K Number
K133177

Validate with FDA (Live)

Date Cleared
2014-02-25

(131 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Modified HD Guide Catheter is indicated for use in facilitating the insertion and guidance of an occlusion catheter, infusion catheter or other appropriate microcatheter into a selected blood vessel in the peripheral, coronary and neuro vascular systems. It may also be used as a diagnostic angiographic catheter and as a conduit for retrieval devices.

Device Description

The Modified HD Guide Catheters are single lumen, braided, variable stiffness shaft catheters designed for use in facilitating the insertion and guidance of an occlusion catheter, infusion catheter or other appropriate microcatheter into a selected blood vessel in the peripheral, coronary or neuro vascular system. The catheters include a radiopaque marker on the distal and for angiographic visualization and a luer hub on the proximal end allowing attachments for flushing and aspiration. The catheter shaft has a hydrophilic coating to reduce friction during use. A rotating hemostatic valve with side-arm adapter is provided with each catheter.

AI/ML Overview

Here's an analysis of the acceptance criteria and study details for the Modified HD Guide Catheter, based on the provided 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance CriteriaReported Device Performance
Tip Patency during AspirationDistal tip of catheter remains patent during vigorous aspiration (to avoid loss of suction).Pass. All samples met acceptance criteria for expanded indications for use. Device continues to meet same design requirements as predicate devices (K090335 and K110483).
Lumen CompatibilityGuidewires and other devices pass through the inner shaft with no more than moderate resistance.Pass. All samples met acceptance criteria for expanded indications for use. Device continues to meet same design requirements as predicate devices (K090335 and K110483).
Lumen and Retriever CompatibilityRetriever and microcatheter (as a system) can be withdrawn through the catheter and completely removed from the model without device fracture.Pass. All samples met acceptance criteria for expanded indications for use. Device continues to meet same design requirements as predicate devices (K090335 and K110483).
Infusion and AspirationUser can aspirate and inject fluid through the Extension Tubing with a 60cc syringe. Catheter maintains mechanical integrity and function.Pass. All samples met acceptance criteria for expanded indications for use. Device continues to meet same design requirements as predicate devices (K090335 and K110483).
Distal Tip StabilityDistal tip does not move (forward or back) during retraction of the Retriever and microcatheter.Pass. All samples met acceptance criteria for expanded indications for use. Device continues to meet same design requirements as predicate devices (K090335 and K110483).
Mechanical and Functional IntegrityCatheter maintains mechanical integrity (e.g., no damage, kink, etc.) and function is not impaired after simulated use.Pass. All samples met acceptance criteria for expanded indications for use. Device continues to meet same design requirements as predicate devices (K090335 and K110483).
Retriever FractureRetriever is free from fractures following simulated use testing.Pass. All samples met acceptance criteria for expanded indications for use. Device continues to meet same design requirements as predicate devices (K090335 and K110483).
Liquid Leak ResistanceCatheter does not leak from the shaft or hub/shaft interface during use.Pass. All samples met acceptance criteria for expanded indications for use. Device continues to meet same design requirements as predicate devices (K090335 and K110483).
Design Verification (various tests listed in document)Performance as designed, suitability for intended use, substantial equivalence to predicate devices.Successfully evaluated and met criteria for all tests (Tip Patency during Aspiration, Air Leak Resistance during Aspiration, Leak Testing, Dimensional Testing, Tensile Testing, Kink Resistance Testing, Flexural Fatigue Testing, Torque Testing, Tip Flexibility Testing, Coating Lubricity and Durability Testing, Flow Rate Testing, Luer Testing).

2. Sample Size Used for the Test Set and Data Provenance

The document categorizes testing into "Testing and Non-Clinical Performance Data" (bench testing) and "Clinical Performance Data" (review of existing clinical studies and post-market data).

  • Non-Clinical Performance Data (Bench Testing):

    • The document states "All samples met acceptance criteria." but does not specify the exact number of samples used for each bench test (e.g., Tip Patency during Aspiration, Lumen Compatibility, etc.). It only implies that multiple samples (denoted by "All samples") were tested for each criterion.
    • Data Provenance: The bench testing utilized a "neurovascular model cast from actual human neurovascular arteries." This is a controlled experimental setup, not data from live patients.
    • Specifically for the expanded indications, "Finished sterilized devices were used for the simulated use testing."
  • Clinical Performance Data:

    • TREVO 2 Study: 99 patients where the DAC (Modified HD Guide Catheter) was used.
      • Provenance: Retrospective analysis of an IDE trial ("TREVO 2 study... designed to support FDA clearance of the Trevo Retriever in the U.S."). Enrolled at 26 sites in the United States and one site in Spain.
    • TREVO Study: 30 patients where aspiration through DAC was applied (a subset of 34 patients where DAC was used).
      • Provenance: Retrospective analysis of a post-marketing prospective, multi-center, single-arm study performed at seven sites in Europe.
    • Post-Market Surveillance: A review of product complaints and Medical Device Reports (MDRs) from 2008 to July 31, 2013. The exact number of complaints reviewed is not given, but it is stated that 10 MDRs were filed.
      • Provenance: Real-world post-market data.
    • Literature Review: 15 published articles pertaining to DAC use.
      • Provenance: Published scientific literature.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Non-Clinical (Bench) Testing: The document does not specify the number or qualifications of experts involved in observing and confirming the "Pass" results for the bench tests. It likely involved internal engineering or research staff.
  • Clinical Performance Data (TREVO and TREVO 2 Studies):
    • The original TREVO 2 and TREVO studies would have had clinical investigators and potentially core labs or adjudication committees, but their specific roles in establishing "ground truth" for catheter performance (as opposed to primary study endpoints for the Retriever device) are not detailed here.
    • The 510(k) summary states that "None of the procedure related or CEC adjudicated events were related to DAC," implying that a Clinical Events Committee (CEC) was used in the TREVO study to adjudicate events. Their specific qualifications are not provided in this 510(k) summary.

4. Adjudication Method for the Test Set

  • Non-Clinical (Bench) Testing: No specific adjudication method is mentioned. Results were simply reported as "Pass" based on visual verification or adherence to specified methods.
  • Clinical Performance Data: The TREVO study mentioned events being "CEC adjudicated," indicating a Clinical Events Committee (CEC) was involved in reviewing adverse events. The specific method within the CEC (e.g., 2+1, 3+1) is not detailed. The TREVO 2 study also implies similar clinical oversight.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There is no MRMC comparative effectiveness study mentioned, nor is there any AI assistance component to this device. This is a physical medical device (catheter), not an AI-powered diagnostic tool. Therefore, this section is not applicable.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable, as this device is a physical catheter and does not involve an algorithm.

7. The Type of Ground Truth Used

  • Non-Clinical (Bench) Testing:
    • For most bench tests, the "ground truth" was based on pre-defined acceptance criteria and direct observation of the device's physical performance characteristics (e.g., patency, resistance, integrity, absence of fracture/leakage) in a simulated environment (neurovascular model).
  • Clinical Performance Data (TREVO 2 and TREVO Studies):
    • The "ground truth" in these studies relates to the clinical outcomes of the overall thrombectomy procedure, specifically revascularization rates and the absence of device-specific adverse events. While the primary ground truth would be patient outcomes as assessed by clinicians (e.g., angiography for revascularization, neurological exams for NIHSS), the review here focuses on whether the DAC was implicated in any negative outcomes.

8. The Sample Size for the Training Set

Not applicable. This is a physical device and not an AI/machine learning model that requires a training set. The existing clinical data reviewed served as real-world performance evidence, not a "training set."

9. How the Ground Truth for the Training Set was Established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

K133177

FEB 2 5 2014

510(k) Summary

Summary Date:January 30, 2014
Submitter Name andAddress:Concentric Medical, Inc.301 E. Evelyn AvenueMountain View, CA 94041Tel 650-938-2100Fax 650-237-5230Facility Registration #2954917
Contact:Rhoda SantosPrincipal Regulatory Affairs SpecialistPhone: 510 413-2269Fax: 510 413-2558Email: rhoda.santos@stryker.com
Trade Name:Modified HD Guide Catheter
Common Name:Percutaneous Catheter
Classification Name:Percutaneous Catheter, 21CFR 870.1250 – Class II
Product Code:DQY and DQO
Legally MarketedPredicate Devices:Reference(Clearance Date)Device
K090335 (May 6, 2009)Concentric HD Guide Catheter
K110483 (April 4, 2011)Modified HD Guide Catheter
K112404 (March 15, 2012)Concentric Balloon Guide Catheter
Device Description:The Modified HD Guide Catheters are single lumen, braided,variable stiffness shaft catheters designed for use in facilitating theinsertion and guidance of an occlusion catheter, infusion catheter orother appropriate microcatheter into a selected blood vessel in theperipheral, coronary or neuro vascular system. The catheters includea radiopaque marker on the distal and for angiographic visualization

adapter is provided with each catheter.

and a luer hub on the proximal end allowing attachments for flushing and aspiration. The catheter shaft has a hydrophilic coating to reduce friction during use. A rotating hemostatic valve with side-arm

{1}------------------------------------------------

Accessories:

The Modified HD Guide Catheter is packaged with a Rotating Hemostasis Valve.

Indications for Use / Intended Use:

The proposed Indications for Use are as follows:

The Modified HD Guide Catheter is indicated for use in facilitating the insertion and guidance of an occlusion catheter, infusion catheter or other appropriate microcatheter into a selected blood vessel in the peripheral, coronary and neuro vascular systems. It may also be used as a diagnostic angiographic catheter and as a conduit for retrieval devices.

Technological Characteristics and Product Feature Comparison:

The subject device has the same technological characteristics as the predicate devices (K090335 and K110483). The device design, materials, fundamental scientific technology, materials and processes for packaging and sterilization have not been changed from the previous predicate devices (K090335 and K110483). The subject device differs from the predicate device, K112404 primarily in that it does not include a balloon at the distal tip or a second lumen.

A tabular comparison of the specific technological characteristics between the predicate devices and subject device is provided below.

{2}------------------------------------------------

Product Feature Comparison of Subject Device with Predicate Devices
FeaturePredicate Device,K112404Predicate Device,K090335Predicate Device,K110483Subject Device
Indications for UseThe Concentric BalloonGuide Catheter is indicatedfor use in facilitating theinsertion and guidance of anintravascular catheter into aselected blood vessel in theperipheral and neurovascular systems. Theballoon provides temporaryvascular occlusion duringthese and other angiographicprocedures. The BalloonGuide Catheter is alsoindicated for use as a conduitfor Retrieval devicesThe HD Guide Catheter isindicated for theremoval/aspiration of fresh,soft emboli and thrombi fromvessels in the arterial system.The Modified HD GuideCatheter is indicated for usein facilitating the insertionand guidance of an occlusioncatheter, infusion catheter orother appropriatemicrocatheter into a selectedblood vessel in theperipheral, coronary andneuro vascular systems. Itmay also be used as adiagnostic angiographiccatheter.The Modified HD GuideCatheter is indicated for usein facilitating the insertionand guidance of anocclusion catheter, infusioncatheter or otherappropriate microcatheterinto a selected blood vesselin the peripheral, coronaryand neuro vascular systems.It may also be used as adiagnostic angiographiccatheter and as a conduit forretrieval devices.
FeatureDevice DescriptionPredicate Device,K112404Predicate Device,K090335Predicate Device,K110483Subject Device
The Concentric BalloonGuide Catheters arecoaxial-lumen, braid-reinforced, variablestiffness cathetersdesigned for use infacilitating the insertionand guidance of anintravascular catheterinto a selected bloodvessel in the peripheraland neuro vascularsystems. A radiopaquemarker is included on thedistal end forangiographicvisualization. Acompliant balloon ismounted on the distal endto provide temporaryvascular occlusion duringangiographic procedures.A bifurcated luer hub onthe proximal end allowsattachments for flushing,inflation and aspiration.Balloon Guide Catheterdimensions andmaximum recommendedballoon inflation volumeare indicated on productlabel. If indicated onproduct label, a dilator isprovided.The HD Guide Catheterconsists of a single lumen,braided, variable stiffnessshaft with a radiopaquemarker on the distal end anda luer hub on the proximalend. The catheter shaft has ahydrophilic coating to reducefriction during use. Arotating hemostatic valvewith side-arm adapter isprovided with each catheter.Same as predicate device,K090335Same as predicate devices,K090335 and K110483
FeaturePredicate Device,K112404Predicate Device,K090335Predicate Device,K110483Subject Device
Outer JacketPebax®Pebax®Same as predicate device,K090335Same as predicate devices,K090335 and K114083
BraidStainless SteelStainless SteelSame as predicate device,K090335Same as predicate devices,K090335 and K114083
Strain ReliefPolyolefinPolyolefinSame as predicate device,K090335Same as predicate devices,K090335 and K114083
Braid distal end securementPTFEPTFESame as predicate device,K090335Same as predicate devices,K090335 and K114083
Catheter HubPolyurethanePebaxSame as predicate device,K090335Same as predicate devices,K090335 and K114083
Marker BandPlatinum/IridiumPlatinum/IridiumSame as predicate device,K090335Same as predicate devices,K090335 and K114083
AdhesiveAcrylic (AcrylatedUrethane)Acrylic (Acrylated Urethane)Same as predicate device,K090335Same as predicate devices,K090335 and K114083
Outer jacket coatingNAhydrophilic coatingSame as predicate device,K090335Same as predicate devices,K090335 and K114083
Labeled Shaft Outer Diameter7F - 9F3.9F - 5.2F6.3FSame as predicate devices,K090335 and K114083
Effective Lengths80cm or 95cm115 cm – 136 cm105-120 cmSame as predicate devices,K090335 and K114083
Accessory Devices ProvidedDilatorRotating Hemostatic ValveSame as predicate device,K090335Same as predicate devices,K090335 and K114083
Packaging Materials andConfigurationPolyethylene Tube andHDPE Packaging CardPolyethylene Tube andHDPE Packaging CardSame as predicate device,K090335Same as predicate devices,K090335 and K114083
Sterilization MethodEO SterilizationSameSameSame
How SuppliedSterile, Single UseSameSameSame

C ... De t Feature Comparison of Subject Device with Produ

:

{3}------------------------------------------------

510(k) Summary (cont

{4}------------------------------------------------

510(k) Summary (cont

{5}------------------------------------------------

Risk Assessment

Risk assessment of the Modified HD Guide Catheter has been conducted in accordance with EN ISO 14971:2012. As a result of the risk assessment, the Instructions for Use have been revised to include recommended aspiration procedure steps for use of the Modified HD Guide Catheter with retrieval devices.

Testing and Non-Clinical Performance Data:

The results of verification and validation conducted on the Modified HD Guide Catheter demonstrate that it performs as designed, is suitable for its intended use and is substantially equivalent to the predicate devices (K090335 and K110483). The same performance standards, specifications and results as those submitted in the predicate devices (K090335 and K110483) are applicable to the subject device because the designs are identical and no changes have been made to the design, materials, processes, or packaging materials as a result of the expanded indications for use. Therefore, the following design verification tests submitted in the predicate devices (K090335 and K110483) are applicable to the subject device and its new indications for use:

  • . Tip Patency during Aspiration: the ability of the device tip to maintain patency during aspiration during simulated use testing was successfully evaluated.
  • Air Leak Resistance during Aspiration: the ability of the device to resist leakage ♥ during aspiration was successfully evaluated.
  • Leak Testing: the ability of the device to resist leakage was successfully evaluated.
  • Dimensional Testing: dimensions of the device were successfully verified. .
  • Tensile Testing: the tensile strength of the device was successfully evaluated. .
  • . Kink Resistance Testing: the ability of the device to withstand curves without kinking was successfully evaluated.
  • . Flexural Fatigue Testing: the flexural fatigue tolerance of the device was successfully evaluated.
  • Torque Testing: the ability of the device to withstand torsional forces was successfully evaluated.
  • Tip Flexibility Testing: the force to deflect the catheter tip was successfully . evaluated.
  • Coating Lubricity and Durability Testing: the durability of the hydrophilic coating . was successfully evaluated.
  • Flow Rate Testing: the rate of flow through the device lumen was successfully . evaluated.
  • . Luer Testing: luer integrity and conformance to luer standards was successfully evaluated.

To support the expansion of the Indications for Use (IFU), performance testing utilized design validation / simulated use testing to confirm that the subject device and accessories meets user needs "as a conduit for retrieval devices" and continue to meet design requirements of the predicate device. Simulated use testing evaluates the device's ability to be used in a neurovascular model per procedural instructions outlined in the modified Instructions for Use. Based on the results of the risk assessment, no other testing was required to demonstrate the device meets intended uses.

{6}------------------------------------------------

Design validation/simulated use testing was conducted using a neurovascular model cast from actual human neurovascular arteries. This bench testing model replicates the tortuosity, diameter and location of the arteries in the neurovasculature. The model incorporates a re-circulating water bath at 37℃ pressurized between 2 - 2.5 psi (100 -126 mm Hg) to simulate the human arterial circulation.

The simulated use tests described in Table 2 were conducted in support of the expanded indications for use. Finished sterilized devices were used for the simulated use testing. All design validation testing followed the procedural instructions outlined in the Instructions for Use which was revised to include a "Recommended Aspiration Procedure" for use with retrieval devices. Based on the successful completion of the testing, the subject device and its accessories have met all of the pre-specified requirements. A summary of the design validation / simulated use testing and the results is described in Table 2 below.

TestTest Method SummaryResults
Tip Patency during AspirationPurpose: Tip patency is evaluated in aneurovascular model to verify that the distal tip ofcatheter remains patent during vigorous aspirationto avoid loss of suction at the distal tip.PassAll samples met acceptancecriteria for expandedindications for use.
Method: A neurovascular model is placed in a re-circulating water bath at 37°C and pressurized tosimulate human arterial circulation. The catheter isplaced in the model to a specified locationfollowing procedural instructions outlined in theInstructions for Use. The catheter is aspiratedvigorously using a 60 cc syringe and the shape ofthe distal tip is visually verified to determine ifdistal tip remains patent during aspiration.Device continues to meet samedesign requirements aspredicate devices (K090335and K110483).
Lumen CompatibilityPurpose: Lumen compatibility is evaluated in aneurovascular model to verify that guidewires andother devices shall pass through the inner shaft ofthe catheter with no more than moderateresistance.PassAll samples met acceptancecriteria for expandedindications for use.
Method: A neurovascular model is placed in a re-circulating water bath at 37°C and pressurized tosimulate human arterial circulation. The catheteris placed in the model to a specified locationfollowing procedural instructions outlined in theInstructions for Use. A "compatible device" isinserted through the inner shaft of the catheter andassessed for degree of resistance through innershaft, i.e., easy, moderate, difficult, or notpossible.Device continues to meet samedesign requirements aspredicate devices (K090335and K110483).

Table 2: Design Validation / Simulated Use Testing of Subject Device and Accessories

{7}------------------------------------------------

510(k) Summary (cont.)

TestTest Method SummaryResults
Lumen and RetrieverCompatibilityPurpose: Lumen and Retriever compatibility isevaluated in a neurovascular model to verify thatthe Retriever and microcatheter (as a system) isable to be withdrawn through the catheter andcompletely removed from the model withoutdevice fracture.Method: A neurovascular model is placed in a re-circulating water bath at 37°C and pressurized tosimulate human arterial circulation. The Retrieverand microcatheter is inserted through the catheterto reach a specified target location followingprocedural instructions outlined in the Instructionsfor Use. The Retriever and microcatheter (as asystem) is pulled back through the catheter andcompletely remove them from the model. TheRetriever is visually inspected for damage.PassAll samples met acceptancecriteria for expandedindications for use.Device continues to meet samedesign requirements aspredicate devices (K090335and K110483).
Infusion and AspirationPurpose: Infusion and aspiration of the catheter isevaluated in a neurovascular model to verify thatthe user is able to aspirate and inject fluid thru theExtension Tubing with 60cc syringe.Method: A neurovascular model is placed in a re-circulating water bath at 37°C and pressurized tosimulate human arterial circulation. The catheter isplaced in the model to a specified locationfollowing procedural instructions outlined in theInstructions for Use. The Retriever andmicrocatheter are advanced (as a system) throughthe BGC. The extension tubing is attached to theY-arm of the rotating hemostasis valve. Fluid isinfused vigorously through extension tubing with a60cc syringe. While retracting both the Retrieverand microcatheter (as a system) through thecatheter to completely remove the system from themodel, a 60cc syringe is used to aspirate throughthe extension tubing. When the Retriever isretracted into the catheter tip, the user aspirates to60cc's vigorously as the Retriever is fully retractedin the catheter. The catheter is visually inspectedfor damage.PassAll samples met acceptancecriteria for expandedindications for use.Device continues to meet samedesign requirements aspredicate devices (K090335and K110483).
TestTest Method SummaryResults
Distal tip stabilityPurpose: Distal tip stability is evaluated in aneurovascular model to verify that the distal tipdoes not move during retraction of the Retrieverand microcatheter (as a system) into the catheter.Method: A neurovascular model is placed in a re-circulating water bath at 37°C and pressurized tosimulate human arterial circulation. The catheter isplaced in the model to a specified locationfollowing procedural instructions outlined in theInstructions for Use. The Retriever andmicrocatehter are advanced (as a system) throughthe catheter. The extension tubing is attached tothe Y-arm of the rotating hemostasis valve. Fluidis infused vigorously through extension tubingwith a 60cc syringe. While retracting both theRetriever and microcatheter (as a system) throughthe catheter to completely remove the system fromthe model, a 60cc syringe is used to aspiratethrough the extension tubing. When the Retrieveris retracted into the catheter tip, the user aspiratesto 60cc's vigorously as the Retriever is fullyretracted in the catheter. The catheter is visuallyinspected for damage. The distal tip of catheter isvisually inspected to verify it does not moveforward or back during retraction of the Retrieverand microcatheter (as a system).PassAll samples met acceptancecriteria for expandedindications for use.Device continues to meet samedesign requirements aspredicate devices (K090335and K110483).
Mechanical and FunctionalIntegrityPurpose: The catheter is inspected for mechanicaland functional integrity following simulated use ina neurovascular model to verify that the cathetermaintains mechanical integrity (e.g., visibledamage, shaft kink, hub joints, shaft joints, tipseparation, corrosion) and that catheter functionwas not impaired by the introducer sheath.Method: A neurovascular model is placed in a re-circulating water bath at 37°C and pressurized tosimulate human arterial circulation. The catheter isplaced in the model to a specified locationfollowing procedural instructions outlined in theInstructions for Use. The Retriever andmicrocatehter are advanced (as a system) throughthe catheter. The extension tubing is attached tothe Y-arm of the rotating hemostasis valve. Fluidis infused vigorously through extension tubingwith a 60cc syringe. While retracting both theRetriever and microcatheter (as a system) throughthe catheter to completely remove the system fromthe model, a 60cc syringe is used to aspiratethrough the extension tubing. When the Retrieveris retracted into the catheter tip, the user aspiratesto 60cc's vigorously as the Retriever is fullyPassAll samples met acceptancecriteria for expandedindications for use.Device continues to meet samedesign requirements aspredicate devices (K090335and K110483).
retracted in the catheter. The catheter is visually
TestTest Method SummaryResults
Retriever FracturePurpose: Retriever fracture is evaluated in aneurovascular model to verify that the Retriever isfree from fractures following simulated usetesting.PassAll samples met acceptancecriteria for expandedindications for use.
Method: A neurovascular model is placed in a re-circulating water bath at 37°C and pressurized tosimulate human arterial circulation. The catheter isplaced in the model to a specified locationfollowing procedural instructions outlined in theInstructions for Use. The Retriever andmicrocatheter are advanced (as a system) throughthe catheter. The extension tubing is attached tothe Y-arm of the rotating hemostasis valve. Fluidis infused vigorously through extension tubingwith a 60cc syringe. While retracting both theRetriever and microcatheter (as a system) throughthe catheter to completely remove the system fromthe model, a 60cc syringe is used to aspiratethrough the extension tubing. When the Retrieveris retracted into the catheter tip, the user aspiratesto 60cc's vigorously as the Retriever is fullyretracted in the catheter. The catheter is visuallyinspected for damage. The Retriever is removedfrom the microcatheter and visually inspected forfractures.Device continues to meet samedesign requirements aspredicate devices (K090335and K110483).
Liquid leak resistancePurpose: Liquid leak resistance of the catheter isevaluated in a neurovascular model to verify thatthe catheter does not leak from the shaft orhub/shaft interface during use.PassAll samples met acceptancecriteria for expandedindications for use.
Method: A neurovascular model is placed in a re-circulating water bath at 37°C and pressurized tosimulate human arterial circulation. The catheter isplaced in the model to a specified locationfollowing procedural instructions outlined in theInstructions for Use. The Retriever andmicrocatheter are advanced (as a system) throughthe catheter. The extension tubing is attached tothe Y-arm of the rotating hemostasis valve. Fluidis infused vigorously through extension tubingwith a 60cc syringe. While retracting both theRetriever and microcatheter (as a system) throughthe catheter to completely remove the system fromthe model, a 60cc syringe is used to aspiratethrough the extension tubing. When the Retrieveris retracted into the catheter tip, the user aspiratesto 60cc's vigorously as the Retriever is fullyretracted in the catheter. The catheter is visuallyinspected for leaks from the shaft or hub/shaftinterface during use.Device continues to meet samedesign requirements aspredicate devices (K090335and K110483).

·

{8}------------------------------------------------

510(k) Summary (cont.)

{9}------------------------------------------------

510(k) Summary (cont.)

Based on conformance with these test requirements, the Modified HD Guide Catheter is as safe, as effective, and performs as well as or better than the legally marketed predicate devices (K090335 and K110483).

{10}------------------------------------------------

Clinical Performance Data:

To demonstrate substantial equivalence, a review of the Thrombectomy REvasculatorization of large Yessel Occlusions in acute ischemic stroke (TREVO) 2 IDE study and the TREVO post market study in Europe was conducted to assess if thrombectomy procedures performed with the Modified HD Guide Catheter as a support catheter have similar revascularization rates as published trials and to identify any Modified HD Guide Catheter specific safety events which may preclude expansion of the indications for use "as a conduit for retrieval devices." Clinical data was also reviewed to identify if any of the procedure events were reported to be directly related to the Modified HD Guide Catheter, also marketed as the Distal Access Catheter (DAC).

TREVO 2 Study

The TREVO 2 study was an IDE trial designed to support FDA clearance of the Trevo Retriever in the U.S. The study enrolled 178 subjects between February 3, 2011 and December 1, 2011 at 26 sites in the United States and one site in Spain. A review of the TREVO 2 data showed that DAC was used in a total of 99 patients. For all 99 patients, the DAC was placed for additional support at the proximal face of the thrombus in the intracranial artery and used as a conduit during the retrieval procedure.

The overall review of the DAC 99 cohort in the TREVO 2 study, showed that safety and efficacy of the thrombectomy procedure was similar to the overall TREVO 2 study population. The total procedure length, time from the first retrieval device to the final angiogram and number of thrombectomy device passes in this DAC cohort are similar to the TREVO 2 data illustrating that the additional step of using the DAC does not affect the procedure time. In the 99 patients where DAC was used during the retriever procedure, no complaints were attributed specifically to the DAC. The microcatheters were able to track through the DAC to the clots in the majority of the cases. The revascularization rate in these patients was 84.8% and showed that successful revascularization was possible using DAC as a conduit for retrieval devices. There were no reports or findings that procedural events were related to DAC therefore supporting safety of DAC during thrombectomy procedures.

TREVO Study

The TREVO study was a post marketing prospective, multi-center, single arm study performed at seven sites in Europe and was designed to determine the revascularization rate of the CE-marked Trevo device in large vessel occlusions in ischemic stroke patients. Sixty (60) patients were enrolled between February, 2010 and August, 2011.

Review of the TREVO data showed DAC was used as an insertion and guidance catheter in a total of 34 patients. Further review of these 34 patients showed that aspiration through DAC was applied in 30 patients during withdrawal of the Trevo device. To support the aspiration step during the withdrawal of the Trevo retriever, data was reviewed for this 30 TREVO patient cohort. For all 30 patients the DAC was placed at or near the proximal face of the thrombus in the intracranial artery.

{11}------------------------------------------------

The overall review of the DAC cohort in the TREVO study, showed that safety and efficacy was similar to the overall study population. The mean procedure time was not increased by the additional step of applying adjunctive aspiration during the withdrawal of the retriever. Although this is a small cohort, a high revascularization rate of 96.7% was observed. The ten point improvement in NIHSS at 24 hours and 7 days also supports the utility of the DAC adjunctive aspiration during withdrawal of the retriever device. None of the procedure related or CEC adjudicated events were related to DAC which supports the safety of the DAC during thrombectomy procedures.

Post-Market Sureillance

In addition to the data review of the TREVO2 and TREVO trials, a review of post market surveillance product complaints and Medical Device Reports (MDRs) was conducted to assess the top complaints reported for DAC and to assess total number of reportable events. The top device complaints for the DAC are catheter kinks, lumen collapse or ovalization which is consistent with complaints routinely reported with this class of access devices. The overall complaint rate for device problem or patient effect is low for DAC (0.21%).

A review of the MDRs for DAC identified a total of 10 MDRs filed from 2008 to July 31, 2013. The overall number of reportable events was small and no new risks were identified from the post market surveillance database and the MAUDE database.

Literature Review

Literature was searched pertaining to DAC use in acute ischemic stroke to support safety of DAC during thrombectomy procedures. A total of 15 published articles were reviewed. The literature search showed that DAC is often used during thrombectomy procedures in acute ischemic stroke patients and is a contributory factor to the overall success of the procedure. DAC provides support in difficult distal anatomy and allows adjunctive aspiration during withdrawal of the retriever from the vessel. Studies showed that the revascularization rates in patients with DAC are similar or better to the acceptable revascularization rates for thrombectomy procedure. There were no unanticipated adverse events reported from using DAC. The literature supports safety and efficacy of DAC during thrombectomy procedures.

Conclusion:

The review of TREVO 2 and TREVO clinical trials, literature and post-market data showed that the Modified HD Guide Catheter is used often during thrombectomy procedures with comparable revascularization rates to previously reported results and no new patient risks were identified. The device provides better access to distal anatomy and support to the microcatheter and retrieval devices during insertion and allows aspiration of clots during withdrawal of the retriever device.

{12}------------------------------------------------

Summary of Substantial Equivalence:

Because the proposed expansion in indications for the Modified HD Guide Catheter does not alter the fundamental scientific technology of the predicate devices; and because risk assessments and successful validation testing and review of the clinical data from TREVO2 and TREVO studies, literature and post-market data raise no new questions of safety and effectiveness, Concentric Medical has determined that the Modified HD Guide Catheter with the expanded indications for use to be as safe, as effective, and performs as well as or better than the predicate devices.

{13}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/13/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure, with three overlapping shapes representing the body, arms, and head. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 25, 2014

Concentric Medical, Inc. % Ms. Rhoda Santos Principal Regulatory Affairs Specialist 301 E. Evelyn Avenue Mountain View, CA 94041

Re: K133177

Trade/Device Name: Concentric HD Guide Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Additional product code: DQO Dated: November 26, 2013 Received: November 27, 2013

Dear Ms. Santos:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{14}------------------------------------------------

Page 2 - Ms. Rhoda Santos

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYowIndustry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Carlos L. Pena -S

Carlos L. Peña, Ph.D., M.S. Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{15}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K133177

Device Name

Concentric HD Guide Catheter

Indications for Use (Describe)

The Concentric HD Guide Cathered for use in facilitating the insertion and guidance of an occlusion catheter, infusion cather or other appropriate microcatheter into a selected blood vessel in the peripheral, coronary and neuro vascular systems. It may also be used as a diagnostic angiographic catheter and as a conduit for retrieval devices.

Type of Use (Select one or both, as applicable)

[x] Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

THE CONTRACT THE CONFORT FOR FORFEDAUSE CONLY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Carlos-DPeña -S

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

See PRA Statement on last page.

{16}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

*An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).