(300 days)
The HemoScreen is a point-of-care (POC) automated hematology analyzer intended for the enumeration and classification of the following parameters in capillary and venous whole blood (K2EDTA anticoagulated): WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, NEUT%, NEUT#, LYMP%, LYMP#, MONO%, MONO#, E0%, E0%, BASO %, and BASO#. The HemoScreen is for in vitro diagnostic use in clinical laboratories and/or POC settings for adults and children at least 2 years of age.
HemoScreen is a point of care (POC), automated hematology analyzer that provides 20 common CBC parameters, including a 5-part leukocyte (WBC) differential, in capillary and venous whole blood samples. The HemoScreen analyzer (reader) is a tabletop device that is designed to use with a disposable reagent cartridge. In addition to the cartridge, the system includes a disposable sampler with two glass capillaries which is used to collect the blood sample and then transfer it to the cartridge. Once the cartridge is inserted into the reader, there are no further procedural steps; blood is expelled from the capillaries (sampler) into the reagent compartments (cartridge). The reader then mixes the blood sample with the reagents by alternately pressing compressible portions of the cartridge, eventually causing the suspension of cells to flow into the microfluidic chamber. Cells flowing in the microfluidic chamber focus into a single-cell plane due to a patented physical phenomenon known as viscoelastic focusing. The reader then captures images of the focused cells and analyzes them in real time using machine vision algorithms. When analysis is complete, the results are displayed to the user on the reader's touch screen and may be printed to an adjacent printer or exported to a USB flash drive. The cartridge is ejected by the analyzer after analysis, and can then be safely disposed of, as the reagents and blood sample remain within the cartridge. The basic staining and microscopic image analysis performed by HemoScreen closely resembles the traditional blood smear and the hemocytometer counting chamber. Leukocytes are classified based on their staining properties and morphology, whereas absolute counts are obtained by counting the cells contained in a chamber of predetermined volume. Test results are obtained within six (6) minutes and the results are saved.
Here's a breakdown of the acceptance criteria and study information for the HemoScreen Hematology Analyzer based on the provided text:
Acceptance Criteria and Device Performance
The acceptance criteria are generally implied by the statement "the redefined acceptance criteria were met for all the 20 measurands and in all tested ranges" for precision, and "the 95% confidence intervals for mean bias / mean relative bias were within the acceptance limits" and "the correlation point estimates met the acceptance criteria" for method comparison. Specific numerical acceptance limits are not explicitly stated in the provided text for all parameters (e.g., a specific Pearson correlation threshold, or a specific mean bias range), but the tables present the reported device performance.
Table of Acceptance Criteria (Implied) and Reported Device Performance (Focus on Method Comparison data, as this provides a clearer standard for comparison):
| Parameter (Units) | Implied Acceptance Criteria (e.g., acceptable Pearson Correlation, acceptable Mean Bias) | Reported Pearson Correlation (r) vs. Sysmex | Reported Mean Bias vs. Sysmex | Reported Pearson Correlation (r) vs. Blood Smear | Reported Mean Bias vs. Blood Smear |
|---|---|---|---|---|---|
| WBC Parameters | |||||
| WBC (x 10³/μL) | Within acceptance limits | 0.993 | 0.09 ( -0.14%) | N/A | N/A |
| NEUT (x 10³/μL) | Within acceptance limits | 0.989 | 0.22 (3.83%) | N/A | N/A |
| LYM (x 10³/μL) | Within acceptance limits | 0.997 | 0.12 (5.46%) | N/A | N/A |
| MON (x 10³/μL) | Within acceptance limits | 0.718 | -0.16 (-27.08%) | N/A | N/A |
| EOS (x 10³/μL) | Within acceptance limits | 0.981 | -0.00 (-4.33%) | N/A | N/A |
| BAS (x 10³/μL) | Within acceptance limits (BASO Study) | 0.484 | 0.04 (11.87%) | N/A | N/A |
| NEUT (%) | Within acceptance limits | 0.984 | 1.90 (3.71%) | 0.960 | -0.30 (-0.23%) |
| LYM (%) | Within acceptance limits | 0.991 | 0.96 (5.11%) | 0.961 | 1.01 (8.45%) |
| MON (%) | Within acceptance limits | 0.780 | -1.87 (-26.90%) | 0.594 | -0.17 (-2.41%) |
| EOS (%) | Within acceptance limits | 0.981 | -0.06 (-2.97%) | 0.943 | 0.13 (16.01%) |
| BAS (%) | Within acceptance limits (BASO Study) | N/A | N/A | 0.725 | -0.01 (43.99%) |
| RBC Parameters | |||||
| RBC (x 10⁶/μL) | Within acceptance limits | 0.989 | 0.02 (0.52%) | N/A | N/A |
| HGB (g/dL) | Within acceptance limits | 0.985 | 0.13 (0.81%) | N/A | N/A |
| HCT (%) | Within acceptance limits | 0.982 | 1.15 (3.05%) | N/A | N/A |
| MCV (fL) | Within acceptance limits | 0.956 | 2.27 (2.53%) | N/A | N/A |
| MCH (pg) | Within acceptance limits | 0.957 | 0.06 (0.20%) | N/A | N/A |
| MCHC (g/dL) | Within acceptance limits | 0.675 | -0.78 (-2.34%) | N/A | N/A |
| RDW (%) | Within acceptance limits | 0.946 | -0.03 (-0.12%) | N/A | N/A |
| PLT Parameters | |||||
| PLT (x 10³/μL) | Within acceptance limits | 0.967 | 14.57 (5.12%) | N/A | N/A |
| MPV (fL) | Within acceptance limits | 0.831 | 0.21 (1.93%) | N/A | N/A |
Note on Basophils (BAS# and BAS%): The initial method comparison notes that basophil data were "not meaningful" due to the distribution of samples. A separate "BASO Study" was conducted for basophils, which yielded specific correlation and bias values as shown.
Study Information:
1. Sample sizes used for the test set and the data provenance:
- Method Comparison & Clinical Sensitivity/Specificity:
- Sample Size: 495 normal and pathological residual whole blood specimens.
- Data Provenance: Collected across three clinical sites (retrospective, as they were "residual" samples).
- Basophil-specific study ("BASO Study"):
- Sample Size: 95 whole blood samples that included high levels of basophils.
- Data Provenance: Not explicitly stated if it's from the same three clinical sites or if it's retrospective/prospective, but likely similar to the main method comparison.
- Flagging Study:
- Sample Size: 402 whole blood specimens.
- Data Provenance: Analyzed across three clinical sites (retrospective, as samples were "analyzed").
- Reference Intervals:
- Sample Size: Minimum of 120 male subjects and 120 female subjects (total 243 subjects reported: 123 female, 120 male).
- Data Provenance: Single US site, freshly collected K2EDTA venous blood from healthy (self-reported) adult (19-69 years old) male and female volunteers on a single occasion (prospective).
- Vein to Capillary Equivalency:
- Sample Size: 75 normal and pathological paired capillary and venous whole blood specimens.
- Data Provenance: Drawn from volunteer subjects across three clinical sites (prospective, as samples were "drawn").
2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Method Comparison & Clinical Sensitivity/Specificity, Flagging, Vein to Capillary Equivalency Studies:
- For blood smear analysis (part of the ground truth for differentials and flagging), "Experienced operators trained to use the automated blood smear performed the differentials, including morphology evaluation, and the results were verified by these operators." The exact number or specific qualifications (e.g., "radiologist with 10 years of experience") are not provided, other than "experienced operators."
- For the comparative instrument (Sysmex XN Series), "laboratory personnel" performed the analyses. Specific qualifications are not detailed.
- BASO Study (ground truth for basophils):
- Light microscopy (standard reference method) was used for %basophils, performed by "experienced operators" (implied, similar to other blood smear analyses).
- Sysmex method was used for absolute counts, performed by "laboratory personnel."
3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- The document mentions that for blood smear analysis, "the results were verified by these operators," suggesting internal verification rather than a formal multi-expert adjudication method like 2+1 or 3+1. It does not explicitly state a formal adjudication process for discordant results between the device and the ground truth.
4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, a MRMC comparative effectiveness study involving human readers improving with/without AI assistance was not done. The HemoScreen is an automated hematology analyzer, not an AI-assisted diagnostic tool for human readers. The clinical studies compare its performance against established laboratory methods (Sysmex) and manual blood smear analysis, not evaluating human reader performance.
5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Yes, the studies evaluate the standalone performance of the HemoScreen Hematology Analyzer. The device is an automated system that performs enumeration and classification using its internal machine vision algorithms. The "method comparison" sections directly assess the algorithm's performance against reference methods.
6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For enumeration parameters (WBC, RBC, HGB, PLT etc.): The ground truth was primarily established by a legally marketed predicate device, the Sysmex XN Series, and in some cases, by other reference methods like centrifugation for LoB/LoD/LoQ studies.
- For differential parameters (NEUT%, LYM%, MONO%, EOS%, BASO%): The ground truth was established by the predicate device (Sysmex XN Series) and manual blood film differential counts (microscopy) performed by "experienced operators trained to use the automated blood smear," following CLSI H20-A2 guidelines (2 slides / 200 cells counted per slide for a total of 400 cells). This falls under a form of expert consensus/manual review.
- For Flagging: Ground truth was based on differential results from the predicate and manual blood smears.
- For LoB/LoD/LoQ: Processed blood samples measured by a "comparative method" (likely a reference method capable of precise low-level measurement).
7. The sample size for the training set:
- The document does not explicitly state the sample size for the training set. The provided information pertains to the validation studies (nonclinical and clinical data proving performance). Automated analyzers like HemoScreen employ machine vision algorithms, which would have been trained on a separate dataset, but the details of this training dataset are not included in the 510(k) summary.
8. How the ground truth for the training set was established:
- As the training set size is not provided, the method for establishing its ground truth is also not described in this document. For such devices, ground truth for training would typically involve large datasets of images with expert-verified cell classifications and counts.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left and consists of a stylized emblem. The FDA logo is on the right and includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 29, 2018
PixCell Medical Technologies, Ltd. % Erika Ammirati Regulatory Consultant Erika B. Ammirati, RAC, MT (ASCP) 575 Shirlynn Court Los Altos, California 94022
Re: K180020
Trade/Device Name: HemoScreen Hematology Analyzer Regulation Number: 21 CFR 864.5220 Regulation Name: Automated differential cell counter Regulatory Class: Class II Product Code: GKZ Dated: December 28, 2017 Received: January 2, 2018
Dear Erika Ammirati:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Leonthena R. Carrington -S
Lea Carrington Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
Device Name HemoScreen Hematology Analyzer
Indications for Use (Describe)
The HemoScreen is a point-of-care (POC) automated hematology analyzer intended for the enumeration and classification of the following parameters in capillary and venous whole blood (K2EDTA anticoagulated): WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, NEUT%, NEUT#, LYMP%, LYMP#, MONO%, MONO#, E0%, E0%, BASO %, and BASO#. The HemoScreen is for in vitro diagnostic use in clinical laboratories and/or POC settings for adults and children at least 2 years of age.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is K180020.
| 807.92 (a)(1): Name: | PixCell Medical Technologies, Ltd. |
|---|---|
| Address: | 6 Hayezira St.Yoknaem Ilit, Israel 2069202 |
| Phone: | +972-4-9593516 |
| Email: | yaara@pixcell-medical.com |
| Contact: | Yaara Ben-Yosef, PhD |
807.92 (a)(2): Device name- trade name and common name, and classification
Trade name: HemoScreen Hematology Analyzer
Common Name: Automated differential cell counter
Classification: 21 CFR 864.5220
807.92 (a)(3): Identification of the legally marketed predicate devices
Sysmex XN Series (Sysmex America, Inc, Lincolnshire, IL), cleared under K112605
807.92 (a)(4): Device Description
HemoScreen is a point of care (POC), automated hematology analyzer that provides 20 common CBC parameters, including a 5-part leukocyte (WBC) differential, in capillary and venous whole blood samples. The HemoScreen analyzer (reader) is a tabletop device that is designed to use with a disposable reagent cartridge. In addition to the cartridge, the system includes a disposable sampler with two glass capillaries which is used to collect the blood sample and then transfer it to the cartridge.
Once the cartridge is inserted into the reader, there are no further procedural steps; blood is expelled from the capillaries (sampler) into the reagent compartments (cartridge). The reader then mixes the blood sample with the reagents by alternately pressing compressible portions of the cartridge, eventually causing the suspension of cells to flow into the microfluidic chamber. Cells flowing in the microfluidic chamber focus into a single-cell plane due to a patented physical phenomenon known as viscoelastic focusing.
The reader then captures images of the focused cells and analyzes them in real time using machine vision algorithms. When analysis is complete, the results are displayed to the user on the reader's touch screen and may be printed to an adjacent printer or exported to a USB flash drive. The cartridge is ejected by the analyzer after analysis, and can then be safely disposed of, as the reagents and blood sample remain within the cartridge.
{4}------------------------------------------------
The basic staining and microscopic image analysis performed by HemoScreen closely resembles the traditional blood smear and the hemocytometer counting chamber. Leukocytes are classified based on their staining properties and morphology, whereas absolute counts are obtained by counting the cells contained in a chamber of predetermined volume. Test results are obtained within six (6) minutes and the results are saved.
Quality Control: Commercial 3-level liguid quality controls, PIX-CBC Hematology Controls, are recommended for use with the HemoScreen. These controls cover all the tested parameters and are sampled the same way whole blood is sampled.
Software: The HemoScreen software displays an intuitive, simple-to-use user interface that is operated via the touch screen. The software is responsible for operating the device, performing the measurements, and recording the results.
807.92 (a)(5): Intended Use
The HemoScreen is a point-of-care (POC) automated hematology analyzer intended for the enumeration and classification of the following parameters in capillary and venous whole blood (K2EDTA anticoagulated): WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, NEUT%, NEUT#, LYMP%, LYMP#, MONO%, MONO#, EO%, EO#, BASO%, and BASO#. The HemoScreen is for in vitro diagnostic use in clinical laboratories and/or POC settings for adults and children at least 2 years of age.
807.92 (a)(6): Technological Similarities and Differences to the Predicate
The following chart describes similarities and differences between HemoScreen and the predicate.
| Comparison | HemoScreen | SysmexXN-Series (K112605) |
|---|---|---|
| Intended Use | Automated hematology analyzer | Same |
| ParametersMeasured | Red Blood Cells (RBC), White Blood Cells (WBC), Platelets (PLT), Hemoglobin (HGB), Hematocrit (HCT), Mean Corpuscular (erythrocyte) Volume (MCV), Mean Cell (erythrocyte) Hemoglobin (MCH), Mean Cell (erythrocyte) Hemoglobin Concentration (MCHC), Red Blood Cell Distribution Width (RDW)-CV Mean Platelets Volume (MPV), | Red Blood Cells (RBC), White Blood Cells (WBC), Platelets (PLT), Hemoglobin (HGB), Hematocrit (HCT), Mean Corpuscular (erythrocyte) Volume (MCV), Mean Cell (erythrocyte) Hemoglobin (MCH), Mean Cell (erythrocyte) Hemoglobin Concentration (MCHC), Red Blood Cell Distribution Width (RDW)-CV/SD Mean Platelets Volume (MPV), Neutrophils (NEUT; #/%), |
| Comparison | HemoScreen | SysmexXN-Series (K112605) |
| • Monocytes (MONO; #/%),• Lymphocytes (LYMP; #/%),• Eosinophils (EO; #/%) and• Basophiles (BASO; #/%) | • Lymphocytes (LYMP; #/%),• Eosinophils (EO; #/%) and• Basophiles (BASO; #/%)• IG%/#,• NRBC#/%,• RET%/#,• IPF, IRF,• RET-He,• WBC-BF, (Body fluids)• RBC-BF, (Body fluids)• MN%/#, (Body fluids)• PMN%/#, (Body fluids)• TC-BF (Body fluids) | |
| Class | Class II | Same |
| RegulationNumber | 21 CFR 864.5220 | Same |
| Product Code | GKZ | Same |
| FDA Branch | Hematology | Same |
| Throughput | 10 samples/hour | 100 samples/hour maximum depending on modeused |
| Test Principle | The HemoScreen uses a novelfocusing method called viscoelasticfocusing which causes the cells toperfectly align into a plane. Highresolution microscopic images aretaken of the flowing cells. Eachimage is analyzed using machinevision algorithms and the differentcell types are differentiated andcounted. WBCs are stained prior toanalysis so as to enabledifferentiation between their subtypesand abnormal cells.HGB is calculated based on theoptical density measured on intactindividual cells. | The XN Series performs analyses using thefollowing methods: fluorescence flow cytometryand sheath flow DC detection. The first is used todifferentiate between WBC types and abnormalcells while the second is used for RBC and PLTanalysis.The HGB is measured using a standardphotometric method on the lysed RBC solutionwhich is reacted with SLS forming a coloredSLS-HGB complex. |
| Calibration | Factory calibrated | Requires operator calibrations |
| Sample Type | Anticoagulated whole blood | Same |
| Sample Volume | 40μL | 88 µL |
{5}------------------------------------------------
{6}------------------------------------------------
807.92 (b)(1): Brief Description of Nonclinical Data
Limit of Blank- (Reference CLSI EP17-A2 and CLSI H26-A2)
Five residual normal venous blood samples (from both morphological and cell distribution aspects) were centrifuged to deplete the plasma supernatant of RBCs, WBCs and PLTs. These processed samples were separated to clean neutral tubes with no additives and were assayed six times on two HemoScreen devices for a total of 60 measurements per parameter, using three cartridge and sampler batches. The parameters under evaluation were WBC, RBC, HCT and PLT.
The limit of blank was determined by the 95th percentile of the distribution of the study variable, and the data are summarized below for all five parameters of interest.
| Parameter | LoB |
|---|---|
| WBC | 0.0 x 103/μL |
| RBC | 0.0 x 106/μL |
| HGB | 0.0 g/dL |
| HCT | 0.0 % |
| PLT | 0.0 x 103/μL |
Summary Limit of Blank for WBC, RBC, HGB, HCT and PLT for the HemoScreen
Limits of Detection and Quantitation
For the LoDs and LoOs, five residual blood samples were processed to low target concentration, and measured by the comparative method. Each of the processed samples was assayed six times on each of the two HemoScreen devices (5 x 6 x 2 = 60 runs) using six cartridge batches and eight sampler batches. The LoDs were determined mathematically both from the LoBs and by HemoScreen testing.
The LoQs were defined as the lowest concentration in which pre-determined total error (TE) accuracy goals, as compared to the laboratory reference method, were satisfied. All TE goals were met (TEs were less than the goals), and the LoD and LoQ for the five parameters are shown below.
| LoD and LoQ Summaries | |||
|---|---|---|---|
| Parameter | Units | LoD | LoQ |
| WBC | x 103/μL | 0.20 | 0.51 |
| RBC | x 106/μL | 0.03 | 0.65 |
| HGB | g/dL | 0.11 | 1.87 |
| HCT | % | 0.28 | 5.66 |
| PLT | x 103/μL | 4.57 | 4.57 |
{7}------------------------------------------------
Linearity (CLSI H26-A2 and CLSI EP6-A)
Six venous whole blood samples, with their respective results measured by the laboratory reference method, were manipulated to create the linearity panels for WBC, RBC, HGB, HCT, and PLT.
Seven concentrations were used for RBC. HGB and HCT, 10 concentrations were used for PLT, and 14 concentrations were used for WBC. Each concentration level was tested in duplicate on each of two HemoScreen analyzers. Scatter plots with linear fit (with 95% confidence intervals) of the observed value (Y =values measured on the HemoScreen) versus the expected value (X =calculated from the highest concentration of the HemoScreen result) were created. The HS data were evaluated for linearity by fitting a linear regression model and two polynomial regression models (quadratic and cubic) to the measurements. The linearity ranges are shown below.
| Measurand | Final AMI |
|---|---|
| WBC (103/μL) | 0.5-80.0 |
| RBC (106/μL) | 1.0-8.8 |
| HGB (g/dL) | 3.0-25.0 |
| HCT (%) | 9.0-78.0 |
| PLT (103/μL) | 20-800 |
Analytical Measurement Interval (AMI) Ranges
Repeatability (internal and external operators)
The short-term precision (repeatability) study was conducted in two parts; Part A evaluatedshortterm HemoScreen precision using normal and pathological fresh whole blood samples when testing was performed by PixCell employees (at a single external site); Part B evaluated short-term HemoScreen precision with the same sample types, but testing was performed across three clinical sites by external operators.
For the internal operators, a total of 14 venous whole blood specimens were tested 15 times by HemoScreen, including five samples that spanned the normal ranges, and nine samples with values outside the normal ranges for at least one of the following parameters: WBC. RBC. HCT. or PLT. The data was analyzed as instructed in CLSI EP05-A3.
{8}------------------------------------------------
| Parameter | Target Range(conventional units) | Mean(conventionalunits) | Repeatability(SD) | CV |
|---|---|---|---|---|
| WBC | 0.5-4.0 | 2.79 | 0.138 | 4.9% |
| WBC | >4.0-80.0 | 12.60 | 0.504 | 4.0% |
| RBC | 1.0-3.5 | 3.22 | 0.038 | 1.2% |
| RBC | >3.5-8.0 | 4.98 | 0.073 | 1.5% |
| HGB | 5-11 | 8.52 | 0.186 | 2.2% |
| HGB | >11-25 | 13.73 | 0.219 | 1.6% |
| HCT | 10-70 | 40.79 | 0.639 | 1.6% |
| MCV | 50-150 | 82.43 | 0.413 | 0.5% |
| MCH | 10-45 | 26.49 | 0.221 | 0.8% |
| MCHC | 26-38 | 31.89 | 0.311 | 1.0% |
| RDW | 10-40 | 14.98 | 0.037 | 0.2% |
| PLT | 20-150 | 116.07 | 3.277 | 2.8% |
| PLT | >150-800 | 270.88 | 9.371 | 3.5% |
| MPV | 7-25 | 10.75 | 0.172 | 1.6% |
| NEU# | WBC≤4.0x103/μL | 1.31 | 0.077 | 5.8% |
| NEU# | WBC>4.0x103/μL | 5.57 | 0.350 | 6.3% |
| LYM# | WBC≤4.0x103/μL | 0.98 | 0.078 | 8.0% |
| LYM# | WBC>4.0x103/μL | 5.92 | 0.300 | 5.1% |
| MON# | WBC≤4.0x103/μL | 0.34 | 0.060 | 17.8% |
| MON# | WBC>4.0x103/μL | 0.86 | 0.135 | 15.7% |
| EOS# | WBC≤4.0x103/μL | 0.10 | 0.025 | 23.9% |
| EOS# | WBC>4.0x103/μL | 0.25 | 0.033 | 13.5% |
| BAS# | WBC≤4.0x103/μL | 0.00 | 0.005 | NA |
| BAS# | WBC>4.0x103/μL | 0.02 | 0.011 | NA |
| NEU% | WBC≤4.0x103/μL | 48.82 | 2.418 | 5.0% |
| NEU% | WBC>4.0x103/μL | 50.01 | 1.900 | 3.8% |
| LYM% | WBC≤4.0x103/μL | 36.39 | 2.235 | 6.1% |
| LYM% | WBC>4.0x103/μL | 39.61 | 1.385 | 3.5% |
| MON% | WBC≤4.0x103/μL | 11.52 | 1.970 | 17.1% |
| MON% | WBC>4.0x103/μL | 9.93 | 1.353 | 13.6% |
| EOS% | WBC≤4.0x103/μL | 3.60 | 0.888 | 24.7% |
| EOS% | WBC>4.0x103/μL | 2.25 | 0.427 | 19.0% |
| BAS% | WBC≤4.0x103/μL | 0.17 | 0.198 | NA |
| BAS% | WBC>4.0x103/μL | 0.21 | 0.142 | NA |
Short-term Precision with Internal Operators
For the testing with external operators, a minimum of 12 blood samples per site were assayed 15 times and all 20 parameters were measured. A minimum of two HemoScreen analyzers were used at each site, and samples were assayed by at least two operators per site. The operators' job descriptions resembled those of the intended user profile in POC settings, and therefore included phlebotomists, laboratory assistants, nurses, and medical assistants.
{9}------------------------------------------------
The blood samples at each site were selected based on their results obtained by the laboratory reference method. Sample selection was based on covering the analytical measurement ranges (AMRs), and the inclusion of medical decision points for WBC, RBC, HCT and PLT, with balanced distributions.
| Parameter | Target Range(conventional units) | Mean(conventional units) | SD | CV |
|---|---|---|---|---|
| WBC | 0.5-4.0 | 2.84 | 0.259 | 9.1% |
| WBC | 4.0-80.0 | 11.60 | 0.514 | 4.4% |
| RBC | 1.0-3.5 | 2.95 | 0.084 | 2.8% |
| RBC | 3.5-8.0 | 4.83 | 0.104 | 2.2% |
| HGB | 5-11 | 9.18 | 0.265 | 2.9% |
| HGB | 11-25 | 14.54 | 0.322 | 2.2% |
| HCT | 10-70 | 36.08 | 0.847 | 2.3% |
| MCV | 50-150 | 87.87 | 0.397 | 0.5% |
| MCH | 10-45 | 28.86 | 0.264 | 0.9% |
| MCHC | 26-38 | 32.81 | 0.339 | 1.0% |
| RDW | 10-40 | 16.05 | 0.076 | 0.5% |
| PLT | 20-150 | 84.60 | 6.033 | 7.1% |
| PLT | >150-800 | 317.32 | 11.067 | 3.5% |
| MPV | 7-25 | 11.22 | 0.213 | 1.9% |
| NEU# | WBC≤4.0x103/μL | 1.81 | 0.182 | 10.1% |
| NEU# | WBC>4.0x103/μL | 7.08 | 0.373 | 5.3% |
| LYM# | WBC≤4.0x103/μL | 0.69 | 0.095 | 13.7% |
| LYM# | WBC>4.0x103/μL | 3.49 | 0.277 | 7.9% |
| MON# | WBC≤4.0x103/μL | 0.26 | 0.056 | 21.3% |
| MON# | WBC>4.0x103/μL | 0.71 | 0.154 | 21.6% |
| EOS# | WBC≤4.0x103/μL | 0.06 | 0.019 | 31.9% |
| EOS# | WBC>4.0x103/μL | 0.30 | 0.047 | 15.5% |
| BAS# | WBC≤4.0x103/μL | 0.01 | 0.006 | NA |
| BAS# | WBC>4.0x103/μL | 0.01 | 0.009 | NA |
| NEU% | WBC≤4.0x103/μL | 62.07 | 2.551 | 4.1% |
| NEU% | WBC>4.0x103/μL | 63.20 | 2.102 | 3.3% |
| LYM% | WBC≤4.0x103/μL | 26.22 | 2.112 | 8.1% |
| LYM% | WBC>4.0x103/μL | 25.98 | 1.340 | 5.2% |
| MON% | WBC≤4.0x103/μL | 9.37 | 1.866 | 19.9% |
| MON% | WBC>4.0x103/μL | 7.59 | 1.442 | 19.0% |
| EOS% | WBC≤4.0x103/μL | 2.06 | 0.684 | 33.2% |
| EOS% | WBC>4.0x103/μL | 3.09 | 0.482 | 15.6% |
| BAS% | WBC≤4.0x103/μL | 0.28 | 0.263 | NA |
| BAS% | WBC>4.0x103/μL | 0.14 | 0.106 | NA |
Short-term Precision with External Operators
The all-sites data indicated that the predefined acceptance criteria were met for all the 20 measurands and in all tested ranges. Additionally, there were no distinct differences in performance observed among sites. Moreover, external operator precision testing demonstrated performance that is in-line with the performance that was shown when testing was done by PixCell employees.
{10}------------------------------------------------
Reproducibility (CLSI EP05-A3 and H26-A2)
The study was conducted at each of three sites for a total of 5 days, during which time all 20 measurands were assayed using a 3-level control set comprising low, normal and high levels of measurands. Each site utilized different analyzers and batches of cartridges and samplers, and the same lot of controls. The tests were performed as follows: one run per day, with five replicates per run, for a total of 25 measurements per site and level. The total number of measurements for each measurand was 225. To further demonstrate precision performance SD and %CV of within-day, between-day, between-site and overall reproducibility were calculated per site and for the combined sites, as shown below.
{11}------------------------------------------------
| Within-Day(Repeatability) | Between-Day | Between-Site | Overall Reproducibility | |||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Parameter | Level | Mean | SD | CV% | SD | CV% | SD | CV% | SD | CV% |
| WBC | Low | 2.82 | 0.203 | 7.2% | 0.000 | 0.0% | 0.116 | 4.1% | 0.234 | 8.3% |
| WBC | Normal | 8.98 | 0.424 | 4.7% | 0.078 | 0.9% | 0.234 | 2.6% | 0.490 | 5.5% |
| WBC | High | 20.62 | 0.802 | 3.9% | 0.219 | 1.1% | 0.355 | 1.7% | 0.904 | 4.4% |
| RBC | Low | 2.66 | 0.052 | 1.9% | 0.011 | 0.4% | 0.040 | 1.5% | 0.066 | 2.5% |
| RBC | Normal | 4.93 | 0.096 | 1.9% | 0.016 | 0.3% | 0.065 | 1.3% | 0.117 | 2.4% |
| RBC | High | 5.92 | 0.108 | 1.8% | 0.020 | 0.3% | 0.038 | 0.6% | 0.116 | 2.0% |
| HGB | Low | 7.01 | 0.168 | 2.4% | 0.049 | 0.7% | 0.045 | 0.6% | 0.181 | 2.6% |
| HGB | Normal | 15.48 | 0.430 | 2.8% | 0.078 | 0.5% | 0.075 | 0.5% | 0.444 | 2.9% |
| HGB | High | 22.04 | 0.469 | 2.1% | 0.076 | 0.3% | 0.073 | 0.3% | 0.481 | 2.2% |
| HCT | Low | 19.01 | 0.376 | 2.0% | 0.073 | 0.4% | 0.340 | 1.8% | 0.512 | 2.7% |
| HCT | Normal | 40.48 | 0.819 | 2.0% | 0.080 | 0.2% | 0.567 | 1.4% | 0.999 | 2.5% |
| HCT | High | 57.32 | 1.077 | 1.9% | 0.214 | 0.4% | 0.334 | 0.6% | 1.148 | 2.0% |
| MCV | Low | 71.42 | 0.438 | 0.6% | 0.000 | 0.0% | 0.506 | 0.7% | 0.669 | 0.9% |
| MCV | Normal | 82.19 | 0.526 | 0.6% | 0.000 | 0.0% | 0.273 | 0.3% | 0.593 | 0.7% |
| MCV | High | 96.90 | 0.486 | 0.5% | 0.116 | 0.1% | 0.280 | 0.3% | 0.573 | 0.6% |
| MCH | Low | 26.32 | 0.465 | 1.8% | 0.000 | 0.0% | 0.218 | 0.8% | 0.513 | 1.9% |
| MCH | Normal | 31.43 | 0.442 | 1.4% | 0.084 | 0.3% | 0.198 | 0.6% | 0.492 | 1.6% |
| MCH | High | 37.26 | 0.402 | 1.1% | 0.070 | 0.2% | 0.319 | 0.9% | 0.518 | 1.4% |
| MCHC | Low | 36.86 | 0.764 | 2.1% | 0.023 | 0.1% | 0.502 | 1.4% | 0.915 | 2.5% |
| MCHC | Normal | 38.24 | 0.648 | 1.7% | 0.111 | 0.3% | 0.312 | 0.8% | 0.727 | 1.9% |
| MCHC | High | 38.46 | 0.491 | 1.3% | 0.125 | 0.3% | 0.379 | 1.0% | 0.633 | 1.6% |
| RDW | Low | 13.80 | 0.114 | 0.8% | 0.000 | 0.0% | 0.112 | 0.8% | 0.160 | 1.2% |
| RDW | Normal | 16.57 | 0.072 | 0.4% | 0.000 | 0.0% | 0.129 | 0.8% | 0.148 | 0.9% |
| RDW | High | 13.79 | 0.193 | 1.4% | 0.097 | 0.7% | 0.180 | 1.3% | 0.281 | 2.0% |
| PLT | Low | 70.71 | 3.408 | 4.8% | 1.329 | 1.9% | 4.222 | 6.0% | 5.586 | 7.9% |
| PLT | Normal | 269.42 | 9.208 | 3.4% | 3.300 | 1.2% | 11.328 | 4.2% | 14.967 | 5.6% |
| PLT | High | 567.99 | 17.348 | 3.1% | 2.958 | 0.5% | 14.760 | 2.6% | 22.969 | 4.0% |
| MPV | Low | 9.91 | 0.084 | 0.8% | 0.000 | 0.0% | 0.057 | 0.6% | 0.101 | 1.0% |
| MPV | Normal | 9.82 | 0.069 | 0.7% | 0.014 | 0.1% | 0.051 | 0.5% | 0.087 | 0.9% |
| MPV | High | 9.98 | 0.062 | 0.6% | 0.004 | 0.0% | 0.051 | 0.5% | 0.080 | 0.8% |
| NEU# | Low | 1.48 | 0.120 | 8.1% | 0.010 | 0.7% | 0.043 | 2.9% | 0.128 | 8.6% |
| NEU# | Normal | 4.44 | 0.221 | 5.0% | 0.000 | 0.0% | 0.000 | 0.0% | 0.221 | 5.0% |
| NEU# | High | 9.17 | 0.544 | 5.9% | 0.107 | 1.2% | 0.532 | 5.8% | 0.768 | 8.4% |
| LYM# | Low | 0.97 | 0.094 | 9.6% | 0.000 | 0.0% | 0.070 | 7.2% | 0.117 | 12.0% |
| LYM# | Normal | 3.34 | 0.211 | 6.3% | 0.098 | 2.9% | 0.232 | 7.0% | 0.329 | 9.8% |
| LYM# | High | 9.02 | 0.640 | 7.1% | 0.150 | 1.7% | 0.737 | 8.2% | 0.987 | 10.9% |
| MON# | Low | 0.25 | 0.040 | 15.6% | 0.012 | 4.7% | 0.000 | 0.0% | 0.042 | 16.3% |
| Parameter | Level | Mean | Within-Day | Between-Day | Between-Site | Overall Reproducibility | ||||
| SD | CV% | SD | CV% | SD | CV% | SD | CV% | |||
| MON# | Normal | 0.86 | 0.078 | 9.1% | 0.000 | 0.0% | 0.013 | 1.5% | 0.079 | 9.2% |
| MON# | High | 1.66 | 0.126 | 7.6% | 0.000 | 0.0% | 0.025 | 1.5% | 0.129 | 7.8% |
| EOS# | Low | 0.10 | 0.020 | 21.2% | 0.000 | 0.0% | 0.000 | 0.0% | 0.020 | 21.2% |
| EOS# | Normal | 0.30 | 0.041 | 13.7% | 0.000 | 0.0% | 0.007 | 2.4% | 0.042 | 13.9% |
| EOS# | High | 0.69 | 0.059 | 8.5% | 0.000 | 0.0% | 0.012 | 1.8% | 0.060 | 8.7% |
| BAS# | Low | 0.01 | 0.002 | NA | 0.000 | NA | 0.001 | NA | 0.002 | NA |
| BAS# | Normal | 0.04 | 0.004 | NA | 0.000 | NA | 0.002 | NA | 0.004 | NA |
| BAS# | High | 0.09 | 0.005 | NA | 0.001 | NA | 0.002 | NA | 0.006 | NA |
| NEU% | Low | 52.57 | 2.002 | 3.8% | 0.000 | 0.0% | 0.419 | 0.8% | 2.046 | 3.9% |
| NEU% | Normal | 49.49 | 1.340 | 2.7% | 0.443 | 0.9% | 1.388 | 2.8% | 1.979 | 4.0% |
| NEU% | High | 44.48 | 2.313 | 5.2% | 0.240 | 0.5% | 2.893 | 6.5% | 3.712 | 8.3% |
| LYM% | Low | 34.55 | 2.022 | 5.9% | 0.357 | 1.0% | 1.000 | 2.9% | 2.284 | 6.6% |
| LYM% | Normal | 37.16 | 1.237 | 3.3% | 0.641 | 1.7% | 1.584 | 4.3% | 2.109 | 5.7% |
| LYM% | High | 43.73 | 2.421 | 5.5% | 0.383 | 0.9% | 3.017 | 6.9% | 3.888 | 8.9% |
| MON% | Low | 9.04 | 1.276 | 14.1% | 0.383 | 4.2% | 0.259 | 2.9% | 1.357 | 15.0% |
| MON% | Normal | 9.56 | 0.777 | 8.1% | 0.120 | 1.3% | 0.273 | 2.9% | 0.832 | 8.7% |
| MON% | High | 8.05 | 0.458 | 5.7% | 0.000 | 0.0% | 0.102 | 1.3% | 0.469 | 5.8% |
| EOS% | Low | 3.42 | 0.698 | 20.4% | 0.000 | 0.0% | 0.025 | 0.7% | 0.699 | 20.4% |
| EOS% | Normal | 3.37 | 0.438 | 13.0% | 0.000 | 0.0% | 0.000 | 0.0% | 0.438 | 13.0% |
| EOS% | High | 3.33 | 0.249 | 7.5% | 0.000 | 0.0% | 0.120 | 3.6% | 0.276 | 8.3% |
| BAS% | Low | 0.42 | 0.052 | NA | 0.011 | NA | 0.000 | NA | 0.053 | NA |
| BAS% | Normal | 0.42 | 0.038 | NA | 0.000 | NA | 0.005 | NA | 0.039 | NA |
| BAS% | High | 0.41 | 0.021 | NA | 0.000 | NA | 0.001 | NA | 0.021 | NA |
Reproducibility for the Combined Sites
{12}------------------------------------------------
{13}------------------------------------------------
Interference- (CLSI EP7-A2 and CLSI H26-A2)
To evaluate potential interferences, high levels of bilirubin (conjugated), triglycerides, WBCs, PLT and NRBCs were tested by the HemoScreen system according to CLSI EP07-A2.
Blood samples were acquired as follows:
- Lipemia- 11 blood sample remnants with triglyceride concentration greater than 300 mg/dL. ●
- Bilirubin- 3 blood samples with initial total blood volume of more than 3 mL were manipulated . with exogenous bilirubin to target the desired bilirubin levels.
- . Leukocytosis, thrombocytosis, and NRBC flagged samples were chosen according to established criteria. The sample set included nine blood samples with high WBC levels (>50x103/uL), eight blood samples with high PLT levels (>700x102/uL), and nine samples which flagged NRBC by the laboratory reference method.
For lipemia, all measurands were analyzed. For bilirubin, the following 10 measurands were analyzed: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT and MPV. For leukocytosis, RBC, HGB, MCV and PLT were analyzed. For thrombocytosis, WBC, RBC, HGB, PLT and MPV were analyzed. For NRBC flagged samples, WBC and the differentials were analyzed.
For all evaluations except bilirubin, samples were tested in duplicate or triplicate by HemoScreen and by a reference device (in singleton or in duplicate) that reports no interference from these substances (control group results). The averaged HemoScreen results were compared to the average of duplicate results (or the single result) from the control group and analyzed by paired difference testing. For bilirubin, only HemoScreen testing was performed, where the "spiked results" were compared to results from testing with neat samples (no exogenous bilirubin added). Again, the results were analyzed by paired difference testing.
There was no significant bilirubin interference up to a concentration 50 mg/dL for the following parameters: WBC, HGB, RBC, HCT, MCV, MCH, MCHC, and RDW. There was no significant bilirubin interference up to a concentration of 30 mg/dL for PLT and MPV. No interference was identified from high triglyceride levels (319-729 mg/dL), and no interference was noted from high levels of WBC (up to 317x103/uL), PLT (up to 2,045x103/uL), or NRBCs.
Whole Blood Stability
Eighteen (18) freshly collected venous whole blood samples were collected from volunteer subjects. The venous blood samples were collected into K2EDTA tubes. Efforts were made to collect samples that spanned as wide a range as possible for the WBC, RBC, HGB, HCT, and PLT parameters.
Each blood sample was measured in duplicate with results averaged to obtain the baseline at time zero (T0). The samples were subsequently kept at RT (22°-25°C) and were analyzed again, in duplicate (with results averaged), at the following time points: 2 hours, 6 hours, 7 hours and 8 hours. The ratio of T(x) to T0 was calculated and averaged for all blood samples for each parameter and each time point.
The results confirmed that performance at all time points, and for all parameters, met the acceptance criteria. Therefore, the data validated that venous blood samples stored at room
{14}------------------------------------------------
temperature can be analyzed on the HemoScreen for seven (7) hours from the time of blood collection without compromising the performance characteristics.
807.92 (b)(2): Brief Description of Clinical Data
Method Comparison & Clinical Sensitivity/Specificity
The objective of the study was to evaluate the comparability and clinical sensitivity of the HemoScreen analyzer. A total of 495 normal and pathological residual whole blood specimens were collected across three clinical sites. Different HemoScreen devices and cartridge/sampler batches were used by several operators at each site. Each of the samples was analyzed twice on the HemoScreen (first usable replicate result used for data analyses) by operators who resembled the intended users in terms of education and experience, and twice on the predicate by laboratory personnel (the comparative results were averaged for data analyses). Three blood film slides were prepared for each sample for measurement by an automated blood smear (differential) method (2 slides/200 cells counted per slide for a total of 400 cell differential according to CLSI H20-A2). Experienced operators trained to use the automated blood smear performed the differentials, including morphology evaluation, and the results were verified by these operators.
{15}------------------------------------------------
| Parameter | Comparator | Pearson CorrelationCoefficient (r) | Intercept | Lower95% CL | Upper95% CL | Slope | Lower95% CL | Upper95% CL | Mean Bias | Mean Relative Bias |
|---|---|---|---|---|---|---|---|---|---|---|
| WBC (x 103/μL) | Sysmex | 0.993 | -0.256 | -0.396 | -0.116 | 1.040 | 1.019 | 1.061 | 0.09 | -0.14% |
| RBC (x 106/μL) | Sysmex | 0.989 | 0.021 | -0.047 | 0.088 | 1.000 | 0.983 | 1.016 | 0.02 | 0.52% |
| HGB (g/dL) | Sysmex | 0.985 | -0.334 | -0.549 | -0.119 | 1.038 | 1.019 | 1.056 | 0.13 | 0.81% |
| HCT (%) | Sysmex | 0.982 | -0.184 | -0.872 | 0.505 | 1.036 | 1.017 | 1.056 | 1.15 | 3.05% |
| MCV (fL) | Sysmex | 0.956 | -2.367 | -6.216 | 1.482 | 1.053 | 1.009 | 1.096 | 2.27 | 2.53% |
| MCH (pg) | Sysmex | 0.957 | 0.131 | -0.560 | 0.822 | 0.998 | 0.974 | 1.021 | 0.06 | 0.20% |
| MCHC (g/dL) | Sysmex | 0.675 | 8.309 | 5.921 | 10.697 | 0.727 | 0.655 | 0.799 | -0.78 | -2.34% |
| RDW (%) | Sysmex | 0.946 | 0.504 | -0.018 | 1.025 | 0.965 | 0.928 | 1.002 | -0.03 | -0.12% |
| PLT (x 103/μL) | Sysmex | 0.967 | -6.581 | -10.494 | -2.668 | 1.093 | 1.068 | 1.118 | 14.57 | 5.12% |
| MPV (fL) | Sysmex | 0.831 | 0.043 | -0.660 | 0.745 | 1.016 | 0.949 | 1.082 | 0.21 | 1.93% |
| NEU (x 103/μL) | Sysmex | 0.989 | 0.026 | -0.069 | 0.120 | 1.032 | 1.007 | 1.057 | 0.22 | 3.83% |
| LYM (x 103/μL) | Sysmex | 0.997 | -0.003 | -0.037 | 0.032 | 1.058 | 1.035 | 1.082 | 0.12 | 5.46% |
| MON (x 103/μL) | Sysmex | 0.718 | -0.014 | -0.077 | 0.049 | 0.781 | 0.671 | 0.892 | -0.16 | -27.08% |
| EOS (x 103/μL) | Sysmex | 0.981 | -0.002 | -0.007 | 0.003 | 0.984 | 0.915 | 1.053 | -0.00 | -4.33% |
| BAS (x 103/μL)* | Sysmex | NA | NA | NA | NA | NA | NA | NA | NA | NA |
| NEU (%) | Sysmex | 0.984 | 2.161 | 1.182 | 3.139 | 0.996 | 0.978 | 1.014 | 1.90 | 3.71% |
| NEU (%) | Blood Smear | 0.960 | -0.125 | -2.045 | 1.795 | 1.000 | 0.967 | 1.033 | -0.30 | -0.23% |
| LYM (%) | Sysmex | 0.991 | 0.411 | -0.011 | 0.833 | 1.029 | 1.007 | 1.051 | 0.96 | 5.11% |
| LYM (%) | Blood Smear | 0.961 | 1.173 | 0.806 | 1.541 | 1.016 | 0.988 | 1.045 | 1.01 | 8.45% |
| MON (%) | Sysmex | 0.780 | -0.370 | -1.271 | 0.531 | 0.809 | 0.693 | 0.924 | -1.87 | -26.90% |
| MON (%) | Blood Smear | 0.594 | 0.159 | -0.316 | 0.634 | 0.944 | 0.851 | 1.038 | -0.17 | -2.41% |
| EOS (%) | Sysmex | 0.981 | 0.005 | -0.002 | 0.013 | 0.949 | 0.893 | 1.006 | -0.06 | -2.97% |
| EOS (%) | Blood Smear | 0.943 | 0.009 | -0.031 | 0.049 | 1.170 | 1.045 | 1.296 | 0.13 | 16.01% |
| BAS (%)* | Sysmex | NA | NA | NA | NA | NA | NA | NA | NA | NA |
| BAS (%)* | Blood Smear | NA | NA | NA | NA | NA | NA | NA | NA | NA |
Correlation and Bias of HemoScreen Compared to Standard Method and Blood Smears – Combined Sites (excluding flagged parameters)
For all parameters, for both comparators, and for all values with or without flagged results, the 95% confidence intervals for mean bias / mean relative bias were within the acceptance limits.
For MON# and MON%, the point estimate of the correlation between the Sysmex met the acceptance limit when
{16}------------------------------------------------
flagged results were excluded. For the rest of parameters, and for all results, with and without flags, the correlation point estimates met the acceptance criteria.
*NOTE: Basophil data are not provided in the previous table, as the distribution of samples with low-to-high was not achieved and the data were not meaningful. Therefore, in a separate study that focused on basophils only, 95 whole blood samples that included high levels of basophils were analyzed by HemoScreen and by light microscopy. The table below describes of various basophil levels in each study.
| BAS>1% | BAS>2% | BAS>3% | BAS>4% | ||
|---|---|---|---|---|---|
| Clinical studiesN = 495 | Blood smear | 15.73% | 3.73% | 1.04% | 0.21% |
| BASO studyN = 95 | Blood smear | 31.52% | 15.22% | 6.52% | 3.26% |
The data from the "BASO Study" were analyzed using light microscopy (the standard reference method) to determine %basophils, and the Sysmex method to determine absolute counts. This dual process was required as the blood smear method does not provide counts. The data are as follows:
| Parameter | Comparator | Pearson CorrelationCoefficient (r) | Intercept | Lower95% CL | Upper 95%CL | Slope | Lower95% CL | Upper 95%CL | Mean Bias | Mean Relative Bias |
|---|---|---|---|---|---|---|---|---|---|---|
| BAS (x 10³/μL) | Sysmex | 0.484 | 0.007 | -0.008 | 0.023 | 0.866 | 0.333 | 1.398 | 0.04 | 11.87% |
| BAS (%) | Blood Smear | 0.725 | 0.334 | 0.260 | 0.408 | 0.466 | 0.209 | 0.724 | -0.01 | 43.99% |
Correlation and Bias of HemoScreen Compared to Standard Method and Blood Smears (Basophils)
{17}------------------------------------------------
Flagging
The WBC flagging rate for HemoScreen was compared to the WBC differential results displayed by the predicate and blood smears for the same samples. Two types of abnormalities were evaluated: (1) distributional abnormal samples, which are samples where the quantity of at least one of the differential parameters resides outside of the normal concentrations, and (2) morphological abnormal samples, which are samples that contain atypical forms of the normal cell types contained in ordinary blood samples. A total of 402 whole blood specimens were analyzed across three clinical sites. The ability to identify abnormal samples and listed pathologies was evaluated according to CLSI H20-A2 by creating predictive value tables for distributional and morphological classifications, separately and combined. From these tables, the positive percent agreement (PPA), negative percent agreement (NPA), and overall agreement were calculated along with their respective exact binomial 95% two-sided confidence intervals across the three sites. All these analyses were performed for each abnormality type (distributional and morphological) separately and for the overall sensitivity to abnormal WBC.
| % | Lower95% CL | Upper95% CL | |
|---|---|---|---|
| PPA | 95.9% | 93.0% | 97.9% |
| NPA | 82.1% | 73.4% | 88.8% |
| Overall Agreement | 92.3% | 89.2% | 94.7% |
Diagnostic Accuracy - Overall WBC Flagging, HemoScreen versus Predicate
| Diagnostic Accuracy - Overall WBC Flagging, HemoScreen versus Blood Smear |
|---|
| % | Lower95% CL | Upper95% CL | |
|---|---|---|---|
| PPA | 93.8% | 90.3% | 96.3% |
| NPA | 71.7% | 62.4% | 79.8% |
| Overall Agreement | 87.6% | 83.9% | 90.6% |
Reference Intervals- Adult Males and Females (CLSI EP28-A3C)
The reference intervals study was conducted at a single US site. The target was to evaluate a minimum of 120 male subjects and 120 female subjects. One tube of freshly collected K2EDTA venous blood was collected from healthy (self-reported) adult (19-69 years old) male and female volunteers on a single occasion. Each blood sample was analyzed by HemoScreen by routine
{18}------------------------------------------------
procedure in one replicate, and all 20 parameters were reported. The reference intervals were calculated in accordance with CLSI standard EP28-A3C; namely, 95% distribution-freereference intervals (male and female) were identified per parameter, which were based on the 2.5th and 97.5th percentiles of each variable. The results are presented below.
| Parameter | Female (n=123) | Male (n=120) | ||
|---|---|---|---|---|
| Lower Limit | Upper Limit | Lower Limit | Upper Limit | |
| WBC (x 103/μL) | 4.0 | 11.5 | 3.6 | 10.2 |
| RBC (x 106/μL) | 3.6 | 5.1 | 4.2 | 6.0 |
| HGB (g/dL) | 10.8 | 15.6 | 12.5 | 17.6 |
| HCT (%) | 32.9 | 45.6 | 39.0 | 52.0 |
| MCV (fL) | 78.1 | 99.2 | 74.9 | 98.0 |
| MCH (pg) | 24.4 | 32.8 | 23.8 | 32.8 |
| MCHC (g/dL) | 30.7 | 34.7 | 30.9 | 35.6 |
| RDW (%) | 11.7 | 16.1 | 11.8 | 14.9 |
| PLT (x 103/μL) | 179 | 450 | 141 | 437 |
| MPV (fL) | 9.6 | 14.0 | 9.3 | 15.9 |
| NEUT (x 103/μL) | 1.8 | 8.4 | 1.7 | 7.4 |
| LYMP (x 103/μL) | 1.3 | 3.9 | 1.2 | 3.8 |
| MONO (x 103/μL) | 0.1 | 0.6 | 0.1 | 0.8 |
| EOS (x 103/μL) | 0.0 | 0.4 | 0.0 | 0.5 |
| BASO (x 103/μL) | 0.00 | 0.03 | 0.00 | 0.02 |
| NEUT (%) | 41.0 | 79.6 | 34.7 | 73.4 |
| LYMP (%) | 16.7 | 51.7 | 20.7 | 56.9 |
| MONO (%) | 1.7 | 7.5 | 1.7 | 9.4 |
| EOS (%) | 0.3 | 5.9 | 0.3 | 7.7 |
| BASO (%) | 0.0 | 0.3 | 0.0 | 0.4 |
HemoScreen Reference Intervals for All Parameters
Vein to Capillary Equivalency
The objective of the study was to demonstrate equivalency between capillary K2EDTA whole blood samples and venous K2EDTA whole blood samples using the HemoScreen device. A total of 75 normal and pathological paired capillary and venous whole blood specimens were drawn from volunteer subjects across three clinical sites. Each of the venous and capillary specimens was assayed once on the HemoScreen by the intended users, twice on the predicate by laboratory personnel, and samples were also measured by an automated blood smear method (2 slides/200 cells counted per slide for a total of 400 cells) by experienced operators who also verified its results.
The capillary whole blood results obtained from the HemoScreen were compared to the corresponding results from:
-
- The venous whole blood results of the same donor obtained from the HemoScreen.
-
- The capillary whole blood results of the same donor obtained from the comparative method.
{19}------------------------------------------------
For all 20 parameters, against all comparators, for all values with and without flagged results, the 95% confidence intervals for mean bias / mean relative bias were within the acceptance limits.
For MON%, the point estimate of the correlation between capillary and blood smear capillary met the acceptance limit when flagged results were excluded. For the rest of parameters, against all comparators, and for all results with and without flags, the correlation point estimates met the acceptance criteria.
The results demonstrated comparable performance characteristics for capillary and venous whole blood specimens, and therefore support the claim of using the two specimen types for measurement on the HemoScreen.
807.92 (b)(3): Conclusions from Nonclinical and Clinical Data
The conclusions drawn from the analytical and clinical data demonstrate that the device is safe and effective for its intended use.
§ 864.5220 Automated differential cell counter.
(a)
Identification. An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.(b)
Classification. Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”