Search Filters

Search Results

Found 14 results

510(k) Data Aggregation

    K Number
    K170872
    Device Name
    Aisys CS2
    Manufacturer
    Date Cleared
    2017-08-15

    (145 days)

    Product Code
    Regulation Number
    868.5160
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aisys CS2 Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide ride range of patients (neonatal, pediatric, adult). The device is intended for volume or pressure control ventilation.

    Device Description

    The GE Datex-Ohmeda Aisys CS2 is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). It represents one of the systems in a long line of products based on the Datex-Ohmeda Excel, Aestiva, Aespire, and Avance Anesthesia Systems. It is to be used only by trained and qualified medical professionals.

    The Aisys CS2 supplies set flows of medical gases to the breathing system using electronic gas mixing. Interfaces to control the system include the touch screen, keypad and rotary controller on the main display unit. Selected gas flows are displayed as electronic flow indicators on the system display unit. The Aisys CS2 is equipped with a pneumatic back-up O2 delivery system and traditional flow tube, as well. A large selection of frames, gases, and vaporizer cassettes are available to give the user control of the system configuration. The Aisys CS4 systems are also available in pendant models. The system shall support a maximum of two-cylinder supply connections mounted inboard on the machine and supported by cylinder yokes. All models have O2. The Aisys CS2 comes with up to two optional gases (air, N20). Safety features and devices within the Aisys CS2 are designed to decrease the risk of hypoxic mixtures, agent mixtures and complete power or sudden gas supply failures. The Aisys CS2 system is available with optional integrated respiratory gas monitoring. When supplied as an option, the integrated respiratory gas monitoring is provided via the CARESCAPE Modules cleared via K123195 (E-sCAiO, EsCAiOV) and K150245 (E-sCAiOVX). The Aisys CS2 is also compatible with legacy M-Gas and E-Gas modules which are in the installed base but are no longer in forward production (M-CAiO and M-CAiOV cleared via K001814, and E-CAiOVX cleared via K051092).

    The above modules can be physically integrated into the Anesthesia device, receive electronic power from the said device and communicate measured values to the said device for display on the system display unit.

    The anesthetic agent delivery for the Aisys CS2 is controlled via an anesthesia computer through user input from the central display. The vaporization technology is based upon the electronic vaporizer cleared as part of the Datex-Ohmeda Anesthesia Delivery Unit (ADU) cleared via K973985. An Aladin 2 is inserted into the active cassette bay. The cassette holds the agent to be delivered - Isoflurane, Desflurane or Sevoflurane. Agent is delivered as a percent volume/volume. The Aisys CS2 is designed to allow only one active cassette at a time. Per the user input into the main display, valves within the active cassette bay will open and allow agent to be delivered. The agent is mixed with gas from the FGC unit. After mixing, the combination of gases and agent is delivered to the breathing system and then onto the patient.

    The Datex-Ohmeda 7900 Anesthesia Ventilator is used in the Aisys CS2. It is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during surgical procedures. The 7900 ventilator is equipped with a built-in monitoring system for inspired oxygen, airway pressure and exhaled volume. Sensors in the breathing circuit are used to control and monitor patient ventilation as well as measure inspired oxygen concentration. This allows for the compensation of compression losses, fresh gas contribution and small leakage in the breathing absorber, bellows and system. User setting and microprocessor calculations control breathing patterns. The user interface keeps settings in memory. The user may change settings with a simple and secure setting sequence. A bellows contains breathing gasses to be delivered to the patient. Positive End Expiratory Pressure (PEEP) is regulated electronically. Positive pressure is maintained in the breathing system so that any leakage that occurs is outward. An RS-232 serial digital communications port connects to and communicates with external devices.

    Ventilator modes for the device include Volume Control (VCV) Mode, Pressure Control (PCV) Mode (Optional), Synchronized Intermittent Mandatory Ventilation with Pressure Control Ventilation -Volume Guaranteed (SIMV/PCV-VG) Mode, Synchronized Intermittent Mandatory Ventilation with Pressure Support Ventilation (SIMV/PSV) Mode, Pressure Support with Apnea Backup (PSVPro) Mode (Optional), Synchronized Intermittent Mandatory Ventilation with Pressure Control (SIMV-PC) Mode (Optional), Pressure Control Ventilation- Volume Guaranteed (PCV-VG) mode (Optional), and Continuous Positive Airway Pressure/ Pressure Support Ventilation (CPAP-PSV).

    Ventilator parameters and measurements are displayed on the system display unit.

    The system display unit is mounted to an arm on the top shelf of the Aisys CS2. The arm is counter balanced and capable of moving vertically and/or horizontally, and tilting the display, enabling the user to position the display to the most advantageous viewing position. The arm length is limited such that the display position is always within the footprint of the Aisys CS2 frame. The arm also supports the mounting of additional display units for a variety of patient monitors.

    Several frame configurations are available, including one that allows for the physical integration of the GE Monitors (cleared Carescape B850 via K092027 and B650 cleared on K102239). This configuration also provides cable management solutions such that the necessary connections from the monitor display unit to the monitor are hidden within the Aisys CS2 frame. An additional option allows the monitor to be linked to the power supply of the Aisys CS2 such that when the Aisys CS2 is turned on, the monitor is also turned on. Additional configurations allow for the mounting of various patient monitors on the top shelf of the Aisys CS2.

    AI/ML Overview

    This document is a 510(k) premarket notification for the Aisys CS2 anesthesia system (version 11). It primarily focuses on demonstrating substantial equivalence to a predicate device (Aisys CS2 version 10) through non-clinical testing. Therefore, it does not contain the detailed information about acceptance criteria and a study that proves the device meets those criteria in the way typically found for a diagnostic or AI-driven device.

    Based on the provided text, here's a breakdown of what is and is not available:

    1. A table of acceptance criteria and the reported device performance

    No explicit table of acceptance criteria and reported device performance in terms of diagnostic metrics (e.g., sensitivity, specificity, AUC) is provided. This is because the submission is for an anesthesia system, and the "performance" is related to its functional capabilities and compliance with safety standards, rather than a diagnostic accuracy.

    The document lists various verification tests and what they verify. For example:

    • Privacy and Security: Verifies functionality including an option to disable viewing patient identifiable information.
    • Duplicate Interface Detection: Verifies functionality including that the system continues communication with its clients even if a duplicate IP condition is detected.
    • Ethernet Interface: Verifies functionality including that the system supports 100Mbps speed and full duplex settings.
    • Network Hazard Mitigation: Verifies that the system has no open ports except for specific clients.
    • Network Requirements: Verifies that the system supports clock synchronization with a network device.
    • Sapphire and HL7: Verifies communication protocols.
    • Address Resolution Protocol Requirements: Verifies correct system subnet mask functionality.
    • Respiratory Gas Monitors: Verifies all requirements related to Respiratory Gas Monitors, including functionality of the Sample Gas Return option.
    • Monitoring Only Mode: Verifies functionality including O2 being administered through the auxiliary O2 port when the mode is enabled.
    • System Hazard Mitigations: Verifies functionality including that the system performs as intended during a recovery state.
    • Materials Testing: Includes Volatile Organic Compounds, Particulate Matter Testing, Bacterial Filter Efficiency Testing, Viral Filter Efficiency Testing.
    • Reprocessing Instructions Validation Testing
    • Verification testing for electrical safety and electromagnetic compatibility: Compliance to AAMI / ANSI ES60601-1:2005/(R)2012, IEC 60601-1-2: 2014, and ISO 80601-2-13: 2011.

    For each of these, the implicit "acceptance criterion" is that the system successfully performs or complies with the stated function or standard. The "reported device performance" is the conclusion that these tests were passed and the system is safe and effective.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not applicable / not provided for this type of submission. This device is an anesthesia machine, not typically tested with "test sets" of patient data in the same way an AI diagnostic device would be. The testing described is functional and safety verification.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable / not provided. There is no "ground truth" establishment in the context of diagnostic accuracy for this device. Ground truth is usually relevant for AI/ML models being validated against expert-labeled data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable / not provided. Adjudication methods are used to resolve discrepancies in expert labeling for ground truth, which is not relevant here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable / not provided. This is an anesthesia machine, not an AI-assisted diagnostic tool for human "readers" (e.g., radiologists interpreting images). An MRMC study would not be relevant for its evaluation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable / not provided. This is an anesthesia machine. While it contains software, it is not a standalone AI algorithm in the diagnostic sense. It is a system intended for use by trained medical professionals. The software updates mentioned are for standards compliance and usability enhancements, not for independent diagnostic decisions.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This information is not applicable / not provided. As explained, there is no "ground truth" akin to diagnostic accuracy for this device. The "ground truth" for its performance is its ability to meet engineering specifications, safety standards, and functional requirements.

    8. The sample size for the training set

    This information is not applicable / not provided. This device is not an AI/ML model that would have a "training set" of data in the typical sense. The software updates are developed and verified through standard software engineering practices.

    9. How the ground truth for the training set was established

    This information is not applicable / not provided. As there is no training set for an AI/ML model, the concept of establishing ground truth for it does not apply.


    In summary, the provided document is a 510(k) summary for an anesthesia system. The "study" proving it meets acceptance criteria consists of a comprehensive set of non-clinical verification and validation tests covering:

    • Software validation (including enhancements like EcoFlow improvements, Monitory Only mode, Network Connectivity, Privacy and Security, etc.)
    • Compliance with electrical safety, EMC, and specific medical device standards (e.g., IEC 60601-1, ISO 80601-2-13).
    • Materials testing.
    • Risk analysis and design reviews.

    The acceptance criteria are implicitly that the system functions as intended, meets its specifications, and complies with all relevant safety and performance standards. The "study" is the extensive report of these non-clinical tests, and the conclusion is that the modified device is substantially equivalent to its predicate. Clinical testing was not deemed necessary for the changes introduced in version 11.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140575
    Manufacturer
    Date Cleared
    2015-02-03

    (334 days)

    Product Code
    Regulation Number
    868.5895
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K050597, K001814, K051092, K023454, K052582

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GE Datex-Ohmeda Engström Carestation and Engström Pro are designed to provide mechanical ventilation for adults and pediatrics weighing 5kg and above having degrees of pulmonary impairment varying from minor to severe. Optional Neonatal capabilities on Engström Carestation and Engström Pro expand the patient range to 0.25 kg.

    The GE Datex-Ohmeda Engström Carestation and Engström Pro are microprocessor based, electronically controlled, pneumatically driven ventilators that include integrated FiO2, airway pressure, spirometry, and volume monitoring. Options include an Aerogen Aeroneb nebulizer, data capture accessory, and an integrated air compressor. Options available on Engström Carestation only include integrated respiratory gas monitoring capabilities via various Datex-Ohmeda patient monitoring modules listed in the product labeling, capabilities to measure SpiroDynamics via an intratracheal pressure sensor in patients using sized 6.5 tracheal tubes and larger, and calculation of functional residual capacity of mechanically ventilated patients using Nitrogen Wash In/Wash Out method.

    Not all features are available with all patient populations.

    The Engström Carestation is not a pulmonary function calculation device.

    The system is designed for facility use, including within-facility transport, and should only be used under the orders of a clinician.

    Device Description

    The GE Datex-Ohmeda Engström Carestation and Engström Pro are flexible, and intuitive, critical care ventilators. A wide selection of performance options gives the user full control of the system configuration. The Engström Carestation and Engström Pro feature patient monitoring, patient ventilation, and the capability of interfacing with central information management systems.

    Both the GE Datex-Ohmeda Engström Carestation and Engström Pro are designed to provide mechanical ventilation for adults and pediatrics weighing 5kg and above having degrees of pulmonary impairment varying from minor to severe. Optional Neonatal capabilities allow the Engström Carestation and Engström Pro to be used with patients weighing 0.25 kg and above.

    The modes of ventilation currently available include:

    1. Volume Controlled (VCV)
    2. Pressure Controlled (PCV)
    3. Pressure Controlled, Volume Guaranteed (PCV-VG)
    4. Synchronized Intermittent Mandatory Ventilation, Volume Controlled (SIMV-VC)
    5. Synchronized Intermittent Mandatory Ventilation, Pressure Controlled (SIMV-PC)
    6. Synchronized Intermittent Mandatory Ventilation, Pressure Controlled Volume Guarantee (SIMV-PCVG)
    7. Bi-level Airway Pressure Ventilation
    8. Constant Positive Airway Pressure Support Ventilation (CPAP/PSV)
    9. Apnea backup (available in SIMV-VC, SIMV-PC, SIMV-PCVG/BiLevel-VG, BiLevel, CPAP/PSV, and VG-PS)
    10. Non-invasive ventilation (NIV), not available in neonatal mode
    11. Infant Nasal CPAP (nCPAP), only available in neonatal mode
    12. Volume Guarantee. Pressure Support (VG-PS), only available in neonatal mode

    The GE Datex-Ohmeda Engström Carestation and Engström Pro are microprocessor based, electronically controlled, pneumatically driven ventilators that include integrated FiO2, airway pressure, spirometry and volume monitoring and an Aerogen Aeroneb Pro nebulizer control board.

    The ventilator consists of two main components: a display and a ventilator unit. The display allows the user to interface with the system and control settings through use of soft keys on the display, a com wheel, and a resistive touch screen. The Engström Carestation also includes a module bay that allows the integration of various Datex-Ohmeda patient monitoring modules with the ventilator.

    The user interface for control of nebulization is provided via the ventilator display unit. The standard nebulizer board is provided with both the Engström Carestation and Engström Pro variants. Users have the option to configure the system to use an external pneumatic nebulizer in place of the standard nebulizer.

    The optional medical air compressor is intended for use as an accessory to provide a dry, filtered, breathable compressed air supply. The compressor is installed in the base of the ventilator cart. The compressor is powered from AC mains only. A source of compressed oxygen is required to be connected to Engström Carestation/Engström Pro equipped with the optional compressor. The use of an integrated air compressor was first cleared in K050597.

    Optional accessories common to both Engström Carestation and Engström Pro include a trolley/cart, integrated air compressor, support arm, humidifier and water trap mounting brackets, and a data capture accessory. Additional optional accessories specific to the Engström Carestation include airway modules, intratracheal pressure sensor, auxiliary electrical outlets, and module bay. Optional functionality specific to the Engström Carestation includes integrated respiratory gas monitoring, capabilities to measure SpiroDynamics via a GE supplied intratracheal pressure sensor in patients using sized 6.5 tracheal tubes and larger, and calculation of functional residual capacity of mechanically ventilated patients using Nitrogen Wash In/Wash Out method. The integrated respiratory gas monitoring is provided via the Datex-Ohmeda Gas Modules, M-C. M-COV. M-COVX, M-CaiO, M-CAiOV, M-CAiOVX, rev 3.2 software and higher (K001814), E-CO, E-COV, E-COVX, E-CAiO, E-CAiOV, E-CAiOVX (K051092), or M-Mini-CO2 Module (K023454) or E-MiniC module (K052582) which are physically integrated into the Engström Carestation, receive electronic power from the Engström Carestation and communicate measured values to the Engström Carestation for display on the system display unit.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the GE Datex-Ohmeda Engström Carestation and Engström Pro ventilators.

    This 510(k) submission (K140575) is primarily for the addition of an alternate integrated air compressor (EVair) to the existing GE Datex-Ohmeda Engström Carestation and Engström Pro ventilators, which were previously cleared under K111116. The submission states "There have been no changes to the intended use or fundamental scientific technology." and "The addition of the alternate compressor, EVair, does not affect the safe or effective use of the ventilator as the compressor interfaces with the ventilator in the same manner, and performs in an equivalent manner. There is no change to the Engstrom performance as a result of this change." Consequently, the acceptance criteria and studies focus specifically on the new compressor and its functional equivalence, rather than a broad re-evaluation of the entire ventilator system.


    1. Table of Acceptance Criteria and Reported Device Performance

    Given the nature of this 510(k) (adding an alternate component), the acceptance criteria are framed around functional equivalence and safety of the new EVair compressor compared to the predicate EVair03.

    Acceptance Criteria CategorySpecific Criteria (Expected Performance)Reported Device Performance (Engström Carestation and Engström Pro with EVair)
    Intended UseIdentical to the predicate device (Engstrom 7.x - K111116): Mechanical ventilation for adults and pediatrics (5kg+), optional neonatal (0.25kg+). Microprocessor-based, electronically controlled, pneumatically driven. Integrated FiO2, airway pressure, spirometry, volume monitoring, Aerogen Aeroneb nebulizer, data capture accessory, integrated air compressor. Integrated respiratory gas monitoring (Carestation only), SpiroDynamics, FRC calculation (Carestation only). For facility use.Confirmed to be identical to the predicate Engstrom 7.x (K111116) Indications for Use, with only a textual change in reference to the product name. No change to the intended use.
    Ventilation ModesIdentical to the predicate device (Engstrom 7.x - K111116) list of 12 ventilation modes, including VCV, PCV, PCV-VG, SIMV-VC, SIMV-PC, SIMV-PCVG, Bi-level, CPAP/PSV, Apnea backup, NIV, Infant Nasal CPAP, and Volume Guarantee/PS.Confirmed to be identical to the predicate Engstrom 7.x (K111116). No change to the ventilation modes.
    Ventilator System SoftwareMinor software updates only, with no change to the features or function of the ventilator. (Version 7.X)Version 7.X. Minor software updates only have been made with no change to the features or function of the ventilator. Considered substantially equivalent.
    Optional Integrated Air Compressor FunctionalityFunctionally equivalent to the predicate (EVair03) such that it does not affect the safe or effective use of the ventilator.Verification has demonstrated that the EVair compressor is functionally equivalent to the predicate EVair03 compressor. The compressor interfaces with the ventilator in the same manner. This results in no change to the Engstrom performance.
    Air Compressor: Transition to Backup Air SupplyPneumatic control to switch from standby to reserve air supply. Backup operation begins when pipeline pressure drops below 250 kPa (36.3 psi) [for predicate EVair03]. Pressure values should fall within the Engstrom ventilator specifications.Software control to switch from standby to reserve air supply. Backup operation begins when pipeline pressure drops below 280 kPa (40.6 psi), which is a factory set-point but can be adjusted by trained service personnel. The backup operation of the EVair is substantially equivalent to the predicate EVair03. Pressure values still fall within the Engstrom ventilator specifications. No functional change to the Engstrom Ventilator; the EVair has been verified to be functionally equivalent to the EVair03 compressor.
    Safety and Performance (New Compressor)Compliance with applicable standards (e.g., electrical safety, EMC, performance standards for ventilators), risk analysis, software validation. The compressor should exhibit improvements in acoustics, vibration, and thermal dissipation compared to the predicate.Thoroughly tested through verifications and validation, including software validation. Verification of compliance with applicable standards has been completed. Risk Analysis, Requirements/Specification Reviews, Design Reviews, Testing on unit level, Integration testing, Performance Testing (Verification), and Safety Testing (Verification) were applied. The EVair compressor contains improvements in acoustics, vibration, and thermal dissipation.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document describes nonclinical testing only.

    • Sample Size: The document does not specify a quantitative "sample size" in terms of units tested for the compressor, nor does it refer to patient data. Testing appears to be on developmental units of the EVair compressor integrated with the Engström ventilators.
    • Data Provenance: The testing was "nonclinical," meaning it did not involve human patients. It was conducted by the manufacturer, Datex-Ohmeda, Inc. (GE Healthcare). There is no mention of country of origin for specific test data, but the company is based in Madison, WI, USA. The testing is prospective in the sense that it evaluates a newly designed component (EVair compressor) before its market release.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • The concept of "ground truth established by experts" is not directly applicable here as this is a nonclinical engineering and functional equivalence study for a medical device component.
    • The ground truth for this type of evaluation is typically defined by engineering specifications, regulatory standards (e.g., ISO, IEC for medical electrical equipment), and the performance characteristics of the legally marketed predicate device.
    • The "experts" involved would be the manufacturer's engineers, quality assurance personnel, and regulatory specialists who designed, tested, and evaluated the device against these objective criteria. Their qualifications would stem from their professional expertise in medical device development, testing, and regulatory requirements.

    4. Adjudication Method for the Test Set

    • Adjudication methods like "2+1" or "3+1" are typically used in clinical studies involving human observers (e.g., radiologists interpreting images) where consensus or a tie-breaking mechanism is needed to establish a "true" label.
    • This submission describes nonclinical engineering testing. Therefore, such adjudication methods are not applicable. The results of the tests (e.g., pressure measurements, acoustic levels, vibration, thermal performance, software function) are objective, quantitative data compared against predefined specifications and predicate performance.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No, an MRMC comparative effectiveness study was not done.
    • MRMC studies are clinical studies designed to evaluate the diagnostic performance of human readers, often comparing performance with and without an AI-assisted device.
    • This 510(k) submission is for an alternate component (air compressor) of a ventilator, not a diagnostic AI device. The submission explicitly states: "The modifications made to the Engström ventilator did not require clinical testing."

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This question is framed for AI/algorithm-based devices. The Engström Carestation and Engström Pro are physical medical devices (ventilators) with microprocessor control, not standalone AI algorithms.
    • The "standalone performance" of the EVair compressor itself was evaluated during its design and verification to meet engineering specifications (acoustics, vibration, thermal dissipation) and then its integrated performance with the ventilator system was verified. This can be considered the equivalent of a "standalone" functional test for the component.

    7. The Type of Ground Truth Used

    • For this nonclinical submission, the ground truth is based on:
      • Engineering Specifications: Detailed technical requirements for component performance (e.g., pressure ranges, flow rates, alarm thresholds, acoustic limits, vibration levels, thermal profiles).
      • Regulatory Standards: Compliance with relevant national and international standards for medical devices, particularly continuous ventilators (e.g., 21 CFR 868.5895, ISO, IEC standards).
      • Predicate Device Performance: The demonstrated safe and effective performance of the previously cleared predicate device (Engstrom Ventilator K111116 and its EVair03 compressor). The new EVair compressor was evaluated against the functional performance of the EVair03.

    8. The Sample Size for the Training Set

    • This question is relevant for machine learning or AI models.
    • Not applicable. This submission is for a physical medical device and its component, not an AI model requiring a training set.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable, as there is no training set for an AI model in this submission.
    Ask a Question

    Ask a specific question about this device

    K Number
    K132530
    Manufacturer
    Date Cleared
    2013-12-20

    (130 days)

    Product Code
    Regulation Number
    868.5160
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GE Datex-Ohmeda Aisys CS2 Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). The device is intended for volume or pressure control ventilation.

    Device Description

    The GE Datex-Ohmeda Aisys CS2 is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). It represents one of the systems in a long line of products based on the Datex-Ohmeda Excel, Aestiva, Aespire, Aisys, and Avance Anesthesia Systems. It is to be used only by trained and qualified medical professionals. The Aisys CS2 supplies set flows of medical gases to the breathing system using electronic gas mixing. Interfaces to control the system include the touch screen, keypad and rotary controller on the main display unit. Selected gas flows are displayed as electronic flow indicators on the system display unit. The Aisys is equipped with a pneumatic back-up O2 delivery system and traditional flow tube, as well. A large selection of frames, gases, and vaporizer cassettes are available to give the user control of the system configuration. The Aisys CS systems are also available in pendant models. The system shall support a maximum of two-cylinder supply connections mounted inboard on the machine and supported by cylinder yokes. All models have O2. The Aisys CS comes with up to two optional gases (air, N2O). Safety features and devices within the Aisys are designed to decrease the risk of hypoxic mixtures and complete power or sudden gas supply failures. The Aisys CS system is available with optional integrated respiratory gas monitoring. When supplied as an option, the integrated respiratory gas monitoring is provided via the Datex-Ohmeda M-Gas Module (M-CAiO and M-CAiOV software revision 3.2 and above cleared via K001814) and E-Gas Module (E-CAiOVX software revision 4.5 and above cleared via K051092). CARESCAPE Modules are also available for Aisys CS' (EsCAiO, E-sCAiOV cleared via K123195). The above modules can be physically integrated into the Anesthesia device, receive electronic power from the said device and communicate measured values to the said device for display on the system display unit. The anesthetic agent delivery for the Aisys CS2 is controlled via an anesthesia computer through user input from the central display. The vaporization technology is based upon the electronic vaporizer cleared as part of the Datex-Ohmeda Anesthesia Delivery Unit (ADU) cleared via K973985. An Aladin cassette (also cleared as part of K973895) or Aladin, is inserted into the active cassette bay. The cassette holds the agent to be delivered - Halothane. Enflurane. Isoflurane. Desflurane or Sevoflurane. Agent is delivered as a percent volume/volume. The Aisys is designed to allow only one active cassette at a time. Per the user input into the main display, valves within the active cassette bay will open and allow agent to be delivered. The agent is mixed with gas from the FGC unit. After mixing, the combination of gases and agent is delivered to the breathing system and then onto the patient. The Datex-Ohmeda 7900 Anesthesia Ventilator is used in the Aisys Anesthesia System. It is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during surgical procedures. The 7900 ventilator is equipped with a built-in monitoring system for inspired oxygen, airway pressure and exhaled volume. Sensors in the breathing circuit are used to control and monitor patient ventilation as well as measure inspired oxygen concentration. This allows for the compensation of compression losses, fresh gas contribution and small leakage in the breathing absorber, bellows and system. User setting and microprocessor calculations control breathing patterns. The user interface keeps settings in memory. The user may change settings with a simple and secure setting sequence. A bellows contains breathing gasses to be delivered to the patient. Positive End Expiratory Pressure (PEEP) is regulated electronically. Positive pressure is maintained in the breathing system so that any leakage that occurs is outward. An RS-232 serial digital communications port connects to and communicates with external devices. Ventilator modes for the device include Volume Mode, Pressure Control Mode, Pressure Support with Apnea Backup Mode (Optional) and Synchronized Intermittent Mandatory Ventilation (SIMV) Mode (Optional) and Continuous Positive Airway Pressure / Pressure Support Ventilation (CPAP/PSV) Mode (Optional). Aisys CS2 also supports optional Pressure Control Ventilation - Volume Guarantee mode with spontaneous breath pressure support (SIMV PCV-VG) mode (Optional). Ventilator parameters and measurements are displayed on the system display unit. The system display unit is mounted to an arm on the top shelf of the Aisys CS. The arm is counter balanced and capable of moving vertically and/or horizontally, and also tilting the display, enabling the user to position the display to the most advantageous viewing position. The arm length is limited such that the display position is always within the footprint of the Aisys CS2 frame. The arm also supports the mounting of additional display units for a variety of patient monitors. Several frame configurations are available, including one that allows for the physical integration of the GE Monitors (cleared Carescape B850 via K092027 and B650 cleared on K102239). This configuration also provides cable management solutions such that the necessary connections from the monitor display unit to the monitor are hidden within the Aisys CS frame. An additional option allows the monitor to be linked to the power supply of the Aisys CS2 such that when the Aisys CS2 is turned on, the monitor is also turned on. Additional configurations allow for the mounting of various patient monitors on the top shelf of the Aisys CS.

    AI/ML Overview

    The provided text is for a 510(k) Premarket Notification for the GE Datex-Ohmeda Aisys CS2 Anesthesia System. This document describes a new version of an already marketed device and focuses on demonstrating substantial equivalence to the predicate device (GE Datex-Ohmeda Aisys, K110213).

    Therefore, the document does not contain the kind of acceptance criteria, study details (like sample sizes, expert qualifications, adjudication methods), or performance metrics associated with a de novo device or a groundbreaking algorithmic performance study. Instead, the focus is on verification and validation of changes relative to a predicate device.

    Here's an analysis based on the information available in the provided text, and explicit statements about what is not available due to the nature of this submission:


    Acceptance Criteria and Device Performance

    The document does not specify quantitative acceptance criteria or numerical performance metrics in the way one would for an AI/algorithm-driven diagnostic device. Instead, the acceptance is based on demonstrating that the updated device continues to meet its specifications and performs as safely and effectively as the predicate device.

    The study that "proves" the device meets acceptance criteria is a series of non-clinical tests.

    1. Table of acceptance criteria and the reported device performance:

    Acceptance Criteria (Implied)Reported Device Performance (Summary from Non-Clinical Testing)
    Compliance with specificationsThoroughly tested through verification of specifications and validation, including software validation.
    Compliance with voluntary standardsVerification of compliance with applicable voluntary standards.
    Safe use in intended environmentApplied quality assurance measures (Risk Analysis, Requirements, Design Reviews, Unit/Integration/Performance/Safety/Simulated Use Testing).
    Substantial Equivalence to PredicateGE Healthcare considers the GE Datex-Ohmeda Aisys CS2 to be as safe, as effective, and to have performance substantially equivalent to the predicate device.

    Study Details

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not applicable/not specified in the context of this 510(k) submission. There isn't a "test set" in the sense of a medical image or patient data set used for algorithmic evaluation. The testing involved various engineering and software validation tests.
    • Data Provenance: Not applicable. The testing was non-clinical (e.g., in-house verification and validation).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of Experts: Not applicable. Ground truth, in the context of an anesthesia machine, would refer to its functional correctness, safety, and performance according to engineering specifications and regulatory standards, not expert interpretation of medical data.
    • Qualifications of Experts: Not applicable. The "ground truth" was established by engineering specifications, regulatory standards, and internal quality assurance processes.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Adjudication Method: Not applicable. This concept is typically relevant for studies involving human interpretation or clinical endpoints, not for the engineering verification and validation of an anesthesia system.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC Study: No, an MRMC study was not done. This device is an anesthesia system, not an AI-assisted diagnostic tool for human readers.

    6. If a standalone (i.e., algorithm only without human-in-the loop performance) was done:

    • Standalone Performance: Not applicable in the context of an anesthesia system as a "standalone algorithm" performance. The device's performance was evaluated through non-clinical verification and validation testing of its hardware and software functions (e.g., gas mixing, ventilation modes, display accuracy, safety features).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Type of Ground Truth: The ground truth for this device's evaluation was based on engineering specifications, regulatory standards compliance, and internal quality assurance requirements. This includes:
      • Functional requirements (e.g., gas flow rates, ventilation parameters).
      • Safety requirements (e.g., hypoxic mixture prevention, alarm accuracy).
      • Performance requirements (e.g., display accuracy, response times).
      • Compliance with voluntary standards.

    8. The sample size for the training set:

    • Training Set Sample Size: Not applicable. This device is an anesthesia system developed through traditional engineering and software development processes, not an AI/machine learning model that requires a "training set" of data.

    9. How the ground truth for the training set was established:

    • Ground Truth for Training Set: Not applicable. There is no "training set" in the context of this device's development.

    Ask a Question

    Ask a specific question about this device

    K Number
    K131945
    Manufacturer
    Date Cleared
    2013-08-08

    (42 days)

    Product Code
    Regulation Number
    868.5160
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K001814, K051092, K123195, K092027, K102239

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GE Datex-Ohmeda Avance CS2 Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). The device is intended for volume or pressure control ventilation.

    Device Description

    The GE Datex-Ohmeda Avance CS2 anesthesia machines are intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). They represent one of the systems in a long line of products based on the Datex-Ohmeda Excel. Aestiva, and Aespire Anesthesia Systems. Avance systems are to be used only by trained and qualified medical professionals trained in the administration of general anesthesia.

    The Avance and Avance CS2 anesthesia systems supply set flows of medical gases to the breathing system using electronic gas mixing. Gas flows are selected by the user using the electronic controls on the main display unit and then displayed as electronic flow meters on the system display unit. The Avance systems are equipped with a pneumatic back-up O2 delivery system and traditional flow tube, as well. A large selection of frames options including central brake or individual caster brakes, gases, and vaporizers are available to give the user control of the system configuration. The Avance systems are also available in pendant models. Avance systems are available with two or three gases, up to three vaporizer positions and up to three cvlinder connections. All models have O2. The Avance systems come with up to two optional gases (air, N2O).

    The Avance systems accept Tec 6 Plus and Tec 7 vaporizers on a Selectatec manifold. Safety features and devices within the Avance systems are designed to decrease the risk of hypoxic mixtures, agent mixtures and complete power or sudden gas supply failures. The Avance systems are available with optional integrated respiratory gas monitoring. When supplied as an option, the integrated respiratory gas monitoring is provided via the GE Compact Gas Airway Modules Series: M-Gas Module (M-CAiO and M-CAiOV software revision 3.2 and above cleared via K001814) and E-Gas Compact Gas Airway Module (E-CAiOVX software revision 3.2 and above cleared via K051092) which can be physically integrated into the Avance, receive electronic power from the Avance and communicate measured values to the Avance for display on the system display unit. In addition to M-Gas and E-gas modules, the Avance CS- can utilize the GE CARESCAPE Respiratory Module (E-sCAiO, E-sCAiOV cleared via K123195).

    The Datex-Ohmeda 7900 Anesthesia Ventilator is used in the Avance Anesthesia Systems. It is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during surgical procedures. The 7900 ventilator is equipped with a built-in monitoring system for inspired oxygen, airway pressure and exhaled volume. Sensors in the breathing circuit are used to control and monitor patient ventilation as well as measure inspired oxygen concentration. This allows for the compensation of compression losses, fresh gas contribution and small leakage in the breathing absorber. bellows and system. User setting and microprocessor calculations control breathing patterns. The user interface keeps settings in memory. The user may change settings with a simple setting sequence. A bellows contains breathing gasses to be delivered to the patient. Positive End Expiratory Pressure (PEEP) is regulated electronically. Positive pressure is maintained in the breathing system so that any leakage that occurs is outward. An RS-232 serial digital communications port connects to and communicates with external devices. Ventilator modes for the device include Volume Control Ventilation (VCV), Pressure Control Ventilation (PCV) (optional), Synchronized Intermittent Mandatory Ventilation/Pressure Support (SIMV/PSV) (optional), Pressure Support Ventilation (PSVPro) (optional), Synchronized Intermittent Mandatory Ventilation-Pressure Control (SIMV-PC) (optional), Pressure Control Ventilation-Volume Guaranteed (PCV-VG) (optional), Constant Positive Airway Pressure/Pressure Support Ventilation (CPAP/PSV), Pressure Control Ventilation-Volume Guaranteed with Pressure Support Ventilation (SIMV PCV-VG) (optional), and Volume Control Ventilation Mode for Cardiac Bypass Mode (optional). Ventilator parameters and measurements are displayed on the system display unit.

    Several frame configurations are available, including one that allows for the physical integration of the GE Monitor Series (cleared Carescape B850 via K092027 and B650 cleared on K102239). These configurations also provide cable management solutions such that the necessary connections from the monitor display unit to the monitor are hidden within the Avance frame. Additional configurations allow for the mounting of various patient monitors on the top shelf of the Avance.

    AI/ML Overview

    The provided text is a 510(k) summary for the GE Datex-Ohmeda Avance CS2 Anesthesia System. It details the device's description, intended use, and a comparison to a predicate device. However, it explicitly states that no clinical testing was required or conducted for the modifications made to this version of the device.

    Therefore, the document does not contain the information requested in your prompt regarding acceptance criteria and the study that proves the device meets them, as these are typically derived from clinical or comprehensive performance studies which were not performed in this case.

    Specifically, the document states:

    • "The modifications made to the GE Datex-Ohmeda Avance CS2 did not require clinical testing. The functionality of the modified features was completely evaluated by performing nonclinical tests of design verification and validation testing."

    Without a clinical study or a study specifically designed to establish performance against acceptance criteria, I cannot fill out the requested table or provide details on sample sizes, expert involvement, ground truth establishment, or comparative effectiveness.

    The only "testing" mentioned is nonclinical verification and validation testing, which includes:

    • Risk Analysis
    • Requirements Reviews
    • Design Reviews
    • Testing on unit level (Module verification)
    • Integration testing (System verification)
    • Performance testing (Verification)
    • Safety testing (Verification)
    • Simulated use testing (Validation)

    This nonclinical testing served to confirm that the changes made (primarily updated labeling to include the use of an optional CARESCAPE respiratory module) did not alter the fundamental scientific technology or indications for use, and that the device remained substantially equivalent to its predicate.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123125
    Manufacturer
    Date Cleared
    2013-02-06

    (125 days)

    Product Code
    Regulation Number
    868.5160
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K001814, K051092, K092027, K102239

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GE Datex-Ohmeda Avance CS2 Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). The device is intended for volume or pressure control ventilation.

    Device Description

    The GE Datex-Ohmeda Avance and Avance CS anesthesia machines are intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). They represent one of the systems in a long line of products based on the Datex-Ohmeda Excel, Aestiva, and Aespire Anesthesia Systems. Avance systems are to be used only by trained and qualified medical professionals trained in the administration of general anesthesia.

    The Avance and Avance CS2 anesthesia systems supply set flows of medical gases to the breathing system using electronic gas mixing. Gas flows are selected by the user using the electronic controls on the main display unit and then displayed as electronic flow meters on the system display unit. The Avance systems are equipped with a pneumatic back-up O2 delivery system and traditional flow tube, as well. A large selection of frames options including central brake or individual caster brakes, gases, and vaporizers are available to give the user control of the system configuration. The Avance systems are also available in pendant models. Avance systems are available with two or three gases, up to three vaporizer positions and up to three cylinder connections. All models have O2. The Avance systems come with up to two optional gases (air, N2O).

    The Avance systems accept Tec 6 Plus and Tec 7 vaporizers on a Selectatec manifold. Safety features and devices within the Avance systems are designed to decrease the risk of hypoxic mixtures, agent mixtures and complete power or sudden gas supply failures. The Avance systems are available with optional integrated respiratory gas monitoring. When supplied as an option, the integrated respiratory gas monitoring is provided via the GE Compact Gas Airway Modules Series: M-Gas Module (M-CAiO and M-CAiOV software revision 3.2 and above cleared via K001814) and E-Gas Compact Gas Airway Module (E-CAiOVX software revision 3.2 and above cleared via K051092) which can be physically integrated into the Avance, receive electronic power from the Avance and communicate measured values to the Avance for display on the system display unit.

    The Datex-Ohmeda 7900 Anesthesia Ventilator is used in the Avance Anesthesia Systems. It is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during surgical procedures. The 7900 ventilator is equipped with a built-in monitoring system for inspired oxygen, airway pressure and exhaled volume. Sensors in the breathing circuit are used to control and monitor patient ventilation as well as measure inspired oxygen concentration. This allows for the compensation of compression losses, fresh gas contribution and small leakage in the breathing absorber, bellows and system. User setting and microprocessor calculations control breathing patterns. The user interface keeps settings in memory. The user may change settings with a simple setting sequence. A bellows contains breathing gasses to be delivered to the patient. Positive End Expiratory Pressure (PEEP) is regulated electronically. Positive pressure is maintained in the breathing system so that any leakage that occurs is outward. An RS-232 serial digital communications port connects to and communicates with external devices. Ventilator modes for the device include Volume Control Ventilation (VCV), Pressure Control Ventilation (PCV) (optional), Synchronized Intermittent Mandatory Ventilation/Pressure Support (SIMV/PSV) (optional), Pressure Support Ventilation (PSVPro) (optional), Synchronized Intermittent Mandatory Ventilation-Pressure Control (SIMV-PC) (optional), Pressure Control Ventilation-Volume Guaranteed (PCV-VG) (optional), Constant Positive Airway Pressure/Pressure Support Ventilation (CPAP/PSV), Pressure Control Ventilation-Volume Guaranteed with Pressure Support Ventilation (SIMV PCV-VG) (optional), and Volume Control Ventilation Mode for Cardiac Bypass Mode (optional). Ventilator parameters and measurements are displayed on the system display unit.

    Several frame configurations are available, including one that allows for the physical integration of the GE Monitor Series(most recently cleared Carescape B850 via K092027 and B650 cleared on K102239). These configurations also provide cable management solutions such that the necessary connections from the monitor display unit to the monitor are hidden within the Avance frame. Additional configurations allow for the mounting of various patient monitors on the top shelf of the Avance.

    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and supporting studies for the GE Datex-Ohmeda Avance CS2 Anesthesia System:

    Based on the provided document, the application is for a 510(k) premarket notification, which focuses on demonstrating substantial equivalence to a predicate device rather than undergoing new clinical trials for efficacy. Therefore, much of the information typically requested for acceptance criteria and a deep dive into study design for novel devices is not present or not applicable in this context.

    The document explicitly states: "The modifications made to the GE Datex-Ohmeda Avance to create the GE Datex-Ohmeda Avance CS2 did not require clinical testing. The functionality of the modified features was completely evaluated by performing nonclinical tests of design verification and validation testing."

    This immediately tells us that there was no standalone clinical study with human patients to prove device performance against specific acceptance criteria in the way one might expect for a new, high-risk device.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and the Reported Device Performance

    Given the nature of a 510(k) for an updated device, the "acceptance criteria" are primarily related to meeting specifications, voluntary standards, and demonstrating that the new features do not negatively impact safety and effectiveness compared to the predicate device. There isn't a table of specific clinical performance metrics (e.g., accuracy of a diagnostic feature, rates of successful treatment) with corresponding target values and achieved results reported in this summary.

    Acceptance Criteria (Implied from 510(k) process and text):

    Acceptance Criteria CategoryDescription (Implied)Reported Device Performance
    Substantial EquivalenceDemonstrated that the Avance CS2 is as safe, as effective, and has performance substantially equivalent to the predicate device (GE Datex-Ohmeda Avance, K112722).Met: The FDA's 510(k) clearance (K123125) confirms substantial equivalence.
    Specifications ComplianceDevice functions (new and existing) meet their defined technical specifications.Met: "thoroughly tested through verification of specifications and validation"
    Voluntary StandardsCompliance with applicable voluntary standards for anesthesia systems.Met: "Verification of compliance with applicable voluntary standards has also been made"
    Risk MitigationIdentified and mitigated risks associated with the device.Met: "Risk Analysis" was applied during development.
    Software ValidationSoftware components are thoroughly validated.Met: "including software validation"
    Intended UseThe device performs as intended for general inhalation anesthesia and ventilatory support for a wide range of patients.Met: The device is cleared for its stated indications for use.

    2. Sample Size Used for the Test Set and the Data Provenance

    • Test Set Sample Size: Not applicable in the context of human clinical data for this specific device. The testing involved "nonclinical tests of design verification and validation testing," including unit-level testing, integration testing, performance testing, safety testing, and simulated use testing. These are not typically quantified with "sample sizes" in the same way as human subject studies; rather, they involve testing various components, configurations, and scenarios.
    • Data Provenance: Not applicable as no clinical data from patients was used. The data provenance would be from internal engineering and quality testing (e.g., lab test data, simulated environment data).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • Not applicable. Since no clinical testing was performed and no ground truth was established from patient data, there's no mention of experts establishing a ground truth for a test set in a diagnostic or outcome-based scenario. The "ground truth" for nonclinical testing would be engineering specifications and functional requirements.

    4. Adjudication Method for the Test Set

    • Not applicable. There was no clinical test set requiring adjudication in the context of human data. Adjudication methods (like 2+1, 3+1) are typically used for resolving disagreements among human readers/experts in interpreting clinical data.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is an anesthesia system, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance would not be relevant and was not performed.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    • Not applicable. This is an anesthesia machine, not an algorithm-only device. Its performance is inherent in its hardware, software, and integrated components, designed to operate with a medical professional as the "human in the loop" administering anesthesia.

    7. The Type of Ground Truth Used

    • For Nonclinical Testing: The ground truth would be based on engineering specifications, design requirements, and established performance benchmarks derived from the predicate device and relevant industry standards.
    • For Substantial Equivalence: The ultimate "ground truth" for the 510(k) process is the safety and effectiveness profile of the predicate device (GE Datex-Ohmeda Avance, K112722). The new device must demonstrate it meets this same standard.

    8. The Sample Size for the Training Set

    • Not applicable. Given the device type and the absence of clinical studies, there is no mention of a "training set" in the context of machine learning. The testing performed (verification and validation) is based on engineering principles and quality assurance, not statistical training sets.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there was no machine learning training set mentioned.
    Ask a Question

    Ask a specific question about this device

    K Number
    K112722
    Manufacturer
    Date Cleared
    2011-12-16

    (88 days)

    Product Code
    Regulation Number
    868.5160
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K001814, K051092, K092027, K102239

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GE Datex-Ohmeda Avance Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). The device is intended for volume or pressure control ventilation.

    Device Description

    The GE Datex-Ohmeda Avance is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). It represents one of the systems in a long line of products based on the Datex-Ohmeda Excel, Aestiva, and Aespire Anesthesia Systems. It is to be used only by trained and qualified medical professionals in the administration of general anesthesia. The Datex-Ohmeda Avance Anesthesia System supplies set flows of medical gases to the breathing system using electronic gas mixing. Gas flows are selected by the user using the keypad and rotary controller on the main display unit and then displayed as electronic flow meters on the system display unit. The Avance is equipped with a pneumatic back-up O2 delivery system and traditional flow tube, as well. A large selection of frames options, gases, and vaporizers are available to give the user control of the system configuration. The Avance is also available in wall-mount and pendant models. It is available with two or three gases, up to two vaporizer positions and up to three cylinder connections. All models have O2. The Avance comes with up to two optional gases (air, N2O). The Avance systems accept Tec 4, Tec 5, Tec 6, and Tec 7 vaporizers on a Selectatec manifold. Safety features and devices within the Avance are designed to decrease the risk of hypoxic mixtures, agent mixtures and complete power or sudden gas supply failures. The Avance system is available with optional integrated respiratory gas monitoring. When supplied as an option, the integrated respiratory gas monitoring is provided via the Datex-Ohmeda M-Gas Module (M-CAiO and M-CAiOV software revision 3.2 and above cleared via K001814) and E-Gas Module (E-CAiOVX software revision 4.5 and above cleared via K051092) which can be physically integrated into the Avance, receive electronic power from the Avance and communicate measured values to the Avance for display on the system display unit. The Datex-Ohmeda 7900 Anesthesia Ventilator is used in the Avance Anesthesia System. It is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during surgical procedures. The 7900 ventilator is equipped with a built-in monitoring system for inspired oxygen, airway pressure and exhaled volume. Sensors in the breathing circuit are used to control and monitor patient ventilation as well as measure inspired oxygen concentration. This allows for the compensation of compression losses, fresh gas contribution and small leakage in the breathing absorber, bellows and system. User setting and microprocessor calculations control breathing patterns. The user interface keeps settings in memory. The user may change settings with a simple and secure setting sequence. A bellows contains breathing gasses to be delivered to the patient. Positive End Expiratory Pressure (PEEP) is regulated electronically. Positive pressure is maintained in the breathing system so that any leakage that occurs is outward. An RS-232 serial digital communications port connects to and communicates with external devices. Ventilator modes for the device include Volume Control Ventilation (VCV), Pressure Control Ventilation (PCV) (optional), Synchronized Intermittent Mandatory Ventilation/Pressure, Support (SIMV/PSV) (optional), Pressure Support Ventilation (PSVPro) (optional), Synchronized Intermittent Mandatory Ventilation-Pressure Control (SIMV-PC) (optional), Pressure Control Ventilation-Volume Guaranteed (PCV-VG) (optional), Constant Positive Airway Pressure/Pressure Support Ventilation (CPAP/PSV), and Volume Control Ventilation Mode for Cardiac Bypass Mode. Ventilator parameters and measurements are displayed on the system display unit. Several frame configurations are available, including one that allows for the physical integration of the GE Monitors (most recently cleared Carescape B850 via K092027 and B650 cleared on K102239). This configuration also provides cable management solutions such that the necessary connections from the monitor display unit to the monitor are hidden within the Avance frame. Additional configurations allow for the mounting of various patient monitors on the top shelf of the Avance.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification summary for the GE Datex-Ohmeda Avance Anesthesia System. This document focuses on demonstrating substantial equivalence to previously cleared predicate devices, rather than establishing de novo acceptance criteria and performing a study to prove they are met.

    Therefore, the requested information categories concerning acceptance criteria and study design are not directly applicable or present in this type of regulatory submission. The document explicitly states: "The modifications made to the GE Datex-Ohmeda Avance did not require clinical testing."

    However, I can extract information related to the non-clinical testing that was performed to support the substantial equivalence claim, which serves as a form of acceptance in the context of a 510(k) submission.

    Here's a breakdown of the available information, addressing as many of your points as possible:

    Acceptance Criteria and Reported Device Performance

    Since this is a 510(k) submission for an updated device, the "acceptance criteria" are implicitly tied to demonstrating that the updated device performs equivalently or better than the predicate devices and meets its own verified specifications. The document does not provide a table for specific performance metrics and their acceptance ranges. Instead, it describes a robust non-clinical testing approach.

    Acceptance Criteria CategoryReported Device Performance / Evaluation Method
    Performance (General)Thorough verification of specifications and validation, including software validation.
    SafetyCompliance with applicable voluntary standards.
    Software FunctionalityComplete evaluation of modified software features (e.g., pediatric improvements, checkout improvements, new ventilation mode, lung mechanics procedures, display enhancements).
    Risk MitigationRisk Analysis conducted.
    Design IntegrityRequirements Reviews, Design Reviews conducted.
    Unit Level TestingModule verification.
    Integration TestingSystem verification.
    Simulated Use TestingValidation performed.

    Study Details (Non-Clinical)

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size: Not specified as a number of "samples" in the clinical sense. The testing involved various modules, integrated systems, and simulated use of the device.
      • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). However, the testing was conducted internally by the manufacturer (Datex-Ohmeda Inc.) as part of their development process. It is by nature "prospective" in the sense that the tests were designed and executed to evaluate the new device's modifications.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Number of Experts: Not specified. This type of non-clinical engineering validation typically involves internal engineers, quality assurance personnel, and potentially subject matter experts who understand anesthesia systems.
      • Qualifications of Experts: Not specified, but implied to be qualified technical and engineering personnel involved in the device's design, development, and testing at Datex-Ohmeda.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Adjudication Method: Not applicable. This concept pertains to clinical studies where independent reviewers agree on ground truth. In non-clinical engineering testing, "adjudication" is managed through formal test protocols, defect tracking, and review processes by the development and quality teams.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • MRMC Study: No. This is not an AI-assisted diagnostic device, but an anesthesia system. Therefore, MRMC studies are not relevant. This document explicitly states: "The modifications made to the GE Datex-Ohmeda Avance did not require clinical testing."
    5. If a standalone (i.e., algorithm only without human-in-the loop performance) was done:

      • Standalone Performance: The non-clinical testing evaluates the device's inherent performance characteristics, including its software logic and hardware functionality, in a standalone manner (i.e., its ability to perform its functions as designed without direct human intervention in the moment of testing, though human users operate the system during its intended use). The testing verifies that the "algorithm only" (software) performs as specified.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • Ground Truth: For non-clinical engineering validation, the "ground truth" is defined by the device's design specifications, industry standards, and regulatory requirements. The device is tested against these predefined criteria to ensure it performs as intended and safely.
    7. The sample size for the training set:

      • Training Set Sample Size: Not applicable. This is not an AI/machine learning device that requires a training set.
    8. How the ground truth for the training set was established:

      • Ground Truth for Training Set: Not applicable, as there is no training set for an AI/ML model.

    In summary: The GE Datex-Ohmeda Avance Anesthesia System underwent a non-clinical verification and validation process. The "acceptance criteria" were met through demonstrating compliance with design specifications, software validation, risk analysis, and adherence to voluntary standards. No clinical studies were deemed necessary due to the nature of the modifications and the substantial equivalence claim.

    Ask a Question

    Ask a specific question about this device

    K Number
    K111116
    Manufacturer
    Date Cleared
    2011-09-09

    (141 days)

    Product Code
    Regulation Number
    868.5895
    Why did this record match?
    Reference Devices :

    K021175, K041775, K001814, K051092, K023454, K052582

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GE Datex-Ohmeda Engström family of ventilators (Engström Carestation and Engström Pro) are designed to provide mechanical ventilation for adults and pediatrics weighing 5kg and above having degrees of pulmonary impairment varying from minor to severe. Optional Neonatal capabilities on Engström family expand the patient range to 0.25 kg.

    The GE Datex-Ohmeda Engström family of ventilators are microprocessor based, electronically controlled, pneumatically driven ventilators that include integrated FiO2, airway pressure, spirometry and volume monitoring. Options include an Aerogen Aeroneb nebulizer, data capture accessory and an integrated air compressor. Options available on Engström Carestation only include integrated respiratory gas monitoring capabilities via various Datex-Ohmeda patient monitoring modules listed in the product labeling, capabilities to measure SpiroDynamics via an intratracheal pressure sensor in patients using sized 6.5 tracheal tubes and larger, and calculation of functional residual capacity of mechanically ventilated patients using Nitrogen Wash In/Wash Out method.

    Not all features are available with all patient populations.

    The Engström Carestation is not a pulmonary function calculation device.

    The system is designed for facility use, including within-facility transport, and should only be used under the orders of a clinician.

    Device Description

    The GE Datex-Ohmeda Engström family of ventilators (Engström Carestation and Engström Pro) are flexible, adaptable, and intuitive critical care ventilators. A wide selection of performance options gives the user full control of the system configuration. The Engström Carestation is a complete system featuring patient monitoring, patient ventilation, and the capability of interfacing with central information management systems. The Engström Pro is a defeatured variant of the Engström Carestation.

    Both the GE Datex-Ohmeda Engström Carestation and Engström Pro are designed to provide mechanical ventilation for adults and pediatrics weighing 5kg and above having degrees of pulmonary impairment varying from minor to severe. Optional Neonatal capabilities expand its patient range to 0.25 kg.

    The modes of ventilation currently available include:

    1. Volume Controlled (VCV)
    2. Pressure Controlled (PCV)
    3. Pressure Controlled, Volume Guaranteed (PCV-VG)
    4. Synchronized Intermittent Mandatory Ventilation, Volume Controlled (SIMV-VC)
    5. Synchronized Intermittent Mandatory Ventilation, Pressure Controlled (SIMV-PC)
    6. Synchronized Intermittent Mandatory Ventilation, Pressure Controlled Volume Guarantee (SIMV-PCVG)
    7. Bi-level Airway Pressure Ventilation
    8. Constant Positive Airway Pressure/Pressure Support Ventilation (CPAP/PSV)
    9. Apnea backup (available in SIMV-VC, SIMV-PC, SIMV-PCVG/BiLevel-VG, BiLevel, CPAP/PSV, and VG-PS)
    10. Non-invasive ventilation (NIV), not available in neonatal mode
    11. Infant Nasal CPAP (nCPAP), only available in neonatal mode
    12. Volume Guarantee, Pressure Support (VG-PS), only available in neonatal mode

    The GE Datex-Ohmeda Engström Carestation and Engström Pro are microprocessor based, electronically controlled, pneumatically driven ventilators that include integrated FiO2, airway pressure, spirometry and volume monitoring and an Aerogen Aeroneb Pro nebulizer control board.

    The ventilator consists of two main components: a display and a ventilator unit. The display allows the user to interface with the system and control settings through use of soft keys on the display, a com wheel, and a resistive touch screen. The ventilator unit controls electrical power, nebulization, and pneumatic gas flow to and from the patient. The Engstrom Carestation also includes a module bay that allows the integration of various Datex-Ohmeda patient monitoring modules with the ventilator.

    The user interface for control of nebulization is provided via the ventilator display unit. The Aerogen Aeroneb Pro Nebulizer board (K021175) is provided standard with the unit. Nebulizers are options for both the Engström Carestation and Engström Pro. Users have the option to configure the system to use an external pneumatic nebulizer in place of the Aerogen.

    Optional accessories common to both Engström Carestation and Engström Pro include a trolley/cart, integrated air compressor, support arm, humidifier and water trap mounting brackets, and a data capture accessory. The GE Datex-Ohmeda EV Air Compressor is intended for use as an accessory to provide a dry, filtered, breathable compressed air supply. The compressor has no alarm functions. The Engström Carestation or Engström Pro provides all alarm functions and reactions to a failure of the compressed gas supply. The compressor is installed in the base of the ventilator cart. The compressor is powered from AC mains only. A source of compressed oxygen is required to be connected to Engström Carestation/Engström Pro equipped with the optional compressor. The compressor was cleared in K041775.

    Additional optional accessories specific to the Engström Carestation include airway modules, intratracheal pressure sensor, auxiliary electrical outlets, and module bay. Optional functionality specific to the Engström Carestation includes integrated respiratory gas monitoring, capabilities to measure SpiroDynamics via a GE supplied intratracheal pressure sensor in patients using sized 6.5 tracheal tubes and larger, and calculation of functional residual capacity of mechanically ventilated patients using Nitrogen Wash In/Wash Out method. The integrated respiratory gas monitoring is provided via the Datex-Ohmeda Gas Modules, M-C, M-CO, M-COV, M-COV, M-COVX, M-CaiO, M-CAiOV, M-CAiOVX, rev 3.2 software and higher (K001814), E-CO, E-COV, E-COVX, E-CAiO, E-CAiOV, E-CAiOVX (K051092), or M-Mini-CO2 Module (K023454) or E-MiniC module (K052582) which are physically integrated into the Engström Carestation, receive electronic power from the Engstrom Carestation and communicate measured values to the Engström Carestation for display on the system display unit.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the GE Datex-Ohmeda Engström Ventilator family (Engström Carestation and Engström Pro). This submission is for an updated version of an existing device, primarily involving software changes.

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not explicitly provide a table of acceptance criteria with specific performance metrics (e.g., accuracy, sensitivity, specificity) for the device's functions. Instead, it focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing and verification.

    However, the general acceptance criteria for this type of submission are implied through the testing methods described:

    Acceptance Criteria CategoryReported Device Performance (Summary from Text)
    Safety and EffectivenessDemonstrated through verification of specifications, software validation, and compliance with applicable standards. No new questions of safety and effectiveness compared to predicate devices.
    Functional EquivalenceOperates with updated software primarily adding touch screen functionality and minor changes based on customer feedback (e.g., nuisance alarms). Performs in a substantially equivalent manner to predicate devices.
    Compliance with StandardsVerification of compliance with applicable standards completed for safe use, including electrical safety and electromagnetic compatibility testing.
    Intended UseContinues to provide mechanical ventilation for adults and pediatrics (5kg and above, with optional neonatal capabilities for 0.25kg) with pulmonary impairment.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated. The text mentions "Testing on unit level," "Integration testing," "Performance Testing (Verification)," and "Safety Testing (Verification)" but does not specify the number of devices or scenarios used for these tests.
    • Data Provenance: The testing appears to be conducted in-house by GE Healthcare/Datex-Ohmeda Inc. as part of their development and verification process. The data is retrospective in the sense that it's based on internal testing conducted on the device prior to submission, rather than prospective clinical trials. There is no mention of data origin by country.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This submission is for a ventilator, and the testing described is primarily engineering verification and validation against technical specifications and predicate device performance. It does not involve interpretation of medical images or diagnostic outputs that would require clinical expert "ground truth" establishment in the way a diagnostic AI device would. "Simulated Use/User Requirements Testing (Validation)" was performed, implying user involvement, but the number and qualifications of these users/experts are not specified, nor is their role in establishing a "ground truth" defined in the typical diagnostic context.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, the testing is against technical specifications and predicate device performance, not against expert-adjudicated ground truth.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

    No MRMC comparative effectiveness study was done. The text explicitly states: "The modifications made to the Engstrom ventilator did not require clinical testing."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The device itself is a standalone medical device (a ventilator), not an algorithm intended for diagnostic interpretation. The software updates were evaluated for their standalone functionality as part of the ventilator system. The non-clinical tests verified and validated the "safety and functionality" of the ventilator with the software changes.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is its adherence to:

    • Technical Specifications: The device's design specifications for ventilation parameters, safety features, alarm conditions, etc.
    • Predicate Device Performance: The established safety and effectiveness of the previously cleared Engstrom Ventilator (K093886) and Hamilton G5 Ventilator (K070513).
    • Applicable Standards: Compliance with relevant medical device standards (e.g., electrical safety, electromagnetic compatibility).

    8. The Sample Size for the Training Set

    Not applicable. This document describes a medical device (ventilator) with software updates, not an AI/ML algorithm that requires a "training set" in the machine learning sense. The software updates were based on "customer feedback" and "minor changes to bring the product in line with current specifications," but this is not equivalent to a supervised learning training process.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there was no AI/ML training set in the context described by the question.

    Ask a Question

    Ask a specific question about this device

    K Number
    K110213
    Manufacturer
    Date Cleared
    2011-07-20

    (176 days)

    Product Code
    Regulation Number
    868.5160
    Why did this record match?
    Reference Devices :

    K001814, K051092, K973985, K973895, K092027, K102239

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GE Datex-Ohmeda Aisys Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). The device is intended for volume or pressure control ventilation. The Aisys is not suitable for use in a MRI environment.

    Device Description

    The GE Datex-Ohmeda Aisys is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). It represents one of the systems in a long line of products based on the Datex-Ohmeda Excel. Aestiva, Aespire, and Avance Anesthesia Systems. It is to be used only by trained and qualified medical professionals.

    The GE Datex-Ohmeda Aisys Carestation supplies set flows of medical gases to the breathing system using electronic gas mixing. Gas flows are selected by the user using the keypad and rotary controller on the main display unit and then displayed as electronic flow indicators on the svstem display unit. The Aisvs is equipped with a pneumatic back-up O2 delivery system and traditional flow tube, as well. A large selection of frames, gases, and vaporizers are available to give the user control of the system configuration. The Aisys is also available in a pendant model. It is available with two or three gases, and up to three cylinder connections. All models have 02. The Aisys comes with up to two optional gases (air, N2O). Safety features and devices within the Aisys are designed to decrease the risk of hypoxic mixtures and complete power or sudden gas supply failures. The Aisys system is available with optional integrated respiratory gas monitoring. When supplied as an option, the integrated respiratory gas monitoring is provided via the Datex-Ohmeda M-Gas Module (M-CAiO and M-CAiOV software revision 3.2 and above cleared via K001814) and E-Gas Module (E-CAiOVX software revision 4.5 and above cleared via K051092) which can be physically integrated into the Aisys, receive electronic power from the Aisys and communicate measured values to the Aisys for display on the system display unit.

    The anesthetic agent delivery for the Aisys is controlled via an anesthesia computer through user input from the central display. The vaporization technology is based upon the electronic vaporizer cleared as part of the Datex-Ohmeda Anesthesia Delivery Unit (ADU) cleared via K973985. An Aladin cassette (also cleared as part of K973895) or Aladin 2 is inserted into the active cassette bay. The cassette holds the agent to be delivered - Halothane, Enflurane, Isoflurane, Desflurane or Sevoflurane. Agent is delivered as a percent volume/volume. The Aisvs is designed to allow only one active cassette at a time. Per the user input into the main display, valves within the active cassette bay will open and allow agent to be delivered. The agent is mixed with gas from the FGC unit. After mixing, the combination of gases and agent is delivered to the breathing system and then on to the patient.

    The Datex-Ohmeda 7900 Anesthesia Ventilator is used in the Aisys Anesthesia System. It is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during surgical procedures. The 7900 ventilator is equipped with a built-in monitoring system for inspired oxygen, airway pressure and exhaled volume. Sensors in the breathing circuit are used to control and monitor patient ventilation as well as measure inspired oxygen concentration. This allows for the compensation of compression losses, fresh gas contribution and small leakage in the breathing absorber, bellows and system. User setting and microprocessor calculations control breathing patterns. The user interface keeps settings in memory. The user may change settings with a simple and secure setting sequence. A bellows contains breathing gasses to be delivered to the patient. Positive End Expiratory Pressure (PEEP) is requlated electronically. Positive pressure is maintained in the breathing system so that any leakage that occurs is outward. An RS-232 serial digital communications port connects to and communicates with external devices. Ventilator modes for the device include Volume Control Ventilation (VCV), Pressure Control Ventilation (PCV) (optional), Synchronized Intermittent Mandatory Ventilation/Pressure Support (SIMV/PSV) (optional), Pressure Support Ventilation (PSVPro) (optional), Synchronized Intermittent Mandatory Ventilation-Pressure Control (SIMV-PC) (optional), Pressure Control Ventilation-Volume Guaranteed (PCV-VG) (optional) and Constant Positive Airway Pressure/Pressure Support Ventilation (CPAP/PSV). Ventilator parameters and measurements are displayed on the system display unit.

    AI/ML Overview

    The provided text describes a premarket notification (510(k)) for the GE Datex-Ohmeda Aisys Anesthesia System. This document focuses on demonstrating substantial equivalence to a predicate device, rather than providing detailed acceptance criteria and a study report as would be found for a novel device.

    The submission states that the modifications made did not require clinical testing, and the functionality was evaluated through nonclinical tests of design verification and validation testing. Therefore, the information requested regarding acceptance criteria and a study proving the device meets those criteria for a new clinical claim is not present in this document.

    However, I can extract information about the type of testing performed and the conclusion regarding substantial equivalence.

    Here's a breakdown of the available information based on your request, with an emphasis on what is not present:

    1. A table of acceptance criteria and the reported device performance

    • Not explicitly provided in the document. The document states "The GE Datex-Ohmeda Aisys has been thoroughly tested through verification of specifications and validation, including software validation," but it does not list specific, quantitative acceptance criteria or corresponding reported performance metrics for clinical efficacy or diagnostic accuracy.
    • The overall "acceptance criteria" for this 510(k) submission is demonstrating substantial equivalence to predicate devices. The conclusion states: "The performance data demonstrates the device is substantially equivalent to the predicates."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable/Not provided. No clinical test set or patient data is mentioned as the basis for the 510(k) in this document. The testing was non-clinical.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable/Not provided. No clinical ground truth was established as no clinical testing was performed for this 510(k). Non-clinical testing would involve engineering and quality assurance experts.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable/Not provided. No clinical test set or adjudication method is mentioned.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is an anesthesia system, not an AI-assisted diagnostic tool. No MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is an anesthesia system. The software updates relate to user interface, control of gas delivery, and ventilator modes, not standalone algorithmic performance in a diagnostic context. "Standalone" performance testing would be more relevant to the device's functional operation (e.g., accuracy of gas delivery, ventilator parameters) which was covered by non-clinical verification and validation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not applicable/Not provided in a clinical sense. For the non-clinical testing, the "ground truth" would be the device's technical specifications and established engineering standards, against which its performance was verified and validated.

    8. The sample size for the training set

    • Not applicable/Not provided. This document is not describing a machine learning or AI model that requires a training set in the conventional sense. The "software updates" are likely improvements to existing control algorithms and user interface logic.

    9. How the ground truth for the training set was established

    • Not applicable/Not provided. As above, no training set for a machine learning model is mentioned.

    Summary of what is present regarding testing:

    The document states:

    • Reason for 510(k): Updates to the software for the GE Datex-Ohmeda Aisys Anesthesia System, introducing several new features (e.g., pediatric improvements, new ventilation mode CPAP/PSV with backup breaths, lung mechanics procedure options, UI enhancements).
    • Basis for Substantial Equivalence: Comparison to legally marketed predicate devices (GE Datex-Ohmeda Aisys (K090233) and GE Datex-Ohmeda Engstrom ventilator (K093886)). No changes to intended use or fundamental scientific technology.
    • Testing Performed:
      • Nonclinical Testing: "thoroughly tested through verification of specifications and validation, including software validation."
      • Quality Assurance Measures applied:
        • Risk Analysis
        • Requirements Reviews
        • Design Reviews
        • Testing on unit level (Module verification)
        • Integration testing (System verification)
        • Performance testing (Verification)
        • Safety testing (Verification)
        • Simulated use testing (Validation)
      • Clinical Testing: "The modifications made to the GE Datex-Ohmeda Aisys did not require clinical testing. The functionality of the modified software features were completely evaluated by performing nonclinical tests of design verification and validation testing."
    • Conclusion: The performance data (nonclinical) demonstrates the device is substantially equivalent to the predicates, with no new questions of safety and effectiveness.

    In essence, this 510(k) leverages the predicate devices' established safety and effectiveness, and for the software updates, relies on rigorous non-clinical engineering verification and validation testing rather than new clinical trials or performance studies against specific acceptance criteria for a novel claim.

    Ask a Question

    Ask a specific question about this device

    K Number
    K093886
    Date Cleared
    2010-05-03

    (136 days)

    Product Code
    Regulation Number
    868.5895
    Why did this record match?
    Reference Devices :

    K021175, K041775, K001814, K023454

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GE Datex-Ohmeda Engström family of ventilators (Engström Carestation and Engström Pro) are designed to provide mechanical ventilation for adults and pediatrics weighing 5kg and above having degrees of pulmonary impairment varying from minor to severe. Optional Neonatal capabilities on Engström family expand the patient range to 0.25 kg.

    The GE Datex-Ohmeda Engström family of ventilators are microprocessor based, electronically controlled, pneumatically driven ventilators that include integrated FiO2, airway pressure, spirometry and volume monitoring. Options include an Aerogen Aeroneb nebulizer, data capture accessory and an integrated air compressor. Options available on Engström Carestation only include integrated respiratory gas monitoring capabilities via various Datex-Ohmeda patient monitoring modules listed in the product labeling, capabilities to measure SpiroDynamics via an intratracheal pressure sensor in patients using sized 6.5 tracheal tubes and larger, and calculation of functional residual capacity of mechanically ventilated patients using Nitrogen Wash In/Wash Out method.

    Not all features are available with all patient populations.

    The Engström Carestation is not a pulmonary function calculation device.

    The system is designed for facility use, including within-facility transport, and should only be used under the orders of a clinician.

    Device Description

    The GE Datex-Ohmeda Engström family of ventilators (Engström Carestation and Engström Pro) are flexible, adaptable, and intuitive critical care ventilators. A wide selection of performance options gives the user full control of the system configuration. The Engström Carestation is a complete system featuring patient monitoring, patient ventilation, and the capability of interfacing with central information management systems. The Engström Pro is a defeatured variant of the Engström Carestation.

    Both the GE Datex-Ohmeda Engström Carestation and Engstrom Pro are designed to provide mechanical ventilation for adults and pediatrics weighing 5kg and above having degrees of pulmonary impairment varying from minor to severe. Optional Neonatal capabilities expand its patient range to 0.25 kg.

    The modes of ventilation currently available include:

    1. Volume Controlled (VCV)
    2. Pressure Controlled (PCV)
    3. Pressure Controlled, Volume Guaranteed (PCV-VG)
    4. Synchronized Intermittent Mandatory Ventilation, Volume Controlled (SIMV-VC)
    5. Synchronized Intermittent Mandatory Ventilation, Pressure Controlled (SIMV-PC)
    6. Synchronized Intermittent Mandatory Ventilation, Pressure Controlled Volume Guarantee (SIMV-PCVG)
    7. Bi-level Airway Pressure Ventilation
    8. Constant Positive Airway Pressure/Pressure Support Ventilation (CPAP/PSV)
    9. Apnea backup (active in Bi-level and CPAP/PSV)
    10. Non-invasive ventilation (NIV), note that NIV is not available in neonatal mode
    11. Infant Nasal CPAP (nCPAP)
    12. Volume Guarantee, Pressure Support (VG-PS)

    The GE Datex-Ohmeda Engström Carestation and Engström Pro are microprocessor based, electronically controlled, pneumatically driven ventilators that include integrated FiO2, airway pressure, spirometry and volume monitoring and an Aerogen Aeroneb Pro nebulizer control board.

    The ventilator consists of two main components: a display and a ventilator unit. The display allows the user to interface with the system and control settings. The ventilator unit controls electrical power, nebulization, and pneumatic gas flow to and from the patient. The Engstrom Carestation also includes a module bay that allows the integration of various Datex-Ohmeda patient monitoring modules with the ventilator.

    The user interface for control of nebulization is provided via the ventilator display unit. The Aerogen Aeroneb Pro Nebulizer board (K021175) is provided standard with the unit. Nebulizers are options for both the Engstrom Carestation and Engstrom Pro. Users have the option to configure the system to use an external pneumatic nebulizer in place of the Aerogen.

    Optional accessories common to both Engström Carestation and Engström Pro include a trolley/cart, integrated air compressor, support arm, humidifier and water trap mounting brackets, and a data capture accessory. The GE Datex-Ohmeda EV Air Compressor is intended for use as an accessory to provide a dry, filtered, breathable compressed air supply. The compressor has no alarm functions. The Engstrom Carestation or Engstrom Pro provides all alarm functions and reactions to a failure of the compressed gas supply. The compressor is installed in the base of the ventilator cart. The compressor is powered from AC mains only. A source of compressed oxygen is required to be connected to Engstrom Carestation/Engström Pro equipped with the optional compressor. The compressor was cleared in K041775.

    Additional optional accessories specific to the Engström Carestation include airway modules, intratracheal pressure sensor, auxiliary electrical outlets, and module bay. Optional functionality specific to the Engstrom Carestation include integrated respiratory gas monitoring, capabilities to measure SpiroDynamics via a GE supplied intratracheal pressure sensor in patients using sized 6.5 tracheal tubes and larger, and calculation of functional residual capacity of mechanically ventilated patients using Nitrogen Wash Out method. The integrated respiratory gas monitoring is provided via the Datex-Ohmeda M-Gas Module (M-C, M-COV, M-COVX, M-CaiO, M-CAiOV, M-CAiOVX, rev 3.2 software and higher, (K001814) or Mini-CO2 Module (K023454) which are physically integrated into the Engstrom Carestation, receive electronic power from the Engström Carestation and communicate measured values to the Engström Carestation for display on the system display unit.

    AI/ML Overview

    This document is a 510(k) premarket notification for the GE Datex-Ohmeda Engström Ventilator, seeking substantial equivalence to previously cleared devices. It does not contain information about acceptance criteria or a study proving the device meets specific performance criteria through clinical data.

    Specifically, the document states:

    • "The modifications made to the Engstrom ventilator did not require clinical testing."

    Therefore, I cannot provide the requested information regarding acceptance criteria and a study proving the device meets those criteria, as no such clinical study data or acceptance criteria related to a performance study are present in the provided text. The submission focuses on verification of specifications, validation (including software validation), electrical safety, and electromagnetic compatibility testing, rather than a clinical performance study with defined acceptance criteria for AI or a detailed comparison of device performance against a benchmark.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090233
    Manufacturer
    Date Cleared
    2009-03-05

    (34 days)

    Product Code
    Regulation Number
    868.5160
    Why did this record match?
    Reference Devices :

    K001814, K051092, K973985, K973895, K030812, K051400

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GE Datex-Ohmeda Aisys Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. The device is intended for volume or pressure control ventilation. The Aisys is not suitable for use in a MRI environment.

    Device Description

    The GE Datex-Ohmeda Aisys is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. It represents one of the systems in a long line of products based on the Datex-Ohmeda Excel, Aestiva, Aespire, and Avance Anesthesia Systems. It is to be used only by trained and qualified medical professionals.

    The GE Datex-Ohmeda Aisys Carestation supplies set flows of medical gases to the breathing system using electronic gas mixing. Gas flows are selected by the user using the keypad and rotary controller on the main display unit and then displayed as electronic flow indicators on the system display unit. The Aisys is equipped with a pneumatic back-up O2 delivery system and traditional flow tube, as well. A large selection of frames, gases, and vaporizers are available to give the user control of the system configuration. The Aisys is also available in a pendant model. It is available with two or three gases, and up to three cylinder connections. All models have O2. The Aisys comes with up to two optional gases (air, N2O). Safety features and devices within the Aisys are designed to decrease the risk of hypoxic mixtures and complete power or sudden gas supply failures. The Aisys system is available with optional integrated respiratory gas monitoring. When supplied as an option, the integrated respiratory gas monitoring is provided via the Datex-Ohmeda M-Gas Module (M-CAiO and M-CAiOV software revision 3.2 and above K# 001814) and E-Gas Module (E-CAiOVX cleared via K051092) which can be physically intecrated into the Aisys, receive electronic power from the Aisys and communicate measured values to the Aisys for display on the system display unit.

    The anesthetic agent delivery for the Aisys is controlled via an anesthesia computer through user input from the central display. The vaporization technology is based upon the electronic vaporizer cleared as part of the Datex-Ohmeda Anesthesia Delivery Unit (ADU) cleared via K973985. An Aladin cassette (also cleared as part of K973895) or Aladin 2 is inserted into the active cassette bay. The cassette holds the agent to be delivered - Halothane, Enflurane, Isoflurane. Desflurane or Sevoflurane. Agent is delivered as a percent volume/volume. The Aisys is designed to allow only one active cassette at a time. Per the user input into the main display, valves within the active cassette bay will open and allow agent to be delivered. The agent is mixed with gas from the FGC unit. After mixing, the combination of gases and agent is delivered to the breathing system and then onto the patient.

    The Datex-Ohmeda 7900 Anesthesia Ventilator is used in the Aisys Anesthesia System. It is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during surgical procedures. The 7900 ventilator is equipped with a built-in monitoring system for inspired oxygen, airway pressure and exhaled volume. Sensors in the breathing circuit are used to control and monitor patient ventilation as well as measure inspired oxygen concentration. This allows for the compensation of compression losses, fresh gas contribution and small leakage in the breathing absorber, bellows and system. User setting and microprocessor calculations control breathing patterns. The user interface keeps settings in memory. The user may change settings with a simple and secure setting sequence. A bellows contains breathing gasses to be delivered to the patient. Positive End Expiratory Pressure (PEEP) is requlated electronically. Positive pressure is maintained in the breathing system so that any leakage that occurs is outward. An RS-232 serial digital communications port connects to and communicates with external devices. Ventilator modes for the device include Volume Mode, Pressure Control Mode, Pressure Support with Apnea Backup Mode (Optional) and Synchronized Intermittent Mandatory Ventilation (SIMV) Mode (Optional). Ventilator parameters and measurements are displayed on the system display unit.

    The system display unit is mounted to an arm on the top shelf of the Aisys. The arm is counter balanced and capable of moving vertically and/or horizontally, and also tilting the display, enabling the user to position the display to the most advantageous viewing position. The arm length is limited such that the display position is always within the footprint of the Aisys frame. The arm also supports the mounting of additional display units for a variety of patient monitors.

    Several frame configurations are available, including one that allows for the physical integration of the Datex-Ohmeda S/5 Anesthesia Monitor (most recently cleared via K030812). This configuration also provides cable management solutions such that the necessary connections from the monitor display unit to the monitor are hidden within the Aisys frame. An additional ootion allows the S/5 AM to be linked to the power supply of the Aisys such that when the Aisys is turned on, the S/5 AM is also turned on. Additional configurations allow for the mounting of various patient monitors on the top shelf of the Aisys.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the GE Datex-Ohmeda Aisys Anesthesia System. The key information regarding acceptance criteria and study details is presented in the "SUMMARY OF NONCLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(1)(3)" and "SUMMARY OF CLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(2)" sections.

    Here's the breakdown of the information requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the GE Datex-Ohmeda Aisys Anesthesia System are compliance with various international and national standards related to medical electrical equipment, anesthesia workstations, and gas cylinders. The reported device performance is that it meets these standards.

    Acceptance Criteria (Standard)Reported Device Performance
    EN 740 :1998 (Anesthesia Workstations and their components)Compliant
    EN 60601-1:1990 (Medical Electrical Equipment Part 1: General Requirements for Safety)Compliant
    EN 60601-1-1:2000 (Medical Electrical Equipment - Medical Electrical Systems)Compliant
    EN 60601-1-2:2001 (Medical Electrical Equipment - Electromagnetic Compatibility)Compliant
    IEC 60601-1-4:2000 (Safety of Programmable Electronic Medical Systems)Compliant
    EN 475:1995 (Electrically Generated Alarm Signals)Compliant
    EN 850:1997 (Small Medical Gas Cylinders - Pin Indexed)Compliant
    EN 980:1997 (Graphical Symbols)Compliant
    EN 1041:1998 (Information to be supplied with medical devices)Compliant
    EN 1089-3:1997 (Color coding for medical gases)Compliant
    ISO5356-1:1996 (Conical Connectors)Compliant
    EN 1820:1997 (Reservoir Bags)Compliant
    IEC 60601-2-13:1998 (Particular requirements for the safety of anaesthetic workstations)Compliant
    Software ValidationCompleted (thoroughly tested through verification and validation)
    Verification of SpecificationsCompleted (thoroughly tested)

    2. Sample Size for the Test Set and Data Provenance

    The document does not specify a "test set" in the context of a clinical study with a defined sample size for evaluating performance against an explicit statistical endpoint. Instead, the evaluation focuses on non-clinical testing, including verification of specifications and software validation, and compliance with established international standards.

    The term "data provenance" is not applicable in the traditional sense of clinical data here, as the testing described is primarily engineering and regulatory compliance testing rather than clinical study data from patients.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This information is not applicable as the evaluation relies on compliance with technical standards and internal verification/validation processes rather than expert-established ground truth from clinical cases. No external "experts" in the context of clinical ground truth determination are mentioned for the non-clinical testing.

    4. Adjudication Method for the Test Set

    This information is not applicable as there is no mention of a test set requiring adjudication in the context of human expert review. The evaluation is based on meeting technical specifications and standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not performed, and is not applicable. The device is an anesthesia system, not an AI-assisted diagnostic tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device itself is a standalone anesthesia system. While it has embedded software and automated functions, the evaluation is of the complete system's compliance with safety and functional standards, not a standalone algorithm's performance in isolation from the hardware or human operator interaction. The "software validation" performed addresses the functionality of the algorithms within the system.

    7. The type of ground truth used

    The "ground truth" for this device's evaluation is primarily compliance with established international and national engineering, safety, and performance standards (e.g., EN, IEC, ISO standards) and internal product specifications. For the software, it's defined by the software requirements and design specifications.

    8. The sample size for the training set

    This information is not applicable. The device is an anesthesia system, and its development and testing do not involve a "training set" in the context of machine learning model development. The software development follows traditional engineering verification and validation processes.

    9. How the ground truth for the training set was established

    This information is not applicable as there is no "training set" in the context of AI/machine learning. The "ground truth" for the device's functionality is established by its design specifications and the requirements of the relevant industry standards.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2