(44 days)
Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording and alparming of multiple physiological parameters of adults, pediatrics and neonates in healthcare facilities. The MP20, MP30, MP40 and MP50. are additionally intended for use in transport situations within healthcare facilities.
ST Segment monitoring is restricted to adult patients only.
The transcutaneous gas measurement (tcp02 / tcpC02) is restricted to neonatal patients only.
The Philips MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90 IntelliVue Patient Monitors. The modification is the introduction of Release E.01 software for the IntelliVue patient monitor devices.
Here's a breakdown of the acceptance criteria and study information for the Philips IntelliVue Patient Monitors, Release E.01, based on the provided 510(k) summary:
1. Table of Acceptance Criteria and Reported Device Performance
The provided 510(k) summary does not explicitly list quantitative acceptance criteria for specific performance metrics (e.g., accuracy, sensitivity, specificity) of the device's functions. Instead, it makes a general statement about meeting predetermined specifications.
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Specifications cleared for the predicate device were met. | "The results demonstrate that the Philips IntelliVue Patient Monitor meets all reliability requirements and performance claims." |
It's important to note that for a medical device such as a patient monitor, "specifications cleared for the predicate device" would typically encompass a wide range of performance parameters for each physiological measurement (e.g., ECG accuracy, NIBP accuracy, SpO2 accuracy, arrhythmia detection performance). However, these specific numerical criteria are not detailed in this summary.
2. Sample Size Used for the Test Set and Data Provenance
The 510(k) summary does not provide specific details on the sample size used for the test set or the data provenance (e.g., country of origin, retrospective/prospective). It only states:
- "Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the modified device with respect to the predicate."
- "Testing involved system level tests, performance tests, and safety testing from hazard analysis."
Without further details, we cannot infer the nature or origin of the test data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The 510(k) summary does not provide information on the number or qualifications of experts used to establish ground truth for any test set. Given the nature of a patient monitor, ground truth for physiological parameters would likely come from validated reference devices or established clinical methods, rather than expert interpretation of raw data.
4. Adjudication Method for the Test Set
The 510(k) summary does not specify any adjudication method for a test set. This type of method (e.g., 2+1, 3+1 consensus) is typically used for subjective assessment tasks, such as image interpretation, which are not directly applicable to the functional testing of a patient monitor.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not discussed or indicated in the provided 510(k) summary. MRMC studies are primarily relevant for diagnostic imaging devices where human readers interpret results, often with and without AI assistance, to assess changes in diagnostic performance. This device is a patient monitor, not an AI-assisted diagnostic imaging tool in that context.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
The testing described is inherently standalone for the monitoring functionalities. A patient monitor's algorithms continuously process physiological signals without direct human intervention in the data processing itself. The summary implies "system level tests, performance tests, and safety testing," which would evaluate the device's algorithms and hardware independently of a human operator, while acknowledging that a human uses the displayed information.
7. The Type of Ground Truth Used
While not explicitly stated for each parameter, the ground truth for a patient monitor's performance tests would typically involve:
- Reference Devices: Using highly accurate and calibrated reference devices for each physiological parameter (e.g., an invasive blood pressure catheter for NIBP accuracy, a calibrated ECG simulator for arrhythmia detection, a gas analyzer for capnography).
- Clinical Measurements: Comparing monitor readings to gold-standard clinical measurements performed by trained professionals.
- Simulated Data: For certain performance aspects like arrhythmia detection, using validated physiological waveforms that contain known arrhythmias.
8. The Sample Size for the Training Set
The 510(k) summary does not mention a training set or its sample size. This is because the device, the Philips IntelliVue Patient Monitor, is a traditional physiological monitoring device, not an AI/machine learning device that requires a distinct "training set" to develop its algorithms. Its algorithms are based on established physiological principles and signal processing techniques.
9. How the Ground Truth for the Training Set Was Established
As no training set is mentioned or implied for this type of device, the concept of establishing ground truth for a training set is not applicable here. The algorithms for a patient monitor are developed based on scientific and medical principles, and then validated (tested) against known reference standards.
{0}------------------------------------------------
510 (k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.
- The submitter of this premarket notification is: Andreas Suchi Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany . Tel: ++49 7031 463-1291 Fax: ++49 7031 463-2442 e-mail: andreas.suchi@philips.com
This summary was prepared on August 4, 2006.
-
- The names of the devices are the Philips MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90 IntelliVue Patient Monitors. Classification names are as follows:
| Device Panel | Classification | ProCode | Description |
|---|---|---|---|
| CirculatorySystem Devices(12625) | $870.1025, II | DSI | Detector and alarm, arrhythmia |
| $870.1025, II | MLD | Monitor, ST Segment with Alarm | |
| $870.1025, II | MHX | Monitor, Physiological, Patient(with arrhythmia detection oralarms) | |
| $870.1100, II | DSJ | Alarm, Blood Pressure | |
| $870.1110, II | DSK | Computer, Blood Pressure | |
| $870.1130, II | DXN | System, Measurement, Blood-Pressure, Non-Invasive | |
| $870.1435, II | DXG | Computer, Diagnostic, Pre-Programmed, Single-Function | |
| $870.1915, II | KRB | Probe, Thermodilution | |
| $870.2060, II | DRQ | Amplifier and SignalConditioner, Transducer Signal | |
| $870.2300, II | DRT | Monitor, Cardiac (incl.Cardiotachometer & Rate Alarm) | |
| $870.2300, II | MSX | System, Network andCommunication, PhysiologicalMonitors | |
| $870.2340, II | DPS | Electrocardiograph | |
| $870.2340, II | MLC | Monitor, ST Segment | |
| $870.2350, II | DRW | Electrocardiograph, LeadSwitching Adapter | |
| $870.2370, II | KRC | Tester, Electrode, Surface,Electrocardiograph | |
| $870.2450, II | DXJ | Display, Cathode-Ray Tube,Medical | |
| $870.2600, I | DRJ | System, Signal Isolation | |
| $870.2700, II | DQA | Oximeter | |
| $870.2770, II | DSB | Plethysmograph, Impedance | |
| $870.2800, II | DSH | Recorder, Magnetic tape,Medical | |
| $870.2810, I | DSF | Recorder, Paper Chart | |
| $870.2850, II | DRS | Extravascular Blood PressureTransducer |
SEP 2 0 2006
{1}------------------------------------------------
| $870.2900, I | DSA | Cable, Transducer and Electrode, incl. Patient Connector | |
|---|---|---|---|
| $870.2910, II | DRG | Transmitters and Receivers, Physiological Signal, Radiofrequency | |
| Anesthesiologyand RespiratoryTherapy (12624) | $868.1400, II | CCK | Analyzer, Gas, Carbon Dioxide, Gaseous-Phase |
| $868.1500, II | CBQ | Analyzer, Gas, Enflurane, Gaseous-Phase (Anesthetic Concentration) | |
| $868.1500, II | NHO | Analyzer, Gas, Desflurane, Gaseous-Phase (Anesthetic Concentration) | |
| $868.1500, II | NHP | Analyzer, Gas, Sevoflurane, Gaseous-Phase (Anesthetic Concentration) | |
| $868.1500, II | NHQ | Analyzer, Gas, Isoflurane, Gaseous-Phase (Anesthetic Concentration) | |
| $868.1620, II | CBS | Analyzer, Gas, Halothane, Gaseous-Phase (Anesthetic Concentration) | |
| $868.1700, II | CBR | Analyzer, Gas, Nitrous Oxide, Gaseous-Phase (Anesthetic Concentration) | |
| $868.1720, II | CCL | Analyzer, Gas, Oxygen, Gaseous-Phase | |
| $868.2375, II | BZQ | Monitor, Breathing Frequency | |
| $868.2480, II | LKD | Monitor, Carbon Dioxide, Cutaneous | |
| $868.2500, II | KLK | Monitor, Oxygen, Cutaneous, for Infant not under Gas Anesthesia | |
| General Hospitaland Personal Use(12520) | $880.2910, II | FLL | Thermometer, Electronic, Clinical |
| Neurological(12513) | $882.1400, II | GWR | |
| $882.1420, I | GWS | Analyzer, Spectrum, Electroencephalogram Signal |
-
- The modification is the introduction of Release E.01 software for the IntelliVue patient monitor devices.
-
- The modified devices have the same intended use as the legally marketed predicate devices. They are intended for the monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare facilities and during transport within healthcare facilities.
-
- The modified devices have the same technological characteristics as the legally marketed predicate devices.
{2}------------------------------------------------
-
- Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the modified device with respect to the predicate. Testing involved system level tests, performance tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence. The results demonstrate that the Philips IntelliVue Patient Monitor meets all reliability requirements and performance claims.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wing-like shapes, representing health, services, and people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 2 0 2006
Philips Medizin Systeme Boblingen GmbH c/o Ms. Linda Stone Philips Medical Systems 3000 Minuteman Road Andover. MA 01810
Re: K062283
Trade Name: The Philips MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90 IntelliVue Patient Monitors, Release E.01 Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (including ST-segment measurement and alarm Regulatory Class: Class II (two) Product Code: MHX Dated: August 21, 2006 Received: August 23, 2006
Dear Ms. Stone:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Linda Stone
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantials squisvalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bram D. Zuckerman, M.D.
Director
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________
Device Name: The Philips MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90 IntelliVue Patient Monitors, Release E.01
Indications for Use: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording and alparming of multiple physiological parameters of adults, pediatrics and neonates in healthcare facilities. The MP20, MP30, MP40 and MP50. are additionally intended for use in transport situations within healthcare facilities.
ST Segment monitoring is restricted to adult patients only.
The transcutaneous gas measurement (tcp02 / tcpC02) is restricted to neonatal patients only.
Prescription Use (Part 21 CFR 801 Subpart D)
yes___________________________________________________________________________________________________________________________________________________________________________
AND/OR
Over-The-Counter Use No (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Bhumanna
§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).
(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.