0.991 (Linear & Deming), Site 3: 0.987 (Linear & Deming), Combined: 0.989 (Linear & Deming). These very high R² values indicate excellent correlation.
    Slope (Linear Regression)Not explicitly stated, but expected to be close to 1 within a tight CISite 1: 0.97 (0.95 to 0.98 CI), Site 2: 0.96 (0.95 to 0.97 CI), Site 3: 0.96 (0.95 to 0.98 CI), Combined: 0.96 (0.95 to 0.97 CI). These are all very close to 1.
    Intercept (Linear Regression)Not explicitly stated, but expected to be close to 0 within a tight CISite 1: 2.42 (-0.10 to 4.95 CI), Site 2: 2.83 (0.28 to 5.37 CI), Site 3: 1.30 (-1.73 to 4.33 CI), Combined: 2.20 (0.64 to 3.76 CI). These are all close to 0, often including 0 in their respective 95% CIs.
    AccuracyCRMLN CertificationCertification by CRMLN (Cholesterol Reference Method Laboratory Network)Accuracy was established by completing the certification process of the CRMLN. (This implies the device meets the accuracy standards required for CRMLN certification.)

    Note: The document implicitly defines successful performance by demonstrating strong correlation, linearity, and acceptable precision in comparison to the predicate device and established reference methods. Explicit numerical acceptance criteria (e.g., "R² must be >0.98") are not stated in this summary, but the reported results clearly meet the standards for substantial equivalence.


    2. Sample Size Used for the Test Set and Data Provenance

    • Linearity Test: Samples across the measuring range were used. The specific number of distinct samples isn't given, but testing was done on 20 Piccolo analyzers. The data provenance is not explicitly stated (e.g., country of origin, retrospective/prospective), but it would typically be prospective for device validation.
    • Precision Test:
      • Serum Samples for Within-Run and Total Precision: 160 measurements for each serum level (Serum 1 & Serum 2).
      • Whole Blood Precision: 5 fresh whole blood samples were tested 7 times each on 4 analyzers over 3 hours. This means a total of 5 * 7 = 35 measurements per analyzer, and 35 * 4 = 140 measurements in total for whole blood precision (though the table states 28 for each sample). The document also states "a total of 20 analyzers were used" for this specific test, which introduces some ambiguity on the total number of data points if each sample was run on all 20 analyzers.
      • The data provenance is not explicitly stated.
    • Method Comparison Test:
      • Site 1: 216 samples
      • Site 2: 210 samples
      • Site 3: 213 samples
      • Combined Data: 639 samples
      • The data provenance (e.g., country of origin, retrospective or prospective) for these clinical samples is not explicitly mentioned in the provided text. Based on the context of a 510(k) submission, these would typically be prospectively collected samples from a clinical laboratory or point-of-care setting.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The ground truth for the device's performance (specifically for method comparison) was established by comparison to the Roche Cholesterol Test on the Cobas 6000 Analyzer.

    For accuracy, the device was certified by the CRMLN (Cholesterol Reference Method Laboratory Network). This implies adherence to established reference methods and standards, which are overseen by expert bodies but doesn't involve individual experts establishing ground truth for each test case in the same way as, for example, image interpretation. The "experts" are essentially the established reference methods and organizations that define accurate cholesterol measurement.


    4. Adjudication Method for the Test Set

    Not applicable. This device is an in-vitro diagnostic (IVD) quantitative test system, not an interpretive device like an AI-powered diagnostic imaging tool that would require human expert adjudication of results. The "ground truth" is analytical (e.g., reference assay results, certified reference materials).


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This type of study design is typically used for interpretive diagnostic devices where human readers (e.g., radiologists) interpret cases with and without AI assistance to measure the AI's impact on human performance. This device is an automated quantitative test system.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the studies presented (Linearity, Precision, Method Comparison) represent the standalone performance of the Piccolo® Total Cholesterol - Capillary Test System as an algorithm/device-only system. Its output is a quantitative value, not an interpretation requiring human input for its direct result.


    7. The Type of Ground Truth Used

    • For Linearity, Precision, and Method Comparison: The ground truth was established by comparison to:
      • Reference materials or samples of known concentration (for linearity, precision).
      • A legally marketed predicate device: The Roche Cholesterol Test on the Cobas 6000 Analyzer (for method comparison).
    • For Accuracy: The ground truth was established by "completing the certification process of the CRMLN" (Cholesterol Reference Method Laboratory Network). CRMLN certification signifies that the method is traceable to the Centers for Disease Control and Prevention (CDC) reference method for cholesterol.

    8. The Sample Size for the Training Set

    The provided document describes validation studies (linearity, precision, method comparison) for a predicate-based 510(k) submission for an IVD device. It does not mention any "training set" in the context of machine learning. The device is a chemical analyzer using enzymatic endpoint reactions, not an AI/ML algorithm that requires a training set in the typical sense.


    9. How the Ground Truth for the Training Set Was Established

    As this is not an AI/ML device, the concept of a "training set" and its associated ground truth establishment is not applicable as described in the document. The device's operational parameters (e.g., reagent formulations, reaction conditions, calibration) would be developed and optimized through standard analytical chemistry and engineering practices, not machine learning training. Calibration is done via a barcode with factory-calibrated lot-specific data.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 6