(30 days)
Not Found
No
The summary explicitly states "Mentions AI, DNN, or ML: Not Found" and the device description focuses on calculating a numeric value (MAC) based on physiological parameters, which is a standard calculation, not indicative of AI/ML.
No.
The device is used for monitoring, recording, and alarming of physiological parameters, not for treating them.
No
The device is described as a patient monitor for physiological parameters, which is used for monitoring, recording, and alarming. While it provides "information about the condition of the patient (physiological monitoring)", it does not independently diagnose diseases or conditions. The output provided by the monitor is physiological data (e.g., ST segment, transcutaneous gas measurements, Minimum Alveolar Concentration), which healthcare professionals interpret in conjunction with other clinical information to make a diagnosis.
No
The device description explicitly states that the modification is the introduction of new software for existing patient monitor devices (Philips MP20, MP30, etc.). These patient monitors are hardware devices that measure physiological parameters. The software enhances the functionality of these hardware devices, but the device itself is not solely software.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use describes monitoring physiological parameters of patients in vivo (within the living body). This includes parameters like ST Segment, transcutaneous gas measurements, and other physiological data.
- Device Description: The device is a patient monitor that calculates and displays a numeric for Minimum Alveolar Concentration (MAC), which is a measure related to anesthetic gas concentration in the body.
- Lack of IVD Characteristics: There is no mention of analyzing samples taken from the body (like blood, urine, tissue) in vitro (outside the body). IVDs are designed to perform tests on these types of samples.
The device is clearly a patient monitoring system used for real-time assessment of a patient's physiological state, which falls under the category of medical devices used for diagnosis and monitoring in vivo, not in vitro.
N/A
Intended Use / Indications for Use
Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics and neonates in hospital environments. The MP20, MP30, MP40 and MP50 are additionally intended for use in transport situations within hospital environments.
ST Segment monitoring is restricted to adult patients only.
The transcutaneous gas measurement (tcp0 2 / tcpCO2) is restricted to neonatal patients only.
Product codes (comma separated list FDA assigned to the subject device)
DSI, MLD, MHX, DSJ, DSK, DXN, DXG, KRB, DRQ, DRT, MSX, DPS, MLC, DRW, KRC, DXJ, DRJ, DQA, DSB, DSH, DSF, DRS, DSA, DRG, CCK, CBQ, NHO, NHP, NHQ, CBS, CBR, CCL, BZQ, LKD, KLK, FLL, GWR, GWS, MXH
Device Description
The names of the devices are the Philips MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90 IntelliVue Patient Monitors.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
adults, pediatrics and neonates
Intended User / Care Setting
health care professionals / hospital environments and during transport within hospital environments
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the modified device with respect to the predicate. Testing involved system level tests, performance tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence. The results demonstrate that the Philips IntelliVue Patient Monitor meets all reliability requirements and performance claims.
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Not Found
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
K014159, K021778, K030038, K032858, K040304, K040183, K040357, K041235, K41741, K042845, K050141, 050762, ко51106
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).
(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
0
510 (k) Summary
NOV - 2 2005
This summary of 510 (k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.
. . . .
- The submitter of this premarket notification is: Hauke Schik Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany Tel: ++49 7031 463-2031 Fax: ++49 7031 463-2442 e-mail: hauke.schik@philips.com
This summary was prepared on September 30, 2005.
-
- The names of the devices are the Philips MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90 IntelliVue Patient Monitors. Classification names are as follows:
Device Panel | Classification | ProCode | Description |
---|---|---|---|
Circulatory | |||
System Devices | |||
(12625) | $870.1025, II | DSI | Detector and alarm, arrhythmia |
$870.1025, II | MLD | Monitor, ST Segment with Alarm | |
$870.1025, II | MHX | Monitor, Physiological, Patient | |
(with arrhythmia detection or | |||
alarms) | |||
$870.1100, II | DSJ | Alarm, Blood Pressure | |
$870.1110, II | DSK | Computer, Blood Pressure | |
$870.1130, II | DXN | System, Measurement, Blood- | |
Pressure, Non-Invasive | |||
$870.1435, II | DXG | Computer, Diagnostic, Pre- | |
Programmed, Single-Function | |||
$870.1915, II | KRB | Probe, Thermodilution | |
$870.2060, II | DRQ | Amplifier and Signal | |
Conditioner, Transducer Signal | |||
$870.2300, II | DRT | Monitor, Cardiac (incl. | |
Cardiotachometer & Rate Alarm) | |||
$870.2300, II | MSX | System, Network and | |
Communication, Physiological | |||
Monitors | |||
$870.2340, II | DPS | Electrocardiograph | |
$870.2340, II | MLC | Monitor, ST Segment | |
$870.2350, II | DRW | Electrocardiograph, Lead | |
Switching Adapter | |||
$870.2370, II | KRC | Tester, Electrode, Surface, | |
Electrocardiograph | |||
$870.2450, II | DXJ | Display, Cathode-Ray Tube, | |
Medical | |||
$870.2600, I | DRJ | System, Signal Isolation | |
$870.2700, II | DQA | Oximeter | |
$870.2770, II | DSB | Plethysmograph, Impedance | |
$870.2800, II | DSH | Recorder, Magnetic tape, | |
Medical | |||
$870.2810, I | DSF | Recorder, Paper Chart | |
$870.2850, II | DRS | Extravascular Blood Pressure | |
Transducer |
1
| $870.2900, I | DSA | Cable, Transducer and
Electrode, incl. Patient
Connector | |
|------------------------------------------------------|----------------|------------------------------------------------------------------------|------------------------------------------------------------------------------|
| $870.2910, II | DRG | Transmitters and Receivers,
Physiological Signal,
Radiofrequency | |
| Anesthesiology
and Respiratory
Therapy (12624) | $868.1400, II | CCK | Analyzer, Gas, Carbon Dioxide,
Gaseous-Phase |
| | $868.1500, II | CBQ | Analyzer, Gas, Enflurane,
Gaseous-Phase (Anesthetic
Concentration) |
| | $868.1500, II | NHO | Analyzer, Gas, Desflurane,
Gaseous-Phase (Anesthetic
Concentration) |
| | $868.1500, II | NHP | Analyzer, Gas, Sevoflurane,
Gaseous-Phase (Anesthetic
Concentration) |
| | $868.1500, II | NHQ | Analyzer, Gas, Isoflurane,
Gaseous-Phase (Anesthetic
Concentration) |
| | $868.1620, II | CBS | Analyzer, Gas, Halothane,
Gaseous-Phase (Anesthetic
Concentration) |
| | $868.1700, II | CBR | Analyzer, Gas, Nitrous Oxide,
Gaseous-Phase (Anesthetic
Concentration) |
| | $868.1720, II | CCL | Analyzer, Gas, Oxygen, Gaseous-
Phase |
| | $868.2375, II | BZQ | Monitor, Breathing Frequency |
| | $868.2480, II | LKD | Monitor, Carbon Dioxide,
Cutaneous |
| | $868.2500, II | KLK | Monitor, Oxygen, Cutaneous, for
Infant not under Gas Anesthesia |
| General Hospital
and Personal Use
(12520) | $880.2910, II | FLL | Thermometer, Electronic,
Clinical |
| Neurological
(12513) | $882.1400, II | GWR | Electroencephalograph |
| | $882.1420, I | GWS | Analyzer, Spectrum,
Electroencephalogram Signal |
-
- The modification is the introduction of Release C.02 software for the IntelliVue patient monitor devices, which enhances the monitor to calculate and display a numeric for the Minimum Alveolar Concentration (MAC) .
-
- The modified devices have the same intended use as the legally marketed predicate devices. They are intended for the monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in hospital environment and during transport within hospital environments.
2
-
- The modified devices have the same technological characteristics as the legally marketed predicate devices.
-
- Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the modified device with respect to the predicate. Testing involved system level tests, performance tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence. The results demonstrate that the Philips IntelliVue Patient Monitor meets all reliability requirements and performance claims.
3
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV - 2 2005
Mr. Hauke Schik Senior Regulatory Affairs Engineer Philips Medizinsysteme Boeblingen GmbH Cardiac and Monitoring Systems Hewlett-Packard Str. 2, D-71034 Boeblingen, GERMANY
Re: K052801
Trade/Device Name: The Phillips MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90 IntelliVue Patient Monitors, Release C.02 Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (including ST-segment measurement and alarm) Regulatory Class: II Product Code: MXH Dated: September 30, 2005 Received: November 2, 2005
Dear Mr. Schik:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
4
Page 2 - Mr. Schik
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Cluts
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
5
Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: The Philips MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90 IntelliVue Patient Monitors, Release C.02
Indications for Use: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics and neonates in hospital environments. The MP20, MP30, MP40 and MP50 are additionally intended for use in transport situations within hospital environments.
ST Segment monitoring is restricted to adult patients only.
The transcutaneous gas measurement (tcp0 2 / tcpCO2) is restricted to neonatal patients only.
Prescription Use Over-The-Counter Use yes No AND/OR ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------(Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Quen Suliom
sion of Anesthesiology, General Hospital, infoction Control, Dental Devices
- (2(k) Number