Search Filters

Search Results

Found 83 results

510(k) Data Aggregation

    K Number
    K251949

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2026-02-05

    (225 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K252612

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-10-16

    (58 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    INDIGO Aspiration Catheters and Separators

    As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems, and for the treatment of pulmonary embolism.

    INDIGO Aspiration Tubing

    As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump.

    Penumbra Aspiration Pump

    The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

    Device Description

    The INDIGO® Aspiration System is comprised of several devices:

    • INDIGO Aspiration Catheter
    • Penumbra Aspiration Pump
    • INDIGO Aspiration Pump Canister
    • INDIGO Aspiration Tubing
    • INDIGO Separator

    The INDIGO Aspiration System is designed to remove thrombus from the vasculature using mechanical aspiration. The INDIGO Aspiration Catheter targets aspiration from the pump directly to the thrombus. The INDIGO Separator may be used to clear the lumen of the INDIGO Aspiration Catheter should it become blocked with thrombus. The INDIGO Aspiration Catheter is introduced through a guide catheter or vascular sheath into the peripheral vasculature and guided over a guidewire to the site of the primary occlusion. The INDIGO Aspiration Catheter is used with the Penumbra Aspiration Pump to aspirate thrombus from an occluded vessel. As needed, an INDIGO Separator may be deployed from the INDIGO Aspiration Catheter to assist with thrombus removal. The INDIGO Separator is advanced and retracted through the INDIGO Aspiration Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the INDIGO Aspiration Catheter tip. The INDIGO Aspiration Catheter may be provided with a steam shaping mandrel, rotating hemostasis valve, and introducer. The INDIGO Separator may be provided with an introducer and torque device. The devices are visible under fluoroscopy. For the aspiration source, the INDIGO Aspiration Catheter is used in conjunction with the Penumbra Aspiration Pump, which is connected using the INDIGO Aspiration Tubing and the INDIGO Aspiration Pump Canister.

    Lightning Flash and Lightning Bolt Aspiration Tubing:

    The Lightning Flash and Lightning Bolt Aspiration Tubing (INDIGO Aspiration Tubing) is designed to serve as a conduit to assist in thrombus removal, facilitating the transfer of vacuum between the Penumbra Aspiration Pump and the INDIGO Aspiration Catheter while providing intermittent, continuous, or modulated aspiration. Modulated aspiration is provided when the Lightning Bolt Aspiration Tubing alternates between connecting the INDIGO Aspiration Catheter to the Penumbra Aspiration Pump and a sterile saline intravenous (IV) bag at ambient pressure.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250079

    Validate with FDA (Live)

    Device Name
    Ruby XL System
    Manufacturer
    Date Cleared
    2025-03-14

    (60 days)

    Product Code
    Regulation Number
    870.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ruby XL System is indicated for arterial and venous embolizations in the peripheral vasculature.

    Device Description

    The Ruby XL System is comprised of a platinum embolization coil attached to a composite delivery pusher, and the Penumbra Detachment Handle. The coil/delivery pusher is packaged separately from the Penumbra Detachment Handle.

    AI/ML Overview

    The provided document is a 510(k) Pre-Market Notification for the Ruby XL System, a vascular embolization device. This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, not for proving a device meets specific acceptance criteria based on a clinical or performance study in the way an AI/ML device would.

    Therefore, the requested information regarding "acceptance criteria" and "study that proves the device meets the acceptance criteria" in the context of an AI/ML device (e.g., sample size, data provenance, expert ground truth, MRMC study, standalone performance) is not applicable to this document.

    This submission focuses on bench-top performance testing, biocompatibility, and shelf-life, establishing substantial equivalence based on the device's physical characteristics, materials, and mechanical performance compared to a predicate embolization coil system. No AI/ML components are mentioned or evaluated in this regulatory filing.

    Here's a breakdown of what is provided and why the AI/ML-specific questions are not applicable:

    1. A table of acceptance criteria and the reported device performance

    • This document does not present acceptance criteria in the typical sense of numerical thresholds for AI/ML performance metrics (e.g., sensitivity, specificity, AUC). Instead, it lists various bench-top performance tests (Dimensional/Visual, Packaging, Radiopacity, Friction, Fatigue Resistance, Torsional Resistance, Stiffness, Simulated Use, Corrosion, Tensile, MRI Compatibility) and states that these tests were "used in support of the subject device to determine substantial equivalence."
    • No specific reported numerical performance values are given for these tests beyond the general statement that they substantiate the device's performance.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable as no AI/ML test set was used. The "test set" here would refer to the physical devices undergoing bench-top testing. Data provenance is not relevant for physical device testing in this context.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. Ground truth, in the AI/ML context, refers to expert labeling or an objective standard against which an algorithm's output is compared. For this device, "ground truth" would be the established physical and mechanical properties and material standards. No human experts are described as defining a "ground truth" for a test set of images or data to evaluate an algorithm.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. Adjudication methods are relevant for resolving discrepancies in expert labeling for AI/ML ground truth, which is not part of this submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is a physical medical device (embolization coil system), not an AI/ML diagnostic or assistive tool. MRMC studies are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No. This is a physical medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not applicable for an AI/ML context. The "ground truth" for this device's performance is established engineering and material science standards and physical measurements.

    8. The sample size for the training set

    • Not applicable as no AI/ML algorithm was trained.

    9. How the ground truth for the training set was established

    • Not applicable as no AI/ML algorithm was trained.

    Summary based on the provided document:

    The Ruby XL System is a medical device (vascular embolization device), not a software/AI/ML product. The 510(k) submission demonstrates substantial equivalence to a predicate device (Penumbra Coil System and POD System) based on:

    • Indications for Use: The subject device is indicated for arterial and venous embolizations in the peripheral vasculature, which is a subset of the predicate's indications.
    • Design and Materials: The document states "SAME" for various material and construction attributes, coil dimensions, detachment mechanism, sterilization method, and method of supply, indicating identical or highly similar characteristics to the predicate.
    • Non-Clinical Safety and Performance Data: Bench-top testing (e.g., Dimensional, Packaging, Radiopacity, Friction, Fatigue, Stiffness, Simulated Use, Corrosion, Tensile, MRI Compatibility) and Biocompatibility testing were performed. These tests are the "study" that substantiates the device's performance and safety, leading to the determination of substantial equivalence. The document explicitly states "No animal or clinical study was conducted as bench testing was determined sufficient for verification and validation purposes."

    Therefore, the requested AI/ML specific criteria and study details are not relevant to this 510(k) submission for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242104

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-02-13

    (210 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Penumbra Reperfusion Catheters and Separators

    As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.

    Penumbra 3D Revascularization Device

    As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.

    Penumbra Aspiration Tubing

    As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump.

    Penumbra Aspiration Pump

    The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

    Device Description

    The Penumbra System is comprised of the following devices:

    • Penumbra Reperfusion Catheter
    • . Penumbra 3D Revascularization Device
    • Penumbra Aspiration Pump
    • . Penumbra Aspiration Pump Canister/Tubing
    • Penumbra Aspiration Tubing
    • Penumbra Separator .

    The Penumbra System is designed to remove thrombus and restore blood flow in the neurovasculature using aspiration. The Reperfusion Catheter delivers aspiration from the pump directly to the site of occlusion to remove the clot. The 3D Revascularization Device is used with Reperfusion Catheters to facilitate aspiration and removal of the thrombus when needed. Alternatively, a Separator may be used to clear the lumen of the Reperfusion Catheter should it become blocked with thrombus. The Reperfusion Catheter is introduced through a guide catheter or long femoral sheath and into the intracranial vasculature and guided over a neurovascular guidewire to the site of the primary occlusion. The Penumbra Reperfusion Catheter is used with the Penumbra Aspiration Pump to aspirate thrombus from an occluded vessel. The Penumbra Reperfusion Catheter is connected to the Penumbra Aspiration Pump using the Penumbra Aspiration Tubing and the Penumbra Aspiration Pump Canister. The Penumbra Reperfusion Catheter is provided with a steam shaping mandrel, rotating hemostasis valve (RHV), peelable sheath, and optionally, SENDit Technology (cleared under K191946). The Penumbra 3D Revascularization Device is provided with an introducer sheath. The Penumbra Separator is provided with an introducer and torque device. The devices are visible under fluoroscopy.

    AI/ML Overview

    The provided text describes specific acceptance criteria and performance data for the Penumbra System (Reperfusion Catheter RED 72). However, it notes that no animal or clinical studies were conducted for this particular device submission (K242104). The evidence for meeting acceptance criteria comes solely from bench and biocompatibility testing. Therefore, information regarding sample sizes for test sets, data provenance, number and qualifications of experts, adjudication methods, MRMC studies, standalone performance, and training set details, which are typically associated with clinical or AI/algorithm performance studies, are not applicable or available in this document.

    Here's a summary of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Dimensional/Visual TestUnits meet all dimensional and visual product specifications.Acceptance Criteria Met
    Friction TestUnits meet product specification related to friction.Acceptance Criteria Met
    Radiopacity TestMarkerband is fluoroscopically visible.Acceptance Criteria Met
    Simulated Use TestFunctionality of units using clinically relevant benchtop model.Acceptance Criteria Met
    Particulate and Coating Integrity TestParticulates generated during simulated use and coating integrity after simulated use (including multiple deployment cycles) were evaluated.Acceptance Criteria Met
    Hub/Air TestUnits have no leaks when tested.Acceptance Criteria Met
    Tensile TestUnits meet product specification related to tensile strength.Acceptance Criteria Met
    Pressure TestUnits meet product specification related to pressure.Acceptance Criteria Met
    Elongation TestUnits meet product specification related to elongation.Acceptance Criteria Met
    Corrosion Resistance TestNo visible corrosion on the units when tested.Acceptance Criteria Met
    Torque Strength TestUnits have sufficient torque strength.Acceptance Criteria Met
    Burst Pressure TestUnits can withstand sufficient pressure.Acceptance Criteria Met
    Distal Tip Stiffness TestUnits have distal tip stiffness comparable to the reference device.Acceptance Criteria Met
    Kink Resistance TestUnits meet product specification related to kink resistance.Acceptance Criteria Met
    Simulated Use Physician Usability TestUsability of units in clinically relevant benchtop model.Acceptance Criteria Met
    Shelf LifeExpiration date based on accelerated aging test studies.Acceptance Criteria Met
    Packaging ValidationPackaging of the units meets all product specifications.Acceptance Criteria Met
    Sterilization TestUnits are sterilized in accordance with ISO 11135+A1 and ISO 10993-7.Acceptance Criteria Met
    Biocompatibility Tests
    Cytotoxicity: MEM ElutionSample extracts must have a cytotoxic reactivity score of Grade 2 or lower.Pass (Non-cytotoxic)
    Sensitization: Magnusson-Kligman MethodTest group shall yield Grade < 1 score on Magnusson and Kligman scale (provided control group yields Grade < 1).Pass (Non-sensitizing)
    Irritation: Intracutaneous ReactivityThe difference between the average scores for the extract of the test article and the control is ≤ 1.0.Pass (Non-irritating)
    Systemic Toxicity: Acute Systemic InjectionSample extracts must not cause significant biological reaction greater than control. That is: Death in 2 or more animals. Signs of toxicity in 2 or more animals (i.e. convulsions, prostration). Weight loss > 10% in 3 or more animals.Pass (Non-toxic)
    Systemic Toxicity: Material-Mediated PyrogenSample extracts must not cause a total rise in body temperature of ≥ 0.5 °C.Pass (Non-pyrogenic)
    Hemocompatibility: In-vitro ThrombogenicityDevice must not be thrombogenic in vitro when compared to a predicate device.Pass (Comparable to the predicate)
    Hemocompatibility: Partial Thromboplastin Time (PTT)Clotting times of test article must be similar to predicate values.Pass (Comparable to the predicate)
    Hemocompatibility: Complement ActivationThe concentration of SC5b-9 of test article must be similar to predicate values.Pass (Non-activator of the complement system)
    Hemocompatibility: Hemolysis (indirect contact)Sample extracts must be non-hemolytic (≤ 2% hemolytic index).Pass (Non-hemolytic)
    Hemocompatibility: Hemolysis (direct contact)Sample must be non-hemolytic (≤ 2% hemolytic index).Pass (Non-hemolytic)
    Hemocompatibility: Platelet and Leukocyte CountSample must meet the requirements of ASTM F2888-19, Standard Practice for Platelet Leukocyte Count.Pass (Similar to the predicate)

    2. Sample size used for the test set and the data provenance: Not applicable. The document states that "No animal or clinical study was conducted as bench and biocompatibility testing were determined sufficient for verification and validation purposes." The tests listed are primarily bench (in-vitro) tests on manufactured units or material samples. The specific number of units or samples tested for each bench test is not provided in this summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. The "ground truth" for these engineering and biocompatibility tests is based on established scientific principles, industry standards, and specified material properties, not expert consensus on medical images or diagnoses.

    4. Adjudication method for the test set: Not applicable. Adjudication methods are typically employed in clinical studies or studies involving human assessment of data, which were not performed in this case.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No MRMC study was performed as this device is a physical medical device (catheter system) and not an AI-powered diagnostic or assistive tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used:
    * Bench Testing: Engineering specifications, material properties, and functionality performance standards.
    * Biocompatibility Testing: Established biological safety standards (ISO 10993 series, USP standards, 21 CFR Part 58 GLP).

    8. The sample size for the training set: Not applicable. This is a physical medical device, not an algorithm that requires a training set.

    9. How the ground truth for the training set was established: Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242033

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-12-11

    (153 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Access25 Delivery Microcatheter is indicated to assist in the delivery of diagnostic agents, such as contrast media, and therapeutic devices, such as occlusion coils, to the peripheral and neuro vasculature.

    Device Description

    The Access25 Delivery Microcatheter (Access25) is a single lumen medical device designed to aid a physician in accessing distal vasculature when used in conjunction with a guide catheter and micro guidewire. The Access25 Delivery Microcatheter is supplied with annealed stainless steel mandrels that can be used to shape the distal tip as desired.

    AI/ML Overview

    This document pertains to the Access25™ Delivery Microcatheter by Penumbra, Inc. This is a medical device and not an AI/ML powered device, therefore some relevant sections of the query will not be applicable.

    Here's an analysis of the acceptance criteria and the study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance CriteriaReported Device Performance
    Dimensional/Visual Inspection TestCatheter outer diameter, inner diameter, length, and coating length meet specifications.Acceptance Criteria met
    Simulated Use TestDevice performs as intended under simulated use conditions in a clinically relevant anatomical model.The device performs as intended under simulated use conditions
    Tensile TestPeak tensile force meets ISO 10555-1 requirements after preconditioning in a simulated use model.Acceptance Criteria met
    Particulate TestParticulate generation during simulated use is comparable to the predicate device.Particulate generation was comparable to predicate device
    Burst Pressure TestUnits can withstand sufficient pressure per ISO 10555-1 after simulated use.Acceptance Criteria met
    Radiopacity TestMarkerband(s) are fluoroscopically visible.Acceptance Criteria met
    Flow Rate TestContrast media can be delivered through the catheter lumen.Acceptance Criteria met
    Friction TestUnits meet product specification related to friction.Friction results were comparable to the predicate
    Corrosion Resistance TestNo visible corrosion on units when tested per ISO 10555-1.Acceptance Criteria met
    Torque Strength TestDevice can withstand a specified number of catheter rotations until failure.Acceptance Criteria met
    Elongation TestUnits meet product specification related to elongation.Acceptance Criteria met
    Hub/Air TestUnits have no leaks when tested per ISO 10555-1.Acceptance Criteria met
    Distal Tip Stiffness TestMaximum compressive force to cause catheter tip buckling is comparable to the predicate.Distal tip stiffness was comparable to the predicate.
    Kink Resistance TestUnits have appropriate kink resistance.Acceptance Criteria met
    Liquid Leakage TestUnits can withstand sufficient pressure.Acceptance Criteria met
    Biocompatibility: CytotoxicityNon-cytotoxic (in accordance with ISO 10993-5).Pass: Non-cytotoxic
    Biocompatibility: SensitizationNon-sensitizing (in accordance with ISO 10993-10).Pass: Non-sensitizing
    Biocompatibility: IrritationNon-irritating (in accordance with ISO 10993-23).Pass: Non-irritating
    Biocompatibility: Systemic ToxicityNon-toxic (in accordance with ISO 10993-11).Pass: Non-toxic
    Biocompatibility: Material-Mediated PyrogenNon-pyrogenic (in accordance with USP <151>).Pass: Non-pyrogenic
    Biocompatibility: Hemolysis (direct contact)Non-hemolytic (in accordance with ISO 10993-4).Pass: Non-hemolytic
    Biocompatibility: Hemolysis (indirect contact)Non-hemolytic (in accordance with ISO 10993-4).Pass: Non-hemolytic
    Biocompatibility: In-vitro ThrombogenicityNon-thrombogenic (in accordance with ISO 10993-4).Pass: Non-thrombogenic
    Biocompatibility: Complement ActivationNon-activator of complement system (in accordance with ISO 10993-4).Pass: Non-activator of complement system
    Biocompatibility: Partial Thromboplastin Time (PTT)Hemocompatible (in accordance with ISO 10993-4).Pass: Hemocompatible
    SterilizationSterilized using a validated EO sterilization process in accordance with ISO 11135-1.Validated EO sterilization process per ISO 11135-1.
    PyrogenicityNon-pyrogenic, meeting < 2.15 EU/device acceptance criteria per USP <151> and ANSI/AAMI ST72.Non-pyrogenic based on material mediated rabbit pyrogen biocompatibility testing per USP <151> and LAL testing per ANSI/AAMI ST72 to meet < 2.15 EU/device acceptance criteria.

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state the exact sample sizes (number of devices) used for each individual test. However, it indicates that the tests were conducted on "units" multiple times ("The device was evaluated for torque strength by measuring the number of catheter rotations until failure").

    The provenance of the data is from bench performance testing and biocompatibility testing. This typically implies laboratory settings rather than patient data, therefore country of origin and retrospective/prospective distinctions are not directly applicable in the same way they would be for clinical trials.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This is a medical device, not an AI/ML algorithm requiring expert ground truth in the context of image analysis or diagnostic interpretation. The "ground truth" for this device's performance is established through adherence to recognized international standards (e.g., ISO, USP, ANSI/AAMI) and engineering specifications, which are developed and validated by experts in their respective fields (e.g., materials science, medical device design, sterilization, biocompatibility).

    4. Adjudication method for the test set:

    Not applicable in the context of this device. The testing described is objective performance measurement against pre-defined engineering and biological acceptance criteria, rather than subjective interpretation requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is a physical medical device (microcatheter), not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies and the concept of human reader improvement with AI assistance are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Not applicable. This is a physical medical device.

    7. The type of ground truth used:

    The "ground truth" for this device's performance is based on established industry standards and regulatory requirements (e.g., ISO 10555-1, ISO 11135-1, ISO 10993 series, USP <151>, ANSI/AAMI ST72) and internal product specifications. These standards and specifications define what constitutes safe and effective performance for such a device.

    8. The sample size for the training set:

    Not applicable. This is a physical medical device, not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable. As this is not an AI/ML algorithm, there is no training set or its associated ground truth establishment process. The "ground truth" in the device development context refers to the established performance requirements and standards against which the device is tested.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242520

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-11-20

    (89 days)

    Product Code
    Regulation Number
    870.1340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Element Vascular Access System is indicated for the introduction of therapeutic or diagnostic devices into the vasculature.

    Device Description

    The Element Vascular Access System is a medical device system designed to provide a conduit for introduction of therapeutic devices to the peripheral, coronary, and neurovascular anatomy. This device is a three-component system comprised of the BMX81 Delivery Catheter, Penumbra (Neuron 5F) Select Catheter, and a Dilator. The BMX81 Delivery Catheter can be used individually with a 0.038 in [0.97 mm] guidewire or together with the Neuron 5F Select Catheter to access the desired anatomy. The sheath component of the device contains an inner diameter of 6 French (Fr) and lengths ranging from 55 - 115 centimeters (cm).

    AI/ML Overview

    The provided text is a 510(k) summary for the Penumbra Element Vascular Access System. It describes the device, its intended use, and comparison to predicate devices, along with a summary of performance data. However, this document does not contain information about studies involving artificial intelligence (AI) or machine learning (ML) components.

    The "Performance Data" section (1.11 and 1.11.1) outlines bench performance testing, which are standard engineering tests for medical devices (e.g., dimensional, tensile, particulate, leakage, fluoroscopy, friction, corrosion, elongation, stiffness, flexibility). There is no mention of an algorithm, AI, or ML being part of the device or its function.

    Therefore, I cannot provide the requested information regarding:

    • A table of acceptance criteria and reported device performance related to AI/ML.
    • Sample sizes, data provenance, expert adjudication, MRMC studies, standalone performance, ground truth types, or training set details for an AI/ML model, as no such model is described in this submission.

    The "K242520" 510(k) is for a physical medical device (catheter introducer) and not for a software as a medical device (SaMD) or a device with an AI/ML component that requires the types of studies you are asking about (like diagnostic or image analysis AI).

    If you have a different document that details an AI/ML component of this device, please provide that. Based solely on the provided text, the device itself is a conventional vascular access system.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242075

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-09-25

    (71 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    INDIGO Aspiration Catheters and Separators: As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems, and for the treatment of pulmonary embolism.

    INDIGO Aspiration Tubing: As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump.

    Penumbra Aspiration Pump: The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

    Device Description

    The INDIGO® Aspiration System is comprised of several devices:

    • INDIGO Aspiration Catheter
    • Penumbra Aspiration Pump
    • INDIGO Aspiration Pump Canister
    • INDIGO Aspiration Tubing
    • INDIGO Separator

    The INDIGO Aspiration System is designed to remove thrombus from the vasculature using mechanical aspiration. The INDIGO Aspiration Catheter targets aspiration from the pump directly to the thrombus. The INDIGO Separator may be used to clear the lumen of the INDIGO Aspiration Catheter should it become blocked with thrombus. The INDIGO Aspiration Catheter is introduced through a guide catheter or vascular sheath into the peripheral vasculature and guided over a guidewire to the site of the primary occlusion. The INDIGO Aspiration Catheter is used with the Penumbra Aspiration Pump to aspirate thrombus from an occluded vessel. As needed, an INDIGO Separator may be deployed from the INDIGO Aspiration Catheter to assist with thrombus removal. The INDIGO Separator is advanced and retracted through the INDIGO Aspiration Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the INDIGO Aspiration Catheter tip. The INDIGO Aspiration Catheter may be provided with a steam shaping mandrel, rotating hemostasis valve, and introducer. The INDIGO Separator may be provided with an introducer and torque devices are visible under fluoroscopy. For the aspiration source, the INDIGO Aspiration Catheter is used in conjunction with the Penumbra Aspiration Pump, which is connected using the INDIGO Aspiration Tubing and the INDIGO Aspiration Pump Canister.

    The Lightning Bolt Aspiration Tubing (INDIGO Aspiration Tubing) is designed to serve as a conduit to assist in thrombus removal, facilitating transfer of vacuum between the Penumbra Aspiration Pump and the INDIGO Aspiration Catheter while providing intermittent, continuous, or modulated aspiration. Modulated aspiration is provided when the Lightning Bolt Aspiration Tubing alternates between connecting the INDIGO Aspiration Catheter to the Penumbra Aspiration Pump and a sterile saline intravenous (IV) bag at ambient pressure.

    AI/ML Overview

    The provided text is a 510(k) summary for the Penumbra Indigo® Aspiration System - Lightning Bolt Aspiration Tubing (K242075). It describes the device, its intended use, and the non-clinical testing performed to establish substantial equivalence to a predicate device (K223186).

    Here's an analysis of the provided information regarding acceptance criteria and the study that proves the device meets them:

    The document focuses on demonstrating substantial equivalence to a predicate device, rather than proving the subject device meets specific clinical performance acceptance criteria in a standalone study against a predefined clinical benchmark. The studies described are primarily bench-top (non-clinical) tests to verify that the subject device performs similarly to the predicate device and functions as intended.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document lists various non-clinical tests performed, stating that the subject device met all acceptance criteria. However, it does not provide a table detailing specific quantitative acceptance criteria or corresponding device performance values for most tests. Instead, it offers general statements of compliance or similarity to the predicate device.

    Test TypeAcceptance CriteriaReported Device Performance
    Dimensional/Visual Inspection(Not explicitly stated, but implied to match design specifications and be free of defects)Met all acceptance criteria
    Pressurization Testing(Not explicitly stated, but implied to withstand pressure without failure)Met all acceptance criteria
    Indigo Aspiration System Compatibility/Simulated Use Testing(Not explicitly stated, but implied to function correctly with the aspiration system)Met all acceptance criteria, performed similarly to predicate
    Thrombus Removal Testing(Not explicitly stated, but implied to be effective in removing thrombus)Met all acceptance criteria, performed similarly to predicate
    Valve Testing(Not explicitly stated, but implied to ensure proper valve function)Met all acceptance criteria
    Tensile Testing(Not explicitly stated, but implied to meet strength requirements)Met all acceptance criteria
    Post Destructive Testing Dimensional Inspection(Not explicitly stated, but implied to assess structural integrity after destructive tests)Met all acceptance criteria
    BiocompatibilityPreviously provided biocompatibility data of the predicate device (K223186)No changes, maintained previous biocompatibility profile
    Shelf-LifeStability for 12 monthsMet 12 months shelf-life based on accelerated aging data
    Electrical Safety/EMC TestingCompliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, AIM 7351731, ISO 10079-1, and ISO 10079-4Complies with specified standards
    Software Verification/ValidationAligned with FDA's "Content of Premarket Submissions for Device Software Functions" guidanceTesting and documentation provided as recommended

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly provide the sample sizes (number of units or test iterations) for each specific bench-top test. It generally references "testing performed on the subject Lightning Bolt Aspiration Tubing device."

    The data provenance is non-clinical bench testing, conducted by the manufacturer, Penumbra, Inc. The data is prospective in the sense that it was generated specifically for this 510(k) submission to demonstrate equivalence. There is no information about country of origin of the data beyond implicitly being from testing conducted by the US-based manufacturer.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    For the non-clinical bench-top tests described, the concept of "ground truth" as established by medical experts (e.g., radiologists) is not applicable. These tests evaluate engineering and material performance against predefined specifications, often based on industry standards or internal design requirements. The "ground truth" for these tests would be the physical properties and performance of the device as measured in the lab.

    4. Adjudication Method for the Test Set

    Since the tests are non-clinical bench tests evaluating engineering performance, an adjudication method (like 2+1, 3+1 for clinical endpoints) is not relevant. Test results are typically objective measurements against a Pass/Fail criterion.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. The document explicitly states: "No animal or clinical study was conducted as bench testing was determined sufficient for verification and validation purposes." Therefore, there is no information on the effect size of human readers improving with or without AI assistance, as this is not an AI device.

    6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study

    No, a standalone study was not done. This device is physical medical tubing, not an AI algorithm.

    7. Type of Ground Truth Used

    For the bench testing, the "ground truth" is based on:

    • Engineering specifications and design requirements for mechanical, dimensional, and material properties.
    • Industry standards (e.g., ISO for sterility, various IEC standards for electrical safety).
    • Performance of the predicate device as a benchmark for comparison (e.g., "performed similarly to the predicate device").

    8. Sample Size for the Training Set

    Not Applicable. This is a physical medical device (tubing), not a machine learning algorithm. Therefore, there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not Applicable. There is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242319

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-09-03

    (29 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    INDIGO Aspiration Catheters and Separators: As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems, and for the treatment of pulmonary embolism.

    INDIGO Aspiration Tubing: As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump.

    Penumbra Aspiration Pump: The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

    Device Description

    The INDIGO® Aspiration System is comprised of the several devices: INDIGO Aspiration Catheter, Penumbra Aspiration Pump, INDIGO Aspiration Pump Canister, INDIGO Aspiration Tubing, INDIGO Separator. The INDIGO Aspiration System is designed to remove thrombus from the vasculature using mechanical aspiration. The INDIGO Aspiration Catheter targets aspiration from the pump directly to the thrombus. The INDIGO Separator may be used to clear the lumen of the INDIGO Aspiration Catheter should it become blocked with thrombus. The INDIGO Aspiration Catheter is introduced through a guide catheter or vascular sheath into the peripheral vasculature and guided over a guidewire to the site of the primary occlusion. TraX may be used with its prepackaged Aspiration Catheter to facilitate access to the primary occlusion. The INDIGO Aspiration Catheter is used with the Penumbra Aspiration Pump to aspirate thrombus from an occluded vessel. As needed, an INDIGO Separator may be deployed from the INDIGO Aspiration Catheter to assist with thrombus removal. The INDIGO Separator is advanced and retracted through the INDIGO Aspiration Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the INDIGO Aspiration Catheter tip. The devices are visible under fluoroscopy. For the aspiration source, the INDIGO Aspiration Catheter is used in conjunction with the Penumbra Aspiration Pump, which is connected using the INDIGO Aspiration Tubing and the INDIGO Aspiration Pump Canister. The INDIGO Aspiration Catheter may be provided with a steam shaping mandrel, rotating hemostasis valve, Select Catheter, introducer, and TraX. TraX compatibility is limited to its prepackaged Aspiration Catheter. The INDIGO Separator may be provided with an introducer and torque device.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA for the Penumbra, Inc. Indigo® Aspiration System - Aspiration Catheter 6X. It details the device, its intended use, and its substantial equivalence to predicate devices based on non-clinical (bench-top) performance testing.

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a specific table of acceptance criteria with corresponding performance metrics like sensitivity, specificity, or AUC, as it is related to a physical medical device (catheter) rather than a diagnostic algorithm or AI. Instead, it lists various bench-top performance tests conducted. For each test, the implicit acceptance criterion is that the device met all requirements.

    Test CategoryReported Device Performance (Acceptance Criteria)
    Particulate TestingMet all requirements
    Dimensional/Visual InspectionMet all requirements
    Friction TestingMet all requirements
    Performance/Simulated Use TestingMet all requirements
    Markerband VisibilityMet all requirements
    Tensile TestingMet all requirements
    Pressure TestingMet all requirements
    Elongation TestingMet all requirements
    Torsion TestingMet all requirements
    Hub Air AspirationMet all requirements
    Corrosion TestingMet all requirements
    Biocompatibility (ISO 10993-1)Met all requirements
    Shelf-LifeEstablished for 12 months based on accelerated aging data

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document primarily describes bench-top testing for a physical device. It does not mention a "test set" in the context of data used for an algorithm or AI. Therefore, concepts like sample size for a test set, country of origin, retrospective or prospective data are not applicable. The data provenance would be from internal laboratory testing conducted by Penumbra, Inc.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable as the document describes bench-top performance testing of a physical medical device (catheter), not a diagnostic algorithm or AI that would require expert-established ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable for the same reason as point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not done. The study described is for a physical medical device, not an AI or diagnostic tool that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    A standalone performance study in the context of an algorithm or AI was not done. The performance evaluation was of the physical device's mechanical, material, and functional properties on a test bench.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the bench-top tests, the "ground truth" would be defined by engineering specifications, material science standards, and functional performance benchmarks for the device. For example, a tensile test would measure the force required to break the catheter, with the "ground truth" being the specified minimum tensile strength that the device must meet according to its design requirements. Similarly, biocompatibility is assessed against ISO 10993-1 standards.

    8. The sample size for the training set

    This information is not applicable as the document describes bench-top performance testing of a physical medical device, not an AI or algorithm that uses a training set.

    9. How the ground truth for the training set was established

    This information is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241399

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-06-10

    (25 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    INDIGO Aspiration Catheters and Separators: As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems, and for the treatment of pulmonary embolism. INDIGO Aspiration Tubing: As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump. Penumbra Aspiration Pump: The Penumbra Aspiration Pump is indicated as a vacuum source for the Penumbra Aspiration Systems.

    Device Description

    The INDIGO® Aspiration System is comprised of the several devices: INDIGO Aspiration Catheter, Penumbra Aspiration Pump, INDIGO Aspiration Pump Canister, INDIGO Aspiration Tubing, INDIGO Separator. The INDIGO Aspiration System is designed to remove thrombus from the vasculature using mechanical aspiration. The INDIGO Aspiration Catheter targets aspiration from the pump directly to the thrombus. The INDIGO Separator may be used to clear the lumen of the INDIGO Aspiration Catheter should it become blocked with thrombus. The INDIGO Aspiration Catheter is introduced through a guide catheter or vascular sheath into the peripheral vasculature and guided over a guidewire to the site of the primary occlusion. The INDIGO Aspiration Catheter is used with the Penumbra Aspiration Pump to aspirate thrombus from an occluded vessel. As needed, an INDIGO Separator may be deployed from the INDIGO Aspiration Catheter to assist with thrombus removal. The INDIGO Separator is advanced and retracted through the INDIGO Aspiration Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the INDIGO Aspiration Catheter tip. The devices are visible under fluoroscopy. For the aspiration source, the INDIGO Aspiration Catheter is used in conjunction with the Penumbra Aspiration Pump, which is connected using the INDIGO Aspiration Tubing and the INDIGO Aspiration Pump Canister. The INDIGO Aspiration Catheter may be provided with a steam shaping mandrel, rotating hemostasis valve, Select Catheter, and introducer. The INDIGO Separator may be provided with an introducer and torque device. The Select+ Catheter is an accessory within kitted configurations of the INDIGO Lightning Flash Aspiration System. The Select+ Catheter is used to aide in access to the site of occlusion and is removed prior to initiating aspiration with the aspiration catheter. The use of the Select+ Catheter to facilitate access to the site of the occlusion is an optional procedural step for INDIGO Lightning Flash Aspiration Catheters. The device is provided sterile and intended for single use only.

    AI/ML Overview

    The provided text is a 510(k) summary for the Indigo® Lightning Flash Aspiration System - Select+™ Catheter. It does not describe a study involving an AI device or software, nor does it provide the detailed information requested in the prompt regarding acceptance criteria and performance of an AI-powered device.

    Instead, this document focuses on establishing substantial equivalence of a new catheter (Select+™ Catheter) to a previously cleared predicate device (Indigo Aspiration System – Lightning Flash Select Catheter) and a reference device (Neuron Max System - Select Catheter).

    Therefore, I cannot provide the requested information from the given text. The text does not contain:

    1. A table of acceptance criteria and reported device performance for an AI device.
    2. Sample size used for a test set for an AI device or data provenance.
    3. Number and qualifications of experts for ground truth establishment for an AI device.
    4. Adjudication method for an AI device.
    5. Information on a multi-reader multi-case (MRMC) comparative effectiveness study for an AI device.
    6. Standalone performance details for an AI algorithm.
    7. Type of ground truth used for an AI device.
    8. Sample size for the training set of an AI device.
    9. How ground truth for the training set of an AI device was established.

    The document states, under "7.5 Summary of Performance Data – Animal, Clinical", that "No animal or clinical study was conducted as bench testing was determined sufficient for verification and validation purposes." This further confirms that the information requested for an AI-based study is not present.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233201

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-03-27

    (181 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MIDWAY Delivery Catheter is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.

    Device Description

    The MIDWAY Delivery Catheters are single lumen percutaneous catheters designed to provide a conduit for introduction of interventional devices to the peripheral, coronary, and neuro vasculature. The MIDWAY Delivery Catheter is a single lumen, coil-reinforced, variable stiffness catheter with a radiopaque marker band on the distal end and a Luer hub on the proximal end. The MIDWAY Delivery Catheter is compatible with sheaths and guide catheters appropriately sized for the outer diameter of the MIDWAY Delivery Catheter. The MIDWAY Delivery Catheter has a hydrophilic coating on the distal segment of the catheter shaft.

    AI/ML Overview

    The document provided is a 510(k) summary for a medical device (MIDWAY Delivery Catheter) seeking substantial equivalence to existing predicate devices. It does not contain information about a study proving the device meets acceptance criteria for a clinical or AI performance study. Instead, it outlines bench performance testing and other non-clinical data to demonstrate substantial equivalence to legally marketed predicate devices.

    Therefore, many of the requested elements for describing an acceptance criteria study (e.g., sample size for test set, data provenance, number of experts, adjudication method, MRMC study, standalone performance, training set size, ground truth for training set) are not applicable to the information provided.

    However, I can extract the acceptance criteria and confirmed performance for the bench performance tests that were conducted:


    1. Table of Acceptance Criteria and Reported Device Performance

    TestTest Method SummaryReported Device Performance
    Simulated Use TestConfirms the functionality of MIDWAY Delivery Catheters using clinically relevant benchtop model, including testing with a marketed stent-retriever.Acceptance Criteria Met (Functionality confirmed)
    Compatibility TestConfirms compatibility of MIDWAY Delivery Catheters with a marketed introducer sheath, guidewire, microcatheter, and stent-retriever.Acceptance Criteria Met (Compatibility confirmed)
    Particulate TestParticulates generated during simulated use were evaluated.Acceptance Criteria Met (Acceptable particulate levels)
    Coating Integrity TestConfirms that coating of MIDWAY Delivery Catheters has not delaminated, peeled, or flaked before or after simulated use.Acceptance Criteria Met (Coating integrity maintained)
    Surface IntegrityThe MIDWAY Delivery Catheter surface is inspected for defects.Acceptance Criteria Met (Surface free of defects)
    Tensile Strength TestConfirms MIDWAY Delivery Catheters meet product specification related to tensile strength after simulated use.Acceptance Criteria Met (Meets tensile strength specifications)
    Burst Pressure TestConfirms MIDWAY Delivery Catheters can withstand sufficient pressure after simulated use.Acceptance Criteria Met (Withstands sufficient burst pressure)

    The following information is not present in the provided document:

    • 2. Sample size used for the test set and the data provenance: The document refers to "bench performance testing" which are laboratory-based tests of physical properties and functionality, not a clinical "test set" of patient data. The sample sizes for these bench tests are not specified, nor is data provenance (country of origin, retrospective/prospective).
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable as this was bench testing, not a study requiring expert-established ground truth on clinical data.
    • 4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
    • 5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. The device is a physical delivery catheter, not an AI-powered diagnostic tool.
    • 6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For the bench tests, the "ground truth" would be established engineering specifications and industry standards for mechanical performance, material integrity, and functionality.
    • 8. The sample size for the training set: Not applicable, as this is bench testing for a physical device, not an AI/ML model that requires a training set.
    • 9. How the ground truth for the training set was established: Not applicable.

    Overall Study Information:

    The study described is a series of bench performance tests and confirmation of existing data from reference devices (MIDWAY 43 Delivery Catheter and MIDWAY 62 Delivery Catheter were compared to Reperfusion Catheter RED 43 (K222808) and Reperfusion Catheter RED 62 (K203440), respectively). The primary purpose was to demonstrate substantial equivalence to predicate devices for a new indication (introduction of interventional devices into peripheral, coronary, and neuro vasculature).

    Key Findings:

    • All listed bench performance tests (Simulated Use, Compatibility, Particulate, Coating Integrity, Surface Integrity, Tensile Strength, Burst Pressure) met their respective acceptance criteria.
    • The subject devices share the same design, materials, and manufacturing as their reference devices.
    • Other performance data for the reference devices (Biocompatibility, Shelf Life, Sterilization, Packaging) were also used to support the subject device.
    • No animal or clinical studies were deemed necessary for the 510(k) submission.
    • The conclusion is that the MIDWAY Delivery Catheters are substantially equivalent to the predicate devices, perform as intended, and do not raise new questions of safety or effectiveness.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 9