K Number
K051092
Device Name
DATEX-OHMEDA S/5 COMPACT AIRWAY MODULE (MODEL FAMILY E-CAIOVX) E-CAIOVX, E-CAIOV, E-CAIO, E-COVX, E-COV, E-CO & ACCESACC
Manufacturer
Date Cleared
2005-05-25

(27 days)

Regulation Number
868.1720
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Datex-Ohmeda S/5™ Compact Airway Module, E-CAiOVX family is intended to be used TIE Datex-Ohmeda 07-5 - Oct plant in monitors for monitoring respiratory, ventilatory and gas exchange parameters of hospitalized patients. The Datex-Ohmeda S/5™ Compact Airway Module, E-CAiOVX family is indicated for The Date.x-Onlineda 075 - Comparation (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiration rate) ventilation (airway pressure, volume and flow) and gas luchange status (Oxygen Consumption VO2, Carbon Dioxide production VCO2, Respiratory Cxchange status (Oxygon Octountiples, EE). Gas exchange status monitoring is not indicated in the Quottent RQ, and Energy Expendhadevice is indicated for use by qualified medical personnel only.
Device Description
The Datex-Ohmeda S/5™ Compact Airway Module, E-CAiOVX is a double-width plug-in parameter module for monitoring respiratory (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiration rate), ventilatory (airway pressure, volume and flow) and gas exchange parameters (Oxygen Consumption VO2, Carbon Dioxide production VCO2, Respiratory Quotient RQ, and Energy Expenditure, EE). The E-CAiOVX is a Compact Airway measuring module for a modular monitoring system. The E-CAiOVX module consists of: - AOVA infrared measuring sensor for measuring CO2, N2O and anaesthetic agents - Paramagnetic O2 sensor . - Side-Stream Spirometry measurement . - Gas Exchange measurement - Halothane, Sevoflurane, Isoflurane and Desflurane).
More Information

Not Found

No
The document describes a standard respiratory and gas monitoring module using established sensor technologies (infrared, paramagnetic, spirometry). There is no mention of AI, ML, or any related concepts in the intended use, device description, or performance studies summary.

No
The device is described as a monitoring module for respiratory, ventilatory, and gas exchange parameters, not for providing therapy.

Yes

The device monitors respiratory, ventilatory, and gas exchange parameters, which are used to assess a patient's physiological state and identify potential issues, thereby contributing to diagnosis.

No

The device description explicitly lists hardware components like infrared measuring sensors, paramagnetic O2 sensors, and side-stream spirometry measurement.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
  • Device Function: The Datex-Ohmeda S/5™ Compact Airway Module, E-CAiOVX family directly monitors respiratory, ventilatory, and gas exchange parameters in vivo (within the living patient's airway). It measures gases and pressures directly from the patient's breathing.
  • Lack of Sample Analysis: The description does not mention the analysis of any biological samples taken from the patient.

Therefore, since the device is measuring physiological parameters directly from the patient's airway rather than analyzing samples taken from the body, it falls outside the definition of an In Vitro Diagnostic device.

N/A

Intended Use / Indications for Use

The Datex-Ohmeda S/5™ Compact Airway Module, E-CAiOVX family is intended to be used TIE Datex-Ohmeda 07-5 - Oct plant in monitors for monitoring respiratory, ventilatory and gas exchange parameters of hospitalized patients.

The Datex-Ohmeda S/5M Compact Airway Module, E-CAiOVX family is indicated for monitoring hospital patient´s respiration (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiration rate) ventilation (airway pressure, volume and flow) and gas luchange status (Oxygen Consumption VO2, Carbon Dioxide production VCO2, Respiratory Cxchange status (Oxygon Octountiples, EE). Gas exchange status monitoring is not indicated in the Quottent RQ, and Energy Expendhadevice is indicated for use by qualified medical personnel only.

Product codes (comma separated list FDA assigned to the subject device)

CCK, CCL, BZK, CAP, CBR, BZL, CBQ, CBS, NHO, NHQ, NHP

Device Description

The Datex-Ohmeda S/5™ Compact Airway Module, E-CAiOVX is a double-width plug-in parameter module for monitoring respiratory (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiration rate), ventilatory (airway pressure, volume and flow) and gas exchange parameters (Oxygen Consumption VO2, Carbon Dioxide production VCO2, Respiratory Quotient RQ, and Energy Expenditure, EE). The E-CAiOVX is a Compact Airway measuring module for a modular monitoring system.

The E-CAiOVX module consists of:

  • AOVA infrared measuring sensor for measuring CO2, N2O and anaesthetic agents
  • Paramagnetic O2 sensor
  • Side-Stream Spirometry measurement
  • Gas Exchange measurement
  • Halothane, Sevoflurane, Isoflurane and Desflurane).
    The E-CAiOVX module uses the same measurement technology and accessories as the predicate I he E-CAIOVX (K001814). The main accessories include airway gas sampling lines, D-fend device, M- CAIO + A (1001011). In tubing and D-lite sensors. The sampling line and the water traps, Spironicily measurement table connectors. The monitor is switched on and the gas spirometry tube and the spirometry tube is attached to the D-lite™ airway adapter. The D-lite™ is sampling the and the spironionly tace is that and moisture exchanger (HIME) of the patient's intubation tube. The monitor displays measurements from the E-CAiOVX's and subtype modules intubation table. The moment and loops. All the calculated paraneters can be selected on in the form of numeric values, our res, and view on the E-CAiOVX moduels are taken the usplay, and action. And for wear interface for alarms in Datex-Ohmeda S/5 patient care of by the nost monitor and forlow the asser merines. There are auditory and visual alarms and user adjustable limits for the gas measurement variables.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

qualified medical personnel / hospitalized patients

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

The Datex-Ohmeda S/5TM Compact Airway Module (model family E-CAiOVX) E-CAiOVX, E-The Datex-Onlineda S75 - Compact In-CO and accessories have been assessed against the CAIOV, L-CAIO, L-CO P.R. B OC T, B OO T, E oo oughly tested through validation and verification of specifications.

  • FDA/DCRND Reviewer Guidance for Premarket Notification Submissions, November 1993
  • I DriDCRISD 110101 1:1991 + Amdt. 2:1995 (Part 1: General requirements for safety)
  • IEC 00001-1:1990+ A1:1993 + A13:1996 + A2:1995 (identical to IEC60601-1:1988 + Amdt. . 1:1991 + Amdt. 2:1995)
  • 1:1994 Aind. 2:1993601.1-M90 + S1:1994 (Canadian deviations to IEC 60601-1:1988 + ◆ Amdt. 1:1991) + S2:1998 (=IEC Amdt 2:1995)
  • Amdt. 1:1971) SE: 27997 (U.S. deviations to IEC 60601-1:1988 + Amdt. 1:1991+ Amdt. . 2:1995)
  • IEC 60601-1-2:2001 (Electromagnetic compatibility Requirements and tests)
  • IEC 60601-1-4:2000 (Programmable medical systems) ◆
  • ISO 9918 (1993) / EN 864 (1996): Capnometers for use with Humans Requirements
  • ISO 7767:1997 Oxygen monitors for monitoring breathing mixtures.
  • EN 12598:1999 Oxygen monitors for monitoring breathing mixtures.
  • ISO 11196:1995 + Corr. 1:1997 /EN ISO 19996 (1997) Anaesthetic gas monitors ◆
  • ASTM F-1456 (2001) Standard specification for Capnometers
  • ADAM I 150 (2007) Datent of Premarket Submission for Software Contained in Medical . Devices. Version 1.0. (May, 29, 1998)
    Conclusion: The summary above shows that there are no new questions of safety and effectiveness for the The summal y a 010 client module (model family E-CAiOVX) E-CAiOVX) E-CAiOVX, E-Datex-Olmedal 8-5-COVX, E-COV, E-CO and accessories as compared to the predicate device.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K001814

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 868.1720 Oxygen gas analyzer.

(a)
Identification. An oxygen gas analyzer is a device intended to measure the concentration of oxygen in respiratory gases by techniques such as mass spectrometry, polarography, thermal conductivity, or gas chromatography. This generic type of device also includes paramagnetic analyzers.(b)
Classification. Class II (performance standards).

0

MAY 2 5 2005

KOS1092

Premarket Notification 510(k) Summary As required by section 807.92 Datex-Ohmeda S/5TM Compact Airway Module (model family E-CAIOVX) E-CAiOVX, E-CAiOV, E-CAiOV, E-CAiO, E-COVX, E-COV, E-CO and accessories.

GENERAL COMPANY INFORMATION as required by 807.92(a)(1)

COMPANY NAME/ADDRESS/PHONE/FAX:

GE Healthcare 86 Pilgrim Road Needham, MA 02492 USA Tel: 781-449-8685 Fax: 781-433-1344

NAME OF CONTACT.

Mr. Joel Kent

DATE:

April 19, 2005

DEVICE NAME as required by 807.92(a)(2)

TRADE NAME:

Datex-Ohmeda S/5™ Compact Airway Module (model family E-CAiOVX) E-CAiOVX, E-CAiOV, E-CAiO, E-COVX, E-COV, E-CO and accessories.

COMMON NAME:

Airway gas, pressure and volume, anesthetic agent and agent identification and gas exchange monitor. Airway gas and Patient Spirometry accessories.

1

CLASSIFICATION NAME:

The following Class II classifications appear applicable:

Product CodeClassification NameCFR Section
CCKAnalyzer, Gas, Carbon-Dioxide, Gaseous-phase868.1400
CCLAnalyzer, Gas, Oxygen, Gaseous-phase868.1720
BZKSpirometer, Monitoring (W/WO alarm)868.1850
CAPMonitor, Airway Pressure (Includes gauge and/or alarm)868.2600
CBRAnalyzer, Gas, Nitrous-Oxide, Gaseous-phase (Anesthetic co)868.1700
BZLComputer, Oxygen-uptake868.1730
CBQAnalyzer, Gas, Enflurane, Gaseous-phase (Anesthetic conc.)868.1500
CBSAnalyzer, Gas, Halothane Gaseous-phase (Anesthetic conc.)868.1500
NHOAnalyzer, Gas, Desflurane, Gaseous-phase (Anesthetic conc.)868.1500
NHQAnalyzer, Gas, Isoflurane Gaseous-phase (Anesthetic conc.)868.1500
NHPAnalyzer, Gas, Sevoflurane, Gaseous-phase (Anesthetic conc)868.1500

NAME OF LEGALLY MARKETED DEVICE FOR WHICH A CLAIM OF SUBSTANTIAL EQUIVALENCE IS MADE as required by 807.92(a)(3)

The Datex-Ohmeda S/5™ Compact Airway Module family, E-CAiOVX is substantially equivalent in safety and effectiveness to the legally marketed (predicate) Datex-Ohmeda M-CAiOVX Module (K001814).

DEVICE DESCRIPTION as required by 807.92(a)(4)

The Datex-Ohmeda S/5™ Compact Airway Module, E-CAiOVX is a double-width plug-in parameter module for monitoring respiratory (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiration rate), ventilatory (airway pressure, volume and flow) and gas exchange parameters (Oxygen Consumption VO2, Carbon Dioxide production VCO2, Respiratory Quotient RQ, and Energy Expenditure, EE). The E-CAiOVX is a Compact Airway measuring module for a modular monitoring system.

The intended use for the modified device is the same as for the predicate, Datex-Ohmeda M-CAiOVX module (K001814) The indications for use are also the same.

There has been no change to the basic technology from the predicate. The E-CAiOVX module is a facelifted version of the predicate M-CAiOVX module. The module cover and mechanics have changed, but the fundamental scientific technology is the same as in the predicate device M-CAiOVX (K001814). The electronic measurement boards are the same in E-CAiOVX and M-CAiOVX, except for one component type change (EEPROM), and the software is basically the same, although both the software code and revision has changed during the five years since the predicate submission. Software changes include mostly improvements simplifying manufacturing and tests during manufacturing. There are also some enhancements in the mechanics of the E-CAiOVX modulc compared with the predicate M-CAiOVX, e.g. EMC cover is better implemented into the mechanics. The EMC specifications for the E-CAiOVX and M-CAiOVX modules are the same. The Datex-Ohmeda Compact Airway Module module, E-CAiOVX can be used with the following Datex-Olımeda modular monitors with any monitor software (*)

S/5TM Anesthesia Monitor (AM) with monitor software S-STD93 or newer (*)

. S/5™ Compact Anesthesia Monitor (CAM) with monitor software S-STD93 or newer (*)

. S/5TM Critical Care Monitor (CCM) with monitor software S-ICU97 or newer (*)

S/5TM Compact Critical Care Monitor (CCCM).with monitor software S ICU97 or newer (*)

(*) The Gas exchange measurement works only with software versions 99 or newer. All monitors can be upgraded to software version 99 by using the U-LIFE U-xxx99(A).

2

The E-CAiOVX module consists of:

  • AOVA infrared measuring sensor for measuring CO2, N2O and anaesthetic agents
  • Paramagnetic O2 sensor .
  • Side-Stream Spirometry measurement .
  • Gas Exchange measurement

. Halothane, Sevoflurane, Isoflurane and Desflurane).

Harothalle, Sevonurane, Isofturane and Decritical).
The E-CAiOVX module uses the same measurement technology and accessories as the predicate I he E-CAIOVX (K001814). The main accessories include airway gas sampling lines, D-fend device, M- CAIO + A (1001011). In tubing and D-lite sensors. The sampling line and the water traps, Spironicily measurement table connectors. The monitor is switched on and the gas spirometry tube and the spirometry tube is attached to the D-lite™ airway adapter. The D-lite™ is sampling the and the spironionly tace is that and moisture exchanger (HIME) of the patient's intubation tube. The monitor displays measurements from the E-CAiOVX's and subtype modules intubation table. The moment and loops. All the calculated paraneters can be selected on in the form of numeric values, our res, and view on the E-CAiOVX moduels are taken the usplay, and action. And for wear interface for alarms in Datex-Ohmeda S/5 patient care of by the nost monitor and forlow the asser merines. There are auditory and visual alarms and user adjustable limits for the gas measurement variables.

INTENDED USE as required by 807.92(a)(5)

Intended use:

The Datex-Ohmeda S/5™ Compact Airway Module, E-CAiOVX family is intended to be used TIE Datex-Ohmeda 07-5 - Oct plant in monitors for monitoring respiratory, ventilatory and gas exchange parameters of hospitalized patients.

Indications for use:

The Datex-Ohmeda S/5M Compact Airway Module, E-CAiOVX family is indicated for The Date.x-Onlineda 075 - Comparation (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiration rate) ventilation (airway pressure, volume and flow) and gas luchange status (Oxygen Consumption VO2, Carbon Dioxide production VCO2, Respiratory Cxchange status (Oxygon Octountiples, EE). Gas exchange status monitoring is not indicated in the Quottent RQ, and Energy Expendhadevice is indicated for use by qualified medical personnel only.

3

SUMMARY OF TECHNOLOGICAL CHARACTERITICS OF DEVICE COMPARED TO THE PREDICATE DEVICE as required by 807.92(a)(6)

The Datex-Ohmeda S/5™ Compact Airway Module family, E-CAiOVX is substantially equivalent in safety and effectiveness to the legally marketed (predicate) Datex-Ohmeda M-CAiOVX Module (K001814).

The E-CAiOVX module has the following similarities compared to the predicate M-CAiOVX (K001814):

  • identical intended use and indications for use .
  • identical fundamental scientific technology .
  • same electronic measurement board (EEPROM type changed) .
  • same module software (version changed from 891977-3.2 to 8001805-4.5) .
  • same algorithms for respiratory, ventilatory and gas exchange parameter calculation .
  • use the same operating principle ●
  • use largely the same accessories ●
  • have the same user interface at the monitor and alarms (can be used with the same . monitor software)
  • the Customer and parameter specifications are the same, except for one minor . modification
  • have the same safety and effectiveness .
  • are manufactured using the same processes t

The main differences between the new E-CAiOVX and the predicate M-CAiOVX (K001814) is primarily due to fact that the new E-CAiOVX module has the following changes:

  • new color and shape and thus differing mechanics .
  • The front panel and labeling have changed .
  • The connector for gas return (front panel and accessories) have changed to be ◆ incompatible with female Luer Lock connectors.
  • The EEPROM type on the electronic measurement boards have changed .
  • The module software has been enhanced and revised several times. Most new revisions . include enhancements for manufacturing and testing during manufacturing.
  • The EMC is better implemented into the mechanics, e.g. some module materials changed . and shielding EMC foam was used.

Based on the analysis and other documentation included in this 510(k) notification and Dused on the and you and of the main features and indications for use of the Datex-Ohmeda S/STM Compact Airway Module, E-CAiOVX are substantially equivalent to the predicate Datex-Ohmeda M-CAiOVX Module (K001814).

4

SUMMARY OF NONCLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(1)(3)

The Datex-Ohmeda S/5TM Compact Airway Module (model family E-CAiOVX) E-CAiOVX, E-The Datex-Onlineda S75 - Compact In-CO and accessories have been assessed against the CAIOV, L-CAIO, L-CO P.R. B OC T, B OO T, E oo oughly tested through validation and verification of specifications.

  • FDA/DCRND Reviewer Guidance for Premarket Notification Submissions, November 1993 .
  • I DriDCRISD 110101 1:1991 + Amdt. 2:1995 (Part 1: General requirements for safety) .
  • IEC 00001-1:1990+ A1:1993 + A13:1996 + A2:1995 (identical to IEC60601-1:1988 + Amdt. . 1:1991 + Amdt. 2:1995)
  • 1:1994 Aind. 2:1993601.1-M90 + S1:1994 (Canadian deviations to IEC 60601-1:1988 + ◆ Amdt. 1:1991) + S2:1998 (=IEC Amdt 2:1995)
  • Amdt. 1:1971) SE: 27997 (U.S. deviations to IEC 60601-1:1988 + Amdt. 1:1991+ Amdt. . 2:1995)
  • IEC 60601-1-2:2001 (Electromagnetic compatibility Requirements and tests) .
  • IEC 60601-1-4:2000 (Programmable medical systems) ◆
  • ISO 9918 (1993) / EN 864 (1996): Capnometers for use with Humans Requirements .
  • ISO 7767:1997 Oxygen monitors for monitoring breathing mixtures. .
  • EN 12598:1999 Oxygen monitors for monitoring breathing mixtures. .
  • ISO 11196:1995 + Corr. 1:1997 /EN ISO 19996 (1997) Anaesthetic gas monitors ◆
  • ASTM F-1456 (2001) Standard specification for Capnometers .
  • ADAM I 150 (2007) Datent of Premarket Submission for Software Contained in Medical . Devices. Version 1.0. (May, 29, 1998)

CONCLUSION:

The summary above shows that there are no new questions of safety and effectiveness for the The summal y a 010 client module (model family E-CAiOVX) E-CAiOVX) E-CAiOVX, E-Datex-Olmedal 8-5-COVX, E-COV, E-CO and accessories as compared to the predicate device.

5

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird symbol, composed of three curved lines that suggest the shape of a bird in flight. The logo is encircled by the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 2 5 2005

Mr. Joel Kent Manager, Quality and Regulatory Affairs GE Healthcare 86 Pilgrim Road Needham, Massachusetts 02492

Re: K051092

Trade/Device Name: Datex-Ohmeda S/5™ Compact Airway Module (model family E-CAiOVX) E-CAiOVX, E-CAiOV, E-CAiOV, E-CAiO, E-COVX, E-COV, E-CO and Accessories Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: II Product Code: CCL, BZK, CAP, CBR, BZL, CBQ, CBS, NHO, NHQ, NHP Dated: April 27, 2005 Received: April 28, 2005

Dear Mr. Kent:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvin uppression (the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 Nr.), it hay of base of Federal Regulations, Title 21, Parts 800 to 898. In your device ear or roublish further announcements concerning your device in the Eederal Register.

6

Page 2 -- Mr. Kent

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. or the Fet of all the Act's requirements, including, but not limited to: registration 1 od intilet 021 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice and instills (21 er retries in the quality systems (QS) regulation (21 CFR Part 820); and if requirements as bet ronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) This tetter "Hiffication. The FDA finding of substantial equivalence of your device to a premaired predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), If you desire specifical of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the may obtain of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Chiu Lin, Ph.D.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

🚗

7

Indications for Use

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: Datex-Ohmeda S/5™ Compact Airway Module (model family E-CAiOVX) E-CAiOVX, E-CAiOV, E-CAiO, E-CAiO, E-COVX, E-COV, E-CO and accessories.

Indications for Use:

The Datex-Ohmeda S/5™ Compact Airway Module, E-CAiOVX family is indicated for monitoring hospital patient´s respiration (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiration rate) ventilation (airway pressure, volume and flow) and gas exchange status (Oxygen Consumption VO2, Carbon Dioxide production VCO2, Respiratory Quotient RQ, and Energy Expenditure, EE). Gas exchange status monitoring is not indicated in the presence of N2O+O2 mixtures. The device is indicated for use by qualified medical personnel only.

Over-The-Counter Use Prescription Use ___ × AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Cluis Ls

Page _ of

(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Device

510(k) Number: K051042