Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K213961

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-07-14

    (206 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TruAbutment DS is a patient-specific CAD/CAM abutment, directly connected to endosseous dental implants and is intended for use as an aid in prosthetic rehabilitation. It is compatible with the following systems:
    • Straumann Tissue Level Implant (K122855, K202942): 4.1(RN), 4.8(RN), 6.5(WN) mm
    All digitally designed abutments and/or coping for use with the TruAbutments are intended to be sent to a TruAbutment-validated milling center for manufacture.

    TruBase is a titanium component that is directly connected to endosseous dental implants to patientspecific prosthetic restorations, such as copings or crowns. It is indicated for a screw-retained single tooth or cementretained single tooth and bridge restorations. It is compatible with the following systems:

    • · Osstem TSIII SA (K121995) 3.5 (3.7) , 4.0 (4.2) , 4.5 (4.6) , 5.0 (5.1) , 6.0 (6.0) , 7.0 (6.8) mm (Mini, Regular)
    • · Astra OsseoSpeed EV (K120414) 3.6, 4.2, 4.8, 5.4 mm
    • · BioHorizon Tapered Internal(K093321, K143022, K071638) 3.0. 3.4, 3.8 mm
    • · Straumann Tissue Level Implant (K122855, K202942): 4.1(RN), 4.8(RN), 6.5(WN) mm
      All digitally designed zirconia superstructure for use with the TruBase are intended to be sent to a TruAbutment-validated milling center for manufacture.
    Device Description

    TruAbutment DS system includes patient-specific abutments which are placed into the dental implant to provide support for the prosthetic restoration. The subject abutments are indicated for cemented or screw-retained restorations. The patient-specific abutment screw are made of Titanium grade Ti-6A1-4V ELI (meets ASTM Standard F-136). Each patient-specific abutment is supplied with two identical screws which are used for:
    (1) For fixing into the endosseous implant
    (2) For dental laboratory use during construction of related restoration.
    The abutment is placed over the implant shoulder and mounted into the implant with the provided screw. The design and manufacturing of the patient-specific abutments take into consideration the shape of the final prosthesis based on the patient's intra-oral indications using CAD/CAM system during the manufacturing. All manufacturing processes of TruAbutment DS are conducted at the TruAbutment milling center and provided to the authorized end-user as a final patient-specific abutment.

    TruBase consists of a two-piece abutment, where the titanium base is a pre-manufactured abutment that will be used to support a CAD/CAM designed superstructure (the second part of the two-piece abutment) that composes the final abutment. The system also includes a TruBase Screw for fixation to the implant body.
    TruBase abutments are made of titanium alloy conforming to ASTM F136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications and are provided in a various prosthetic platform diameters (OSSTEM TSIII SA 3.5 (3.7), 4.0 (4.2), 4.5 (4.6), 5.0 (5.1), 6.0 (6.0), 7.0 (6.8) (Mini, Regular) and Astra EV 3.6, 4.2, 4.8, 5.4mm and BioHorizons Internal 3.0. 3.5. 4.5. 5.7mm and Straumann Tissue Level: 4.1(RN), 4.8(RN), 6.5(WN). The TruBase Screws are composed of titanium alloy per ASTM F136.
    CAD/CAM customized superstructure that composes the final abutment intended to be sent to a TruAbutment-validated milling center to be designed and milled, according to the prosthetic planning and patient clinical situation. The superstructure is cemented to the TruBase in the lab. Use "RelyX Unicem 2Automix" as an adhesive extra orally to connect. TruBase is provided non-sterile therefore must be sterilized after the cementation of the customized superstructure on the TruBase.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for two dental devices, TruAbutment DS and TruBase. The submission asserts substantial equivalence to a predicate device based on material, intended use, and mechanical testing. The document focuses on regulatory compliance and mechanical performance, rather than clinical efficacy measured by expert assessments or comparative effectiveness studies with human readers.

    Here's an analysis of the provided information concerning acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are primarily based on established international standards for dental implants and abutments and ensuring mechanical integrity. The reported performance indicates that the devices met these criteria.

    Acceptance Criteria CategorySpecific Criteria/StandardReported Device Performance
    Mechanical PerformanceFatigue Test per ISO 14801:2016Met the criteria of the standard. Fatigue limit data for all implant lines demonstrated construct strengths suitable for intended use.
    SterilizationEnd User Steam Sterilization Test per ISO 17665-1:2006, 17665-2:2009, and ANSI/AAMI ST79:2010Met the criteria of the standard (leveraged from predicate device). The performed validation method is the same as the predicate devices.
    BiocompatibilityBiocompatibility tests per ISO 10993-1:2009, ISO 10993-5:2009, and ISO 10993-10:2010Met the criteria of the standard (leveraged from predicate device).
    Material CompositionTi-6A1-4V ELI (meets ASTM Standard F-136) for abutment and screwsDevices are made of Ti-6Al-4V ELI, conforming to ASTM F136.
    Dimensional CompatibilityAssessment of maximum and minimum dimensions of critical design aspects, tolerances, and cross-sectional images for implant-to-abutment connection.Testing demonstrated implant to abutment compatibility and established substantial equivalency.
    MRI SafetyWorst-case MRI review using scientific rationale and published literature (e.g., Woods et al. 2019) addressing magnetically induced displacement force and torque based on FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment."Rationale addressed parameters and supports safety in the MRI environment (no specific performance metric, but rather a successful review).

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document does not explicitly state a sample size for the mechanical fatigue testing beyond "worst-case constructs." For other tests (sterilization, biocompatibility, MRI review), specific sample sizes are not provided, though these are typically laboratory-based tests rather than patient-data-based studies.
    • Data Provenance: The document does not specify the country of origin for any data or whether the data is retrospective or prospective. The studies mentioned (e.g., fatigue testing, sterilization) are engineering and laboratory tests, not clinical studies involving patient data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable to the provided document. The studies described are non-clinical (mechanical, sterilization, biocompatibility, MRI safety) and rely on testing against established engineering and safety standards, rather than expert interpretation of a 'ground truth' in a clinical context.

    4. Adjudication Method for the Test Set

    This information is not applicable as the described studies are non-clinical laboratory tests and material/design assessments, not studies involving human judgment or adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    This information is not applicable. The submission relates to dental abutments, which are physical medical devices, and does not involve AI or image-based diagnostics requiring human reader performance studies.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This information is not applicable. The device is not an algorithm or AI system.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    The "ground truth" for the non-clinical tests is based on:

    • Established international and national standards (e.g., ISO 14801, ISO 17665, ISO 10993, ASTM F-136, ANSI/AAMI ST79).
    • Engineering specifications and design limits (e.g., minimum thickness, maximum angle for abutments).
    • Physical measurements and compatibility assessments (e.g., dimensional analysis and reverse engineering of implant-to-abutment connections).
    • Scientific rationale and published literature for MRI safety.

    8. The Sample Size for the Training Set

    This information is not applicable. The devices are physical dental abutments, not AI models that require training data.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no training set for an AI model.

    In summary, the provided document focuses on demonstrating the substantial equivalence of physical medical devices through adherence to established engineering and safety standards, rather than clinical performance based on human reader assessments or AI model evaluations.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203649

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-09-15

    (275 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TruAbutment DS is a patient-specific CAD/CAM abutment, directly connected to endosseous dental implants and is intended for use as an aid in prosthetic rehabilitation. It is compatible with the following systems:

    • · MIS C1 Conical Connection Implant (K172505, K112162)
    • : 3.3 (NP) 3.75, 4.2, 5.0 (SP, WP)
    • · Neodent Implant System GM Helix (K163194, K180536)
    • : 3.5, 3.75, 4.0, 4.3, 5.0 (3.0) 6.0 (3.0)
    • · Nobel Biocare Groovy Implants (K050258)
    • : 3.5. 4.3, 5.0, 6.0 (NP, RP, WP, 6.0)
    • · Straumann BLX Implant (K173961, K181703, K191256)
    • : 3.5, 3.75, 4.0, 4.5, 5.0, 5.5, 6.5 (RB, WB)
    • · Straumann Tissue Level Implant (K111357)
    • : 3.3(NNC)

    All digitally designed abutments and/or coping for use with the TruAbutments are intended to be sent to a TruAbutment-validated milling center for manufacture.

    TruBase is a titanium component that is directly connected to endosseous dental implants to provide support for patient-specific prosthetic restorations, such as copings or crowns. It is indicated for screw-retained single tooth or cement-retained single tooth and bridge restorations. It is compatible with the following systems:

    • · MIS C1 Conical Connection Implant (K172505, K112162)
    • : 3.3 (NP) 3.75, 4.2, 5.0 (SP, WP)
    • · Neodent Implant System GM Helix (K163194, K180536)
    • : 3.5, 3.75, 4.0, 4.3, 5.0 (3.0) 6.0 (3.0)
    • · Nobel Biocare Groovy Implants (K050258)
    • : 3.5. 4.3, 5.0, 6.0 (NP, RP, WP, 6.0)
    • · Straumann BLX Implant (K173961, K181703, K191256)
    • : 3.5, 3.75, 4.0, 4.5, 5.0, 5.5, 6.5 (RB, WB)
    • · Straumann Tissue Level Implant (K111357)
    • : 3.3(NNC)

    All digitally designed zirconia superstructure for use with the TruBase are intended to be sent to a TruAbutment-validated milling center for manufacture.

    Device Description

    TruAbutment DS system includes patient-specific abutments which are placed into the dental implant to provide support for the prosthetic restoration. The subject abutments are indicated for cemented or screw-retained restorations. The patient-specific abutment and abutment screw are made of Titanium grade Ti-6A1-4V ELI (meets ASTM Standard F-136). Each patientspecific abutment is supplied with two identical screws which are used for:

    (1) For fixing into the endosseous implant

    (2) For dental laboratory use during construction of related restoration.

    The abutment is placed over the implant shoulder and mounted into the implant with the provided screw. The design and manufacturing of the patient-specific abutments take into consideration the shape of the final prosthesis based on the patient's intra-oral indications using CAD/CAM system during the manufacturing. All manufacturing processes of TruAbutment DS are conducted at the TruAbutment milling center and provided to the authorized end-user as a final patient-specific abutment.

    TruBase consists of a two-piece abutment, where the titanium base is a pre-manufactured abutment that will be used to support a CAD/CAM designed superstructure (the second part of the two-piece abutment) that composes the final abutment. The system also includes a TruBase screw for fixation to the implant body.

    TruBase abutments are made of titanium alloy conforming to ASTM F136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications and are provided in various prosthetic platform diameters. The TruBase screws are composed of titanium alloy per ASTM F136.

    CAD/CAM customized superstructure that composes the final abutment is intended to be sent to a TruAbutment-validated milling center to be designed and milled, according to the prosthetic planning and patient clinical situation. The superstructure is cemented to the TruBase in the lab.

    AI/ML Overview

    The provided text is a 510(k) summary for the TruAbutment DS and TruBase devices, which are endosseous dental implant abutments. It primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting detailed "acceptance criteria" and results from a study proving the device meets those specific criteria in the way one might find for a novel AI/software medical device.

    The document discusses performance in terms of mechanical resistance and dimensional compatibility to ensure long-term functional performance with dental implants. However, it does not outline distinct "acceptance criteria" that are then verified by a specific study with a defined sample size, ground truth, or expert adjudication as typically seen in AI/ML validation. Instead, it relies on demonstrating compliance with recognized standards and comparability to a predicate device.

    Given the information provided, here's an attempt to answer the questions based on the context of this 510(k) summary, interpreting "acceptance criteria" as the performance expectations set by the applicable standards and "study" as the non-clinical testing performed:

    1. A table of acceptance criteria and the reported device performance

    The document doesn't explicitly state "acceptance criteria" in a tabular format with corresponding reported performance for a comparative AI study. Instead, it refers to compliance with established standards for dental implant abutments. The closest equivalent to "acceptance criteria" for the mechanical performance is meeting the requirements of ISO 14801.

    Acceptance Criteria (Derived from Standards/Comparisons)Reported Device Performance
    Mechanical Performance: Fatigue limit based on ISO 14801:2016 for worst-case constructs of compatible OEM implant lines."The fatigue limit data for all other implant lines demonstrated the construct strengths to be sufficient for their intended use." (Implies compliance with ISO 14801, i.e., meeting or exceeding the minimum required fatigue strength as per the standard).
    Dimensional Compatibility: Assessment of maximum and minimum dimensions of critical design aspects, tolerances, and cross-sectional images of the submission device and compatible OEM implant body, OEM abutment, and OEM fixation screw."The testing demonstrated implant to abutment compatibility"
    Sterilization: Compliance with ISO 17665-1:2006, 17665-2:2009 and ANSI/AAMI ST79:2010."Sterilization validation for the subject devices (TruAbutment DS, TruBase) was performed according to ISO 17665-1 and ISO 17665-2. This sterilization method is the same as the primary predicate devices (K202579)."
    Biocompatibility: Compliance with ISO 10993-1:2009, ISO 10993-5:2009, and ISO 10993-10:2010."Biocompatibility tests according to ISO 10993-1:2009, ISO 10993-5:2009, and ISO 10993-10:2010. The results of the above tests have met the criteria of the standard..."

    2. Sample size used for the test set and the data provenance

    The document does not specify a "test set" sample size in the context of an AI/ML model for image analysis or diagnostics. The testing described is primarily non-clinical mechanical testing, dimensional analysis, and biocompatibility testing of physical devices.

    • Sample Size: Not specified in terms of "number of cases" or "patients" for a diagnostic study. For mechanical testing, samples would be physical devices/constructs. The number of samples for ISO 14801 fatigue testing is typically defined by the standard itself (e.g., typically 3 samples for static and 15 samples for fatigue for each configuration).
    • Data Provenance: Not applicable in the sense of patient data or images. The "data" comes from physical testing of manufactured devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable to the type of device and testing described. The "ground truth" for mechanical properties is established by physical measurement against engineering specifications and performance under defined load conditions, not by expert human interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. There is no human interpretation or subjective assessment involved that would require an adjudication method.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The device is a physical dental implant abutment, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The device is a physical dental implant abutment, not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this medical device is based on engineering specifications, material science standards (e.g., ASTM F136), and mechanical testing standards (e.g., ISO 14801). For example, mechanical strength and fatigue life are measured directly, and dimensional compatibility is assessed against design tolerances.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device that requires a training set. The "design" of the abutments is based on CAD/CAM systems informed by engineering principles, not machine learning from large datasets.

    9. How the ground truth for the training set was established

    Not applicable. As this is not an AI/ML device, there is no "training set" or "ground truth for the training set" in the context of machine learning. The knowledge base for the CAD/CAM design and manufacturing is derived from established dental and engineering principles, material science, and regulatory standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202579

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-07-08

    (303 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TruAbutment DS is a patient-specific CAD/CAM abutment, directly connected to endosseous dental implants and is intended for use as an aid in prosthetic rehabilitation.
    TruBase is a titanium component that is directly connected to endosseous dental implants to provide support for patient-specific prosthetic restorations, such as copings or crowns. It is indicated for screw-retained single tooth or cement-retained single tooth and bridge restorations.

    Device Description

    TruAbutment DS system includes patient-specific abutments which are placed into the dental implant to provide support for prosthetic restoration. The subject abutments are indicated for cemented or screw-retained restorations. The patient-specific abutment screw are made of Titanium grade Ti-6A1-4V ELI (meets ASTM Standard F-136). Each patient-specific abutment is supplied with two identical screws which are used for: (1) For fixing into the endosseous implant (2) For dental laboratory use during construction of related restoration. The abutment is placed over the implant shoulder and mounted into the implant with the provided screw. The design and manufacturing of the patient-specific abutments take into consideration the shape of the final prosthesis based on the patient's intra-oral indications using CAD/CAM system during the manufacturing. All manufacturing processes of TruAbutment DS are conducted at the TruAbutment milling center and provided to the authorized end-user as a final patient-specific abutment.
    TruBase consists of a two-piece abutment, where the titanium base is a pre-manufactured abutment that will be used to support a CAD/CAM designed superstructure (the second part of the two-piece abutment) that composes the final abutment. TruBase abutments are made of titanium alloy conforming to ASTM F136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications and are provided in various prosthetic platform diameters. CAD/CAM customized superstructure that composes the final abutment must be designed and milled through the 3Shape Abutment Designer Software, according to the prosthetic planning and patient clinical situation. The superstructure is cemented to the TruBase in the lab. Use "RelyX Unicem 2Automix" as an adhesive extra orally to connect. TruBase is provided non-sterile therefore must be sterilized after the cementation of the customized superstructure on the TruBase.

    AI/ML Overview

    The provided text describes the regulatory clearance for the TruAbutment DS and TruBase devices, citing substantial equivalence to predicate devices, and includes details about non-clinical testing. However, it does not contain information about a study proving device performance against specific acceptance criteria for AI or algorithmic performance. The studies mentioned are focused on mechanical and biocompatibility testing of the physical dental implant abutments.

    Therefore, for aspects related to AI/algorithmic performance, sample sizes, expert ground truth, adjudication, and MRMC studies, the information is not present in the provided document.

    Here's a breakdown of the available information:

    1. Acceptance Criteria and Reported Device Performance

    The document describes "Design Limitations" for both TruAbutment DS and TruBase, which can be interpreted as design acceptance criteria. It also references compatibility and mechanical performance tests.

    TruAbutment DS Design Limitations / Reported Performance:

    Design parameterAcceptance Criteria (Design Limit)Reported Device Performance
    Minimum and Maximum abutment angle0~25°- (Meets criteria as per "Conclusion")
    Minimum and Maximum Cuff Height0.5~6.0mm- (Meets criteria as per "Conclusion")
    Minimum and Maximum diameter at the abutment/implant interfaceØ3.8mm~Ø8.0mm- (Meets criteria as per "Conclusion")
    Minimum Thickness0.4 mm- (Meets criteria as per "Conclusion")
    Minimum and Maximum length of abutment post (length above the abutment collar/gingival height)4~7 mm- (Meets criteria as per "Conclusion")

    TruBase / Zirconia Superstructure Design Limitations / Reported Performance:

    Design parameterAcceptance Criteria (Design Limit)Reported Device Performance
    Minimum and Maximum abutment angle0~15°- (Meets criteria as per "Conclusion")
    Minimum and Maximum Cuff Height0.5~5 mm- (Meets criteria as per "Conclusion")
    Minimum and Maximum diameter at the abutment/implant interfaceØ5.0mm~Ø8.0mm- (Meets criteria as per "Conclusion")
    Minimum Thickness0.4 mm- (Meets criteria as per "Conclusion")
    Minimum and Maximum length of abutment post (length above the abutment collar/gingival height)4~6 mm- (Meets criteria as per "Conclusion")

    Mechanical Performance:

    • Acceptance Criteria: Met the criteria of ISO 14801:2016 and FDA Guidance "Class II Special Controls Guidance Document: Rootform Endosseous Dental Implants and Endosseous Dental Implant Abutments."
    • Reported Device Performance: "The fatigue limit data for all other implant lines demonstrated the construct strengths to be sufficient for their intended use."

    Sterilization Validation:

    • Acceptance Criteria: Met the criteria of ISO 17665-1:2006, 17665-2:2009, and ANSI/AAMI ST79:2010.
    • Reported Device Performance: Achieved through end-user moist steam sterilization.

    Biocompatibility:

    • Acceptance Criteria: Met the criteria of ISO 10993-1:2009, ISO 10993-5:2009, and ISO 10993-10:2010.
    • Reported Device Performance: Not explicitly stated as "passed" for the subject device, but mentioned it was leveraged from predicate device testing and supports substantial equivalence.

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size: For mechanical testing (ISO 14801), "worst-case constructs" were subjected to static compression and compression fatigue testing for each compatible OEM implant line. The exact number of samples is not specified.
    • Data Provenance: Not specified; the context implies laboratory testing. The document does not refer to real-world patient data for performance evaluation in the context of AI.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The tests performed are non-clinical (mechanical, sterilization, biocompatibility) of physical devices, not assessments requiring expert interpretation of data or images.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable for non-clinical, physical device testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. No AI component is described in the performance evaluation section for either TruAbutment DS or TruBase. The software mentioned (3Shape Abutment Designer Software) is for CAD/CAM design, which is a tool for designing the physical device, not an AI for diagnosis or interpretation that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. As noted above, the device itself is a physical dental abutment, and its design process involves CAD/CAM software, not a standalone AI algorithm for medical decision-making or image analysis.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Mechanical Testing: Ground truth is established by the specified ISO standards (ISO 14801) and FDA guidance, defining acceptable force and fatigue limits for dental implant systems.
    • Dimensional Analysis: Ground truth is against design specifications and reverse engineering of OEM implant-to-abutment connections.
    • Sterilization: Ground truth is against ISO 17665-1:2006, 17665-2:2009, and ANSI/AAMI ST79:2010 standards for sterilization efficacy.
    • Biocompatibility: Ground truth is against ISO 10993 standards for biological safety.

    8. The sample size for the training set

    Not applicable. The document describes a physical medical device and its manufacturing process, not an AI/ML algorithm that requires a training set. The CAD/CAM software aids in design, but the document does not detail its internal development or any associated training sets.

    9. How the ground truth for the training set was established

    Not applicable, as no training set for an AI/ML algorithm is described.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201197

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-03-26

    (326 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TruAbutment DS is a patient-specific CAD/CAM abutment, directly connected to endosseous dental implants and is intended for use as an aid in prosthetic rehabilitation. It is compatible with the following systems:
    · Zimmer 3.1mmD Dental Implant System (K142082)

    • · Screw Vent® and Tapered Screw Vent® (K013227)
    • · Nobel Active 3.0 (K102436)
    • · Nobel Active Internal Connection Implant (K071370)
      All digitally designed abutments and/or coping for use with the TruAbutments are intended to be sent to a TruAbutment-validated milling center for manufacture.

    TruBase S is a titanium component that is directly connected to endosseous dental implants to provide support for custom prosthetic restorations, such as copings or crowns. It is indicated for screw-retained single tooth or cement-retained single tooth and bridge restorations. It is compatible with the following systems:
    · Zimmer 3.1mmD Dental Implant System (K142082)

    • · Screw Vent® and Tapered Screw Vent® (K013227)
      TruBase S is intended to be sent to a TruAbutment-validated milling center for manufacture.
    Device Description

    TruAbutment DS system includes patient-specific abutments that are placed into the dental implant to provide support for prosthetic restoration. The subject abutments are indicated for cemented or screwretained restorations. The patient-specific abutment screws are made of Titanium grade Ti-6A1-4V ELI (meets ASTM Standard F-136). Each patient-specific abutment is supplied with two identical screws which are used for:

    • (1) For fixing into the endosseous implant
    • (2) For dental laboratory use during construction of related restoration.
      The abutment is placed over the implant shoulder and mounted into the implant with the provided screw. The design and manufacturing of the patient-specific abutments take into consideration the shape of the final prosthesis based on the patient's intra-oral indications using the CAD/CAM system during the manufacturing. All manufacturing processes of TruAbutment DS are conducted at the TruAbutment milling center and provided to the authorized end-user as a final patient-specific abutment.

    TruBase S consists of a two-piece abutment, where the titanium base is a pre-manufactured abutment that will be used to support a CAD/CAM designed superstructure (the second part of the two-piece abutment) that composes the final abutment. TruBase S abutments are made of titanium alloy conforming to ASTM F136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications and are provided in various prosthetic platform diameters (Zimmer 3.1mmD Dental Implant System: 2.9mm and Screw Vent® and Tapered Screw Vent®: 3.5, 4.5, 5.7mm). They also feature:

    • . cylindrical shape
    • . hexagonal indexing at the apical end of the connection
    • . indexing guide in the cementable portion for coping fitting
      CAD/CAM customized superstructure that composes the final abutment is intended to be sent to a TruAbutment-validated milling center to be designed and milled, according to the prosthetic planning and patient clinical situation. The superstructure is cemented to the TruBase S in the lab. Use "RelyX Unicem 2Automix" as an adhesive extra orally to connect.
      TruBase S is provided non-sterile therefore must be sterilized after the cementation of the customized superstructure on the TruBase S.
    AI/ML Overview

    The provided text is a 510(k) summary for a dental device, TruAbutment DS and TruBase S. It primarily focuses on demonstrating substantial equivalence to predicate devices based on design specifications, materials, and non-clinical testing (mechanical and biocompatibility).

    This document does not contain information about studies involving human-in-the-loop performance, expert ground truth establishment, or clinical effectiveness studies in the context of AI/ML device performance. The "acceptance criteria" discussed are related to the mechanical properties and biocompatibility of the physical dental abutment and base, rather than the performance of an AI/ML algorithm.

    Therefore, many of the requested points, such as sample size for test/training sets, experts for ground truth, adjudication methods, MRMC studies, or standalone algorithm performance, cannot be extracted from this document as they are not relevant to the type of device and testing described.

    However, I can provide information based on the mechanical and material testing criteria that are present.

    Here's an analysis of the provided text in relation to your request, with a clear indication of what information is not available:

    Information Pertaining to Device Acceptance (Mechanical/Biocompatibility):

    1. A table of acceptance criteria and the reported device performance:

      The document describes design limitations and states that non-clinical testing (fatigue and biocompatibility) met the criteria of the standard, demonstrating substantial equivalence. It does not provide specific numerical performance results for the acceptance criteria, but rather states whether the device "met the criteria of the standard" (ISO 14801:2016 for fatigue, ISO 17665-1:2006, 17665-2:2009 and ANSI/AAMI ST79:2010 for sterilization, and ISO 10993 series for biocompatibility).

      For TruAbutment DS:

      Design ParameterDesign Limit (Acceptance Criteria)Reported Performance (Qualitative)
      Abutment Angle0~25°Met design limit
      Cuff Height0.5~6.0mmMet design limit
      Diameter at abutment/implant interfaceØ3.5mm~Ø8.0mmMet design limit
      Minimum Thickness0.4 mmMet design limit
      Length of abutment post4~7 mmMet design limit

      For TruBase S (Zirconia Superstructure):

      Design ParameterDesign Limit (Acceptance Criteria)Reported Performance (Qualitative)
      Abutment Angle0~15°Met design limit
      Cuff Height0.5~5 mmMet design limit
      Diameter at abutment/implant interfaceØ5.0mm~Ø8.0mmMet design limit
      Minimum Thickness0.4 mmMet design limit
      Length of abutment post4~6 mmMet design limit

      Non-Clinical Testing:

      Acceptance Criteria (Standard)Reported Performance
      Fatigue Test (ISO 14801:2016)"The fatigue limit data for all other implant lines demonstrated the construct strengths to be sufficient for their intended use."
      End User Steam Sterilization Test (ISO 17665-1:2006, 17665-2:2009, ANSI/AAMI ST79:2010)"The results of the above tests have met the criteria of the standard, and demonstrated the substantial equivalence with the predicate device."
      Biocompatibility (ISO 10993-5:2009, ISO 10993-10:2010)"The results of the above tests have met the criteria of the standard, and demonstrated the substantial equivalence with the predicate device."
    2. Sample sized used for the test set and the data provenance:

      • Sample Size: Not explicitly stated for specific tests (e.g., how many abutments were fatigued). The document mentions "worst-case constructs" were subjected to testing.
      • Data Provenance: Not specified (e.g., country of origin). The studies appear to be non-clinical (laboratory testing) rather than human subject data. The studies are retrospective as they leverage data from previous 510(k) clearances (K152559).
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable. The "ground truth" here is based on engineering standards (e.g., ISO 14801 for mechanical properties, ISO 10993 for biocompatibility) and direct physical measurements, not expert human interpretation of medical images or clinical outcomes.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. This relates to human expert review and consensus, which is not part of the physical and mechanical testing described.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This device is a physical dental implant component, not an AI/ML diagnostic or assistive imaging device.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is not an AI/ML algorithm. It's a medical device that is CAD/CAM manufactured.
    7. The type of ground truth used:

      • For mechanical performance: Compliance with ISO 14801:2016 standard. This involves objective physical measurements and material science.
      • For sterilization: Compliance with ISO 17665-1:2006, 17665-2:2009 and ANSI/AAMI ST79:2010 standards.
      • For biocompatibility: Compliance with ISO 10993-5:2009, ISO 10993-10:2010 standards.
    8. The sample size for the training set:

      • Not applicable. This is not an AI/ML algorithm that requires a "training set" in the computational sense. The "training" for manufacturing is implied through the "TruAbutment-validated milling center."
    9. How the ground truth for the training set was established:

      • Not applicable. As above, no computational "training set" is described. The manufacturing process relies on validated design parameters and material specifications.

    In summary, this 510(k) submission successfully demonstrates substantial equivalence based on engineering design parameters, material properties, and adherence to established mechanical and biological safety standards for a physical dental implant component. It does not involve AI/ML technology or associated human-in-the-loop performance studies.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1