(28 days)
No
The summary explicitly states "Mentions AI, DNN, or ML: Not Found" and the description focuses on software updates for interfacing with a gas analyzer, not on AI/ML capabilities.
No.
The device is described as a patient monitor intended for monitoring, recording, and alarming of physiological parameters, not for providing therapy or treatment.
No
Explanation: The device is described as a patient monitor that monitors, records, and alarms multiple physiological parameters. While this information is useful for healthcare professionals, the primary purpose is monitoring, not diagnosing a disease or condition. The document also states that it is for "monitoring the physiological parameters" and "recording and alarming of multiple physiological parameters." There is no mention of output that helps in forming a diagnosis.
No
The device description explicitly states the names of the devices are "Philips MP20, MP30, MP40, MP50, MP60, MP70, and MP90 IntelliVue Patient Monitor," which are hardware devices. The modification is a software release for these existing hardware monitors.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use describes monitoring physiological parameters of patients (ECG, ST Segment, transcutaneous gas measurement). This involves measuring signals directly from the patient's body, not analyzing samples taken from the body (like blood, urine, or tissue).
- Device Description: The device is a patient monitor that interfaces with a gas analyzer. While the gas analyzer might be involved in measuring gases, the overall system is focused on monitoring the patient's physiological state.
- Lack of IVD Characteristics: The description does not mention any analysis of biological samples, reagents, or diagnostic tests performed on samples outside of the body.
IVD devices are specifically designed to examine specimens derived from the human body to provide information for diagnostic, monitoring, or compatibility purposes. This device's function is to monitor the patient's vital signs directly.
N/A
Intended Use / Indications for Use
The modified devices have the same intended use as the leqally marketed predicate devices. They are intended for the monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in hospital environment and during transport within hospital environments.
Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters patients. Intended for monitoring, recording and alarming of of multiple physiological parameters of adults, pediatrics and neonates in hospital environments. The MP20, MP30, MP40 and MP50 are additionally intended for use in transport situations within hospital environments.
EASI 12-lead ECG is only for use on adult and pediatric patients.
ST Seqment monitoring is restricted to adult patients only.
The transcutaneous gas measurement (tcp02/tcpCO2) is restricted to neonatal patients only.
Product codes (comma separated list FDA assigned to the subject device)
DSI, MLD, MHX, DSJ, DSK, DXN, DXG, KRB, DRQ, DRT, DPS, MLC, DRW, KRC, DXJ, DRJ, DQA, DSB, DSH, DSF, DRS, DSA, MSX, CCK, CBQ, NHO, NHP, NHQ, CBS, CBR, CCL, BZQ, LKD, KLK, FLL, GWR, GWS
Device Description
The modification is the introduction of Release B.1.1 software for the IntelliVue patient monitor devices, MP20, MP30, MP40, MP50 interfacing the Gas Analyzer model M1013A EGM (K041956).
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
adults, pediatrics, and neonates
Intended User / Care Setting
health care professionals in hospital environment and during transport within hospital environments.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the modified device with respect to the predicate. Testing involved system level tests, performance tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence.
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Not Found
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
K014159, K021778, K032858, K040304, K040183, K040259, K040357, K041235
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).
(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).
0
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.
KO42845
- The submitter of this premarket notification is: Egon Pfeil Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany Tel: ++49 7031 463-1926 Fax: ++49 7031 463-2442 e-mail: egon.pfeil@philips.com
This summary was prepared on October 13, 2004.
-
- The names of the devices are the Philips MP20, MP30, MP40, MP50, MP60, MP70, and MP90 IntelliVue Patient Monitor. Classification names are as follows:
Device Panel | Classification | ProCode | Description |
---|---|---|---|
Circulatory | |||
System Devices | |||
(12625) | $870.1025, III | DSI | Detector and alarm, arrhythmia |
$870.1025, III | MLD | Monitor, ST Segment with Alarm | |
$870.1025, III | MHX | Monitor, Physiological, Patient | |
(with arrhythmia detection or | |||
alarms) | |||
$870.1100, II | DSJ | Alarm, Blood Pressure | |
$870.1110, II | DSK | Computer, Blood Pressure | |
$870.1130, II | DXN | System, Measurement, Blood- | |
Pressure, Non-Invasive | |||
$870.1435, II | DXG | Computer, Diagnostic, Pre- | |
Programmed, Single-Function | |||
$870.1915, II | KRB | Probe, Thermodilution | |
$870.2060, II | DRQ | Amplifier and Signal | |
Conditioner, Transducer Signal | |||
$870.2300, II | DRT | Monitor, Cardiac (incl. | |
Cardiotachometer & Rate Alarm) | |||
$870.2340, II | DPS | Electrocardiograph | |
$870.2340, II | MLC | Monitor, ST Segment | |
$870.2350, II | DRW | Electrocardiograph, Lead | |
Switching Adapter | |||
$870.2370, II | KRC | Tester, Electrode, Surface, | |
Electrocardiograph | |||
$870.2450, II | DXJ | Display, Cathode-Ray Tube, | |
Medical | |||
$870.2600, I | DRJ | System, Signal Isolation | |
$870.2700, II | DQA | Oximeter | |
$870.2770, II | DSB | Plethysmograph, Impedance | |
$870.2800, II | DSH | Recorder, Magnetic tape, | |
Medical | |||
$870.2810, I | DSF | Recorder, Paper Chart | |
$870.2850, II | DRS | Extravascular Blood Pressure | |
Transducer | |||
$870.2900, I | DSA | Cable, Transducer and | |
Electrode, incl. Patient | |||
Connector |
1