Search Filters

Search Results

Found 208 results

510(k) Data Aggregation

    K Number
    K250943
    Date Cleared
    2025-06-25

    (89 days)

    Product Code
    Regulation Number
    864.5220
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    GKZ

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The XR-Series module (XR-10) is a quantitative multi-parameter automated hematology analyzer intended for in vitro diagnostic use in screening patient populations found in clinical laboratories.

    The XR-Series module classifies and enumerates the following parameters in whole blood: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT (PLT-I, PLT-F), NEUT%/#, LYMPH%/#, MONO%/#, EO%/#, BASO%/#, IG%/#, RDW-CV, RDW-SD, MPV, NRBC%/#, RET%/#, IPF, IPF#, IRF, RET-He and has a Body Fluid mode for body fluids. The Body Fluid mode enumerates the WBC-BF, RBC-BF, MN%/#, PMN%/#, and TC-BF# parameters in cerebrospinal fluid (CSF), serous fluids (peritoneal, pleural) and synovial fluids. Whole blood should be collected in K2EDTA or K3EDTA anticoagulant, and serous and synovial fluids in K2EDTA anticoagulant to prevent clotting of fluid. The use of anticoagulants with CSF specimens is neither required nor recommended.

    Device Description

    The Sysmex XR-Series module (XR-10) is a quantitative multi-parameter hematology analyzer intended to perform tests on whole blood samples collected in K2 or K3EDTA and body fluids (pleural, peritoneal and synovial) collected in K2EDTA anticoagulant. The analyzers can also perform tests on CSF, which should not be collected in any anticoagulant. The XR-Series analyzer consist of four principal units: (1) One Main Units (XR-10) which aspirates, dilutes, mixes, and analyzes blood and body fluid samples; (2) Two Auto Sampler Units (SA-10, SA-01) which supply samples to the Main Unit automatically; (3) IPU (Information Processing Unit) which processes data from the Main Unit and provides the operator interface with the system; (4) Pneumatic Unit which supplies pressure and vacuum from the Main Unit.

    AI/ML Overview

    This document describes the acceptance criteria and the studies conducted to prove that the Sysmex XR-Series (XR-10) Automated Hematology Analyzer meets these criteria, demonstrating substantial equivalence to its predicate device, the Sysmex XN-20.

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA 510(k) clearance letter does not explicitly present a neatly formatted table of acceptance criteria alongside the reported performance for all parameters. Instead, it describes various performance studies (Precision, Linearity, Analytical Specificity/Interferences, Sample Stability, Detection Limit, Carry-Over, Comparison Studies, Matrix Studies, Bridging Studies, Clinical Studies, and Expected Values/Reference Range) and states that the XR-10 met "manufacturer's specifications or predefined acceptance criteria requirements" for each.

    However, based on the provided data, we can infer some general acceptance criteria, particularly from the method comparison study which uses correlation coefficient (r) and percent bias (%Bias) as metrics. The clinical sensitivity and specificity tables also present implied acceptance criteria based on the demonstrated performance.

    Inferred Acceptance Criteria & Reported Performance (Selection of Key Metrics)

    Study Type / Parameter CategoryAcceptance Criteria (Inferred)Reported Device Performance (Summary from text)
    Whole Blood Precision (Analyte-specific %CV)Met manufacturer's specifications or predefined acceptance criteria requirements.WBC: 0.30% to 2.76% CV (Repeatability); 0.97% to 1.98% (Reproducibility, within run)
    RBC: 0.45% to 0.97% CV (Repeatability); 0.73% to 1.03% (Reproducibility, within run)
    HGB: 0.38% to 0.79% CV (Repeatability); 0.40% to 0.98% (Reproducibility, within run)
    PLT-I: 1.30% to 8.32% CV (Repeatability); 1.59% to 3.70% (Reproducibility, within run)
    Body Fluid Precision (Analyte-specific %CV)Met manufacturer's specifications or predefined acceptance criteria requirements.WBC-BF: 2.01% to 3.91% CV (Repeatability); 2.01% to 3.91% (Reproducibility, within run)
    RBC-BF: 1.87% to 3.49% CV (Repeatability); 1.87% to 3.49% (Reproducibility, within run)
    Linearity (Whole Blood & Body Fluid)Linear from lower limit to upper limit and within measured maximum allowable deviation from linearity for each interval. (All results met predefined acceptance criteria).WBC (WB): 0.03 – 440.00 x10³/μL
    RBC (WB): 0.01 – 8.60 x10⁶/μL
    HGB (WB): 0.1 – 26.0 g/dL
    PLT (WB): 2 – 5,000 x10³/μL
    WBC-BF: 0.003 – 10.000 x10³/μL
    Method Comparison (Whole Blood: r-value)≥0.95 (explicitly stated for HGB, implied for others)WBC: 0.9997
    RBC: 0.9900
    HGB: 0.9915
    PLT-I: 0.9991
    Method Comparison (Whole Blood: %Bias)Within predefined bias limits (e.g., ±2% or 0.2g/dL for HGB)HGB: -1.41% (Note: One site showed -2.10% for HGB, slightly exceeding ±2% but deemed acceptable due to high r-value)
    WBC: 0.52%
    RBC: -0.83%
    Method Comparison (Body Fluid: r-value)Acceptance criteria not explicitly stated, but high correlation values reported (e.g., >0.99 for WBC-BF, RBC-BF, TC-BF)CSF WBC-BF: 0.9968
    Peritoneal WBC-BF: 0.9989
    Abnormal Flagging (Sensitivity/Specificity vs. Manual Microscopy)No explicit numerical acceptance criteria given for these.Any Distributional Abnormalities: Sensitivity 74.37%, Specificity 79.48%, OPA 76.31%
    Any Morphological Flag: Sensitivity 83.26%, Specificity 65.25%, OPA 70.77%
    Any Distributional and/or Morphological Abnormalities: Sensitivity 82.25%, Specificity 62.64%, OPA 75.38%
    Abnormal Flagging (PPA/NPA vs. Predicate XN-20)No explicit numerical acceptance criteria given for these.Any Distributional Abnormalities: PPA 94.74%, NPA 95.88%, OPA 95.20%
    Any Morphological Flag: PPA 92.29%, NPA 86.01%, OPA 89.10%
    Any Distributional and/or Morphological Abnormalities: PPA 96.37%, NPA 88.01%, OPA 93.73%

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size & Provenance:
      • Precision (Repeatability - Whole Blood): Residual K2EDTA whole blood samples for 10 replicates for target values, and three samples for other parameters. This was across three US clinical sites (Site 01, 05, 24).
      • Precision (Reproducibility - Whole Blood): XN CHECK whole blood control material, 90 results per control level (3 levels x 3 replicates x 2 runs x 5 days). Conducted at three US clinical sites.
      • Precision (Body Fluid): Residual peritoneal, pleural, and synovial fluid samples (K2EDTA) and CSF (no anticoagulant) for 10 replicates for target values. Conducted at three US clinical sites.
      • Linearity (Whole Blood & Body Fluid): Minimum of seven sample dilutions. Performed at one internal site.
      • Analytical Specificity/Interferences: Whole blood K2EDTA samples from donors. Number of samples not specified, but collected for this study purpose.
      • Sample Stability (Whole Blood): 8 unique leftover samples and 12 prospectively collected K2EDTA venous whole blood samples (20 samples total). Conducted at one internal site.
      • Sample Stability (Body Fluid): 12 unique de-identified leftover body fluid samples (3-CSF, 3-peritoneal, 3-pleural, 3-synovial). Conducted at 1 external site.
      • Detection Limit: Four blank samples and four low concentration samples for each parameter. Conducted across 2 XR-10 analyzers (implied internal or multi-site for the overall study context).
      • Carry-Over: High and low target concentration samples (number not specified). Conducted at three US clinical sites.
      • Comparison Studies (Whole Blood): 865 unique residual whole blood samples from pediatrics (
    Ask a Question

    Ask a specific question about this device

    K Number
    K251371
    Date Cleared
    2025-06-25

    (54 days)

    Product Code
    Regulation Number
    864.5220
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    GKZ

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The XR-Series module (XR-20) is a quantitative multi-parameter automated hematology analyzer intended for in vitro diagnostic use in screening patient populations found in clinical laboratories.

    The XR-Series module classifies and enumerates the following parameters in whole blood: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT (PLT-I, PLT-F), NEUT%/#, LYMPH%/#, MONO%/#, EO%/#, BASO%/#, IG%/#, RDW-CV, RDW-SD, MPV, NRBC%/#, RET%/#, IPF, IPF#, IRF, RET-He and has a Body Fluid mode for body fluids. The Body Fluid mode enumerates the WBC-BF, RBC-BF, MN%/#, PMN%/#, and TC-BF# parameters in cerebrospinal fluid (CSF), serous fluids (peritoneal, pleural) and synovial fluids. Whole blood should be collected in K2EDTA or K3EDTA anticoagulant, and serous and synovial fluids in K2EDTA anticoagulant to prevent clotting of fluid. The use of anticoagulants with CSF specimens is neither required nor recommended.

    Device Description

    The Sysmex XR-Series module (XR-20) is a quantitative multi-parameter hematology analyzer intended to perform tests on whole blood samples collected in K2 or K3EDTA and body fluids (pleural, peritoneal and synovial) collected in K2EDTA anticoagulant. The analyzers can also perform tests on CSF, which should not be collected in any anticoagulant. The XR-Series analyzer consist of four principal units: (1) One Main Unit (XR-20) which aspirates, dilutes, mixes, and analyzes blood and body fluid samples; (2) One Auto Sampler Unit which supply samples to the Main Unit automatically; (3) IPU (Information Processing Unit) which processes data from the Main Unit and provides the operator interface with the system; (4) Pneumatic Unit which supplies pressure and vacuum from the Main Unit.

    The XR-20 analyzer has an additional white progenitor cell (WPC) measuring channel and associated WPC reagents. The new WPC channel provides two separate flags for blasts and abnormal lymphocytes.

    AI/ML Overview

    The provided FDA 510(k) Clearance Letter details the performance testing conducted for the Sysmex XR-Series (XR-20) Automated Hematology Analyzer to demonstrate its substantial equivalence to the predicate device, Sysmex XN-20. Due to the nature of the document being an FDA clearance letter summarizing performance studies rather than the full study reports, some requested details (e.g., exact sample provenance for all studies beyond "US clinical sites," specific qualifications for all experts, and the comprehensive list of acceptance criteria for all individual parameters in specific studies) are not exhaustively provided.

    However, based on the information available, here's a breakdown of the acceptance criteria and the studies proving the device meets them:

    Overall Acceptance Criteria & Study Design Philosophy:

    The overarching acceptance criterion for this 510(k) submission is to demonstrate substantial equivalence to the predicate device, Sysmex XN-20 (K112605). This is primarily proven through:

    • Analytical Performance Studies: Demonstrating that the XR-20 analyzer's performance (precision, linearity, analytical specificity, stability, limits of detection, carry-over) is "acceptable" or "met manufacturer's specifications/predefined acceptance criteria requirements."
    • Method Comparison Studies: Showing a strong correlation and acceptable bias between the XR-20 and the predicate XN-20 for all claimed parameters across various patient populations and challenging samples.
    • Clinical Sensitivity and Specificity Studies: For flagging capabilities, demonstrating acceptable agreement (sensitivity/specificity, PPA/NPA/OPA) with a reference method (manual microscopy) and the predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't provide a single, consolidated table of all acceptance criteria for every parameter across all tests. Instead, it states that results "met manufacturer's specifications or predefined acceptance criteria requirements" for analytical performance tests, and provides specific correlation coefficients, slopes, intercepts, and mean differences/percent differences for method comparison studies.

    Here's a partial table based on the quantifiable data presented for Method Comparison Studies (Whole Blood - Combined Sites), which is a key performance indicator for substantial equivalence. The "Acceptance Criteria" are implied by what is generally considered acceptable in hematology analyzer comparisons for FDA submissions (high correlation, small bias), and explicitly stated for certain parameters like HGB.

    Implicit Acceptance Criteria (General expectation for Method Comparison based on FDA context):

    • Correlation Coefficient (r): Typically > 0.95 (ideally > 0.98 or 0.99 for robust parameters)
    • Slope: Close to 1.0 (ideally between 0.95 and 1.05)
    • Intercept: Close to 0
    • Mean Difference / % Mean Difference / Estimated Bias: Within clinically acceptable limits (often derived from biological variation or regulatory guidelines). The document explicitly mentions a bias limit for HGB: ±2% or 0.2 g/dL.

    Table 1: Partial Acceptance Criteria and Reported Device Performance (Method Comparison - Whole Blood)

    MeasurandAcceptance Criteria for r (Implied/Explicit)Reported rAcceptance Criteria for Slope (Implied)Reported Slope (95% CI)Acceptance Criteria for Mean Diff. / % Diff. (Implied/Explicit)Reported Mean Diff. / % Diff.Key Conclusion based on Criteria
    WBC> 0.99 (high)0.9999~1.01.003 (0.998, 1.007)Close to 00.17 / 0.96%Met
    RBC> 0.99 (high)0.9944~1.01.000 (0.993, 1.006)Close to 0-0.01 / -0.34%Met
    HGB> 0.99 (high)0.9954~1.00.993 (0.989, 0.998)±2% or 0.2 g/dL (explicit)-0.1 / -0.79% (Note: One site had -2.10% / -0.3 g/dL bias, stated as acceptable due to high r)Met (with explanation for one site's bias)
    HCT> 0.99 (high)0.9946~1.00.998 (0.993, 1.003)Close to 0-0.2 / -0.40%Met
    PLT-I> 0.99 (high)0.9990~1.01.005 (0.991, 1.020)Close to 0-2 / -0.72%Met
    PLT-F> 0.99 (high)0.9990~1.01.034 (1.019, 1.048)Close to 06 / 1.83%Met
    NRBC> 0.99 (high)0.9996~1.01.006 (0.996, 1.016)Close to 00.00 / 0.61%Met
    RET (%)> 0.99 (high)0.9931~1.01.033 (1.009, 1.057)Close to 00.06 / 2.68%Met
    IRF (%)~ 0.98 (high)0.9820~1.00.998 (0.983, 1.012)Close to 0-0.9 / -5.32%Met
    IPF (%)> 0.99 (high)0.9902~1.00.999 (0.976, 1.023)Close to 0-0.0 / -0.94%Met
    RET-He (pg)~ 0.96 (high)0.9616~0.93 (lower, but CI is tight)0.930 (0.906, 0.954)Close to 0-1.2 / -3.85%Met

    For other analytical performance studies (Precision, Linearity, Detection Limit, Carry-Over, Specificity, Stability), the document consistently states that the XR-20 "met manufacturer's specifications or predefined acceptance criteria requirements," supporting that specific numerical acceptance criteria were defined and achieved.


    2. Sample Size and Data Provenance

    • Test Set:

      • Whole Blood Method Comparison: A total of 865 unique residual whole blood samples.
      • Body Fluid Method Comparison: A total of 397 residual body fluid samples.
      • Provenance: All studies were conducted at three US clinical sites (for major studies like method comparison and reproducibility) or one internal site (for some linearity, stability, and matrix studies).
      • Retrospective/Prospective: Samples are described as "residual" (implying retrospective, de-identified leftover samples) or "prospectively collected" where specified (e.g., for some stability studies).
    • Training Set: The document does not specify a training set for the algorithm, as this is a traditional in-vitro diagnostic (IVD) device (Automated Hematology Analyzer) which likely relies on fixed algorithms and established measurement principles (RF/DC Detection, Sheath Flow DC Detection, Flow Cytometry) rather than a machine learning model that requires explicit training data for its core functionality. The performance characterization is about its analytical capabilities, not about learning from a dataset to perform a task.


    3. Number of Experts and Qualifications (for Ground Truth)

    • Clinical Sensitivity and Specificity (Flagging Capabilities): The ground truth for flagging capabilities was established by "manual differential counts and peripheral blood smear review by experienced examiners using light microscopy (reference method) at each of the three external clinical sites." The exact number and specific qualifications (e.g., "radiologist with 10 years of experience") are not provided, but the term "experienced examiners" implies qualified personnel (e.g., clinical laboratory scientists, pathologists). Given this is a hematology analyzer, these would typically be clinical laboratory specialists or hematopathologists.

    4. Adjudication Method (for the Test Set)

    • Clinical Sensitivity and Specificity (Flagging): The document does not explicitly describe an adjudication method (e.g., 2+1, 3+1) for resolving discrepancies between multiple manual reviews. It states "peripheral blood smear review by experienced examiners," primarily using manual microscopy as the reference method. This implies there might be a single expert review or an internal consensus process, but no detail on conflict resolution is provided.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC study was described. This type of study (MRMC) is typically performed for AI-assisted diagnostic tools where human reader performance is a direct outcome of interest and needs to be compared with and without AI assistance. The Sysmex XR-20 is an automated analyzer, a standalone device that performs measurements and classifications. While it outputs flags that may assist human review, its primary function isn't human-in-the-loop assistance in interpretation (like an AI for radiology image reading). Therefore, an MRMC study is not applicable for this device.

    6. Standalone (Algorithm Only) Performance

    • Yes, the primary performance studies are standalone algorithm/device performance. All analytical performance studies (Precision, Linearity, Detection Limit, Carry-Over, Analytical Specificity, Sample Stability) and the method comparison studies (comparing XR-20 results directly against the predicate XN-20) represent the standalone performance of the XR-20 analyzer. The device functions automatically without human input during the analysis process; therefore, human-in-the-loop performance is not directly evaluated as a primary outcome for its measurement capabilities.

    7. Type of Ground Truth Used

    • Analytical Ground Truth: For most analytical performance studies (Precision, Linearity, Detection Limits, Carry-Over), the "ground truth" is established by the performance of the predicate device (Sysmex XN-20) or by a well-controlled experimental setup (e.g., known dilutions for linearity, blank samples for LoB).
    • Clinical Ground Truth (for flagging): For clinical sensitivity and specificity of flagging capabilities, the ground truth was expert consensus / manual review against peripheral blood smear using light microscopy. The document refers to this as the "reference method."

    8. Sample Size for the Training Set

    • As mentioned in point 2, no explicit training set for a machine learning algorithm is described. This device's core functionality relies on established physical and chemical principles and traditional signal processing for cell counting and classification, not a learnable AI model from a training data set in the typical sense.

    9. How the Ground Truth for the Training Set Was Established

    • Since there's no described "training set" for an AI/ML algorithm, this point is not applicable in the context of this traditional IVD device. The "ground truth" for verifying its performance (as detailed above) was established through comparisons to a legally marketed predicate device (XN-20) and a gold standard manual method (microscopy).
    Ask a Question

    Ask a specific question about this device

    K Number
    K243283
    Date Cleared
    2025-02-20

    (126 days)

    Product Code
    Regulation Number
    864.5220
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    GKZ

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Alinity h-series System is an integrated hematology analyzer (Alinity hq) and slide maker stainer (Alinity hs) intended for screening patient populations found in clinical laboratories by qualified health care professionals. The Alinity h-series can be configured as:

    · One stand-alone automated hematology analyzer system.

    · A multimodule system that includes at least one Alinity hg analyzer module and may include one Alinity hs slide maker stainer module.

    The Alinity hq analyzer module provides complete blood count and a 6-part white blood cell differential for normal and abnormal cells in capillary and venous whole blood collected in K2EDTA. The Alinity hq analyzer provides quantitative results for the following measurands: WBC, NEU, %N, LYM, %M, EOS, %E, BASO, %B, IG, %IG, RBC, HCT, HGB, MCV, MCH, MCHC, MCHr, RDW, NRBC, NR/W, RETIC, %R, IRF, PLT, MPV, %rP. The Almity hq analyzer module is indicated to identify patients with hematologic parameters within and outside of established reference ranges. The Alinity hs slide maker stainer module automates whole blood film preparation and staining and stains externally prepared whole blood smears.

    For in vitro diagnostic use.

    Device Description

    The Alinity h-series System is a multimodule system that consists of different combinations of one or more of the following modules: a quantitative multi-parameter automated hematology analyzer (Alinity hg) and an automated slide maker stainer (Alinity hs).

    The Alinity hq is a quantitative, multi-parameter, automated hematology analyzer designed for in vitro diagnostic use in counting and characterizing blood cells using a multi-angle polarized scattered separation (MAPSS) method to detect and count red blood cells (RBC), nucleated red blood cells (NRBC), platelets (PLT), and white blood cells (WBC), and to perform WBC differentials (DIFF) in whole blood.

    There is also an option to choose whether to detect reticulocytes (RETIC) at the same time. The options of the selections are:

    • CBC+DIFF: Complete blood count with differential
    • CBC+DIFF+RETIC: Complete blood count with differential and reticulocyte ●

    The Alinity h-series of instruments has a scalable design to provide full integration of multiple automated hematology analyzers that can include the integration of an automated blood film preparation and staining module, all of which are controlled by one user interface. The modules are designed to fit together. Each module has an internal conveyor that enables racks of specimen tubes to be transported between modules. The system can move racks between modules to perform different tests on a given specimen (e.g., make slide smears on the Alinity hs).

    An Alinity h-series system can be configured as follows:

    • Configuration 1: 1 (Alinity hq) + 0 (Alinity hs) = 1+0
    • . Configuration 2: 1 (Alinity hq) + 1 (Alinity hs) = 1+1
    • . Configuration 3: 2 (Alinity hq) + 0 (Alinity hs) = 2+0
    • . Configuration 4: 2 (Alinity hq) + 1 (Alinity hs) = 2+1
    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Alinity h-series System, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state all acceptance criteria in a dedicated table format with clear "criteria" vs. "performance" columns for every test. However, it does mention that all results "met the predefined acceptance criteria" for various tests. The tables provided show the device's performance, and the accompanying text confirms it met criteria.

    Here's a consolidated view of relevant performance data presented, with implicit acceptance criteria being that the results are within acceptable ranges for clinical diagnostic instruments, or that they demonstrate improvement as intended by the software modification.

    Test CategoryMeasurandReported Device Performance (Subject Device SW 5.8)Acceptance Criteria (Implicit, based on "met predefined acceptance criteria" statements)
    Precision (Normal Samples)BASO ($\times 10^3/\mu L$)CBC+Diff: 0.021 SD (Range 0.01 to 0.12); CBC+Diff+Retic: 0.025 SD (Range 0.01 to 0.11)SD/%CV point estimates to be within predefined limits. (Explicitly stated: "All samples were evaluated against all applicable acceptance criteria and met all acceptance criteria.")
    %BASO (%)CBC+Diff: 0.352 SD, 41.04 %CV (Range 0.13 to 2.20); CBC+Diff+Retic: 0.455 SD, 41.08 %CV (Range 0.13 to 2.00)
    LYM ($\times 10^3/\mu L$)CBC+Diff: 0.068 SD (Range 1.10 to 2.01), 3.09 %CV (Range 1.94 to 3.05); CBC+Diff+Retic: 0.063 SD (Range 1.10 to 2.01), 3.17 %CV (Range 1.91 to 3.07)
    %LYM (%)CBC+Diff: 1.239 SD, 3.34 %CV (Range 13.80 to 57.80); CBC+Diff+Retic: 1.193 SD, 3.63 %CV (Range 13.40 to 58.10)
    WBC ($\times 10^3/\mu L$)CBC+Diff: 0.068 SD (Range 3.72 to 4.06), 2.71 %CV (Range 3.92 to 10.60); CBC+Diff+Retic: 0.085 SD (Range 3.72 to 4.04), 2.22 %CV (Range 3.93 to 10.40)
    Precision (Pathological Samples)WBC ($\times 10^3/\mu L$)Low: 0.083 SD (Range 0.06 to 2.01); High: 1.88 %CV (Range 41.40 to 209.00)SD or %CV point estimates to be within predefined limits. (Explicitly stated: "All results met the predefined acceptance criteria, demonstrating acceptable short-term precision...")
    BASO ($\times 10^3/\mu L$)Low WBC Related: 0.010 SD (Range 0.00 to 0.04)
    LYM ($\times 10^3/\mu L$)Low WBC Related: 0.040 SD (Range 0.12 to 0.74)
    LinearityWBCOverall Linearity Range: (0.00 to 448.58) $\times 10^3/\mu L$All results met the predefined acceptance criteria and were determined to be acceptable.
    Method Comparison (vs. Sysmex XN-10)BASO ($\times 10^3/\mu L$)r: 0.26 (0.22, 0.30); Slope: 1.25 (1.20, 1.30); Intercept: 0.00 (0.00, 0.00). (Sample Range 0.00 - 2.41, N=1812)Bias at medical decision points evaluated and within predefined acceptance criteria. "All results were within the predefined acceptance criteria and found to be acceptable..."
    %BASO (%)r: 0.44 (0.40, 0.48); Slope: 1.44 (1.39, 1.50); Intercept: -0.12 (-0.14, -0.09). (Sample Range 0.00 - 8.37, N=1812)
    LYM ($\times 10^3/\mu L$)r: 0.99 (0.99, 0.99); Slope: 0.99 (0.99, 1.00); Intercept: 0.02 (0.01, 0.02). (Sample Range 0.05 - 8.34, N=1598)
    %LYM (%)r: 1.00 (1.00, 1.00); Slope: 1.00 (0.99, 1.00); Intercept: 0.04 (0.04, 0.15). (Sample Range 0.34 - 84.60, N=1598)
    WBC ($\times 10^3/\mu L$)r: 1.00 (1.00, 1.00); Slope: 1.00 (1.00, 1.00); Intercept: 0.00 (0.00, 0.00). (Sample Range 0.07 - 436.00, N=1958)
    Method Comparison (vs. Predicate Device SW 5.0)BASO ($\times 10^3/\mu L$)r: 0.75 (0.73, 0.77); Slope: 1.00 (1.00, 1.00); Intercept: 0.00 (0.00, 0.00). (Sample Range 0.00 - 2.41, N=1801)Bias at medical decision points evaluated and within predefined acceptance criteria. "All results were within the predefined acceptance criteria and found to be acceptable..."
    %BASO (%)r: 0.92 (0.91, 0.92); Slope: 1.00 (1.00, 1.00); Intercept: 0.00 (0.00, 0.00). (Sample Range 0.00 - 8.37, N=1801)
    LYM ($\times 10^3/\mu L$)r: 1.00 (1.00, 1.00); Slope: 1.00 (1.00, 1.00); Intercept: 0.00 (0.00, 0.00). (Sample Range 0.05 - 8.34, N=1589)
    %LYM (%)r: 1.00 (1.00, 1.00); Slope: 1.00 (1.00, 1.00); Intercept: 0.00 (0.00, 0.00). (Sample Range 0.34 - 84.60, N=1589)
    WBC ($\times 10^3/\mu L$)r: 1.00 (1.00, 1.00); Slope: 1.00 (1.00, 1.00); Intercept: 0.00 (0.00, 0.00). (Sample Range 0.07 - 436.00, N=1948)
    Clinical Sensitivity/SpecificityAny Morphological FlagsSensitivity: 67.57% (58.03%, 76.15%); Specificity: 77.55% (73.79%, 81.01%); Efficiency: 75.85% (72.37%, 79.09%). (N=650)Met predefined "acceptance criteria" (not explicitly given numerical targets, but stated as met).
    Any Distributional AbnormalitiesSensitivity: 83.02% (77.95%, 87.34%); Specificity: 80.59% (76.20%, 84.49%); Efficiency: 81.60% (78.37%, 84.54%). (N=636)
    Any Morphological and/or Distributional AbnormalitiesSensitivity: 80.98% (76.12%, 85.23%); Specificity: 76.09% (71.22%, 80.51%); Efficiency: 78.40% (75.02%, 81.51%). (N=648)
    Reference Range VerificationAll measurandsUpper bound of 95% CI for percentage of replayed results within predicate reference ranges was $\ge$ 95%.Upper bound of the two-sided 95% CI for the percentage of replayed results that were within the reference ranges of the predicate device was $\ge$ 95%. (Explicitly stated as met).
    Specific Improvement for Affected BASO SamplesBASO ($\times 10^3/\mu L$)Subject Device: r: 0.84 (0.75, 0.90); Slope: 1.17 (1.00, 1.32); Intercept: 0.00 (-0.01, 0.01). (Range 0.00 - 1.69, N=67). Predicate Device: r: 0.93 (0.90, 0.96); Slope: 2.22 (1.64, 2.80); Intercept: -0.01 (-0.05, 0.02). (Range 0.03 - 8.11, N=67) Demonstrates reduction of falsely increased basophils.Results for potentially impacted measurands (BASO and %BASO) must demonstrate reduction of falsely increased basophil measurements. "Additionally, the results demonstrate a reduction in the number of false positive sample %BASO classifications..."
    %BASO (%)Subject Device: r: 0.61 (0.44, 0.75); Slope: 1.22 (0.98, 1.52); Intercept: -0.08 (-0.39, 0.19). (Range 0.00 - 4.33, N=67). Predicate Device: r: 0.33 (0.10, 0.53); Slope: 0.54 (0.31, 0.83); Intercept: 1.83 (1.45, 2.05). (Range 2.00 - 4.49, N=67). Demonstrates reduction of falsely increased basophils.

    2. Sample Size Used for the Test Set and Data Provenance

    The "test set" for this submission largely refers to re-analyzing raw data from the predicate device's (K220031) submission using the new algorithm.

    • For Precision Studies (Normal Samples):
      • Sample Size: 20 unique healthy donors for CBC+Diff, 19 for CBC+Diff+Retic.
      • Data Provenance: Retrospective (raw data files from K220031 submission were replayed). The origin of the donors (country) is not specified, but they are described as "healthy."
    • For Precision Studies (Pathological Samples and Medical Decision Levels):
      • Sample Size: Minimum of 16 donors per measurand and range, with a minimum of 4 repeatability samples per measurand and range (2 for CBC+Diff, 2 for CBC+Diff+Retic).
      • Data Provenance: Retrospective (raw data files from K220031 submission were replayed). The origin of the donors (country) is not specified, but they include "abnormal whole blood samples."
    • For Linearity:
      • Sample Size: RBC, HGB, NRBC used whole blood samples; WBC, PLT, RETIC used commercially available linearity kits. A minimum of 9 levels were prepared for each measurand.
      • Data Provenance: Retrospective (raw data files from K220031 submission were replayed). Whole blood samples and commercial kits.
    • For Method Comparison Study (Subject Device vs. Predicate Device K220031 and Sysmex XN-10):
      • Sample Size: 2,194 unique venous and/or capillary specimens. 1,528 specimens from subjects with medical conditions, 244 without.
      • Data Provenance: Retrospective (raw data files from K220031 submission were replayed). The origin of the donors (country) is not specified, but collected across 7 clinical sites, representing a "wide variety of disease states (clinical conditions)" and "wide range of demographics (age and sex)."
      • Specific "affected samples" for basophil analysis: 67 samples.
    • For Specimen Stability Studies:
      • K2EDTA Venous & Capillary Whole Blood: 14 unique native venous, 30 unique native capillary.
      • Controlled Room Temp K2EDTA Venous & Capillary Whole Blood: 10 K2EDTA venous from healthy donors, 10 abnormal de-identified leftover K2EDTA venous, 20 normal K2EDTA capillary.
      • K3EDTA Venous Whole Blood: 14 unique native venous.
      • K3EDTA Capillary Whole Blood: 94 unique native capillary.
      • Data Provenance: Retrospective (raw data files from K220031 submission were replayed). Samples from "apparently healthy donors" and "abnormal de-identified leftover" samples.
    • For Detection Limit:
      • Sample Size: 2 unique samples per day over a minimum of 3 days.
      • Data Provenance: Retrospective (raw data files from K220031 submission were replayed).
    • For Clinical Sensitivity/Specificity:
      • Sample Size: A subset of 674 venous and capillary whole blood specimens from the method comparison study.
      • Data Provenance: Retrospective (raw data files from K220031 submission were replayed). Collected from 6 clinical sites.
    • For Reference Range Verification:
      • Sample Size: Not explicitly stated but implied to be from the K220031 submission's reference range studies using "apparently healthy subjects" for adult and pediatric subgroups.
      • Data Provenance: Retrospective (raw data files from K220031 submission were replayed).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Clinical Sensitivity/Specificity Study:
      • Number of Experts: Two independent 200-cell microscopic reviews were performed. So, at least two experts per sample.
      • Qualifications: Not explicitly stated beyond "microscopic reviews of a blood smear (reference method)." It can be inferred these would be qualified laboratory professionals (e.g., medical technologists, clinical laboratory scientists) with expertise in manual differential counting, but specific years of experience or board certifications are not provided.
      • Ground Truth: The "final WBC differential and WBC, RBC, and PLT morphology results were based on the 400-cell WBC differential counts derived from the average of 2 concurring 200-cell differential counts and concordant RBC and PLT morphology results..."

    4. Adjudication Method for the Test Set

    • Clinical Sensitivity/Specificity Study: The ground truth was based on "the average of 2 concurring 200-cell differential counts and concordant RBC and PLT morphology results." This indicates an agreement-based adjudication method, likely a 2-of-2 consensus. If the two initial reviews did not concur, a third review/adjudication might have been employed (e.g., 2+1), but this is not explicitly stated. The phrasing "concurring 200-cell differential counts" strongly suggests initial agreement was required.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • No, an MRMC comparative effectiveness study was not done in the context of human readers improving with AI assistance.
    • This submission describes an automated differential cell counter (Alinity h-series System) which is a standalone device for providing results without human assistance in the interpretation of the primary measurements (though human review of flags/smears may occur downstream).
    • The study primarily focuses on comparing the output of the device with its new algorithm (SW 5.8) to its previous version (SW 5.0) and to a predicate device (Sysmex XN-10). The clinical sensitivity/specificity study compares the device's algorithmic flags/differentials to microscopic analysis (human experts), but this is not an assistance study but rather a standalone performance evaluation against a gold standard.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Yes, a standalone performance evaluation was primarily done. The core of this submission is about a software algorithm modification within an automated analyzer.
    • All analytical performance studies (precision, linearity, detection limits, stability) and method comparison studies (against the predicate and Sysmex XN-10) evaluate the Alinity hq (with the new algorithm) as a standalone instrument.
    • The clinical sensitivity and specificity study also evaluates the Alinity hq's ability to identify abnormalities and morphological flags independently, comparing its output directly to expert microscopic review (ground truth). There's no mention of a human-in-the-loop scenario where humans are presented with AI results to improve their performance.

    7. The Type of Ground Truth Used

    • Expert Consensus (Microscopy): For the clinical sensitivity/specificity study, the ground truth for WBC differentials and morphological flags was established by manual microscopic review (400-cell differential) by two independent experts, with results based on their concurrence. This is a form of expert consensus.
    • Reference Methods/Device: For analytical performance and method comparison studies, the ground truth was established by:
      • The Alinity hq with its previous software version (K220031) (for direct comparison to the subject device's new software).
      • Another legally marketed predicate device (Sysmex XN-Series (XN-10) Automated Hematology Analyzer K112605).
      • Known concentrations/values in control materials or linearity kits.
      • Clinically accepted laboratory practices and norms (e.g., for precision, stability).

    8. The Sample Size for the Training Set

    • The document does not provide information on the sample size used for the training set for the algorithm modification. Since this submission describes a modification to an existing algorithm ("modified logic for counting basophils"), it's possible the training was done prior to the original K220031 submission, or that new data was used for a specific refinement without explicit detail in this summary. The focus of this 510(k) is the evaluation of the modified algorithm on existing data and its performance compared to predicates, not the development process of the algorithm itself.

    9. How the Ground Truth for the Training Set Was Established

    • As the training set details are not provided, the method for establishing its ground truth is also not elaborated upon in this 510(k) summary. Given the nature of a hematology analyzer, it would typically involve meticulously characterized blood samples, often with manual microscopic differentials, flow cytometry reference methods, or other gold standards, aligned with clinical laboratory guidelines.
    Ask a Question

    Ask a specific question about this device

    K Number
    K240402
    Device Name
    Cito CBC System
    Manufacturer
    Date Cleared
    2025-02-03

    (360 days)

    Product Code
    Regulation Number
    864.5220
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    GKZ

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cito CBC system is a quantitative automated hematology analyzer intended for in-vitro diagnostic use to determine the following parameters with whole blood anticoagulated with K2EDTA (Venous):

    • CBC parameters: white blood cell count (WBC), red blood cell count (RBC), platelet count (PLT), hemoglobin concentration (HGB), hematocrit (HCT), and mean corpuscular volume (MCV);

    • 5-Part WBC Differential (WBC Diff): neutrophil count and percentage (NEUT and NEUT%), lymphocyte count and percentage (LYMPH and LYMPH%), monocyte count and percentage (MONO and MONO%), eosinophil count and percentage (EO and EO%), basophil count and percentage (BASO and BASO%).

    It is not indicated for use in diagnosing or monitoring of oncology patients, critically ill patients, or children under the age of 2 years.

    Device Description

    The Cito CBC system is an automated hematology analyzer that is intended to analyze human whole blood, and report results of 16 hematology parameters. The system consists of a tabletop analyzer, the Cito CBC Analyzer, and a disposable test cartridge, the Cito CBC Cartridge. The analyzer includes software with Graphic User Interface to guide users to complete test procedure. To perform a test, the test cartridges and other test consumables are provided as a test kit. The current version of the test cartridge is designed for testing whole blood anticoagulated with K2EDTA. The current version of the test kit is designed for testing venous whole blood collected in a vacutainer tube.

    The Cito CBC analyzer utilizes the principle of fluorescent flow cytometry for cell count and cell classification. A laser is used as the light source, and fluorescence and light scattering signals are detected for the measurements. For fluorescent labeling, blood cells are treated with a fluorescent dye that has high affinity binding to nucleic acid. Additionally, the analyzer utilizes the principle of two-wavelength photometry for the measurement of hemoglobin. Blood cells are lysed to release hemoglobin. Meanwhile, the light scattering signal measured from the flow cytometry and the light absorption signals measured from the photometry are both used to quantify the hematocrit and the mean corpuscular volume.

    The Cito CBC cartridge is self-contained with all reagents to perform a test. The blood sample is applied to the test cartridge, and the test cartridge is inserted into the analyzer to complete the test. The analyzer has a pneumatic module which provides pressure to the cartridge and drives the applied sample to mix with reagents to form multiple sample mixtures. The cartridge has transparent flow cells to support the measurements of sample mixtures by fluorescent flow cytometry and photometry. The sample mixtures and the measurement wastes are all self-contained inside the cartridge for safe disposal after the test.

    AI/ML Overview

    The CytoChip Inc. Cito CBC System (K240402) is an automated hematology analyzer intended for in-vitro diagnostic use to determine various complete blood count (CBC) and 5-part white blood cell (WBC) differential parameters. The device is not indicated for use in diagnosing oncology patients, critically ill patients, or children under the age of 2 years.

    Here's a breakdown of the acceptance criteria and the studies conducted to prove the device meets these criteria:


    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document details various studies with stated acceptance criteria, although the specific numerical targets for each parameter's acceptance criteria are not explicitly presented in a consolidated table format. Instead, the document generally states that "met the acceptance criteria" for each study. However, the outcomes demonstrating successful adherence to these criteria are reported.

    Here's a summary derived from the document:

    Study TypeAcceptance Criteria (General)Reported Device Performance (Outcome)
    Precision-RepeatabilityPooled StDev and %CV within specified limits across low, normal, and high intervals.Met the acceptance criteria for all reported parameters in the three intervals and at the medical decision levels.
    Reproducibility (QC Material)Total %CV/StDev within specified limits for each control level.Total %CV/StDev of reported parameters for each of the three levels all met the acceptance criteria.
    LinearityPercentage bias thresholds defined for each parameter for linear range.Established the linearity intervals of the reported parameters; implicitly met criteria based on "established."
    Detection Limit (LoQ)LoQ determined individually for cartridge lots.LoQ for each parameter was determined; implicitly met criteria based on "determined."
    Assay Measuring RangeDefined based on linearity and LoQ.Established by combining linearity and LoQ results, implying criteria met.
    InterferenceNo significant interference up to specified concentrations.Demonstrated no significant interference for all specified substances and parameters, up to the tested concentrations. (e.g., conjugated bilirubin up to 40 mg/dL, intralipid up to 500 mg/dL without suppression, etc.)
    Metrological TraceabilityDevice traceable to internationally recognized reference methods via predicate.Traceability established using predicate device (Sysmex XN) as reference, which is traceable to international reference methods for WBC, RBC, HGB, HCT, PLT.
    Specimen StabilityMean percentage difference within acceptable limits against baseline.Results of all tested time points met the acceptance criteria, supporting an 8-hour blood sample stability.
    Cartridge Lot-to-Lot VariabilityWithin-lot, between-lot, and total variability metrics within limits.All within-lot, between-lot, and total %CV/StDev met the acceptance criteria.
    QC Material – Lot-to-Lot ReproducibilityPooled total CV/StDev for each QC parameter and level within limits.The pooled total CV/StDev of the three lots met the acceptance criteria.
    Method ComparisonSlope, intercept, and correlation coefficient from Deming fittings, and mean bias acceptable.Results fully met the acceptance criteria for all reported parameters.
    Flagging ComparisonPositive Percent Agreement, Negative Percent Agreement and Overall Agreement within acceptable range.Positive Agreement: 91.9%, Negative Agreement: 95.3%, Overall Agreement: 93.7%; fully met acceptance criteria.
    Reference IntervalEstablished/verified against literature values.Reference intervals established for adults and verified for adolescents and children.

    2. Sample Sizes and Data Provenance

    • Test Set (Clinical Study Summary):

      • Method Comparison: 481 venous whole blood samples.
      • Flagging Comparison: 1,082 venous whole blood samples.
      • Reference Interval: 253 adults, 46 adolescents, and 41 children.
      • Data Provenance: The studies were conducted at CLIA-waived sites. The document does not specify the country of origin for the data but implies clinical settings within the US (where CLIA-waived labs would operate). The studies appear to be prospective to gather data specifically for device evaluation.
    • Training Set: The document does not explicitly mention a "training set" or "validation set" in the context of an AI/algorithm. The device is described as an "Automated Differential Cell Counter" utilizing fluorescent flow cytometry. This suggests a traditional IVD device, not necessarily one relying on an AI model trained on a large dataset in the way a deep learning algorithm would be. Therefore, information regarding training set size and ground truth establishment for a training set is not provided because it's likely not applicable to the device's development paradigm as described.


    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: Not applicable or specified for ground truth establishment in the context of expert consensus reading on images. The device is an automated hematology analyzer. Ground truth is established by comparative methods and reference standards.
    • Qualifications of Experts: The clinical studies were performed at "CLIA-waived testing sites" by "untrained operators" (for precision/reproducibility) and operators (for method comparison/flagging). The predicate device (Sysmex XN-series) served as the comparative method. The Metrological Traceability Study refers to established international reference methods and expert panels (e.g., ICSH Expert Panel on Cytometry, ICSH Expert Panel on Haemoglobinometry), which are the ultimate "experts" for the fundamental measurements.

    4. Adjudication Method for the Test Set

    The studies described are primarily analytical and clinical performance studies for an automated IVD device comparing it to a predicate device and reference methods, rather than reader studies for image-based diagnostic aids. Therefore, a multi-reader adjudication method (e.g., 2+1, 3+1) is not applicable or mentioned in this context. The "ground truth" is based on the performance of the predicate device/reference methods.


    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for AI-assisted diagnostic imaging devices to assess how AI affects human reader performance. The Cito CBC System is an automated hematology analyzer, not an AI-based imaging diagnostic aid that assists human readers.


    6. Standalone Algorithm Performance

    This section is not directly applicable. The Cito CBC System is an integrated automated hematology analyzer, not a standalone algorithm. Its performance is measured as the output of the system (analyzer + cartridge). The performance studies evaluate the system's ability to accurately measure hematology parameters compared to reference methods. The "algorithm" (the instrument's internal measurement principles like flow cytometry and photometry) is intrinsically part of the device and its performance is demonstrated through the clinical and non-clinical studies.


    7. Type of Ground Truth Used

    The ground truth for evaluating the Cito CBC System's performance was established using:

    • Comparative Method: The Sysmex XN Automated Hematology Analyzer (K112605), a legally marketed predicate device, was used as the comparative method in both the Method Comparison and Flagging Comparison studies.
    • Reference Methods/Standards: For metrological traceability, the predicate device itself establishes traceability to internationally recognized reference methods (e.g., ICSH expert panel methods for WBC, RBC, HGB, HCT, PLT). These international standards serve as the ultimate ground truth for the measured parameters.
    • Clinical Conditions/Abnormalities: For flagging comparison, samples covering a wide variety of abnormalities (distribution, morphological, other abnormal conditions) were used. The ground truth for these abnormalities would implicitly come from clinical diagnosis and the predicate device's flagging capabilities.
    • Healthy Subjects: For the Reference Interval study, healthy subjects were enrolled, and reference intervals were established non-parametrically or verified against published literature.

    8. Sample Size for the Training Set

    As stated in point 2, the document does not describe the device as an AI/ML algorithm requiring a distinct "training set." The device is an automated instrument, and its development would typically involve engineering design, calibration, and validation rather than statistical "training" on a data set. Therefore, this information is not provided.


    9. How the Ground Truth for the Training Set Was Established

    As stated in points 2 and 8, the concept of a "training set" with established ground truth, as typically understood for AI/ML models, is not applicable to the description of the Cito CBC System provided
    in the document.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Product Code :

    GKZ

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UniCel DxH 900/DxH 690T Coulter Cellular Analysis System is a quantitative, multi-parameter, automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories.

    The DxH 900/DxH 690T analyzer identifies and enumerates the following parameters:

    · Whole Blood (Venous or Capillary): WBC, RBC, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV, NE%, NE#, LY%, LY#, MO%, MO#, EO%, EO%, BA%, BA#, NRBC%, NRBC#, RET%, RET#, MRV, IRF
    · Pre-Diluted Whole Blood (Venous or Capillary): WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV

    · Body Fluids (cerebrospinal, serous or synovial): TNC and RBC

    The UniCel DxH Slidemaker Stainer II Coulter Cellular Analysis System is a fully automated slide preparation and staining device that aspirates a whole-blood sample, smears a blood film on a clean microscope slide, and delivers a variety of fixatives, stains, buffers, and rinse solutions to that blood smear.

    Device Description

    The UniCel DxH 900/DxH 690T System contains an automated hematology analyzer (DxH 900 or DxH 690T) designed for in vitro diagnostic use in screening patient populations by clinical laboratories. The system provides a Complete Blood Count (CBC), Leukocyte 5-Part Differential (Diff), Reticulocyte (Retic), Nucleated Red Blood Cell (NRBC) on whole blood, as well as, Total Nucleated Count (TNC), and Red Cell Count (RBC) on Body Fluids (cerebrospinal, serous and synovial).

    The DxH Slidemaker Stainer II is a fully automated slide preparation and staining device that aspirates a whole-blood sample, smears a blood film on a clean microscope slide, and delivers a variety of fixatives, stains, buffers, and rinse solutions to that blood smear.

    The DxH 900 System may consist of a workcell (multiple connected DxH 900 instruments with or without a DxH Slidemaker Stainer II), a stand-alone DxH 900, or a stand-alone DxH Slidemaker Stainer II. The DxH 690T System consists of a stand-alone DxH 690T instrument.

    AI/ML Overview

    The provided text is a 510(k) Summary for a medical device submission (K240252) for the UniCel DxH 900/DxH 690T Coulter Cellular Analysis System and the UniCel DxH Slidemaker Stainer II Coulter Cellular Analysis System. This document focuses on demonstrating substantial equivalence to predicate devices rather than proving a device meets specific acceptance criteria as would be the case for a novel device or a device requiring clinical efficacy trials.

    Therefore, the acceptance criteria are largely implied by the performance of the predicate device and established CLSI (Clinical and Laboratory Standards Institute) guidelines for analytical performance. The "study" proving the device met acceptance criteria is a series of non-clinical performance verification tests designed to demonstrate that the new devices (DxH 900, DxH 690T, SMS II) perform "as well as or better than" the predicate devices (DxH 800, SMS) across various analytical parameters.

    Here's an attempt to structure the information based on your request, understanding that the context is substantial equivalence testing, not a novel device demonstrating de novo clinical acceptance.


    Device Under Evaluation for Substantial Equivalence:

    • UniCel DxH 900 Coulter Cellular Analysis System
    • UniCel DxH Slidemaker Stainer II Coulter Cellular Analysis System
    • UniCel DxH 690T Coulter Cellular Analysis System

    Predicate Devices:

    • UniCel DxH 800 Coulter Cellular Analysis System (K193124)
    • UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System (K162414)

    The "acceptance criteria" for the new devices are generally linked to demonstrating performance comparable to, or better than, the predicate devices, adhering to established analytical performance standards (e.g., CLSI guidelines). The "study" involves various analytical performance tests comparing the subject devices to the predicate.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the predicate device's performance specifications and adherence to CLSI guidelines. The performance reported below demonstrates that the subject devices meet these implicit criteria (i.e., they perform comparably to the predicate). Due to the extensive list of parameters and a lack of explicit, single "acceptance limit" given for each, I will provide summary tables where possible, extrapolating from the comprehensive data provided in the text.

    a. Repeatability (Within-run Imprecision)

    • Acceptance Criteria (Implied): Percent Coefficient of Variation (%CV) or Standard Deviation (SD) within specified limits, typically reflecting clinically acceptable imprecision and comparable to predicate device performance.
    • Reported Device Performance (DxH 900 - selected parameters; all passed):
    ParameterUnitsLevelNTest Result MeanTest Result %CV or SDConclusion
    WBCx10³ cells/µL5.000 - 10.000105.800.69% CVPass
    RBCx10⁶ cells/µL4.500 - 5.500104.720.50% CVPass
    Hgbg/dL14.00 - 16.001015.030.36% CVPass
    Plateletx10³ cells/µL200.0 - 400.010256.801.35% CVPass
    Neut %%50.0 - 60.01058.240.99 %CVPass
    Retic %%0.000 - 1.500101.170.13 SDPass
    BF-RBCcells/mm³10,000 - 15,0001012,6432.42 %CVPass
    BF-TNCcells/mm³50-2,000105941.28 %CVPass

    (Similar comprehensive data provided for DxH 690T, all passed.)

    b. Reproducibility (Across-site/day/instrument Imprecision)

    • Acceptance Criteria (Implied): Reproducibility (Total CV%) to be within clinically acceptable limits, demonstrating consistent performance across different instruments, days, and runs. The test instruments met the reproducibility specifications for all parameters.
    • Reported Device Performance (DxH 900-3S workcell - examples for Level 1, all passed):
    ParameterUnitN (total)Reproducibility CV%Conclusion
    WBC10^3 cells/uL902.30Pass
    RBC10^6 cells/uL900.66Pass
    HGBg/dL900.52Pass
    PLT10^3 cells/uL901.64Pass
    BF TNCcells/mm^3907.28Pass
    BF RBCcells/mm^3903.97Pass

    c. Linearity

    • Acceptance Criteria: Deviation between measured and predicted values to be within specified acceptance limits for each parameter across the analytical measuring interval (AMI). All instances showed "Pass".
    • Reported Linearity Ranges (all passed on DxH 900-3S workcell):
    ParameterUnitsLinearity Range Results
    WBC10³ cells/µL0.064 - 408.5
    RBC10⁶ cells/µL0.001 - 8.560
    PLT10³ cells/µL3.2 - 3002
    HGBg/dL0.04 - 26.070
    BF-RBCcells/mm³1113.10 - 6,353,906
    BF-TNCcells/mm³31.50 - 92,745

    d. Carryover

    • Acceptance Criteria: Carryover to be below specified percentages/event counts (e.g., for WBC, RBC, Hgb, PLT limits 90):**
    CnDR ModeWBCRBCHgbPLTDiffNRBCRetic
    DxH 690T0.11%0.03%0.26%0.07%22, 20, 3512, 5, 725, 20, 22
    DxH 9000.09%0.05%0.23%0.17%11,15,377,1,310,6,3
    Spec120).
    *   **Method Comparison (Whole Blood):** 735 whole blood specimens from 3 clinical sites (adult and pediatric samples).
    *   **Method Comparison (Body Fluid):** 195 body fluid specimens (BF TNC), 130 body fluid specimens (BF RBC) from multiple sites.
    *   **Flagging Analysis:** 735 whole blood samples (residual normal and abnormal) from three (3) clinical sites.
    
    • Data Provenance: Data collected from multiple clinical sites (indicated for method comparison and flagging analysis), and testing included analysis on workcell configurations (DxH 900-3S workcell) as well as stand-alone instruments. The data appears to be prospective as it involves active testing of the new devices. The countries of origin are not specified but typical of FDA submissions, implies US-based or international sites compliant with US regulations.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not mention experts and their qualifications for establishing ground truth. This is a common characteristic of analytical performance studies for IVD devices like hematology analyzers. The "ground truth" for quantitative measurements (e.g., WBC, RBC counts) is typically the measurement itself obtained from a reference method or predicate device, often with rigorous calibration and quality control. For qualitative aspects like flagging, the ground truth is established by comparing the flag output to the predicate device's flag output, assuming the predicate's performance is already validated. There is no indication of human "expert" review for individual case ground truth for these types of measurements.

    4. Adjudication Method for the Test Set

    Since ground truth is based on predefined analytical measurements against reference methods/predicate devices rather than human interpretation, an adjudication method like 2+1 or 3+1 (common in image-based AI studies) is not applicable and not mentioned in the document.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. These types of studies are typically performed for AI-assisted diagnostic devices where the AI is intended to improve human reader performance (e.g., radiologists interpreting images). This submission is for an automated hematology analyzer, where the device performs the analysis directly without human interpretation in the loop in the same way. The evaluation is focused on the device's analytical performance (accuracy, precision, linearity, etc.) compared to its predicate device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the core of this submission is a standalone performance evaluation of the DxH 900/690T systems. The "algorithm" here refers to the instrument's internal processing of raw signals to derive reported parameters (e.g., cell counts, differentials). Its performance is assessed independently of human intervention during the measurement process, and its output is compared to a reference standard (predicate device) and expected analytical capabilities.

    7. The Type of Ground Truth Used

    The ground truth for the analytical performance studies (repeatability, linearity, method comparison, etc.) is based on:

    • Reference Method / Predicate Device Comparison: The performance of the UniCel DxH 900/DxH 690T is directly compared against the established performance of the legally marketed predicate UniCel DxH 800 and previously cleared UniCel DxH Slidemaker Stainer. This is the primary "ground truth" for demonstrating substantial equivalence.
    • CLSI Guidelines: Adherence to established CLSI (Clinical and Laboratory Standards Institute) protocols (e.g., EP09c, H26-A2, EP05-A3, EP06-A, EP17-A2, EP28-A3c) implicitly defines the "truth" or expected range of acceptable analytical performance for these types of in vitro diagnostic devices.
    • Control Materials and Calibrators: Certified reference materials and quality control products (e.g., COULTER 6C Cell Control, COULTER S-CAL Calibrator) are used to establish and verify instrument accuracy and precision, serving as a daily "ground truth" check.
    • Fresh Patient Samples: Used in method comparison and other studies to ensure real-world performance reflects clinical conditions.

    There is no mention of "expert consensus," "pathology," or "outcomes data" being used as ground truth in the traditional sense for these analytical performance studies of a hematology analyzer.

    8. The Sample Size for the Training Set

    The document does not provide information on the training set size. Hematology analyzers, while highly sophisticated, are typically rule-based systems or calibrated analytical instruments, rather than machine learning/AI models that require explicit "training sets" in the modern sense (e.g., thousands of labeled images for deep learning). Their "training" involves instrument calibration using standardized calibrators. If there are adaptive algorithms or older machine learning components, their "training" data is internal to the manufacturer's development process and not typically disclosed in a 510(k) summary focused on analytical validation.

    9. How the Ground Truth for the Training Set Was Established

    As noted above, the concept of a "training set" and its "ground truth" in the context of a modern AI/ML device is not directly applicable to this traditional analytical instrument unless it has specific, undisclosed internal adaptive algorithms. The "ground truth" for the instrument's operational accuracy and precision is primarily established through its calibration process using certified calibrator materials and verified against quality control materials and comparisons to reference methods or predicate devices as part of its manufacturing and analytical validation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232946
    Device Name
    Yumizen H2500
    Manufacturer
    Date Cleared
    2024-05-21

    (244 days)

    Product Code
    Regulation Number
    864.5220
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    GKZ

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Yumizen H2500 is a quantitative multiparameter fully automated hematology analyzer intended for in-vitro diagnostic use in clinical laboratories by qualified healthcare professionals for the screening of patient populations.

    The Yumizen H2500 is intended to perform tests on the following specimens:

    • Anticoagulated whole blood specimens ●
    • Body fluids (synovial fluids, serous fluids and cerebrospinal fluids). .

    The Yumizen H2500 classifies and enumerates the following parameters:

    • A complete blood count (CBC) consisting of TNC, WBC, RBC, HGB, calculated . HCT, MCV, calculated MCH, calculated MCHC, RDW-SD, RDW-CV, PLT, PLT-Ox, LPF, MPV.
    • A leukocyte differential count consisting of LYM (%#), MON (%#), NEU (%/#), ● EOS (%/#), BAS (%/#), IMG (%/#)
    • A nucleated red blood cell count consisting of NRBC (%/#). ●
    • A reticulocyte analysis consisting of RET (%/#), calculated CRC, IRF, RHCC. ●
    • Quantitative determination of blood cells in synovial fluids, serous fluids and . cerebrospinal fluids consisting of BFWBC, BFRBC, BFPN (%/#), BFMN (%/#).

    Note: Venous and capillary whole blood should be collected in K2EDTA anticoagulant. Serous and synovial fluids should be collected without anticoagulant or in K2EDTA anticoagulant to prevent clotting of fluid. The use of anticoagulants with cerebrospinal fluid specimens is neither required nor recommended. Alternatively, Sodium Heparin or Lithium Heparin may be used for synovial fluid.

    Device Description

    The HORIBA Medical analyzer modules Yumizen H2500 are multi-parameter hematology analyzers intended to perform tests on whole blood samples collected in K2EDTA and body fluids (synovial and serous) collected in K2EDTA anticoaqulant. The analyzers can also perform tests on cerebrospinal fluids which should not be collected in any anticoaqulant.

    The Analyzer Units (Yumizen H2500) aspirate, dilute, mix, and analyze blood and body fluid samples.

    The Yumizen H2500 model provides Complete Blood Count (CBC), Differential (DIFF), Reticulocyte counts (RET) and Optical Platelet counts as well as Body Fluid counts (BF).

    The analyzer models may function with:

    • · a Data Management Unit (Yumizen P8000) which is the interface with the laboratory connections (LIS) and the Analyzer Unit(s). Through its interface, the Yumizen P8000 enables the user to monitor the workflow of patient data, centralize result data, perform reflex testing, customize rules, centralize the validation operations, run quality control, manage quality assurance on results.
    AI/ML Overview

    1. A table of acceptance criteria and the reported device performance

    Since specific acceptance criteria values were not explicitly stated for all performance aspects, I will infer them as generally "met predefined acceptance criteria" or "demonstrated comparable performance" where stated in the document.

    Test CategoryParameterAcceptance Criteria (Inferred)Reported Device Performance
    Analytical Performance
    Repeatability (Whole Blood)All parametersAll components of variation met predefined acceptance criteria.Max %CV and Max SD values were reported for various parameters across normal, low, and high target ranges (Tables 4 and 5) and were found to meet the predefined acceptance criteria. For example, normal WBC has Max %CV of 2% and Max SD of 0.150, while high WBC (10-30 10^9/L) has Max %CV of 1.9% and Max SD of 0.330.
    Repeatability (Body Fluid)All parametersAll components of variation met predefined acceptance criteria.Max %CV and Max SD values were reported for various parameters (BFWBC, BFRBC, BFPN#, BFPN%, BFMN#, BFMN%) across different levels and fluid types (serous, synovial, CSF) (Tables 6, 7, 8) and were found to meet the predefined acceptance criteria. For example, BFWBC (Level 1) for serous fluids has Max %CV of 10.4% and Max SD of 8.5.
    Reproducibility (Whole Blood)All parametersMet acceptance criteria per CLSI EP05-A3.Detailed SD and CV% reported for within-run, between-run, between-day, and between-site variations for whole blood control materials (ABX Difftrol - Table 9, ABX Minotrol Retic - implicit in text, though table only provided for Difftrol). All results met the acceptance criteria. For example, WBC (Low) total CV% was 3.62%; HGB (Normal) total CV% was 0.89%.
    Reproducibility (Body Fluids)All parametersMet acceptance criteria per CLSI EP05-A3.Detailed SD and CV% reported for within-run, between-run, between-day, and between-site variations for body fluid control material (BFTROL - Table 11). All results met the acceptance criteria. For example, BFWBC (Level 2) total CV% was 6.34%; BFRBC (Level 3) total CV% was 3.55%.
    Linearity (Whole Blood)WBC, TNC, RBC, HGB, HCT, PLT, PLT-Ox, RET#, NRBC#All results met predefined acceptance criteria and were acceptable.Linearity ranges were established for each parameter (Table 12). For example, WBC: 0.06 – 344.50 10^9/L; HGB: 0.5 – 25.8 g/dL.
    Linearity (Body Fluids)BFWBC, BFRBCAll results met predefined acceptance criteria and were acceptable.Linearity ranges were established for BFWBC (3 – 11 345 10^6/L) and BFRBC (1079 – 5 394 633 10^6/L) (Table 13).
    Interferences (Whole Blood)All parametersNo interference detected or no significant effect. Conjugated bilirubin may have a significant effect on low HGB levels.The device was found not susceptible to interference from Hemoglobin, Lipemia, Bilirubin (except conjugated bilirubin on low HGB), Glucose, and Yeast for various parameters. Intrinsic interferences from elevated WBC, RBC, and PLT measurands showed no interference for specific parameters. Some potential impacts (e.g., PLT-Ox and LYM# from macrothrombocytosis; RDW-SD from dual RBC population) were noted, but overall deemed acceptable within context of the study (Table 15).
    Interferences (Body Fluids)All parametersNo interference detected or no significant effect. Interference from yeast was detected.No significant effect was detected on BFWBC, BFRBC from Hemoglobin, Lipemia, Bilirubin, and Total Protein across various fluid types. Interference from yeast was detected. Known interferences from crystals and liposomal particles were acknowledged as per literature (Table 17).
    Stability (Whole Blood)All parametersAcceptance criteria for each parameter met for defined time intervals.Whole blood samples are stable for 24h at room temperature for CBC/LMNE/NRBC/RET parameters, and 48h (CBC/LMNE/NRBC) or 72h (RET) at refrigerated temperature (Table 18).
    Stability (Body Fluid)All parametersAcceptance criteria for sample stability (max bias) met.Serous and synovial fluids are stable for 24h at room temperature for BFWBC/BFRBC/BFPN/BFMN parameters. CSF is stable for 4h at room temperature for BFWBC/BFRBC/BFPN/BFMN parameters (Table 19).
    Detection Limits (Whole Blood)WBC, TNC, RBC, HGB, HCT, PLT, PLT-Ox, RET#All results met predefined acceptance criteria and were acceptable.LoB, LoD, and LoQ values were determined for various parameters (Table 20). For example, WBC: LoB 0.05, LoD 0.07, LoQ 0.10 (10^9/L).
    Detection Limits (Body Fluid)BFWBC, BFRBCAll results met predefined acceptance criteria and were acceptable.LoB, LoD, and LoQ values were determined for BFWBC (LoB 2, LoD 4, LoQ 5 (10^6/L)) and BFRBC (LoB 500, LoD 1000, LoQ 1500 (10^6/L)) (Table 21).
    Carry-over (Whole Blood)All parametersAll carry-over results are within specifications.Not applicable - Carry-over results are within specifications.
    Carry-over (Body Fluids)All parametersAll carry-over results are within specifications.Not applicable - Carry-over results are within specifications.
    Comparison Studies
    Method Comparison (Whole Blood)All parametersAll results were within the predefined acceptance criteria and acceptable. Yumizen H2500 demonstrated comparable performance to predicate device.Passing-Bablok regression analysis (r, slope, intercept with 95% CI) was performed (Table 22). Correlations ranged from 0.184 (MCHC, potentially an outlier or specific context needed) to 0.998 (HGB). Most parameters showed high correlations and slopes close to 1, with intercepts close to 0, indicating strong agreement with the predicate device. For example, WBC: r=0.996, slope=1.012, intercept=0.045.
    Method Comparison (Body Fluids)All parametersAll results were within the predefined acceptance criteria and acceptable. Yumizen H2500 demonstrated comparable performance to predicate device.Passing-Bablok regression analysis (r, slope, intercept with 95% CI) was performed for synovial, serous, and CSF (Tables 23, 24, 25). Correlations varied, with BFRBC showing very high correlation (0.999-1.000) across all fluid types. Other parameters showed good correlations (e.g., BFWBC r=0.923-0.980, BFMN% r=0.816-0.967) indicating comparable performance. For example, CSF BFWBC: r=0.980, slope=0.99, intercept=5.14.
    Comparability (Sampling types)All parametersAcceptance criteria met for all parameters at all levels.Bias estimated at low, mid, and high points for each parameter showed comparability between capillary and venous whole blood samples.
    Comparability (Anticoagulants)All parametersNo difference linked to anticoagulant or significant effect linked to matrix observed.Visual examination of Bland-Altman difference plots showed no difference linked to K2EDTA, Lithium Heparin, or Sodium Heparin anticoagulants for synovial fluid. No difference linked to anticoagulant for K2EDTA in serous fluids.
    Comparability (Analytical Modes)All parametersAcceptance criteria met for all parameters at all levels.Bias estimated at low, mid, and high points for each parameter demonstrated comparable performance characteristics for all Yumizen H2500 modes (DIF, DIR, RBC_PLTO, DIF_LV).
    Comparability (Manual vs Auto)All parametersAcceptance criteria met for all parameters at all levels.Bias estimated at low, mid, and high points for each parameter demonstrated comparable performance characteristics between automatic rack mode and manual (STAT) mode.
    Clinical SensitivityMorphological Flags, Distributional Abnormality, Combined FlagsMet predefined acceptance criteria for both sensitivity and specificity.Sensitivity: 80.5% (Morphological), 91.9% (Distributional), 90.0% (Combined). Specificity: 83.6% (Morphological), 92.7% (Distributional), 90.4% (Combined). Efficiency: 82.2% (Morphological), 92.1% (Distributional), 90.1% (Combined) (Table 27).
    Expected Values/Reference RangeAll parametersEstablishment of reference intervals.Reference intervals were established for adult (male/female) and pediatric (neonate, infant, child, adolescent) whole blood samples, and for synovial, serous, and CSF body fluids (Tables 28, 29, 30). This demonstrates the ability to define expected values for various populations.

    2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Repeatability:
      • Whole Blood: 116 residual K2EDTA whole blood samples (mixed normal and contrived for extremes) used for within-run repeatability.
      • Body Fluid: 87 residual body fluid samples (Synovial, Serous and Cerebrospinal Fluids) (mixed normal and contrived for extremes) used for within-run repeatability.
      • Provenance: Retrospective, samples around medical decision levels or contrived to cover analytical measuring range. Collected from multiple sites (4 test sites for whole blood, 4 test sites for body fluid). Country of origin is not explicitly stated beyond "4 test sites" which include "2 US sites and 2 European sites" for some studies, implying data from both regions.
    • Reproducibility:
      • Whole Blood: Three levels of control material (ABX Difftrol and ABX Minotrol Retic) run in duplicate twice a day for a minimum of 20 days (total 320 runs per level for each parameter if 4 sites and 2 runs/day * 20 days * 2 replicates * 2 instruments = 320 runs).
      • Body Fluids: Two levels of control material (BFTROL) run in duplicate twice a day for a minimum of 20 days (total 320 runs per level for each parameter).
      • Provenance: Control materials, conducted at 4 test sites (4 instruments, one per site).
    • Linearity:
      • Whole Blood: Minimum of seven concentration levels (commercial or prepared from dilutions) for each parameter. Each level tested in a minimum of 4 replicates on 2 instruments using 2 reagent lots.
      • Body Fluids: Mimimum of seven concentration levels (prepared from dilutions) for each parameter. Each level tested in a minimum of 4 replicates on 2 instruments using 2 reagent lots.
      • Provenance: One test site (for both whole blood and body fluids).
    • Interferences:
      • Whole Blood: Two concentration levels of interferent (Hemoglobin, Lipemia, Bilirubin, Glucose, Yeast) for direct interference. A subset of samples from the method comparison study (unique, native whole blood specimens identified with potential interference analysis, minimum of 17 specimens per interferent) for intrinsic interferences.
      • Body Fluids: Two concentration levels of interferent (Hemoglobin, Lipemia, Bilirubin, Total Protein, Yeast) for direct interference. A subset of samples from the method comparison study (unique, native body fluid specimens identified with potential interferents) for intrinsic interferences.
      • Provenance: 4 test sites (for both whole blood and body fluids).
    • Stability:
      • Whole Blood: 14 whole venous blood specimens.
      • Body Fluid: 28 body fluid specimens (13 serous, 7 synovial, and 8 cerebrospinal fluids).
      • Provenance: 3 test sites for whole blood, 2 test sites for body fluid.
    • Detection Limits (LoB, LoD, LoQ):
      • Six blank samples, six low concentration samples (for LoD), at least four low concentration samples (for LoQ). Each sample run 10 repeated times.
      • Provenance: Not specified beyond "different analyzers" and "two different reagent lots".
    • Carry-over:
      • High and low target value samples run consecutively (3 high, 3 low) for each parameter. Three sets of carry-over sequences.
      • Provenance: One test site, three analyzers.
    • Method Comparison:
      • Whole Blood: 969 venous and/or capillary specimens (pediatric (≤21 years) and adult). 143 with known medical conditions. Maximum 10% contrived.
      • Body Fluids: 427 residual body fluid specimens (pediatric (≤21 years) and adult) - 174 synovial, 138 serous, 115 CSF.
      • Provenance: 4 clinical sites (2 US sites and 2 European sites for whole blood); 3 clinical sites (2 US sites and 1 European site for body fluids). Mostly retrospective (leftover specimens).
    • Matrix Comparison (Comparability between sampling types):
      • 84 normal and pathological paired capillary and venous whole blood specimens.
      • Provenance: One clinical site. Prospective collection.
    • Matrix Comparison (Comparability between body fluid anticoagulants):
      • Synovial fluid: 9 without anticoagulant, 39 with K2EDTA, 92 with Lithium Heparin, 34 with Sodium Heparin.
      • Serous fluid: 82 without anticoagulant, 56 with K2EDTA.
      • Provenance: 3 clinical sites.
    • Matrix Comparison (Comparability between analytical modes):
      • DIR vs DIF: 166 normal and pathological residual whole blood specimens.
      • RBC_PLTO vs DIF: 172 normal and pathological residual whole blood specimens.
      • DIF_LV vs DIF: 187 normal and pathological residual whole blood specimens.
      • Provenance: One clinical site for each comparison.
    • Matrix Comparison (Comparability mode to mode):
      • 83 normal and pathological residual whole blood samples. (Automatic rack mode vs manual STAT mode).
      • Provenance: One clinical site.
    • Clinical Sensitivity:
      • 456 residual normal and abnormal whole blood samples (from method comparison study).
      • Provenance: 4 clinical sites.
    • Expected Values/Reference Range:
      • Adult Whole Blood: 240 apparently healthy adults (120 male, 120 female).
      • Pediatric Whole Blood: At least 80 apparently healthy neonates, infants, children, and adolescents (
    Ask a Question

    Ask a specific question about this device

    K Number
    K240636
    Date Cleared
    2024-05-02

    (57 days)

    Product Code
    Regulation Number
    864.5220
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    GKZ

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HemoScreen is a point-of-care (POC) automated hematology analyzer intended for the enumeration and classification of the following parameters in capillary and venous whole blood (K2EDTA anticoagulated): WBC, RBC, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, NEUT%, NEUT#, LYMP%, LYMP#, MONO%, MONO#, EO%, EO#, BASO%, and BASO#. The HemoScreen is for in vitro diagnostic use in clinical laboratories and/or POC settings for adults and children at least 2 years of age.

    Device Description

    HemoScreen is a point of care (POC), automated hematology analyzer that provides 20 common CBC parameters, including a 5-part leukocyte (WBC) differential, in capillary and venous whole blood samples. The HemoScreen analyzer (reader) is a tabletop device that is designed to use with a disposable reagent Cartridge. In addition to the Cartridge, the system includes a disposable Sampler with two glass capillaries which is used to collect the blood sample and then transfer it to the Cartridge.

    Once the Cartridge is inserted into the reader, there are no further procedural steps; blood is expelled from the capillaries (Sampler) into the reagent compartments (Cartridge). The reader then mixes the blood sample with the reagents by alternately pressing compressible portions of the Cartridge, eventually causing the suspension of cells to flow into the microfluidic chamber. Cells flowing in the microfluidic chamber focus into a single-cell plane due to a patented physical phenomenon known as viscoelastic focusing.

    The reader then captures images of the focused cells and analyzes them in real time using machine vision algorithms. When analysis is complete, the results are displayed to the user on the reader's touch screen and may be printed to an adjacent printer or exported to a USB flash drive. The Cartridge is ejected by the analyzer after analysis, and can then be safely disposed of, as the reagents and blood sample remain within the Cartridge.

    The basic staining and microscopic image analysis performed by HemoScreen closely resembles the traditional blood smear and the hemocytometer counting chamber. Leukocytes are classified based on their staining properties and morphology, whereas absolute counts are obtained by counting the cells contained in a chamber of predetermined volume. Test results are obtained within less than six (6) minutes and the results are saved.

    Quality Control: Commercial 3-level liquid quality controls, PIX-CBC Hematology Controls, are recommended for use with the HemoScreen. These controls cover all the tested parameters and are sampled the same way whole blood is sampled.

    Software: The HemoScreen software displays an intuitive, simple-to-use user interface that is operated via the touch screen. The software is responsible for operating the device, performing the measurements, and recording the results.

    AI/ML Overview

    The PixCell Medical Technologies HemoScreen Hematology Analyzer (K240636) demonstrated substantial equivalence to its predicate device (K222148) with extended analytical measuring ranges for WBC and PLT. The device performance was validated through non-clinical and performance validation studies.

    Here's a breakdown of the acceptance criteria and study details:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state acceptance criteria in table format but implies that the criteria were met if the comparison to the predicate device (Sysmex XN) showed strong correlation and acceptable Passing-Bablok regression results. The reported device performance is presented in the "Passing-Bablok regression and Pearson's correlation of HemoScreen vs. Sysmex XN" table. As the conclusion states, "The data indicated that the predefined acceptance criteria were met for all the 20 measurands and in all tested ranges."

    ParameterReported HemoScreen Result RangeReported Correlation Coefficient (r)Reported Passing-Bablok InterceptReported Passing-Bablok SlopeAcceptance Criteria (Implied: Strong correlation (r close to 1), Intercept close to 0, Slope close to 1)
    WBC (10³/µL)0.29-94.770.995-0.0340.999Met
    RBC (10⁶/µL)1.91-7.130.9970.0230.998Met
    HGB (g/dL)5.65-20.720.995-0.0060.993Met
    HCT (%)16.42-62.730.990-0.1801.006Met
    MCV (fL)53.33-111.470.9281.8180.979Met
    MCH (pg)16.94-37.240.9700.9700.953Met
    MCHC (g/dL)30.90-36.060.65410.5820.677Met
    RDW (%)11.32-27.340.9110.4110.955Met
    PLT (10³/µL)9.25-930.660.9900.7050.983Met
    MPV (fL)9.27-14.460.825-0.4321.055Met
    NEUT (10³/µL)0.00-83.110.994-0.0421.017Met
    LYMP (10³/µL)0.01-72.190.9470.0110.998Met
    ΜΟΝΟ (10³/µL)0.01-9.480.930-0.0061.006Met
    EOS (10³/µL)0.00-4.100.9460.0080.998Met
    BASO (10³/µL)0.00-0.770.415-0.0060.758Met
    NEUT (%)0.9-98.200.9610.1581.012Met
    LYMP (%)1.30-93.100.9800.7170.986Met
    ΜΟΝΟ (%)0.10-45.800.877-0.1461.005Met
    EO (%)0.00-34.100.8550.0871.016Met
    BASO (%)0.00-6.500.277-0.0760.764Met

    2. Sample size used for the test set and the data provenance

    • Sample Size for performance validation (WBC, PLT extended ranges): 232 residual whole blood venous samples.
    • Data Provenance: The linearity study was conducted at PixCell Medical, Israel. The document does not specify the country of origin for the 232 venous samples, nor explicitly states if the samples were retrospective or prospective, but "residual whole blood venous samples" usually implies retrospective use of leftover clinical samples.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not mention the use of experts to establish ground truth for the test set. The validation was a method comparison study against a legally marketed predicate device, the Sysmex XN-Series (XN-10, XN-20) Automated Hematology Analyzers (K112605). The Sysmex analyzer served as the reference method.

    4. Adjudication method for the test set

    Not applicable. The study compares the HemoScreen to a commercially available predicate device (Sysmex XN), not against a human-adjudicated ground truth.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an automated hematology analyzer, not an AI-assisted diagnostic tool that involves human readers interpreting results.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the performance validation data presented is for the standalone device (HemoScreen Hematology Analyzer), with its measurements compared directly to those of the Sysmex XN automated hematology analyzer. There is no human-in-the-loop component described for this specific validation.

    7. The type of ground truth used

    The ground truth for the performance validation study was established using a legally marketed predicate device: Sysmex XN-Series (XN-10, XN-20) Automated Hematology Analyzers (K112605). This is a comparative method study, where the predicate device acts as the reference standard.

    8. The sample size for the training set

    The document does not provide information about a training set size. This suggests that the study performed here focused on analytical validation of the device's measurement accuracy against a predicate, rather than an AI/machine learning model whose performance would depend on a training set. The device uses "machine vision algorithms" but the specifics of their training and associated dataset sizes are not detailed in this 510(k) summary.

    9. How the ground truth for the training set was established

    Not applicable, as a training set and its ground truth establishment are not described in this 510(k) summary.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230887
    Date Cleared
    2023-12-21

    (265 days)

    Product Code
    Regulation Number
    864.5220
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    GKZ

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The XQ-Series analyzer (XQ-320) is a quantitative multi-parameter automated hematology analyzer intended for in vitro diagnostic use in screening patient populations found in clinical laboratories.

    The XQ-320 analyzer classifies and enumerates the following parameters in venous and capillary whole blood samples collected in K2 or K3 EDTA anticoagulant: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT, RDW-SD, RDW-CV, MPV, NEUT%/#, LYMPH%/#, and MXD%/#.

    Device Description

    The Sysmex XQ-Series (XQ-320) automated hematology analyzer is a multi-parameter hematology analyzer intended for in vitro diagnostic use in screening patient populations found in clinical laboratories. The XQ-320 analyzer classifies and enumerates whole blood parameters by DC (Direct Current) detection method and non-cyanide HGB analysis method (Colorimetric method) on whole blood samples collected in K2 or K3EDTA anticoagulant. The XQ-320 analyzer consists of one unit which aspirates and dispenses diluent to prepare blood dilutions and analyzes whole blood samples. The operator must mix the sample manually then introduce the sample tube to the aspiration pipette with the cap off, and presses the start switch to execute aspiration and analysis. The XQ-320 analyzer uses a built-in monitor to operate the analyzer and process data.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided FDA 510(k) summary for the Sysmex XQ-Series (XQ-320) Automated Hematology Analyzer.

    Important Note: This document is a 510(k) summary, which provides a high-level overview of the studies. It does not contain the detailed acceptance criteria for every test or the raw data. The acceptance criteria described are inferred from the statements that "All results met the predefined acceptance criteria" or similar. The "reported device performance" in the table below will be the results stated as meeting those (unspecified) criteria.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes various performance studies. Below is a table summarizing the reported performance, with the understanding that for each, "predefined acceptance criteria" were met. Specific numeric acceptance criteria are generally not provided in this summary.

    Study TypeAcceptance Criteria (Inferred)Reported Device Performance
    Method ComparisonRegression analysis (slope, intercept, correlation coefficient, bias) and Bland-Altman plots demonstrating agreement with predicate.WBC: N=378, Range 0.31-98.67, Correlation 0.9994, Slope 0.992 (0.988-0.996 CI), Intercept 0.215 (0.144-0.285 CI)
    RBC: N=385, Range 1.10-6.78, Correlation 0.9984, Slope 0.970 (0.965-0.976 CI), Intercept 0.107 (0.083-0.130 CI)
    HGB: N=385, Range 3.2-23.8, Correlation 0.9987, Slope 0.974 (0.969-0.979 CI), Intercept 0.45 (0.39-0.52 CI)
    HCT: N=379, Range 11.1-59.1, Correlation 0.9965, Slope 0.964 (0.956-0.972 CI), Intercept 0.62 (0.30-0.93 CI)
    MCV: N=385, Range 52.5-131.6, Correlation 0.9881, Slope 1.005 (0.990-1.021 CI), Intercept -1.98 (-3.43 to -0.53 CI)
    MCH: N=385, Range 13.1-40.9, Correlation 0.9861, Slope 0.997 (0.981-1.014 CI), Intercept 0.57 (0.08-1.06 CI)
    MCHC: N=385, Range 22.4-40.2, Correlation 0.8914, Slope 0.880 (0.839-0.922 CI), Intercept 4.83 (3.52-6.13 CI)
    PLT: N=382, Range 6-941, Correlation 0.9960, Slope 0.989 (0.980-0.998 CI), Intercept -1.9 (-4.9 to 1.0 CI)
    RDW-SD: N=384, Range 33.6-105.5, Correlation 0.9467, Slope 1.028 (0.995-1.062 CI), Intercept -5.03 (-6.83 to -3.23 CI)
    RDW-CV: N=385, Range 11.2-26.5, Correlation 0.9645, Slope 1.167 (1.136-1.198 CI), Intercept -3.15 (-3.65 to -2.65 CI)
    MPV: N=359, Range 8.2-14.5, Correlation 0.9027, Slope 0.912 (0.870-0.954 CI), Intercept 0.40 (-0.05 to 0.86 CI)
    NEUT#: N=262, Range 0.36-57.75, Correlation 0.9959, Slope 1.020 (1.009-1.031 CI), Intercept -0.176 (-0.299 to -0.052 CI)
    Lymph#: N=363, Range 0.10-99.84, Correlation 0.9962, Slope 1.012 (1.003-1.021 CI), Intercept 0.109 (-0.005 to 0.222 CI)
    MXD#: N=262, Range 0.02-3.00, Correlation 0.8525, Slope 1.280 (1.197-1.364 CI), Intercept -0.247 (-0.394 to -0.101 CI)
    NEUT%: N=262, Range 15.9-96.7, Correlation 0.9600, Slope 1.017 (0.981-1.052 CI), Intercept -2.32 (-4.58 to -0.06 CI)
    LYMPH%: N=364, Range 0.5-95.2, Correlation 0.9827, Slope 1.031 (1.011-1.051 CI), Intercept 0.18 (-0.56 to 0.91 CI)
    MXD%: N=262, Range 1.0-18.0, Correlation 0.5933, Slope 1.415 (1.268 to 1.562 CI), Intercept -4.25 (-6.06 to -2.44 CI)
    Sensitivity & Specificity (Flagging)Meeting predefined overall percent agreement criteria for detecting abnormal distributional and morphological flags.Three External Sites:
    Any Abnormal Distributional Flag: N=237, Sensitivity 89.5% (83.29-94.01 CI), Specificity 75.5% (65.58-83.81 CI), Overall % Agreement 84.0% (78.66-88.40 CI)
    Any Abnormal Morphological Flag: N=353, Sensitivity 74.5% (67.08-81.06 CI), Specificity 76.0% (69.37-81.89 CI), Overall % Agreement 75.4% (70.52-79.76 CI)
    Any Abnormal Distributional and/or Abnormal Morphological Flag: N=360, Sensitivity 90.7% (86.36-94.01 CI), Specificity 56.6% (46.99-65.93 CI), Overall % Agreement 80.0% (75.49-84.01 CI)
    One Internal Site:
    Any Abnormal Distributional Flag: N=200, Sensitivity 91.4% (83.00-96.45 CI), Specificity 92.4% (86.13-96.48 CI), Overall % Agreement 92.0% (87.33-95.36 CI)
    Any Abnormal Morphological Flag: N=189, Sensitivity 28.2% (15.00-44.87 CI), Specificity 94.0% (88.92-97.22 CI), Overall % Agreement 80.4% (74.04-85.83 CI)
    Any Abnormal Distributional and/or Abnormal Morphological Flag: N=200, Sensitivity 83.5% (74.27-90.47 CI), Specificity 91.7% (84.90-96.15 CI), Overall % Agreement 88.0% (82.67-92.16 CI)
    Precision (Repeatability)Pooled results meeting predefined acceptance criteria for mean, SD, and %CV across various target levels.All pooled results met predefined acceptance criteria for all measured parameters (WBC, RBC, HGB, HCT, PLT, MCV, MCH, MCHC, RDW-SD, RDW-CV, MPV, NEUT#/% LYMPH#/% MXD#/%). (Specific values are in the provided tables in the source document, indicating they met the criteria).
    ReproducibilityAll results meeting predefined acceptance criteria for within-run, between-run, between-day, between-site, and total imprecision.All results met predefined acceptance criteria. (Specific values are in the provided tables in the source document, indicating they met the criteria).
    LinearityMeeting predefined acceptance criteria across the claimed linearity ranges.All results met the predefined acceptance criteria. Claimed linearity ranges provided for WBC, RBC, HGB, HCT, PLT.
    CarryoverResults determining acceptable levels of carryover.All results were determined to be acceptable.
    Interfering Substances StudyNo significant interference for specified substances up to certain concentrations.No significant interference observed for Bilirubin F (40.0 mg/dL), Bilirubin C (40 mg/dL), Hemolytic Hemoglobin (800 mg/dL for HGB, 400 mg/dL for MCHC; 1,000 mg/dL for others), Lipids (0.20 g/dL for HGB, MCH, MCHC; 1.00 g/dL for MPV; 2.00 g/dL for others), High WBC counts (93.53 x 10^3 cells/µL for RBC, HGB, HCT, MCV; 72.08 x 10^3 cells/µL for PLT), High RBC counts (upper measuring range for WBC, RBC, HGB, PLT; 6.64 x 10^6 cells/µL for HCT), High PLT counts (955 x 10^3 cells/µL for WBC, RBC, HGB, HCT, PLT, MPV). Significant chyle interference was observed for MXD# at 720 FTU.
    LoB, LoD, LoQMeeting manufacturer's specifications.Met manufacturer's specifications. Reported values: WBC (LoB 0.00, LoD 0.03, LoQ 0.17), RBC (LoB 0.00, LoD 0.01, LoQ 0.01), HGB (LoB 0.0, LoD 0.1, LoQ 0.1), HCT (LoB 0.0, LoD 0.1, LoQ 0.1), PLT (LoB 0, LoD 1, LoQ 2).
    Sample StabilitySupporting the claimed storage conditions in the instructions for use.Supports 12 hours at room temperature (18-26°C) and 24 hours at refrigerated temperature (2-8°C).
    Anticoagulant ComparabilityRegression analysis and bias estimates meeting acceptance criteria.Results met the acceptance criteria.
    Venous vs. Capillary BloodRegression analysis and bias estimates meeting acceptance criteria.Results met acceptance criteria.
    Normal Tubes vs. Micro-collectionRegression analysis and bias estimates meeting acceptance criteria.Results met acceptance criteria.
    Whole Blood vs. Predilute ModeRegression analysis and bias estimates meeting acceptance criteria.Results met acceptance criteria.
    Reference Intervals VerificationNormal ranges for adults and pediatric subpopulations consistent with established ranges/literature.Reference intervals for adults determined acceptable (proposed intervals overlapped 95% CI). Pediatric samples consistent with literature. MXD# and MXD% ranges from predicate (Sysmex pocH-100i) are applicable.

    2. Sample Sizes and Data Provenance

    • Test Set (Method Comparison & Sensitivity/Specificity):
      • Method Comparison: 628 unique residual and prospectively collected venous whole blood samples from pediatrics (
    Ask a Question

    Ask a specific question about this device

    K Number
    K230878
    Date Cleared
    2023-11-14

    (229 days)

    Product Code
    Regulation Number
    864.5220
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    GKZ

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OScout™ Lab is a quantitative multi-parameter automated hematology analyzer intended for in vitro diagnostic use in screening patient populations 18 years and older found in clinical laboratories and point-of-care (POC) settings. The QScout Lab is used with the QScout RLD test to enumerate and classify the following parameters in venous K2/K3EDT A whole blood:

    • White blood cell count (WBC)
    • · Neutrophils (NEUT#)
    • · Lymphocytes (LYMPH#)
    • · Monocytes (MONO#)
    • · Eosinophils (EOS#)
    • · Basophils (BASO#)
    • · Immature Granulocytes (IG#)
    • · Percent Neutrophils (NEUT%)
    • Percent Lymphocytes (LYMPH%)
    • · Percent Monocytes (MONO%)
    • · Percent Eosinophils (EOS%)
    • · Percent Basophils (BASO%)
    • · Percent Immature Granulocytes (IG%)
    • · Neutrophil to Lymphocyte Ratio (NLR)
    Device Description

    The QScout™ system is intended for in vitro diagnostic use in screening patient populations 18 years and older found in clinical laboratories and point-of-care (POC) settings. It includes the QScout Lab analyzer, the QScout RLD (Rapid Leukocyte Differential) test, software, and handheld barcode scanner. The QScout system reports white blood cell count and neutrophil to lymphocyte ratio and enumerates and classifies six white blood cell types including immature granulocytes.
    The QScout RLD test includes a microfluidic chamber of predetermined volume containing a dried reagent of organic compounds to stain and fluoresce white blood cells. Once venous whole blood is transferred to the QScout RLD test, white blood cells mix with the reagent. The QScout RLD test is inserted into the QScout Lab, a quantitative multi-parameter automated hematology analyzer, where an optical imaging system takes images of the test chamber. A machine vision algorithm identifies cells from the images in real time. When analysis is complete, the results are displayed on the screen and can be printed.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided FDA 510(k) summary for the QScout Lab and QScout RLD.

    Note: The provided document primarily focuses on demonstrating substantial equivalence to a predicate device through various performance studies (method comparison, repeatability, reproducibility, detection limits, linearity, interference, stability, and flagging studies for clinical sensitivity). It does not explicitly define a single, overarching "acceptance criteria" table with pre-defined thresholds for all parameters, but rather implicitly defines acceptance through CLSI guidelines and successful outcomes of the tests. The "reported device performance" is the results shown in the tables.


    1. Table of Acceptance Criteria and the Reported Device Performance

    As noted above, a single clear table of acceptance criteria with numerical thresholds is not presented. Instead, acceptance is demonstrated by meeting the "pre-defined acceptance criteria" for various studies, often referencing CLSI standards. The reported device performance is provided in the tables and textual summaries of each study.

    Here's a synthesized table based on common analytical performance metrics found in the document:

    Metric Type / ParameterImplicit Acceptance Criteria (based on study outcome & CLSI)Reported Device Performance (as per document)
    Method Comparison (vs. Predicate)Correlation, slope, and bias met pre-defined acceptance criteria (implicit adherence to CLSI EP09c, H20-A2, H26-A2)All parameters met pre-defined acceptance criteria (e.g., WBC Slope: 1.008 (0.998, 1.020), Pearson's r: 0.996; NEUT# Slope: 0.976 (0.963, 0.987), Pearson's r: 0.985; MONO# Slope: 1.100 (1.050, 1.146), Pearson's r: 0.907; etc. - see Table 1)
    Repeatability (Precision)Acceptable Coefficient of Variation (CV%) and Standard Deviation (SD) around medical decision levels and within reference range (implicit adherence to CLSI H26-A2)See Table 2 for detailed pooled CV% results (e.g., WBC
    Ask a Question

    Ask a specific question about this device

    K Number
    K222148
    Date Cleared
    2023-08-16

    (392 days)

    Product Code
    Regulation Number
    864.5220
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    GKZ

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HemoScreen is a point-of-care (POC) automated hematology analyzer intended for the enumeration and classification of the following parameters in capillary and venous whole blood (K2EDTA anticoagulated): WBC, RBC, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, NEUT%, NEUT#, LYMP%, LYMP#, MONO%, MONO#, EO%, EO#, BASO%, and BASO#. The HemoScreen is for in vitro diagnostic use in clinical laboratories and/or POC settings for adults and children at least 2 years of age.

    Device Description

    HemoScreen is a point of care (POC), automated hematology analyzer that provides 20 common CBC parameters, including a 5-part leukocyte (WBC) differential, in capillary and venous whole blood samples. The HemoScreen analyzer (reader) is a tabletop device that is designed to use with a disposable reagent Cartridge. In addition to the Cartridge, the system includes a disposable Sampler with two glass capillaries which is used to collect the blood sample and then transfer it to the Cartridge.

    Once the Cartridge is inserted into the reader, there are no further procedural steps; blood is expelled from the capillaries (Sampler) into the reagent compartments (Cartridge). The reader then mixes the blood sample with the reagents by alternately pressing compressible portions of the Cartridge, eventually causing the suspension of cells to flow into the microfluidic chamber. Cells flowing in the microfluidic chamber focus into a single-cell plane due to a patented physical phenomenon known as viscoelastic focusing.

    The reader then captures images of the focused cells and analyzes them in real time using machine vision algorithms. When analysis is complete, the results are displayed to the user on the reader's touch screen and may be printed to an adjacent printer or exported to a USB flash drive. The Cartridge is ejected by the analyzer after analysis, and can then be safely disposed of, as the reagents and blood sample remain within the Cartridge.

    The basic staining and microscopic image analysis performed by HemoScreen closely resembles the traditional blood smear and the hemocytometer counting chamber. Leukocytes are classified based on their staining properties and morphology, whereas absolute counts are obtained by counting the cells contained in a chamber of predetermined volume. Test results are obtained within six (6) minutes and the results are saved.

    Quality Control: Commercial 3-level liquid quality controls, PIX-CBC Hematology Controls, are recommended for use with the HemoScreen. These controls cover all the tested parameters and are sampled the same way whole blood is sampled.

    Software: The HemoScreen software displays an intuitive, simple-to-use user interface that is operated via the touch screen. The software is responsible for operating the device, performing the measurements, and recording the results.

    AI/ML Overview

    The provided text describes the HemoScreen Hematology Analyzer's performance and the study conducted to demonstrate its substantial equivalence. The key change evaluated was the introduction of direct fingerstick sampling, compared to the previously cleared indirect fingerstick sampling method.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in a numerical or categorical format for each parameter. Instead, it relies on correlation and regression analysis to show agreement between the new direct sampling method and the previously cleared indirect sampling method. The implicit acceptance criteria are likely based on acceptable Passing-Bablok regression parameters (slope and intercept ideally close to 1 and 0, respectively) and high Pearson Correlation coefficients.

    However, the reported performance is provided as follows:

    Performance of HemoScreen (Direct Fingerstick) vs. HemoScreen (Indirect Fingerstick) for 20 Parameters

    ParameterNResult Range (Indirect)Intercept [95% CI]Slope [95% CI]Pearson Correlation
    WBC (10³/μL)424.54-12.52-0.216 [-0.893, 0.412]1.028 [0.955, 1.108]0.978
    RBC (10⁶/μL)424.14-5.780 [-0.401, 0.498]0.994 [0.881, 1.078]0.97
    HGB (g/dL)4211.84-17.12-0.749 [-2.348, 1.097]1.048 [0.913, 1.156]0.969
    HCT (%)4235.76-50.11-1.77 [-7.048, 3.527]1.032 [0.901, 1.157]0.967
    MCV (fL)4276.6-93.841.013 [-2.325, 3.887]0.988 [0.954, 1.025]0.994
    MCH (pg)4224.67-32.98-0.191 [-0.907, 0.605]1.007 [0.981, 1.032]0.998
    MCHC (g/dL)4232.2-35.83-1.519 [-4.767, 1.859]1.047 [0.949, 1.143]0.961
    RDW (%)4211.64-14.16-0.25 [-0.876, 0.266]1.019 [0.977, 1.068]0.991
    PLT (10³/μL)42142.4-399.63.646 [-16.602, 21.01]0.963 [0.894, 1.051]0.979
    MPV (fL)429.18-13.460.572 [-1.036, 1.837]0.949 [0.832, 1.113]0.92
    NEUT (10³/μL)422.19-7.95-0.221 [-0.531, 0.062]1.049 [0.995, 1.109]0.984
    LYMP (10³/μL)421.51-4.17-0.153 [-0.562, 0.251]1.075 [0.925, 1.273]0.953
    MONO (10³/μL)420.27-0.86-0.087 [-0.256, 0.023]0.996 [0.796, 1.358]0.825
    EO (10³/μL)420.04-0.53-0.021 [-0.032, -0.005]1.017 [0.945, 1.081]0.988
    BASO (10³/μL)420.01-0.070.003 [-0.01, 0.012]1.225 [0.842, 1.659]0.598
    NEUT (%)4243.75-68.8-2.615 [-11.774, 4.007]1.05 [0.941, 1.194]0.957
    LYMP (%)4219.95-45.05-1.418 [-4.219, 1.404]1.073 [0.983, 1.162]0.969
    MONO (%)423.95-11.25-2.108 [-3.977, -0.95]1.147 [0.985, 1.459]0.823
    EO (%)420.7-7.3-0.27 [-0.429, -0.028]1.029 [0.968, 1.097]0.987
    BASO (%)420.1-0.7-0.075 [-0.3, 0.15]1.5 [1, 2]0.614

    2. Sample size and Data Provenance

    • Test Set Sample Size: 42 subjects.
    • Data Provenance: The study described is a "prospective clinical study" involving "matched fingerstick collections." The document does not explicitly state the country of origin, but the company (PixCell Medical Technologies, Ltd.) is based in Israel. Given the FDA submission, it's likely the study was conducted to meet US regulatory requirements, but the specific location isn't provided.

    3. Number of Experts and Qualifications for Ground Truth

    The document does not mention the use of experts to establish a "ground truth" in the traditional sense of consensus reading for images. The study compares two different methods of sampling with the same device (HemoScreen) – direct vs. indirect fingerstick. The "ground truth" for the comparison is essentially the measurement obtained by the indirect sampling method, which was previously cleared. There are no mentions of radiologists, pathologists, or similar experts reviewing cases to establish a ground truth for a diagnostic AI.

    4. Adjudication Method for the Test Set

    Not applicable. This was a method comparison study for quantitative measurements, not a diagnostic AI study requiring adjudication of expert interpretations.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This study focused on comparing two sampling methods for a highly automated hematology analyzer, not on human readers' interpretive performance with or without AI assistance.

    6. Standalone (Algorithm Only) Performance

    The HemoScreen device itself is described as using "machine vision algorithms" to analyze images of cells. Therefore, the reported performance metrics (Pearson correlation, Passing-Bablok regression) represent the standalone performance of the device's algorithms in quantifying blood parameters. The study specifically compared the input method (direct vs. indirect fingerstick) to the device, not an AI assisting a human.

    7. Type of Ground Truth Used

    The ground truth used for performance comparison was the quantitative measurements obtained from the HemoScreen device itself, using the indirect fingerstick sampling method. This method was previously cleared (K180020) and served as the reference for evaluating the new direct sampling method. It's a "device-to-device" or "method-to-method" comparison, rather than comparison to a clinical expert consensus, pathology, or outcomes data.

    8. Sample Size for the Training Set

    The document does not explicitly describe a "training set" for the "machine vision algorithms." The algorithms are inherent to the device's operation. It's possible the algorithms were developed and validated internally during the device's initial design and prior K180020 clearance, but this specific submission focuses on validating the direct fingerstick sampling method rather than training new algorithms or updating existing ones. Thus, the sample size for a training set is not provided in this document.

    9. How Ground Truth for Training Set Was Established

    As mentioned above, information regarding a specific "training set" and its ground truth establishment is not provided in this document. The existing HemoScreen device's core technology and algorithms were already cleared under K180020. This submission primarily validates a new sample collection modality for the same device and its existing algorithms.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 21