K Number
K092487

Validate with FDA (Live)

Date Cleared
2009-11-10

(89 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The HALOFLEX Energy Generator is indicated for use for the coagulation of soft tissue.

The HALOFLEX Energy Generator is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including, but not limited to, the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, and Angiodysplasia.

Device Description

The HALOFLEX Energy Generator is a modification to the HALO360 Energy Generator and HALO® Energy Generators that have already been cleared by the Food and Drug Administration ("FDA" or the "Agency") for use in coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to the esophagus (K051168 and K060169, K062441 respectively). The Generator has the same intended use and fundamental scientific technology as the cleared HALO360 and HALO90 Energy Generators. In addition, the intended use was expanded to include "coagulation of soft tissue", for this indication for use the predicate device is the RF 3000 Radiofrequency Generator (K000241).

The design changes between the HALO360 and HALO® Energy Generators and the HALOFLEX Energy Generator do not affect the clinical parameters employed in the energy delivery or the safety mechanisms that control those parameters.

The HALOFLEX Energy Generator has a limited number of design changes compared to the predicate devices. These changes are minor and do not raise no questions or effectiveness. Further, bench testing demonstrated comparable performance and safety. These modifications were as follows:

  • Add to the indications for use the "coagulation of soft tissue", to reflect the . general nature of the device.
  • Modification of the CPU board, for increased memory .
  • Consolidation of the Pressure Monitoring Board and RF MUX Board in the Hand Piece Interface Board
  • Integration on the HALO360 and HALO90 Energy Generators on the same . hardware platform as HALO360 Energy Generator
  • . Update the software codes to reflect the specific performance characteristics for each Energy Generator.
  • Optimization of components for reliability, and/or obsolescence
AI/ML Overview

Here's an analysis of the provided text regarding the HALOFLEX Energy Generator, addressing the specific points requested:

1. Table of Acceptance Criteria and Reported Device Performance

The provided 510(k) summary for the HALOFLEX Energy Generator does not explicitly define quantitative acceptance criteria or provide specific numerical device performance metrics in the way one might expect for a diagnostic or AI-driven device.

Instead, the submission relies on the concept of substantial equivalence to predicate devices. This means the acceptance criteria are implicitly that the HALOFLEX Energy Generator performs at least as safely and effectively as the predicate devices.

Acceptance Criterion (Implicit)Reported Device Performance
Safety equivalent to predicate devicesPerformance data demonstrated that, when used in accordance with the instructions, the HALOFLEX is at least as safe as the predicate devices (HALO360 and HALO90 Energy Generators).
Effectiveness equivalent to predicate devicesPerformance data demonstrated that, when used in accordance with the instructions, the HALOFLEX is at least as effective as the predicate devices (HALO360 and HALO90 Energy Generators).
Technological characteristics equivalent to predicate devicesSame intended use, fundamental scientific technology, and principles of operation as predicate devices. Minor design changes do not affect clinical parameters or safety mechanisms.
No new issues of safety or effectivenessMinor technological differences raise no new issues of safety or effectiveness.
Compliance with applicable international and domestic requirementsPerformance testing conducted in compliance with applicable international and domestic requirements and certifications for harmonized standards.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a "test set" in the context of a clinical study with a defined sample size. The performance data mentioned refers to "bench testing" and compliance with standards rather than a clinical trial where patients or specific datasets are evaluated.

Therefore:

  • Sample size for the test set: Not applicable/Not provided. The submission focuses on substantial equivalence through design and bench testing, not a clinical test set.
  • Data provenance: Not explicitly stated as retrospective or prospective clinical data. The "bench testing" implies laboratory-based data, not patient data from a specific country of origin.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This information is not applicable/not provided. The submission does not describe a process of establishing ground truth by medical experts for a test set, as it is not a diagnostic device requiring such evaluation. The safety and effectiveness are established through comparison with predicate devices and bench testing.

4. Adjudication Method for the Test Set

This information is not applicable/not provided. There is no described "test set" or clinical evaluation requiring adjudication by experts.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This type of study is typically conducted for diagnostic devices (e.g., imaging software) to evaluate interpretative performance by human readers with and without AI assistance. The HALOFLEX Energy Generator is a therapeutic device (electrosurgical coagulation system), and its evaluation focuses on safety, fundamental scientific technology, and equivalent performance to existing devices, not on improving human reader performance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The concept of "standalone performance" for an algorithm is not applicable to the HALOFLEX Energy Generator. This device is an energy generator for electrosurgical coagulation, which is inherently a human-in-the-loop therapeutic procedure. Its performance is evaluated on its ability to deliver energy safely and effectively, not as an autonomous algorithm. The "performance data" mentioned refers to bench testing to ensure the device's functional integrity.

7. The Type of Ground Truth Used

The concept of "ground truth" as typically applied to diagnostic algorithms (e.g., pathology, expert consensus on imaging) is not directly applicable here. The "ground truth" for affirming the device's safety and effectiveness relies on:

  • Bench testing results: Demonstrating compliance with engineering specifications and safety standards.
  • Comparison to predicate devices: The established safety and effectiveness of the existing HALO360, HALO90, and RF 3000 generators serve as a de facto "ground truth" for what is considered safe and effective performance for this class of device.
  • Regulatory standards: Compliance with international and domestic requirements acts as a standard for acceptable performance.

8. The Sample Size for the Training Set

This information is not applicable/not provided. The HALOFLEX Energy Generator is a hardware device with embedded software, not a machine learning model that undergoes "training" on a dataset. The software updates mentioned are for "specific performance characteristics" and "optimization of components," not for an AI algorithm trained on a dataset.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable/not provided for the same reasons as point 8.

{0}------------------------------------------------

K092487)

NOV 1 0 2009

510(k) SUMMARY

BÂRRX Medical Inc.'s HALORLEX Energy Generator

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

BARRX Medical Inc. 540 Oakmead Parkway Sunnyvale, CA 94085

Phone:408-328-7302
Facsimile:408-328-7395

Viorica (Vivi) Filimon Contact Person:

Date Prepared: August 12, 2009

Name of Device and Name/Address of Sponsor

HALOFLEX Energy Generator

BÂRRX Medical Inc. 540 Oakmead Parkway Sunnyvale, CA 94085

Common or Usual Name:

Electrosurgical Coagulation System

Classification Name

878.4400 Electrosurgical, cutting & coagulation & accessories, product code GEI

Predicate Devices

HALO360 Energy Generator model 1100C-115B

HALO® Energy Generator model 1100C-115C

HALO90 Energy Generator model 90-9000

RF 3000 Radiofrequency Generator

BÂRRX Medical Inc.

BARRX Medical Inc.

BARRX Medical Inc.

Radio Therapeutics Corporation (Boston Scientific Inc)

{1}------------------------------------------------

Intended Use / Indications for Use

The HALOFLEX Energy Generator is indicated for use for the coagulation of soft tissue

The HALOFLEX Energy Generator intended use is for coagulation of soft tissue, and bleeding on non bleeding sites in the gastrointestinal tract.

The HALORLEX Energy Generator is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including, but not limited to, the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, and Angiodysplasia.

Technological Characteristics

The HALORLEX Energy Generator is a modification to the HALO360 Energy Generator and HALO® Energy Generators that have already been cleared by the Food and Drug Administration ("FDA" or the "Agency") for use in coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to the esophagus (K051168 and K060169, K062441 respectively). The Generator has the same intended use and fundamental scientific technology as the cleared HALO360 and HALO90 Energy Generators. In addition, the intended use was expanded to include "coagulation of soft tissue", for this indication for use the predicate device is the RF 3000 Radiofrequency Generator (K000241).

The design changes between the HALO360 and HALO® Energy Generators and the HALOFLEX Energy Generator do not affect the clinical parameters employed in the energy delivery or the safety mechanisms that control those parameters.

The HALOFLEX Energy Generator has a limited number of design changes compared to the predicate devices. These changes are minor and do not raise no questions or effectiveness. Further, bench testing demonstrated comparable performance and safety. These modifications were as follows:

  • Add to the indications for use the "coagulation of soft tissue", to reflect the . general nature of the device.
  • Modification of the CPU board, for increased memory .
  • Consolidation of the Pressure Monitoring Board and RF MUX Board in the Hand Piece Interface Board

{2}------------------------------------------------

  • Integration on the HALO360 and HALO90 Energy Generators on the same . hardware platform as HALO360 Energy Generator
  • . Update the software codes to reflect the specific performance characteristics for each Energy Generator.
  • Optimization of components for reliability, and/or obsolescence

Performance Data

The HALOFLEX Energy Generator is technologically equivalent and clinically identical to the cleared HALO360 and HALO90 Energy Generators.

Performance testing was conducted in compliance with the applicable international and domestic requirements and the certifications (for harmonized standards) and the reports for performance testing were provided to FDA to support this submission. That data demonstrated that, when used in accordance with the instructions, the HALOFLEX is at least as safe and effective as the predicate devices HALO360 and HALO90 Energy Generators.

Substantial Equivalence

The HALOFLEX Energy Generator is as safe and effective as the HALO360 Energy Generator (K051168) and HALO90 Energy Generator (K060169 and K062441), RF 3000 Radiofrequency Generator (K000241) the predicate devices. The HALOFLEX Energy Generator has the same intended uses and indications, technological characteristics, and principles of operation as its predicate device. The minor technological differences between the HALORLES Energy Generator and its predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that the HALOFLEX Energy Generator is as safe and effective as HALO360 Energy Generator (K051168) and HALO90 Energy Generator (K060169 and K062441), RF 3000 Radiofrequency Generator the predicate devices. Thus, the HALOFLEX Energy Generator is substantially equivalent.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized human figure with outstretched arms.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

BARRX Medical, Inc. % Ms. Viorica Filimon Vice President of Quality/Regulatory Affairs 540 Oakmead Parkway Sunnyvale, California 94085

NOV 1 0 2009

Re: K092487

Trade/Device Name: HALOFLEX Energy Generator Model 1190A-115A Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: August 12, 2009 Received: August 13, 2009

Dear Ms. Filimon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 – Ms. Viorica Filimon

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Sincerely yours,

Mark N Milliken

Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use Statement

510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________

Device Name:

Indications for Use:

The HALOFLEX Energy Generator is indicated for use for the coagulation of soft tissue.

The HALOFLEX Energy Generator is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including, but not limited to, the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Die Hotor Lesions, and Angiodysplasia.

AND/OR

Prescription Use _ X Use (Part 21 C.F.R. 801 Subpart D)

Over-The-Counter

(21 C.F.R. 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

FOR M. MELKERSON

Division of Surgical, Orthopedic, and Restorative Devices

Page _ of _

510(k) Number K01248

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.