510(k) Data Aggregation

    K Number
    K251660

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-06-27

    (28 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K122300, K063286, K082639, K220978

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The T3® Pro Implants are intended for surgical placement in the upper or lower jaw to provide a means for prosthetic attachment in single tooth restorations and in partially or fully edentulous spans with multiple single teeth utilizing delayed loading, or with a terminal or intermediary abutment for fixed or removable bridgework, and to retain overdentures.

    The T3® Pro Implants may also utilize immediate loading for these indications. The T3® Pro Implants are intended for immediate function on single tooth and/or multiple tooth applications when good primary stability is achieved, with appropriate occlusal loading, in order to restore chewing function.

    Device Description

    The T3® Pro Implants with the new ZimVie Implant Packaging System. The material and design of the subject device is identical to the primary predicate device T3® Pro Implants cleared in K213672.

    The ZimVie Implant Packaging System configuration, similarly to the primary predicate device, consists of a titanium sleeve that is inserted into a polypropylene inner tray, covered with a Tyvek lid and heat-sealed. This assembly is then placed in a heat-sealed outer tray, covered with a Tyvek® lid. The outer tray is placed inside a cardboard box, and gamma sterilized. The new ZimVie Packaging System has been standardized to align the packaging across all ZimVie implant families. The subject device in the new ZimVie Packaging System has a shelf-life of two (2) years.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and associated documentation for the "T3 Pro Implants- ZimVie Packaging System" (K251660) do not describe acceptance criteria or a study proving the device meets those criteria in the context of an AI/human reader performance study.

    Instead, this submission is for a medical device (endosseous dental implant) and its new packaging system. The "performance data" section explicitly states that no clinical data was included in the submission for the substantial equivalence determination. The focus of the provided document is to demonstrate equivalence to a previously cleared predicate device for the dental implant itself, with the primary change being the packaging system.

    Therefore, I cannot fulfill your request to create a table of acceptance criteria and reported device performance related to an AI-assisted or standalone AI system, nor can I provide details on sample size, expert adjudication, MRMC studies, or ground truth establishment, as these concepts are not discussed or relevant to the provided documentation for this specific 510(k) clearance.

    The information provided describes non-clinical performance data related to the physical device and its packaging:

    • Sterilization validation: Gamma irradiation sterilization to a sterility assurance level of 10^-6 according to ISO 13004 and ISO 11137.
    • Aging studies: Demonstrating a shelf life of two (2) years for the implants, with packaging sterile barrier testing per ASTM F1980.
    • Packaging studies: Simulated distribution and environmental conditioning per ISTA 3A and ASTM D4169.
    • Biological evaluation: Leveraged from the primary predicate device, demonstrating no increased risks regarding biocompatibility. The material is Commercially Pure Titanium conforming to ASTM F67.
    • MR Safety testing: Conducted according to ASTM F2503-13 on the primary predicate device (identical to the subject device) for MR Conditional determination.
    • Bacterial Endotoxin Testing: Routinely performed using LAL assay in accordance with ISO 11737-3.

    The acceptance criteria for this submission would primarily revolve around demonstrating that the new packaging system maintains the sterility, integrity, and safety of the dental implant for its stated shelf life, and that the dental implant itself remains identical to the previously cleared predicate device in all other aspects (materials, design, intended use, operating principle).

    In summary, your request pertains to an AI/Software as a Medical Device (SaMD) study, while the provided document describes the 510(k) clearance process for a physical dental implant device and its new packaging.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241753

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-10-23

    (127 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K220978, K013227, K150571, K210523, K191054

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Immediate Molar Dental Implants are intended for surgical placement in the upper or lower jaw to provide a means for prosthetic attachment in single tooth restorations and in partially or fully edentulous spans with multiple single teeth utilizing delayed loading, or with a terminal or intermediary abutment for fixed or removable bridgework, and to retain overdentures.

    Immediate Molar Dental Implants may also utilize immediate loading for these indications. Immediate Molar Dental Implants are intended for immediate function on single tooth and/or multiple tooth applications when good primary stability is achieved, with appropriate occlusal loading, in order to restore chewing function. The Immediate Molar Dental Implants may be placed immediately following an extraction or loss of natural teeth provided there is sufficient volume of alveolar bone to provide good primary stability. The Immediate Molar Dental Implants are intended for implantation in the maxillary or mandibular molar region where bone exists and the surgeon has determined that the placement of a narrower diameter implant would increase the probability of failure due to poor primary stability, or increased surgical procedures leading to complications.

    The Immediate Molar Dental Implants are also indicated for compatibility with the following OEM abutment systems:

    Abutment System NameModelsPlatform Diameters
    Abutments manufactured byTerrats Medical SLTitanium Abutments4.1, 5.0 and 5.7mm
    Abutments manufactured byZfx GmbHTitanium Abutments4.1, 5.0 and 5.7mm
    Device Description

    The Immediate Molar Implants are basic screw-type designs available in tapered body geometry. The devices are manufactured from Commercially Pure Titanium (ASTM F67) and feature a roughened apex and traditional OSSEOTITE® coronal surface. The device is packaged in a Titanium sleeve that is inserted into a polypropylene inner tray, covered with a Tyvek lid and heat-sealed. This assembly is then placed inside a larger polyethylene thermoformed outer tray, covered with a Tyvek lid and heat-sealed. The outer tray is packaged inside a box. The device is sold sterile. The shelf life of the Immediate Molar Implants is 5 years and they are intended for single use only. The device is sterilized using gamma irradiation method. The implants are available in various platform options and feature an internal hex connection for mating with associated Biomet 3i internal connection restorative components and also a TSV connection for mating with associated Zimmer Dental TSV connection restorative components. The implants are also compatible with titanium abutments manufactured by Terrats Medical SL and Zfx GmbH. The implants are offered in a variety of diameters and lengths to accommodate varying patient anatomy. The T3 PRO Immediate Molar Implants are offered in 8, 10 and 11.5mm implant lengths for each of the implant body diameter sizes of 7, 8 and 9mm. The TSX Immediate Molar Implants are offered in 6, 8, 10 and 11.5mm implant lengths for each of the implant body diameter sizes of 7 and 8mm. The TSX Immediate Molar Implants are offered in 8, 10 and 11.5mm lengths for implant body diameter size of 9mm.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (Immediate Molar Implants) and does not contain information about acceptance criteria or a study proving that an AI/software device meets acceptance criteria.

    The document explicitly states: "No clinical data were included in this submission." and focuses on demonstrating substantial equivalence to predicate devices based on technological characteristics, non-clinical testing (fatigue, MR compatibility, pull-out tests, surface area analysis, sterilization, biocompatibility, shelf life), and contractual agreements for compatible components.

    Therefore, I cannot provide the requested information about acceptance criteria and a study proving a device meets acceptance criteria using clinical or performance data, as this information is not present in the provided text. The document describes a dental implant, not an AI or software-based device that would typically have acceptance criteria presented in the manner requested.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.

    All digitally designed custom abutments for use with DESS Ti Base abutments or Pre-Milled Blank abutments are to be sent to a Terrats Medical validated milling center for manufacture.

    Device Description

    The purpose of this submission is to add components to the DESS Dental Smart Solutions system, which includes dental implants, abutments, and prosthetic components cleared previously in various submissions. The previously cleared abutments and prosthetic components are compatible with a variety of original equipment manufacturer (OEM) dental implants as well as DESS Dental Smart Solutions dental implants.

    This submission adds various abutments to the DESS and OEM implant lines as summarized on the following pages in Table 1 Summary of Subject Device Abutment Designs, and Table 2 Summary of Subject Device Abutment Sizes.

    The subject device abutment designs include Cover Screws, Healing Abutments, Temporary Abutments, Ti Base Abutments, AURUM Base Abutments (Ti Base abutments with a screw channel design that allows easier instrument access to the abutment screw), CoCr Base Abutments, Pre-Milled Blank Abutments, Multi-Unit Abutments (straight and angled), DESSLoc Abutments (Locator-type abutments), and abutment screws.

    AI/ML Overview

    This document is a 510(k) summary for the DESS Dental Smart Solutions, a line of endosseous dental implant abutments. It details the device's characteristics, intended use, and a comparison to predicate devices to establish substantial equivalence.

    Based on the provided text, the device itself is a physical medical device (dental implant abutments and associated components), not a software or AI-driven system. Therefore, the questions related to AI acceptance criteria, training/test sets, expert adjudication, MRMC studies, and ground truth establishment for AI would not be directly applicable to this product as described.

    The document focuses on establishing substantial equivalence to previously cleared predicate devices through:

    • Identical Intended Use: The device is intended for "functional and esthetic rehabilitation of the edentulous mandible or maxilla" by providing support for prosthetic restorations, which is the same as the predicate devices.
    • Similar Technological Characteristics: The device utilizes similar designs, materials (Ti-6Al-4V, Co-Cr-Mo alloy), manufacturing processes, and sterilization methods as its predicates.
    • Performance Data: Non-clinical testing (mechanical testing per ISO 14801, MR environment assessment) and comparison to existing data from predicate devices are used to demonstrate safety and effectiveness.

    Therefore, many of the requested elements for describing AI acceptance criteria and studies are not present or applicable in this document.

    However, I can extract information relevant to the device's performance assessment and criteria for its type of submission.

    Here's an attempt to answer the questions based solely on the provided text, acknowledging that the nature of the device (a physical implant component) means many AI-specific questions will be answered as "Not Applicable" (N/A):

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state quantifiable "acceptance criteria" in a table format with corresponding "reported device performance" metrics for this specific 510(k) submission. Instead, it relies on demonstrating substantial equivalence to existing predicate devices.

    The underlying "acceptance criteria" for demonstrating substantial equivalence for this device type are primarily through:

    • Mechanical Testing (ISO 14801): This is a critical performance standard for dental implants and abutments. The document states that mechanical testing was "conducted according to ISO 14801 to support the performance." The acceptance criteria would be successful completion of these tests, demonstrating the device's mechanical strength and fatigue resistance are comparable to or better than predicate devices. The reported performance is simply that the tests supported the performance.
    • Material Conformance: Materials must conform to specific ASTM standards (e.g., ASTM F136 for Ti-6Al-4V, ASTM F1537 for Co-Cr-Mo). The reported performance is that the materials conform to these standards.
    • Biocompatibility: While not detailed in this excerpt, the mention of "biocompatibility" in relation to predicates implies conformance to relevant biocompatibility standards (e.g., ISO 10993 series). The reported performance is that it is compatible.
    • Sterilization Validation: Demonstrated sterility assurance level (SAL) of 10⁻⁶ via validated methods (moist heat or gamma irradiation). The reported performance is that validation was performed and met this SAL.
    • Dimensional Compatibility: The abutments must fit the corresponding OEM implants correctly. The reported performance is that reverse engineering dimensional analysis confirmed compatibility.

    Due to the nature of the document being a 510(k) summary focusing on substantial equivalence rather than a full study report, specific numerical performance results for the device tests are not provided in this text.

    2. Sample sizes used for the test set and the data provenance

    • Sample Size for Mechanical Testing: The document states "mechanical testing conducted according to ISO 14801." For such tests, ISO 14801 typically specifies minimum sample sizes (e.g., 10-11 samples for static strength, typically more for fatigue). The exact number of samples used for this specific submission is not explicitly stated, but it would have followed the standard's requirements.
    • Data Provenance: The mechanical testing and material analyses are assumed to be "non-clinical data submitted or referenced" by the manufacturer, Terrats Medical SL, based in Barcelona, Spain. The "reverse engineering dimensional analysis" was done by Terrats Medical SL or through contractual agreement. This is prospective testing performed to support the 510(k). The document itself does not specify the country of origin for the underlying OEM implant data used for reverse engineering, although the OEM companies are listed (e.g., Astra Tech AB, BioHorizons).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    N/A. This is a physical device. Ground truth, in the context of AI, refers to validated labels for data used to train and test an algorithm. For a physical device, performance is evaluated through engineering and biocompatibility testing against defined standards. There are no "experts" establishing ground truth in the AI sense. Testing would be performed by qualified engineers and technicians.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    N/A. Adjudication methods are typically used in clinical studies involving interpretation (e.g., by radiologists) to resolve discrepancies. This document describes non-clinical performance testing of a physical device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    N/A. This product is a dental implant abutment, not an AI software intended to assist human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    N/A. This is a physical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    N/A. For engineering tests of physical devices, the "ground truth" is typically derived from established engineering principles, international standards (e.g., ISO 14801 for mechanical properties, ASTM for materials), and the physical properties of the materials and designs themselves. There isn't "expert consensus" or "pathology" in the AI or clinical trials sense.

    8. The sample size for the training set

    N/A. This is a physical device; there's no "training set" in the machine learning sense. The device is manufactured based on established engineering designs and material specifications.

    9. How the ground truth for the training set was established

    N/A. No training set for AI. For device manufacturing, the "ground truth" for design and production parameters comes from established engineering best practices, prior successful device designs (predicate devices), and adherence to quality systems regulations (21 CFR Part 820).

    Ask a Question

    Ask a specific question about this device

    K Number
    K231915

    Validate with FDA (Live)

    Device Name
    Zfx Abutments
    Date Cleared
    2024-03-01

    (246 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Zfx Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandbular arch to provide support for prosthetic restorations.

    All digitally designed custom abutments or superstructures and/or hybrid abutment crowns for use with Zfx TiBase or Pre-Milled Blank Abutments are intended to be sent to a Zfx validated milling center for manufacture.

    The Zfx TiBase abutments for the 03.25 mm Certain implant bodies, and 03.1 mm Eztetic implant bodies are indicated for maxillary lateral and mandibular central/lateral incisors only.

    Device Description

    The subject device, Zfx Abutments, consists of two abutment types, a TiBase and a Pre-milled Abutment Blank, with corresponding retaining screws. Zfx Abutments are offered in a variety of connections compatible with ZimVie dental implants. All abutments and screws are provided non-sterile.

    TiBase Abutments: Subject device TiBases are two-piece abutments. The pre-manufactured titanium base component is the apical part. The coronal part of the two-piece abutment is a CAD/CAM designed and manufactured superstructure. The subject device TiBase abutments are available in hexed/engaging and non-hexed/nonengaging configurations. The engaging TiBase abutments are intended for single and multi-unit restorations and the non-engaging TiBases are intended for multi-unit restorations. The superstructure is intended to be manufactured at a Zfx validated milling center.

    Pre-milled Abutment Blanks: The Pre-milled Abutment Blank is a cylindrical titanium alloy abutment designed for patient-specific abutment fabrication with CAD/CAM technology. The patient-specific abutment milled from a Pre-milled Abutment Blank is secured directly to the implant using a retaining screw. Pre-milled Abutment Blanks are available in an engaging/hexed design only for single-unit and multi-unit restorations.

    Retaining Screws: Corresponding retaining screws are packaged with the abutment and replacement screws are available individually. All screws are compatible with the corresponding ZimVie dental implants.

    Material Composition: All subject device abutments are made of titanium alloy Ti-6Al-4V ELI conforming to ASTM F136. Screws are made of either titanium alloy Ti-6A1-4V ELI conforming to ASTM F136, or Gold-Tite® screws with stainless steel conforming to ASTM F138. Gold-Tite screws have a gold-plating conforming to ASTM B488 and ASTM B571. Zirconia superstructures for use with the TiBase Abutments are made of zirconia conforming to ISO 13356.

    AI/ML Overview

    The provided text is an FDA 510(k) clearance letter for Zfx Abutments, and as such, it focuses on demonstrating substantial equivalence to a predicate device rather than presenting a performance study with detailed acceptance criteria and supporting data in the way a clinical study report or a formal performance evaluation report would.

    Therefore, the information required to answer many of the questions regarding acceptance criteria, sample sizes, expert involvement, and ground truth establishment is not present in this document. The document primarily outlines the device's technical specifications, indications for use, and a comparison to predicate devices, supported by non-clinical testing data (mechanical testing, biocompatibility, sterilization, etc.).

    However, I can extract the information that is available and explain why other requested details are missing.


    Acceptance Criteria and Device Performance (Based on "Equivalence to Marketed Devices" and "Performance Data" sections):

    This document describes a substantial equivalence submission, meaning the acceptance criteria are primarily demonstrated by showing the device is as safe and effective as a legally marketed predicate device. The performance is assessed through non-clinical testing to ensure it meets established standards comparable to the predicate.

    Acceptance Criterion (Implicit for Substantial Equivalence)Reported Device Performance (Non-Clinical/Design)
    Indications for Use Equivalence"Zfx Abutments (subject device) are substantially equivalent in design, function, material, and Indications for Use to the Primary Predicate, DESS Dental Smart Solutions abutments cleared in K222288. All are intended for use with endosseous dental implants in the maxilla and mandible to provide support for single and multi-unit restorations."
    Material Equivalence"All subject device abutments are made of titanium alloy Ti-6Al-4V ELI conforming to ASTM F136." "Screws are made of either titanium alloy Ti-6A1-4V ELI conforming to ASTM F136, or Gold-Tite® screws with stainless steel conforming to ASTM F138." "Zirconia superstructures for use with the TiBase Abutments are made of zirconia conforming to ISO 13356." This is compared to comparable materials in the predicate.
    Product Design and Manufacturing Process Equivalence"All digitally designed subject device abutments and primary predicate abutments are to a validated milling center for manufacture. Subject device abutments are similar in range of sizes, connections, and technological characteristics to the DESS Dental Smart Solutions, K22288." "Design parameters for the TiBase CAD/CAM zirconia superstructure are listed in the following tables (Tables 2-5)." "Design parameters for the CAD/CAM patient-specific abutment using a Pre-milled Abutment Blank are included in Table 6."
    Sterilization Efficacy (Non-clinical)"sterilization validation according to ISO 17665-1"
    Biocompatibility (Non-clinical)"biocompatibility testing according to ISO 10993-1 Table A-1, ISO 10993-5, and ISO 10993-10"
    Reprocessing Validation (Non-clinical)"reprocessing validation according to ISO 17665-2"
    Mechanical Strength/Fatigue (Non-clinical)"mechanical testing according to ISO 14801 to determine that the subject device has sufficient strength for its intended use."
    MR Safety (Non-clinical)"MR Safety testing was conducted according to ASTM F2052, ASTM F2213, ASTM F2182, and ASTM F2119 for a determination of MR Conditional."
    Compatibility with Existing Systems"reverse engineering analysis to confirm compatibility with the Sirona inCoris Meso Blocks" (for TiBase).

    Missing Information and Explanations:

    1. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

      • Not provided. This document details a 510(k) submission based on non-clinical testing for substantial equivalence, not a clinical trial or performance study with human data. The "Performance Data" section explicitly states: "No clinical data were included in this submission." Therefore, there is no "test set" in the context of clinical images or patient data to analyze. The "samples" used were physical samples for mechanical, sterilization, and biocompatibility testing. The data provenance would be from laboratory tests.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

      • Not applicable/Not provided. As no clinical data or clinical "test set" was used, there was no need for expert adjudication to establish ground truth from patient data. The ground truth for mechanical and material properties is established by engineering specifications and industry standards.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable/None. No clinical test set to adjudicate.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This device (Zfx Abutments) is a physical dental implant component, not an AI software/device. Therefore, an MRMC study or AI assistance is not relevant to its performance evaluation for this submission.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. As above, this is a physical medical device, not an algorithm or AI.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for demonstrating substantial equivalence for this device relies on engineering specifications, material standards (e.g., ASTM, ISO), mechanical test results (e.g., forces to fracture), and comparative analysis with the predicate device's established safe and effective performance. There is no clinical outcomes ground truth cited here.
    7. The sample size for the training set:

      • Not applicable. This refers to a dataset for training an AI model. This device is not an AI/software.
    8. How the ground truth for the training set was established:

      • Not applicable. As above, no training set for an AI model.

    In summary, this document is a regulatory clearance letter focused on substantial equivalence for a physical dental device. The "study" proving it meets acceptance criteria consists of various non-clinical bench tests and a comparison to a predicate device's specifications and performance, rather than a clinical study with patients or an AI model's performance evaluation.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.

    Device Description

    The purpose of this submission is to expand the DESS Dental Smart Solutions abutment system by a change in sterilization status to provide products sterile to the end user that were previously cleared to be provided non-sterile. The subject device abutments and abutment screws were cleared previously to be provided non-sterile to the end user in K170588, K191986, K212628, and K22288. All subject device components will now be provided sterile.

    The subject device components include Healing Abutments, Multi-Unit Abutments (0, 17°, and 30°), and abutment screws.

    AI/ML Overview

    The provided text is a 510(k) summary for the DESS Dental Smart Solutions, an endosseous dental implant abutment. It details the device, its intended use, and its substantial equivalence to previously cleared predicate and reference devices. However, this document does not contain the acceptance criteria or a study proving the device meets those criteria in the context of an AI/ML medical device.

    The 510(k) submission for this dental abutment focuses on establishing substantial equivalence based on:

    • Design and Material: The subject device components are identical in design, material (Ti-6Al-4V alloy, DLC coating), and technological characteristics to previously cleared devices.
    • Manufacturing: The manufacturing process is consistent with previously cleared devices.
    • Biocompatibility: Referenced from previous K-clearances.
    • Sterilization: The main change in this submission is expanding the system to provide products sterile to the end-user via gamma irradiation, which was validated by referencing a previous K-clearance (K212538).
    • MR Environment Testing: Non-clinical analysis was performed to evaluate the subject devices in the MR environment, referencing published literature and FDA guidance.
    • Shelf Life Testing: Referenced from a previous K-clearance (K212538) for samples after accelerated aging.

    Therefore, I cannot fulfill your request to describe the acceptance criteria and a study proving an AI/ML device meets them based on the provided text, as this document is not about an AI/ML medical device. It's about dental implant abutments and establishes substantial equivalence through non-clinical performance data and comparison to predicate devices, not through AI/ML performance metrics.

    To provide the information you requested, I would need a document detailing the validation of an AI/ML medical device, which would include definitions of acceptance criteria (e.g., accuracy, sensitivity, specificity), details of training and test datasets, ground truth establishment, and potentially clinical effectiveness studies if applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231434

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-08-14

    (89 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    Why did this record match?
    Reference Devices :

    K142082, K013227, K072589, K220978, K063286, K111216, K212538, K170588, K22269, K213063

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.

    Device Description

    The purpose of this submission is to add components to the DESS Dental Smart Solutions system, which includes abutments compatible with a variety of original equipment manufacturers (OEM) of dental implants. This submission adds abutments, designated DESS Multi-Unit Abutments, for implant lines from Zimmer Dental and Biomet 31 (now collectively, ZimVie Dental). The subject device abutments include Multi-Unit Abutments in straight, 17° angled, and 30° angled designs, which are compatible with implants having internal hex connections. This submission is also to change how previously cleared devices are provided; the change is from previously provided nonsterile to now provided sterile. All abutments are provided with the appropriate abutment screw (if applicable) for attachment to the corresponding implant. The subject device is only intended for multi-unit restorations such as bridges and bars.

    AI/ML Overview

    It appears there might be a misunderstanding of the provided text. The document is a 510(k) premarket notification for a dental implant abutment, not an AI/ML medical device. Therefore, the information typically required for describing an AI/ML device's acceptance criteria and study data (like sample sizes for test and training sets, expert qualifications, ground truth establishment, MRMC studies, or standalone performance) is not present.

    The document discusses the substantial equivalence of the DESS Dental Smart Solutions abutment to legally marketed predicate devices. The "performance data" section refers to non-clinical testing to demonstrate the device's physical properties and compatibility, not a study evaluating an AI algorithm's diagnostic or predictive performance.

    Let's break down what can be extracted from the provided text relevant to the device's acceptance, and then explain why the AI/ML specific questions cannot be answered.

    Understanding the Device and its "Acceptance":

    The "acceptance" in this context refers to the FDA's determination that the new DESS Dental Smart Solutions abutments are "substantially equivalent" to predicate devices already on the market. This determination is based on the device having the same intended use and similar technological characteristics (design, materials, performance via non-clinical testing).

    Here's a summary of the information available in the document:

    1. Table of Acceptance Criteria (or rather, Equivalence Criteria) and Reported Device Performance:

    The document doesn't present acceptance criteria in a quantitative table with performance metrics like sensitivity, specificity, or accuracy, as would be common for an AI/ML device. Instead, "acceptance" is based on demonstrating substantial equivalence to predicate devices across several characteristics. The performance is shown through non-clinical testing to ensure safety and functionality.

    CharacteristicSubject Device (DESS Dental Smart Solutions, K231434)Primary Predicate (K230143, DESS Dental Smart Solutions)Reference Device (K213063, TLX SRAs and TLX Gold Abutments)Reported Performance (Non-Clinical Data)
    Intended UseDESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.DESS Multi-Unit Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.Functional and esthetic rehabilitation of the edentulous mandible or maxillaStated as "same intended use" as predicates.
    Abutment Designs/TypesMulti-Unit, straight (0°), 17° angled, 30° angledMulti-Unit, straight (0°), 17° angled, 30° angledMulti-Unit, TLX SRA: 17°, 30°; TLX Gold: 0°, and up to 30°Similar to predicates.
    Prosthesis AttachmentScrew RetainedScrew RetainedScrew RetainedSimilar to predicates.
    RestorationMulti-unitMulti-unitMulti-unitSimilar to predicates.
    Prosthetic Interface ConnectionsInternal hex, External hexInternal hex, Internal conicalInternal conicalSimilar to predicates, with some variations noted and justified.
    Abutment/Implant Platform Diameter2.9 - 5.7 mm3.0 - 4.5 mmTLX SRA: 6 mm; TLX Gold: 4.0 (NT), 5.0 (RT), and 7.0 (WT)Within similar ranges or justified compatibility.
    Prosthetic Platform Diameter4.8 mm4.8 mmTLX SRA: 4.6 mm; TLX Gold: not provided in 510(k) SummarySimilar to predicates.
    Gingival Height1 mm - 5 mm1 mm - 5 mmNot provided in 510(k) SummarySimilar to predicates.
    Abutment MaterialTi-6Al-4V ELI (Titanium alloy)Ti-6Al-4V ELI (Titanium alloy)TLX SRA: Ti-6Al-7Nb; TLX Gold: Ceramicor®Similar (titanium alloys). Biocompatibility tested (ISO 10993-5 referenced).
    Abutment Screw MaterialTi-6Al-4V ELI (Titanium alloy)Ti-6Al-4V ELI (Titanium alloy)Ti-6Al-7NbSimilar (titanium alloys).
    How Provided (Sterilization)Non-sterile, and sterile by gamma irradiationNon-sterileTLX SRA: Sterile by gamma irradiation; TLX Gold: non-sterileBoth non-sterile and sterile options justified via reference devices/testing (e.g., K212538 for gamma, K22269 for moist heat).
    UsageSingle patient, single useSingle patient, single useSingle patient, single useIdentical.
    Mechanical PerformanceEvaluated according to ISO 14801 (mechanical testing); Engineering analysis for new worst-case constructs.Referenced-Demonstrated to not create a new worst-case or met ISO 14801.
    MR SafetyEvaluated according to ASTM F2052, ASTM F2213, ASTM F2182, ASTM F2119.--Non-clinical analysis provided to evaluate metallic devices in MR environment.
    Sterility & Shelf LifeRefer to K22269 (moist heat), K212538 (gamma irradiation, bacterial endotoxin, shelf life).Referenced-Validated to SAL of 10^-6, met bacterial endotoxin limits, and 5-year shelf life after accelerated aging.

    Important Note: The "acceptance criteria" here are qualitative statements of substantial equivalence, backed by quantitative non-clinical engineering and materials testing, not performance metrics from a clinical study or an AI model's accuracy.


    Regarding the AI/ML Specific Questions:

    2. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Not applicable. This document is for a physical medical device (dental implant abutment), not an AI/ML algorithm. There is no "test set" in the context of data for an AI/ML model. The "testing" refers to mechanical, material, and sterilization validation.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. Ground truth establishment by experts is relevant for AI/ML diagnostic devices. This device's performance is assessed through engineering standards and material properties. Compatibility with implant systems is established through design agreements and reverse engineering data (referenced from K170588).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. Adjudication methods are used in clinical studies or for establishing ground truth labels for AI/ML datasets.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. MRMC studies are for evaluating diagnostic efficacy, particularly with AI assistance. This device is a passive prosthetic component.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. There is no algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not applicable. The "ground truth" for this device's performance is its adherence to engineering standards (e.g., ISO 14801 for mechanical properties, ASTM for materials/MR safety) and its demonstrated compatibility with specified implant systems, rather than a clinical ground truth like disease presence.

    8. The sample size for the training set:

    • Not applicable. There is no AI model or training set.

    9. How the ground truth for the training set was established:

    • Not applicable. As above, no AI model or training set.

    Conclusion from the document:

    The provided document describes a 510(k) submission for a Class II medical device, DESS Dental Smart Solutions abutments. The "acceptance criteria" revolve around demonstrating substantial equivalence to existing legally marketed predicate devices. This is achieved through:

    • Identical intended use.
    • Similar technological characteristics: This includes materials (Titanium alloy), design (multi-unit, screw-retained), and geometric parameters (diameters, gingival heights).
    • Performance via non-clinical testing: This includes mechanical testing (ISO 14801), MR safety testing (ASTM standards), sterilization validation (ANSI/AAMI/ISO standards), bacterial endotoxin testing (ANSI/AAMI ST72), shelf-life testing (ASTM F1980), and biocompatibility testing (ISO 10993-5).
    • Compatibility with specified implant systems: This is supported by contractual agreements with OEMs (ZimVie/Terrats Medical SL) and referenced reverse engineering data.

    The document explicitly states: "No clinical data were included in this submission." This further confirms that the evaluation framework is entirely non-clinical and does not involve AI/ML performance metrics, studies with human readers, or expert ground truthing as one would see for an AI-powered diagnostic tool.