K Number
K122300

Validate with FDA (Live)

Manufacturer
Date Cleared
2013-01-30

(183 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

BIOMET 3/ Dental implants are intended for surgical placement in the upper or lower jaw to provide a means for prosthetic attachment in single tooth restorations and in partially or fully edentulous spans with multiple single teeth, or as a terminal or intermediary abutment for fixed or removable bridgework, and to retain overdentures.

Device Description

The 3i T3 Dental Implants are manufactured from Commercially Pure Grade 4 titanium and feature a roughened apex and traditional OSSEOTITE® coronal surface. The dental implants will consist of a straight wall or tapered body type with a basic screw-type design in various platform options and feature an internal connection/anti-rotation feature; 3.25 and 4/3mm has a 12pt dual hex; 4, 5, 6, 5/4 & 6/5mm has a six-point hex at the top and lower 12-point dual hex. The 3i T3 Dental Implants are available with either the Prevail platform switching feature or standard collar. In addition, the implants are offered with and without the nano-scale discrete crystalline deposition (DCD) calcium phosphate (CaP) surface treatment.

AI/ML Overview

The provided document is a 510(k) Summary for the BIOMET 3i T3 Dental Implants. This document focuses on demonstrating substantial equivalence to previously cleared devices through non-clinical performance data. It does not describe an AI medical device or a study involving AI with human readers. Therefore, most of the requested information regarding AI device acceptance criteria and study design is not available in this document.

However, I can extract information related to the acceptance criteria and the studies performed for this non-AI medical device.

1. Table of Acceptance Criteria and Reported Device Performance

Characteristic / TestAcceptance Criteria (Requirement)Reported Device Performance (Results Summary)
New Device (3i T3 & 3i T3 with DCD Implants)
Cyclic Fatigue TestingISO 14801:2007 (Standard/Test/FDA Guidance)Met indications
Print VerificationBiomet 3i procedureComparison of the original and design verification test models confirms the prints convey the design intent.
Comparative Performance (New Device vs. Predicate)
3i T3 vs. 3i T3 with DCD Implants - Cyclic Fatigue TestingMeet or exceed parametersMet parameters of K100724 (predicate)
3i T3 with DCD Nano-Scale Calcium Phosphate Adhesion StrengthMeet or exceed parametersExceeded parameters of K051461 (predicate)
Tolerance Analysis 3iT3 & 3iT3 with DCD ImplantsMeet or exceed parametersMet parameters of K100724 (predicate)
3iT3 and 3iT3 with DCD Implants - Torque TestingMeet or exceed parametersMet parameters of K100724 (predicate)
3iT3 & 3iT3 with DCD Implants - Fit Check/Mating AnalysisMeet or exceed parametersMet parameters of K100724 (predicate)

2. Sample Size Used for the Test Set and Data Provenance

This document does not specify exact sample sizes for the non-clinical tests (e.g., how many implants were cycle-tested). The clinical data mentioned refers to postmarket clinical studies of predicate implants, not the new device being tested for equivalence.

  • Clinical Data (from predicate implants): Data on over 6,829 implants placed in the posterior from 1996 - 2011.
  • Data Provenance: The provenance of the 6,829 implants is "sponsored or supported by Biomet 3i," implying it's from their own studies or collected data. No specific country of origin is mentioned, but Biomet 3i is a US-based company (Palm Beach Gardens, Florida). The studies are explicitly described as "postmarket clinical studies" of predicate implants, indicating retrospective data collection for the clinical portion. The non-clinical tests were performed on the new device, as part of its premarket submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. This device is a dental implant, and the testing involves mechanical and material properties, not interpretation of images or data by human experts for establishing ground truth as would be relevant for an AI diagnostic device.

4. Adjudication Method for the Test Set

Not applicable. This is not an AI diagnostic device requiring expert adjudication for ground truth.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

Not applicable. This is not an AI diagnostic device.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Not applicable. This is not an AI medical device. The "standalone" performance here refers to the device itself meeting mechanical and material standards without human interpretation.

7. Type of Ground Truth Used

For the non-clinical tests, the "ground truth" is based on established engineering standards, test procedures, and design specifications. For example:

  • Mechanical Integrity: ISO 14801:2007 for cyclic fatigue testing.
  • Design Intent: Biomet 3i internal procedures for print verification, ensuring manufacturing matches design.
  • Material Properties: Adhesion strength, tolerance analysis, and torque testing against established parameters for the predicate devices.

For the postmarket clinical data, the ground truth is clinical survival rates of predicate implants, based on actual patient outcomes over several years.

8. Sample Size for the Training Set

Not applicable. There is no training set mentioned, as this is not an AI device.

9. How the Ground Truth for the Training Set Was Established

Not applicable. There is no training set for an AI model.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the text "BIOMET 3i" in a stylized, bold font. The letters are outlined, giving them a three-dimensional appearance. The "3i" portion of the text is smaller and has a trademark symbol next to it.

K 1.22300

JAN 3 0 2013

510(k) SUMMARY

A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92

Submitter Information
NameBIOMET 3i
Address4555 Riverside DrivePalm Beach Gardens, Florida 33410
Phone number(561) 776-6840
Fax number(561) 514-6316
EstablishmentRegistration Number1038806
Name of contactpersonJacquelyn Hughes
Date preparedJanuary 17, 2012
Name of device
Trade or proprietaryname3i T3 Dental Implant
Common or usualnameEndosseous Dental Implants
Classification nameImplant, Endosseous, Root-Form
Classification panelDental
Regulation21CFR §872.3640
Product Code(s)DZE
Legally marketeddevice(s) to whichequivalence is claimedK100724 OSSEOTITE® 2 Certain ImplantsK063341 Certain® ImplantsK051461 NanoTite® ImplantsK00321 ITI Dental Implant System
Reason for 510(k)submissionAddition to BIOMET 3i dental implant product line to includean implant with a multi-level surface topography by adding aCalcium Phosphate (CaP) media-blasted roughened surfaceon the apical aspect of existing BIOMET 3i Certain internalconnection OSSEOTITE 2 (K100724) and internalconnection Tapered (K063341) product lines.
Device descriptionThe 3i T3 Dental Implants are manufactured fromCommercially Pure Grade 4 titanium and feature aroughened apex and traditional OSSEOTITE® coronalsurface. The dental implants will consist of a straight wallor tapered body type with a basic screw-type design invarious platform options and feature an internalconnection/anti-rotation feature; 3.25 and 4/3mm has a12pt dual hex; 4, 5, 6, 5/4 & 6/5mm has a six-point hex atthe top and lower 12-point dual hex. The 3i T3 DentalImplants are available with either the Prevail platform
Intended use of the deviceswitching feature or standard collar. In addition, the implants are offered with and without the nano-scale discrete crystalline deposition (DCD) calcium phosphate (CaP) surface treatment.
The 3/T3 Dental Implants are intended for immediate function on single tooth and/or multiple tooth applications when good primary stability is achieved, with appropriate occlusal loading, in order to restore chewing function.
Indications for useBIOMET dental implants are intended for surgical placement in the upper or lower jaw to provide a means for prosthetic attachment in single tooth restorations and in partially or fully edentulous spans with multiple single teeth utilizing delayed or immediate loading, or with a terminal or intermediary abutment for fixed or removable bridgework, and to retain overdentures.

4555 Riverside Drive
Palm Beach Gardens, FL 33410 Office: 561.776.6700 Fax: 561.776.6762
www.3i -online.com

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for BIOMET 3i. The logo is in black and white and features a stylized font. The "i" in "3i" has a small box above it with the letters "TM" inside.

Summary of the technological characteristics of the device compared to the predicate

CharacteristicNew DeviceK100724K063341K051461K003271
Description3i T3 DentalOsseotite 2CertainNanotiteDental
ImplantsDentalImplantsImplantsImplant
ImplantsSystem
MaterialCP4 TitaniumCP4 TitaniumCP TitaniumCP TitaniumCP4
(ASTM F67)(ASTM F67)(ASTM F)(ASTM F67)Titanium
Ti -6AL-4VTi -6AL-4V(ASTM F67)
(ASTM F136)(ASTM F136)
SurfaceGrit BlastAcid-etchAcid-etchAcid-etchGrit Blast
FinishAcid-etchDCDAcid-etch
DCD
ImplantStraight-WallStraight-wallStraight-WallStraight-WallSolid self-
DesignTaperedTaperedTaperedtapping
Collar DesignStandardStandardStandardStandardUnknown
PrevailPrevailPrevail
DiameterStandard:Standard:Standard:Standard:various
3.25- 6mm3.25- 6mm3.25- 6mm3.25- 6mm
Prevail:Prevail:Prevail:
4/3-6/5mm4/3-6/5mm4/3-6/5mm
Length8.5-18mm8.5-18mm7-20mm8.5-20mmvarious
ConnectionInternalInternalInternalInternalExternal
LabelingSterileSterileSterileSterileSterile
PERFORMANCE DATA

4555 Riverside Drive
Palm Beach Gardens, FL 33410
Office: 561.7776.6700
Fax: 561.7776.6762
www.3i -online.com

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for BIOMET 3i. The logo is in black and white and features a stylized font. The "3i" portion of the logo is smaller and to the right of the word "BIOMET".

SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF
SUBSTANTIAL EQUIVALENCE
Performance Test Summary-New Device
CharacteristicStandard/Test/FDA GuidanceResults Summary
3/T3 & 3/T3 with DCDISO 14801:2007Cyclic Fatigue testing met
Implants - Cyclic Fatigueindications
Testing
3/T3 & 3/T3 with DCDBiomet 3i procedureComparison of the original
Implants - Print Verificationand design verification test
models confirms the prints
convey the design intent.
Comparative Performance Information Summary
CharacteristicRequirementNew DevicePredicate Device
31 T3 vs. 31T3 with DCDMeet or exceedMeetK100724
Implants - Cyclic Fatigueparameters
Testing
3/T3 with DCD Nano-ScaleMeet or exceedExceedK051461
Calcium Phosphateparameters
Adhesion Strength
Tolerance Analysis 3iT3 &Meet or exceedMeetK100724
3/T3 with DCD Implants -parameters
Tolerance Analysis
3/T3 and 3/T3 with DCDMeet or exceedMeetK100724
Implants - Torque Testingparameters
3iT3 & 3iT3 with DCDMeet or exceedMeetK100724
Implants - Fit Check/parameters
Mating Analysis
SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF
SUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION
Clinical Performance Data/Information: Data was provided from postmarket clinical studies
of the prodicato implants cronented or cunnorted by Biomet 22. Data on over 6. 200

sponsored or supported by Biomet 31. Data on over 6,829 implants placed in the posterior from 1996 -2011 were available to demonstrate the difference in clinical survival rates for the various diameters of implants. None of the implant cases included fracture as an etiology for implant failure.

CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA

No additional clinical testing was necessary for a determination of substantial equivalence.

The results of mechanical testing indicated the devices performed within the intended use, did not raise any new safety and efficacy issues and were found to be substantially equivalent to the predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The logo is black and white.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 30, 2013

Ms. Jacquelyn A. Hughes, RAC Director, Regulatory Affairs & Clinical Research Biomet 31 4555 Riverside Drive PALM BEACH GARDENS FL 33410

Re: K122300

i

Trade/Device Name: 3i T3TM Dental Implants Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE Dated: January 9, 2013

Received: January 14, 2013

Dear Ms. Hughes:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

K122300 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: 3i T3™ Dental Implants

Indications For Use:

BIOMET 3/ Dental implants are intended for surgical placement in the upper or lower jaw to provide a means for prosthetic attachment in single tooth restorations and in partially or fully edentulous spans with multiple single teeth, or as a terminal or intermediary abutment for fixed or removable bridgework, and to retain overdentures.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use NO (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Susan Runner DDS, MA2013.01.24
15:26:11 -05'00'

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number:

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.