(89 days)
The XR-Series module (XR-10) is a quantitative multi-parameter automated hematology analyzer intended for in vitro diagnostic use in screening patient populations found in clinical laboratories.
The XR-Series module classifies and enumerates the following parameters in whole blood: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT (PLT-I, PLT-F), NEUT%/#, LYMPH%/#, MONO%/#, EO%/#, BASO%/#, IG%/#, RDW-CV, RDW-SD, MPV, NRBC%/#, RET%/#, IPF, IPF#, IRF, RET-He and has a Body Fluid mode for body fluids. The Body Fluid mode enumerates the WBC-BF, RBC-BF, MN%/#, PMN%/#, and TC-BF# parameters in cerebrospinal fluid (CSF), serous fluids (peritoneal, pleural) and synovial fluids. Whole blood should be collected in K2EDTA or K3EDTA anticoagulant, and serous and synovial fluids in K2EDTA anticoagulant to prevent clotting of fluid. The use of anticoagulants with CSF specimens is neither required nor recommended.
The Sysmex XR-Series module (XR-10) is a quantitative multi-parameter hematology analyzer intended to perform tests on whole blood samples collected in K2 or K3EDTA and body fluids (pleural, peritoneal and synovial) collected in K2EDTA anticoagulant. The analyzers can also perform tests on CSF, which should not be collected in any anticoagulant. The XR-Series analyzer consist of four principal units: (1) One Main Units (XR-10) which aspirates, dilutes, mixes, and analyzes blood and body fluid samples; (2) Two Auto Sampler Units (SA-10, SA-01) which supply samples to the Main Unit automatically; (3) IPU (Information Processing Unit) which processes data from the Main Unit and provides the operator interface with the system; (4) Pneumatic Unit which supplies pressure and vacuum from the Main Unit.
This document describes the acceptance criteria and the studies conducted to prove that the Sysmex XR-Series (XR-10) Automated Hematology Analyzer meets these criteria, demonstrating substantial equivalence to its predicate device, the Sysmex XN-20.
1. Table of Acceptance Criteria and Reported Device Performance
The FDA 510(k) clearance letter does not explicitly present a neatly formatted table of acceptance criteria alongside the reported performance for all parameters. Instead, it describes various performance studies (Precision, Linearity, Analytical Specificity/Interferences, Sample Stability, Detection Limit, Carry-Over, Comparison Studies, Matrix Studies, Bridging Studies, Clinical Studies, and Expected Values/Reference Range) and states that the XR-10 met "manufacturer's specifications or predefined acceptance criteria requirements" for each.
However, based on the provided data, we can infer some general acceptance criteria, particularly from the method comparison study which uses correlation coefficient (r) and percent bias (%Bias) as metrics. The clinical sensitivity and specificity tables also present implied acceptance criteria based on the demonstrated performance.
Inferred Acceptance Criteria & Reported Performance (Selection of Key Metrics)
| Study Type / Parameter Category | Acceptance Criteria (Inferred) | Reported Device Performance (Summary from text) |
|---|---|---|
| Whole Blood Precision (Analyte-specific %CV) | Met manufacturer's specifications or predefined acceptance criteria requirements. | WBC: 0.30% to 2.76% CV (Repeatability); 0.97% to 1.98% (Reproducibility, within run) |
| RBC: 0.45% to 0.97% CV (Repeatability); 0.73% to 1.03% (Reproducibility, within run) | ||
| HGB: 0.38% to 0.79% CV (Repeatability); 0.40% to 0.98% (Reproducibility, within run) | ||
| PLT-I: 1.30% to 8.32% CV (Repeatability); 1.59% to 3.70% (Reproducibility, within run) | ||
| Body Fluid Precision (Analyte-specific %CV) | Met manufacturer's specifications or predefined acceptance criteria requirements. | WBC-BF: 2.01% to 3.91% CV (Repeatability); 2.01% to 3.91% (Reproducibility, within run) |
| RBC-BF: 1.87% to 3.49% CV (Repeatability); 1.87% to 3.49% (Reproducibility, within run) | ||
| Linearity (Whole Blood & Body Fluid) | Linear from lower limit to upper limit and within measured maximum allowable deviation from linearity for each interval. (All results met predefined acceptance criteria). | WBC (WB): 0.03 – 440.00 x10³/μL |
| RBC (WB): 0.01 – 8.60 x10⁶/μL | ||
| HGB (WB): 0.1 – 26.0 g/dL | ||
| PLT (WB): 2 – 5,000 x10³/μL | ||
| WBC-BF: 0.003 – 10.000 x10³/μL | ||
| Method Comparison (Whole Blood: r-value) | ≥0.95 (explicitly stated for HGB, implied for others) | WBC: 0.9997 |
| RBC: 0.9900 | ||
| HGB: 0.9915 | ||
| PLT-I: 0.9991 | ||
| Method Comparison (Whole Blood: %Bias) | Within predefined bias limits (e.g., ±2% or 0.2g/dL for HGB) | HGB: -1.41% (Note: One site showed -2.10% for HGB, slightly exceeding ±2% but deemed acceptable due to high r-value) |
| WBC: 0.52% | ||
| RBC: -0.83% | ||
| Method Comparison (Body Fluid: r-value) | Acceptance criteria not explicitly stated, but high correlation values reported (e.g., >0.99 for WBC-BF, RBC-BF, TC-BF) | CSF WBC-BF: 0.9968 |
| Peritoneal WBC-BF: 0.9989 | ||
| Abnormal Flagging (Sensitivity/Specificity vs. Manual Microscopy) | No explicit numerical acceptance criteria given for these. | Any Distributional Abnormalities: Sensitivity 74.37%, Specificity 79.48%, OPA 76.31% |
| Any Morphological Flag: Sensitivity 83.26%, Specificity 65.25%, OPA 70.77% | ||
| Any Distributional and/or Morphological Abnormalities: Sensitivity 82.25%, Specificity 62.64%, OPA 75.38% | ||
| Abnormal Flagging (PPA/NPA vs. Predicate XN-20) | No explicit numerical acceptance criteria given for these. | Any Distributional Abnormalities: PPA 94.74%, NPA 95.88%, OPA 95.20% |
| Any Morphological Flag: PPA 92.29%, NPA 86.01%, OPA 89.10% | ||
| Any Distributional and/or Morphological Abnormalities: PPA 96.37%, NPA 88.01%, OPA 93.73% |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size & Provenance:
- Precision (Repeatability - Whole Blood): Residual K2EDTA whole blood samples for 10 replicates for target values, and three samples for other parameters. This was across three US clinical sites (Site 01, 05, 24).
- Precision (Reproducibility - Whole Blood): XN CHECK whole blood control material, 90 results per control level (3 levels x 3 replicates x 2 runs x 5 days). Conducted at three US clinical sites.
- Precision (Body Fluid): Residual peritoneal, pleural, and synovial fluid samples (K2EDTA) and CSF (no anticoagulant) for 10 replicates for target values. Conducted at three US clinical sites.
- Linearity (Whole Blood & Body Fluid): Minimum of seven sample dilutions. Performed at one internal site.
- Analytical Specificity/Interferences: Whole blood K2EDTA samples from donors. Number of samples not specified, but collected for this study purpose.
- Sample Stability (Whole Blood): 8 unique leftover samples and 12 prospectively collected K2EDTA venous whole blood samples (20 samples total). Conducted at one internal site.
- Sample Stability (Body Fluid): 12 unique de-identified leftover body fluid samples (3-CSF, 3-peritoneal, 3-pleural, 3-synovial). Conducted at 1 external site.
- Detection Limit: Four blank samples and four low concentration samples for each parameter. Conducted across 2 XR-10 analyzers (implied internal or multi-site for the overall study context).
- Carry-Over: High and low target concentration samples (number not specified). Conducted at three US clinical sites.
- Comparison Studies (Whole Blood): 865 unique residual whole blood samples from pediatrics (<21 years) and adults (≥21 years), including various disease states. Collected across 3 US clinical sites. Samples were "residual" (retrospective).
- Comparison Studies (Body Fluid): 397 residual body fluid samples (CSF, peritoneal, pleural, and synovial). Collected across three US sites. Samples were "residual" (retrospective).
- Matrix Studies (Anticoagulant Comparison): 46 paired K2 and K3EDTA venous whole blood samples. Conducted at 1 internal site.
- Matrix Studies (Venous vs. Capillary): 70 paired venous whole blood samples. Conducted at one internal site.
- Matrix Studies (Normal vs. Micro-collection Tubes): 70 paired venous whole blood samples. Conducted at one internal site.
- Bridging Studies (Whole Blood Mode to Pre-dilute Mode): 45 de-identified residual whole blood samples. Conducted at 1 internal site.
- Bridging Studies (Predilute Mode Normal Tube to Micro-collection Tube): 40 de-identified residual whole blood samples. Conducted at one internal site.
- Bridging Studies (Low WBC Mode Normal Tube to Micro collection Tube): 43 residual de-identified venous whole blood samples. Conducted at one internal site.
- Clinical Studies (Sensitivity and Specificity): Patient samples representing a variety of abnormal conditions. 705 to 780 samples for manual microscopy comparison; 834 to 845 samples for XN-20 predicate comparison. Conducted at three external clinical sites (same as method comparison). This data appears retrospective as it involves "patient samples" and "manual differential counts and peripheral blood smear review".
- Expected Values/Reference Range (Adult): 132 samples (58 males, 74 females). Data provenance implied from "verification of adult reference intervals" which suggests comparison against pre-existing intervals for a predicate device.
- Expected Values/Reference Range (Pediatric): 196 pediatric samples (40 neonates, 55 infants, 60 children, 41 adolescents). Data provenance implied from "verification study" using established literature.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- For the Clinical Sensitivity and Specificity study (Abnormal Flagging – Manual Microscopy), ground truth was established by "manual differential counts and peripheral blood smear review by experienced examiners using light microscopy (reference method) at each of the three external clinical sites".
- The exact number of experts and their specific qualifications (e.g., "radiologist with 10 years of experience") are not specified in the provided text. It merely states "experienced examiners."
4. Adjudication Method for the Test Set
- For the Clinical Sensitivity and Specificity study, an adjudication method for discrepancies between examiners or between the device and manual microscopy is not explicitly described. It states that the manual differential counts and peripheral blood smear review were the "reference method," implying these were the accepted ground truth. No mention of 2+1, 3+1, or any other formal adjudication process is present.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, a traditional MRMC comparative effectiveness study where human readers' performance with and without AI assistance is measured and compared was not performed, nor is it applicable to this type of device (an automated hematology analyzer). This device classifies and enumerates blood parameters automatically, without human "reading" or AI assistance in the human interpretation loop. The "clinical studies" described evaluate the device's flagging capabilities against manual microscopy (human expert consensus) and against a predicate automated device, not the improvement of human readers using the device's AI capabilities.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Yes, the entire performance evaluation of the Sysmex XR-Series (XR-10) is essentially a standalone (algorithm only) performance assessment. It is an automated hematology analyzer that quantifies and classifies blood parameters without direct human intervention or interpretive assistance during the measurement process. The "Comparison Studies" directly assess the XR-10's performance against a predicate automated device (XN-20), representing a standalone comparison. The "Clinical Sensitivity and Specificity" also evaluates the XR-10's automated flagging against manual microscopy (human expert consensus of ground truth slides), demonstrating its standalone diagnostic performance.
7. The Type of Ground Truth Used
- The ground truth varied depending on the type of study:
- For Analytical Performance (Precision, Linearity, Detection Limit, Carry-Over): The ground truth was based on the performance of the reference methods, control materials, calibrators, and linearity materials, as well as manufacturer's specifications and CLSI guidelines. For LoD/LoQ, the Sysmex XN-20 (the predicate device) was used to assign reference values to low-level samples.
- For Method Comparison Studies: The ground truth was implicitly the predicate device (Sysmex XN-20). The study aimed to show substantial equivalence between the new device and the predicate, not necessarily against a "true" biological gold standard across all parameters.
- For Clinical Sensitivity and Specificity: The ground truth for abnormal flagging was established by manual differential counts and peripheral blood smear review by experienced examiners using light microscopy. This represents an expert consensus (from potentially multiple experts, though not detailed) based on visual examination.
- For Reference Interval Verification: The ground truth was established by literature sources (Wong et al., 2021 for pediatric; Kjeldsberg's Body Fluids, 1993 for body fluids) or pre-established reference intervals for a predicate device (Sysmex XE-5000 for adult whole blood).
8. The Sample Size for the Training Set
- The FDA 510(k) clearance letter does not specify the sample size used for the training set for the Sysmex XR-Series (XR-10) Automated Hematology Analyzer. It focuses solely on the validation/test performance data, which is typical for 510(k) submissions where the device's performance is demonstrated against established methods or predicate devices. This indicates that information on the internal development and training of the device's algorithms is not part of the publicly available clearance letter.
9. How the Ground Truth for the Training Set Was Established
- Since the training set size is not provided, the method for establishing its ground truth is also not described in this document. It is generally assumed that manufacturers use robust internal processes for data collection, annotation, and ground truth establishment during the development phase of such automated diagnostic systems (e.g., using a combination of expert review, reference methods, and potentially pathological or clinical outcomes data, depending on the specific parameter and its clinical relevance). However, this specific information is not included in the 510(k) summary.
U.S. Food & Drug Administration 510(k) Clearance Letter
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.07.05
June 25, 2025
Sysmex America, Inc.
Yvonne Doswell
Senior Scientist
577 Aptakisic Road
Lincolnshire, Illinois 60069
Re: K250943
Trade/Device Name: Sysmex XR-Series (XR-10) Automated Hematology Analyzer
Regulation Number: 21 CFR 864.5220
Regulation Name: Automated differential cell counter
Regulatory Class: Class II
Product Code: GKZ
Dated: March 28, 2025
Received: March 28, 2025
Dear Yvonne Doswell:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Takeesha Taylor-bell -S
Takeesha Taylor-Bell
Deputy Director
Division of Immunology and Hematology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
Indications for Use
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known): K250943
Device Name: Sysmex XR-Series (XR-10) Automated Hematology Analyzer
Indications for Use (Describe)
The XR-Series module (XR-10) is a quantitative multi-parameter automated hematology analyzer intended for in vitro diagnostic use in screening patient populations found in clinical laboratories.
The XR-Series module classifies and enumerates the following parameters in whole blood: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT (PLT-I, PLT-F), NEUT%/#, LYMPH%/#, MONO%/#, EO%/#, BASO%/#, IG%/#, RDW-CV, RDW-SD, MPV, NRBC%/#, RET%/#, IPF, IPF#, IRF, RET-He and has a Body Fluid mode for body fluids. The Body Fluid mode enumerates the WBC-BF, RBC-BF, MN%/#, PMN%/#, and TC-BF# parameters in cerebrospinal fluid (CSF), serous fluids (peritoneal, pleural) and synovial fluids. Whole blood should be collected in K2EDTA or K3EDTA anticoagulant, and serous and synovial fluids in K2EDTA anticoagulant to prevent clotting of fluid. The use of anticoagulants with CSF specimens is neither required nor recommended.
Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 5
510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K250943.
Submitter's name, address, telephone number, contact person, and date the summary was prepared:
Submitter's Name: Sysmex America, Inc.
Submitter's Address: 577 Aptakisic Road
Lincolnshire, IL 60069
Submitter's Contact: Yvonne Doswell
Senior Scientist, Regulatory Affairs
E-Mail: doswelly@sysmex.com
Phone: (678) 274-8024
Date 510(k) Summary Prepared: March 23, 2025
Name of the device, including the trade or proprietary name, the common or usual name, and the classification name:
Proprietary Name: Sysmex XR-Series (XR-10) Automated Hematology Analyzer
Common Name: Automated Hematology Analyzer
Regulation Description: Automated Differential Cell Counter
Regulation Section: 21 CFR 864.5220
Device Class: 2
Product Code: GKZ
Related Reagents, Controls and Calibrator:
Product Code: 81GIF
- CELLPACK DCL (Diluent)
- CELLPACK DST (Diluent)
- CELLPACK DFL (Diluent)
Product Code: 81GGK
- Lysercell WNR (Lyse)
- Lysercell WDF II (Lyse)
- SULFOLYSER (Lyse)
Product Code: 81KQC
- Fluorocell WNR (Stain)
- Fluorocell WDF (Stain)
- Fluorocell RET (Stain)
- Fluorocell PLT (Stain)
Product Code: 81KSA
Page 6
- XN CAL (Calibrator)
- XN CAL PF (Calibrator)
Product Code: 81JPK
- XN CHECK (Quality Control)
- XN CHECK BF (Quality Control)
Product Code: 81JCB
- CELLCLEAN AUTO (Detergent)
Predicate Device and 510(k) number: Sysmex XN-Series (XN-10, XN-20) Automated Hematology Analyzer, K112605
Description of the Device:
The Sysmex XR-Series module (XR-10) is a quantitative multi-parameter hematology analyzer intended to perform tests on whole blood samples collected in K2 or K3EDTA and body fluids (pleural, peritoneal and synovial) collected in K2EDTA anticoagulant. The analyzers can also perform tests on CSF, which should not be collected in any anticoagulant. The XR-Series analyzer consist of four principal units: (1) One Main Units (XR-10) which aspirates, dilutes, mixes, and analyzes blood and body fluid samples; (2) Two Auto Sampler Units (SA-10, SA-01) which supply samples to the Main Unit automatically; (3) IPU (Information Processing Unit) which processes data from the Main Unit and provides the operator interface with the system; (4) Pneumatic Unit which supplies pressure and vacuum from the Main Unit.
Principles of Operation:
The XR-10 analyzer performs analysis using the following methods: RF/DC Detection Method, Sheath Flow DC Detection Method, and Flow Cytometry Methods using a Semiconductor Laser and SLS-hemoglobin. The RF/DC detection method detects the size of the cells by changes in direct-current resistance and the density of the cell interior by changes in radio-frequency resistance. In the sheath flow method, cells pass through the aperture of the detector surrounded by sheath fluid. Flow cytometry is also used where a semiconductor laser beam is emitted to the cells passing through the flow cell. The forward scattered light is received by the photodiode, and the lateral scattered light and lateral fluorescent light are received by the photo multiplier tube. This light is converted into electrical pulses, thus making it possible to obtain cell information. Particle characterization and identification is based on detection of forward scatter, fluorescence and adaptive cluster analysis. The system carries out all processes automatically from aspiration of the sample to outputting results and uses Microsoft Windows Operating System.
The body fluid analysis mode of the XR-10 analyzer uses the 4 part differential scattergram and the RBC distribution obtained from a specialized analysis sequence to calculate and display the WBC (WBC-BF) counts, mononuclear cell (MN) / polymorphonuclear cell (PMN) counts and percentages, TC-BF (Total Count) & RBC (RBC-BF) counts found in the body fluid.
Analysis results and graphics are displayed on the IPU screen. They can be printed on any of the available printers or transmitted to a host computer.
Page 7
Statement of Intended Use:
The XR-Series module (XR-10) is a quantitative multi-parameter automated hematology analyzer intended for in vitro diagnostic use in screening patient populations found in clinical laboratories.
The XR-Series module classifies and enumerates the following parameters in whole blood: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT (PLT-I, PLT-F), NEUT%/#, LYMPH%/#, MONO%/#, EO%/#, BASO%/#, IG%/#, RDW-CV, RDW-SD, MPV, NRBC%/#, RET%/#, IPF, IPF#, IRF, RET-He and has a Body Fluid mode for body fluids. The Body Fluid mode enumerates the WBC-BF, RBC-BF, MN%/#, PMN%/#, and TC-BF# parameters in cerebrospinal fluid (CSF), serous fluids (peritoneal, pleural) and synovial fluids. Whole blood should be collected in K2EDTA or K3EDTA anticoagulant, and serous and synovial fluids in K2EDTA anticoagulant to prevent clotting of fluid. The use of anticoagulants with CSF specimens is neither required nor recommended.
Summary of Substantial Equivalence:
Table 1 compares the Sysmex XR-Series (XR-10) Automated Hematology Analyzer with the XN-20 Automated Hematology analyzer.
Table 1: Comparison of the Predicate XN-20 and the Proposed Sysmex XR-Series (XR-10) Automated Hematology Analyzer
| Item | Predicate Analyzer XN-Series(XN-20) K112605 | Proposed Analyzer XR-Series (XR-10) K250943 |
|---|---|---|
| Similarities | ||
| Intended Use | The XN-Series modules (XN-10, XN-20) are quantitative multi-parameter automated hematology analyzers intended for in vitro diagnostic use in screening patient populations found in clinical laboratories. The XN-Series modules classify and enumerate the following parameters in whole blood: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT, NEUT%/#, LYMPH%/#, MONO%/#, EO%/#, BASO%/#, IG%/#, RDW-CV, RDW-SD, MPV, NRBC%/#, RET%/#, IPF, IRF, RET-He and has a Body Fluid mode for body fluids. The Body Fluid mode enumerates the WBC-BF, RBC-BF, MN%/#, PMN%/# and TC-BF parameters in cerebrospinal fluid (CSF), serous fluids (peritoneal, pleural) and synovial fluids. Whole blood should be collected in K₂ or K₃EDTA anticoagulant and, Serous and Synovial fluids in K2EDTA anticoagulant to prevent clotting of fluid. The use of anticoagulants | The XR-Series module (XR-10) is a quantitative multi-parameter automated hematology analyzer intended for in vitro diagnostic use in screening patient populations found in clinical laboratories. The XR-Series module classifies and enumerates the following parameters in whole blood: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT (PLT-I, PLT-F), NEUT%/#, LYMPH%/#, MONO%/#, EO%/#, BASO%/#, IG%/#, RDW-CV, RDW-SD, MPV, NRBC%/#, RET%/#, IPF, IPF#, IRF, RET-He and has a Body Fluid mode for body fluids. The Body Fluid mode enumerates the WBC-BF, RBC-BF, MN%/#, PMN%/#, and TC-BF# parameters in cerebrospinal fluid (CSF), serous fluids (peritoneal, pleural) and synovial fluids. Whole blood should be collected in K2EDTA or K3EDTA anticoagulant, and serous and synovial fluids in K2EDTA anticoagulant to prevent clotting of fluid. The use of |
Page 8
| Item | Predicate Analyzer XN-Series(XN-20) K112605 | Proposed Analyzer XR-Series (XR-10) K250943 |
|---|---|---|
| with CSF specimens is neither required nor recommended. | anticoagulants with CSF specimens is neither required nor recommended. | |
| Specimen Type | Whole Blood collected in K₂ or K₃EDTA | SAME |
| Measurement Principle | Performs hematology analyses according to the Hydro Dynamic Focusing (DC Detection), Flow cytometry method using semiconductor laser SLS-Hemoglobin Method | SAME |
| Parameters | Whole Blood Mode: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT (PLT-I, PLT-F), NEUT%/#, LYMPH%/#, MONO%/#, EO%/#, BASO%/#, IG%/#, RDW-CV,RDW-SD, MPV, NRBC%/#, RET%/#, IRF, IPF, IPF#¹, RET-He Body Fluid Mode: WBC-BF, RBC-BF, MN%/#, PMN%/#, TC-BF# | SAME |
| Reagents | CELLPACK DCL (Diluent) CELLPACK DST (Diluent) CELLPACK DFL (Diluent) Lysercell WNR (Lyse) SULFOLYSER (Lyse) Fluorocell WNR (Stain) Fluorocell WDF (Stain) Fluorocell RET (Stain) Fluorocell PLT (Stain) CELLCLEAN AUTO (Detergent) | SAME |
| Controls/ Calibrators | XN CAL (Calibrator) XN CAL PF (Calibrator) XN CHECK (Quality Control) XN CHECK BF (Quality Control) | SAME |
| Types of Analysis | Manual analysis Sampler analysis | SAME |
| Analysis Modes | Whole Blood mode Low WBC mode Pre-Dilution mode Body Fluid mode | SAME |
| Sample Aspiration/ Fluidic Pathway | Single Pathway | SAME |
| Measuring Channels | WNR, WDF, RBC/PLT, RET, PLT-F | SAME |
Page 9
| Item | Predicate Analyzer XN-Series(XN-20) K112605 | Proposed Analyzer XR-Series (XR-10) K250943 |
|---|---|---|
| Throughput | Pre-Dilution mode: Approximately 90 samples/hour Body Fluid 40 samples/hour maximum | SAME |
| Sample Aspiration Volumes | Sampler Mode – 88 μL Manual (Closed tube) Mode – 88 μL Manual (Open tube) Mode – 88 μL Dilution Mode – 70 μL Body Fluid Mode – 88 μL | SAME |
| Software Specifications | Samples stored: 100,000 samples Patient information: 10,000 records Wards registered: 200 wards Doctor names registered: 200 names Analysis registration function: 2,000 records QC files: 99 files per analysis module (300 plots per file) Reagent replacement history: 5,000 records Maintenance history: 5,000 records | SAME |
| Differences | ||
| Measuring Channels | WPC | Not Available |
| Reagents | Lysercell WDF (Lyse) Lysercell WPC (Lyse) Fluorocell WPC (Stain) | Lysercell WDF II (Lyse) Not Available Not Available |
| Throughput | Whole Blood Mode: 100 samples/hour maximum depending on mode used. | Whole Blood Mode: 110 samples/hour maximum depending on mode used. |
Please note: Intended use for the predicate analyzer was cleared in submission K112605 (XN-Series modules XN-10, XN-20). All other information listed for the predicate analyzer refers to the XN 20 module. ¹ IPF# was added to the XN-Series under K141964.
The XR-Series (XR-10) analyzer's Indications for Use statement is similar to the predicate device with minor variation. The XR-Series (XR-10) analyzer also has similar technological characteristics to the predicate device with minor variations. Both devices measure similar parameters and utilize most of the same reagents, controls, calibrators, and cleaning detergent. The data collection software functionality, communication method with data management software functionality, monitor software, connectivity, and communication are similar to the predicate device.
In order to demonstrate that differences in technological characteristics between the subject device and predicate device do not impact safety and effectiveness, the following clinical performance studies were conducted utilizing the XR-Series (XR-10) analyzer.
Page 10
Summary of Performance Testing:
Clinical testing was conducted on the XR-Series (XR-10) analyzer to show equivalent performance to the XN-20 analyzer. Testing included:
Analytical Performance
Precision (Repeatability) – Whole Blood Mode
Whole blood repeatability studies were performed to evaluate within-run imprecision of whole blood parameter results when measured in replicates of ten on Sysmex XR-Series (XR-10) analyzers at three US clinical sites. The study was conducted using residual K2EDTA whole blood samples with concentrations targeting medical decision levels, normal and high measurement range of WBC, HGB and PLT parameters and the low, normal, and high measurement range of RBC and HCT parameters. For all other claimed parameters, testing was completed using three samples (1-3) per parameter. Testing was conducted in accordance with the CLSI EP05-A3 approved guideline. The mean, standard deviation (SD), and coefficient of variation (%CV) were calculated for each parameter. The %CV estimates were evaluated against acceptance criteria. The XR-10 met manufacturer's specifications or predefined acceptance criteria requirements. The table below summarizes results from XR-10.
Page 11
| Measurand | Sample | Site | N | Range | Mean | SD | CV % |
|---|---|---|---|---|---|---|---|
| WBC (10³/μL) | MDL | 01 | 10 | 0.70-0.74 | 0.73 | 0.01 | 1.97 |
| 05 | 10 | 1.32-1.44 | 1.36 | 0.04 | 2.76 | ||
| 24 | 10 | 1.17-1.28 | 1.23 | 0.03 | 2.44 | ||
| Normal | 01 | 10 | 4.79-5.21 | 5.02 | 0.10 | 2.09 | |
| 05 | 10 | 5.98-6.21 | 6.13 | 0.08 | 1.29 | ||
| 24 | 10 | 7.44-7.86 | 7.63 | 0.14 | 1.89 | ||
| High | 01 | 10 | 101.91-103.43 | 102.88 | 0.44 | 0.43 | |
| 05 | 10 | 103.71-104.75 | 104.08 | 0.31 | 0.30 | ||
| 24 | 10 | 81.04-82.39 | 81.88 | 0.49 | 0.60 | ||
| RBC (10⁶/μL) | Low | 01 | 10 | 1.96-1.99 | 1.98 | 0.01 | 0.55 |
| 05 | 10 | 1.97-2.03 | 2.01 | 0.02 | 0.79 | ||
| 24 | 10 | 1.51-1.55 | 1.53 | 0.01 | 0.97 | ||
| Normal | 01 | 10 | 4.04-4.15 | 4.11 | 0.03 | 0.81 | |
| 05 | 10 | 3.84-3.90 | 3.87 | 0.02 | 0.48 | ||
| 24 | 10 | 4.69-4.85 | 4.75 | 0.05 | 0.96 | ||
| High | 01 | 10 | 6.10-6.24 | 6.20 | 0.04 | 0.61 | |
| 05 | 10 | 6.04-6.27 | 6.12 | 0.07 | 1.17 | ||
| 24 | 10 | 6.40-6.50 | 6.45 | 0.03 | 0.45 | ||
| HGB (g/dL) | MDL | 01 | 10 | 7.0-7.1 | 7.0 | 0.04 | 0.60 |
| 05 | 10 | 6.2-6.3 | 6.3 | 0.05 | 0.77 | ||
| 24 | 10 | 6.5-6.6 | 6.6 | 0.05 | 0.79 | ||
| Normal | 01 | 10 | 13.1-13.3 | 13.2 | 0.06 | 0.48 | |
| 05 | 10 | 14.7-14.9 | 14.8 | 0.06 | 0.38 | ||
| 24 | 10 | 14.3-14.6 | 14.4 | 0.11 | 0.73 | ||
| High | 01 | 10 | 17.4-17.6 | 17.5 | 0.07 | 0.39 | |
| 05 | 10 | 18.7-18.9 | 18.9 | 0.07 | 0.38 | ||
| 24 | 10 | 21.2-21.6 | 21.4 | 0.13 | 0.59 |
Page 12
| Measurand | Sample | Site | N | Range | Mean | SD | CV % |
|---|---|---|---|---|---|---|---|
| MCHC (g/dL) | 1 | 01 | 10 | 28.9-29.6 | 29.2 | 0.23 | 0.80 |
| 05 | 10 | 27.7-28.9 | 28.3 | 0.32 | 1.13 | ||
| 24 | 10 | 28.5-29.3 | 28.8 | 0.26 | 0.92 | ||
| 2 | 01 | 10 | 31.3-32.2 | 31.7 | 0.28 | 0.88 | |
| 05 | 10 | 30.4-31.4 | 30.8 | 0.31 | 1.02 | ||
| 24 | 10 | 36.6-37.6 | 37.2 | 0.28 | 0.75 | ||
| 3 | 01 | 10 | 33.8-35.1 | 34.5 | 0.36 | 1.03 | |
| 05 | 10 | 31.9-32.6 | 32.2 | 0.29 | 0.89 | ||
| 24 | 10 | 38.5-40.6 | 39.5 | 0.75 | 1.91 | ||
| PLT-I (10³/μL) | MDL | 01 | 10 | 25-29 | 27 | 1.10 | 4.06 |
| 05 | 10 | 27-33 | 30 | 1.7 | 5.62 | ||
| 24 | 10 | 19-25 | 22 | 1.81 | 8.32 | ||
| Normal | 01 | 10 | 210-240 | 222 | 8.50 | 3.83 | |
| 05 | 10 | 156-164 | 159 | 3.2 | 2.01 | ||
| 24 | 10 | 224-241 | 235 | 6.04 | 2.57 | ||
| High | 01 | 10 | 885-931 | 911 | 15.62 | 1.71 | |
| 05 | 10 | 947-992 | 969 | 12.62 | 1.30 | ||
| 24 | 10 | 754-799 | 768 | 13.69 | 1.78 | ||
| PLT-F (10³/μL) | MDL | 01 | 10 | 7-8 | 8 | 0.42 | 5.41 |
| 05 | 10 | 21-23 | 22 | 0.79 | 3.62 | ||
| 24 | 10 | 13-15 | 13 | 0.70 | 5.22 | ||
| Normal | 01 | 10 | 207-225 | 217 | 6.54 | 3.01 | |
| 05 | 10 | 167-179 | 172 | 3.03 | 1.76 | ||
| 24 | 10 | 149-155 | 152 | 1.69 | 1.11 | ||
| High | 01 | 10 | 1020-1052 | 1040 | 9.19 | 0.88 | |
| 05 | 10 | 853-886 | 875 | 10.74 | 1.23 | ||
| 24 | 10 | 871-907 | 889 | 10.10 | 1.14 |
Page 13
| Measurand | Sample | Site | N | Range | Mean | SD | CV % |
|---|---|---|---|---|---|---|---|
| RDW-SD (fL) | 1 | 01 | 10 | 42.4-43.2 | 42.7 | 0.25 | 0.59 |
| 05 | 10 | 43.4-44.7 | 44.0 | 0.34 | 0.78 | ||
| 24 | 10 | 40.6-42.3 | 41.4 | 0.63 | 1.52 | ||
| 2 | 01 | 10 | 50.3-53.0 | 52.2 | 0.71 | 1.37 | |
| 05 | 10 | 65.4-67.1 | 66.2 | 0.51 | 0.77 | ||
| 24 | 10 | 50.4-52.1 | 51.1 | 0.71 | 1.38 | ||
| 3 | 01 | 10 | 90.2-91.7 | 90.7 | 0.48 | 0.53 | |
| 05 | 10 | 81.6-84.5 | 83.5 | 0.87 | 1.04 | ||
| 24 | 10 | 68.5-72.5 | 70.5 | 1.13 | 1.60 | ||
| RDW-CV (%) | 1 | 01 | 10 | 13.0-13.2 | 13.1 | 0.06 | 0.43 |
| 05 | 10 | 12.9-13.2 | 13.1 | 0.10 | 0.74 | ||
| 24 | 10 | 13.2-13.4 | 13.2 | 0.07 | 0.51 | ||
| 2 | 01 | 10 | 17.2-17.6 | 17.4 | 0.14 | 0.79 | |
| 05 | 10 | 15.4-15.6 | 15.5 | 0.07 | 0.43 | ||
| 24 | 10 | 17.9-18.8 | 18.4 | 0.28 | 1.50 | ||
| 3 | 01 | 10 | 22.2-22.7 | 22.5 | 0.13 | 0.59 | |
| 05 | 10 | 20.7-21.2 | 21.0 | 0.14 | 0.65 | ||
| 24 | 10 | 23.9-24.8 | 24.3 | 0.36 | 1.46 | ||
| MPV (fL) | 1 | 01 | 10 | 9.1-9.5 | 9.3 | 0.13 | 1.34 |
| 05 | 10 | 9.1-9.2 | 9.2 | 0.05 | 0.56 | ||
| 24 | 10 | 8.8-9.0 | 8.9 | 0.07 | 0.76 | ||
| 2 | 01 | 10 | 10.7-10.9 | 10.8 | 0.08 | 0.76 | |
| 05 | 10 | 11.3-12.3 | 11.7 | 0.32 | 2.77 | ||
| 24 | 10 | 10.7-11.2 | 11.0 | 0.15 | 1.41 | ||
| 3 | 01 | 10 | 14.3-14.6 | 14.5 | 0.09 | 0.66 | |
| 05 | 9 | 13.2-14.0 | 13.7 | 0.26 | 1.91 | ||
| 24 | 10 | 11.5-12.4 | 12.0 | 0.28 | 2.35 |
Page 14
| Measurand | Sample | Site | N | Range | Mean | SD | CV % |
|---|---|---|---|---|---|---|---|
| NRBC (10³/μL) | 1 | 01 | 10 | 0.02-0.05 | 0.04 | 0.01 | 24.28 |
| 05 | 10 | 0.03-0.05 | 0.04 | 0.01 | 15.06 | ||
| 24 | 10 | 0.02-0.05 | 0.04 | 0.01 | 24.28 | ||
| 2 | 01 | 10 | 0.82-0.90 | 0.85 | 0.03 | 3.17 | |
| 05 | 10 | 0.03-0.06 | 0.05 | 0.01 | 24.00 | ||
| 24 | 10 | 0.10-0.18 | 0.14 | 0.03 | 20.76 | ||
| 3 | 01 | 10 | 0.25-0.35 | 0.31 | 0.03 | 11.07 | |
| 05 | 10 | 0.96-1.03 | 0.99 | 0.02 | 2.41 | ||
| 24 | 10 | 1.28-1.35 | 1.31 | 0.02 | 1.79 | ||
| NRBC (%) | 1 | 01 | 10 | 0.5-1.1 | 0.8 | 0.18 | 22.12 |
| 05 | 10 | 0.3-0.5 | 0.4 | 0.06 | 15.06 | ||
| 24 | 10 | 0.1-0.1 | 0.1 | 0.00 | 0.00 | ||
| 2 | 01 | 10 | 1.5-1.7 | 1.6 | 0.07 | 4.30 | |
| 05 | 10 | 0.7-1.1 | 0.8 | 0.15 | 18.92 | ||
| 24 | 10 | 0.1-0.2 | 0.2 | 0.04 | 23.42 | ||
| 3 | 01 | 10 | 2.5-3.3 | 2.9 | 0.31 | 10.62 | |
| 05 | 10 | 1.8-2.0 | 1.9 | 0.06 | 3.29 | ||
| 24 | 10 | 9.1-9.5 | 9.3 | 0.14 | 1.56 | ||
| NEUT (10³/μL) | 1 | 01 | 10 | 0.34-0.43 | 0.39 | 0.03 | 7.93 |
| 05 | 10 | 0.49-0.57 | 0.52 | 0.02 | 4.62 | ||
| 24 | 10 | 0.55-0.66 | 0.59 | 0.04 | 6.40 | ||
| 2 | 01 | 10 | 11.57-12.00 | 11.79 | 0.14 | 1.18 | |
| 05 | 10 | 12.87-13.60 | 13.24 | 0.22 | 1.67 | ||
| 24 | 10 | 6.42-6.65 | 6.54 | 0.08 | 1.22 | ||
| 3 | 01 | 10 | 88.11-90.27 | 88.92 | 0.73 | 0.82 | |
| 05 | 10 | 92.04-95.03 | 94.09 | 1.00 | 1.06 | ||
| 24 | 10 | 52.36-54.67 | 53.58 | 0.65 | 1.21 |
Page 15
| Measurand | Sample | Site | N | Range | Mean | SD | CV % |
|---|---|---|---|---|---|---|---|
| NEUT (%) | 1 | 01 | 10 | 14.5-14.9 | 14.6 | 0.13 | 0.88 |
| 05 | 10 | 11.1-11.7 | 11.4 | 0.18 | 1.59 | ||
| 24 | 10 | 49.4-54.5 | 51.9 | 1.55 | 2.99 | ||
| 2 | 01 | 10 | 73.8-76.4 | 75.0 | 0.92 | 1.23 | |
| 05 | 10 | 74.2-76.8 | 75.5 | 0.89 | 1.18 | ||
| 24 | 10 | 71.3-73.9 | 72.4 | 0.77 | 1.06 | ||
| 3 | 01 | 10 | 90.9-93.9 | 92.5 | 1.04 | 1.12 | |
| 05 | 10 | 88.3-91.3 | 90.4 | 0.85 | 0.94 | ||
| 24 | 10 | 86.0-89.7 | 87.5 | 1.13 | 1.29 | ||
| LYMPH (10³/μL) | 1 | 01 | 10 | 0.52-0.61 | 0.55 | 0.03 | 5.04 |
| 05 | 10 | 0.45-0.51 | 0.49 | 0.02 | 4.19 | ||
| 24 | 10 | 0.58-0.71 | 0.64 | 0.05 | 7.13 | ||
| 2 | 01 | 10 | 1.12-1.25 | 1.18 | 0.04 | 3.35 | |
| 05 | 10 | 1.44-1.75 | 1.59 | 0.09 | 5.71 | ||
| 24 | 10 | 2.02-2.23 | 2.13 | 0.06 | 2.97 | ||
| 3 | 01 | 10 | 65.70-67.84 | 66.67 | 0.77 | 1.16 | |
| 05 | 10 | 57.42-58.32 | 57.90 | 0.30 | 0.52 | ||
| 24 | 10 | 5.35-5.74 | 5.59 | 0.12 | 2.24 | ||
| LYMPH (%) | 1 | 01 | 10 | 2.7-3.0 | 2.8 | 0.10 | 3.40 |
| 05 | 10 | 6.1-6.8 | 6.4 | 0.26 | 4.06 | ||
| 24 | 10 | 2.9-3.6 | 3.3 | 0.22 | 6.69 | ||
| 2 | 01 | 10 | 11.6-13.8 | 12.5 | 0.66 | 5.25 | |
| 05 | 10 | 38.4-40.3 | 39.4 | 0.60 | 1.53 | ||
| 24 | 10 | 26.3-29.9 | 28.0 | 1.01 | 3.61 | ||
| 3 | 01 | 10 | 80.5-92.6 | 87.1 | 3.99 | 4.58 | |
| 05 | 10 | 85.9-86.4 | 86.2 | 0.17 | 0.20 | ||
| 24 | 10 | 41.5-44.4 | 43.2 | 1.08 | 2.50 |
Page 16
| Measurand | Sample | Site | N | Range | Mean | SD | CV % |
|---|---|---|---|---|---|---|---|
| MONO (10³/μL) | 1 | 01 | 10 | 0.07-0.09 | 0.08 | 0.01 | 10.81 |
| 05 | 10 | 0.22-0.35 | 0.27 | 0.04 | 15.26 | ||
| 24 | 10 | 0.04-0.07 | 0.06 | 0.01 | 17.81 | ||
| 2 | 01 | 10 | 0.92-1.09 | 1.01 | 0.05 | 5.31 | |
| 05 | 10 | 0.67-0.75 | 0.71 | 0.03 | 3.73 | ||
| 24 | 10 | 0.61-0.82 | 0.71 | 0.07 | 10.09 | ||
| 3 | 01 | 10 | 2.80-3.27 | 3.01 | 0.15 | 4.98 | |
| 05 | 10 | 1.83-2.00 | 1.91 | 0.06 | 3.00 | ||
| 24 | 10 | 13.90-18.48 | 15.62 | 1.40 | 8.94 | ||
| MONO (%) | 1 | 01 | 10 | 2.7-3.2 | 2.9 | 0.15 | 5.24 |
| 05 | 10 | 1.4-2.3 | 1.8 | 0.29 | 16.14 | ||
| 24 | 10 | 4.2-7.6 | 5.1 | 1.00 | 19.61 | ||
| 2 | 01 | 10 | 9.5-12.3 | 11.2 | 1.20 | 10.77 | |
| 05 | 10 | 8.7-11.4 | 9.8 | 0.82 | 8.42 | ||
| 24 | 10 | 6.5-8.4 | 7.4 | 0.70 | 9.48 | ||
| 3 | 01 | 10 | 51.2-54.8 | 53.7 | 1.00 | 1.87 | |
| 05 | 10 | 14.8-19.3 | 17.4 | 1.41 | 8.09 | ||
| 24 | 10 | 25.0-30.2 | 27.3 | 1.83 | 6.71 | ||
| EO (10³/μL) | 1 | 01 | 10 | 0.08-0.12 | 0.10 | 0.01 | 12.09 |
| 05 | 10 | 0.12-0.16 | 0.14 | 0.01 | 7.92 | ||
| 24 | 10 | 0.09-0.15 | 0.12 | 0.02 | 20.41 | ||
| 2 | 01 | 10 | 0.08-0.13 | 0.10 | 0.02 | 16.47 | |
| 05 | 10 | 0.23-0.38 | 0.29 | 0.05 | 17.10 | ||
| 24 | 10 | 0.19-0.28 | 0.23 | 0.03 | 13.12 | ||
| 3 | 01 | 10 | 0.25-0.31 | 0.28 | 0.02 | 7.98 | |
| 05 | 10 | 1.96-2.15 | 2.03 | 0.05 | 2.67 | ||
| 24 | 10 | 0.86-1.27 | 1.10 | 0.16 | 14.18 | ||
| EO (%) | 1 | 01 | 10 | 1.2-2.1 | 1.5 | 0.33 | 21.58 |
| 05 | 10 | 0.5-0.6 | 0.6 | 0.05 | 9.22 | ||
| 24 | 10 | 0.1-0.1 | 0.1 | 0.00 | 0.00 | ||
| 2 | 01 | 10 | 1.8-2.9 | 2.3 | 0.39 | 17.15 | |
| 05 | 10 | 2.7-4.4 | 3.4 | 0.56 | 16.49 | ||
| 24 | 10 | 2.2-3.1 | 2.7 | 0.32 | 11.98 |
Page 17
| Measurand | Sample | Site | N | Range | Mean | SD | CV % |
|---|---|---|---|---|---|---|---|
| EO (%) | 3 | 01 | 10 | 6.1-10.6 | 8.0 | 1.61 | 20.06 |
| 05 | 10 | 5.9-7.5 | 6.8 | 0.46 | 6.85 | ||
| 24 | 10 | 3.2-5.5 | 4.4 | 0.92 | 20.99 | ||
| IG (10³/μL) | 1 | 01 | 10 | 0.03-0.08 | 0.06 | 0.01 | 22.56 |
| 05 | 10 | 0.03-0.06 | 0.04 | 0.01 | 24.43 | ||
| 24 | 10 | 0.03-0.06 | 0.05 | 0.01 | 18.89 | ||
| 2 | 01 | 10 | 0.33-0.55 | 0.48 | 0.07 | 15.67 | |
| 05 | 10 | 0.44-0.72 | 0.63 | 0.09 | 14.84 | ||
| 24 | 10 | 0.10-0.17 | 0.13 | 0.02 | 15.87 | ||
| 3 | 01 | 10 | 2.81-4.55 | 3.52 | 0.61 | 17.36 | |
| 05 | 10 | 4.79-6.21 | 5.50 | 0.37 | 6.66 | ||
| 24 | 10 | 19.32-21.21 | 20.22 | 0.61 | 3.02 | ||
| IG (%) | 1 | 01 | 10 | 0.7-1.8 | 1.4 | 0.33 | 22.95 |
| 05 | 10 | 0.5-1.0 | 0.7 | 0.16 | 24.42 | ||
| 24 | 10 | 0.4-0.7 | 0.6 | 0.08 | 15.45 | ||
| 2 | 01 | 10 | 2.2-3.9 | 3.4 | 0.56 | 16.34 | |
| 05 | 10 | 1.6-2.4 | 2.1 | 0.27 | 13.19 | ||
| 24 | 10 | 2.8-5.9 | 4.1 | 1.01 | 25.00 | ||
| 3 | 01 | 10 | 9.2-11.7 | 10.2 | 0.86 | 8.42 | |
| 05 | 10 | 9.3-12.0 | 10.7 | 0.71 | 6.62 | ||
| 24 | 10 | 23.8-25.8 | 24.7 | 0.72 | 2.90 | ||
| IPF (%) | 1 | 01 | 10 | 1.7-1.9 | 1.8 | 0.08 | 4.43 |
| 05 | 10 | 1.2-1.4 | 1.4 | 0.07 | 5.14 | ||
| 24 | 10 | 0.9-1.0 | 0.9 | 0.03 | 3.48 | ||
| 2 | 01 | 10 | 4.2-4.8 | 4.5 | 0.17 | 3.86 | |
| 05 | 10 | 6.2-6.7 | 6.4 | 0.16 | 2.57 | ||
| 24 | 10 | 3.9-4.3 | 4.1 | 0.13 | 3.15 | ||
| 3 | 01 | 10 | 18.1-18.8 | 18.5 | 0.21 | 1.14 | |
| 05 | 10 | 14.1-16.3 | 15.3 | 0.84 | 5.50 | ||
| 24 | 10 | 27.6-29.4 | 28.5 | 0.46 | 1.63 |
Page 18
| Measurand | Sample | Site | N | Range | Mean | SD | CV % |
|---|---|---|---|---|---|---|---|
| IPF (10³/μL) | 1 | 01 | 10 | 0.6-1.1 | 0.8 | 0.15 | 17.92 |
| 05 | 10 | 1.2-1.5 | 1.3 | 0.11 | 8.11 | ||
| 24 | 10 | 0.7-1.0 | 0.8 | 0.10 | 12.28 | ||
| 2 | 01 | 10 | 10.5-12.4 | 11.2 | 0.54 | 4.83 | |
| 05 | 10 | 6.2-7.2 | 6.7 | 0.32 | 4.80 | ||
| 24 | 10 | 13.1-14.4 | 13.8 | 0.45 | 3.26 | ||
| 3 | 01 | 10 | 80.5-88.3 | 83.6 | 2.42 | 2.89 | |
| 05 | 10 | 25.6-27.5 | 26.3 | 0.62 | 2.35 | ||
| 24 | 10 | 34.7-37.0 | 35.8 | 0.62 | 1.73 | ||
| RET (%) | 1 | 01 | 10 | 0.25-0.36 | 0.31 | 0.03 | 11.12 |
| 05 | 10 | 0.77-0.96 | 0.84 | 0.06 | 6.66 | ||
| 24 | 10 | 0.28-0.37 | 0.31 | 0.03 | 10.33 | ||
| 2 | 01 | 10 | 2.67-2.97 | 2.81 | 0.10 | 3.61 | |
| 05 | 10 | 2.23-2.44 | 2.32 | 0.07 | 2.95 | ||
| 24 | 10 | 2.43-2.74 | 2.60 | 0.11 | 4.15 | ||
| 3 | 01 | 10 | 5.31-5.88 | 5.59 | 0.16 | 2.84 | |
| 05 | 10 | 4.65-4.96 | 4.80 | 0.10 | 2.03 | ||
| 24 | 10 | 8.99-10.18 | 9.58 | 0.42 | 4.36 | ||
| RET (10⁶/μL) | 1 | 01 | 10 | 0.0049 - 0.0071 | 0.0061 | 0.00 | 11.36 |
| 05 | 10 | 0.0289 - 0.0340 | 0.0303 | 0.00 | 5.14 | ||
| 24 | 10 | 0.0068 - 0.0093 | 0.0077 | 0.00 | 10.79 | ||
| 2 | 01 | 10 | 0.0913 - 0.1025 | 0.0964 | 0.00 | 3.47 | |
| 05 | 10 | 0.0828 - 0.0921 | 0.0875 | 0.00 | 3.07 | ||
| 24 | 10 | 0.0774 - 0.0903 | 0.0829 | 0.00 | 4.49 | ||
| 3 | 01 | 10 | 0.2523 - 0.2744 | 0.2597 | 0.01 | 2.94 | |
| 05 | 10 | 0.1379 - 0.1561 | 0.1445 | 0.01 | 3.82 | ||
| 24 | 10 | 0.1694 - 0.1960 | 0.1826 | 0.01 | 3.94 |
Page 19
| Measurand | Sample | Site | N | Range | Mean | SD | CV % |
|---|---|---|---|---|---|---|---|
| IRF (%) | 1 | 01 | 10 | 4.4-6.4 | 5.4 | 0.56 | 10.26 |
| 05 | 10 | 2.3-3.6 | 3.0 | 0.43 | 14.23 | ||
| 24 | 10 | 3.2-5.2 | 4.5 | 0.56 | 12.57 | ||
| 2 | 01 | 10 | 23.1-25.8 | 24.3 | 0.95 | 3.90 | |
| 05 | 10 | 15.5-18.8 | 17.0 | 1.25 | 7.34 | ||
| 24 | 10 | 18.4-22.3 | 20.7 | 1.38 | 6.66 | ||
| 3 | 01 | 10 | 41.2-44.6 | 42.7 | 1.17 | 2.75 | |
| 05 | 10 | 29.9-32.4 | 31.3 | 0.91 | 2.90 | ||
| 24 | 10 | 41.8-47.0 | 43.8 | 2.00 | 4.55 | ||
| RET-He (pg) | 1 | 01 | 10 | 23.6-24.8 | 24.3 | 0.39 | 1.59 |
| 05 | 10 | 24.5-25.4 | 25.0 | 0.27 | 1.10 | ||
| 24 | 10 | 24.6-25.0 | 24.8 | 0.14 | 0.57 | ||
| 2 | 01 | 10 | 33.6-33.9 | 33.7 | 0.09 | 0.28 | |
| 05 | 10 | 33.5-34.5 | 34.1 | 0.33 | 0.97 | ||
| 24 | 10 | 33.9-35.3 | 34.7 | 0.48 | 1.39 | ||
| 3 | 01 | 10 | 40.3-40.7 | 40.4 | 0.14 | 0.35 | |
| 05 | 10 | 39.3-39.9 | 39.6 | 0.19 | 0.48 | ||
| 24 | 10 | 42.8-43.6 | 43.1 | 0.29 | 0.67 |
Precision (Reproducibility) – Whole Blood Mode
Reproducibility studies were performed to evaluate within-run, between-run, between-day, between-site and total imprecision of whole blood parameter results when measured in the whole blood mode of Sysmex XR-Series (XR-10) analyzers. The study was conducted using 3 levels (Low, Normal and High) of XN CHECK whole blood control material. Each level of control material was run in triplicate, twice a day for 5 days (3 levels x 3 replicates x 2 runs x 5 days = 90 results). Testing was conducted by a minimum of two operators at three US clinical sites in accordance with the CLSI EP05-A3 approved guideline. The results were analyzed by analysis of variance (ANOVA) method. The XR-10 results met the acceptance criteria. The table below summarizes results from XR-10.
Page 20
| Parameter | Control Level | N | Mean | Within Run | Between Run | Between Day | Between Site | Total Precision | |||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | ||||
| WBC (10³/μL) | 1 | 90 | 3.05 | 0.060 | 1.98 | 0.000 | 0.00 | 0.000 | 0.00 | 0.021 | 0.68 | 0.064 | 2.09 |
| 2 | 90 | 6.84 | 0.083 | 1.21 | 0.024 | 0.35 | 0.012 | 0.17 | 0.055 | 0.80 | 0.103 | 1.50 | |
| 3 | 90 | 16.63 | 0.161 | 0.97 | 0.040 | 0.24 | 0.000 | 0.00 | 0.128 | 0.77 | 0.209 | 1.26 | |
| RBC (10⁶/μL) | 1 | 90 | 2.32 | 0.024 | 1.03 | 0.000 | 0.00 | 0.012 | 0.51 | 0.032 | 1.36 | 0.041 | 1.78 |
| 2 | 90 | 4.23 | 0.031 | 0.73 | 0.005 | 0.12 | 0.018 | 0.42 | 0.043 | 1.02 | 0.056 | 1.33 | |
| 3 | 90 | 4.97 | 0.037 | 0.75 | 0.017 | 0.33 | 0.000 | 0.00 | 0.032 | 0.65 | 0.052 | 1.05 | |
| HGB (g/dL) | 1 | 90 | 5.2 | 0.05 | 0.98 | 0.01 | 0.22 | 0.01 | 0.20 | 0.01 | 0.17 | 0.05 | 1.04 |
| 2 | 90 | 10.9 | 0.07 | 0.68 | 0.01 | 0.10 | 0.00 | 0.00 | 0.01 | 0.13 | 0.08 | 0.70 | |
| 3 | 90 | 14.4 | 0.06 | 0.40 | 0.01 | 0.06 | 0.02 | 0.12 | 0.06 | 0.38 | 0.08 | 0.57 | |
| HCT (%) | 1 | 90 | 16.1 | 0.21 | 1.30 | 0.08 | 0.51 | 0.08 | 0.52 | 0.36 | 2.25 | 0.43 | 2.70 |
| 2 | 90 | 33.0 | 0.29 | 0.88 | 0.19 | 0.56 | 0.13 | 0.39 | 0.62 | 1.89 | 0.72 | 2.19 | |
| 3 | 90 | 42.1 | 0.37 | 0.89 | 0.25 | 0.59 | 0.00 | 0.00 | 0.61 | 1.44 | 0.76 | 1.79 | |
| MCV (fL) | 1 | 90 | 69.3 | 0.46 | 0.66 | 0.32 | 0.47 | 0.00 | 0.00 | 0.65 | 0.94 | 0.86 | 1.24 |
| 2 | 90 | 78.0 | 0.37 | 0.47 | 0.35 | 0.45 | 0.00 | 0.00 | 0.68 | 0.87 | 0.85 | 1.09 | |
| 3 | 90 | 84.8 | 0.33 | 0.39 | 0.23 | 0.27 | 0.07 | 0.08 | 0.72 | 0.85 | 0.83 | 0.97 |
Page 21
| Parameter | Control Level | N | Mean | Within Run | Between Run | Between Day | Between Site | Total Precision | |||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | ||||
| MCH (pg) | 1 | 90 | 22.6 | 0.26 | 1.15 | 0.00 | 0.00 | 0.11 | 0.49 | 0.25 | 1.09 | 0.37 | 1.66 |
| 2 | 90 | 25.9 | 0.20 | 0.76 | 0.02 | 0.07 | 0.11 | 0.41 | 0.29 | 1.14 | 0.37 | 1.43 | |
| 3 | 90 | 29.0 | 0.22 | 0.77 | 0.07 | 0.24 | 0.03 | 0.11 | 0.26 | 0.89 | 0.35 | 1.21 | |
| MCHC (g/dL) | 1 | 90 | 32.6 | 0.47 | 1.46 | 0.08 | 0.25 | 0.15 | 0.45 | 0.67 | 2.05 | 0.84 | 2.57 |
| 2 | 90 | 33.1 | 0.30 | 0.90 | 0.14 | 0.42 | 0.14 | 0.41 | 0.65 | 1.96 | 0.74 | 2.24 | |
| 3 | 90 | 34.2 | 0.32 | 0.92 | 0.16 | 0.48 | 0.00 | 0.00 | 0.60 | 1.74 | 0.69 | 2.03 | |
| PLT-I (10³/μL) | 1 | 90 | 95 | 3.5 | 3.70 | 2.4 | 2.57 | 0.0 | 0.00 | 4.3 | 4.49 | 6.0 | 6.36 |
| 2 | 90 | 248 | 5.1 | 2.06 | 1.8 | 0.73 | 2.9 | 1.16 | 5.7 | 2.29 | 8.3 | 3.37 | |
| 3 | 90 | 583 | 9.3 | 1.59 | 1.9 | 0.33 | 0.6 | 0.11 | 9.4 | 1.62 | 13.4 | 2.30 | |
| PLT-F (10³/μL) | 1 | 90 | 89 | 1.4 | 1.58 | 0.3 | 0.35 | 0.0 | 0.00 | 4.6 | 5.24 | 4.9 | 5.48 |
| 2 | 90 | 260 | 2.4 | 0.92 | 1.0 | 0.40 | 1.4 | 0.52 | 16.1 | 6.19 | 16.4 | 6.30 | |
| 3 | 90 | 584 | 4.7 | 0.80 | 0.7 | 0.12 | 1.3 | 0.23 | 32.0 | 5.47 | 32.4 | 5.54 | |
| RDW-SD (fL) | 1 | 90 | 51.9 | 0.50 | 0.86 | 0.00 | 0.00 | 0.20 | 0.41 | 0.30 | 0.65 | 0.60 | 1.15 |
| 2 | 90 | 46.3 | 0.33 | 0.70 | 0.08 | 0.17 | 0.00 | 0.00 | 0.73 | 1.58 | 0.80 | 1.74 | |
| 3 | 90 | 52.0 | 0.33 | 0.64 | 0.17 | 0.33 | 0.11 | 0.21 | 0.58 | 1.11 | 0.69 | 1.34 |
Page 22
Precision (Repeatability) – Body Fluid Mode
Body fluid repeatability studies were conducted using residual peritoneal, pleural and synovial fluid samples collected in K2EDTA anticoagulant and CSF without anticoagulant with concentrations targeting the low, and high end of the measurement range of direct measured parameters (WBC-BF, RBC-BF, TC-BF). Each sample was thoroughly mixed by gentle hand inversion and measured in replicates of ten in the body fluid mode on XR-10 analyzer at three US clinical sites. Testing was conducted in accordance with the CLSI EP05-A3 approved guideline. The mean, standard deviation (SD), and coefficient of variation (%CV) were calculated for each parameter. The %CV estimates were evaluated against acceptance criteria. The calculated coefficient of variation (%CV) of all body fluid parameters on XR-10 analyzer across all sites met manufacturer's specifications.
Precision (Reproducibility) – Body Fluid Mode
Reproducibility studies were performed to evaluate within-run, between-run, between-day, between-site and total imprecision of body fluid parameter results when measured in the body fluid mode of Sysmex XR-Series (XR-10) analyzer. The study was conducted using 2 levels (Low and High) of XN CHECK BF body fluid control material. Testing was conducted by a minimum of two operators at three US clinical sites in accordance with the CLSI EP05-A3 approved guideline. Each level of control was mixed thoroughly prior to sampling in triplicate, twice a day for 5 days (2 levels x 3 replicates x 2 runs x 5 days = 60 results per parameter) in the body fluid analysis mode on XR-10 analyzer at each site. The results were analyzed by analysis of variance (ANOVA) method. The XR-10 results met manufacturer's specifications. The table below summarizes results from XR-10.
Page 23
| Parameter | Control Level | N | Mean | Within Run | Between Run | Between Day | Between Site | Total Precision | |||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| SD | %CV | SD | %CV | SD | %CV | SD | %CV | %SD | %CV | ||||
| WBC-BF (10³/μL) | 1 | 90 | 0.073 | 0.0029 | 3.91 | 0.0009 | 1.18 | 0.0000 | 0.00 | 0.0006 | 0.82 | 0.0030 | 4.16 |
| 2 | 90 | 0.303 | 0.0061 | 2.01 | 0.0000 | 0.00 | 0.0016 | 0.54 | 0.0032 | 1.07 | 0.0071 | 2.34 | |
| RBC-BF (10⁶/μL) | 1 | 90 | 0.024 | 0.0008 | 3.49 | 0.0000 | 0.00 | 0.0002 | 0.86 | 0.0004 | 1.57 | 0.0009 | 3.93 |
| 2 | 90 | 0.073 | 0.0014 | 1.87 | 0.0000 | 0.00 | 0.0004 | 0.53 | 0.0011 | 1.51 | 0.0018 | 2.46 | |
| MN (10³/μL) | 1 | 90 | 0.021 | 0.0016 | 7.33 | 0.0003 | 1.38 | 0.0000 | 0.00 | 0.0001 | 0.49 | 0.0016 | 7.47 |
| 2 | 90 | 0.088 | 0.0032 | 3.61 | 0.0000 | 0.00 | 0.0008 | 0.90 | 0.0011 | 1.23 | 0.0035 | 3.92 | |
| MN % (%) | 1 | 90 | 29.1 | 1.74 | 5.98 | 0.34 | 1.17 | 0.00 | 0.00 | 0.28 | 0.97 | 1.80 | 6.17 |
| 2 | 90 | 29.1 | 0.90 | 3.08 | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 | 0.90 | 3.08 | |
| PMN (10³/μL) | 1 | 90 | 0.052 | 0.0023 | 4.50 | 0.0007 | 1.41 | 0.0000 | 0.00 | 0.0006 | 1.16 | 0.0025 | 4.86 |
| 2 | 90 | 0.215 | 0.0052 | 2.40 | 0.0000 | 0.00 | 0.0000 | 0.00 | 0.0021 | 0.97 | 0.0056 | 2.59 | |
| PMN (%) | 1 | 90 | 70.9 | 1.74 | 2.45 | 0.34 | 0.48 | 0.00 | 0.00 | 0.28 | 0.40 | 1.80 | 2.53 |
| 2 | 90 | 70.9 | 0.90 | 1.26 | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 | 0.90 | 1.26 | |
| TC-BF (10³/μL) | 1 | 90 | 0.073 | 0.0029 | 3.91 | 0.0009 | 1.18 | 0.0000 | 0.00 | 0.0006 | 0.82 | 0.0030 | 4.16 |
| 2 | 90 | 0.303 | 0.0061 | 2.01 | 0.0000 | 0.00 | 0.0016 | 0.54 | 0.0032 | 1.07 | 0.0071 | 2.34 |
Page 24
Linearity:
Whole Blood
Linearity studies were performed using WRP CHECK EX whole blood linearity material and system diluent (CELLPACK DCL). A minimum of seven sample dilutions were created with concentrations which span the full measurement range (1-level below, 5-levels within and 1-level above) of WBC, RBC, HGB, HCT, PLT-I and PLT-F directly measured parameters. All samples were mixed thoroughly prior to testing in replicates of three in the whole blood mode of three XR-10 analyzers at one internal site. Testing and analysis of the data were conducted in accordance with CLSI EP06-ED2:2020. The results of whole blood linearity for measured WBC, RBC, HGB, HCT, PLT-I and PLT-F parameters by XR-10 analyzers demonstrate the method to be linear from the lower limit to upper limit and within the measured maximum allowable deviation from linearity for each interval. All results met predefined acceptance criteria.
| Parameter | Linear Range |
|---|---|
| WBC (x10³/μL) | 0.03 – 440.00 |
| RBC (x10⁶/μL) | 0.01 – 8.60 |
| HGB (g/dL) | 0.1 – 26.0 |
| HCT (%) | 0.1 – 75.0 |
| PLT(x10³/μL) | 2 – 5,000 |
| RET (%) | 0.00 – 30.00 |
Body Fluid
Linearity studies were performed using a minimum of seven sample dilutions created with concentrations which spanned the full measurement range (1-level below, 5-levels within and 1-level above) of WBC-BF, RBC-BF and TC-BF direct measured parameters. All samples were mixed thoroughly prior to testing in replicates of three in the body fluid mode of three XR-10 analyzers at one internal site. Testing and analysis of the data were conducted in accordance with CLSI EP06-ED2:2020. The results of body fluid linearity for measured WBC-BF, RBC-BF and TC-BF parameters by XR-10 demonstrate the method to be linear from the lower limit to upper limit and is within the measured maximum allowable deviation from linearity for each interval. All results met predefined acceptance criteria.
| Parameter | Linear Range |
|---|---|
| WBC-BF (x10³/μL) | 0.003 – 10.000 |
| RBC-BF (x10⁶/μL) | 0.002 – 5.000 |
| TC-BF (x10³/μL) | 0.003 – 10.000 |
Page 25
Analytical Specificity/Interferences:
Interfering substances studies were conducted for Bilirubin F, Bilirubin C, Chyle, Hemolytic Hemoglobin, Lipids, and high WBC, RBC, and platelet counts and microcytic RBCs to determine the concentration that impact all claimed parameters on the Sysmex XR-Series Automated Hematology analyzers. Whole blood K2EDTA samples were collected from donors for this study and varying concentrations of interferent was added to obtain concentrations. The tubes were mixed, and measurements were repeated in four consecutive batches on the XR-Series automated hematology analyzer. The following table includes the results of the study:
| Interferent | Conclusion |
|---|---|
| Bilirubin F | There was no significant Bilirubin F interference up to a concentration of 40 mg/dL for all parameters. |
| Bilirubin C | There was no significant Bilirubin C interference up to a concentration of 40 mg/dL for all parameters. |
| Chyle | There was no significant Chyle interference up to a concentration of 3,600 FTU for all parameters except for the MCHC parameter. For the MCHC parameter, there was no significant interference up to a concentration of 2,880 FTU. |
| Hemolytic Hemoglobin | There was no significant Hemolysis interference up to a concentration of 1,000 mg/dL for all parameters except for MCH, MCHC. For the MCH parameter, there was no significant interference observed up to a concentration of 800mg/dL. For the MCHC parameter, no significant interference was observed up to a concentration of 400mg/dL. |
| Lipids | There was no significant Lipemia (Intralipos) interference up to a concentration of 2.00 g/dL for all parameters except for HGB, MCH, and MCHC parameters. For HGB and MCH parameters, there was no significant interference up to a concentration of 0.20 g/dL. For the MCHC parameter, there was no significant interference up to a concentration of 0.10 g/dL. |
| High white blood cell counts | There was no significant WBC interference up to a concentration of 360.63 x 10³ cells/μL for all parameters. |
| High red blood cell counts | There was no significant RBC interference observed up to a concentration of 8.04 x 10⁶ cells/µL for all parameters. |
| High platelet counts | There was no significant PLT interference observed up to a concentration of 1513 (PLT-I) or 1563 (PLT-F) x 10³ cells/µL for all parameters. |
| Microcytic RBCs | There was no significant interference from microcytic RBCs observed for RBC, HCT or PLT (I/F) parameters with MCV values ≤70 fL. |
Page 26
Sample Stability:
Whole Blood Stability
The evaluation of whole blood stability was conducted at one internal site using normal and abnormal (8 unique leftover samples and 12 prospectively collected) de-identified K2EDTA venous whole blood samples. Samples were split into two sets stored at room temperature (18-26°C) and refrigerated temperature (2-8°C). Samples were tested in duplicate at baseline (T0), 6, 24, 25, 48, 49, 72 and 73 hours at both conditions. The mean, standard deviation, mean difference and percent difference from the baseline mean of each sample result were calculated for each parameter at each time interval for both conditions. The data supports a whole blood sample stability of 24 hours at room temperature (18-26°C) and 48 hours at refrigerated temperature (2-8°C) for XR-Series (XR-10) claimed whole blood parameters.
Body Fluid Short-term Stability
The evaluation of body fluid stability was conducted at 1 external site using 12 unique de-identified leftover body fluid samples (3-CSF, 3-peritioneal, 3-pleural and 3-synovial) when stored at controlled room temperature (RT). Peritoneal, pleural and synovial body fluid samples collected in K2EDTA anticoagulant and CSF without anticoagulant. Samples were thoroughly mixed by gentle hand inversion at least ten times, before analyzing in singlet on XR-10 analyzer. Samples were tested at baseline or zero (0) time, 1, 2, and 4 hours at RT (18-26°C) in the body fluid analysis mode.
Body fluid samples should be analyzed within 1 hour of collection on the XR-Series (XR-10) analyzer as demonstrated in the short term stability study.
Detection Limit:
Whole Blood Mode
Limits of Blank (LoB), Limit of Detection (LoD), and the Limit of Quantitation (LoQ) were determined for the direct measured WBC, RBC, HGB, HCT and PLT parameters on Sysmex XR-10 Automated Hematology Analyzers.
In LoB testing, four blank samples were measured in replicates of five, over a period of three days using two reagent lots, to yield 120 total measurement results per parameter. To determine the LoD and LoQ, four low concentration samples were analyzed on the Sysmex XN-20 automated hematology analyzer (K112605) to assign the reference value. The low-level samples were then measured in replicates of five over a period of three days using two reagent lots, to yield 120 total measurement results per parameter across 2-XR-10 (4 samples x 5 replicates x 3 days x 1 reagent lot per analyzer = 60 results per analyzer).
The results of the LoB, LoD, and LoQ are provided in the table below.
Page 27
| Parameter (Unit) | Limit of Blank (LoB) | Limit of Detection (LoD) | Limit of Quantitation (LoQ) |
|---|---|---|---|
| WBC (x 10³/µL) | 0.00 | 0.01 | 0.02 |
| RBC (x 10⁶/µL) | 0.00 | 0.01 | 0.01 |
| HGB (g/dL) | 0.0 | 0.1 | 0.1 |
| HCT (%) | 0.0 | 0.1 | 0.1 |
| PLT-I (x 10³/µL) | 0 | 1 | 2 |
| PLT-F (x 10³/µL) | 0 | 1 | 2 |
Body Fluid Mode
Limits of Blank (LoB), Limit of Detection (LoD), and the Limit of Quantitation (LoQ) were determined for the direct measured WBC-BF, RBC-BF and TC-BF parameters on Sysmex XR-10 Automated Hematology Analyzers.
In LoB testing, four blank samples were measured in replicates of five, over a period of three days using two reagent lots, to yield 120 total measurement results per parameter. To determine the LoD and LoQ, four low concentration samples were analyzed on the Sysmex XN-20 automated hematology analyzer (K112605) to assign the reference value. The low-level samples were then measured in replicates of five over a period of three days using two reagent lots, to yield 120 total measurement results per parameter across 2-XR-10 (4 samples x 5 replicates x 3 days x 1 reagent lot per analyzer = 60 results per analyzer).
The results of the LoB, LoD, and LoQ are provided in the table below.
| Parameter | LoB (N=120) | LoD (N=120) | LoQ (N=120) |
|---|---|---|---|
| WBC-BF (10³/μL) | 0.001 | 0.002 | 0.002 |
| RBC-BF (10⁶/μL) | 0.000 | 0.002 | 0.002 |
| TC-BF (10³/μL) | 0.001 | 0.002 | 0.002 |
Page 28
Carry-Over:
Whole Blood
Three sets of carryover sequences were run on Sysmex XR-10 analyzers for each applicable parameter at three US clinical sites using de-identified leftover venous whole blood samples collected in K2EDTA anticoagulant. For each parameter, high target concentration samples were run in replicates of three (H1, H2, H3) followed by three replicates of low target concentration samples (L1, L2, L3) in XR-10 whole blood mode. The study was conducted in accordance with CLSI H26-A2. The results of the whole blood carryover on XR-10 analyzers show all applicable parameters met the manufacturer's specifications.
Body Fluid
Carryover was conducted using de-identified peritoneal, pleural and synovial fluids collected in K2EDTA and CSF samples without anticoagulant with high target and low target WBC-BF, RBC-BF, and TC-BF. Three sets of carryover sequences were run on Sysmex XR-10 analyzers for each applicable parameter at three US clinical sites using de-identified leftover peritoneal, pleural and synovial fluid samples collected in K2EDTA anticoagulant and CSF without anticoagulant. For each parameter, high target concentration samples were run in replicates of three (H1, H2, H3) followed by three replicates of low target concentration samples (L1, L2, L3) in XR-Series body fluid mode. The study was conducted in accordance with CLSI H26-A2. The results of body fluid carryover on XR-10 analyzers show all applicable parameters met the manufacturer's specifications.
Comparison Studies:
Whole Blood - Method Comparison with Predicate Device
A method comparison study was conducted to assess the performance of the Sysmex XR-Series (XR-10) Automated Hematology analyzers compared to the predicate device, Sysmex XN-20 (K112605). A total of 865 unique residual whole blood samples in K2EDTA anticoagulant from pediatrics (<21 years) and adult (≥21 years) subjects including a variety of disease states (e.g., pathological WBCs, lipemia, hyperbilirubinemia, hemoglobinopathies, thrombocytopenia, thrombocytosis, etc.) were tested across 3 US clinical sites. Twenty samples were excluded due to insufficient sample volume and operator error.
Sample demographics from all sites included 310 pediatric subjects and 551 adults and 4 subjects with age not reported. Of this total, 53.7% were male, 45.8% female and 0.5% with sex not reported. A total of 362 normal (no flags, marked as negative) whole blood samples and 500 abnormal whole blood samples (contained flags, marked as positive) were tested.
Page 29
The results of the linear regression and bias analyses between the XR-10 Whole Blood Mode and XN-20 met the acceptance criteria for all applicable parameters for correlation coefficient (r) and %Bias with the exception of the calculated bias for HGB on XR-10 at one site. The bias (-2.10%, -0.3 g/dL) was slightly above the predefined bias limits (±2% or 0.2g/dL) but is well within the predefined correlation and coefficient value limits of ≥0.95 (0.9932) and is determined acceptable. An example of the results between XN-20 and XR-10 is shown below.
Correlation and Estimated Bias (Whole Blood – Combined Sites) – XN-20 vs XR-10
| Measurand | N | Result Range | r | Slope (95% CI) | Intercept (95% CI) | Mean Diff. | Mean %Diff. |
|---|---|---|---|---|---|---|---|
| WBC (10³/μL) | 841 | 0.03 - 410.88 | 0.9997 | 1.002 (0.993, 1.010) | 0.066 (-0.031, 0.163) | 0.09 | 0.52 |
| RBC (10⁶/μL) | 845 | 0.03 - 8.50 | 0.9900 | 1.000 (0.991, 1.009) | -0.035 (-0.075, 0.005) | -0.03 | -0.83 |
| HGB (g/dL) | 843 | 0.2 - 25.6 | 0.9915 | 0.994 (0.986, 1.002) | -0.100 (-0.213, 0.013) | -0.2 | -1.41 |
| HCT (%) | 842 | 0.3 – 75.0 | 0.9901 | 0.999 (0.989, 1.009) | -0.338 (-0.743, 0.067) | -0.4 | -0.98 |
| MCV (fL) | 845 | 62.7 - 141.5 | 0.9922 | 0.980 (0.971, 0.989) | 1.711 (0.881, 2.540) | -0.2 | -0.19 |
| MCH (pg) | 844 | 17.2 - 46.6 | 0.9778 | 0.995 (0.973, 1.017) | -0.006 (-0.650, 0.637) | -0.2 | -0.54 |
| MCHC (g/dL) | 844 | 22.2 - 41.6 | 0.9286 | 0.961 (0.921, 1.002) | 1.104 (-0.181, 2.390) | -0.1 | -0.38 |
| PLT-I (10³/μL) | 842 | 3 - 4930 | 0.9991 | 0.989 (0.977, 1.001) | 0.840 (-2.580, 4.260) | -3.0 | -0.81 |
| PLT-F (10³/μL) | 824 | 4 - 4748 | 0.9987 | 1.017 (0.999, 1.035) | 0.393 (-4.754, 5.540) | 6.0 | 1.81 |
| RDW-SD (fL) | 840 | 29.8 - 121.5 | 0.9959 | 0.965 (0.953, 0.977) | 1.891 (1.306, 2.476) | 0.1 | 0.12 |
Page 30
| Measurand | N | Result Range | r | Slope (95% CI) | Intercept (95% CI) | Mean Diff. | Mean %Diff. |
|---|---|---|---|---|---|---|---|
| RDW-CV (%) | 841 | 10.4 - 34.8 | 0.9947 | 0.985 (0.974, 0.995) | 0.229 (0.078, 0.380) | -0.0 | -0.06 |
| MPV (fL) | 780 | 8.1 - 14.7 | 0.9338 | 0.972 (0.944, 0.999) | 0.085 (-0.204, 0.374) | -0.2 | -2.04 |
| NRBC (10³/μL) | 843 | 0 - 11.95 | 0.9994 | 1.001 (0.988, 1.014) | -0.000 (-0.002, 0.002) | 0.00 | 0.00 |
| NRBC (%) | 843 | 0 - 114.3 | 0.9991 | 0.969 (0.937, 1.000) | 0.026 (0.006, 0.047) | -0.0 | -0.47 |
| NEUT (10³/μL) | 834 | 0 - 300.11 | 0.9957 | 0.999 (0.960, 1.038) | 0.063 (-0.206, 0.332) | 0.05 | 0.52 |
| LYMPH (10³/μL) | 834 | 0 - 389.35 | 0.9985 | 1.026 (0.988, 1.065) | 0.120 (-0.017, 0.256) | 0.24 | 5.19 |
| MONO (10³/μL) | 834 | 0 - 71.69 | 0.9692 | 0.939 (0.823, 1.056) | -0.001 (-0.128, 0.126) | -0.11 | -6.15 |
| EO (10³/μL) | 833 | 0.00 - 8.11 | 0.9760 | 1.036 (0.938, 1.134) | -0.004 (-0.021, 0.014) | 0.01 | 2.09 |
| BASO (10³/μL) | 834 | 0 - 30.10 | 0.9832 | 0.836 (0.670, 1.001) | 0.009 (-0.013, 0.031) | -0.04 | -15.07 |
| NEUT (%) | 834 | 0 - 96.3 | 0.9925 | 1.001 (0.992, 1.010) | -0.077 (-0.676, 0.521) | -0.0 | -0.03 |
| LYMPH (%) | 834 | 0 - 96.7 | 0.9938 | 1.013 (1.003, 1.023) | 0.188 (-0.057, 0.433) | 0.6 | 2.01 |
| MONO (%) | 834 | 0 - 79.6 | 0.9658 | 0.940 (0.898, 0.981) | 0.148 (-0.201, 0.496) | -0.4 | -4.50 |
| EO (%) | 833 | 0.0 - 47.5 | 0.9897 | 1.001 (0.970, 1.031) | 0.028 (-0.030, 0.085) | 0.0 | 1.33 |
| BASO (%) | 834 | 0 - 8.4 | 0.9417 | 0.865 (0.796, 0.935) | 0.014 (-0.027, 0.055) | -0.1 | -11.55 |
| IG (10³/μL) | 834 | 0 - 155.97 | 0.9980 | 0.995 (0.967, 1.023) | -0.000 (-0.033, 0.033) | -0.01 | -0.42 |
| IG (%) | 834 | 0 - 44.8 | 0.9599 | 0.977 (0.936, 1.018) | 0.016 (-0.057, 0.090) | -0.0 | -0.26 |
Page 31
| Measurand | N | Result Range | r | Slope (95% CI) | Intercept (95% CI) | Mean Diff. | Mean %Diff. |
|---|---|---|---|---|---|---|---|
| RET (%) | 824 | 0.12 - 23.10 | 0.9839 | 0.997 (0.959, 1.035) | 0.031 (-0.036, 0.098) | 0.02 | 1.06 |
| RET (10⁶/μL) | 793 | 0.01 - 0.5460 | 0.9727 | 0.978 (0.913, 1.042) | 0.002 (-0.003, 0.007) | 0.0004 | 0.40 |
| IRF (%) | 824 | 0 - 56.7 | 0.9729 | 0.976 (0.955, 0.997) | -1.660 (-1.927, -1.392) | -2.1 | -12.62 |
| RET-He (pg) | 822 | 17 - 51.3 | 0.9689 | 0.923 (0.902, 0.944) | 1.294 (0.619, 1.969) | -1.2 | -3.68 |
| IPF (%) | 824 | 0.4 - 37.9 | 0.9891 | 0.983 (0.958, 1.009) | -0.072 (-0.175, 0.031) | -0.2 | -3.12 |
| IPF (10³/μL) | 819 | 0.2 - 256.4 | 0.9876 | 1.036 (0.950, 1.123) | -0.666 (-1.632, 0.300) | -0.2 | -1.46 |
Body Fluid - Method Comparison with Predicate Device
A method comparison study was conducted to assess the performance of Sysmex XR-10 Body Fluid Mode when compared to the predicate device, Sysmex XN-20 (K112605). A total of 397 residual body fluid samples at three US sites. Two samples were excluded due to insufficient sample volume. All body fluids (peritoneal, pleural, and synovial) were collected in K2EDTA anticoagulant with exception of CSF. All Samples were run in the Body Fluid Mode in singlet on the XN-20 and within two hours on the XR- 10 analyzer. Samples covered clinical medical decision levels, and to the extent possible, of the full measuring ranges of the XR-Series analyzers. The results of the linear regression and bias analyses between XR-10, and XN-20 Body Fluid Mode for all body fluids (CSF, peritoneal, pleural and synovial) are shown below.
Page 32
| Fluid Type | Measurand | N | Result Range | R | Slope (95% CI) | Intercept (95% CI) | Mean Diff. | % Mean Diff. Or SD |
|---|---|---|---|---|---|---|---|---|
| CSF | WBC-BF (10³/μL) | 76 | 0.003 - 9.178 | 0.9968 | 0.996(0.966, 1.027) | 0.001(-0.025, 0.028) | -0.008 | -0.31 |
| RBC-BF (10⁶/μL) | 61 | 0.002 - 4.875 | 0.9970 | 0.997(0.986, 1.008) | 0.023(-0.020, 0.066) | 0.019 | 1.29 | |
| MN (10³/μL) | 70 | 0.003 - 4.074 | 0.8750 | 1.734(0.729, 2.738) | -0.459(-1.198, 0.280) | 0.244 | 1 SD | |
| MN (%) | 90 | 0.0 - 100.0 | 0.7984 | 1.013(0.880, 1.145) | -2.053(-6.793, 2.688) | -1.5 | 1 SD | |
| PMN (10³/μL) | 61 | 0.003 - 8.393 | 0.9862 | 0.952(0.884, 1.021) | -0.044(-0.122, 0.034) | -0.147 | 1 SD | |
| PMN% (%) | 90 | 0.0 - 100.0 | 0.7984 | 1.013(0.880, 1.145) | 0.792(-9.206, 10.791) | 1.5 | 1 SD | |
| TC-BF (10³/μL) | 78 | 0.003 - 9.192 | 0.9968 | 0.997(0.967, 1.027) | 0.001(-0.024, 0.027) | -0.006 | -0.24 | |
| Peritoneal | WBC-BF (10³/μL) | 107 | 0.004 - 9.766 | 0.9989 | 0.996(0.982, 1.010) | 0.012(-0.003, 0.027) | 0.006 | 0.31 |
| RBC-BF (10⁶/μL) | 70 | 0.002 - 4.801 | 0.9996 | 0.999(0.980, 1.017) | 0.004(-0.003, 0.010) | 0.003 | 0.31 | |
| MN (10³/μL) | 108 | 0.003 - 5.689 | 0.9571 | 1.120(0.930, 1.310) | -0.019(-0.089, 0.052) | 0.048 | 1 SD | |
| MN (%) | 109 | 0.6 - 100.0 | 0.9543 | 0.981(0.940, 1.021) | 1.411(-0.736, 3.559) | 0.3 | 1 SD | |
| PMN (10³/μL) | 97 | 0.003 - 8.469 | 0.9936 | 0.955(0.902, 1.008) | 0.017(-0.015, 0.050) | -0.046 | 1 SD | |
| PMN% (%) | 109 | 0.0 - 99.4 | 0.9543 | 0.981(0.940, 1.021) | 0.512(-2.413, 3.437) | -0.3 | 1 SD | |
| TC-BF (10³/μL) | 107 | 0.004 - 9.774 | 0.9988 | 0.997(0.982, 1.011) | 0.011(-0.005, 0.026) | 0.005 | 0.26 |
Page 33
| Fluid Type | Measurand | N | Result Range | r | Slope (95% CI) | Intercept (95% CI) | Mean Diff. | % Mean Diff. Or SD |
|---|---|---|---|---|---|---|---|---|
| Pleural | WBC-BF (10³/μL) | 95 | 0.005 - 9.996 | 0.9982 | 1.017(0.996, 1.038) | -0.006(-0.033, 0.021) | 0.032 | 1.42 |
| RBC-BF (10⁶/μL) | 72 | 0.002 - 4.691 | 0.9998 | 0.994(0.986, 1.003) | 0.001(-0.002, 0.004) | -0.003 | -0.41 | |
| MN (10³/μL) | 97 | 0.004 - 5.240 | 0.9786 | 1.225(1.099, 1.350) | -0.127(-0.212, -0.041) | 0.112 | 1 SD | |
| MN (%) | 99 | 9.2 - 100.0 | 0.9154 | 1.003(0.962, 1.044) | 0.317(-3.802, 4.437) | 0.5 | 1 SD | |
| PMN (10³/μL) | 92 | 0.006 - 8.862 | 0.9882 | 0.937(0.864, 1.011) | 0.007(-0.047, 0.060) | -0.083 | 1 SD | |
| PMN% (%) | 99 | 0.0 - 90.8 | 0.9154 | 1.003(0.962, 1.044) | -0.637(-2.494, 1.221) | -0.5 | 1 SD | |
| TC-BF (10³/μL) | 94 | 0.006 - 9.437 | 0.9978 | 1.013(0.989, 1.036) | -0.005(-0.035, 0.024) | 0.024 | 1.04 | |
| Synovial | WBC-BF (10³/μL) | 55 | 0.004 - 9.780 | 0.9980 | 1.003(0.978, 1.027) | 0.032(-0.008, 0.072) | 0.041 | 1.21 |
| RBC-BF (10⁶/μL) | 56 | 0.002 - 4.801 | 0.9998 | 0.987(0.975, 0.999) | 0.003(0.000, 0.006) | -0.007 | -0.90 | |
| MN (10³/μL) | 61 | 0.004 - 4.803 | 0.9342 | 1.248(0.943, 1.554) | -0.156(-0.345, 0.033) | 0.094 | 1 SD | |
| MN (%) | 65 | 0.0 - 93.1 | 0.9460 | 0.995(0.900, 1.090) | 0.306(-1.595, 2.207) | 0.1 | 1 SD | |
| PMN (10³/μL) | 56 | 0.003 - 8.694 | 0.9836 | 0.991(0.960, 1.023) | -0.033(-0.144, 0.077) | -0.057 | 1 SD | |
| PMN% (%) | 65 | 6.9 - 100.0 | 0.9460 | 0.995(0.900, 1.090) | 0.202(-7.956, 8.361) | -0.1 | 1 SD | |
| TC-BF (10³/μL) | 55 | 0.004 - 9.783 | 0.9979 | 1.003(0.978, 1.028) | 0.032(-0.010, 0.074) | 0.042 | 1.22 |
Page 34
Matrix Studies:
Whole Blood Anticoagulant Comparison (K2EDTA vs. K3EDTA)
Comparability between K2EDTA vs. K3EDTA anticoagulated whole blood samples on the Sysmex XR-Series (XR-10) analyzer were conducted at 1 internal site using 46 paired K2 and K3EDTA venous whole blood samples. Testing was run in singlet between K2 and K3 EDTA anticoagulant tubes from a single donor and completed within a target of 2 hours. The study was conducted in accordance with CLSI EP09-A3 approved guidelines. The results of the regression analysis and bias analyses between K2EDTA and K3EDTA venous whole blood samples met predefined correlation and coefficient and/or bias limits for all applicable parameters and demonstrate equivalency between the two anticoagulants
Venous Whole Blood vs. Capillary Whole Blood (K2EDTA)
Comparability between venous whole blood and capillary whole blood samples on the Sysmex XR-Series (XR-10) analyzer was conducted using residual and prospectively collected K2EDTA (4mL tubes) venous whole blood samples with concentrations representative of patient samples, across medical decision levels, and to the extent possible of the full analytical measurement range of direct measured parameters (WBC, RBC, HGB, HCT and PLT). The comparison study was conducted at one internal site using seventy paired venous whole blood samples. Within 2 hours of analysis of samples in the normal tube position, samples were mixed and transferred to micro-collection tubes (without anticoagulant). Then the samples were placed in the micro collection tube holder position and run in singlet in the manual mode using the same discrete test selections used for normal tube samples on XR-10 analyzer. The study was conducted in accordance with CLSI EP09-A3 approved guidelines. The K2EDTA sample results from the 4mL tubes were compared to the corresponding results of the micro collection tube. The results of the regression analysis and bias analyses between K2EDTA 4mL tube samples and micro collection tube samples met predefined correlation and coefficient and/or bias limits for all applicable parameters and demonstrate equivalency between the two collection tubes.
Whole Blood K2EDTA Normal Tubes vs. Micro-collection Tube
The study was conducted using residual and prospectively collected K2EDTA (4mL tubes) venous whole blood samples with concentrations representative of patient samples, across medical decision levels, and to the extent possible of the full analytical measurement range of direct measured parameters (WBC, RBC, HGB, HCT and PLT) of the Sysmex XR-10 analyzer. The comparison study was conducted at one internal site using seventy paired venous whole blood samples. K2EDTA venous whole blood samples were thoroughly mixed, then placed in the normal tube holder position and run in singlet in the manual mode. Within 2 hours of analysis of samples in the normal tube position, samples were mixed and transferred to micro-collection tubes (without anticoagulant). Then the samples were placed in the micro collection tube holder position and run in singlet in the manual mode. The study was conducted in accordance with CLSI EP09-A3 approved guidelines. The results of the regression analysis and bias analyses between K2EDTA 4mL tube samples and micro collection tube samples met predefined correlation and coefficient and/or bias limits for all applicable parameters and demonstrate equivalency between the two collection tubes.
Page 35
Bridging Studies:
Whole Blood Mode to Pre-dilute Mode Comparison
Comparability between whole blood mode to pre-dilute mode was conducted at 1 internal site using 45 de-identified residual whole blood samples and system diluent to create 1:7 dilution samples. Each whole blood sample was run in singlet in the whole blood mode of the Sysmex XR-10 analyzer. Within 2 hours of analysis in the whole blood mode, a 1:7 dilution was prepared for each sample by pipetting 600 µL of system diluent (CELLPACK DCL) into 4mL plain top tubes using calibrated displacement pipettes and adding 100 µL of whole blood using a new tip to create a 1:7 dilution. Each dilution sample was mixed by gentle hand inversion and run in singlet in the pre-dilution mode. Results in the pre-dilution mode are automatically multiplied by 7 before results are displayed, no additional calculation is required. The study was conducted in accordance with CLSI EP09-A3 approved guidelines. The results from the 4mL whole blood mode sample tubes were compared to the corresponding results of the 4 mL dilution sample tubes for the same patient sample. The results of the comparison data met predefined correlation and coefficient and/or bias limits for all applicable parameters and demonstrate equivalency between the whole blood and pre-dilution modes.
Predilute Mode Normal Tube to Micro-collection Tube Comparison
The comparison study was conducted at one internal site using 40 de-identified residual whole blood samples and system diluent to create dilution samples. A 1:7 dilution was prepared for each whole blood sample by pipetting 600 µL of system diluent (CELLPACK DCL) into 4mL plain top tubes using calibrated displacement pipettes. Using a new pipette tip, 100 µL of whole blood was added to 4 mL tubes to create a 1:7 dilution for each whole blood sample. Each sample dilution was mixed by gentle hand inversion a minimum of 10 times then placed in the Normal tube holder position and analyzed in singlet in the Pre-dilution mode of the Sysmex XR-10 analyzer. Within 2 hours of analysis, the 4 mL tube dilution samples were remixed then transferred to micro collection tubes without additives, then placed in the Micro collection tube holder position and analyzed in singlet in the Pre-dilution mode. Results in the pre-dilution mode are automatically multiplied by 7 before results are displayed, no additional calculation is required. The study was conducted in accordance with CLSI EP09-A3 approved guidelines. The results from the 4mL dilution sample tubes were compared to the corresponding results of the micro collection dilution sample tubes for the same patient sample. The results of the comparison data met predefined correlation and coefficient and/or bias limits for all applicable parameters and demonstrate equivalency between the Normal and Micro collection tube holder positions on XR-Series (XR-10) analyzer.
Low WBC Mode Normal Tube to Micro collection Tube Comparison
Comparability between Low WBC Mode Normal Tube to Micro collection was conducted at one internal site using 43 residual de-identified venous whole blood samples with WBC concentrations less than 4.50 x 10³/µL. Each whole blood sample was mixed by gentle hand inversion a minimum of 10 times then placed in the Normal tube holder position and run in singlet in the Low WBC whole blood mode of the XR-10 analyzer. Within 2 hours of analysis in the Low WBC whole blood mode, the whole blood samples were remixed then transferred to micro collection tubes without anticoagulant then placed in the Micro collection tube holder position and run in singlet in the Low WBC whole blood mode. The study was conducted in accordance with CLSI EP09-A3 approved guidelines. The results from the Low WBC whole blood Normal tube holder sample position were compared to the corresponding results of the Micro collection tube holder sample position for the same patient sample. The results of the comparison
Page 36
data met predefined correlation and coefficient and/or bias limits for all applicable parameters and demonstrate equivalency between Low WBC mode Normal tube and Micro collection tube holder positions on the XR-Series (XR-10) analyzer.
Clinical Studies:
Clinical Sensitivity and Specificity
Sensitivity/specificity studies were conducted to evaluate the flagging capabilities of the Sysmex XR-Series (XR-10) Automated Hematology analyzer using patient samples representing a variety of abnormal conditions in comparison to manual differential counts and peripheral blood smear review by experienced examiners using light microscopy (reference method) at each of the three external clinical sites from the method comparison study. Three blood film slides were prepared for each sample for manual measurement. The flagging results from Sysmex XR-10 Automated Hematology Analyzer for normal (no flags) and abnormal (flags present) were compared to manual differential counts and peripheral blood smear review using light microscopy. The distributional and morphological flagging results were also compared to those of the XN-20 (predicate device). The sensitivity, positive percent agreement (PPA), specificity, negative percent agreement (NPA) and overall percent agreement (OPA) with 95% confidence intervals (CI) were calculated for all sites combined and are presented in the following tables.
Abnormal Flagging – Manual Microscopy (Reference Method)
| Category | N | True Positive (TP) | False Positive (FP) | True Negative (TN) | False Negative (FN) | Sensitivity (95% CI) | Specificity (95% CI) | OPA (95% CI) |
|---|---|---|---|---|---|---|---|---|
| Any Distributional Abnormalities | 705 | 325 | 55 | 213 | 112 | 74.37 (70.01 – 78.40) | 79.48 (74.14 – 84.15) | 76.31 (73.00 – 79.41) |
| Any Morphological Flag | 780 | 199 | 188 | 353 | 40 | 83.26 (77.91 - 87.77) | 65.25 (61.07 - 69.26) | 70.77 (67.44 - 73.94) |
| Any Distributional and/or Morphological Abnormalities | 780 | 417 | 102 | 171 | 90 | 82.25 (78.62 – 85.48) | 62.64 (56.60 – 68.39) | 75.38 (72.20 – 78.37) |
Abnormal Flagging – XN-20 (Predicate Method)
| Category | N | True Positive (TP) | False Positive (FP) | True Negative (TN) | False Negative (FN) | PPA (95% CI) | NPA (95% CI) | OPA (95% CI) |
|---|---|---|---|---|---|---|---|---|
| Any Distributional Abnormalities | 834 | 468 | 14 | 326 | 26 | 94.74 (92.38 – 96.53) | 95.88 (93.19 – 97.73) | 95.20 (93.53 – 96.55) |
| Any Morphological Flag | 844 | 383 | 60 | 369 | 32 | 92.29 (89.29 – 94.67) | 86.01 (82.37 – 89.15) | 89.10 (86.80 – 91.12) |
| Any Distributional and/or Morphological Abnormalities | 845 | 557 | 32 | 235 | 21 | 96.37 (94.50 – 97.74) | 88.01 (83.50 – 91.66) | 93.73 (91.88 – 95.27) |
Page 37
Expected Values/Reference Range:
Whole Blood - Verification of Adult Reference Intervals
Verification of adult reference intervals was conducted on the Sysmex XR-Series (XR-10) Automated Hematology analyzer to demonstrate comparability of whole blood reference intervals for an adult population (>21 years) to ranges established for a predicate device Sysmex XE-5000 (K071967). One hundred and thirty-two samples (58 males and 74 females) were tested and compared to pre-established reference intervals to determine if the ranges were applicable for use with Sysmex XR-10 Automated Hematology analyzers. The results of the proposed reference intervals overlapped the 95% confidence intervals (lower and upper limit) of the adult male and female and were determined to be acceptable.
Whole Blood - Verification of Pediatric Reference Intervals
Using Pediatric Reference Interval literature source (Wong, E., Brugnara, C., Straseski, J., Kellogg, M., & Adeli, K. 2021. Pediatric Reference Intervals. 8th ed., Hematology Tests (pp. 209-267), Academic Press.), reference interval verification study was performed for the pediatric population. A total of 196 pediatric samples including each subpopulation: 40 neonates (birth–28 days); 55 infants (>28 days–2 years); 60 children (>2 years–12 years); and 41 adolescents (>12 years–21 years) were used in the study. The results of the proposed reference intervals overlapped the 95% confidence intervals (lower and upper limit) of the pediatric datasets from XR-10 for all parameters.
Body Fluid – Verification of Reference Intervals
A verification study was conducted using a minimum of 20 normal CSF and 20 normal Synovial fluids to verify normal reference ranges cited from published literature (Kjeldsberg's Body Fluids, Third Edition (1993). Reference intervals for other body fluid types have not been established. According to Kjeldsberg (1193), reference intervals for RBC counts are not applicable in body fluid, and therefore, no reference interval for RBC has been established.
Conclusions:
The XR-Series (XR-10) Automated Hematology analyzer and its predicate device, XN-20 Automated Hematology analyzer (K112605), have similar indications for use, fundamental technology, and principles of operation.
Performance, verification, and validation testing were conducted to characterize the performance of the XR-Series (XR-10) analyzer using predetermined acceptance criteria. Results of this testing have demonstrated that the XR-Series (XR-10) analyzer is substantially equivalent to the XN-20 analyzer.
The differences in the XR-Series (XR-10) analyzer and the predicate device (XN-20 analyzer) do not raise any questions regarding safety and effectiveness.
§ 864.5220 Automated differential cell counter.
(a)
Identification. An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.(b)
Classification. Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”