(65 days)
No
The summary describes a hardware modification for wireless connectivity and does not mention any AI/ML capabilities or related performance metrics.
No
This device is a patient monitor, which is used for monitoring and recording physiological parameters, not for treating a condition.
No.
The device is indicated for monitoring, recording, and alarming of physiological parameters, which is a patient management function rather than a diagnostic one.
No
The device description explicitly states that the modification is primarily a hardware-based change involving an externally mounted wireless network connection to existing patient monitor devices.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly states monitoring physiological parameters of patients. This involves measuring vital signs and other bodily functions directly from the patient, not analyzing samples taken from the patient (which is the core of IVD).
- Device Description: The device is described as a patient monitor that measures physiological parameters. The modification is related to network connectivity, not the analysis of biological samples.
- Lack of IVD Indicators: There is no mention of analyzing biological samples (blood, urine, tissue, etc.), reagents, or any processes typically associated with in vitro diagnostics.
Therefore, this device falls under the category of a patient monitoring device, not an IVD.
N/A
Intended Use / Indications for Use
Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics and neonates in hospital environments.
Product codes
DSI, MLD, MHX, DSJ, DSK, DXN, DXG, KRB, DRQ, DRT, DPS, MLC, DRW, KRC, DXJ, DRJ, DQA, DSB, DSH, DSF, DRS, DSA, MSX, CCK, CBQ, NHO, NHP, NHQ, CBS, CBR, CCL, BZQ, LKD, KLK, FLL, GWR, GWS
Device Description
The modification is primarily a hardware based change that offers, as an option, the addition of a an externally mounted wireless network connection to the Philips Medical System MP60, MP70 and MP90 IntelliVue patient monitor devices.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
adults, pediatrics, and neonates
Intended User / Care Setting
health care professionals / hospital environments
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Verification testing activities were conducted to establish the performance and reliability characteristics of the new device. Testing involved functional level tests and safety testing from the risk analysis.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).
(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
0
MAR 1 2 2003
510 (k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.
- The submitter of this premarket notification is: David Osborn Philips Medical Systems 3000 Minuteman Road Andover, MA 01810-1099 Tel: 978 659 3178 Fax: 978 685 5624 e-mail: d.g.osborn@ieee.org
This summary was prepared on December 31, 2002.
-
- The names of the devices are the Philips MP60, MP70, and MP90 IntelliVue Patient Monitor with Portal Technology and Wireless LAN. Classification names are as follows:
Device Panel | Classification | ProCode | Description |
---|---|---|---|
Circulatory | |||
System Devices | |||
(12625) | $870.1025, III | DSI | Detector and alarm, arrhythmia |
$870.1025, III | MLD | Monitor, ST Segment with Alarm | |
$870.1025, III | MHX | Monotor, Physiological, Patient | |
(with arrhythmia detection or | |||
alarms) | |||
$870.1100, II | DSJ | Alarm, Blood Pressure | |
$870.1110, II | DSK | Computer, Blood Pressure | |
$870.1130, II | DXN | System, Measurement, Blood- | |
Pressure, Non-Invasive | |||
$870.1435, II | DXG | Computer, Diagnostic, Pre- | |
Programmed, Single-Function | |||
$870.1915, II | KRB | Probe, Thermodilution | |
$870.2060, II | DRQ | Amplifier and Signal | |
Conditioner, Transducer Signal | |||
$870.2300, II | DRT | Monitor, Cardiac (incl. | |
Cardiotachometer & Rate Alarm) | |||
$870.2340, II | DPS | Electrocardiograph | |
$870.2340, II | MLC | Monitor, ST Segment | |
$870.2350, II | DRW | Electrocardiograph, Lead | |
Switching Adapter | |||
$870.2370, II | KRC | Tester, Electrode, Surface, | |
Electrocardiograph | |||
$870.2450, II | DXJ | Display, Cathode-Ray Tube, | |
Medical | |||
$870.2600, I | DRJ | System, Signal Isolation | |
$870.2700, II | DQA | Oximeter | |
$870.2770, II | DSB | Plethysmograph, Impedance | |
$870.2800, II | DSH | Recorder, Magnetic tape, | |
Medical | |||
$870.2810, I | DSF | Recorder, Paper Chart | |
$870.2850, II | DRS | Extravascular Blood Pressure | |
Transducer | |||
$870.2900, I | DSA | Cable, Transducer and | |
Electrode, incl. Patient | |||
Connector | |||
- | MSX | System, Network and | |
Communication, Physiological | |||
Monitors |
1
| Anesthesiology
and Respiratory
Therapy (12624) | $868.1400, II | CCK | Analyzer, Gas, Carbon Dioxide,
Gaseous-Phase |
|------------------------------------------------------|----------------|-----|------------------------------------------------------------------------------|
| | $868.1500, II | CBQ | Analyzer, Gas, Enflurane,
Gaseous-Phase (Anesthetic
Concentration) |
| | $868.1500, II | NHO | Analyzer, Gas, Desflurane,
Gaseous-Phase (Anesthetic
Concentration) |
| | $868.1500, II | NHP | Analyzer, Gas, Sevoflurane,
Gaseous-Phase (Anesthetic
Concentration) |
| | $868.1500, II | NHQ | Analyzer, Gas, Isoflurane,
Gaseous-Phase (Anesthetic
Concentration) |
| | $868.1620, II | CBS | Analyzer, Gas, Halothane,
Gaseous-Phase (Anesthetic
Concentration) |
| | $868.1700, II | CBR | Analyzer, Gas, Nitrous Oxide,
Gaseous-Phase (Anesthetic
Concentration) |
| | $868.1720, II | CCL | Analyzer, Gas, Oxygen, Gaseous-
Phase |
| | $868.2375, II | BZQ | Monitor, Breathing Frequency |
| | $868.2480, II | LKD | Monitor, Carbon Dioxide,
Cutaneous |
| | $868.2500, II | KLK | Monitor, Oxygen, Cutaneous, for
Infant not under Gas Anesthesia |
| General Hospital
and Personal Use
(12520) | $880.2910, II | FLL | Thermometer, Electronic,
Clinical |
| Neurological
(12513) | $882.1400, II | GWR | Electroencephalograph |
| | $882.1420, I | GWS | Analyzer, Spectrum,
Electroencephalogram Signal |
- The modification is primarily a hardware based change that offers, as 4 . an option, the addition of a an externally mounted wireless network connection to the Philips Medical System MP60, MP70 and MP90 IntelliVue patient monitor devices.
-
- The new devices have the same intended use as the legally marketed predicate devices. When used in the hospital environment, they are intended for the monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates.
-
- The new devices have the same technological characteristics as the leqally marketed predicate devices.
-
- Verification testing activities were conducted to establish the performance and reliability characteristics of the new device. Testing involved functional level tests and safety testing from the risk analysis.
2
Public Health Service
Image /page/2/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three swooping lines forming its body and wings. The eagle is enclosed in a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper half of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 2 2003
Phillips Medical Systems, Inc. c/o Mr. David Osborn Ouality Program Manager 3000 Minuteman Road Andover, MA 01810-1099
Re: K030038
Trade/Device Name: IntelliVue Models MP60, MP70 and MP90 Patient Monitors with Portal Technology and M3185 opt. C41 Wireless Bedside Adapter Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm Regulatory Class: Class III (three) Product Code: MHX Dated: February 24, 2003 Received: February 25, 2003
Dear Mr. Osborn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
3
Page 2 – Mr. David Osborn
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Bram D. Zuckerman, M.D.
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
4
Indications for Use
Page
of
KO30038 510(k) Number (if known):
Device Name: The Philips IntelliVue MP60, MP70, and MP90 Patient Monitors with Portal Technology and M3185 opt C41 Wireless Bedside Adapter.
Indications for Use: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics and neonates in hospital environments.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
(Optional Format 1-2-96)
OR
Over-The-Counter
VCelkatil
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number_
9