K Number
K123125

Validate with FDA (Live)

Manufacturer
Date Cleared
2013-02-06

(125 days)

Product Code
Regulation Number
868.5160
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GE Datex-Ohmeda Avance CS2 Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). The device is intended for volume or pressure control ventilation.

Device Description

The GE Datex-Ohmeda Avance and Avance CS anesthesia machines are intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). They represent one of the systems in a long line of products based on the Datex-Ohmeda Excel, Aestiva, and Aespire Anesthesia Systems. Avance systems are to be used only by trained and qualified medical professionals trained in the administration of general anesthesia.

The Avance and Avance CS2 anesthesia systems supply set flows of medical gases to the breathing system using electronic gas mixing. Gas flows are selected by the user using the electronic controls on the main display unit and then displayed as electronic flow meters on the system display unit. The Avance systems are equipped with a pneumatic back-up O2 delivery system and traditional flow tube, as well. A large selection of frames options including central brake or individual caster brakes, gases, and vaporizers are available to give the user control of the system configuration. The Avance systems are also available in pendant models. Avance systems are available with two or three gases, up to three vaporizer positions and up to three cylinder connections. All models have O2. The Avance systems come with up to two optional gases (air, N2O).

The Avance systems accept Tec 6 Plus and Tec 7 vaporizers on a Selectatec manifold. Safety features and devices within the Avance systems are designed to decrease the risk of hypoxic mixtures, agent mixtures and complete power or sudden gas supply failures. The Avance systems are available with optional integrated respiratory gas monitoring. When supplied as an option, the integrated respiratory gas monitoring is provided via the GE Compact Gas Airway Modules Series: M-Gas Module (M-CAiO and M-CAiOV software revision 3.2 and above cleared via K001814) and E-Gas Compact Gas Airway Module (E-CAiOVX software revision 3.2 and above cleared via K051092) which can be physically integrated into the Avance, receive electronic power from the Avance and communicate measured values to the Avance for display on the system display unit.

The Datex-Ohmeda 7900 Anesthesia Ventilator is used in the Avance Anesthesia Systems. It is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during surgical procedures. The 7900 ventilator is equipped with a built-in monitoring system for inspired oxygen, airway pressure and exhaled volume. Sensors in the breathing circuit are used to control and monitor patient ventilation as well as measure inspired oxygen concentration. This allows for the compensation of compression losses, fresh gas contribution and small leakage in the breathing absorber, bellows and system. User setting and microprocessor calculations control breathing patterns. The user interface keeps settings in memory. The user may change settings with a simple setting sequence. A bellows contains breathing gasses to be delivered to the patient. Positive End Expiratory Pressure (PEEP) is regulated electronically. Positive pressure is maintained in the breathing system so that any leakage that occurs is outward. An RS-232 serial digital communications port connects to and communicates with external devices. Ventilator modes for the device include Volume Control Ventilation (VCV), Pressure Control Ventilation (PCV) (optional), Synchronized Intermittent Mandatory Ventilation/Pressure Support (SIMV/PSV) (optional), Pressure Support Ventilation (PSVPro) (optional), Synchronized Intermittent Mandatory Ventilation-Pressure Control (SIMV-PC) (optional), Pressure Control Ventilation-Volume Guaranteed (PCV-VG) (optional), Constant Positive Airway Pressure/Pressure Support Ventilation (CPAP/PSV), Pressure Control Ventilation-Volume Guaranteed with Pressure Support Ventilation (SIMV PCV-VG) (optional), and Volume Control Ventilation Mode for Cardiac Bypass Mode (optional). Ventilator parameters and measurements are displayed on the system display unit.

Several frame configurations are available, including one that allows for the physical integration of the GE Monitor Series(most recently cleared Carescape B850 via K092027 and B650 cleared on K102239). These configurations also provide cable management solutions such that the necessary connections from the monitor display unit to the monitor are hidden within the Avance frame. Additional configurations allow for the mounting of various patient monitors on the top shelf of the Avance.

AI/ML Overview

Here's an analysis of the provided text regarding acceptance criteria and supporting studies for the GE Datex-Ohmeda Avance CS2 Anesthesia System:

Based on the provided document, the application is for a 510(k) premarket notification, which focuses on demonstrating substantial equivalence to a predicate device rather than undergoing new clinical trials for efficacy. Therefore, much of the information typically requested for acceptance criteria and a deep dive into study design for novel devices is not present or not applicable in this context.

The document explicitly states: "The modifications made to the GE Datex-Ohmeda Avance to create the GE Datex-Ohmeda Avance CS2 did not require clinical testing. The functionality of the modified features was completely evaluated by performing nonclinical tests of design verification and validation testing."

This immediately tells us that there was no standalone clinical study with human patients to prove device performance against specific acceptance criteria in the way one might expect for a new, high-risk device.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and the Reported Device Performance

Given the nature of a 510(k) for an updated device, the "acceptance criteria" are primarily related to meeting specifications, voluntary standards, and demonstrating that the new features do not negatively impact safety and effectiveness compared to the predicate device. There isn't a table of specific clinical performance metrics (e.g., accuracy of a diagnostic feature, rates of successful treatment) with corresponding target values and achieved results reported in this summary.

Acceptance Criteria (Implied from 510(k) process and text):

Acceptance Criteria CategoryDescription (Implied)Reported Device Performance
Substantial EquivalenceDemonstrated that the Avance CS2 is as safe, as effective, and has performance substantially equivalent to the predicate device (GE Datex-Ohmeda Avance, K112722).Met: The FDA's 510(k) clearance (K123125) confirms substantial equivalence.
Specifications ComplianceDevice functions (new and existing) meet their defined technical specifications.Met: "thoroughly tested through verification of specifications and validation"
Voluntary StandardsCompliance with applicable voluntary standards for anesthesia systems.Met: "Verification of compliance with applicable voluntary standards has also been made"
Risk MitigationIdentified and mitigated risks associated with the device.Met: "Risk Analysis" was applied during development.
Software ValidationSoftware components are thoroughly validated.Met: "including software validation"
Intended UseThe device performs as intended for general inhalation anesthesia and ventilatory support for a wide range of patients.Met: The device is cleared for its stated indications for use.

2. Sample Size Used for the Test Set and the Data Provenance

  • Test Set Sample Size: Not applicable in the context of human clinical data for this specific device. The testing involved "nonclinical tests of design verification and validation testing," including unit-level testing, integration testing, performance testing, safety testing, and simulated use testing. These are not typically quantified with "sample sizes" in the same way as human subject studies; rather, they involve testing various components, configurations, and scenarios.
  • Data Provenance: Not applicable as no clinical data from patients was used. The data provenance would be from internal engineering and quality testing (e.g., lab test data, simulated environment data).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

  • Not applicable. Since no clinical testing was performed and no ground truth was established from patient data, there's no mention of experts establishing a ground truth for a test set in a diagnostic or outcome-based scenario. The "ground truth" for nonclinical testing would be engineering specifications and functional requirements.

4. Adjudication Method for the Test Set

  • Not applicable. There was no clinical test set requiring adjudication in the context of human data. Adjudication methods (like 2+1, 3+1) are typically used for resolving disagreements among human readers/experts in interpreting clinical data.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This device is an anesthesia system, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance would not be relevant and was not performed.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

  • Not applicable. This is an anesthesia machine, not an algorithm-only device. Its performance is inherent in its hardware, software, and integrated components, designed to operate with a medical professional as the "human in the loop" administering anesthesia.

7. The Type of Ground Truth Used

  • For Nonclinical Testing: The ground truth would be based on engineering specifications, design requirements, and established performance benchmarks derived from the predicate device and relevant industry standards.
  • For Substantial Equivalence: The ultimate "ground truth" for the 510(k) process is the safety and effectiveness profile of the predicate device (GE Datex-Ohmeda Avance, K112722). The new device must demonstrate it meets this same standard.

8. The Sample Size for the Training Set

  • Not applicable. Given the device type and the absence of clinical studies, there is no mention of a "training set" in the context of machine learning. The testing performed (verification and validation) is based on engineering principles and quality assurance, not statistical training sets.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable, as there was no machine learning training set mentioned.

{0}------------------------------------------------

K) 2 3125

Page 1 of 4

FEB 6 2013

Premarket Notification 510(k) Summary As required by section 807.92 GE Datex-Ohmeda Avance CS2

GENERAL COMPANY INFORMATION as required by 807.92(a)(1)

COMPANY NAME/ADDRESS/PHONE/FAX:

Datex-Ohmeda Inc. 3030 Ohmeda Drive PO Box 7550 Madison, WI 53707 USA Tel: 608-221-1551 Fax: 608-299-2132

NAME OF CONTACT:

Mr. James P. Raskob Ms. Monica Morrison (alternate)

DATE:

October 1, 2012

DEVICE NAME as required by 807.92(a)(2)

TRADE NAME:

GE Datex-Ohmeda Avance CS2 Anesthesia System

COMMON NAME:

Gas Machine, Anesthesia

CLASSIFICATION NAME:

Anesthesiology, 73 BSZ, 21 CFR 868.5160 Gas Machine, Anesthesia

NAME OF LEGALLY MARKETED DEVICE FOR WHICH A CLAIM OF SUBSTANTIAL EQUIVALENCE IS MADE as required by 807.92(a)(3)

The GE Datex-Ohmeda Avance CS2 is substantially equivalent in safety and effectiveness to the legally marketed (predicate) GE Datex-Ohmeda Avance (K112722).

{1}------------------------------------------------

DEVICE DESCRIPTION as required by 807.92(a)(4)

The GE Datex-Ohmeda Avance and Avance CS anesthesia machines are intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). They represent one of the systems in a long line of products based on the Datex-Ohmeda Excel, Aestiva, and Aespire Anesthesia Systems. Avance systems are to be used only by trained and qualified medical professionals trained in the administration of general anesthesia.

The Avance and Avance CS2 anesthesia systems supply set flows of medical gases to the breathing system using electronic gas mixing. Gas flows are selected by the user using the electronic controls on the main display unit and then displayed as electronic flow meters on the system display unit. The Avance systems are equipped with a pneumatic back-up O2 delivery system and traditional flow tube, as well. A large selection of frames options including central brake or individual caster brakes, gases, and vaporizers are available to give the user control of the system configuration. The Avance systems are also available in pendant models. Avance systems are available with two or three gases, up to three vaporizer positions and up to three cylinder connections. All models have O2. The Avance systems come with up to two optional gases (air, N2O).

The Avance systems accept Tec 6 Plus and Tec 7 vaporizers on a Selectatec manifold. Safety features and devices within the Avance systems are designed to decrease the risk of hypoxic mixtures, agent mixtures and complete power or sudden gas supply failures. The Avance systems are available with optional integrated respiratory gas monitoring. When supplied as an option, the integrated respiratory gas monitoring is provided via the GE Compact Gas Airway Modules Series: M-Gas Module (M-CAiO and M-CAiOV software revision 3.2 and above cleared via K001814) and E-Gas Compact Gas Airway Module (E-CAiOVX software revision 3.2 and above cleared via K051092) which can be physically integrated into the Avance, receive electronic power from the Avance and communicate measured values to the Avance for display on the system display unit.

The Datex-Ohmeda 7900 Anesthesia Ventilator is used in the Avance Anesthesia Systems. It is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during surgical procedures. The 7900 ventilator is equipped with a built-in monitoring system for inspired oxygen, airway pressure and exhaled volume. Sensors in the breathing circuit are used to control and monitor patient ventilation as well as measure inspired oxygen concentration. This allows for the compensation of compression losses, fresh gas contribution and small leakage in the breathing absorber, bellows and system. User setting and microprocessor calculations control breathing patterns. The user interface keeps settings in memory. The user may change settings with a simple setting sequence. A bellows contains breathing gasses to be delivered to the patient. Positive End Expiratory Pressure (PEEP) is regulated electronically. Positive pressure is maintained in the breathing system so that any leakage that occurs is outward. An RS-232 serial digital communications port connects to and communicates with external devices. Ventilator modes for the device include Volume Control Ventilation (VCV), Pressure Control Ventilation (PCV) (optional), Synchronized Intermittent Mandatory Ventilation/Pressure Support (SIMV/PSV) (optional), Pressure Support Ventilation (PSVPro) (optional), Synchronized Intermittent Mandatory Ventilation-Pressure Control (SIMV-PC) (optional), Pressure Control Ventilation-Volume Guaranteed (PCV-VG) (optional), Constant Positive Airway Pressure/Pressure Support Ventilation (CPAP/PSV), Pressure Control Ventilation-Volume Guaranteed with Pressure Support Ventilation (SIMV PCV-VG) (optional), and Volume Control Ventilation Mode for Cardiac Bypass Mode (optional). Ventilator parameters and measurements are displayed on the system display unit.

Several frame configurations are available, including one that allows for the physical integration of the GE Monitor Series(most recently cleared Carescape B850 via K092027 and B650 cleared on K102239). These configurations also provide cable management solutions such that the necessary connections from the monitor display unit to the monitor are hidden within the Avance

{2}------------------------------------------------

frame. Additional configurations allow for the mounting of various patient monitors on the top shelf of the Avance.

INTENDED USE as required by 807.92(a)(5)

The GE Datex-Ohmeda Avance CS2 is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). The device is intended for volume or pressure control ventilation.

SUMMARY OF TECHNOLOGICAL CHARACTERISTICS OF DEVICE COMPARED TO THE PREDICATE DEVICE as required by 807.92(a)(6)

The GE Datex-Ohmeda Avance has been updated from the predicate version (K112722) to create the GE Datex-Ohmeda Avance CS2. There has been no change to the indications for use or the fundamental scientific technology.

The Avance CS2 introduces several new features. The primary new features include the following items.

Display Unit

  • 15" Display with integrated touchscreen .
  • Revised user interface with flattened menu structure and touchscreen support
  • Supports VGA output of the display image

Software Features

  • "ecoFLOW" display option .
  • Quick Picks, for settings adjustments .
  • Additional Ventilation mode PCV-VG mode with pressure support (SIMV PCV-VG) .
  • Improvements in screen configurability / customization .

Hardware Features

  • Revised color scheme .
  • Supports an optional third vaporizer .
  • Central brake for locking casters �
  • Arm design that supports display movement (swivel, tilting, movement outside frame)
  • Improved work surface lighting with variable settings
  • Additional 3rd cylinder gas option

SUMMARY OF NONCLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as reguired by 807.92(b)(1)(3)

The GE Datex-Ohmeda Avance CS2 has been thoroughly tested through verification of specifications and validation, including software validation. Verification of compliance with applicable voluntary standards has also been made to support safe use of the device in its intended environment. The following quality assurance measures were applied to the development of the system:

  • Risk Analysis .
  • Requirements Reviews .
  • Design Reviews ●
  • Testing on unit level (Module verification) ●

{3}------------------------------------------------

  • Integration testing (System verification)
  • Performance testing (Verification)
  • Safety testing (Verification)
  • Simulated use testing (Validation)

SUMMARY OF CLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as reguired by 807 92(b)(2)

The modifications made to the GE Datex-Ohmeda Avance to create the GE Datex-Ohmeda Avance CS2 did not require clinical testing. The functionality of the modified features was completely evaluated by performing nonclinical tests of design verification and validation testing.

CONCLUSION:

GE Healthcare considers the GE Datex-Ohmeda Avance CS2 to be as safe, as effective, and have performance substantially equivalent to the predicate device.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of a human figure embracing a globe, composed of three curved lines.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 6, 2013

Mr. James P. Raskob Regulatory Affairs Leader Datex-Ohmeda, Incorporated 3030 Ohmeda Drive PO Box 7550 MADISON WI 53707-7550

Re: K123125

Trade/Device Name: GE Datex-Ohmeda Avance CS4 Anesthesia System Regulation Number: 21 CFR 868.5160 Regulation Name: Gas Machine for Anesthesia or Analgesia Regulatory Class: II Product Code: BSZ Dated: January 7, 2013 Received: January 9, 2013

Dear Mr. Raskob:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Ms. Raskob

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Cintron de m

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known): |( | 2 S | 2 S | 2 S

Device Name: GE Datex-Ohmeda Avance CS2 Anesthesia System

Indications For Use:

The GE Datex-Ohmeda Avance CS2 Anesthesia System is intended to provide The GE Datex-Offineda Avance O Alleonoon of the a wide range of patients
general inhalation anesthesia and ventilatory support a wide range of patients general inhalation anesthesia and ventilatory oupport on more or pressure control ventilation.

Prescription Use _XXX (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of __ 1_

(Division Sign-Off)Division of Anesthesiology, General HospitalInfection Control, Dental Devices

Albert E. MoyalAlbert E. Moyal
2013.02.04 13:02:54-05'00'
for LS

510(k) Number: K123125

§ 868.5160 Gas machine for anesthesia or analgesia.

(a)
Gas machine for anesthesia —(1)Identification. A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The device may include a gas flowmeter, vaporizer, ventilator, breathing circuit with bag, and emergency air supply.(2)
Classification. Class II (performance standards).(b)
Gas machine for analgesia —(1)Identification. A gas machine for analgesia is a device used to administer to a patient an analgesic agent, such as a nitrous oxide-oxygen mixture (maximum concentration of 70 percent nitrous oxide).(2)
Classification. Class II (performance standards).