(233 days)
HALO is a notification only cloud-based image processing software artificial intelligence algorithms to analyze patient imaging data in parallel to the standard of care imaging interpretation. Its intended use is to identify suggestive imaging patterns of a pre-specified clinical condition and to directly notify an appropriate medical specialist.
HALO's indication is to facilitate the evaluation of the brain vasculature on patients suspected of stroke by processing and analyzing contrast enhanced CT angiograms of the brain acquired in an acute setting. After completion of the data analysis. HALO sends a notification if a pattern suggestive for a suspected intracranial Large Vessel Occlusion (LVO) of the anterior circulation (ICA, M1 or M2) has been identified in an image.
The intended users of HALO are defined as appropriate medical specialists that are involved in the diagnosis and care of stroke patients at emergency department where stroke patients are administered. They include physicians such as neurologists, and/or other emergency department physicians.
HALO's output should not be used for primary diagnosis or clinical decisions; the final diagnosis is always decided upon by the medical specialist. HALO is indicated for CT scanners from GE Healthcare.
HALO is a notification only, cloud-based clinical support tool which identifies image features and communicates the analysis results to a specialist in parallel to the standard of care workflow.
HALO is designed to process CT angiograms of the brain and facilitate evaluation of these images using artificial intelligence to detect patterns suggestive of an intracranial large vessel occlusion (LVO) of the anterior circulation.
A copy of the original CTA images is sent to HALO cloud servers for automatic image processing. After analyzing the images, HALO sends a notification regarding a suspected finding to a specialist, recommending review of these images. The specialist can review the results remotely in a compatible DICOM web viewer.
Here's a summary of the acceptance criteria and study details for the HALO device, based on the provided FDA 510(k) summary:
HALO Device Acceptance Criteria and Study Details
1. Table of Acceptance Criteria and Reported Device Performance
| Metric | Acceptance Criteria (Implicit) | Reported Device Performance (HALO) |
|---|---|---|
| Sensitivity | Sufficiently high for LVO detection (comparable to predicate) | 91.1% (95% CI, 86.0%-94.8%) |
| Specificity | Sufficiently high for LVO detection (comparable to predicate) | 87.0% (95% CI, 81.2%-91.5%) |
| AUC | High (indicative of good discriminative power) | 0.97 |
| Notification Time | Fast enough for acute stroke setting (comparable to predicate) | Median: 4 minutes 31 seconds. Range: 3:47 to 7:12 |
| Substantial Equivalence | Equivalent to predicate device ContaCT in terms of indications for use, technological characteristics, and safety and effectiveness. | Concluded to be substantially equivalent. |
2. Sample Size and Data Provenance for Test Set
- Sample Size: 348 CTA scans were initially collected, with 364 patients included for further analysis after exclusion. It's unclear if the "348 CTA scans" and "364 patients" refer to the same dataset or if some patients had multiple scans or if there was an expansion of the dataset. Assuming 364 cases (patients with at least one scan) were used.
- Data Provenance: Retrospective evaluation in a consecutive patient cohort. Data was collected from US comprehensive stroke centers.
3. Number and Qualifications of Experts for Ground Truth
- Number of Experts: 3 neuro radiologists.
- Qualifications: "Neuro radiologists" implies specialized training and experience in interpreting neurological imaging, which is appropriate for stroke diagnosis. Specific years of experience are not mentioned.
4. Adjudication Method for Test Set
The adjudication method is not explicitly stated. It says "Ground truth was established by an expert panel consisting of 3 neuro radiologists," which suggests a consensus-based approach, but the specific rule (e.g., majority vote, unanimous agreement, review by a lead expert if disagreement) is not provided.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not explicitly mentioned or conducted as detailed in the summary. The study focused on the standalone performance of the HALO algorithm.
6. Standalone Performance (Algorithm Only)
Yes, a standalone study was performed. The clinical study retrospectively evaluated the performance of the HALO clinical decision support algorithm for LVO detection using the collected CTA scans. The reported sensitivity, specificity, AUC, and notification time are all measures of the algorithm's standalone performance.
7. Type of Ground Truth Used
Expert Consensus. The ground truth for the test set was established by an expert panel consisting of 3 neuro radiologists.
8. Sample Size for Training Set
The sample size for the training set is not explicitly mentioned in the provided document. The document only covers the evaluation of the algorithm.
9. How Ground Truth for Training Set was Established
How the ground truth was established for the training set is not explicitly mentioned in the provided document. ("database of images" is stated for the core algorithm, but not how ground truth was applied to them).
{0}------------------------------------------------
November 20, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
NiCo-Lab B.V. % Roujuan Zhang Consultant MD Squared B.V. High Tech Campus 29 Eindhoven, 5858AE NETHERLANDS
Re: K200873
Trade/Device Name: HALO Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological computer aided triage and notification software Regulatory Class: Class II Product Code: QAS Dated: October 1, 2020 Received: October 5, 2020
Dear Roujuan Zhang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K200873
Device Name HALO
Indications for Use (Describe)
HALO is a notification only cloud-based image processing software artificial intelligence algorithms to analyze patient imaging data in parallel to the standard of care imaging interpretation. Its intended use is to identify suggestive imaging patterns of a pre-specified clinical condition and to directly notify an appropriate medical specialist.
HALO's indication is to facilitate the evaluation of the brain vasculature on patients suspected of stroke by processing and analyzing contrast enhanced CT angiograms of the brain acquired in an acute setting. After completion of the data analysis. HALO sends a notification if a pattern suggestive for a suspected intracranial Large Vessel Occlusion (LVO) of the anterior circulation (ICA, M1 or M2) has been identified in an image.
The intended users of HALO are defined as appropriate medical specialists that are involved in the diagnosis and care of stroke patients at emergency department where stroke patients are administered. They include physicians such as neurologists, and/or other emergency department physicians.
HALO's output should not be used for primary diagnosis or clinical decisions; the final diagnosis is always decided upon by the medical specialist. HALO is indicated for CT scanners from GE Healthcare.
| Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) | Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|---|---|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | |||
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K200873
510(k) Summary
| Date Prepared: | March 27, 2020 | ||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Manufacturer: | NICo-Lab B.V.Paasheuvelweg 25, 1105BP, AmsterdamThe Netherlands | ||||||||||||||||
| Contact Person: | Merel BoersCEOPhone: +31202440852E-mail: mboers@nico-lab.com | ||||||||||||||||
| Device: | |||||||||||||||||
| Trade Name:HALOClassification Name:Radiological Computer-Assissted Triage andNotification SoftwareClassification Regulation:21 CFR 892.2080Classification Panel:RadiologyDevice Class:Class IIPrimary Product Code:QAS | Trade Name: | HALO | Classification Name: | Radiological Computer-Assissted Triage andNotification Software | Classification Regulation: | 21 CFR 892.2080 | Classification Panel: | Radiology | Device Class: | Class II | Primary Product Code: | QAS | |||||
| Trade Name: | HALO | ||||||||||||||||
| Classification Name: | Radiological Computer-Assissted Triage andNotification Software | ||||||||||||||||
| Classification Regulation: | 21 CFR 892.2080 | ||||||||||||||||
| Classification Panel: | Radiology | ||||||||||||||||
| Device Class: | Class II | ||||||||||||||||
| Primary Product Code: | QAS | ||||||||||||||||
| Primary Predicate Device: | |||||||||||||||||
| Trade Name:ContaCTManufacturer:Viz.Al, Inc.DeNovo Clearance:DEN170073Classification Name:Radiological Computer-Assissted Triage andNotification SoftwareClassification Regulation:21 CFR 892.2080Classification Panel:RadiologyDevice Class:Class IIProduct Code:QAS | Trade Name: | ContaCT | Manufacturer: | Viz.Al, Inc. | DeNovo Clearance: | DEN170073 | Classification Name: | Radiological Computer-Assissted Triage andNotification Software | Classification Regulation: | 21 CFR 892.2080 | Classification Panel: | Radiology | Device Class: | Class II | Product Code: | QAS | |
| Trade Name: | ContaCT | ||||||||||||||||
| Manufacturer: | Viz.Al, Inc. | ||||||||||||||||
| DeNovo Clearance: | DEN170073 | ||||||||||||||||
| Classification Name: | Radiological Computer-Assissted Triage andNotification Software | ||||||||||||||||
| Classification Regulation: | 21 CFR 892.2080 | ||||||||||||||||
| Classification Panel: | Radiology | ||||||||||||||||
| Device Class: | Class II | ||||||||||||||||
| Product Code: | QAS |
{4}------------------------------------------------
Device description:
HALO is a notification only, cloud-based clinical support tool which identifies image features and communicates the analysis results to a specialist in parallel to the standard of care workflow.
HALO is designed to process CT angiograms of the brain and facilitate evaluation of these images using artificial intelligence to detect patterns suggestive of an intracranial large vessel occlusion (LVO) of the anterior circulation.
A copy of the original CTA images is sent to HALO cloud servers for automatic image processing. After analyzing the images, HALO sends a notification regarding a suspected finding to a specialist, recommending review of these images. The specialist can review the results remotely in a compatible DICOM web viewer.
Indications for Use: HALO is a notification only cloud-based image processing software application using artificial intelligence algorithms to analyze patient imaging data in parallel to the standard of care imaging interpretation. Its intended use is to identify suggestive imaging patterns of a pre-specified clinical condition and to directly notify an appropriate medical specialist.
HALO's indication is to facilitate the evaluation of the brain vasculature on patients suspected of stroke by processing and analyzing CT angiograms of the brain acquired in an acute setting. After completion of the data analysis, HALO sends a notification if a pattern suggestive for a suspected intracranial Large Vessel Occlusion (LVO) of the anterior circulation (ICA, M1 or M2) has been identified in an image.
The intended users of HALO are defined as medical specialists or a team of specialists that are involved in the diagnosis and care of stroke patients at emergency departments or other department where stroke patients are administered. They include physicians such as neurologists, radiologists, and/or other emergency department physicians.
HALO's output should not be used for primary diagnosis or clinical decisions; the final diagnosis is always decided upon by the medical specialist. HALO is indicated for CT scanners from GE Healthcare.
Technological Both HALO and the predicate device ContaCT are intended as a characteristics: notification-only, parallel workflow tool to identify and communicate
{5}------------------------------------------------
images of specific patients to a specialist, independent of standard of care workflow. After completion of the data analysis, both devices send a notification if a pattern suggestive for a suspected Large Vessel Occlusion (LVO) has been identified in the dataset.
The technological characteristic of HALO and how it is comparable to the currently marketed and predicate device ContaCT are summarized below.
| Predicate: ContaCT | Subject: HALO | ||
|---|---|---|---|
| Clinical condition | Large vessel | Large vessel | |
| occlusion | occlusion | ||
| Anatomical region of | Head | Head | |
| interest | |||
| Data acquisition | CT angiogram images | CT angiogram images | |
| protocol | of the brain | of the brain | |
| Segmentation of | No; device does not | No; device does not | |
| region of interest | mark, highlight, or | mark, highlight, or | |
| direct users' attention | direct users' attention | ||
| to a specific location | to a specific location | ||
| in the original image. | in the original image. | ||
| Core Algorithm | Artificial intelligence | Artificial intelligence | |
| algorithm with | |||
| database of images | database of images | ||
| Device Output/ | The software sends a | The software sends a | |
| notification to the | notification email to | ||
| Notification | specialist identifying | the specialist | |
| the study of interest. | identifying the study | ||
| Additionally, the | of interest. | ||
| device also provides a | Additionally, the | ||
| DICOM viewing | device provides user | ||
| mobile application | with a link to DICOM | ||
| allow users preview | Web viewer allowing | ||
| the images. | users review the | ||
| images. | |||
| Triage effectiveness | Notification time is | Notification time is | |
| defined as the time | defined as the time | ||
| from CTA to | from CTA to | ||
| notification. | notification. | ||
| Independent of | No cases are removed | No cases are removed | |
| standard of care | from worklist | from worklist | |
| workflow |
Based on the information provided above, non-clinical and clinical performance tests provided in this 510(k) premarket notification demonstrate the proposed HALO is considered substantially equivalent to
{6}------------------------------------------------
| Summary of Non-Clinical Performance Data: | Non-Clinical software testing has been performed in accordance with the its predicate device ContaCT in terms of fundamental scientific technology."Guidance for Industry and FDA Staff – Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (issued May 11, 2005, document number 337) and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices” (issued October 2, 2014, document number 1825).All these tests were used to support substantial equivalence of the subject device and demonstrate that the performance of HALO is according to predefined user requirements, system level requirements and the risk control measures. The device meets the acceptance criteria and is adequate for its intended use. |
|---|---|
| Summary of Clinical Performance Data: | In a multi-center clinical study, the performance of the HALO clinical decision support algorithm for LVO detection was retrospectively evaluated in a consecutive patient cohort admitted to US comprehensive stroke centers.Three hundred forty-eight CTA scans of the brain were collected. After exclusion, 364 patients were included for further analyses. Ground truth was established by an expert panel consisting of 3 neuro radiologists.For the primary endpoint: calculation of the performance of the HALO algorithm showed a sensitivity and specificity for LVO detection of respectively 91.1% (95% CI, 86.0%-94.8%) and 87.0% (95% CI, 81.2%-91.5%). The area under the curve (AUC) is 0.97.For the secondary endpoints the median notification time for the detected LVO cases was 4 minutes 31 seconds, with a minimum of 3:47 and maximal 7:12.The HALO performance with regard to sensitivity and specificity, and the notification time are both equivalent to that of the selected predicate device: ContaCT of Viz.AI.Therefore, the HALO algorithm fulfills the requirement of a suitable screening tool to support diagnosis of LVOs by flagging these scans as requiring urgent radiologist review. |
{7}------------------------------------------------
| Substantial |
|---|
| Equivalence |
| Conclusion: |
The non-clinical and clinical performance tests provided in this 510(k) premarket notification demonstrate HALO is substantially equivalence to the currently marketed predicate device ContaCT (DEN 170073) in terms of indications for use, technological characteristics and safety and effectiveness. The differences between the subject device and predicate device do not alter the triage use of the device and do not affect its safety and effectiveness.
N/A