(237 days)
For In Vitro Diagnostic Use
HALO AP Dx is a software only device intended as an aid to the pathologist to review, interpret and manage digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue for the purposes of pathology primary diagnosis. HALO AP Dx is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens.
It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and, where necessary, use conventional light microscopy review when making a diagnostic decision. HALO AP Dx is intended for use with the Hamamatsu NanoZoomer S360MD Slide scanner and the JVC Kenwood JD-C240BN01A display.
HALO AP Dx, version 2.1 is a browser-based software-only device intended to aid pathology professionals in viewing, manipulating, management and interpretation of digital pathology whole slide images (WSI) of glass slides obtained from the Hamamatsu Photonics K.K. NanoZoomer S360MD scanner and viewed on the JVC Kenwood JD-C240BN01A display.
HALO AP Dx is typically operated as follows:
- Image acquisition is performed using the predicate device, NanoZoomer S360MD Slide scanner according to its Instructions for Use. The operator performs quality control of the digital slides per the instructions of the NanoZoomer and lab specifications to determine if re-scans are necessary.
- Once image acquisition is complete, the unaltered image is saved by the scanner's software to an image storage location. HALO AP Dx ingests the image, and a copy of image metadata is stored in the subject device's database to improve viewing response times.
- Scanned images are reviewed by scanning personnel such as histotechnicians to confirm image quality and initiate any re-scans before making it available to the pathologist.
- The reading pathologist selects a patient case from a selected worklist within HALO AP Dx whereby the subject device fetches the associated images from external image storage.
- The reading pathologist uses the subject device to view the images and can perform the following actions, as needed:
a. Zoom and pan the image.
b. Measure distances and areas in the image.
c. Annotate images.
d. View multiple images side by side in a synchronized fashion.
The above steps are repeated as necessary.
After viewing all images belonging to a particular case (patient), the pathologist will make a diagnosis which is documented in another system, such as a Laboratory Information System (LIS).
The interoperable components of HALO AP Dx are provided in table 1 below:
Table 1. Interoperable Components for Use with HALO AP Dx
| Components | Manufacturer | Model |
|---|---|---|
| Scanner | Hamamatsu | NanoZoomer S360MD Slide scanner |
| Display | JVC | JD-C240BN01A |
This FDA 510(k) clearance letter pertains to HALO AP Dx, a software-only device for digital pathology image review. The documentation indicates that the device has been deemed substantially equivalent to a predicate device, the Hamamatsu NanoZoomer S360MD Slide scanner system (K213883).
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The provided document defines the performance data points primarily through "Performance Data" and "Summary of Studies" sections, focusing on comparisons to the predicate device and usability. There are no explicit quantitative "acceptance criteria" presented as specific thresholds, but rather statements of adequacy and similarity to the predicate.
| Acceptance Criterion (Implicit) | Reported Device Performance |
|---|---|
| Image Reproduction Quality (Color Accuracy) | Criteria: Identical image reproduction compared to the predicate device (NZViewMD viewer), specifically regarding pixel-wise color accuracy.Performance: Pixel-level comparisons demonstrated that the 95th percentile CIEDE2000 values across all Regions of Interest (ROIs) from varied tissue types and diagnoses were less than 3 ΔE00. This was determined to be "identical image reproduction." |
| Turnaround Time (Image Loading - Case Selection) | Criteria: "When selecting a case, it should not take longer than 4 seconds until the image is fully loaded."Performance: Determined to be "adequate for the intended use of the subject device." (No specific value reported, but implies <= 4 seconds). |
| Turnaround Time (Image Loading - Panning) | Criteria: "When panning the image, it should not take longer than 3 seconds until the image is fully loaded."Performance: Determined to be "adequate for the intended use of the subject device." (No specific value reported, but implies <= 3 seconds). |
| Measurement Accuracy | Criteria: Ability to perform accurate measurements (distance and area).Performance: "The subject device has been found to perform accurate measurements with respect to its intended use." (Verified using a test image with known sizes). |
| System Responsiveness under Load | Criteria: Maintain responsiveness under constant utilization.Performance: "Concurrent multi-user load testing confirms HALO AP Dx performance remains responsive under constant utilization over a long time period." |
| Human Factors/Usability (Safety and Effectiveness for Users) | Criteria: User interface is intuitive, safe, and effective for intended users.Performance: "Task-based usability tests verified the HALO AP Dx user interface to be intuitive, safe, and effective for the range of intended users." (Conducted per FDA's Guidance on Applying Human Factors and Usability Engineering to Medical Devices (2016)). |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: The document does not explicitly state the numerical sample size for the test set (e.g., number of whole slide images). It mentions "multiple tiles at multiple magnification levels" and "all ROIs taken from images with varied tissue types and diagnoses" for image reproduction testing, and "a test image containing objects with known sizes" for measurement accuracy. This suggests a varied, though unspecified, set of images or data points were used for testing.
- Data Provenance: Not explicitly stated regarding country of origin. The study appears to be an internal non-clinical performance evaluation. The type of tissue used for image reproduction testing is "varied tissue types and diagnoses," implying real FFPE tissue samples, but it doesn't specify if they were retrospective or prospectively collected for the study.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- The document describes non-clinical performance testing (e.g., pixel-wise comparison, turnaround times, measurement accuracy, human factors testing). These tests typically rely on predefined objective metrics or reference standards rather than expert consensus on diagnostic interpretation.
- For the human factors/usability testing, the study "verified the HALO AP Dx user interface to be intuitive, safe, and effective for the range of intended users." This implies involvement of intended users (pathologists or similar professionals), but neither the specific number nor their qualifications are detailed.
4. Adjudication Method for the Test Set
- Adjudication methods (e.g., 2+1, 3+1) are typically relevant for studies where a "ground truth" is established by multiple human readers for diagnostic accuracy.
- Since this document focuses on technical performance and usability, rather than diagnostic accuracy (which would involve human pathologists making diagnoses with and without AI assistance), traditional adjudication methods were not applicable or described. The "ground truth" in these tests consists of objective technical parameters or usability feedback.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No MRMC comparative effectiveness study was described. The study primarily focuses on showing the technical performance of HALO AP Dx is substantially equivalent to the viewing software component of the predicate device (NZViewMD), not on improving human reader performance or diagnostic accuracy with AI assistance.
- The device is a "viewer" intended to "aid the pathologist," not an AI-powered diagnostic algorithm. Therefore, an MRMC study comparing human readers with and without AI assistance to measure effect size is not relevant to this submission, which focuses on the viewing platform itself.
6. Standalone (Algorithm Only) Performance
- This device, HALO AP Dx, is a standalone software-only device in the sense that it functions as a digital pathology image viewer.
- However, it does not involve a diagnostic AI algorithm where "standalone performance" (e.g., sensitivity/specificity of the AI itself) would be measured. Its "performance" is about accurate image reproduction, speed, and usability of the viewing functions.
7. Type of Ground Truth Used
The ground truth used for the technical performance evaluations was objective and predefined:
- For Image Reproduction: A pixel-wise comparison to the predicate device's viewer (NZViewMD) was performed. The "ground truth" was essentially the image rendered by the predicate's software.
- For Turnaround Times: Time taken for specific actions (loading, panning) against predefined numerical thresholds (4 seconds, 3 seconds).
- For Measurement Accuracy: A "test image containing objects with known sizes." The known sizes were the ground truth.
- For Human Factors: User feedback and task completion during usability tests against predefined criteria for intuitiveness, safety, and effectiveness.
8. Sample Size for the Training Set
- The document does not mention a "training set" in the context of machine learning. This is because HALO AP Dx is described as a "software only device intended as an aid to the pathologist to review, interpret and manage digital images," not an AI/ML-based diagnostic algorithm that would require a training set.
- Its core functionality is image display and manipulation, not pattern recognition or classification that would necessitate machine learning.
9. How the Ground Truth for the Training Set Was Established
- As no machine learning training set is mentioned or implied for this device's functionality, this question is not applicable.
FDA 510(k) Clearance Letter - HALO AP Dx
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID # 04017.06.09
Silver Spring, MD 20993
www.fda.gov
Indica Labs, Inc.
Michele Fredrickson
Sr Manager, Quality and Regulatory
8700 Education Place NW
Building B
Albuquerque, New Mexico 87114
Re: K232833
Trade/Device Name: HALO AP Dx
Regulation Number: 21 CFR 864.3700
Regulation Name: Whole slide imaging System
Regulatory Class: Class II
Product Code: QKQ
Dated: September 13, 2023
Received: September 13, 2023
Dear Michele Fredrickson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
May 7, 2024
Page 2
U.S. FOOD & DRUG ADMINISTRATION
Indica Labs, Inc.
Michele Fredrickson
Sr Manager, Quality and Regulatory
8700 Education Place NW
Building B
Albuquerque, New Mexico 87114
May 7, 2024
Re: K232833
Trade/Device Name: HALO AP Dx
Regulation Number: 21 CFR 864.3700
Regulation Name: Whole slide imaging System
Regulatory Class: Class II
Product Code: QKQ
Dated: September 13, 2023
Received: September 13, 2023
Dear Michele Fredrickson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
K232833 - Michele Fredrickson Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shyam Kalavar -S
Shyam Kalavar
Deputy Branch Chief
Division of Molecular Genetics and Pathology
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 3
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K232833
Device Name
HALO AP Dx
Indications for Use (Describe)
For In Vitro Diagnostic Use
HALO AP Dx is a software only device intended as an aid to the pathologist to review, interpret and manage digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue for the purposes of pathology primary diagnosis. HALO AP Dx is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens.
It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and, where necessary, use conventional light microscopy review when making a diagnostic decision. HALO AP Dx is intended for use with the Hamamatsu NanoZoomer S360MD Slide scanner and the JVC Kenwood JD-C240BN01A display.
Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF
Page 4
510(k) Summary
HALO AP Dx
1. Submitter
Indica Labs, Inc.
8700 Education Place, Building B
Albuquerque, NM, 87114 USA
Contact Person: Michele Fredrickson, Sr. Manager Quality & Regulatory
Tel: 505-492-0972
Date Prepared: May 6, 2024
2. Device
Device Trade Name: HALO AP Dx
Version: 2.1
Regulation Name: Whole Slide Imaging System
Regulation: 21 CFR §864.3700
Regulatory Class: Class II
Product Classification Code: QKQ
Review Panel: 88 – Pathology
510 (k) Submission Number: K232833
3. Predicate Device
Predicate Manufacturer: Hamamatsu Photonics K.K.
Predicate Trade Name: NanoZoomer S360MD Slide scanner system
Predicate 510(k): K213883
No reference devices were used in this submission.
4. Device Description
HALO AP Dx, version 2.1 is a browser-based software-only device intended to aid pathology professionals in viewing, manipulating, management and interpretation of digital pathology whole slide images (WSI) of glass slides obtained from the Hamamatsu Photonics K.K. NanoZoomer S360MD scanner and viewed on the JVC Kenwood JD-C240BN01A display.
HALO AP Dx is typically operated as follows:
- Image acquisition is performed using the predicate device, NanoZoomer S360MD Slide scanner according to its Instructions for Use. The operator performs quality control of the digital slides per the instructions of the NanoZoomer and lab specifications to determine if re-scans are necessary.
Page 5
-
Once image acquisition is complete, the unaltered image is saved by the scanner's software to an image storage location. HALO AP Dx ingests the image, and a copy of image metadata is stored in the subject device's database to improve viewing response times.
-
Scanned images are reviewed by scanning personnel such as histotechnicians to confirm image quality and initiate any re-scans before making it available to the pathologist.
-
The reading pathologist selects a patient case from a selected worklist within HALO AP Dx whereby the subject device fetches the associated images from external image storage.
-
The reading pathologist uses the subject device to view the images and can perform the following actions, as needed:
a. Zoom and pan the image.
b. Measure distances and areas in the image.
c. Annotate images.
d. View multiple images side by side in a synchronized fashion. -
The above steps are repeated as necessary.
After viewing all images belonging to a particular case (patient), the pathologist will make a diagnosis which is documented in another system, such as a Laboratory Information System (LIS).
The interoperable components of HALO AP Dx are provided in table 1 below:
Table 1. Interoperable Components for Use with HALO AP Dx
| Components | Manufacturer | Model |
|---|---|---|
| Scanner | Hamamatsu | NanoZoomer S360MD Slide scanner |
| Display | JVC | JD-C240BN01A |
5. Intended Use
For In Vitro Diagnostic Use
HALO AP Dx is a software only device intended as an aid to the pathologist to review, interpret and manage digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue for the purposes of pathology primary diagnosis. HALO AP Dx is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens.
It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and, where necessary, use conventional light microscopy review when making a diagnostic decision. HALO AP Dx is intended for use with the Hamamatsu NanoZoomer S360MD Slide scanner and the JVC Kenwood JD-C240BN01A display.
6. Summary of Technological Characteristics
| Item | Predicate (K213883) | Subject Device |
|---|---|---|
| Device Trade Name | Hamamatsu NanoZoomer S360MD Slide scanner system | HALO AP Dx |
Page 6
| Item | Predicate (K213883) | Subject Device |
|---|---|---|
| Product Code | PSY | QKQ |
| Regulation | 864.3700 | 864.3700 |
| Regulation Name | Whole Slide Imaging System | Whole Slide Imaging System |
| Classification | Class II | Class II |
| Indications for Use | The NanoZoomer S360MD Slide scanner system ("NanoZoomer System") is an automated digital slide creation, viewing and management system. The NanoZoomer System is intended for in vitro diagnostic use as an aid to the pathologist to review and interpret digital images of surgical pathology slides prepared from formalin-fixed paraffin embedded ("FFPE") tissue for the purposes of pathology primary diagnosis. The NanoZoomer System is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens.The NanoZoomer System comprises the NanoZoomer S360MD Slide scanner, the NZViewMD Software and the JVC Kenwood JD-C240BN01A display. The NanoZoomer System is for creation and viewing of digital images of scanned glass slides that would otherwise be appropriate for manual visualization by conventional light microscopy. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using NanoZoomer System. | HALO AP Dx is a software only device intended as an aid to the pathologist to review, interpret and manage digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue for purposes of pathology primary diagnosis. HALO AP Dx is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens.It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and, where necessary, use conventional light microscopy review when making a diagnostic decision. HALO AP Dx is intended for use with the Hamamatsu NanoZoomer S360MD Slide scanner and the JVC Kenwood JD-C240BN01A display. |
| Specimen Type | Surgical pathology slides prepared from FFPE tissue | Same |
| Diagnostic Image File Format | Hamamatsu NDPi File | Same |
| Image Manipulation | Functions for continuous panning and zooming, annotations, distance/area measurements, track visited areas, | Functions for continuous panning and zooming, annotations, image adjustments, distance/area measurements, organize |
Page 7
| Item | Predicate (K213883) | Subject Device |
|---|---|---|
| and Review Functions | export images, discrete Z-axis displacement, and display of diagnostic status of images. | workload and view patient data, export images, and display of diagnostic status of images. |
| Type of Software Application | PC-based installed application | Internet browser-based application:Google Chrome version 116 and aboveMicrosoft Edge version 116 and above |
| Principle of Operation | During review, the pathologist opens the WSI images acquired with the scanner from the network storage, performs further QC, and reviews and interprets the WSI images to make a diagnosis. | Same |
| Image Storage | User-supplied network attached storage | Same |
| End User's Interface | NZViewMD | HALO AP Dx |
| Scanner | Hamamatsu S360MD | Same |
| Display Monitor | JVC Kenwood JD-C240BN01A | Same |
7. Performance Data
| Performance data | Description |
|---|---|
| Identical Image Reproduction | Based on pixel-wise comparison to NZViewMD viewer, across multiple tiles at multiple magnification levels, it was determined that color images reproduced by HALO AP Dx demonstrated identical image reproduction compared to the predicate device. |
| Turnaround time | Turnaround times for image loading, panning, and zooming have been determined and found to be adequate for the intended use of the subject device. The system requirements have been fulfilled:• When selecting a case, it should not take longer than 4 seconds until the image is fully loaded.• When panning the image, it should not take longer than 3 seconds until the image is fully loaded. |
| Measurements | Measurement accuracy has been verified using a test image containing objects with known sizes. |
| Human factors testing | HALO AP Dx has been found to be safe and effective for the intended users, uses, and use environments. |
Page 8
8. Substantial Equivalence Comparison
The major difference between the subject and predicate device is that the predicate includes the Hamamatsu S360MD scanner and the JVC Kenwood JD-C240BN01A monitor, whereas the subject device is indicated for use with the same scanner and monitor. Therefore, the indications for use for the predicate contain a section detailing the creation of the WSI, while HALO AP Dx is only focused on viewing those images.
As proposed, when HALO AP Dx is used with the Hamamatsu S360MD scanner and the JVC Kenwood JD-C240BN01A monitor, it has similar indications for use, inputs, outputs, intended user types, and technological characteristics as the NZViewMD viewing software included with the predicate device. Therefore, HALO AP Dx is substantially equivalent to the predicate device (K213883).
9. Summary of Studies
Non-clinical test results
Conducted per FDA's Guidance on Technical Performance Assessment of Digital Pathology Whole Slide Imaging Devices (2016):
• Pixelwise comparison testing was conducted in the Windows 10 environment to support identical image reproduction. Pixel-level comparisons were made across multiple tiles at multiple magnification levels, to compare color images reproduced by HALO AP Dx and Nanozoomer NZViewMD. After alignment of the image pair, the pixelwise differences were computed. The mean, minimum and maximum of the 95th percentile were reported compositely and separately, per Region of Interest (ROI), image, magnification level, scanner, and browser. The 95th percentile values (CIEDE2000) across all ROIs taken from images with varied tissue types and diagnoses are less than 3 ΔE00. It was determined that color images reproduced by HALO AP Dx demonstrated identical image reproduction compared to the predicate device.
• Turnaround times for panning and zooming have been determined and found to be adequate for the intended use of the subject device. Concurrent multi-user load testing confirms HALO AP Dx performance remains responsive under constant utilization over a long time period.
• The subject device has been found to perform accurate measurements with respect to its intended use.
Conducted per FDA's Guidance on Applying Human Factors and Usability Engineering to Medical Devices (2016):
• Task-based usability tests verified the HALO AP Dx user interface to be intuitive, safe, and effective for the range of intended users.
10. Conclusion
When HALO AP Dx is used with the Hamamatsu Photonics K.K. NanoZoomer S360MD scanner and JVC Kenwood JD-C240BN01A display, it has similar Indications for Use, Functional, and Technological Characteristics as the predicate's viewing software. The results of non-clinical testing demonstrate the device is substantially equivalent to the predicate device (K213883).
§ 864.3700 Whole slide imaging system.
(a)
Identification. The whole slide imaging system is an automated digital slide creation, viewing, and management system intended as an aid to the pathologist to review and interpret digital images of surgical pathology slides. The system generates digital images that would otherwise be appropriate for manual visualization by conventional light microscopy.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the following information:
(i) The indications for use must specify the tissue specimen that is intended to be used with the whole slide imaging system and the components of the system.
(ii) A detailed description of the device and bench testing results at the component level, including for the following, as appropriate:
(A) Slide feeder;
(B) Light source;
(C) Imaging optics;
(D) Mechanical scanner movement;
(E) Digital imaging sensor;
(F) Image processing software;
(G) Image composition techniques;
(H) Image file formats;
(I) Image review manipulation software;
(J) Computer environment; and
(K) Display system.
(iii) Detailed bench testing and results at the system level, including for the following, as appropriate:
(A) Color reproducibility;
(B) Spatial resolution;
(C) Focusing test;
(D) Whole slide tissue coverage;
(E) Stitching error; and
(F) Turnaround time.
(iv) Detailed information demonstrating the performance characteristics of the device, including, as appropriate:
(A) Precision to evaluate intra-system and inter-system precision using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included.
(B) Reproducibility data to evaluate inter-site variability using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included.
(C) Data from a clinical study to demonstrate that viewing, reviewing, and diagnosing digital images of surgical pathology slides prepared from tissue slides using the whole slide imaging system is non-inferior to using an optical microscope. The study should evaluate the difference in major discordance rates between manual digital (MD) and manual optical (MO) modalities when compared to the reference (
e.g., main sign-out diagnosis).(D) A detailed human factor engineering process must be used to evaluate the whole slide imaging system user interface(s).
(2) Labeling compliant with 21 CFR 809.10(b) must include the following:
(i) The intended use statement must include the information described in paragraph (b)(1)(i) of this section, as applicable, and a statement that reads, “It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using this device.”
(ii) A description of the technical studies and the summary of results, including those that relate to paragraphs (b)(1)(ii) and (iii) of this section, as appropriate.
(iii) A description of the performance studies and the summary of results, including those that relate to paragraph (b)(1)(iv) of this section, as appropriate.
(iv) A limiting statement that specifies that pathologists should exercise professional judgment in each clinical situation and examine the glass slides by conventional microscopy if there is doubt about the ability to accurately render an interpretation using this device alone.