Search Filters

Search Results

Found 1223 results

510(k) Data Aggregation

    K Number
    K253188
    Manufacturer
    Date Cleared
    2025-10-24

    (28 days)

    Product Code
    Regulation Number
    864.7750
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CoaguChek XS Plus System is intended for use by professional healthcare providers for quantitative prothrombin time testing for the monitoring of warfarin therapy. The system uses fresh capillary or non-anticoagulated venous whole blood.

    Device Description

    The CoaguChek XS Plus system is a portable coagulation monitoring system to monitor prothrombin time (PT) in patients receiving oral anticoagulant therapy. The system uses the amperometric detection of thrombin in the blood sample. A test strip is used to determine a PT value from 8 µL of whole blood. Onboard quality control is available on every test strip and the system also features an optional external quality control material (CoaguChek XS PT Control).

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252580
    Device Name
    iQ200 Series
    Date Cleared
    2025-09-10

    (26 days)

    Product Code
    Regulation Number
    864.5200
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The iQ200 System is an in-vitro diagnostic device used to automate the complete urinalysis profile, including urine test strip chemistry panel, and microscopic sediment analysis. Optionally, the iQ200 Analyzer can be used as a stand-alone unit, or the results from the iQ200 analyzer can be combined with other urine chemistry results received from an LIS. It produces quantitative or qualitative counts of all formed sediment elements present in urine, including cells, casts, crystals, and organisms. A competent human operator can set criteria for auto-reporting and flagging specimens for review. All instrument analyte image decisions may be reviewed and overridden by a trained technologist.

    Device Description

    The iQ200 Series Automated Urine Microscopy system utilizes a specimen sandwiched between lamina layers presented to a microscope with a CCD video camera. This ensures the specimen is precisely within the microscope's focus and field of view. The system automates sample handling and analyte classification for improved data reporting and management. Specimens are aspirated by an autosampler, and individual particle images are isolated in each frame. The Auto-Particle Recognition (APR) software classifies images into 12 categories, and more, with 27 additional sub-classifications available. Particle concentration is determined by the number of images and the analyzed volume. Results are checked against user-defined criteria and sent for operator review or directly uploaded to the LIS. Specimen results can be edited, imported, and exported.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251404
    Manufacturer
    Date Cleared
    2025-08-25

    (111 days)

    Product Code
    Regulation Number
    864.5430
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The QStat Cartridge is a multi-channel cartridge that provides semi-quantitative indications of the coagulation and clot lysis state of a 3.2% citrated venous or arterial whole blood sample using the Quantra Hemostasis Analyzer. The QStat Cartridge includes tests to assess coagulation via the intrinsic and extrinsic pathways and includes a test with tranexamic acid to evaluate clot lysis characteristics.

    The QStat Cartridge is intended for in vitro diagnostic use by trained professionals at the point-of-care and in clinical laboratories to evaluate the viscoelastic properties of whole blood by means of the following functional parameters: Clot Time (CT), Clot Stiffness (CS), Fibrinogen Contribution to Clot Stiffness (FCS), Platelet Contribution to Clot Stiffness (PCS), and Clot Stability to Lysis (CSL).

    The QStat Cartridge is indicated for the evaluation of blood coagulation and clot lysis in patients age 18 years and older to assess possible hypocoagulable and hypercoagulable conditions in trauma, liver transplantation, and peripartum obstetric procedures.

    Results obtained with the QStat Cartridge should not be the sole basis for patient diagnosis.

    For prescription use only.

    Device Description

    The QStat Cartridge is a single-use, multi-channel disposable plastic cartridge used with the Quantra Hemostasis Analyzer for the evaluation of blood coagulation and clot lysis. The measurements are performed in four test channels of the disposable cartridge which enable differential testing with different sets of reagents without the need for any reagent preparation or controlled pipetting. The cartridge utilizes a citrated evacuated blood collection tube filled with a patient whole blood sample The proprietary technology SEER Sonorheometry measures the evolution of shear modulus (i.e., clot stiffness) in all four channels as a function of time. The QStat Cartridge is intended for use in patients 18 years or older by professionals in a hospital setting (point of care or laboratory) to assess possible hypocoagulable and hypercoagulable conditions in trauma and liver transplantation procedures.

    Each QStat Cartridge is pre-filled with lyophilized reagent beads individually sealed in an airtight pouch. After a QStat Cartridge is removed from its primary packaging, it is inserted into the instrument dock. A whole blood sample, collected in a 3.2% sodium citrate anticoagulant blood collection tube (minimum volume 2.7 mL), is attached directly to the cartridge and the test is initiated using the touch screen interface on the Quantra Hemostasis Analyzer. The cartridge is the only component of the Quantra System that is in direct contact with blood. The fluidic system within the instrument draws the sample into the cartridge where it is warmed to 37°C, aliquoted, introduced and mixed with the lyophilized reagents, and analyzed. When the test is complete, the cartridge is released from the dock to be disposed of in an appropriate biosafety sharps container.

    The analyzer displays the test results (n=5) in three different views: dial display screen, stiffness curves data, and trend screen. The dial display screen is the primary viewing screen and has a dial for each of the five output parameters. Each dial shows the reference range, assay measurement range, parameter abbreviation, and the numerical result for the corresponding parameter. The stiffness curves are a graphical display of shear modulus measurements over time that enable the user to view the development of clot stiffness over time. The trends screen displays results from a patient for up to six time points.

    There are two levels of external QStat Controls (QSL1 and QSL2) that are supplied separately (required but not provided materials) for testing on the Quantra System when changing cartridge lots, changing control lots, or after significant changes are made to the Quantra instrument (e.g., software update).

    AI/ML Overview

    The FDA 510(k) clearance letter for the Quantra QStat Cartridge describes its intended use for evaluating blood coagulation and clot lysis, specifically extending its indication to peripartum obstetric patients. The submission refers to non-clinical and clinical tests to demonstrate the device meets acceptance criteria.

    Here's a breakdown of the requested information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state quantitative acceptance criteria in a table format. However, it describes acceptable performance in relation to interference and clinical agreement.

    Acceptance Criteria (Implied from Study Description)Reported Device Performance
    No significant interference from common obstetric medications (hemabate, methergine, misoprostol, oxytocin)The highest concentration of each substance tested showed no significant interference in whole blood samples collected in 3.2% sodium citrate anticoagulant collection tubes.
    Clinical agreement between QStat CSL and ROTEM delta EXTEM ML for identifying fibrinolytic samples.Overall agreement of patient sample assignments into lysis-positive and lysis-negative based on data for QStat CSL and ROTEM delta EXTEM ML was 92%.
    Correlation between QStat FCS and fibrinogen levels.Passing-Bablok regression analysis showed good agreement between the methods (slope = 1, r = 0.815), albeit with a constant bias.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Clinical Study): 322 subjects
    • Data Provenance: Prospective observational study, conducted across seven clinical sites in the US. The study involved parturients (women in labor) aged 18 years or older with concerns for coagulopathy.
    • Sample Size (Interference Study): Not explicitly stated with a specific number for the test set, but it mentioned "normal and hypercoagulable whole blood specimens" and that the "number of replicates at each level of a screening study was targeted to provide a 95% confidence interval (2-sided), per CLSI EP07-A2 Guideline."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not specify the number or qualifications of experts used to establish ground truth for the clinical test set. Instead, ground truth for some parameters appears to be established by comparison to other legally marketed and established devices (ROTEM delta, TEG 5000) and conventional coagulation testing (aPTT, PT/INR, fibrinogen level, platelet count).

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method involving experts for establishing ground truth. The comparison is made against existing, accepted methods and devices.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    There is no mention of a multi-reader multi-case (MRMC) comparative effectiveness study, nor any effect size regarding human reader improvement with or without AI assistance. This device is an in vitro diagnostic (IVD) device that measures coagulation parameters, not an AI-assisted diagnostic imaging or interpretation tool.

    6. Standalone (Algorithm Only) Performance Study

    Yes, the studies described are standalone performance studies. The clinical and analytical specificity studies evaluate the performance of the QStat Cartridge and Quantra Hemostasis Analyzer directly, not in conjunction with human interpretation for the primary measurement. The comparisons are to other standalone diagnostic methods (ROTEM delta, TEG 5000, conventional coagulation tests).

    7. Type of Ground Truth Used

    The ground truth used for the clinical performance evaluation was based on:

    • Comparison to legally marketed viscoelastic testing devices: ROTEM delta or TEG 5000.
    • Comparison to conventional coagulation testing: aPTT, PT/INR, fibrinogen level, platelet count.
    • The "concern for coagulopathy" as a trigger for testing suggests clinical suspicion as an initial selection criterion, with the aforementioned tests serving as the gold/reference standard for comparison.

    8. Sample Size for the Training Set

    The document does not provide information about a separate training set or its sample size. The studies described appear to be focused on performance validation.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is mentioned, there is no information on how its ground truth was established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K251440
    Date Cleared
    2025-08-25

    (108 days)

    Product Code
    Regulation Number
    864.7290
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CRYOcheck Chromogenic Factor VIII is for clinical laboratory use in the quantitative determination of factor VIII activity in 3.2 % citrated human plasma. It is intended to be used in identifying factor VIII deficiency and as an aid in the management of hemophilia A in individuals aged 2 years and older. For in vitro diagnostic use.

    Device Description

    CRYOcheck Chromogenic Factor VIII is used for determination of FVIII activity and contains the following four components, packaged in glass vials and provided frozen to preserve the integrity of the components:

    • Reagent 1: Bovine FX and a fibrin polymerization inhibitor, with activators and stabilizer.
    • Reagent 2: Human FIIa, bovine FIXa, calcium chloride and phospholipids.
    • Reagent 3: FXa substrate containing EDTA and a thrombin inhibitor.
    • Diluent Buffer: Tris buffer solution containing 1% BSA and a heparin antagonist.

    In the first stage of the chromogenic assay, test plasma (containing an unknown amount of functional FVIII) is added to a reaction mixture comprised of calcium, phospholipids, human purified thrombin and FIXa, and bovine FX (Reagent 1 and Reagent 2). This mixture swiftly activates FVIII to FVIIIa, which works in concert with FIXa to activate FX. When the reaction is stopped, FXa production is assumed to be proportional to the amount of functional FVIII present in the sample. The second stage of the assay is to measure FXa through cleavage of a FXa-specific peptide nitroanilide substrate (FXa Substrate). P-nitroaniline is produced, giving a color that can be measured spectrophotometrically by absorbance at 405 nm.

    AI/ML Overview

    Based on the provided FDA 510(k) Clearance Letter, the device in question is the CRYOcheck Chromogenic Factor VIII. This document details the clearance of a modified version of an existing device, emphasizing the differences from the previous version regarding interference claims and recovery of Factor VIII replacement therapies.

    Here's an analysis of the acceptance criteria and study proving the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" for each performance claim in a quantified manner (e.g., "Interference must be less than X%"). Instead, it reports the limits of non-interference found in their studies, implying these served as the de facto acceptance criteria. For the Factor VIII replacement therapy recovery, the acceptance criterion appears to be "accurate evaluation" across a range of concentrations, with specific over/under recovery noted.

    Performance CharacteristicAcceptance Criteria (Implied)Reported Device Performance
    Interference:
    HemoglobinMust show no interference up to the concentration indicated.No interference observed up to ≤1000 mg/dL (increased from ≤500 mg/dL)
    IntralipidMust show no interference up to the concentration indicated.No interference observed up to ≤830 mg/dL (increased from ≤500 mg/dL)
    Bilirubin (unconjugated)Must show no interference up to the concentration indicated.No interference observed up to ≤40 mg/dL (increased from ≤29 mg/dL)
    Bilirubin (conjugated)Must show no interference up to the concentration indicated.No interference observed up to ≤11 mg/dL (increased from ≤2 mg/dL)
    von Willebrand factorMust show no interference up to the concentration indicated.No interference observed up to ≤20 µg/mL (same)
    Unfractionated heparinMust show no interference up to the concentration indicated.No interference observed up to ≤3.3 IU/mL (increased from ≤2 IU/mL)
    Low molecular weight heparinMust show no interference up to the concentration indicated.No interference observed up to ≤5 IU/mL (increased from ≤2 IU/mL)
    FondaparinuxMust show no interference up to the concentration indicated.No interference observed up to ≤0.2 mg/L (decreased from ≤1.25 mg/L)
    Lupus AnticoagulantMust show no interference up to the concentration indicated.No interference observed up to ≤1.8 dRVVT ratio (same)
    EmicizumabMust show no interference up to the concentration indicated.No interference observed up to ≤150 µg/mL (new claim)
    Mim8Must show no interference up to the concentration indicated.No interference observed up to ≤8 µg/mL (new claim)
    WarfarinMust show no interference up to the concentration indicated.No interference observed up to INR ≤7 (new claim)
    RivaroxabanMust not interfere.Interfered with quantification of FVIII activity.
    DabigatranMust not interfere.Interfered with quantification of FVIII activity.
    Recovery of FVIII Replacement Therapy:Must accurately evaluate potency.Accurately evaluated potency for ADVATE, ADYNOVATE, AFSTYLA, ALTUVIIO, ESPEROCT, HUMATE-P, JIVI, KOVALTRY, Novoeight, Nuwiq, and wilate at 0.05-1.0 IU/mL; ELOCTATE, and XYNTHA at 0.05-0.6 IU/mL (with over recovery at 0.8 & 1.0 IU/mL); Underestimation for OBIZUR.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Interference Studies: Plasma samples were "spiked with possible interferents," and "10 replicates were tested alongside 10 replicates of the corresponding blank matrix control." The total number of individual patient samples from which this plasma was derived is not specified, nor is the country of origin. The study design implies a prospective spiking experiment in a laboratory setting.
    • Recovery of Factor VIII Replacement Therapy: "Congenital FVIII deficient plasma was spiked with 14 FVIII replacement therapies at seven concentrations." The number of individual patient plasma units or lots of deficient plasma used is not specified. The study design implies a prospective spiking experiment in a laboratory setting.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    N/A. This is an in vitro diagnostic device for quantitative determination of factor VIII activity, not an AI/imaging device requiring expert human readers for ground truth generation. The ground truth for these studies is established by the known concentrations of spiked interferents or FVIII replacement therapies, and the intrinsic properties of the FVIII deficient plasma.

    4. Adjudication Method for the Test Set

    N/A. As this is a quantitative in vitro diagnostic device, an adjudication method in the context of human expert review of imaging or clinical data is not applicable. The results are measured spectrophotometrically.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC study was not done. This type of study is relevant for AI imaging devices where human readers interpret medical images with and without AI assistance. This document describes an in vitro diagnostic device.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

    Yes, this entire submission describes the standalone performance of the CRYOcheck Chromogenic Factor VIII assay. The device itself performs the quantitative determination of FVIII activity, entirely without a "human-in-the-loop" once the sample is loaded and the assay run according to protocol.

    7. The Type of Ground Truth Used

    • Interference Studies: The ground truth was the known concentration of the spiked interferent (e.g., Hemoglobin, Intralipid, Bilirubin, etc.) added to plasma samples, and the corresponding blank matrix control.
    • Recovery of Factor VIII Replacement Therapy: The ground truth was the known concentration of the spiked FVIII replacement therapy added to congenital FVIII deficient plasma at various concentrations.

    8. The Sample Size for the Training Set

    N/A. This document describes an in vitro diagnostic assay based on chromogenic principles, not an AI/ML algorithm that requires a "training set" in the computational sense. The device's components (reagents, diluent buffer) and their interaction define the assay, which is then validated through performance studies.

    9. How the Ground Truth for the Training Set was Established

    N/A. See point 8. The "ground truth" for developing and optimizing such a chromogenic assay would stem from extensive biochemical research, characterization of reagents, and titrations against known standards, which is inherent in the development of any diagnostic assay, but not referred to as a "training set" or "ground truth establishment" in the AI/ML context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K251087
    Manufacturer
    Date Cleared
    2025-08-04

    (117 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Daxor BVA-200 is an automated system that is used to measure/calculate the red cell mass (mL), plasma volume (mL) and total blood volume (mL), along with the related deviations from ideal values by amount (mL) and percentage (%) in adults. In addition, the Normalized Hematocrit (%) and Albumin Transudation Rate (%/min) are calculated. It is an in vitro medical device composed of a microprocessor, software, touchscreen, and gamma counter and accessory convenience kit.

    The Daxor BVA-200 is intended to calculate human blood volumes by the method of tracer diffusion (Indicator dilution technique) with I-131 as the tracer after injection of I-131 Human Serum Albumin. The Daxor BVA-200 provides a Quantitative Assessment of total blood and plasma volumes using an automated system.

    Data inputs to the software come from the measured characteristics of patient venous whole blood samples collected in K3EDTA vacutainer tubes (hematocrit and tracer concentration) and tracer calibration standards. The patient blood samples and the calibration standards are measured in a gamma counter, whose output is automatically input to this calculation program. The package also calculates the patient expected (or ideal) blood volume from physical parameters. Hyper- or hypovolemia, and associated red cell volumes are reported, with statistics showing the quality of the results.

    For in vitro diagnostic use in a Clinical Laboratory setting and operated by laboratory technicians.

    Rx use only.

    Device Description

    The Daxor BVA-200 is an automated system that is used to calculate the red cell mass, plasma volume and total blood volume. It is an in vitro medical device composed of a microprocessor, software, touchscreen, and gamma counter. The accessory convenience kit includes single-use whole blood cartridges and protective sleeves.

    The Daxor BVA-200 is designed to calculate human blood volume, using the method of tracer dilution, utilizing tagged serum albumin (I-131, resulting in "I-HSA"). Data inputs to the software come from the measured characteristics of patient blood samples (hematocrit and tracer concentration) and tracer calibration standards. The package also calculates the patient expected (or ideal) blood volume from physical parameters. Hyper- or hypovolemia, and associated red cell volumes, are reported, with statistics showing the quality of the results.

    The patient blood samples and the calibration standards are measured in a gamma counter, whose output is automatically input to this calculation program.

    The BVA-200 has a touchscreen for operator interaction, and provides clear instructions and prompts for the steps necessary for performing the test.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Daxor Blood Volume Analyzer (200), based on the provided FDA 510(k) clearance letter.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in a tabulated format. However, it provides performance metrics from a clinical comparison study and precision, linearity, carryover, and stability studies. The implicit acceptance criteria are that the device performance should be substantially equivalent to the predicate device and demonstrate acceptable precision and linearity.

    Implicit Acceptance Criteria and Reported Device Performance for BVA-200 (vs. BVA-100)

    MeasureAcceptance Criteria (Implicit, based on substantial equivalence)Reported Device Performance (vs. Predicate BVA-100)
    Clinical Comparison (Regression Analysis)Strong correlation, slope near 1, intercept near 0, narrow 95% CI covering 1 for slope and 0 for intercept.
    TBV (Total Blood Volume)-Passing-Bablok Slope: 1.031 (95% CI: 1.014 to 1.049), Intercept: -138 (-235 to -59), Pearson r: 0.99Deming Slope: 1.033 (95% CI: 1.017 to 1.050), Intercept: -153 (-241 to -69), Pearson r: 0.99
    RCV (Red Cell Volume)-Passing-Bablok Slope: 1.016 (95% CI: 1.002 to 1.030), Intercept: -25 (-50 to 2), Pearson r: 0.99Deming Slope: 1.024 (95% CI: 1.008 to 1.040), Intercept: -37 (-68 to -10), Pearson r: 0.99
    PV (Plasma Volume)-Passing-Bablok Slope: 1.033 (95% CI: 1.016 to 1.051), Intercept: -99 (-155 to -46), Pearson r: 0.99Deming Slope: 1.029 (95% CI: 1.014 to 1.046), Intercept: -83 (-138 to -32), Pearson r: 0.99
    nHct (Normalized Hematocrit)-Passing-Bablok Slope: 1.007 (95% CI: 0.982 to 1.032), Intercept: -0.19 (-1.18 to 0.77), Pearson r: 0.98Deming Slope: 1.022 (95% CI: 0.999 to 1.047), Intercept: -0.76 (-1.71 to 0.21), Pearson r: 0.98
    Precision (Total Precision %CV)(Implicitly) Low %CV, indicating high precision.All total precision %CV results were below 3.6%.
    Linearity(Implicitly) Strong correlation with expected values, low bias, and CV for repeatability ≤5%, max deviation from linearity ≤8%.Global Slope near 1 (range 1.00-1.02), Mean Bias Percent low (range 0.37-0.41).For contrived samples (2000-18000 ml): CV for repeatability ≤5%, Max deviation from linearity ≤8%.
    CarryoverNo carryover observed.No carryover observed.
    Sample StabilityImpervious to temperature/humidity, decay matches I-131, withstands shipping, non-level operation up to 15 degrees, settling does not affect readings, stability for 4 days (refrigerated/room temp).All stated conditions were met. "No significant differences were observed, nor was there any significant drift of measurements observed."
    Handling (Cartridge drop test)Performance criteria still met after drop.None of the average measured parameters differed from pre-drop by more than 2.7%.
    Handling (Shipping & Handling)Resistance to breakage and radio-emission stability. Sample and device integrity maintained.None of the average measured parameters differed from pre-ship by more than 3.1%.
    Handling (Tilt position)Stable measurement at non-level positioning up to 15 degrees.None of the average measured parameters differed from pre-ship by more than 3.1%.
    Interference (Hemolysis)Results not affected by hemolysis.Partial hemolysis: -0.72% ± 0.77% difference from baseline. Gross hemolysis: -0.14% ± 2.16% difference from baseline.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 319 unique, independent, comparable measurements.
    • Data Provenance:
      • Blood derived from patients undergoing blood volume measurement with the BVA-100 (predicate device) as part of their clinical treatment or as part of a research study.
      • The document implies the data is retrospective for the clinical comparison, as it used blood derived from patients undergoing measurement with the predicate device.
      • For precision, linearity, sample stability, handling, and interference studies, contrived samples and venous whole blood samples from male and female volunteers were used. No specific country of origin is mentioned.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not mention the use of experts to establish ground truth for the test set in the context of a reader study or image interpretation. The ground truth for the clinical comparison study was established by the predicate device (Daxor BVA-100) measurements.

    4. Adjudication Method for the Test Set

    Not applicable. The ground truth for the clinical comparison was based on measurements from the predicate device, not expert adjudication of subjective assessments.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This device is a quantitative measurement system, not an AI software intended for interpretation by human readers. The clinical study focused on comparing the performance of the new device (BVA-200) to its predicate (BVA-100) on quantitative measurements.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, the studies performed focused on the standalone performance of the BVA-200. The device is an automated system that measures and calculates blood volumes. The performance data presented (clinical comparison, precision, linearity, stability, etc.) represents the device's output independently, without human interpretation as part of a "human-in-the-loop" workflow being evaluated.

    7. The Type of Ground Truth Used

    The ground truth for the primary clinical comparison study was the measurements obtained from the predicate device, the Daxor BVA-100. For other studies (precision, linearity, stability, etc.), the ground truth was based on known properties of contrived samples or established physical/biological principles (e.g., known decay rate of I-131).

    8. The Sample Size for the Training Set

    The document does not provide information on a training set sample size. This suggests that the BVA-200, similar to its predicate BVA-100, is likely based on established physiological and radiological principles (indicator dilution technique, I-131 decay) rather than a machine learning model that requires a distinct "training set." The software performs calculations based on these principles and measured inputs, not through a learned model from a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as a "training set" in the context of machine learning is not implied or described in the document for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K251968
    Date Cleared
    2025-07-24

    (28 days)

    Product Code
    Regulation Number
    864.7340
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For the quantitative determination of fibrinogen, based on the Clauss method, in human citrated plasma on IL Coagulation Systems.

    Device Description

    Several congenital abnormalities of fibrinogen result in impaired conversion of fibrinogen to fibrin during blood coagulation. Fibrinogen is also a useful marker in the evaluation of several disease states including disseminated intravascular coagulation, liver disease, inflammatory diseases and malignancy. High levels of fibrinogen are associated with an increased risk for cardiovascular disease. Increased levels are also found during pregnancy and oral contraceptive use, while reduced levels are found during thrombolytic therapy.

    The HemosIL Fibrinogen-C kit and HemosIL Fibrinogen-C XL kit use an excess of thrombin to convert fibrinogen to fibrin in diluted plasma. At high thrombin and low fibrinogen concentration, the rate of reaction is a function of fibrinogen concentration.

    AI/ML Overview

    This document, an FDA 510(k) clearance letter for HemosIL Fibrinogen-C and HemosIL Fibrinogen-C XL, does not contain the information requested in your prompt regarding acceptance criteria and a study proving the device meets those criteria.

    The 510(k) summary clearly states:

    • Reason for Submission: "This Special 510(k) is being submitted to remove the reconstituted reagent frozen stability claim of 1 month at -20°C in the original vial for HemosIL Fibrinogen-C and HemosIL Fibrinogen-C XL."
    • Data Requirement: "No new performance data are needed to remove the reconstituted reagent frozen stability claim."
    • No new performance claims: "Changes to labeled performance claims, except to remove the reconstituted reagent frozen stability claim from the Instructions for Use and add 'Do not freeze reconstituted reagent.'"

    Therefore, the document explicitly states that no new performance data or studies were required or submitted for this particular 510(k) clearance. The clearance is based on a minor labeling change related to reagent stability, not on new performance validation.

    To provide the information you requested (acceptance criteria, study details, sample sizes, expert involvement, ground truth, etc.), you would need a different type of submission document, such as an original 510(k) application or a PMA, where initial and comprehensive performance validation studies are typically included.

    Ask a Question

    Ask a specific question about this device

    K Number
    K251564
    Device Name
    microINR System
    Date Cleared
    2025-07-21

    (60 days)

    Product Code
    Regulation Number
    864.7750
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The microINR System measures prothrombin time (PT) expressed in International Normalized Ratio (INR), for monitoring oral anticoagulant therapy with warfarin.

    The microINR System consists of a meter and chips (test strips) and uses fresh capillary whole blood from a fingerstick. The microINR System is intended for patient self-testing use as well as for healthcare professionals at Point of Care settings.

    The microINR System is intended for use in patients 18 years old or older. Patients must be stable on warfarin medication for at least 6 weeks before starting to use the microINR System.

    For self-testing use: The system is intended for properly trained users under specific prescription of a physician.

    Caution: The microINR System is not intended for use in patients who are transitioning from heparin treatment to warfarin therapy. The microINR System is not intended to be used for screening purposes.

    Device Description

    The microINR System is comprised of a portable measuring device (microINR, microINR Link and microINR Expert meter) and test strips (microINR Chips) in which the capillary blood sample flows through capillary action.

    The microINR Chip contains a reagent in dried form which consists of thromboplastin, and contains two symmetrical regions, the measuring channel and a control channel. The microINR meters measure International Normalized Ratio (INR) based on a Prothrombin Time (PT) assay carried out in the microINR Chip based on microfluidic technology with machine vision detection.

    The microINR System has a multi-level On-board Quality Control. Multiple key functions and elements of the system are checked and if deviations are detected, error messages are displayed and test results are not reported.

    AI/ML Overview

    The microINR System measures prothrombin time (PT) expressed in International Normalized Ratio (INR) for monitoring oral anticoagulant therapy with warfarin.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) clearance letter:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined "acceptance criteria" in terms of specific thresholds for slope, intercept, Pearson correlation, SD, or CV%. Instead, it presents the performance characteristics demonstrated by the studies. The comparison to predicate devices, and the conclusion of substantial equivalence, imply that these reported performance values were considered acceptable.

    MetricAcceptance Criteria (Implied)Reported Device Performance
    Accuracy (microINR System vs. Lab Reference)Strong correlation (r ≈ 1), small bias (slope ≈ 1, intercept ≈ 0)N: 1016, Slope: 1.00, Intercept: 0.08, Pearson (r): 0.97
    Accuracy (Self-testing patients vs. Lab Reference)Strong correlation (r ≈ 1), small bias (slope ≈ 1, intercept ≈ 0)N: 248, Slope: 0.95, Intercept: 0.13, Pearson (r): 0.97
    Accuracy (Self-testing patients vs. HCPs using microINR System)Strong correlation (r ≈ 1), small bias (slope ≈ 1, intercept ≈ 0)N: 461, Slope: 0.99, Intercept: 0.02, Pearson (r): 0.98
    Precision (INR < 2.0)Low variability (small SD, CV%)N test pairs: 131, Mean: 1.46, SD: 0.07, CV%: 4.9
    Precision (2.0 ≤ INR < 4.5)Low variability (small SD, CV%)N test pairs: 337, Mean: 2.79, SD: 0.14, CV%: 5.0
    Precision (INR ≥ 4.5)Low variability (small SD, CV%)N test pairs: 34, Mean: 5.30, SD: 0.26, CV%: 4.9
    Precision (Trained PSTs)Low variability (small SD, CV%)N: 111, Mean: 2.63, SD: 0.13, CV%: 4.9
    Precision (Untrained PSTs with high INR)Low variability (small SD, CV%)N: 13, Mean: 5.64, SD: 0.26, CV%: 4.6

    Note: The "Acceptance Criteria (Implied)" are derived from the typical expectations for such medical devices and the fact that these results led to FDA clearance, indicating they met regulatory standards.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Accuracy Study (microINR System vs. Laboratory Reference Device):

      • N: 1016 microINR System results.
      • Data Provenance: Capillary blood samples from control subjects and patients recruited at ten clinical sites. This suggests a prospective collection from multiple geographical locations, though specific countries are not mentioned.
    • Accuracy Study (Self-testing patients vs. Laboratory Reference Device):

      • N: 248 INR test results obtained by self-testing patients (compared to laboratory systems).
      • Data Provenance: Collected at seven clinical sites. This also implies prospective data collection.
    • Accuracy Study (Self-testing patients vs. Healthcare Professionals using microINR System):

      • N: 461 INR test results.
      • Data Provenance: Collected at seven clinical sites, comparing self-testers to healthcare professionals. This is likely prospective.
    • Precision Study (microINR System):

      • N: Duplicate measurements from 502 control subjects and patients.
      • Data Provenance: Performed at ten clinical sites, suggesting prospective collection.
    • Precision Study (Trained PSTs):

      • N: Duplicate measurements from 111 patients ("Paired results").
      • Data Provenance: Performed at four clinical sites. Implies prospective collection.
    • Precision Study (Untrained PSTs with high INR values):

      • N: Duplicate measurements from 13 patients ("Paired results").
      • Data Provenance: Performed at three clinical sites. Implies prospective collection.

    The general provenance is from multiple clinical sites, suggesting a diverse patient population, but specific countries are not detailed. All studies appear to be prospectively collected for the purpose of validating the device performance.

    3. Number of Experts Used to Establish the Ground Truth and Qualifications

    • The document mentions comparisons to "laboratory systems ACL TOP 500 or ACL TOP 750" as reference methods. These are established laboratory analyzers.
    • For the "Accuracy of the microINR System by self-testing patients" study, the comparison also includes "Healthcare professionals (HCP) using the microINR System." While HCPs are involved, specific qualifications beyond their professional role are not detailed for the purpose of establishing a "ground truth" other than operating the device.
    • The ground truth for the clinical method comparison studies seems to be based on the results from the ACL TOP 500 or ACL TOP 750 laboratory systems, which are considered the gold standard for PT/INR measurement, rather than human experts adjudicating individual cases.
    • The document does not explicitly describe "experts" in the sense of radiologists reviewing images; rather, it refers to standardized laboratory methods and healthcare professionals. Therefore, the specific "number of experts" or their "qualifications" for establishing ground truth, beyond the inherent accuracy of the reference laboratory devices, is not provided.

    4. Adjudication Method for the Test Set

    No explicit adjudication method (e.g., 2+1, 3+1) is mentioned. Given that the ground truth appears to be established by comparison to highly accurate and standardized laboratory reference instruments (ACL TOP 500/750), a human adjudication process for individual test results is typically not required or described in such device submissions. The comparison is quantitative against these reference devices.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study, detailing how much human readers improve with AI vs. without AI assistance, was performed or presented. This device is a standalone diagnostic system (a meter and test strips) and not an AI-assisted diagnostic tool that aids human interpretation of complex medical images or data. The "readers" in this context are the device users (patients or healthcare professionals), and the study compares their results directly to a laboratory reference, or compares self-testers to healthcare professionals using the same device, rather than measuring an improvement in human diagnostic accuracy with an AI assist.

    6. Standalone Performance

    Yes, standalone performance was done. The entire set of performance characteristics, particularly the "Accuracy of the microINR System" (comparing the microINR System against a laboratory reference device) and "Precision of the microINR System," represents the standalone performance of the device without explicit human-in-the-loop interpretation beyond operating the device as intended. The system itself measures and displays the INR.

    7. Type of Ground Truth Used

    The primary ground truth used is comparison to established laboratory reference methods, specifically ACL TOP 500 or ACL TOP 750 laboratory systems, which measure PT/INR from citrate venous plasma samples. This is a highly standardized and accepted method for determining INR.

    8. Sample Size for the Training Set

    The document does not provide details on a specific "training set" for the device, as it is not an AI/machine learning model in the typical sense that requires a training and a separate test set. The changes implemented are "only SW changes... for increasing the INR measuring range... and the removal of the limitation for LVAD patients," implying a modification to an existing algorithm rather than a completely new, data-trained AI model. The studies presented are for validation of these changes.

    9. How the Ground Truth for the Training Set Was Established

    As noted above, no specific "training set" is mentioned in the context of a machine learning model. The device's underlying principle relies on microfluidic technology and machine vision detection of a prothrombin time assay. The calibration of the chips is done against "International Reference Thromboplastin of the World Health Organization," which serves as a fundamental standard rather than a "ground truth" for a training set in data science terms.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243144
    Manufacturer
    Date Cleared
    2025-06-27

    (270 days)

    Product Code
    Regulation Number
    864.5260
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The X100/X100HT with Full Field Peripheral Blood Smear Application is intended to locate and display images of white cells, red cells, and platelets acquired from fixed and stained peripheral blood smears and assists a qualified technologist in conducting a WBC differential, RBC morphology evaluation, and platelet estimate using those images. For in-vitro diagnostic use only. For professional use only.

    Device Description

    The X100 / X100HT with Full Field Peripheral Blood Smear (PBS) Application ("Full Field PBS") is a digital cell morphology solution, presenting high resolution digital images of fixed and stained microscopy Peripheral Blood Smears. The Full Field PBS was previously cleared by the Agency on October 2, 2020, throughout the review of K201301 and on May 3, 2022, throughout the review of K220013. The system automatically locates and presents images of peripheral blood cells and streamlines the PBS analysis process with a review workflow composed of four steps: (1) full field review, (2) white blood cells (WBC) review (DSS for WBC Pre-Classification is available), (3) red blood cells (RBC) review and (4) platelet review (DSS for platelet estimation is available).

    Under the proposed modification, subject of this 510(k) submission, additional DSS component is added to the RBC and Platelet review steps. Concerning RBC analysis, system-suggested RBC morphological pre-gradings are added as proposed DSS to the user by means of a dotted line around a suggested grading selection box. Notably, the user is still required to review the slide and actively mark the final grading, exactly as performed in the cleared workflow review.

    The same approach is used with regard to the update to the platelet review step; A system-suggested platelet clump indication is presented to the user by means of a dotted line around a selection box. Notably, the user is still required to manually mark whether platelet clumps were detected, exactly as currently performed in the cleared workflow review.

    The changes under discussion do not affect the cleared indications for use or intended use; the user's workflow of scanning and analyzing peripheral blood smears using the Full Field PBS Application remains otherwise unchanged as well.

    AI/ML Overview

    The Scopio Labs X100/X100HT with Full Field Peripheral Blood Smear (PBS) Application has received FDA 510(k) clearance (K243144) based on a modification that adds system-suggested pre-gradings for RBC morphology and platelet clump indications. The study involved a method comparison study, repeatability study, reproducibility study, and software verification and validation.

    1. Acceptance Criteria and Reported Device Performance

    The FDA clearance document does not explicitly state the numerical acceptance criteria for the method comparison study. However, it indicates that "The results met the pre-defined acceptance criteria." The reported performance metrics are presented as overall agreement, Positive Percent Agreement (PPA), and Negative Percent Agreement (NPA).

    CategoryOverall AgreementPPANPA
    RBC Color97.88% (97.29% to 98.42%)98.33% (97.48% to 99.10%)97.61% (96.81% to 98.33%)
    RBC Inclusions97.90% (97.50% to 98.27%)86.73% (81.66% to 91.23%)98.41% (98.06% to 98.78%)
    RBC Shape96.22% (95.92% to 96.50%)95.35% (94.50% to 96.12%)96.40% (96.06% to 96.71%)
    RBC Size95.58% (95.06% to 96.13%)99.42% (99.03% to 99.75%)92.72% (91.82% to 93.70%)
    PLT Clumping87.08% (85.25% to 88.92%)86.11% (82.13% to 89.91%)87.39% (85.39% to 89.39%)

    The precision studies (repeatability and reproducibility) also "met the pre-defined acceptance criteria." However, the specific numerical criteria for these studies are not detailed in the provided document.

    2. Sample Size and Data Provenance for the Test Set

    • Sample Size: A total of 1200 anonymized PBS slides were used for the method comparison study.
    • Data Provenance: The slides were collected from the laboratory routine workload of three medical centers. The country of origin is not specified, but the submitter information lists "Tel Aviv, Israel" as the sponsor address, which may suggest the data originated from Israel, though this is not explicitly stated for the clinical evaluation. The data is retrospective, as it was collected from "routine workload."

    3. Number of Experts and Qualifications for Ground Truth Establishment (Test Set)

    The document does not explicitly state the number of experts used to establish the ground truth for the test set or their specific qualifications (e.g., years of experience). It implies that the ground truth was established by referring to "pre-defined acceptance criteria" and a "method comparison study," which typically involves comparison against human expert interpretation. However, the details of this expert interpretation are not provided.

    4. Adjudication Method for the Test Set

    The adjudication method used to establish the ground truth for the test set is not specified in the provided document.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study where human readers' performance with and without AI assistance was evaluated. The study described is a method comparison study between the modified device (with added DSS) and either the previous cleared device or human expert consensus, but not a comparative effectiveness study measuring improvement in human reader performance.

    6. Standalone Performance Study (Algorithm Only)

    The document describes the device as providing "system-suggested RBC morphological pre-gradings" and "system-suggested platelet clump indication" by means of a dotted line, and that "the user is still required to review the slide and actively mark the final grading." This indicates that the device functions as an assistive tool (human-in-the-loop) rather than a completely standalone algorithm making final diagnoses. The performance metrics presented (Overall Agreement, PPA, NPA) are likely comparisons of the system's suggestions against ground truth, which implicitly reflects a standalone component, but the final reported performance is within the context of assisting a qualified technologist, not solely the algorithm's output. Therefore, a purely standalone (algorithm-only without human-in-the-loop) performance is not explicitly presented or claimed for clinical use.

    7. Type of Ground Truth Used

    Based on the description of the "method comparison study" and the nature of the device (assisting a qualified technologist), the ground truth was likely established by expert consensus or through a reference method performed by qualified experts in hematology/morphology. The context of "laboratory routine workload" and comparison implies an existing gold standard interpretation.

    8. Sample Size for the Training Set

    The document does not provide information on the sample size used for the training set of the AI/DSS components. It only details the test set used for performance evaluation of the modified device.

    9. How the Ground Truth for the Training Set Was Established

    The document does not provide information on how the ground truth for the training set was established. It states that "Cell images are analysed using standard mathematical methods, including deterministic artificial neural networks (ANN's) trained to distinguish between classes of white blood cells." This implies a training process, but the details of ground truth establishment for that training are omitted.

    Ask a Question

    Ask a specific question about this device

    K Number
    K251371
    Date Cleared
    2025-06-25

    (54 days)

    Product Code
    Regulation Number
    864.5220
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The XR-Series module (XR-20) is a quantitative multi-parameter automated hematology analyzer intended for in vitro diagnostic use in screening patient populations found in clinical laboratories.

    The XR-Series module classifies and enumerates the following parameters in whole blood: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT (PLT-I, PLT-F), NEUT%/#, LYMPH%/#, MONO%/#, EO%/#, BASO%/#, IG%/#, RDW-CV, RDW-SD, MPV, NRBC%/#, RET%/#, IPF, IPF#, IRF, RET-He and has a Body Fluid mode for body fluids. The Body Fluid mode enumerates the WBC-BF, RBC-BF, MN%/#, PMN%/#, and TC-BF# parameters in cerebrospinal fluid (CSF), serous fluids (peritoneal, pleural) and synovial fluids. Whole blood should be collected in K2EDTA or K3EDTA anticoagulant, and serous and synovial fluids in K2EDTA anticoagulant to prevent clotting of fluid. The use of anticoagulants with CSF specimens is neither required nor recommended.

    Device Description

    The Sysmex XR-Series module (XR-20) is a quantitative multi-parameter hematology analyzer intended to perform tests on whole blood samples collected in K2 or K3EDTA and body fluids (pleural, peritoneal and synovial) collected in K2EDTA anticoagulant. The analyzers can also perform tests on CSF, which should not be collected in any anticoagulant. The XR-Series analyzer consist of four principal units: (1) One Main Unit (XR-20) which aspirates, dilutes, mixes, and analyzes blood and body fluid samples; (2) One Auto Sampler Unit which supply samples to the Main Unit automatically; (3) IPU (Information Processing Unit) which processes data from the Main Unit and provides the operator interface with the system; (4) Pneumatic Unit which supplies pressure and vacuum from the Main Unit.

    The XR-20 analyzer has an additional white progenitor cell (WPC) measuring channel and associated WPC reagents. The new WPC channel provides two separate flags for blasts and abnormal lymphocytes.

    AI/ML Overview

    The provided FDA 510(k) Clearance Letter details the performance testing conducted for the Sysmex XR-Series (XR-20) Automated Hematology Analyzer to demonstrate its substantial equivalence to the predicate device, Sysmex XN-20. Due to the nature of the document being an FDA clearance letter summarizing performance studies rather than the full study reports, some requested details (e.g., exact sample provenance for all studies beyond "US clinical sites," specific qualifications for all experts, and the comprehensive list of acceptance criteria for all individual parameters in specific studies) are not exhaustively provided.

    However, based on the information available, here's a breakdown of the acceptance criteria and the studies proving the device meets them:

    Overall Acceptance Criteria & Study Design Philosophy:

    The overarching acceptance criterion for this 510(k) submission is to demonstrate substantial equivalence to the predicate device, Sysmex XN-20 (K112605). This is primarily proven through:

    • Analytical Performance Studies: Demonstrating that the XR-20 analyzer's performance (precision, linearity, analytical specificity, stability, limits of detection, carry-over) is "acceptable" or "met manufacturer's specifications/predefined acceptance criteria requirements."
    • Method Comparison Studies: Showing a strong correlation and acceptable bias between the XR-20 and the predicate XN-20 for all claimed parameters across various patient populations and challenging samples.
    • Clinical Sensitivity and Specificity Studies: For flagging capabilities, demonstrating acceptable agreement (sensitivity/specificity, PPA/NPA/OPA) with a reference method (manual microscopy) and the predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't provide a single, consolidated table of all acceptance criteria for every parameter across all tests. Instead, it states that results "met manufacturer's specifications or predefined acceptance criteria requirements" for analytical performance tests, and provides specific correlation coefficients, slopes, intercepts, and mean differences/percent differences for method comparison studies.

    Here's a partial table based on the quantifiable data presented for Method Comparison Studies (Whole Blood - Combined Sites), which is a key performance indicator for substantial equivalence. The "Acceptance Criteria" are implied by what is generally considered acceptable in hematology analyzer comparisons for FDA submissions (high correlation, small bias), and explicitly stated for certain parameters like HGB.

    Implicit Acceptance Criteria (General expectation for Method Comparison based on FDA context):

    • Correlation Coefficient (r): Typically > 0.95 (ideally > 0.98 or 0.99 for robust parameters)
    • Slope: Close to 1.0 (ideally between 0.95 and 1.05)
    • Intercept: Close to 0
    • Mean Difference / % Mean Difference / Estimated Bias: Within clinically acceptable limits (often derived from biological variation or regulatory guidelines). The document explicitly mentions a bias limit for HGB: ±2% or 0.2 g/dL.

    Table 1: Partial Acceptance Criteria and Reported Device Performance (Method Comparison - Whole Blood)

    MeasurandAcceptance Criteria for r (Implied/Explicit)Reported rAcceptance Criteria for Slope (Implied)Reported Slope (95% CI)Acceptance Criteria for Mean Diff. / % Diff. (Implied/Explicit)Reported Mean Diff. / % Diff.Key Conclusion based on Criteria
    WBC> 0.99 (high)0.9999~1.01.003 (0.998, 1.007)Close to 00.17 / 0.96%Met
    RBC> 0.99 (high)0.9944~1.01.000 (0.993, 1.006)Close to 0-0.01 / -0.34%Met
    HGB> 0.99 (high)0.9954~1.00.993 (0.989, 0.998)±2% or 0.2 g/dL (explicit)-0.1 / -0.79% (Note: One site had -2.10% / -0.3 g/dL bias, stated as acceptable due to high r)Met (with explanation for one site's bias)
    HCT> 0.99 (high)0.9946~1.00.998 (0.993, 1.003)Close to 0-0.2 / -0.40%Met
    PLT-I> 0.99 (high)0.9990~1.01.005 (0.991, 1.020)Close to 0-2 / -0.72%Met
    PLT-F> 0.99 (high)0.9990~1.01.034 (1.019, 1.048)Close to 06 / 1.83%Met
    NRBC> 0.99 (high)0.9996~1.01.006 (0.996, 1.016)Close to 00.00 / 0.61%Met
    RET (%)> 0.99 (high)0.9931~1.01.033 (1.009, 1.057)Close to 00.06 / 2.68%Met
    IRF (%)~ 0.98 (high)0.9820~1.00.998 (0.983, 1.012)Close to 0-0.9 / -5.32%Met
    IPF (%)> 0.99 (high)0.9902~1.00.999 (0.976, 1.023)Close to 0-0.0 / -0.94%Met
    RET-He (pg)~ 0.96 (high)0.9616~0.93 (lower, but CI is tight)0.930 (0.906, 0.954)Close to 0-1.2 / -3.85%Met

    For other analytical performance studies (Precision, Linearity, Detection Limit, Carry-Over, Specificity, Stability), the document consistently states that the XR-20 "met manufacturer's specifications or predefined acceptance criteria requirements," supporting that specific numerical acceptance criteria were defined and achieved.


    2. Sample Size and Data Provenance

    • Test Set:

      • Whole Blood Method Comparison: A total of 865 unique residual whole blood samples.
      • Body Fluid Method Comparison: A total of 397 residual body fluid samples.
      • Provenance: All studies were conducted at three US clinical sites (for major studies like method comparison and reproducibility) or one internal site (for some linearity, stability, and matrix studies).
      • Retrospective/Prospective: Samples are described as "residual" (implying retrospective, de-identified leftover samples) or "prospectively collected" where specified (e.g., for some stability studies).
    • Training Set: The document does not specify a training set for the algorithm, as this is a traditional in-vitro diagnostic (IVD) device (Automated Hematology Analyzer) which likely relies on fixed algorithms and established measurement principles (RF/DC Detection, Sheath Flow DC Detection, Flow Cytometry) rather than a machine learning model that requires explicit training data for its core functionality. The performance characterization is about its analytical capabilities, not about learning from a dataset to perform a task.


    3. Number of Experts and Qualifications (for Ground Truth)

    • Clinical Sensitivity and Specificity (Flagging Capabilities): The ground truth for flagging capabilities was established by "manual differential counts and peripheral blood smear review by experienced examiners using light microscopy (reference method) at each of the three external clinical sites." The exact number and specific qualifications (e.g., "radiologist with 10 years of experience") are not provided, but the term "experienced examiners" implies qualified personnel (e.g., clinical laboratory scientists, pathologists). Given this is a hematology analyzer, these would typically be clinical laboratory specialists or hematopathologists.

    4. Adjudication Method (for the Test Set)

    • Clinical Sensitivity and Specificity (Flagging): The document does not explicitly describe an adjudication method (e.g., 2+1, 3+1) for resolving discrepancies between multiple manual reviews. It states "peripheral blood smear review by experienced examiners," primarily using manual microscopy as the reference method. This implies there might be a single expert review or an internal consensus process, but no detail on conflict resolution is provided.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC study was described. This type of study (MRMC) is typically performed for AI-assisted diagnostic tools where human reader performance is a direct outcome of interest and needs to be compared with and without AI assistance. The Sysmex XR-20 is an automated analyzer, a standalone device that performs measurements and classifications. While it outputs flags that may assist human review, its primary function isn't human-in-the-loop assistance in interpretation (like an AI for radiology image reading). Therefore, an MRMC study is not applicable for this device.

    6. Standalone (Algorithm Only) Performance

    • Yes, the primary performance studies are standalone algorithm/device performance. All analytical performance studies (Precision, Linearity, Detection Limit, Carry-Over, Analytical Specificity, Sample Stability) and the method comparison studies (comparing XR-20 results directly against the predicate XN-20) represent the standalone performance of the XR-20 analyzer. The device functions automatically without human input during the analysis process; therefore, human-in-the-loop performance is not directly evaluated as a primary outcome for its measurement capabilities.

    7. Type of Ground Truth Used

    • Analytical Ground Truth: For most analytical performance studies (Precision, Linearity, Detection Limits, Carry-Over), the "ground truth" is established by the performance of the predicate device (Sysmex XN-20) or by a well-controlled experimental setup (e.g., known dilutions for linearity, blank samples for LoB).
    • Clinical Ground Truth (for flagging): For clinical sensitivity and specificity of flagging capabilities, the ground truth was expert consensus / manual review against peripheral blood smear using light microscopy. The document refers to this as the "reference method."

    8. Sample Size for the Training Set

    • As mentioned in point 2, no explicit training set for a machine learning algorithm is described. This device's core functionality relies on established physical and chemical principles and traditional signal processing for cell counting and classification, not a learnable AI model from a training data set in the typical sense.

    9. How the Ground Truth for the Training Set Was Established

    • Since there's no described "training set" for an AI/ML algorithm, this point is not applicable in the context of this traditional IVD device. The "ground truth" for verifying its performance (as detailed above) was established through comparisons to a legally marketed predicate device (XN-20) and a gold standard manual method (microscopy).
    Ask a Question

    Ask a specific question about this device

    K Number
    K250943
    Date Cleared
    2025-06-25

    (89 days)

    Product Code
    Regulation Number
    864.5220
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The XR-Series module (XR-10) is a quantitative multi-parameter automated hematology analyzer intended for in vitro diagnostic use in screening patient populations found in clinical laboratories.

    The XR-Series module classifies and enumerates the following parameters in whole blood: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT (PLT-I, PLT-F), NEUT%/#, LYMPH%/#, MONO%/#, EO%/#, BASO%/#, IG%/#, RDW-CV, RDW-SD, MPV, NRBC%/#, RET%/#, IPF, IPF#, IRF, RET-He and has a Body Fluid mode for body fluids. The Body Fluid mode enumerates the WBC-BF, RBC-BF, MN%/#, PMN%/#, and TC-BF# parameters in cerebrospinal fluid (CSF), serous fluids (peritoneal, pleural) and synovial fluids. Whole blood should be collected in K2EDTA or K3EDTA anticoagulant, and serous and synovial fluids in K2EDTA anticoagulant to prevent clotting of fluid. The use of anticoagulants with CSF specimens is neither required nor recommended.

    Device Description

    The Sysmex XR-Series module (XR-10) is a quantitative multi-parameter hematology analyzer intended to perform tests on whole blood samples collected in K2 or K3EDTA and body fluids (pleural, peritoneal and synovial) collected in K2EDTA anticoagulant. The analyzers can also perform tests on CSF, which should not be collected in any anticoagulant. The XR-Series analyzer consist of four principal units: (1) One Main Units (XR-10) which aspirates, dilutes, mixes, and analyzes blood and body fluid samples; (2) Two Auto Sampler Units (SA-10, SA-01) which supply samples to the Main Unit automatically; (3) IPU (Information Processing Unit) which processes data from the Main Unit and provides the operator interface with the system; (4) Pneumatic Unit which supplies pressure and vacuum from the Main Unit.

    AI/ML Overview

    This document describes the acceptance criteria and the studies conducted to prove that the Sysmex XR-Series (XR-10) Automated Hematology Analyzer meets these criteria, demonstrating substantial equivalence to its predicate device, the Sysmex XN-20.

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA 510(k) clearance letter does not explicitly present a neatly formatted table of acceptance criteria alongside the reported performance for all parameters. Instead, it describes various performance studies (Precision, Linearity, Analytical Specificity/Interferences, Sample Stability, Detection Limit, Carry-Over, Comparison Studies, Matrix Studies, Bridging Studies, Clinical Studies, and Expected Values/Reference Range) and states that the XR-10 met "manufacturer's specifications or predefined acceptance criteria requirements" for each.

    However, based on the provided data, we can infer some general acceptance criteria, particularly from the method comparison study which uses correlation coefficient (r) and percent bias (%Bias) as metrics. The clinical sensitivity and specificity tables also present implied acceptance criteria based on the demonstrated performance.

    Inferred Acceptance Criteria & Reported Performance (Selection of Key Metrics)

    Study Type / Parameter CategoryAcceptance Criteria (Inferred)Reported Device Performance (Summary from text)
    Whole Blood Precision (Analyte-specific %CV)Met manufacturer's specifications or predefined acceptance criteria requirements.WBC: 0.30% to 2.76% CV (Repeatability); 0.97% to 1.98% (Reproducibility, within run)
    RBC: 0.45% to 0.97% CV (Repeatability); 0.73% to 1.03% (Reproducibility, within run)
    HGB: 0.38% to 0.79% CV (Repeatability); 0.40% to 0.98% (Reproducibility, within run)
    PLT-I: 1.30% to 8.32% CV (Repeatability); 1.59% to 3.70% (Reproducibility, within run)
    Body Fluid Precision (Analyte-specific %CV)Met manufacturer's specifications or predefined acceptance criteria requirements.WBC-BF: 2.01% to 3.91% CV (Repeatability); 2.01% to 3.91% (Reproducibility, within run)
    RBC-BF: 1.87% to 3.49% CV (Repeatability); 1.87% to 3.49% (Reproducibility, within run)
    Linearity (Whole Blood & Body Fluid)Linear from lower limit to upper limit and within measured maximum allowable deviation from linearity for each interval. (All results met predefined acceptance criteria).WBC (WB): 0.03 – 440.00 x10³/μL
    RBC (WB): 0.01 – 8.60 x10⁶/μL
    HGB (WB): 0.1 – 26.0 g/dL
    PLT (WB): 2 – 5,000 x10³/μL
    WBC-BF: 0.003 – 10.000 x10³/μL
    Method Comparison (Whole Blood: r-value)≥0.95 (explicitly stated for HGB, implied for others)WBC: 0.9997
    RBC: 0.9900
    HGB: 0.9915
    PLT-I: 0.9991
    Method Comparison (Whole Blood: %Bias)Within predefined bias limits (e.g., ±2% or 0.2g/dL for HGB)HGB: -1.41% (Note: One site showed -2.10% for HGB, slightly exceeding ±2% but deemed acceptable due to high r-value)
    WBC: 0.52%
    RBC: -0.83%
    Method Comparison (Body Fluid: r-value)Acceptance criteria not explicitly stated, but high correlation values reported (e.g., >0.99 for WBC-BF, RBC-BF, TC-BF)CSF WBC-BF: 0.9968
    Peritoneal WBC-BF: 0.9989
    Abnormal Flagging (Sensitivity/Specificity vs. Manual Microscopy)No explicit numerical acceptance criteria given for these.Any Distributional Abnormalities: Sensitivity 74.37%, Specificity 79.48%, OPA 76.31%
    Any Morphological Flag: Sensitivity 83.26%, Specificity 65.25%, OPA 70.77%
    Any Distributional and/or Morphological Abnormalities: Sensitivity 82.25%, Specificity 62.64%, OPA 75.38%
    Abnormal Flagging (PPA/NPA vs. Predicate XN-20)No explicit numerical acceptance criteria given for these.Any Distributional Abnormalities: PPA 94.74%, NPA 95.88%, OPA 95.20%
    Any Morphological Flag: PPA 92.29%, NPA 86.01%, OPA 89.10%
    Any Distributional and/or Morphological Abnormalities: PPA 96.37%, NPA 88.01%, OPA 93.73%

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size & Provenance:
      • Precision (Repeatability - Whole Blood): Residual K2EDTA whole blood samples for 10 replicates for target values, and three samples for other parameters. This was across three US clinical sites (Site 01, 05, 24).
      • Precision (Reproducibility - Whole Blood): XN CHECK whole blood control material, 90 results per control level (3 levels x 3 replicates x 2 runs x 5 days). Conducted at three US clinical sites.
      • Precision (Body Fluid): Residual peritoneal, pleural, and synovial fluid samples (K2EDTA) and CSF (no anticoagulant) for 10 replicates for target values. Conducted at three US clinical sites.
      • Linearity (Whole Blood & Body Fluid): Minimum of seven sample dilutions. Performed at one internal site.
      • Analytical Specificity/Interferences: Whole blood K2EDTA samples from donors. Number of samples not specified, but collected for this study purpose.
      • Sample Stability (Whole Blood): 8 unique leftover samples and 12 prospectively collected K2EDTA venous whole blood samples (20 samples total). Conducted at one internal site.
      • Sample Stability (Body Fluid): 12 unique de-identified leftover body fluid samples (3-CSF, 3-peritoneal, 3-pleural, 3-synovial). Conducted at 1 external site.
      • Detection Limit: Four blank samples and four low concentration samples for each parameter. Conducted across 2 XR-10 analyzers (implied internal or multi-site for the overall study context).
      • Carry-Over: High and low target concentration samples (number not specified). Conducted at three US clinical sites.
      • Comparison Studies (Whole Blood): 865 unique residual whole blood samples from pediatrics (<21 years) and adults (≥21 years), including various disease states. Collected across 3 US clinical sites. Samples were "residual" (retrospective).
      • Comparison Studies (Body Fluid): 397 residual body fluid samples (CSF, peritoneal, pleural, and synovial). Collected across three US sites. Samples were "residual" (retrospective).
      • Matrix Studies (Anticoagulant Comparison): 46 paired K2 and K3EDTA venous whole blood samples. Conducted at 1 internal site.
      • Matrix Studies (Venous vs. Capillary): 70 paired venous whole blood samples. Conducted at one internal site.
      • Matrix Studies (Normal vs. Micro-collection Tubes): 70 paired venous whole blood samples. Conducted at one internal site.
      • Bridging Studies (Whole Blood Mode to Pre-dilute Mode): 45 de-identified residual whole blood samples. Conducted at 1 internal site.
      • Bridging Studies (Predilute Mode Normal Tube to Micro-collection Tube): 40 de-identified residual whole blood samples. Conducted at one internal site.
      • Bridging Studies (Low WBC Mode Normal Tube to Micro collection Tube): 43 residual de-identified venous whole blood samples. Conducted at one internal site.
      • Clinical Studies (Sensitivity and Specificity): Patient samples representing a variety of abnormal conditions. 705 to 780 samples for manual microscopy comparison; 834 to 845 samples for XN-20 predicate comparison. Conducted at three external clinical sites (same as method comparison). This data appears retrospective as it involves "patient samples" and "manual differential counts and peripheral blood smear review".
      • Expected Values/Reference Range (Adult): 132 samples (58 males, 74 females). Data provenance implied from "verification of adult reference intervals" which suggests comparison against pre-existing intervals for a predicate device.
      • Expected Values/Reference Range (Pediatric): 196 pediatric samples (40 neonates, 55 infants, 60 children, 41 adolescents). Data provenance implied from "verification study" using established literature.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • For the Clinical Sensitivity and Specificity study (Abnormal Flagging – Manual Microscopy), ground truth was established by "manual differential counts and peripheral blood smear review by experienced examiners using light microscopy (reference method) at each of the three external clinical sites".
    • The exact number of experts and their specific qualifications (e.g., "radiologist with 10 years of experience") are not specified in the provided text. It merely states "experienced examiners."

    4. Adjudication Method for the Test Set

    • For the Clinical Sensitivity and Specificity study, an adjudication method for discrepancies between examiners or between the device and manual microscopy is not explicitly described. It states that the manual differential counts and peripheral blood smear review were the "reference method," implying these were the accepted ground truth. No mention of 2+1, 3+1, or any other formal adjudication process is present.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, a traditional MRMC comparative effectiveness study where human readers' performance with and without AI assistance is measured and compared was not performed, nor is it applicable to this type of device (an automated hematology analyzer). This device classifies and enumerates blood parameters automatically, without human "reading" or AI assistance in the human interpretation loop. The "clinical studies" described evaluate the device's flagging capabilities against manual microscopy (human expert consensus) and against a predicate automated device, not the improvement of human readers using the device's AI capabilities.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Yes, the entire performance evaluation of the Sysmex XR-Series (XR-10) is essentially a standalone (algorithm only) performance assessment. It is an automated hematology analyzer that quantifies and classifies blood parameters without direct human intervention or interpretive assistance during the measurement process. The "Comparison Studies" directly assess the XR-10's performance against a predicate automated device (XN-20), representing a standalone comparison. The "Clinical Sensitivity and Specificity" also evaluates the XR-10's automated flagging against manual microscopy (human expert consensus of ground truth slides), demonstrating its standalone diagnostic performance.

    7. The Type of Ground Truth Used

    • The ground truth varied depending on the type of study:
      • For Analytical Performance (Precision, Linearity, Detection Limit, Carry-Over): The ground truth was based on the performance of the reference methods, control materials, calibrators, and linearity materials, as well as manufacturer's specifications and CLSI guidelines. For LoD/LoQ, the Sysmex XN-20 (the predicate device) was used to assign reference values to low-level samples.
      • For Method Comparison Studies: The ground truth was implicitly the predicate device (Sysmex XN-20). The study aimed to show substantial equivalence between the new device and the predicate, not necessarily against a "true" biological gold standard across all parameters.
      • For Clinical Sensitivity and Specificity: The ground truth for abnormal flagging was established by manual differential counts and peripheral blood smear review by experienced examiners using light microscopy. This represents an expert consensus (from potentially multiple experts, though not detailed) based on visual examination.
      • For Reference Interval Verification: The ground truth was established by literature sources (Wong et al., 2021 for pediatric; Kjeldsberg's Body Fluids, 1993 for body fluids) or pre-established reference intervals for a predicate device (Sysmex XE-5000 for adult whole blood).

    8. The Sample Size for the Training Set

    • The FDA 510(k) clearance letter does not specify the sample size used for the training set for the Sysmex XR-Series (XR-10) Automated Hematology Analyzer. It focuses solely on the validation/test performance data, which is typical for 510(k) submissions where the device's performance is demonstrated against established methods or predicate devices. This indicates that information on the internal development and training of the device's algorithms is not part of the publicly available clearance letter.

    9. How the Ground Truth for the Training Set Was Established

    • Since the training set size is not provided, the method for establishing its ground truth is also not described in this document. It is generally assumed that manufacturers use robust internal processes for data collection, annotation, and ground truth establishment during the development phase of such automated diagnostic systems (e.g., using a combination of expert review, reference methods, and potentially pathological or clinical outcomes data, depending on the specific parameter and its clinical relevance). However, this specific information is not included in the 510(k) summary.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 123