(93 days)
Rodo Abutment System is intended to be used in conjunction with compatible implant systems in the maxillary or mandibular arch to provide support for crowns, bridges or overdentures.
The Rodo Abutment System includes the Rodo Abutment, Smileloc Sleeve, Titanium Coping, Temporary Cap, abutment screws, the Smileloc Activator (or Smileloc Remover) (all cleared under K160786) and Smilekey (cleared under K180609). The Smileloc Sleeve is used to lock and unlock the Titanium Coping for final restoration to or from the abutment. This makes the prosthesis removable. The Rodo Abutment System eliminates the need for an access hole on the occlusal surface of a screw-retained restoration and also eliminates the possibility of prosthetic screw loosening. The Smilekey is an induction heating device for dental prosthesis removal of the Smileloc Sleeve in the Rodo Abutment System. The Smilekey was cleared as an accessory to the Abutment System in K180609, and there have been no changes to the Smilekey since this clearance.
The Rodo Abutment is provided in five series designs (100 F, 200 P, 300 S, 400 M, 500 D) with the 200 P and 500 D series having angled abutments (17°, 30°), for a total of nine designs. The 300 S series is designed for limited occlusal space and the 400 M series is designed for large interproximal spaces. Abutments are available in sizes ranging from 3.0 mm to 6.0 mm depending on the compatible implant system in use. Designs are available with engaging and non-engaging implant-abutment interfaces.
The provided text describes a 510(k) premarket notification for the Rodo Abutment System, focusing on adding compatibility with new implant lines. However, it does not contain the detailed information typically found in a study proving a device meets specific performance acceptance criteria for an AI/ML medical device.
The Rodo Abutment System is a hardware dental device, not an AI/ML software device. The "performance data" section discusses non-clinical testing such as dimensional analysis, reverse engineering, and fatigue testing according to ISO 14801. It does not refer to AI/ML model performance metrics, ground truth establishment, or human reader studies.
Therefore, I cannot extract the requested information regarding acceptance criteria, study details, sample sizes, expert qualifications, or multi-reader multi-case studies, as the provided document pertains to the clearance of a mechanical dental device and not an AI/ML-driven diagnostic tool.
Specifically, the document does NOT contain information on:
- A table of acceptance criteria and reported device performance for an AI/ML model (e.g., sensitivity, specificity, AUC).
- Any sample sizes used for a test set in the context of AI/ML performance.
- Data provenance for AI/ML data (country of origin, retrospective/prospective).
- Number of experts used to establish ground truth for an AI/ML model.
- Adjudication method for an AI/ML test set.
- Multi-reader multi-case (MRMC) comparative effectiveness studies.
- Standalone (algorithm-only) performance of an AI/ML model.
- Type of ground truth used (pathology, outcomes data, etc.) for an AI/ML model.
- Sample size for a training set in the context of AI/ML.
- How ground truth for a training set was established for an AI/ML model.
The document's "Performance Data" section solely refers to non-clinical bench testing for a mechanical device.
Ask a specific question about this device
(143 days)
The Spiraltech Dental Implants are endosseous implants intended to be surgically placed in the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patients esthetics and chewing function. Spiraltech implants are intended for single or multiple unit restorations on splinted or non-splinted applications. The implants ESi Dynamic and Ultimate are intended for immediate loading when good primary stability is achieved, and with appropriate occlusive loading. These implants [along with Premium and One Piece] can also be used for loading after a conventional healing period.
Solo One Piece 3.0 and 3.3 implants, Ultimate (conical) 3.0 implants, and ESi (conical) 3.0 implants are intended to replace a lateral incisor in the maxilla and/or a central or lateral incisor in the mandible. Mandibular central and lateral incisors must be splinted if using two or more 3.0 and/or 3.3 implants adjacent to one another.
The SpiralTech implant system is a comprehensive product line that includes implants, corresponding abutments, and closure screws. SpiralTech dental implants are grade 5 titanium (Ti 6Al-4V ELI, conforms to ASTM F136) implants that come in 2 different surface treatments - SLA, and RBM.
SpiralTech dental implants come in five product lines with four based on their thread designs. The ESi has sharp, square, and rounded threads. The Ultimate and Dynamic lines also have sharp threads with the Dynamic having a reverse buttress thread design. The Premium line features square and sharper threads in a more conventional design. The implants have diameters ranging from 3.0 mm, and the lengths from 8mm to 15 mm. ESi, Ultimate, and Dynamic are intended to be used for immediate loading. The fifth implant type is the One Piece which comes with an abutment which cannot be used with low mechanical stability cases.
Abutments are available in various types including straight, shoulder, angulated, ball attachments, multiunit, temporary and healing. All abutments come in both hex and conical connections. Temporary abutments come in PEEK and zirconia. Healing abutments come in titanium alloy and zirconia. Abutments come in titanium alloy and as a titanium base with zirconia abutments and multi-units are titanium alloy. No SpiralTech abutments are intended to be modified.
The SpiralTech Dental Implant System's acceptance criteria and studies are focused on demonstrating substantial equivalence to predicate devices, rather than establishing direct clinical performance metrics against a defined standard. Therefore, the device performance is reported in terms of equivalence to established devices and compliance with relevant standards.
Here's an breakdown based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
This section is derived from the "Non-Clinical Testing" and "Substantial Equivalence" sections, as well as the comparison tables. The acceptance criteria essentially reflect conformance to established standards and similarity to predicate devices.
Acceptance Criterion | Reported Device Performance |
---|---|
Sterilization Validation | Complies with ISO 11737-1, ISO 11737-2, ISO 11137-1, ISO 11137-2, ISO 11137-3 (for gamma radiation), ISO 17665-1, and ISO 17665-2 (for moist heat). Includes USP 85, USP 161, and LAL method testing. Shelf life validated for 5 years based on accelerated testing (ASTM F1980) with one year of real-time data (ASTM F1929-12, ASTM F88/F88M). Predicate/reference devices underwent the same type of sterilization testing. |
Fatigue Testing | Designs of SpiralTech Dental Implants were fatigue tested according to ISO 14801. Justification provided to show the tested design was the worst-case for that connection type. Predicate/reference devices underwent the same ISO 14801 testing. |
Surface Analysis | Implants with SLA and RBM surface treatments evaluated by SEM and EDS for surface analysis. (Implied acceptance: surfaces are comparable to predicate devices and suitable for dental implants). |
Biocompatibility | Cytotoxicity (MEM elution) provided to demonstrate biocompatibility for all subject devices composed of Ti6Al4V, ZrO2, and PEEK. (Implied acceptance: materials are biocompatible and comparable to predicate devices). |
Material Composition | Implants: Grade 5 titanium (Ti 6Al-4V ELI, conforms to ASTM F136). Abutments: Ti6Al4V or ZrO2. Temporary Abutments: PEEK. Chemical characterization of ZrO2 provided. Materials used are commonly found in predicate and reference devices. |
Dimensions (Diameters) | Implants: 3.0-6.0mm (Hex: 3.3, 3.5, 4.3, 5.0, 6.0; Conical: 3.0, 3.5, 4.3, 5.0, 6.0; Solo: 3.0, 3.3, 3.75, 4.2). The widths are in the same range as predicate and reference devices (e.g., NobelActive: 3.5-5.5mm, Tapered SwissPlus: 3.7, 4.8mm). |
Dimensions (Lengths) | Implants: 8-15mm (specific lengths vary by design and diameter). The lengths are in the same range as predicate and reference devices (e.g., NobelActive: 8.5-11.5mm, Biohorizons: 7.5-15mm). |
Thread Designs | ESi (sharp, square, rounded threads), Ultimate (sharp threads), Dynamic (sharp threads, reverse buttress), Premium (square, sharper threads). These thread designs incorporate types of threads used in predicate and reference devices (specific references to Biohorizons, NobelReplace and NobelActive for various thread designs). |
Connection Types | Hex and conical. These are types used in several predicate devices (e.g., NobelActive: hex, NobelReplace: conical). |
Surface Treatments | SLA and RBM. These are the same types as those used on predicate devices (e.g., NobelActive: acid etched, NobelReplace: HA, acid etched, Biohorizons: RBM, MIS: SLA). |
Indications for Use | Consistent with the predicate device, with more restrictive placement for smaller diameter implants (Solo One Piece 3.0/3.3, Ultimate conical 3.0, ESi conical 3.0 for lateral incisors in maxilla and/or central/lateral incisors in mandible, with splinting requirement for mandibular central/lateral incisors if two or more 3.0/3.3 implants are adjacent). The general indications are substantially equivalent to Nobel Biocare's NobelActive implants. |
2. Sample size used for the test set and the data provenance
The document describes non-clinical laboratory testing, not clinical studies involving patient data. Therefore, there isn't a "test set" in the sense of patient data.
- Sample Size for Testing: The specific sample sizes for fatigue testing, sterilization validation, SEM/EDS analysis, and biocompatibility testing are not explicitly stated in the provided text. They are implied to be sufficient for compliance with the referenced ISO and ASTM standards.
- Data Provenance: The studies are non-clinical, laboratory-based tests conducted following international standards (ISO, ASTM, USP). There is no country of origin for "data" in the context of patient information, nor is it retrospective or prospective. The data originates from the manufacturer's testing or contracted laboratories.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This device's approval is based on substantial equivalence through non-clinical testing and comparison to predicate devices, not on interpretation of patient data by a panel of human experts.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable, as there is no "test set" requiring expert adjudication.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This document pertains to a dental implant system, which is a physical medical device, not an AI-powered diagnostic or decision support system.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable, as this is not an algorithm-based device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" in this context is the established performance and safety requirements outlined in international standards (ISO, ASTM, USP) for dental implants and materials, as well as the documented characteristics of the predicate and reference devices. For example:
- Sterility: Demonstrated by meeting specified sterility assurance levels (SAL) according to ISO standards.
- Mechanical Strength: Demonstrated by meeting fatigue resistance criteria according to ISO 14801.
- Biocompatibility: Demonstrated by passing cytotoxicity tests (MEM elution) and material characterization (SEM, EDS) in accordance with relevant standards.
- Design & Materials: Comparison to the known, established designs and materials of legally marketed predicate devices.
8. The sample size for the training set
Not applicable. There is no "training set" as this is not an AI/machine learning device.
9. How the ground truth for the training set was established
Not applicable. There is no "training set."
Ask a specific question about this device
(211 days)
BioHorizons CAD/CAM Abutments are dental abutments placed onto a dental implant to provide support for dental prosthetic restorations. The abutments include: 1) Titanium abutment blanks with a pre-machined implant connection where the upper portion may be custom-milled in accordance with a patient-specific design using CAD/CAM techniques; and 2) Titanium bases with a pre-machined implant connection upon which a CAD/CAM designed superstructure may be fitted to complete a two-piece dental abutments include an abutment screw for fixation to the underlying implant. The abutments may be used for single-unit (single-tooth) or multiple-unit (bridges and bars) restorations and are compatible for use with BioHorizons Internal and Tapered Internal implant systems and Zimmer® Dental Screw-Vent® and Tapered Screw-Vent® implants with 3.5mm, 4.5mm and 5.7mm internal hex-connection mating platform diameters.
All digitally designed abutments and/or copings for use with BioHorizons CAD/CAM Abutments are intended to be sent to a BioHorizons-validated milling center for manufacture. BioHorizons abutments designed using CAD/CAM techniques must fulfill the BioHorizons allowable range of design parameters.
BioHorizons CAD/CAM Abutments are dental implant final restorative abutments supplied in platform diameters of 3.0mm, 3.5mm, 4.5mm and 5.7mm. The abutments are intended to provide support for dental prosthetic restorations. Each abutment includes an abutment screw for fixation to the underlying implant. Abutment material is titanium alloy as specified in ASTM F136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy (UNS R56401) for Surgical Implant Applications.
Select abutments are further processed by applying patterns of micro-machined grooves or channels, known as Laser-Lok, to a specified region of the abutment margin. The abutments are provided non-sterile, and they are packaged using materials known in the industry to be appropriate for medical device packaging.
Here's an analysis of the provided text regarding acceptance criteria and supporting studies for the BioHorizons CAD/CAM Abutments:
The document primarily focuses on demonstrating substantial equivalence to predicate devices for regulatory approval (510(k)). This means the "acceptance criteria" are largely based on showing the device performs comparably to existing, legally marketed devices, rather than establishing de novo performance targets. The "studies" are tests conducted to support this claim of equivalence.
Acceptance Criteria and Reported Device Performance
The document doesn't explicitly list numerical acceptance criteria in a typical pass/fail format with specific thresholds. Instead, the acceptance criteria are implicitly defined by the performance of the predicate devices and relevant industry standards. The reported device performance is demonstrated through various tests designed to show that the BioHorizons CAD/CAM Abutments perform in accordance with their intended use and are substantially equivalent to predicate devices.
Acceptance Criterion (Implicit) | Reported Device Performance |
---|---|
Mechanical Fatigue Strength (per ISO 14801 & FDA Guidance) | "Results of the testing demonstrate that the fully milled, worst-case 30° angled configurations of both the 3.0mm and 3.5mm prosthetic platform abutments and the titanium bases with bonded zirconia superstructure perform in accordance with their intended use." (Implies meeting or exceeding the performance established by the standard and predicate devices for similar angled abutments). |
Implant Compatibility | "Compatibility testing was performed on a representative subset of Zimmer® Screw-Vent® and Tapered Screw-Vent® implants... This testing verifies compatibility of BioHorizons Abutments for Zimmer® with all Zimmer® Screw-Vent® and Tapered Screw-Vent® items based on equivalent mating platform geometry." (Implies successful and reliable fit/function with listed implant systems). |
Sterilization Efficacy (for non-sterile devices requiring sterilization by user) | "Steam sterilization validation testing was performed new in accordance with AAMI/ANSVISO 17665-1:2006... Test results demonstrate a sterility assurance level (SAL) of 10⁻⁶." (Implies achievement of the industry-standard sterility level after a validated sterilization process). |
Software Validation (for CAD/CAM design parameters) | "Software intended use validation testing of the software systems utilized in the manufacture of the CAD/CAM abutments was performed to ensure that the program design limitations prevent the user from milling abutments that do not fulfill the BioHorizons design criteria." (Implies the software correctly enforces design constraints to ensure proper abutment manufacturing). |
Substantial Equivalence (Overall Safety & Effectiveness) | "The data presented in this submission demonstrates that the proposed devices are substantially equivalent with respect to performance and intended use. The proposed devices perform as well as the legally marketed predicate devices. Furthermore, the proposed devices do not pose any new or increased risks as compared to the legally marketed predicate devices." (This is the overarching conclusion of the submission, supported by all the aforementioned tests and comparisons). |
Further Details on the Study:
Given the nature of this 510(k) submission for a dental abutment, it's highly unlikely that components like "test sets," "ground truth experts," "adjudication methods," or "MRMC comparative effectiveness studies" were part of the presented data. These concepts are more typically associated with diagnostic imaging AI/ML devices or clinical trials involving complex human interpretation.
Therefore, many of the requested items below will be marked as "Not Applicable" or "Not Provided" based on the document's content and the typical scope of device approval for implant components.
-
Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size (Test Set): Not explicitly stated for each test, but implied to be sufficient for compliance with the respective standards (e.g., ISO 14801 typically requires 12 samples per group for fatigue testing). The document mentions "worst-case 30° angled configurations" and "representative subset" for compatibility testing.
- Data Provenance: Not specified, but generally assumed to be internal laboratory testing by the manufacturer or contracted labs. The document does not indicate data from human subjects or clinical sites.
- Retrospective or Prospective: Not applicable as these are laboratory and validation tests, not clinical studies involving patient data.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable. The "ground truth" for mechanical and sterilization testing is defined by the physical properties measured against engineering standards (e.g., fracture strength, SAL level). For software validation, it's the correct implementation of defined design parameters. No human expert "ground truthing" in the sense of clinical interpretation is relevant here.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set
- None. Adjudication methods are relevant for ambiguous human interpretations, which is not the case for material property or mechanical performance testing.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. This is not a diagnostic device involving human readers or AI assistance.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- Not Directly Applicable. While software validation for the CAD/CAM milling process was performed (an "algorithm only" component), this isn't a standalone diagnostic AI. Its "performance" is about correctly enforcing design parameters, not making a diagnosis or prediction. The core device is a physical product, not a software algorithm presented for standalone performance evaluation in a clinical context.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Engineering Standards / Defined Specifications:
- For Dynamic Mechanical Fatigue: ISO 14801 standards and BioHorizons' internal performance specifications for expected strength and durability.
- For Compatibility: Physical fit and function with the specified implant systems.
- For Sterilization: AAMI/ANSVISO 17665-1:2006, specifically achieving a Sterility Assurance Level (SAL) of 10⁻⁶.
- For Software: BioHorizons' allowable range of design parameters and criteria.
- Engineering Standards / Defined Specifications:
-
The sample size for the training set
- Not Applicable. As a physical medical device submission, there is no "training set" in the context of machine learning algorithms. The design process for the abutments is based on engineering principles and existing product lines, not data training.
-
How the ground truth for the training set was established
- Not Applicable. (See above, no training set for this type of device submission).