    Intended Use

    The Millipede 088 Access Catheter is indicated for use in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the neurovasculature.

    Device Description

    The Millipede 088 Access Catheter consists of the catheter, a rotating hemostasis valve (RHV) and a valve crossing tool. The catheter, RHV and valve crossing tool are provided sterile. They are sterilized by ethylene oxide (EO).

    The Millipede 088 Access Catheter is a single lumen, coil-reinforced, variable stiffness catheter. The distal segment has a hydrophilic coating for navigation through the vasculature. The catheter has a radiopaque marker located at its distal end for visualization under fluoroscopy. The valve crossing tool is used to open the valve of the access sheath and to facilitate insertion of the Millipede 088 Access Catheter through the access sheath without damage. The RHV is assembled onto the hub of the Millipede 088 Access Catheter and is used to maintain hemostasis during infusion of saline and contrast agent and insertion of other devices through the Millipede 088 Access Catheter.

    AI/ML Overview

    The provided document is a 510(k) summary for the Millipede 088 Access Catheter. This type of regulatory submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than providing detailed acceptance criteria and a standalone clinical study to prove the device meets those criteria.

    Therefore, the document does not contain the information requested in points 2, 3, 4, 5, 8, and 9, and only partially addresses points 1, 6, and 7.

    Here's a breakdown of what can be extracted:

    1. Acceptance Criteria and Reported Device Performance

    The document lists various performance tests and their conclusions, indicating that the device met "established specifications" or was "suitable for its intended use." However, the specific quantitative acceptance criteria for each test are not provided. The reported device performance is qualitative rather than quantitative in most cases.

    TestTest MethodReported Device Performance (Conclusions)
    Dimensional InspectionDevice dimensions were measured to confirm conformance to the specifications.The device met established specifications.
    Tip StiffnessTest specimens were tested for tip flexibility and compared to predicate and reference devices.The device met established specifications.
    Visual InspectionDevice surface characteristics were assessed to confirm freedom from defects.The device surface characteristics are suitable for its intended use.
    Simulated Use TestingDeliverability and compatibility with accessory devices were evaluated in a neurovascular model.The device performs as intended under simulated use conditions.
    Hydrophilic Coating IntegrityThe integrity of the hydrophilic coating was evaluated after multiple insertion and withdrawal cycles.The hydrophilic coating integrity is suitable for its intended use.
    Particulate RecoveryQuantify the particulate size and count generated by simulated use of the test article.The particulate size and count were similar to control devices.
    Tensile StrengthThe tensile strength was evaluated for the bonds between sections of the catheter.The device met established specifications.
    Air LeakageTested per ISO 10555-1:2013 Annex D.The device integrity is suitable for its intended use.
    Liquid LeakageTested per ISO 10555-1:2013 Annex C.The device integrity is suitable for its intended use.
    Static BurstTested per ISO 10555-1:2013 Annex F.The device integrity is suitable for its intended use.
    Luer IntegrityThe luers were evaluated for compliance to relevant standards.The luers on the device are suitable for their intended use.
    Kink ResistanceTest specimen segments were formed into a defined bend diameter to evaluate kink resistance.The device met established specifications.
    Torque StrengthThe test specimens were rotated in a simulated use model to evaluate integrity after rotation.The device met established specifications.
    Flow Rate CharacterizationThe flow rate of saline and a contrast-saline solution was characterized when injected through the catheter.The flow rate was characterized.
    RadiopacityRadiopacity of the device was evaluated in an animal model under fluoroscopy.The radiopacity of the Millipede 088 Access Catheter was similar to a control device.
    Biocompatibility (various tests)ISO 10993-1, 10993-4 (Hemocompatibility), 10993-5 (Cytotoxicity), 10993-10 (Irritation, Sensitization), 10993-11 (Systemic Toxicity, Pyrogenicity)Non-cytotoxic, no sensitization response, met intracutaneous reactivity, met acute systemic injection, non-pyrogenic, not a complement activator, not an intrinsic coagulation pathway activator, non-hemolytic, similar thromboresistance to controls.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the sample sizes for the individual performance tests beyond implicitly suggesting multiple samples were tested (e.g., "test specimens," "test articles"). For the Animal Testing, it states "two studies in a porcine model."

    The data provenance for the in vitro and animal studies isn't explicitly stated beyond "Good Laboratory Practices" for the animal studies, which is a standard of conduct rather than a geographic origin. No human data is presented.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided as the submission relies on objective performance testing, biocompatibility studies, and animal studies rather than expert-derived ground truth from human data for this type of device.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. The evaluation methods described are objective performance tests and animal studies, not human data requiring expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an access catheter, not an AI-powered diagnostic tool. Therefore, an MRMC study or AI assistance is not relevant to its evaluation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the performance testing described (mechanical, dimensional, coating integrity, etc.) and the animal studies represent "standalone" evaluations of the device's physical and functional properties without human interpretation of data in a clinical context.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the performance and biocompatibility tests, the "ground truth" is defined by the established specifications, relevant ISO standards, and comparison to control devices or predicate devices. For example:

    • "The device met established specifications."
    • "The test article is non-cytotoxic."
    • "The particulate size and count were similar to control devices."
    • "The radiopacity of the Millipede 088 Access Catheter was similar to a control device."
    • For animal testing, "Usability, radiopacity, thromboresistance, and vessel injury were assessed," and "The results for the subject device were comparable to a control device."

    8. The sample size for the training set

    Not applicable. This device is a medical catheter and does not involve a "training set" in the context of machine learning or AI. Its performance is evaluated through engineering and biological testing.

    9. How the ground truth for the training set was established

    Not applicable (as above).

    Ask a Question

    Ask a specific question about this device

    K Number
    K211654

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-08-16

    (80 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K190010

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Penumbra Reperfusion Catheters and Separators

    As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA therapy are candidates for treatment.

    Penumbra 3D Revascularization Device

    As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

    Penumbra Aspiration Tubing

    As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump.

    Penumbra Aspiration Pump

    The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

    Device Description

    The Penumbra System® is comprised of the following devices:

    • Penumbra Reperfusion Catheter
    • Penumbra Aspiration Pump
    • Penumbra Aspiration Pump/Canister Tubing
    • Penumbra Aspiration Tubing
    • Penumbra Separator
    • Penumbra 3D Revascularization Device

    The Penumbra System is designed to remove thrombus from the vasculature using continuous aspiration. The Reperfusion Catheter targets aspiration from the pump directly to the thrombus. The 3D Revascularization Device is used with Reperfusion Catheters to facilitate aspiration and removal of the thrombus when needed. The Separator may be used to clear the lumen of the Reperfusion Catheter should it become blocked with thrombus. The use of the Separator may not be necessary when using a Reperfusion Catheter with an ID of 0.054 in. or larger. The Reperfusion Catheter is introduced through a guide catheter or long femoral sheath and into the intracranial vasculature and guided over a neurovascular guidewire to the site of the primary occlusion. The Penumbra Reperfusion Catheter is used with the Aspiration Pump to aspirate thrombus from an occluded vessel. As needed, a Penumbra Separator may be deployed from the Reperfusion Catheter to assist with thrombus removal. The Penumbra Separator is advanced and retracted through the Penumbra Reperfusion Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the Reperfusion Catheter tip. For the aspiration source, the Penumbra Reperfusion Catheter is used in conjunction with the Aspiration Pump, which is connected using the Penumbra Aspiration Tubing and the Penumbra Pump/Canister Tubing. The Penumbra Reperfusion Catheter is provided with a steam shaping mandrel and rotating hemostasis valve, and a peelable sheath. The Penumbra 3D Revascularization Device is provided with an introducer sheath. The Penumbra Separator is provided with an introducer and torque device. The Penumbra Reperfusion Catheters, 3D Revascularization Device and Separators are visible under fluoroscopy.

    AI/ML Overview

    This document describes the acceptance criteria and study proving the performance of the Penumbra System® (Reperfusion Catheter RED™ 72). It is a 510(k) summary, indicating that the device is being cleared based on substantial equivalence to a legally marketed predicate device, rather than requiring a full premarket approval (PMA) with extensive clinical trials.

    It's important to note that because this is a 510(k) submission based on substantial equivalence to an existing device (the Penumbra System (Reperfusion Catheter JET 7)), the studies primarily focus on engineering and biocompatibility performance data against established requirements, rather than clinical efficacy studies often seen with novel AI devices. There are no AI/ML components described in this document, thus the requested information regarding MRMC studies, standalone AI performance, human expert ground truth for imaging, and training/test set sample sizes for AI models are not applicable.

    Here's the breakdown of the acceptance criteria and study details provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document provides a series of design verification tests and biocompatibility tests. The acceptance criteria for each test are stated, and the conclusion is consistently that the "Acceptance Criteria Met" or "Pass."

    Design Verification Testing:

    TestTest Method SummaryAcceptance Criteria Met (Reported Performance)
    Dimensional/Visual TestConfirms the units meet all dimensional and visual product specifications.Yes
    Friction TestConfirms units meet product specification related to friction.Yes
    Fluoroscopy TestConfirms the markerband is fluoroscopically visible.Yes
    Simulated Use TestConfirms the functionality of units using clinically relevant benchtop model.Yes
    Particulate TestParticulates generated during simulated use (including multiple deployment cycling) were evaluated.Yes
    Hub/Air TestConfirms units have no leaks when tested.Yes
    Tensile TestConfirms units meet product specification related to tensile strength.Yes
    Pressure TestConfirms units meet product specification related to pressure.Yes
    Elongation TestConfirms units meet product specification related to elongation.Yes
    Corrosion Resistance TestConfirms there is no visible corrosion on the units when tested.Yes
    Torque Strength TestConfirms units have sufficient torque strength.Yes
    Burst Pressure TestConfirms units can withstand sufficient pressure.Yes
    Shelf-LifeConfirms expiration date based on accelerated aging test studies.Yes
    Sterilization TestConfirms the units are sterilized in accordance with ISO 11135 and ISO 10993-7.Yes

    Biocompatibility Testing (Based on predicate/reference devices):

    TestsAcceptance CriteriaPass/Conclusion (Reported Performance)
    Cytotoxicity: MEM Elution (ISO 10993-5)Sample extracts must have a cytotoxic reactivity score of grade 2 or lower.Pass (Non-cytotoxic)
    Sensitization: Magnusson-Kligman Method (ISO 10993-10)Test Group shall yield Grade < 1 score on Magnusson and Kligman scale (provided Control Group yields Grade < 1).Pass (Non-sensitizing)
    Irritation: Intracutaneous Reactivity (ISO 10993-10)The difference between the average scores for the extract of the test article and the control is ≤ 1.0.Pass (Non-irritating)
    Systemic Toxicity: Acute Systemic Injection (ISO 10993-11)Sample extracts must not cause significant biological reaction greater than control. That is: • Death in 2 or more animals. • Signs of toxicity in 2 or more animals (i.e. convulsions, prostration). • Weight loss > 10% in 3 or more animals.Pass (Non-toxic)
    Systemic Toxicity: Material Mediated Pyrogen (ISO 10993-11)Sample extracts must not cause a total rise in body temperature of ≥ 0.5°C.Pass (Non-pyrogenic)
    Hemocompatibility: Thrombogenicity (ISO 10993-4)The test article must have similar or less thrombus formation than predicate.Pass (Non-thrombogenic)
    Hemocompatibility: Partial Thromboplastin Time (PTT) (ISO 10993-4)Clotting times of test article must be similar to predicate values.Pass (Hemocompatible)
    Hemocompatibility: Complement Activation (ISO 10993-4)The concentration of SC5b-9 of test article must be similar to predicate values.Pass (Hemocompatible)
    Hemocompatibility: Hemolysis, indirect contact (ISO 10993-4)Sample extracts must be non-hemolytic (≤ 2% hemolytic index).Pass (Non-hemolytic)
    Hemocompatibility: Hemolysis, direct contact (ISO 10993-4)Sample must be non-hemolytic (≤ 2% hemolytic index).Pass (Non-hemolytic)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes (number of units) used for each design verification bench test, as is typical for 510(k) summaries of mechanical devices. However, it indicates that units were tested. The data provenance would be from internal testing conducted by Penumbra, Inc. (the manufacturer). This is a retrospective evaluation of the device's physical and functional properties. There is no indication of country of origin for specific test data beyond the manufacturer being in Alameda, California, USA.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. This submission is for a mechanical medical device (catheter system), not an AI/ML device that requires human expert annotation for ground truth in image analysis. The "ground truth" for this device's performance is established by objective physical and chemical testing against engineering specifications and international standards (e.g., ISO, USP).

    4. Adjudication Method for the Test Set

    Not applicable. As there are no human expert readings or interpretations involved in the performance data presented (it's benchmark and material testing), there is no need for an adjudication method.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No. An MRMC study is for evaluating the impact of AI on human reader performance, typically in diagnostic imaging. This device is a mechanical catheter system for stroke intervention, not an AI/ML diagnostic tool. Therefore, no such study was performed or required.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    No. This refers to the standalone performance of an AI algorithm. Since this is a mechanical device, this concept is not applicable. The standalone "performance" refers to the device meeting its physical and functional specifications during bench testing.

    7. The Type of Ground Truth Used

    The "ground truth" used for this device's performance evaluation comes from engineering specifications, international standards (e.g., ISO 10993 series), and validated test methods. This includes:

    • Pre-defined dimensional and visual product specifications.
    • Measured physical properties (friction, tensile strength, pressure, elongation, torque, burst pressure).
    • Chemical and biological properties (cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility) determined through laboratory assays adhering to GLP (Good Laboratory Practices) and ISO standards.
    • Functionality in simulated clinical use benchtop models.

    8. The Sample Size for the Training Set

    Not applicable. This refers to a training set for an AI model. This device is a mechanical medical device, not an AI/ML product.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As there is no AI training set, there is no ground truth establishment for it. The "ground truth" for the device's design and manufacturing is based on established engineering principles, materials science, and medical device regulations.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211411

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-07-01

    (56 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K190010

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Penumbra Reperfusion Catheters and Separators

    As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA therapy are candidates for treatment.

    Penumbra 3D Revascularization Device

    As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

    Penumbra Aspiration Tubing

    As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump.

    Penumbra Aspiration Pump

    The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

    Device Description

    The Penumbra System® is comprised of the following devices:

    • . Penumbra Reperfusion Catheter
    • Penumbra Aspiration Pump .
    • Penumbra Aspiration Pump/Canister Tubing .
    • Penumbra Aspiration Tubing ●
    • . Penumbra Separator
    • Penumbra 3D Revascularization Device .

    The Penumbra System is designed to remove thrombus from the vasculature using continuous aspiration. The Reperfusion Catheter targets aspiration from the pump directly to the thrombus. The 3D Revascularization Device is used with Reperfusion Catheters to facilitate aspiration and removal of the thrombus when needed. The Separator may be used to clear the lumen of the Reperfusion Catheter should it become blocked with thrombus. The use of the Separator may not be necessary when using a Reperfusion Catheter with an ID of 0.054 in. or larger. The Reperfusion Catheter is introduced through a guide catheter or long femoral sheath and into the intracranial vasculature and guided over a neurovascular guidewire to the site of the primary occlusion. The Penumbra Reperfusion Catheter is used with the Aspiration Pump to aspirate thrombus from an occluded vessel. As needed, a Penumbra Separator may be deployed from the Reperfusion Catheter to assist with thrombus removal. The Penumbra Separator is advanced and retracted through the Penumbra Reperfusion Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the Reperfusion Catheter tip. For the aspiration source, the Penumbra Reperfusion Catheter is used in conjunction with the Aspiration Pump, which is connected using the Penumbra Aspiration Tubing and the Penumbra Pump/Canister Tubing. The Penumbra Reperfusion Catheter is provided with a steam shaping mandrel and rotating hemostasis valve, and a peelable sheath. The Penumbra 3D Revascularization Device is provided with an introducer sheath. The Penumbra Separator is provided with an introducer and torque device. The Penumbra Reperfusion Catheters, 3D Revascularization Device and Separators are visible under fluoroscopy.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting studies for the Penumbra System (RED 68 Reperfusion Catheter):

    1. Table of Acceptance Criteria and Reported Device Performance

    Test NameTest Method SummaryAcceptance CriteriaReported Device Performance
    Dimensional/Visual TestConfirms the units meet all dimensional and visual product specifications.Not explicitly stated in terms of specific dimensions/visuals, but implied as meeting product specifications.Acceptance Criteria Met
    Kink Resistance TestConfirms units meet product specification related to kink resistance when formed in a defined radius.Implied as meeting product specifications related to kink resistance.Acceptance Criteria Met
    Friction TestConfirms units meet product specification related to friction.Implied as meeting product specifications related to friction.Acceptance Criteria Met
    Fluoroscopy TestConfirms the markerband is fluoroscopically visible.Implied: Markerband is fluoroscopically visible.Acceptance Criteria Met
    Simulated Use TestConfirms the functionality of units using clinically relevant benchtop model.Implied: Functionality of units confirmed in a clinically relevant benchtop model.Acceptance Criteria Met
    Particulate TestParticulates generated during simulated use (including multiple deployment cycling) were evaluated.Implied: Particulates generated during simulated use are within acceptable limits or deemed safe.Acceptance Criteria Met
    Hub/Air TestConfirms units have no leaks when tested.Implied: No leaks when tested.Acceptance Criteria Met
    Tensile TestConfirms units meet product specification related to tensile strength.Implied as meeting product specifications related to tensile strength.Acceptance Criteria Met
    Pressure TestConfirms units meet product specification related to pressure.Implied as meeting product specifications related to pressure.Acceptance Criteria Met
    Elongation TestConfirms units meet product specification related to elongation.Implied as meeting product specifications related to elongation.Acceptance Criteria Met
    Corrosion Resistance TestConfirms there is no visible corrosion on the units when tested.No visible corrosion on the units when tested.Acceptance Criteria Met
    Torque Strength TestConfirms units have sufficient torque strength.Implied: Sufficient torque strength.Acceptance Criteria Met
    Burst Pressure TestConfirms units can withstand sufficient pressure.Implied: Can withstand sufficient pressure.Acceptance Criteria Met
    Shelf-LifeConfirms expiration date based on accelerated aging test studies.Implied: Expiration date confirmed based on accelerated aging test studies.Acceptance Criteria Met
    Sterilization TestConfirms the units are sterilized in accordance with ISO 11135 and ISO 10993-7.Units are sterilized in accordance with ISO 11135 and ISO 10993-7.Acceptance Criteria Met
    Biocompatibility Tests
    Cytotoxicity (MEM Elution)Sample extracts must have a cytotoxic reactivity score of grade 2 or lower.Pass (Non-cytotoxic)
    Sensitization (Magnusson-Kligman Method)Test Group shall yield Grade < 1 score on Magnusson and Kligman scale (provided Control Group yields Grade < 1).Pass (Non-sensitizing)
    Irritation (Intracutaneous Toxicity)The difference between the average scores for the extract of the test article and the control is ≤ 1.0.Pass (Non-irritating)
    Systemic Toxicity (Acute Systemic Injection)Sample extracts must not cause significant biological reaction greater than control: • Death in 2 or more animals • Signs of toxicity in 2 or more animals (i.e. convulsions, prostration) • Weight loss > 10% in 3 or more animals.Pass (Non-toxic)
    Systemic Toxicity (Material Mediated Pyrogen)Sample extracts must not cause a total rise in body temperature of ≥ 0.5°C.Pass (Non-pyrogenic)
    Hemocompatibility (Thrombogenicity)The test article must have similar or less thrombus formation compared to predicate device.Pass (Non-thrombogenic)
    Hemocompatibility (Partial Thromboplastin Time - PTT)Clotting times of test article must be similar to predicate values.Pass (Hemocompatible)
    Hemocompatibility (Complement Activation)The concentration of SC5b-9 of test article must be similar to predicate values.Pass (Hemocompatible)
    Hemocompatibility (Hemolysis, indirect contact)Sample extracts must be non-hemolytic (≤ 2% hemolytic index).Pass (Non-hemolytic)
    Hemocompatibility (Hemolysis, direct contact)Sample must be non-hemolytic (≤ 2% hemolytic index).Pass (Non-hemolytic)

    2. Sample Size Used for the Test Set and Data Provenance

    • Design Verification Testing: The document does not specify a quantitative sample size for each individual design verification test. It states "the units" were tested.
    • Biocompatibility Testing: The sample sizes for animal tests (e.g., systemic toxicity, irritation, sensitization) are not explicitly stated in this document but are implied to be sufficient for GLP compliance.
    • Data Provenance: Not explicitly stated for each test, but the tests were performed by Penumbra, Inc. as part of their 510(k) submission. For biocompatibility, it states "previous biocompatibility testing on the predicate and reference devices applies to the subject device" and "All studies were conducted pursuant to 21 CFR, Part 58, Good Laboratory Practices (GLP)." This suggests a mix of historical data from predicate/reference devices and possibly new testing. The country of origin is not mentioned, but is likely the USA, given the FDA submission. All data appears to be retrospective in the sense that the studies were already completed for the submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    There is no mention of "experts" being used to establish ground truth in the context of the design verification or biocompatibility testing described. These tests rely on predefined engineering specifications, standardized test methods (e.g., ISO standards), and laboratory procedures, rather than expert interpretation of a test set.

    4. Adjudication Method for the Test Set

    Not applicable. The tests performed are objective, measurable physical and biological tests against established criteria, not subjective interpretations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, What was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance?

    Not applicable. This device is a medical catheter and does not involve AI or human readers for diagnostic interpretation. The studies are for physical and biological performance of the device itself.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done?

    Not applicable, as this is a physical medical device, not an AI algorithm.

    7. The Type of Ground Truth Used

    • Design Verification Tests: The "ground truth" for these tests is based on engineering specifications and performance requirements defined for the device. For example, a catheter must not kink at a certain radius, or it must withstand a certain pressure.
    • Biocompatibility Tests: The "ground truth" is based on established biological safety standards and criteria (e.g., ISO 10993 guidelines, USP requirements). These standards define what constitutes non-cytotoxic, non-sensitizing, non-irritating, etc.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device that requires a "training set."

    9. How the Ground Truth for the Training Set was Established

    Not applicable. There is no training set mentioned or implied.