K Number
K982839
Device Name
CADD-LEGACY PCA AMBULATORY INFUSION SYSTEM, MODEL 6300
Manufacturer
Date Cleared
1998-11-02

(82 days)

Product Code
Regulation Number
880.5725
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The CADD-Legacy™ pump is suitable for intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural space, or subarachnoid space infusion.
Device Description
The CADD-Legacy™ PCA Model 6300 ambulatory infusion pump is similar in design, function, and intended use to Deltec's CADD-PCA® Ambulatory Infusion Pump and the CADD-Prizm® VIP Model 6100 Ambulatory Infusion System . The Model 6300 pump provides measured drug therapy for intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural space or subarachnoid space delivery to patients in hospital or outpatient settings. The pump is capable of storing one delivery application. The delivery application resident in the Model 6300 pump and the subject of this 510(k) Notification is the continuous infusion with bolus (PCA) delivery application.
More Information

K/DEN numbers are not found in the provided text.

No
The summary describes a standard infusion pump with programmed delivery applications and does not mention any AI/ML capabilities or related concepts.

Yes
This device is an ambulatory infusion pump that delivers measured drug therapy to patients, which is a therapeutic function.

No

The device is described as an "ambulatory infusion pump" used for "measured drug therapy." Its function is to deliver substances, not to diagnose conditions.

No

The device description explicitly states it is an "ambulatory infusion pump," which is a hardware device. The software mentioned is resident within and controls the hardware.

Based on the provided information, the CADD-Legacy™ pump is not an IVD (In Vitro Diagnostic) device.

Here's why:

  • Intended Use: The intended use clearly states that the pump is for delivering drug therapy to patients via various routes (intravenous, intra-arterial, etc.). This is a therapeutic function, not a diagnostic one.
  • Device Description: The description reinforces that the pump provides "measured drug therapy."
  • Lack of IVD Characteristics: IVD devices are used to examine specimens (like blood, urine, tissue) from the human body to provide information for diagnosis, monitoring, or screening. The CADD-Legacy™ pump does not perform any such examination of specimens.

Therefore, the CADD-Legacy™ pump is a medical device used for drug delivery, not an in vitro diagnostic device.

N/A

Intended Use / Indications for Use

"The CADD-Legacy™ pump is suitable for intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural space, or subarachnoid space infusion."

Product codes

MEA

Device Description

The CADD-Legacy™ PCA Model 6300 ambulatory infusion pump is similar in design, function, and intended use to Deltec's CADD-PCA® Ambulatory Infusion Pump and the CADD-Prizm® VIP Model 6100 Ambulatory Infusion System . The Model 6300 pump provides measured drug therapy for intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural space or subarachnoid space delivery to patients in hospital or outpatient settings. The pump is capable of storing one delivery application. The delivery application resident in the Model 6300 pump and the subject of this 510(k) Notification is the continuous infusion with bolus (PCA) delivery application.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural space, or subarachnoid space

Indicated Patient Age Range

Not Found

Intended User / Care Setting

patients in hospital or outpatient settings.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies

Functional Testing: Test plans associated with software validation, verification of software controlled programming functions, and software related to proper pump operation were certified for the CADD-Legacy™ PCA Model 6300 Ambulatory Infusion System.
Clinical Studies: Clinical studies were not deemed necessary regarding the use of the CADD-Legacy™ PCA Model 6300 Ambulatory Infusion System.

Key Metrics

Not Found

Predicate Device(s)

CADD-PCA® Ambulatory Infusion Pump, CADD-Prizm® VIP Model 6100 Ambulatory Infusion System

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 880.5725 Infusion pump.

(a)
Identification. An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.(b)
Classification. Class II (performance standards).

0

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS for the CADD-Legacy™ PCA MODEL 6300 AMBULATORY INFUSION SYSTEM

NOV 2 1998

Image /page/0/Picture/2 description: The image shows a handwritten alphanumeric string that appears to be a code or identifier. The string is "K98 2839", with a forward slash preceding the 'K'. The numbers and letters are written in a cursive style, giving the impression of a quick or informal notation. The writing is clear and legible against the background.

GENERAL INFORMATION I.

Common/Usual Name:Ambulatory Infusion Pump
Proprietary Name:CADD-Legacy™ PCA Model 6300
Ambulatory Infusion System
Applicant's name and address:SIMS Deltec, Inc.
1265 Grey Fox Road
St. Paul, MN 55112
Equivalence device comparison:CADD-PCA® Ambulatory Infusion Pump
and CADD-Prizm® VIP Model 6100
Ambulatory Infusion System

DEVICE DESCRIPTION II.

The CADD-Legacy™ PCA Model 6300 ambulatory infusion pump is similar in design, function, and intended use to Deltec's CADD-PCA® Ambulatory Infusion Pump and the CADD-Prizm® VIP Model 6100 Ambulatory Infusion System . The Model 6300 pump provides measured drug therapy for intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural space or subarachnoid space delivery to patients in hospital or outpatient settings. The pump is capable of storing one delivery application. The delivery application resident in the Model 6300 pump and the subject of this 510(k) Notification is the continuous infusion with bolus (PCA) delivery application.

III. ALTERNATIVES

Alternatives to the CADD-Legacy™ PCA Model 6300 pump include the use of other commercially available ambulatory infusion pumps, such as Deltec's CADD-PCA® and CADD-Prizm® VIP Ambulatory Infusion Pumps.

IV. POTENTIAL ADVERSE EFFECTS

The potential direct adverse effects that may occur when using the CADD-Legacy™ PCA Model 6300 pump, as well as other commercially available non-implantable ambulatory infusion pumps, include the possibility of overinfusion, under-infusion, or no infusion.

1

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS for the CADD-Legacy™ PCA MODEL 6300 AMBULATORY INFUSION SYSTEM

SUMMARY OF STUDIES V.

Functional Testing A.

Test plans associated with software validation, verification of software controlled programming functions, and software related to proper pump operation were certified for the CADD-Legacy™ PCA Model 6300 Ambulatory Infusion System.

B. Clinical Studies

Clinical studies were not deemed necessary regarding the use of the CADD-Legacy™ PCA Model 6300 Ambulatory Infusion System.

CONCLUSIONS DRAWN FROM THESE STUDIES VI.

Based upon the information provided above, the CADD-Legacy™ PCA Model 6300 Ambulatory Infusion System is substantially equivalent to other commercially available ambulatory infusion systems.

2

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with three horizontal lines above it, representing feathers or wings.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 2 1998

Mr. Edward W. Numainville Vice President, Regulatory Affairs and Quality Systems SIMS Deltec, Incorporated 1265 Grey Fox Road St. Paul, Minnesota 55112

K982839 Re : CADD-Legacy™ PCA Ambulatory Infusion Trade Name: System, Model 6300 Regulatory Class: II Product Code: MEA Dated: August 12, 1998 August 12, 1998 Received:

Dear Mr. Numainville:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual reqistration, listing of devices, qood manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਬ substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

3

Page 2 - Mr. Numainville

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

Tim A Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

4

INDICATIONS FOR USE

510(k) Number (if known):

K982839

Device Name:

CADD-Legacy™ PCA Model 6300 Ambulatory Infusion Pump

Indications For Use:

"The CADD-Legacy™ pump is suitable for intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural space, or subarachnoid space infusion."

(PLEASE DO NOT WRITE BELUW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109) OR Over-The-Counter Use

(Division Sign-Off)
Division of Dental, Infection Control,
and General Hospital Devices
510(k) Number. K982839

(Optional Format 1/2/96)

M:\510(k)ISDSubmissions1998\CADD-Legacy\IndicationsLegacyPCA