K Number
K090233

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-03-05

(34 days)

Product Code
Regulation Number
868.5160
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GE Datex-Ohmeda Aisys Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. The device is intended for volume or pressure control ventilation. The Aisys is not suitable for use in a MRI environment.

Device Description

The GE Datex-Ohmeda Aisys is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. It represents one of the systems in a long line of products based on the Datex-Ohmeda Excel, Aestiva, Aespire, and Avance Anesthesia Systems. It is to be used only by trained and qualified medical professionals.

The GE Datex-Ohmeda Aisys Carestation supplies set flows of medical gases to the breathing system using electronic gas mixing. Gas flows are selected by the user using the keypad and rotary controller on the main display unit and then displayed as electronic flow indicators on the system display unit. The Aisys is equipped with a pneumatic back-up O2 delivery system and traditional flow tube, as well. A large selection of frames, gases, and vaporizers are available to give the user control of the system configuration. The Aisys is also available in a pendant model. It is available with two or three gases, and up to three cylinder connections. All models have O2. The Aisys comes with up to two optional gases (air, N2O). Safety features and devices within the Aisys are designed to decrease the risk of hypoxic mixtures and complete power or sudden gas supply failures. The Aisys system is available with optional integrated respiratory gas monitoring. When supplied as an option, the integrated respiratory gas monitoring is provided via the Datex-Ohmeda M-Gas Module (M-CAiO and M-CAiOV software revision 3.2 and above K# 001814) and E-Gas Module (E-CAiOVX cleared via K051092) which can be physically intecrated into the Aisys, receive electronic power from the Aisys and communicate measured values to the Aisys for display on the system display unit.

The anesthetic agent delivery for the Aisys is controlled via an anesthesia computer through user input from the central display. The vaporization technology is based upon the electronic vaporizer cleared as part of the Datex-Ohmeda Anesthesia Delivery Unit (ADU) cleared via K973985. An Aladin cassette (also cleared as part of K973895) or Aladin 2 is inserted into the active cassette bay. The cassette holds the agent to be delivered - Halothane, Enflurane, Isoflurane. Desflurane or Sevoflurane. Agent is delivered as a percent volume/volume. The Aisys is designed to allow only one active cassette at a time. Per the user input into the main display, valves within the active cassette bay will open and allow agent to be delivered. The agent is mixed with gas from the FGC unit. After mixing, the combination of gases and agent is delivered to the breathing system and then onto the patient.

The Datex-Ohmeda 7900 Anesthesia Ventilator is used in the Aisys Anesthesia System. It is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during surgical procedures. The 7900 ventilator is equipped with a built-in monitoring system for inspired oxygen, airway pressure and exhaled volume. Sensors in the breathing circuit are used to control and monitor patient ventilation as well as measure inspired oxygen concentration. This allows for the compensation of compression losses, fresh gas contribution and small leakage in the breathing absorber, bellows and system. User setting and microprocessor calculations control breathing patterns. The user interface keeps settings in memory. The user may change settings with a simple and secure setting sequence. A bellows contains breathing gasses to be delivered to the patient. Positive End Expiratory Pressure (PEEP) is requlated electronically. Positive pressure is maintained in the breathing system so that any leakage that occurs is outward. An RS-232 serial digital communications port connects to and communicates with external devices. Ventilator modes for the device include Volume Mode, Pressure Control Mode, Pressure Support with Apnea Backup Mode (Optional) and Synchronized Intermittent Mandatory Ventilation (SIMV) Mode (Optional). Ventilator parameters and measurements are displayed on the system display unit.

The system display unit is mounted to an arm on the top shelf of the Aisys. The arm is counter balanced and capable of moving vertically and/or horizontally, and also tilting the display, enabling the user to position the display to the most advantageous viewing position. The arm length is limited such that the display position is always within the footprint of the Aisys frame. The arm also supports the mounting of additional display units for a variety of patient monitors.

Several frame configurations are available, including one that allows for the physical integration of the Datex-Ohmeda S/5 Anesthesia Monitor (most recently cleared via K030812). This configuration also provides cable management solutions such that the necessary connections from the monitor display unit to the monitor are hidden within the Aisys frame. An additional ootion allows the S/5 AM to be linked to the power supply of the Aisys such that when the Aisys is turned on, the S/5 AM is also turned on. Additional configurations allow for the mounting of various patient monitors on the top shelf of the Aisys.

AI/ML Overview

The provided text describes the 510(k) premarket notification for the GE Datex-Ohmeda Aisys Anesthesia System. The key information regarding acceptance criteria and study details is presented in the "SUMMARY OF NONCLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(1)(3)" and "SUMMARY OF CLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(2)" sections.

Here's the breakdown of the information requested:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the GE Datex-Ohmeda Aisys Anesthesia System are compliance with various international and national standards related to medical electrical equipment, anesthesia workstations, and gas cylinders. The reported device performance is that it meets these standards.

Acceptance Criteria (Standard)Reported Device Performance
EN 740 :1998 (Anesthesia Workstations and their components)Compliant
EN 60601-1:1990 (Medical Electrical Equipment Part 1: General Requirements for Safety)Compliant
EN 60601-1-1:2000 (Medical Electrical Equipment - Medical Electrical Systems)Compliant
EN 60601-1-2:2001 (Medical Electrical Equipment - Electromagnetic Compatibility)Compliant
IEC 60601-1-4:2000 (Safety of Programmable Electronic Medical Systems)Compliant
EN 475:1995 (Electrically Generated Alarm Signals)Compliant
EN 850:1997 (Small Medical Gas Cylinders - Pin Indexed)Compliant
EN 980:1997 (Graphical Symbols)Compliant
EN 1041:1998 (Information to be supplied with medical devices)Compliant
EN 1089-3:1997 (Color coding for medical gases)Compliant
ISO5356-1:1996 (Conical Connectors)Compliant
EN 1820:1997 (Reservoir Bags)Compliant
IEC 60601-2-13:1998 (Particular requirements for the safety of anaesthetic workstations)Compliant
Software ValidationCompleted (thoroughly tested through verification and validation)
Verification of SpecificationsCompleted (thoroughly tested)

2. Sample Size for the Test Set and Data Provenance

The document does not specify a "test set" in the context of a clinical study with a defined sample size for evaluating performance against an explicit statistical endpoint. Instead, the evaluation focuses on non-clinical testing, including verification of specifications and software validation, and compliance with established international standards.

The term "data provenance" is not applicable in the traditional sense of clinical data here, as the testing described is primarily engineering and regulatory compliance testing rather than clinical study data from patients.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This information is not applicable as the evaluation relies on compliance with technical standards and internal verification/validation processes rather than expert-established ground truth from clinical cases. No external "experts" in the context of clinical ground truth determination are mentioned for the non-clinical testing.

4. Adjudication Method for the Test Set

This information is not applicable as there is no mention of a test set requiring adjudication in the context of human expert review. The evaluation is based on meeting technical specifications and standards.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

An MRMC comparative effectiveness study was not performed, and is not applicable. The device is an anesthesia system, not an AI-assisted diagnostic tool for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The device itself is a standalone anesthesia system. While it has embedded software and automated functions, the evaluation is of the complete system's compliance with safety and functional standards, not a standalone algorithm's performance in isolation from the hardware or human operator interaction. The "software validation" performed addresses the functionality of the algorithms within the system.

7. The type of ground truth used

The "ground truth" for this device's evaluation is primarily compliance with established international and national engineering, safety, and performance standards (e.g., EN, IEC, ISO standards) and internal product specifications. For the software, it's defined by the software requirements and design specifications.

8. The sample size for the training set

This information is not applicable. The device is an anesthesia system, and its development and testing do not involve a "training set" in the context of machine learning model development. The software development follows traditional engineering verification and validation processes.

9. How the ground truth for the training set was established

This information is not applicable as there is no "training set" in the context of AI/machine learning. The "ground truth" for the device's functionality is established by its design specifications and the requirements of the relevant industry standards.

{0}------------------------------------------------

K090233

5 2009 MAR

Premarket Notification 510(k) Summary As required by section 807.92 GE Datex-Ohmeda Aisys

GENERAL COMPANY INFORMATION as required by 807.92(a){1}

COMPANY NAME/ADDRESS/PHONE/FAX:

Datex-Ohmeda, Inc. PO Box 7550 Madison, WI 53707 USA Tel: 608-221-1551 Fax: 608-223-2476

NAME OF CONTACT:

Mr. James P. Raskob Ms. Adrienne Lenz, RAC (alternate)

DATE:

March 4, 2009

DEVICE NAME as required by 807.92(a)(2)

TRADE NAME:

GE Datex-Ohmeda Aisys Anesthesia System

COMMON NAME:

Gas Machine, Anesthesia

CLASSIFICATION NAME:

Anesthesiology, 73 BSZ, 21 CFR 868.5160 Gas Machine, Anesthesia

NAME OF LEGALLY MARKETED DEVICE FOR WHICH A CLAIM OF SUBSTANTIAL EQUIVALENCE IS MADE as required by 807.92(a)(3)

The GE Datex-Ohmeda Aisys is substantially equivalent in safety and effectiveness to the legally marketed (predicate) GE Datex-Ohmeda Avance (K081844) and the GE Datex-Ohmeda Aisys (K073707).

{1}------------------------------------------------

DEVICE DESCRIPTION as required by 807.92(a)(4)

The GE Datex-Ohmeda Aisys is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. It represents one of the systems in a long line of products based on the Datex-Ohmeda Excel, Aestiva, Aespire, and Avance Anesthesia Systems. It is to be used only by trained and qualified medical professionals.

The GE Datex-Ohmeda Aisys Carestation supplies set flows of medical gases to the breathing system using electronic gas mixing. Gas flows are selected by the user using the keypad and rotary controller on the main display unit and then displayed as electronic flow indicators on the system display unit. The Aisys is equipped with a pneumatic back-up O2 delivery system and traditional flow tube, as well. A large selection of frames, gases, and vaporizers are available to give the user control of the system configuration. The Aisys is also available in a pendant model. It is available with two or three gases, and up to three cylinder connections. All models have O2. The Aisys comes with up to two optional gases (air, N2O). Safety features and devices within the Aisys are designed to decrease the risk of hypoxic mixtures and complete power or sudden gas supply failures. The Aisys system is available with optional integrated respiratory gas monitoring. When supplied as an option, the integrated respiratory gas monitoring is provided via the Datex-Ohmeda M-Gas Module (M-CAiO and M-CAiOV software revision 3.2 and above K# 001814) and E-Gas Module (E-CAiOVX cleared via K051092) which can be physically intecrated into the Aisys, receive electronic power from the Aisys and communicate measured values to the Aisys for display on the system display unit.

The anesthetic agent delivery for the Aisys is controlled via an anesthesia computer through user input from the central display. The vaporization technology is based upon the electronic vaporizer cleared as part of the Datex-Ohmeda Anesthesia Delivery Unit (ADU) cleared via K973985. An Aladin cassette (also cleared as part of K973895) or Aladin 2 is inserted into the active cassette bay. The cassette holds the agent to be delivered - Halothane, Enflurane, Isoflurane. Desflurane or Sevoflurane. Agent is delivered as a percent volume/volume. The Aisys is designed to allow only one active cassette at a time. Per the user input into the main display, valves within the active cassette bay will open and allow agent to be delivered. The agent is mixed with gas from the FGC unit. After mixing, the combination of gases and agent is delivered to the breathing system and then onto the patient.

The Datex-Ohmeda 7900 Anesthesia Ventilator is used in the Aisys Anesthesia System. It is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during surgical procedures. The 7900 ventilator is equipped with a built-in monitoring system for inspired oxygen, airway pressure and exhaled volume. Sensors in the breathing circuit are used to control and monitor patient ventilation as well as measure inspired oxygen concentration. This allows for the compensation of compression losses, fresh gas contribution and small leakage in the breathing absorber, bellows and system. User setting and microprocessor calculations control breathing patterns. The user interface keeps settings in memory. The user may change settings with a simple and secure setting sequence. A bellows contains breathing gasses to be delivered to the patient. Positive End Expiratory Pressure (PEEP) is requlated electronically. Positive pressure is maintained in the breathing system so that any leakage that occurs is outward. An RS-232 serial digital communications port connects to and communicates with external devices. Ventilator modes for the device include Volume Mode, Pressure Control Mode, Pressure Support with Apnea Backup Mode (Optional) and Synchronized Intermittent Mandatory Ventilation (SIMV) Mode (Optional). Ventilator parameters and measurements are displayed on the system display unit.

The system display unit is mounted to an arm on the top shelf of the Aisys. The arm is counter balanced and capable of moving vertically and/or horizontally, and also tilting the display, enabling the user to position the display to the most advantageous viewing position. The arm length is limited such that the display position is always within the footprint of the Aisys frame. The arm also supports the mounting of additional display units for a variety of patient monitors.

{2}------------------------------------------------

Several frame configurations are available, including one that allows for the physical integration of the Datex-Ohmeda S/5 Anesthesia Monitor (most recently cleared via K030812). This configuration also provides cable management solutions such that the necessary connections from the monitor display unit to the monitor are hidden within the Aisys frame. An additional ootion allows the S/5 AM to be linked to the power supply of the Aisys such that when the Aisys is turned on, the S/5 AM is also turned on. Additional configurations allow for the mounting of various patient monitors on the top shelf of the Aisys.

INTENDED USE as required by 807.92(a)(5)

The GE Datex-Ohmeda Aisys Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. The device is intended for volume or pressure control ventilation. The Aisys is not suitable for use in a MRI environment.

SUMMARY OF TECHNOLOGICAL CHARACTERISTICS OF DEVICE COMPARED TO THE PREDICATE DEVICE as required by 807.92(a)(6)

The GE Datex-Ohmeda Aisys has been updated from the predicate version (K073707). There have been no changes to the intended use or fundamental scientific technology.

The software for the Aisys has been updated to introduce several new features. A summary of the changes to the software include:

  • Alarm Improvements . Improvements are made to the alarms to reduce nuisance alarms. ◆ These changes are the same as have been implemented in Avance 6.0 (K081844). The Vaporizer failure hardware alarm was improved to match the functionality of the existing Aisys Mixer hardware alarm. An audio alarm was added to the current Alternate O2 Button Pressed video message.
  • User Configurations. Addition of more default configurations. .
  • Usability Improvements Waves/Pages. The system will have more screen configurability . for waveforms and the numerics.
  • . Checkout Improvements. Low pressure leak check, breathing circuit leak check, system check flow sensor and agent delivery checkout tests were enhanced.
  • MAC Age. The mathematical calculation for MAC will now be able to be made ◆ accounting for the patient age. This is the same as is displayed on the S/5 Anesthesia Monitor (K030812, K051400).
  • . Vaporizer Settings. To aid the filling process the system defaults to automatically depressurize the vaporizer cassette when the agent is turned off.
  • Preset ventilator settings based on patient weight for PCV-VG mode. The preset . ventilator settings at the start of a case can now be based on patient weight for PCV-VG mode, similar to the way preset vent settings can be been based on patient weight for VCV mode in the previous Aisys.
  • . Auto exit PSVPro backup mode. The system will now automatically exit SIMV-PC vent mode upon recognition of spontaneous breathing and resume PSVPro™ when patient resumes spontaneous breathing (change was previously made manually).
  • . Calibrate flow sensors reminder. An informational alarm message will appear when the system is not in therapy between cases to remind the user to calibrate the flow sensors if flows sensors have not been calibrated for 24 hours. Currently the manuals give this instruction.

{3}------------------------------------------------

SUMMARY OF NONCLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(1)(3)

The GE Datex-Ohmeda Aisys has been thoroughly tested through verification of specifications and validation, including software validation. Verification of compliance with the following standards has also been made to support safe use of the device in its intended environment.

StandardTitle
EN 740 :1998Anesthesia Workstations and their components
EN 60601-1:1990Medical Electrical Equipment Part 1: General Requirements for SafetyIncorporates Corrigendum July 1994; Includes Amendments A1:1993,A11:1993, A12:1993, A2:1995, A13:1996, IEC 601-1:1998 + A2:1995+ Corrigendum 1995, Modified
EN 60601-1-1:2000Medical Electrical Equipment - Medical Electrical Systems
EN 60601-1-2:2001Medical Electrical Equipment - Electromagnetic Compatibility
IEC 60601-1-4:2000Safety of Programmable Electronic Medical Systems
EN 475:1995Electrically Generated Alarm Signals
EN 850:1997Small Medical Gas Cylinders - Pin Indexed
EN 980:1997Graphical Symbols
EN 1041:1998Information to be supplied with medical devices
EN 1089-3:1997Color coding for medical gases
ISO5356-1:1996Conical Connectors
EN 1820:1997Reservoir Bags
IEC 60601-2-13:1998Particular requirements for the safety of anaesthetic workstations

SUMMARY OF CLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(2)

The modifications made to the GE Datex-Ohmeda Aisys did not require clinical testing.

CONCLUSION:

The summary above shows that there are no new questions of safety and effectiveness for the GE Datex-Ohmeda Aisys as compared to the predicate device.

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the three levels of government: federal, state, and local. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the eagle.

MAR 5 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. James Raskob Regulatory Affairs Leader Datex-Ohmeda, Incorporated Life Support Solutions 3030 Ohmeda Drive P.O. Box 7550 Madison, Wisconsin 53707

Re: K090233

Trade/Device Name: GE Datex-Ohmeda Aisys Anesthesia System Regulation Number: 21 CFR 868.5160 Regulation Name: Gas Machine for Anesthesia or Analgesia Regulatory Class: II Product Code: BSZ Dated: January 29, 2009 Received: February 3, 2009

Dear Mr. Raskob:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Raskob

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Siste H. Mcchan Oms.

Ginette Y. Michaud, M.D. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known): K

Device Name: GE Datex-Ohmeda Aisys Anesthesia System

Indications For Use:

The GE Datex-Ohmeda Aisys Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. The device is intended for volume or pressure control ventilation. The Aisys is not suitable for use in a MRI environment.

Prescription Use XXX (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Susan Reinger

(Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K090233

§ 868.5160 Gas machine for anesthesia or analgesia.

(a)
Gas machine for anesthesia —(1)Identification. A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The device may include a gas flowmeter, vaporizer, ventilator, breathing circuit with bag, and emergency air supply.(2)
Classification. Class II (performance standards).(b)
Gas machine for analgesia —(1)Identification. A gas machine for analgesia is a device used to administer to a patient an analgesic agent, such as a nitrous oxide-oxygen mixture (maximum concentration of 70 percent nitrous oxide).(2)
Classification. Class II (performance standards).