K Number
K214048

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-09-20

(267 days)

Product Code
Regulation Number
870.1250
Age Range
All
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Millipede 088 Access Catheter is indicated for use in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the neurovasculature.

Device Description

The Millipede 088 Access Catheter consists of the catheter, a rotating hemostasis valve (RHV) and a valve crossing tool. The catheter, RHV and valve crossing tool are provided sterile. They are sterilized by ethylene oxide (EO).

The Millipede 088 Access Catheter is a single lumen, coil-reinforced, variable stiffness catheter. The distal segment has a hydrophilic coating for navigation through the vasculature. The catheter has a radiopaque marker located at its distal end for visualization under fluoroscopy. The valve crossing tool is used to open the valve of the access sheath and to facilitate insertion of the Millipede 088 Access Catheter through the access sheath without damage. The RHV is assembled onto the hub of the Millipede 088 Access Catheter and is used to maintain hemostasis during infusion of saline and contrast agent and insertion of other devices through the Millipede 088 Access Catheter.

AI/ML Overview

The provided document is a 510(k) summary for the Millipede 088 Access Catheter. This type of regulatory submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than providing detailed acceptance criteria and a standalone clinical study to prove the device meets those criteria.

Therefore, the document does not contain the information requested in points 2, 3, 4, 5, 8, and 9, and only partially addresses points 1, 6, and 7.

Here's a breakdown of what can be extracted:

1. Acceptance Criteria and Reported Device Performance

The document lists various performance tests and their conclusions, indicating that the device met "established specifications" or was "suitable for its intended use." However, the specific quantitative acceptance criteria for each test are not provided. The reported device performance is qualitative rather than quantitative in most cases.

TestTest MethodReported Device Performance (Conclusions)
Dimensional InspectionDevice dimensions were measured to confirm conformance to the specifications.The device met established specifications.
Tip StiffnessTest specimens were tested for tip flexibility and compared to predicate and reference devices.The device met established specifications.
Visual InspectionDevice surface characteristics were assessed to confirm freedom from defects.The device surface characteristics are suitable for its intended use.
Simulated Use TestingDeliverability and compatibility with accessory devices were evaluated in a neurovascular model.The device performs as intended under simulated use conditions.
Hydrophilic Coating IntegrityThe integrity of the hydrophilic coating was evaluated after multiple insertion and withdrawal cycles.The hydrophilic coating integrity is suitable for its intended use.
Particulate RecoveryQuantify the particulate size and count generated by simulated use of the test article.The particulate size and count were similar to control devices.
Tensile StrengthThe tensile strength was evaluated for the bonds between sections of the catheter.The device met established specifications.
Air LeakageTested per ISO 10555-1:2013 Annex D.The device integrity is suitable for its intended use.
Liquid LeakageTested per ISO 10555-1:2013 Annex C.The device integrity is suitable for its intended use.
Static BurstTested per ISO 10555-1:2013 Annex F.The device integrity is suitable for its intended use.
Luer IntegrityThe luers were evaluated for compliance to relevant standards.The luers on the device are suitable for their intended use.
Kink ResistanceTest specimen segments were formed into a defined bend diameter to evaluate kink resistance.The device met established specifications.
Torque StrengthThe test specimens were rotated in a simulated use model to evaluate integrity after rotation.The device met established specifications.
Flow Rate CharacterizationThe flow rate of saline and a contrast-saline solution was characterized when injected through the catheter.The flow rate was characterized.
RadiopacityRadiopacity of the device was evaluated in an animal model under fluoroscopy.The radiopacity of the Millipede 088 Access Catheter was similar to a control device.
Biocompatibility (various tests)ISO 10993-1, 10993-4 (Hemocompatibility), 10993-5 (Cytotoxicity), 10993-10 (Irritation, Sensitization), 10993-11 (Systemic Toxicity, Pyrogenicity)Non-cytotoxic, no sensitization response, met intracutaneous reactivity, met acute systemic injection, non-pyrogenic, not a complement activator, not an intrinsic coagulation pathway activator, non-hemolytic, similar thromboresistance to controls.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify the sample sizes for the individual performance tests beyond implicitly suggesting multiple samples were tested (e.g., "test specimens," "test articles"). For the Animal Testing, it states "two studies in a porcine model."

The data provenance for the in vitro and animal studies isn't explicitly stated beyond "Good Laboratory Practices" for the animal studies, which is a standard of conduct rather than a geographic origin. No human data is presented.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided as the submission relies on objective performance testing, biocompatibility studies, and animal studies rather than expert-derived ground truth from human data for this type of device.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. The evaluation methods described are objective performance tests and animal studies, not human data requiring expert adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is an access catheter, not an AI-powered diagnostic tool. Therefore, an MRMC study or AI assistance is not relevant to its evaluation.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, the performance testing described (mechanical, dimensional, coating integrity, etc.) and the animal studies represent "standalone" evaluations of the device's physical and functional properties without human interpretation of data in a clinical context.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the performance and biocompatibility tests, the "ground truth" is defined by the established specifications, relevant ISO standards, and comparison to control devices or predicate devices. For example:

  • "The device met established specifications."
  • "The test article is non-cytotoxic."
  • "The particulate size and count were similar to control devices."
  • "The radiopacity of the Millipede 088 Access Catheter was similar to a control device."
  • For animal testing, "Usability, radiopacity, thromboresistance, and vessel injury were assessed," and "The results for the subject device were comparable to a control device."

8. The sample size for the training set

Not applicable. This device is a medical catheter and does not involve a "training set" in the context of machine learning or AI. Its performance is evaluated through engineering and biological testing.

9. How the ground truth for the training set was established

Not applicable (as above).

{0}------------------------------------------------

September 20, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

Perfuze Ltd. Anne-Marie Gannon Director of Regulatory Affairs Unit 6, Galway Business Park, Dangan Galway, H91 W7CP Ireland

Re: K214048

Trade/Device Name: Millipede 088 Access Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: QJP Dated: August 18, 2022 Received: August 22, 2022

Dear Anne-Marie Gannon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K214048

Device Name Millipede 088 Access Catheter

Indications for Use (Describe)

The Millipede 088 Access Catheter is indicated for use in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the neurovasculature.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K214048 - 510(k) Summary

Submitter Information

Submitter's Name: Address:

Contact Person: Telephone: Date Prepared:

Subject Device

Proprietary Name: Common/Usual Name: Classification Name: Regulatory Class: Regulation: Product Code:

Predicate Device

Proprietary Name:

Common/Usual Name: Classification Name: Regulatory Class: Regulation: Product Code:

Manufacturer: 510(k) Number: Perfuze Ltd. Unit 6, Galway Business Park, Dangan, Galway, H91 W7CP, Ireland Anne-Marie Gannon +353 91 428083 September 20, 2022

Millipede 088 Access Catheter Guide Catheter Catheter, Percutaneous, Neurovasculature = 21 CFR 870.1250 QJP

TracStar™ Large Distal Platform Zoom™ 88 Large Distal Platform Zoom™ 88-T Large Distal Platform Guide Catheter Catheter, Percutaneous, Neurovasculature . 21 CFR 870.1250 QJP DQY Imperative Care K203764

Reference Devices

The following table lists the reference devices that were used to support the substantial equivalence determination in this submission.

510(k)NumberProductCodeName of DeviceDevice Manufacturer
K190010NRYPenumbra System ReperfusionCatheter JET 7Penumbra Inc.
K161152DQYNavien Intracranial SupportCatheterMicro Therapeutics Inc. d/b/aev3 Neurovascular
K152541NRYPenumbra System ACE 64Reperfusion CatheterPenumbra Inc.
K193034DQYAXS Infinity LS Plus LongSheathStryker Neurovascular
K203840QJP, DQYBOSS 8F Balloon GuideCatheterMarblehead Medical LLC
K182097DQYReact 71 CatheterMicro Therapeutics Inc. d/b/aev3 Neurovascular

{4}------------------------------------------------

510(k)NumberProductCodeName of DeviceDevice Manufacturer
K183464NRYAXS Catalyst 7 Distal AccessCatheterStryker Neurovascular
K133177DQY, DQOModified HD Guide CatheterConcentric Medical Inc.
K090752NRYPenumbra ReperfusionCatheter 054Penumbra Inc.

Device Description

The Millipede 088 Access Catheter consists of the catheter, a rotating hemostasis valve (RHV) and a valve crossing tool. The catheter, RHV and valve crossing tool are provided sterile. They are sterilized by ethylene oxide (EO).

The Millipede 088 Access Catheter is a single lumen, coil-reinforced, variable stiffness catheter. The distal segment has a hydrophilic coating for navigation through the vasculature. The catheter has a radiopaque marker located at its distal end for visualization under fluoroscopy. The valve crossing tool is used to open the valve of the access sheath and to facilitate insertion of the Millipede 088 Access Catheter through the access sheath without damage. The RHV is assembled onto the hub of the Millipede 088 Access Catheter and is used to maintain hemostasis during infusion of saline and contrast agent and insertion of other devices through the Millipede 088 Access Catheter.

Indications for Use

The Millipede 088 Access Catheter is indicated for use in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the neurovasculature.

The Indications for Use statement for the Millipede 088 Access Catheter is not identical to the predicate device. However, both the Millipede 088 Access Catheter and the predicate device can be used to facilitate insertion and guidance of microcatheters into the neurovasculature. The differences in the Indications for Use statement do not raise new questions of safety and effectiveness for the Millipede 088 Access Catheter relative to the predicate device.

Comparison to the Predicate Device

The subject and predicate devices have similar technological characteristics as shown in the following table.

AttributePredicate DeviceTracStar™ Large Distal PlatformZoom™ 88 Large Distal PlatformZoom™ 88-T Large Distal Platform(K203764)Subject DeviceMillipede 088 Access Catheter
RegulationNumber21 CFR 870.1250Same
Regulation NamePercutaneous catheterSame
ClassificationClass IISame
Product CodeQJP, DQYQJP
AttributePredicate DeviceSubject Device
TracStar™ Large Distal PlatformZoom™ 88 Large Distal PlatformZoom™ 88-T Large Distal Platform(K203764)Millipede 088 Access Catheter
Indications for useThe TracStar Large Distal Platform is indicated for the introduction ofinterventional devices into theperipheral, coronary, and neurovasculature.The ZOOM 88 and ZOOM 88-TLarge Distal Platform are indicatedfor the introduction of interventionaldevices into the peripheral, coronary,and neuro vasculature.The Millipede 088 Access Catheteris indicated for use in facilitatingthe insertion and guidance ofmicrocatheters into a selectedblood vessel in theneurovasculature.
Prescription/over-the-counter usePrescriptionSame
Device DescriptionSingle-use, variable stiffness, wire-reinforced catheters with a singlelumen. The catheters are comprisedof a hollow cylindrical tube bonded atthe proximal end to a standard luerfitting. The wall of the tube isconstructed using metals andpolymers. A radiopaque markerprovides visual confirmation of thedistal tip location under fluoroscopy.Same
Principle ofOperationMay be used with support cathetersto assist in accessing the targetvasculature.Same
Techniques forUseStandard percutaneous interventionaltechniques, including access sitepreparation, introduction of thecatheter into the access vessel,advancing the catheter underfluoroscopy, withdrawing thecatheter, and closing the access site.Same
MaterialsPolymers and metals commonly usedin the manufacture of medicaldevices.Same
Distal TipBeveled edge, soft, flexible, andatraumaticSquare edge, soft, flexible, andatraumatic
Catheter WallConstructionCoil-reinforcedCoil-reinforced with ribbed surfaceat distal section
CoatingHydrophilic CoatingSame
Catheter Profile8 FrSame
Inner DiameterDistal: 0.088"Proximal: 0.088"Distal: 0.088"Proximal: 0.087"
Outer DiameterDistal: 0.106"Proximal: 0.108"Distal: 0.104"Proximal: 0.108"
Effective Length80cm -110cm115cm
PackagedAccessoriesRHVRHV and Valve Crossing Tool
ConditionSterile and Single UseSame
AttributePredicate DeviceTracStarTM Large Distal PlatformZoomTM 88 Large Distal PlatformZoomTM 88-T Large Distal Platform(K203764)Subject DeviceMillipede 088 Access Catheter
Sterilization MethodEthylene Oxide (EO), SterilityAssurance Level 10-6Same
Packaging ConfigurationThe catheters are placed in aprotective polyethylene tube,mounted with accessory RHV onto apolyethylene packaging card, placedinto a pouch, sealed, and labeled.The sealed pouch and IFU areplaced in a labeled shelf carton box.The catheters are placed in aprotective polyethylene tube,mounted with accessory RHV andvalve crossing tool onto acardboard packaging card, placedinto a pouch, sealed, and labeled.The sealed pouch and IFU areplaced in a labeled shelf cartonbox.

{5}------------------------------------------------

{6}------------------------------------------------

Biocompatibility Testing

The Millipede 088 Access Catheter is constructed using materials that are commonly used in the medical device industry. All patient contacting components have been evaluated for biocompatibility in accordance with ISO 10993-1, "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process." The Millipede 088 Access Catheter is classified per ISO 10993-1 as an externally communicating device that contacts circulating blood for a limited (< 24 hours) duration. A summary of the biocompatibility testing is outlined below.

TestResults
Cytotoxicity - ISO MEM ElutionThe test article is non-cytotoxic.
Sensitization - ISO Guinea PigMaximization Sensitization TestThe test article did not elicit a sensitization response.
Irritation - ISO Intracutaneous ReactivityRequirements of the ISO intracutaneous reactivitytest were met for the test article.
Acute Systemic Toxicity – ISO AcuteSystemic InjectionRequirements of the ISO acute systemic injectiontest were met for the test article.
Material-Mediated PyrogenicityThe test article is non-pyrogenic.
Hemocompatibility - ComplementActivation (SC5b-9)The test article is not considered to be a potentialactivator of the complement system.
Hemocompatibility - PartialThromboplastin TimeThe test article is not considered to be an activatorof the intrinsic coagulation pathway.
Hemocompatibility – ASTM HemolysisThe test article is considered non-hemolytic.
Hemocompatibility - ThromboresistanceThe test articles have similar thromboresistancecharacteristics as the control devices.

Performance Testing

The successful completion of the performance testing listed in the following table demonstrates that the Millipede 088 Access Catheter is suitable for its intended use.

TestTest MethodConclusions
DimensionalInspectionDevice dimensions were measured toconfirm conformance to thespecifications.The device met establishedspecifications.
Tip StiffnessTest specimens were tested for tipflexibility and compared to predicate andreference devices.The device met establishedspecifications.

{7}------------------------------------------------

TestTest MethodConclusions
Visual InspectionDevice surface characteristics wereassessed to confirm freedom fromdefects.The device surfacecharacteristics are suitablefor its intended use.
Simulated UseTestingDeliverability and compatibility withaccessory devices were evaluated in aneurovascular model.The device performs asintended under simulated useconditions.
Hydrophilic CoatingIntegrityThe integrity of the hydrophilic coatingwas evaluated after multiple insertion andwithdrawal cycles.The hydrophilic coatingintegrity is suitable for itsintended use.
ParticulateRecoveryThe purpose of this test was to quantifythe particulate size and count generatedby simulated use of the test article.The particulate size and countwere similar to controldevices.
Tensile StrengthThe tensile strength was evaluated forthe bonds between sections of thecatheter.The device met establishedspecifications.
Air LeakageTested per ISO 10555-1:2013 Annex D.The device integrity issuitable for its intended use.
Liquid LeakageTested per ISO 10555-1:2013 Annex C.The device integrity issuitable for its intended use.
Static BurstTested per ISO 10555-1:2013 Annex F.The device integrity issuitable for its intended use.
Luer IntegrityThe luers were evaluated for complianceto relevant standards.The luers on the device aresuitable for their intendeduse.
Kink ResistanceTest specimen segments were formedinto a defined bend diameter to evaluatekink resistance.The device met establishedspecifications.
Torque StrengthThe test specimens were rotated in asimulated use model to evaluate integrityafter rotation.The device met establishedspecifications.
Flow RateCharacterizationThe flow rate of saline and a contrast-saline solution was characterized wheninjected through the catheter.The flow rate wascharacterized.
RadiopacityRadiopacity of the device was evaluatedin an animal model under fluoroscopy.The radiopacity of theMillipede 088 AccessCatheter was similar to acontrol device.

Animal Testing

The in vivo performance and safety of the device was assessed in two studies in a porcine model at 3-day and 30-day time points. Studies were conducted under Good Laboratory Practices. Usability, radiopacity, thromboresistance, and vessel injury were assessed. The results for the subject device were comparable to a control device, demonstrating acceptable results.

Sterilization

The Millipede 088 Access Catheter is sterilized using a validated EO process with a sterility assurance level of 1x10-9. The validation was conducted using the overkill method according to ISO 11135, "Sterilization of Health-Care Products - Ethylene Oxide - Requirements for the Development, Validation and Routine Control of a Sterilization Process for Medical Devices".

Shelf Life and Packaqinq

Accelerated aging testing based on ASTM F1980 was conducted to verify packaged device performance. A real time aging equivalent of 8 months was used to support an 8-month shelf-life claim. Device performance was verified by functional and performance testing.

{8}------------------------------------------------

Substantial Equivalence

The intended use of the Millipede 088 Access Catheter is similar to the intended use of the predicate device. The Millipede 088 Access Catheter and the predicate device use the same operating principles and have a similar design. The minor technological differences identified do not raise different questions of safety or effectiveness for the two devices. The successful completion of biocompatibility testing and performance testing demonstrates that the Millipede 088 Access Catheter is substantially equivalent to the predicate device.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).