(17 days)
Not Found
No
The document explicitly states "Mentions AI, DNN, or ML: Not Found" and provides no other indication of AI/ML use.
No
The device is described as a patient monitor, which is used for monitoring, recording, and alarming of physiological parameters, not for treating conditions.
No
Explanation: The device is described as a patient monitor that tracks physiological parameters and alarms for them. While monitoring and alarming are part of patient care, directly diagnosing a condition is not stated as its primary function. It provides data that a healthcare professional might use for diagnosis, but it does not perform the diagnosis itself.
No
The device description explicitly states that the modification is the introduction of new software for existing patient monitor devices (Philips MP20, MP30, MP40, MP50, MP60, MP70, and MP90 IntelliVue Patient Monitors) and incorporating a wireless interface. This indicates the software is part of a larger hardware system (the patient monitors) and not a standalone software-only device.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use describes monitoring physiological parameters of patients. This involves measuring vital signs and other bodily functions directly from the patient, not analyzing samples of bodily fluids or tissues in vitro (outside the body).
- Device Description: The description refers to patient monitors, which are devices used to observe and display a patient's physiological data in real-time.
- Lack of IVD Characteristics: There is no mention of analyzing samples, reagents, or any processes typically associated with in vitro diagnostics.
IVD devices are specifically designed to examine specimens derived from the human body (like blood, urine, tissue) to provide information for diagnosis, monitoring, or screening. This device's function is to monitor the patient directly.
N/A
Intended Use / Indications for Use
Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics and neonates in hospital environments. The MP20, MP30, MP40 and MP50 are intended for use in transport situations within additionally hospital environments.
ST Segment monitoring is restricted to adult patients only.
The transcutaneous gas measurement (tcp02 / tcpCO2) is restricted to neonatal patients only.
Product codes (comma separated list FDA assigned to the subject device)
DSI, MLD, MHX, DSJ, DSK, DXN, DXG, KRB, DRQ, DRT, MSX, DPS, MLC, DRW, KRC, DXJ, DRJ, DQA, DSB, DSH, DSF, DRS, DSA, DRG, CCK, CBQ, NHO, NHP, NHQ, CBS, CBR, CCL, BZQ, LKD, KLK, FLL, GWR, GWS
Device Description
The modification is the introduction of Release B.1.2 software for the IntelliVue patient monitor devices and incorporating a ITS wireless interface.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
adults, pediatrics and neonates
Intended User / Care Setting
health care professionals, hospital environments, transport within hospital environments
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the modified device with respect to the predicate. Testing involved system level tests, performance tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence. The results demonstrate that the Philips IntelliVue Patient Monitor meets all reliability requirements and performance claims.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
K014159, K021778, K030038, K032858, K040304, K040183, K040357, K041235, K41741, K042845
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).
(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
0
510 (k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Sate Medical Devices Act of 1990 and 21 C.F.R. §807.92.
- The submitter of this premarket notification is: Hauke Schik Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany Tel: ++49 7031 463-2031 Fax: ++49 7031 463-2442 e-mail: hauke.schik@philips.com
This summary was prepared on January 21, 2005.
-
- The names of the devices are the Philips MP20, MP30, MP40, MP50, MP60, MP70, and MP90 IntelliVue Patient Monitors. Classification names are as follows:
Device Panel | Classification | ProCode | Description |
---|---|---|---|
Circulatory | |||
System Devices | |||
(12625) | $870.1025, II | DSI | Detector and alarm, arrhythmia |
$870.1025, II | MLD | Monitor, ST Segment with Alarm | |
$870.1025, II | MHX | Monitor, Physiological, Patient | |
(with arrhythmia detection or | |||
alarms) | |||
$870.1100, II | DSJ | Alarm, Blood Pressure | |
$870.1110, II | DSK | Computer, Blood Pressure | |
$870.1130, II | DXN | System, Measurement, Blood- | |
Pressure, Non-Invasive | |||
$870.1435, II | DXG | Computer, Diagnostic, Pre- | |
Programmed, Single-Function | |||
$870.1915, II | KRB | Probe, Thermodilution | |
$870.2060, II | DRQ | Amplifier and Signal | |
Conditioner, Transducer Signal | |||
$870.2300, II | DRT | Monitor, Cardiac (incl. | |
Cardiotachometer & Rate Alarm) | |||
$870.2300, II | MSX | System, Network and | |
Communication, Physiological | |||
Monitors | |||
$870.2340, II | DPS | Electrocardiograph | |
$870.2340, II | MLC | Monitor, ST Segment | |
$870.2350, II | DRW | Electrocardiograph, Lead | |
Switching Adapter | |||
$870.2370, II | KRC | Tester, Electrode, Surface, | |
Electrocardiograph | |||
$870.2450, II | DXJ | Display, Cathode-Ray Tube, | |
Medical | |||
$870.2600, I | DRJ | System, Signal Isolation | |
$870.2700, II | DQA | Oximeter | |
$870.2770, II | DSB | Plethysmograph, Impedance | |
$870.2800, II | DSH | Recorder, Magnetic tape, | |
Medical | |||
$870.2810, I | DSF | Recorder, Paper Chart | |
$870.2850, II | DRS | Extravascular Blood Pressure | |
Transducer |
1
| | $870.2900, I | DSA | Cable, Transducer and
Electrode, incl. Patient
Connector |
|------------------------------------------------------|----------------|-----|------------------------------------------------------------------------------|
| | $870.2910, II | DRG | Transmitters and Receivers,
Physiological Signal,
Radiofrequency |
| Anesthesiology
and Respiratory
Therapy (12624) | $868.1400, II | CCK | Analyzer, Gas, Carbon Dioxide,
Gaseous-Phase |
| | $868.1500, II | CBQ | Analyzer, Gas, Enflurane,
Gaseous-Phase (Anesthetic
Concentration) |
| | $868.1500, II | NHO | Analyzer, Gas, Desflurane,
Gaseous-Phase (Anesthetic
Concentration) |
| | $868.1500, II | NHP | Analyzer, Gas, Sevoflurane,
Gaseous-Phase (Anesthetic
Concentration) |
| | $868.1500, II | NHQ | Analyzer, Gas, Isoflurane,
Gaseous-Phase (Anesthetic
Concentration) |
| | $868.1620, II | CBS | Analyzer, Gas, Halothane,
Gaseous-Phase (Anesthetic
Concentration) |
| | $868.1700, II | CBR | Analyzer, Gas, Nitrous Oxide,
Gaseous-Phase (Anesthetic
Concentration) |
| | $868.1720, II | CCL | Analyzer, Gas, Oxygen, Gaseous-
Phase |
| | $868.2375, II | BZQ | Monitor, Breathing Frequency |
| | $868.2480, II | LKD | Monitor, Carbon Dioxide,
Cutaneous |
| | $868.2500, II | KLK | Monitor, Oxygen, Cutaneous, for
Infant not under Gas Anesthesia |
| General Hospital
and Personal Use
(12520) | $880.2910, II | FLL | Thermometer, Electronic,
Clinical |
| Neurological
(12513) | $882.1400, II | GWR | Electroencephalograph |
| | $882.1420, I | GWS | Analyzer, Spectrum |
-
- The modification is the introduction of Release B.1.2 software for the IntelliVue patient monitor devices and incorporating a ITS wireless interface.
-
- The modified devices have the same intended use as the legally marketed predicate devices. They are intended for the monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in hospital environment and during transport within hospital environments.
-
- The modified devices have the same technological characteristics as the legally marketed predicate devices.
Electroencephalogram Signal
2
K050141
P³/₃
-
- Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the modified device with respect to the predicate. Testing involved system level tests, performance tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence. The results demonstrate that the Philips IntelliVue Patient Monitor meets all reliability requirements and performance claims.
3
Image /page/3/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. In the center of the seal is an abstract image of an eagle.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 1 2005
Phillips Medizin Systeme Boeblingen GmbH c/o Mr. Hauke Schik Sr. Regulatory Affairs Engineer Hewlett-Packard Str. 2 D-71034 Boeblingen GERMANY
Re: K050141
Trade Name: The Phillips MP20, MP30, MP40, MP50, MP60, MP70 and MP90 Intellivue Patient Monitors, Release B.1.2. Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (including ST-segment measurement and alarm) Regulatory Class: II (two) Product Code: MHX Dated: January 21, 2005 Received: January 24 2005
Dear Mr. Hauke Schik:
We have reviewed your Section 510(k) premarket notification of intent to market the device w & no re review a your we determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, de nees mat have been require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
4
Page 2 - Mr. Hauke Schik
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Brimmer for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
5
Indications for Use
510(k) Number (if known):_ Ko5o14 l
Device Name: The Philips MP20, MP30, MP40, MP50, MP60, MP70 and MP90 IntelliVue Patient Monitors, Release B.1.2.
Indications for Use: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics and neonates in hospital environments. The MP20, MP30, MP40 and MP50 are intended for use in transport situations within additionally hospital environments.
ST Segment monitoring is restricted to adult patients only.
The transcutaneous gas measurement (tcp02 / tcpCO2) is restricted to neonatal patients only.
Prescription Use yes___________________________________________________________________________________________________________________________________________________________________________ Over-The-Counter Use _ No (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Blinums