(18 days)
Not Found
No
The summary does not mention AI, ML, or related terms, nor does it describe features or processes typically associated with these technologies.
No
The device is indicated for monitoring, recording, and alarming of physiological parameters, not for providing therapy.
Yes
Explanation: The device is indicated for monitoring, recording, and alarming of multiple physiological parameters, which are functions commonly associated with diagnostic devices that provide information about a patient's health status.
No
The device description explicitly lists "Philips MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90 IntelliVue Patient Monitors," which are hardware devices. The summary also mentions "system level tests, performance tests, and safety testing from hazard analysis," indicating hardware-related testing.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use is for "monitoring, recording and alarming of multiple physiological parameters of patients." This involves measuring parameters directly from the patient's body (e.g., heart rate, blood pressure, oxygen saturation) using sensors attached to the patient.
- IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- Lack of Mention of Samples: The description does not mention the analysis of any biological samples.
- Focus on Physiological Parameters: The device focuses on monitoring real-time physiological signals from the patient.
Therefore, the Philips IntelliVue Patient Monitors described here are patient monitoring devices, not in vitro diagnostic devices.
N/A
Intended Use / Indications for Use
Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics and neonates in healthcare facilities. The MP20, MP30, MP40 and MP50 are additionally intended for use in transport situations within healthcare facilities.
ST Segment monitoring is restricted to adult patients only.
The transcutaneous gas measurement (tcp02 / tcpCO2) is restricted to neonatal patients only.
Product codes (comma separated list FDA assigned to the subject device)
DSI, MLD, MHX, DSJ, DSK, DXN, DXG, KRB, DRQ, DRT, MSX, DPS, MLC, DRW, KRC, DXJ, DRJ, DQA, DSB, DSH, DSF, DRS, DSA, DRG, CCK, CBQ, NHO, NHP, NHQ, CBS, CBR, CCL, BZQ, LKD, KLK, FLL, GWR, GWS
Device Description
The modification is the introduction of software Release E.03 for the IntelliVue patient monitor devices.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
adults, pediatrics, and neonates
Intended User / Care Setting
health care professionals; in healthcare facilities and during transport within healthcare facilities.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Testing involved system level tests, performance tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence.
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the modified device with respect to the predicate. Testing involved system level tests, performance tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence. The results demonstrate that the Philips IntelliVue Patient Monitor meets all reliability requirements and performance claims.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
K021778, K030038, K032858, K033513, K040304, K041235, K042845, K050141, K050762, K051106, K052801, K053522, K060221, K060541, K061052, K061610 and K062283
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).
(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
0
510 (k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.
- The submitter of this premarket notification is: Andreas Suchi Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany Tel: ++49 7031 463-1291 Fax: ++49 7031 463-2442 e-mail: andreas.suchi@philips.com
This summary was prepared on October 31, 2006.
-
- The names of the devices are the Philips MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90 IntelliVue Patient Monitors. Classification names are as follows:
Device Panel | Classification | ProCode | Description |
---|---|---|---|
Circulatory | |||
System Devices | |||
(12625) | $870.1025, II | DSI | Detector and alarm, arrhythmia |
$870.1025, II | MLD | Monitor, ST Segment with Alarm | |
$870.1025, II | MHX | Monitor, Physiological, Patient | |
(with arrhythmia detection or | |||
alarms) | |||
$870.1100, II | DSJ | Alarm, Blood Pressure | |
$870.1110, II | DSK | Computer, Blood Pressure | |
$870.1130, II | DXN | System, Measurement, Blood- | |
Pressure, Non-Invasive | |||
$870.1435, II | DXG | Computer, Diagnostic, Pre- | |
Programmed, Single-Function | |||
$870.1915, II | KRB | Probe, Thermodilution | |
$870.2060, II | DRQ | Amplifier and Signal | |
Conditioner, Transducer Signal | |||
$870.2300, II | DRT | Monitor, Cardiac (incl. | |
Cardiotachometer & Rate Alarm) | |||
$870.2300, II | MSX | System, Network and | |
Communication, Physiological | |||
Monitors | |||
$870.2340, II | DPS | Electrocardiograph | |
$870.2340, II | MLC | Monitor, ST Segment | |
$870.2350, II | DRW | Electrocardiograph, Lead | |
Switching Adapter | |||
$870.2370, II | KRC | Tester, Electrode, Surface, | |
Electrocardiograph | |||
$870.2450, II | DXJ | Display, Cathode-Ray Tube, | |
Medical | |||
$870.2600, I | DRJ | System, Signal Isolation | |
$870.2700, II | DQA | Oximeter | |
$870.2770, II | DSB | Plethysmograph, Impedance | |
$870.2800, II | DSH | Recorder, Magnetic tape, | |
Medical | |||
$870.2810, I | DSF | Recorder, Paper Chart | |
$870.2850, II | DRS | Extravascular Blood Pressure |
1
| | $870.2900, I | DSA | Cable, Transducer and
Electrode, incl. Patient
Connector |
|------------------------------------------------------|----------------|-----|------------------------------------------------------------------------------|
| | $870.2910, II | DRG | Transmitters and Receivers,
Physiological Signal,
Radiofrequency |
| Anesthesiology
and Respiratory
Therapy (12624) | $868.1400, II | CCK | Analyzer, Gas, Carbon Dioxide,
Gaseous-Phase |
| | $868.1500, II | CBQ | Analyzer, Gas, Enflurane,
Gaseous-Phase (Anesthetic
Concentration) |
| | $868.1500, II | NHO | Analyzer, Gas, Desflurane,
Gaseous-Phase (Anesthetic
Concentration) |
| | $868.1500, II | NHP | Analyzer, Gas, Sevoflurane,
Gaseous-Phase (Anesthetic
Concentration) |
| | $868.1500, II | NHQ | Analyzer, Gas, Isoflurane,
Gaseous-Phase (Anesthetic
Concentration) |
| | $868.1620, II | CBS | Analyzer, Gas, Halothane,
Gaseous-Phase (Anesthetic
Concentration) |
| | $868.1700, II | CBR | Analyzer, Gas, Nitrous Oxide,
Gaseous-Phase (Anesthetic
Concentration) |
| | $868.1720, II | CCL | Analyzer, Gas, Oxygen, Gaseous-
Phase |
| | $868.2375, II | BZQ | Monitor, Breathing Frequency |
| | $868.2480, II | LKD | Monitor, Carbon Dioxide,
Cutaneous |
| | $868.2500, II | KLK | Monitor, Oxygen, Cutaneous, for
Infant not under Gas Anesthesia |
| General Hospital
and Personal Use
(12520) | $880.2910, II | FLL | Thermometer, Electronic,
Clinical |
| Neurological
(12513) | $882.1400, II | GWR | Electroencephalograph |
| | $882.1420, I | GWS | Analyzer, Spectrum,
Electroencephalogram Signal |
-
- The modification is the introduction of software Release E.03 for the IntelliVue patient monitor devices.
-
- The modified devices have the same intended use as the legally marketed predicate devices. They are intended for the monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare facilities and during transport within healthcare facilities.
-
- The modified devices have the same technological characteristics as the legally marketed predicate devices.
2
K04.3315
P 3/3
-
- Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the modified device with respect to the predicate. Testing involved system level tests, performance tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence. The results demonstrate that the Philips IntelliVue Patient Monitor meets all reliability requirements and performance claims.
3
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, often used as a symbol of medicine. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC - 7 2006
Andreas Suchi Ouality and Regulatory Philips Medizinsysteme Boeblingen GMBH Hewlett-Packard-Str. 2 71034 Boblingen, Germany
Re: K063315
Trade/Device Name: INTELLIVUE PATIENT MONITORS, MODELS MP20, MP30, MP40, MP50, MP60, MP70, MP80 AND MP90 Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm). Regulatory Class: II
Product Code: MHX Dated: October 31, 2006 Received: November 2, 2006
Dear Mr. Suchi:
This letter corrects our substantially equivalent letter of November 20, 2006.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a the rederal 1 000, 2008). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your (1 will ), w that of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
4
Page 2 - Mr. Suchi
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
fr
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation
Center for Devices and Radiological Health
Sincerely vours.
5
Indications for Use
510(k) Number (if known): ドレビ331S
Device Name: The Philips MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90 IntelliVue Patient Monitors, Release E.03
Indications for Use: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics and neonates in healthcare facilities. The MP20, MP30, MP40 and MP50 are additionally intended for use in transport situations within healthcare facilities.
ST Segment monitoring is restricted to adult patients only.
The transcutaneous gas measurement (tcp02 / tcpCO2) is restricted to neonatal patients only.
Prescription Use yes___________________________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Dung R. be dines
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number_K063315
11
6
3.2 Philips Medical Systems Draft Labeling
The following labeling documents are found in Annex A:
- · A draft of the relevant chapters of the Instructions for Use for the Philips MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90 IntelliVue Patient Monitors
- o Chapter 13 Monitoring Cardiac Output
- o Chapter 27 Installation and Specifications
in the state of the time