Search Results
Found 18 results
510(k) Data Aggregation
(237 days)
For In Vitro Diagnostic Use
HALO AP Dx is a software only device intended as an aid to the pathologist to review, interpret and manage digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue for the purposes of pathology primary diagnosis. HALO AP Dx is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens.
It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and, where necessary, use conventional light microscopy review when making a diagnostic decision. HALO AP Dx is intended for use with the Hamamatsu NanoZoomer S360MD Slide scanner and the JVC Kenwood JD-C240BN01A display.
HALO AP Dx, version 2.1 is a browser-based software-only device intended to aid pathology professionals in viewing, manipulating, management and interpretation of digital pathology whole slide images (WSI) of glass slides obtained from the Hamamatsu Photonics K.K. NanoZoomer S360MD scanner and viewed on the JVC Kenwood JD-C240BN01A display.
HALO AP Dx is typically operated as follows:
- Image acquisition is performed using the predicate device, NanoZoomer S360MD Slide scanner according to its Instructions for Use. The operator performs quality control of the digital slides per the instructions of the NanoZoomer and lab specifications to determine if re-scans are necessary.
- Once image acquisition is complete, the unaltered image is saved by the scanner's software to an image storage location. HALO AP Dx ingests the image, and a copy of image metadata is stored in the subject device's database to improve viewing response times.
- Scanned images are reviewed by scanning personnel such as histotechnicians to confirm image quality and initiate any re-scans before making it available to the pathologist.
- The reading pathologist selects a patient case from a selected worklist within HALO AP Dx whereby the subject device fetches the associated images from external image storage.
- The reading pathologist uses the subject device to view the images and can perform the following actions, as needed:
a. Zoom and pan the image.
b. Measure distances and areas in the image.
c. Annotate images.
d. View multiple images side by side in a synchronized fashion.
The above steps are repeated as necessary.
After viewing all images belonging to a particular case (patient), the pathologist will make a diagnosis which is documented in another system, such as a Laboratory Information System (LIS).
The interoperable components of HALO AP Dx are provided in table 1 below:
Table 1. Interoperable Components for Use with HALO AP Dx
Components | Manufacturer | Model |
---|---|---|
Scanner | Hamamatsu | NanoZoomer S360MD Slide scanner |
Display | JVC | JD-C240BN01A |
This FDA 510(k) clearance letter pertains to HALO AP Dx, a software-only device for digital pathology image review. The documentation indicates that the device has been deemed substantially equivalent to a predicate device, the Hamamatsu NanoZoomer S360MD Slide scanner system (K213883).
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The provided document defines the performance data points primarily through "Performance Data" and "Summary of Studies" sections, focusing on comparisons to the predicate device and usability. There are no explicit quantitative "acceptance criteria" presented as specific thresholds, but rather statements of adequacy and similarity to the predicate.
Acceptance Criterion (Implicit) | Reported Device Performance |
---|---|
Image Reproduction Quality (Color Accuracy) | Criteria: Identical image reproduction compared to the predicate device (NZViewMD viewer), specifically regarding pixel-wise color accuracy. |
Performance: Pixel-level comparisons demonstrated that the 95th percentile CIEDE2000 values across all Regions of Interest (ROIs) from varied tissue types and diagnoses were less than 3 ΔE00. This was determined to be "identical image reproduction." | |
Turnaround Time (Image Loading - Case Selection) | Criteria: "When selecting a case, it should not take longer than 4 seconds until the image is fully loaded." |
Performance: Determined to be "adequate for the intended use of the subject device." (No specific value reported, but implies |
Ask a specific question about this device
(29 days)
HALO is a notification only cloud-based image processing software artificial intelligence algorithms to analyze patient imaging data in parallel to the standard of care imaging interpretation. Its intended use is to identify suggestive imaging patterns of a pre-specified condition and to directly notify an appropriate medical specialist.
HALO's indication is to facilitate the evaluation of the brain vasculature on patients suspected of stroke by processing and analyzing CT angiograms of the brain acquired in an acute setting. After completion of the data analysis, HALO sends a notification if a pattern suggestive for a suspected intracranial Large Vessel Occlusion (LVO) of the anterior circulation (ICA, M1 or M2) has been identified in an image.
The intended users of HALO are defined as medical specialists or a team of specialists that are involved in the diagnosis and care of stroke patients at emergency department where stroke patients are administered. The include physicians such as neurologists, radiologists, and/or other emergency department physicians.
HALO's output should not be used for primary diagnosis or clinical decisions; the final diagnosis is always decided upon by the medical specialist. HALO is indicated for CT scanners from GE Healthcare and Philips.
HALO is a notification only, cloud-based clinical support tool which identifies image features and communicates the analysis results to a specialist in parallel to the standard of care workflow.
HALO is designed to process CT angiograms of the brain and facilitate evaluation of these images using artificial intelligence to detect patterns suggestive of an intracranial large vessel occlusion (LVO) of the anterior circulation.
A copy of the original CTA images is sent to HALO cloud servers for automatic image processing. After analyzing the images, HALO sends a notification regarding a suspected finding to a specialist, recommending review of these images. The specialist can review the results remotely in a compatible DICOM web viewer.
Here's a detailed breakdown of the acceptance criteria and study proving the device meets them, based on the provided FDA 510(k) summary for HALO:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria | Reported Device Performance |
---|---|
Primary Endpoints: | |
LVO Detection Sensitivity | 91.3% (95% CI, 86.6%-94.8%) |
LVO Detection Specificity | 85.9% (95% CI, 80.6%-90.2%) |
Area Under the Curve (AUC) for LVO Detection | 0.97 |
Secondary Endpoints: | |
Median Notification Time for Detected LVOs | 4 minutes 29 seconds (minimum 3:47, maximum 7:12) |
The document states that "The HALO performance with regard to sensitivity and specificity, and the notification time are both equivalent to that of the selected predicate device." This implies that the reported performance metrics met or exceeded the established criteria for substantial equivalence to the predicate.
2. Sample Size and Data Provenance
- Test Set Sample Size: 427 patients after exclusions (originally 434 CTA scans).
- Data Provenance: Retrospective, multi-center clinical study. Patients were admitted to US comprehensive stroke centers.
3. Number and Qualifications of Experts for Ground Truth
- Number of Experts: 3 neuro radiologists.
- Qualifications: "Expert panel consisting of 3 neuro radiologists." Specific details on years of experience or board certification are not provided in this document.
4. Adjudication Method for the Test Set
The document states: "Ground truth was established by an expert panel consisting of 3 neuro radiologists." While it doesn't explicitly detail the adjudication method (e.g., 2+1, 3+1, consensus discussion), the wording suggests a consensus-based approach among the three experts. "Established by" implies a final, agreed-upon determination, not individual readings.
5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study
No MRMC comparative effectiveness study involving human readers with vs. without AI assistance is mentioned in the provided text for this specific device clearance. The study described focuses on the standalone performance of the AI algorithm.
6. Standalone (Algorithm Only) Performance
Yes, a standalone performance study was done. The reported sensitivity, specificity, and AUC are all metrics of the algorithm's performance without human intervention in the diagnosis and notification process. The intended use of HALO is to "directly notify an appropriate medical specialist" if a suspected finding is identified, running "in parallel to the standard of care imaging interpretation." This means its function is to flag cases for specialist review, not to replace it.
7. Type of Ground Truth Used
The ground truth used was expert consensus among three neuro radiologists, based on their interpretation of the CTA scans.
8. Sample Size for the Training Set
The document does not specify the sample size used for the training set. It only mentions the test set of 427 patients. It alludes to the algorithm using "a database of images" for its AI model but provides no numbers for this database's size or composition regarding training.
9. How Ground Truth for the Training Set Was Established
The document does not explicitly state how the ground truth for the training set was established. It only details the ground truth establishment for the test set. It is common practice for training data ground truth to be established through expert labeling or other robust methods, but this information is not provided here.
Ask a specific question about this device
(233 days)
HALO is a notification only cloud-based image processing software artificial intelligence algorithms to analyze patient imaging data in parallel to the standard of care imaging interpretation. Its intended use is to identify suggestive imaging patterns of a pre-specified clinical condition and to directly notify an appropriate medical specialist.
HALO's indication is to facilitate the evaluation of the brain vasculature on patients suspected of stroke by processing and analyzing contrast enhanced CT angiograms of the brain acquired in an acute setting. After completion of the data analysis. HALO sends a notification if a pattern suggestive for a suspected intracranial Large Vessel Occlusion (LVO) of the anterior circulation (ICA, M1 or M2) has been identified in an image.
The intended users of HALO are defined as appropriate medical specialists that are involved in the diagnosis and care of stroke patients at emergency department where stroke patients are administered. They include physicians such as neurologists, and/or other emergency department physicians.
HALO's output should not be used for primary diagnosis or clinical decisions; the final diagnosis is always decided upon by the medical specialist. HALO is indicated for CT scanners from GE Healthcare.
HALO is a notification only, cloud-based clinical support tool which identifies image features and communicates the analysis results to a specialist in parallel to the standard of care workflow.
HALO is designed to process CT angiograms of the brain and facilitate evaluation of these images using artificial intelligence to detect patterns suggestive of an intracranial large vessel occlusion (LVO) of the anterior circulation.
A copy of the original CTA images is sent to HALO cloud servers for automatic image processing. After analyzing the images, HALO sends a notification regarding a suspected finding to a specialist, recommending review of these images. The specialist can review the results remotely in a compatible DICOM web viewer.
Here's a summary of the acceptance criteria and study details for the HALO device, based on the provided FDA 510(k) summary:
HALO Device Acceptance Criteria and Study Details
1. Table of Acceptance Criteria and Reported Device Performance
Metric | Acceptance Criteria (Implicit) | Reported Device Performance (HALO) |
---|---|---|
Sensitivity | Sufficiently high for LVO detection (comparable to predicate) | 91.1% (95% CI, 86.0%-94.8%) |
Specificity | Sufficiently high for LVO detection (comparable to predicate) | 87.0% (95% CI, 81.2%-91.5%) |
AUC | High (indicative of good discriminative power) | 0.97 |
Notification Time | Fast enough for acute stroke setting (comparable to predicate) | Median: 4 minutes 31 seconds. Range: 3:47 to 7:12 |
Substantial Equivalence | Equivalent to predicate device ContaCT in terms of indications for use, technological characteristics, and safety and effectiveness. | Concluded to be substantially equivalent. |
2. Sample Size and Data Provenance for Test Set
- Sample Size: 348 CTA scans were initially collected, with 364 patients included for further analysis after exclusion. It's unclear if the "348 CTA scans" and "364 patients" refer to the same dataset or if some patients had multiple scans or if there was an expansion of the dataset. Assuming 364 cases (patients with at least one scan) were used.
- Data Provenance: Retrospective evaluation in a consecutive patient cohort. Data was collected from US comprehensive stroke centers.
3. Number and Qualifications of Experts for Ground Truth
- Number of Experts: 3 neuro radiologists.
- Qualifications: "Neuro radiologists" implies specialized training and experience in interpreting neurological imaging, which is appropriate for stroke diagnosis. Specific years of experience are not mentioned.
4. Adjudication Method for Test Set
The adjudication method is not explicitly stated. It says "Ground truth was established by an expert panel consisting of 3 neuro radiologists," which suggests a consensus-based approach, but the specific rule (e.g., majority vote, unanimous agreement, review by a lead expert if disagreement) is not provided.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not explicitly mentioned or conducted as detailed in the summary. The study focused on the standalone performance of the HALO algorithm.
6. Standalone Performance (Algorithm Only)
Yes, a standalone study was performed. The clinical study retrospectively evaluated the performance of the HALO clinical decision support algorithm for LVO detection using the collected CTA scans. The reported sensitivity, specificity, AUC, and notification time are all measures of the algorithm's standalone performance.
7. Type of Ground Truth Used
Expert Consensus. The ground truth for the test set was established by an expert panel consisting of 3 neuro radiologists.
8. Sample Size for Training Set
The sample size for the training set is not explicitly mentioned in the provided document. The document only covers the evaluation of the algorithm.
9. How Ground Truth for Training Set was Established
How the ground truth was established for the training set is not explicitly mentioned in the provided document. ("database of images" is stated for the core algorithm, but not how ground truth was applied to them).
Ask a specific question about this device
(200 days)
The Halo Ambulatory Infusion System is intended to deliver medications and/or fluids to a patient under the direction or supervision of physician or other certified healthcare professional. The device is intended for subcutaneous, perineural, epidural and intravenous infusion. The device is intended for 240 hours or 1500mL of infusion, whichever limit is reached first.
The Halo Ambulatory Infusion System includes Halo Ambulatory Infusion Pump ("Halo Pump'') and Halo Administration Set. The system provides measured drug / fluid therapy for subcutaneous, perineural, evidural and intravenous delivery to patient under the direction or supervision of physician or other certified health care professional. The design philosophy was "Simplicity = Safety". Verification and validation data have demonstrated that this is a simple, safe and reliable infusion device for patient in the clinical and non-clinical environments, including but not limited to hospital, homecare, and outpatient settings.
Halo Ambulatory Infusion Pump: The Halo Ambulatory Infusion Pump ("Halo Pump") has a microprocessor controlled motor that drives a mechanical pumping mechanism to deliver fluid at a controlled rate, same as the CADD-Legacy PCA 6300 Ambulatory Infusion Pump (K982839) and ambIT Continuous Ambulatory Infusion Pump (K033325). The Halo Pump interfaces with user via keypad, LED and LCD. Infusion parameters can be set up by operating the keypad, and displayed on the LCD. The pump includes sensors for detection of upstream occlusion, downstream occlusion, cassette loading error, and pumping mechanism malfunction. Error conditions can be displayed on LCD, and redundantly indicated by a LED light and audio alert. The electronics is powered by a battery. The materials of construction for Halo Pump components are widely used in the medical industry. The Halo Pump is intended for therapies that require a continuous infusion with optional Patient Controlled Analgesia (PCA).
Halo Administration Set: The Halo Administration Set is designed to administer fluids / medication from a container to a patient through a needle or a catheter. Major components include spike, cassette, slide clamp, filter, and luer lock. The air-eliminating filter prevents air from entering into a patient. The Halo Administration Set can be used with Halo Pump or for gravity infusion. It is similar to the predicate device, Zyno Medical Administration Sets(K120685), which can be used with Z-800 Infusion Pump or gravity infusion.
This document is a 510(k) summary for the Halo Ambulatory Infusion System, encompassing the Halo Ambulatory Infusion Pump and the Halo Administration Set. It seeks to demonstrate substantial equivalence to previously cleared predicate devices.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are generally established for each test based on relevant standards (ISO, AAMI, ASTM, IEC) and FDA guidance documents. The document states that "All testing met pre-established specifications," and provides a summary table of the verification and validation tests performed.
Testing Performed | Acceptance Criteria Source (Implicit) | Reported Device Performance |
---|---|---|
Halo Administration Set testing (Biocompatibility, Pyrogenicity, Physical & Mechanical, Chemical, Sterility, Aging, DEHP, Integration with Halo Pump) | Reference B (Intravascular Administration Sets Premarket Notification Submissions), ISO / AAMI / ASTM standards | Pass |
Halo Pump software verification and validation testing | Reference E (Software Contained in Medical Device) | Pass |
Halo Ambulatory Infusion System basic safety and essential performance testing | IEC 60601-1 and 60601-2-24 | Pass |
Halo Pump Electromagnetic Compatibility Testing | IEC 60606-1-2 | Pass |
Halo Ambulatory Infusion System Alarm System Testing | IEC 60601-1-8 | Pass |
Verification and Validation of Halo Ambulatory Infusion System used in the home healthcare environment | Reference C (Design Considerations for Devices Intended for Home Use), IEC 60601-1-11 | Pass |
Additional Testing for Verification of Safety and Effectiveness | Reference A (Infusion Pump Premarket Notification Submissions), design requirements, hazard analysis | Pass |
Usability Validation Testing of Halo Ambulatory Infusion System | References A, B, C, D (Human Factors and Usability Engineering), E; IEC 6060-1-6; IEC 62366 | Pass |
Volumetric Accuracy (Halo Ambulatory Infusion Pump) | Predicate Devices (CADD-Legacy, ambIT, CADD-Solis VIP) all have +/- 6% | +/- 5% |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample sizes used for each specific test (e.g., number of infusion sets tested for biocompatibility, number of pumps for volumetric accuracy). It only mentions that "Zyno Medical performed extensive verification and validation testing."
The provenance of the data is implicit: it was generated through the manufacturer's (Zyno Medical, LLC, Natick, MA, USA) own verification and validation testing, which is generally conducted under controlled, prospective conditions for a new device submission. There is no mention of country of origin for the data or whether it was retrospective or prospective in the context of clinical studies, as clinical studies were "deemed unnecessary."
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not provided in the document. Given that the testing was primarily lab-based verification and validation against technical standards and safety requirements, clinical "experts" in the traditional sense for establishing ground truth on patient data were not involved as "Human clinical studies were deemed unnecessary." The "ground truth" for these tests would be the established performance specifications and regulatory standards.
4. Adjudication Method for the Test Set
This information is not applicable and not provided. Adjudication methods (like 2+1 or 3+1) are typically used in clinical studies involving interpretation of medical images or patient outcomes by multiple readers, which was not performed here.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The document explicitly states: "Human clinical studies were deemed unnecessary to evaluate the safety or effectiveness of the Halo Ambulatory Infusion System." Therefore, there is no information on the effect size of human readers improving with or without AI assistance, as AI is not mentioned as a component.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
This question is not applicable as the device is a physical medical device (infusion pump and administration set), not an algorithm or AI system. Its performance is inherent to its mechanical and electrical function.
7. The Type of Ground Truth Used
The "ground truth" for the device's performance was established via:
- Technical Specifications and Regulatory Standards: Compliance with various international standards (ISO, AAMI, ASTM, IEC) and FDA guidance documents related to infusion pumps and administration sets.
- Predicate Device Performance: The "Equivalency Matrix" tables (Table 1 and 2) compare the Halo system's features and performance parameters (e.g., volumetric accuracy) against legally marketed predicate devices, implying that the established performance of these predicates served as a benchmark for equivalence.
- Design Requirements and Hazard Analysis: The testing included verification against internal design requirements and findings from hazard analysis.
8. The Sample Size for the Training Set
This question is not applicable. The Halo Ambulatory Infusion System is a conventional medical device (infusion pump), not a machine learning or AI system that requires a "training set."
9. How the Ground Truth for the Training Set Was Established
This question is not applicable as there is no training set for this type of device.
Ask a specific question about this device
(27 days)
"Catalyst" is a diagnostic ultrasound system designed to be used for investigating disorders of the pelvic floor. An ultrasonographic crystal within the probe records images of the organ, muscle, and tissue structures of the pelvic region. MidCRYSTL and HALO probes allow for ultrasonography of the following: 1) on the surface of the perineum and/or abdomen, 2) endocavity, by inserting the endovaginal probe into the vagina, and 3) endocavity, by inserting the endoanal probe into the anal canal.
The Halo Medical Technologies, Catalyst™, MidCRYSTL™, HALO™ Ultrasound System is a self contained portable, single-mode, and singleapplication ultrasound imaging system. The system contains an ultrasound generator/receiver, analog to digital converter, microcontroller, control logic, USB 2.0 interface and control offering a full complement of conventional operating modes, software-based parameter controls, and video recording.
User-customized parameter settings for the MidCRYSTL™ Endoanal, HALO™ Endovaginal and HALO™ Surface Ultrasound Probes may be inserted by the operator and stored for recall as needed via the system control panel. Customization includes transmit power, images controls selection, and Time Gain Compensation (TGC). Controls are also provided to select display format and to utilize the cine function.
More detailed explanations of these functions and controls are included in the User Manuals, and in the software/firmware documentation included in this 510(k) Notification. Patient contact materials have been used in accordance to their intended use and are described below for each individual transducer. The transducers were previously cleared for use on other Systems (K951976 and referenced in K070907).
The Halo Medical Technologies Catalyst™, MidCRYSTL™, HALO™ Ultrasound System is a B-Mode ultrasound scanner which provides high resolution, high penetration performance. Probes are supported in frequencies from 3.5 MHz to 24.0 MHz. The probes are indicated expressly for endoanal, endovaginal, and/or abdominal/transperineal (surface) application.
The Halo Medical Technologies Catalyst™, MidCRYSTL™, HALO™ Ultrasound System provides various measuring functions. It can measure distances and calculate angles. Halo Medical Technologies Catalyst™, MidCRYSTL™, HALO™ Ultrasound System supports the Cine function (capable of storing up to 512 sequential images). Management of patient history is possible by image-storage function. High-resolution images are provided by utilizing a technology called digital dynamic receive focusing. The same clinical uses were cleared for the predicate device(s), Laborie NuWAV (K081781).
The provided 510(k) summary for the Halo Medical Technologies, Catalyst™, MidCRYSTL™, HALO™ Ultrasound System focuses on demonstrating substantial equivalence to a predicate device through nonclinical testing and comparison of technological characteristics. It explicitly states that clinical studies were NOT required. Therefore, the information requested for acceptance criteria and a study proving the device meets criteria, particularly those related to clinical performance, a test set, ground truth, and expert evaluation, is not present in the given document.
However, I can extract the nonclinical acceptance criteria and the methods used to prove the device meets these criteria.
1. A table of acceptance criteria and the reported device performance
Since this is a 510(k) summary demonstrating substantial equivalence for an ultrasound system, the "acceptance criteria" are primarily that the new device's performance in specific nonclinical tests (related to safety and efficacy) is comparable to established standards and the predicate device. The "reported device performance" is implicitly that it met these standards and was found to be comparable.
Acceptance Criteria (Nonclinical Standards/Requirements) | Reported Device Performance (Implicitly Met) |
---|---|
Acoustic Output Measurement (in accordance with AIUM/NEMA UD-2 and UD-3) | Compliant; Performance comparable to predicate (ISPTA.3 94 mW/cm2, MI 1.9 (Maximum)) |
Electrical Safety (IEC 60601-1) | Compliant |
Electromagnetic Compatibility (IEC/EN 60601-1-2) | Compliant |
Ultrasound Equipment Safety (IEC 60601-2-37) - includes Thermal Safety Validation | Compliant |
Biological Safety (ISO 10993-1 Compliant, Selection of Tests) | Compliant |
Cytotoxicity Study (ISO 10993-5 Compliant, using ISO Elution Method) | Compliant |
Vaginal Irritation Study (ISO 10993-10 Compliant) | Compliant |
Maximum Sensitization Study - Extract (ISO 10993-10 Compliant) | Compliant |
Sample Preparation (ISO 10993-12 Compliant) | Compliant |
Technological Characteristics (Materials, Display Mode, Measurements, Principle of Operation, Transducer Probe Design, Acoustic Output Limits) | Comparable to predicate device (Laborie Medical Technologies, NuWav Ultrasound System (K081781)) and established safety limits |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not applicable as the submission explicitly states, "The subject of this premarket submission... did not require clinical studies to support substantial equivalence." Therefore, no clinical test set data from actual patients was used. The provenances are the nonclinical test labs that conducted the specified standards testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable as no clinical studies requiring expert ground truth were conducted.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable as no clinical studies requiring adjudication were conducted.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable as no clinical studies, including MRMC studies, were conducted. This device is an ultrasound imaging system, not an AI-powered diagnostic algorithm.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This information is not applicable as this is an ultrasound imaging system, not an algorithm, and no clinical performance studies were conducted.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the nonclinical tests:
- Acoustic Output: Measured against established engineering and physics standards (AIUM/NEMA UD-2 and UD-3).
- Electrical Safety, EMC, Ultrasound Equipment Safety: Tested against established safety standards (IEC 60601-1, IEC/EN 60601-1-2, IEC 60601-2-37).
- Biological Safety: Tested against established biocompatibility standards (ISO 10993 series).
- Technological Characteristics: Direct comparison to the specifications of the legally marketed predicate device (Laborie NuWav, K081781).
8. The sample size for the training set
This information is not applicable as no machine learning algorithm requiring a training set is mentioned or implied for this device's 510(k) submission.
9. How the ground truth for the training set was established
This information is not applicable as no machine learning algorithm requiring a training set is mentioned or implied, and thus no training set ground truth was established.
Ask a specific question about this device
(80 days)
The HALO® ULTRA Ablation Catheter (used with the HALOEEK Energy Generator, model 1190A-115A) is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, Angiodysplasia, Gastric Antral Vascular Ectasia (GAVE) and Radiation Proctitis (RP).
The subject device, HALO® ULTRA Ablation Catheter (hereafter referred to as "modified ULTRA") is a sterile single-use bipolar device that delivers radiofrequency (RF) energy to the treatment tissue within the gastrointestinal tract through a copper electrode. It is used exclusively with HALOFES Energy Generator model 1190A-115A (cleared under K092487).
The provided text describes a 510(k) premarket notification for a medical device, the HALO90 ULTRA Ablation Catheter. It focuses on demonstrating substantial equivalence to predicate devices through technical characteristic comparisons and bench testing.
Therefore, the following information regarding acceptance criteria and a study proving the device meets these criteria cannot be extracted from the document, as it is a 510(k) submission, not a study report demonstrating clinical performance against specific acceptance criteria.
The 510(k) process is about demonstrating substantial equivalence to a legally marketed predicate device, not necessarily proving a device meets new performance acceptance criteria through clinical studies in the same way a PMA (Premarket Approval) would.
Here's why the requested information is absent:
- Acceptance Criteria & Reported Performance: The document doesn't define specific "acceptance criteria" for clinical performance (e.g., sensitivity, specificity, accuracy) that an AI device would typically have. It's comparing the technological characteristics and bench test results of the modified device to the predicate.
- Sample Size, Ground Truth, Adjudication, MRMC, Standalone Performance, Training Set: These are all concepts relevant to studies evaluating the clinical performance of a diagnostic or AI-driven device, particularly when assessing its ability to detect or diagnose conditions. This document is for an ablation catheter, which treats conditions, and its filing is based on demonstrating safety and efficacy through equivalence to existing devices, supported by bench testing of physical characteristics.
Specifically addressed, based on the provided text:
- A table of acceptance criteria and the reported device performance: Not applicable. The submission is for substantial equivalence based on technological characteristics and bench testing (migration, deflection, catheter distal integrity, detachment). No clinical performance acceptance criteria are mentioned.
- Sample sized used for the test set and the data provenance: Not applicable. The "test set" refers to bench testing, not a clinical data set. The document lists the specific bench tests performed without providing sample sizes as would be relevant for clinical data, nor does it discuss data provenance.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth, in the context of clinical studies, refers to definitive diagnoses, often established by expert consensus or pathology. This document describes bench testing where "ground truth" would be engineering specifications or physical measurements, not expert clinical interpretation.
- Adjudication method for the test set: Not applicable. Adjudication is used to resolve discrepancies in expert interpretation of clinical data. This is a bench test submission.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is for an ablation catheter, not an AI diagnostic tool involving human readers.
- If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This is for a physical medical device (catheter), not an algorithm.
- The type of ground truth used: For the bench testing, the "ground truth" would be the expected physical properties and performance metrics of the catheter as measured against engineering standards. This is not explicitly detailed but is implied by the nature of bench tests like "Migration," "Deflection," "Catheter Distal Integrity," and "Detachment."
- The sample size for the training set: Not applicable. No training set is mentioned as this is not an AI/algorithm-based device.
- How the ground truth for the training set was established: Not applicable. See point 8.
Summary of Device and Evidence Presented in the 510(k):
The HALO90 ULTRA Ablation Catheter is a modification of an existing device (original ULTRA). The submission aims to prove substantial equivalence to its predicate devices, not to establish new performance criteria through a de novo clinical study.
Evidence presented for substantial equivalence:
- Technological Characteristics: The modified ULTRA has the "same construction, principles of operation, materials and energy density" as the original ULTRA. The differences are described as "a slight change in the manufacturing process of the endoscope mounting strap, as well as dimensional modification on the pivot mechanism components."
- Bench Testing: The minor differences were evaluated via the following bench tests to ensure no new questions of safety and effectiveness were raised:
- Migration
- Deflection
- Catheter Distal Integrity
- Detachment
Ask a specific question about this device
(133 days)
The HALO® Ablation Catheter (used with the HALOFER Energy Generator, model 1190A-115A) is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, Angiodysplasia, Gastric Antral Vascular Ectasia (GAVE) and Radiation Proctitis (RP).
The HALO® (subject device) is a single-use bipolar device that delivers radiofrequency energy to the treatment tissue within the gastrointestinal tract. It is used solely with the HALOFES Energy Generator (model 1190A-115A), which provides the radiofrequency energy.
The HALO60 Ablation Catheter is a modification of the HALO90 Ablation Catheter, primarily involving a reduction in the copper electrode surface area. The submission is a Special 510(k) as the modification does not change the intended use or alter the fundamental scientific technology.
Here's an analysis of the provided information regarding acceptance criteria and the study:
1. Table of Acceptance Criteria and Reported Device Performance
The provided document does not explicitly state specific quantitative acceptance criteria or numerical performance metrics for the HALO60 Ablation Catheter. Instead, it focuses on demonstrating substantial equivalence to a predicate device (HALO90 Ablation Catheter) based on the absence of new questions of safety or effectiveness.
The "performance" is implicitly demonstrated through the modifications and the conclusion of substantial equivalence.
Acceptance Criteria | Reported Device Performance |
---|---|
No new questions of safety or effectiveness raised by the modifications | The primary modification (dimensional reduction of copper electrode surface area from 2.6cm² to 1.5cm²) and secondary modifications (dimensional reduction of supporting components) do not raise new questions of safety or effectiveness. |
Same intended use as predicate device | The HALO60 Ablation Catheter has the same intended use as the HALO90 Ablation Catheter. |
Same fundamental scientific technology as predicate device | The HALO60 Ablation Catheter has the same fundamental scientific technology as the HALO90 Ablation Catheter. |
Compatible with HALOFER Energy Generator | The HALO60 Ablation Catheter is compatible with the HALOFER Energy Generator. No changes were made to the generator's software for HALO60. |
Materials remained unchanged | The materials used in the HALO60 Ablation Catheter are unchanged from the predicate device. |
Amount of energy delivered remained unchanged | The amount of energy delivered by the HALO60 Ablation Catheter is unchanged from the predicate device. |
Depth of tissue treated remained unchanged | The depth of tissue treated by the HALO60 Ablation Catheter is unchanged from the predicate device. |
Design verification activities demonstrate meeting design input requirements | Design verification activities were conducted to demonstrate that the design outputs of the modified device meet the design input requirements. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not describe a specific clinical "test set" with a sample size in the traditional sense of evaluating a new device's performance against clinical endpoints. The submission relies on design verification activities and a comparison to the predicate device's established safety and effectiveness.
- Sample Size for Test Set: Not applicable as a clinical test set for performance evaluation is not described.
- Data Provenance: Not applicable as a clinical test set for performance evaluation is not described. The document pertains to regulatory submission data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable to this submission. The document focuses on engineering and design verification rather than clinical outcomes requiring expert consensus for ground truth.
4. Adjudication Method for the Test Set
This information is not applicable to this submission as a clinical test set requiring adjudication is not described.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable. The HALO60 Ablation Catheter is a medical device for radiofrequency energy delivery, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study related to AI assistance would not be relevant.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
This information is not applicable. The HALO60 Ablation Catheter is a physical medical device, not an algorithm, and operates with human intervention during its use.
7. The Type of Ground Truth Used
The "ground truth" in this context is established through design verification activities against design input requirements, and based on the established safety and effectiveness of the predicate device (HALO90). The document highlights:
- Design Input Requirements: The modified device's design outputs were verified to meet these.
- Predicate Device's Established Performance: The HALO90's prior clearances (K093008, K083737, K062723) serve as the basis for demonstrating substantial equivalence.
- Failure Modes, Effects, and Criticality Analysis (FMECA): Used to analyze risks, implying that safety standards and risk mitigation served as a form of "ground truth" for design decisions.
8. The Sample Size for the Training Set
This information is not applicable. The HALO60 Ablation Catheter is a physical medical device; there is no "training set" of data in the context of machine learning or AI.
9. How the Ground Truth for the Training Set was Established
This information is not applicable for the reasons stated in point 8.
Ask a specific question about this device
(58 days)
The HALO90 ULTRA Ablation Catheter model 90-9200 (when used with HALOFEN System) is indicated for use in the coagulation of bleeding and nonbleeding sites in the gastrointestinal tract including but not limited to the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, Angiodysplasia, Gastric Antral Vascular Ectasia (GAVE) and Radiation Proctitis (RP).
The HALO® ULTRA Ablation Catheter model 90-9200 is a modification . of the predicate device HALO" Ablation Catheter model 90-9100 currently commercialized in USA. Both catheters have the same constructions, design, principle of operation, materials and energy density. The differences between HALO" ULTRA and its predicate consist in the electrode surface increase and the associated components electrode cap and base. HALO90 ULTRA Ablation Catheter model 90-9200 is used in conjunction with HALOFER Energy Generator There are no changes associated to the HALOFLEX Energy Generator software, hardware and accessories.
The provided text describes a 510(k) premarket notification for the HALO® ULTRA Ablation Catheter Model 90-9200. This is a submission demonstrating substantial equivalence to a predicate device, not a study designed to establish new acceptance criteria or definitively prove performance against specific clinical endpoints. Therefore, much of the requested information (e.g., acceptance criteria table, sample sizes, expert qualifications, MRMC studies, ground truth establishment) is not typically part of a 510(k) submission focused on minor modifications to an already cleared device.
However, based on the information provided, here's a breakdown of what can be extracted:
1. Table of Acceptance Criteria and Reported Device Performance
The submission does not define specific performance acceptance criteria in the typical sense of a clinical study, as it's a modification of a predicate device. Instead, the "acceptance criteria" are implied by the successful completion of a series of bench tests showcasing that the changes did not negatively impact safety and efficacy compared to the predicate. The "reported device performance" is that these tests were passed successfully, indicating equivalence.
Acceptance Criterion (Implied) | Reported Device Performance |
---|---|
Maintain Apposition | HALO® ULTRA Ablation Catheter Model 90-9200 Design Verification - Apposition (Passed) |
Tensile Strength (Base-cap & Litz wires) | Tensile strengths of Base-cap & Litz wires (Passed) |
Attachment Strength (Endoscope-elastomeric strap) | Attachment strength endoscope-elastomeric strap (Passed) |
Endoscope Compatibility | Endoscope compatibility with the HALO® ULTRA Ablation Catheter Model 90-9200 (Passed) |
Insertion-Retraction Performance | Insertion-Retraction performance in a tissue model (Passed) |
2. Sample Sizes Used for the Test Set and Data Provenance
The document mentions "Bench Tests." These are typically laboratory-based tests and do not involve human subjects. Therefore, the concept of a "test set" in the clinical sense with human subjects, data provenance (country of origin, retrospective/prospective), and sample sizes for clinical data are not applicable or provided here. For bench tests, sample sizes would typically refer to the number of devices or components tested, but this specific detail is not given.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
Not applicable. Bench tests do not typically involve human expert "ground truth" establishment in the way clinical diagnostic studies do. The "ground truth" for bench tests is defined by engineering specifications and objective measurements.
4. Adjudication Method
Not applicable, as there is no clinical data or expert review process described that would require adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. This type of study evaluates human reader performance, usually in diagnostic imaging, and is not relevant to a bench testing submission for an electrosurgical ablation catheter.
6. Standalone Performance Study
Yes, in a way. The "Bench Tests" described represent a standalone evaluation of the modified device's physical and functional characteristics. However, this is not a standalone clinical performance study, but rather a standalone engineering/design verification study. The tests evaluate the device itself without human-in-the-loop performance being the primary focus.
7. Type of Ground Truth Used
For the bench tests, the "ground truth" would be established by:
- Engineering specifications and design requirements: For tests like tensile strength, apposition, and insertion-retraction, the "truth" is whether the device meets pre-defined, measurable specifications.
- Predicate device performance: The ultimate "ground truth" for demonstrating substantial equivalence is that the modified device performs similarly to or better than the legally marketed predicate device on the critical parameters.
8. Sample Size for the Training Set
Not applicable. This is a 510(k) submission for a physical medical device, not an AI/ML algorithm. There is no concept of a "training set" in this context. The manufacturing of the device would follow established quality control processes, but that's different from an algorithm's training data.
9. How the Ground Truth for the Training Set Was Established
Not applicable for the same reasons as point 8.
Ask a specific question about this device
(30 days)
HALO360 + Sizing Balloon model 3441C is indicated for the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to, the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, and Angiodysplasia.
The HALO360+ Sizing Balloon model 3441C is used in conjunction with either the HALO360 Energy Generator models 1100C-115B (or 1100C-230B), or HALORLEX Energy Generator model 1190A-115A (or 1190A-230A) for assessing the size of the esophageal lumen, and facilitate the selection of the disposable single-use HALO360+ Coagulation Catheter. The HALO360+ Sizing Balloon model 3441C, like the predicate device HALO360 Sizing Balloon model 3441B is comprised of a sizing balloon, a catheter shaft with markings, and an electrical connector.
Here's an analysis of the acceptance criteria and study information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria | Reported Device Performance |
---|---|
Sizing accuracy | Met the same specifications as predicate devices. |
Structural strength | Met the same specifications as predicate devices. |
Material compatibility | Met the same specifications as predicate devices. |
Sterility | Met the same specifications as predicate devices. |
Esophageal diameter measurement capability | Increased to measure diameters higher than 33.7 mm (allowing identification of migration into the stomach). |
Note: The document explicitly states the modified device "met the same specifications requirements as the HALO360 Sizing Balloon and HALO360+ Ablation Catheter" for the first four criteria. The fifth criterion is a noted improvement in capability.
2. Sample Size Used for the Test Set and Data Provenance
The provided text does not contain information regarding the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective nature of the data). The information provided focuses on the device's technical specifications and a comparison to predicate devices, rather than a clinical study.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
The provided text does not contain information about the number of experts used or their qualifications to establish ground truth. As this appears to be a submission based on engineering and performance criteria for substantial equivalence, clinical expert assessment for ground truth is not detailed.
4. Adjudication Method
The provided text does not contain information about any adjudication method.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done or at least not mentioned in the provided text. The submission focuses on demonstrating substantial equivalence based on material and performance specifications, not on comparing human reader performance with or without AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
This question is not applicable. The HALO360+ Sizing Balloon is a medical device, not an AI algorithm. Therefore, a standalone performance study in the context of AI is not relevant. The device itself is designed to be used by a human operator (a physician) to size the esophagus and select a coagulation catheter.
7. The Type of Ground Truth Used
The "ground truth" for the device's performance appears to be established through engineering specifications and testing based on the predicate devices. For example, "sizing accuracy" would be compared against a known, precise measurement standard in a laboratory or simulated environment, rather than a clinical ground truth like pathology or outcomes data.
8. The Sample Size for the Training Set
The provided text does not contain information about a training set since this is not an AI/machine learning device. The testing described appears to be for device validation against established specifications.
9. How the Ground Truth for the Training Set was Established
This question is not applicable as there is no mention of a training set for an AI algorithm.
Ask a specific question about this device
(102 days)
The HALO® System (including HALO® Ablation Catheter model 90-9100 and HALO® Energy Generator model 90-9000) is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's esophagus, Dieulafoy Lesions, Angiodysplasia, Gastric Antral Vascular Ectasia (GAVE) and Radiation Proctitis (RP).
The HALO® Ablation Catheter model 90-9100 operates in conjunction with HALO® Energy Generator model 90-9000. There are no changes to the HALO® Ablation Catheter or HALO® Energy Generator implemented since the devices were cleared by K083737, K062723 and K062441. The product is also identical in principle of operation and energy density to the treatment site with the predicate devices HALO360 Ablation Catheter, Stellartech Coagulation Catheter. HALO® is substantially equivalent in construction with the predicate devices HALO360 Ablation Catheter, Stellartech Coagulation Catheter and Microvasive Gold Probe. The product is similar in performance with Olympus Heat Probe.
The provided 510(k) summary for BARRX's HALO System (K093008) indicates that no new performance testing was conducted for this submission.
The submission focuses on expanding the indications for use for the HALO® Ablation Catheter to include Gastric Antral Vascular Ectasia (GAVE) and Radiation Proctitis (RP) based on the substantial equivalence to previously cleared devices and existing clinical data.
Therefore, the following information cannot be extracted from the provided text as the study described is a review of existing clinical data, not a new performance study with acceptance criteria.
- A table of acceptance criteria and the reported device performance: Not applicable, as no new performance testing was conducted.
- Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable for new performance testing. The clinical data supporting the expanded indications was "derived from publications and case studies." The provenance details are not specified.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable, as no new test set was created with expert-established ground truth.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI/device assisting human readers.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc): For the expanded indications, the clinical data relied upon was from "publications and case studies" which demonstrated the safety and effectiveness of the device when used for GAVE and RP. This implies that outcomes data from these existing sources served as the basis for substantiating safety and effectiveness for these new indications.
- The sample size for the training set: Not applicable, as no new training set was explicitly mentioned for algorithmic development in this submission.
- How the ground truth for the training set was established: Not applicable.
Summary of the study that proves the device meets the acceptance criteria (expanded indications):
The submission argues for the substantial equivalence of the HALO® Ablation Catheter for the treatment of GAVE and RP by referencing:
- Technological Identity and Substantial Equivalence: The HALO® Ablation Catheter model 90-9100 is "technologically identical" to its cleared predicate (K083737, K062723). It also highlights substantial equivalence in principle of operation, energy density, construction, and performance with other predicate devices (HALO360 Ablation Catheter, Stellartech Coagulation Catheter, Microvasive Gold Probe, Olympus Heat Probe).
- Clinical Data Review: Clinical data was provided to FDA "to support the changes in the Instructions for Use associated with the use in the coagulation of bleeding sites for the HALO® Ablation Catheter for the treatment of GAVE and Radiation Proctitis." This data "demonstrated that, when used in accordance with those instructions, the HALO® Ablation Catheter used for the treatment of GAVE and Radiation Proctitis respectively is at least as safe and effective as the cleared HALO® Ablation Catheter for the treatment of bleeding sites in the gastrointestinal tract."
- Basis of Clinical Data: The clinical data supporting the additional indications (GAVE and RP) was "derived from publications and case studies" and facilitated "Physician's Instructions recommending specific treatment settings and selection criteria for the patients (Contraindications, Warnings and Cautions)."
- Similarity of Conditions: GAVE and RP are described as being similar to other already indicated bleeding conditions (esophageal ulcers, Mallory-Weiss tears, etc.) in that they are confined to the mucosal layer, associated with hemorrhage, and treated with coagulative therapy.
In essence, the "study" proving the device meets the acceptance criteria for expanded indications was a retrospective review of existing clinical data from publications and case studies, and an argument for substantial equivalence to previously cleared predicate devices for similar applications. No new pre-market performance study with defined acceptance criteria was conducted.
Ask a specific question about this device
Page 1 of 2