K Number
K203440

Validate with FDA (Live)

Manufacturer
Date Cleared
2021-05-20

(178 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump.

The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

Device Description

The Penumbra System® is comprised of the following devices:

  • Penumbra Reperfusion Catheter .
  • . Penumbra Aspiration Pump
  • Penumbra Aspiration Pump/Canister Tubing .
  • . Penumbra Aspiration Tubing
  • Penumbra Separator .
  • Penumbra 3D Revascularization Device .

The Penumbra System is designed to remove thrombus from the vasculature using continuous aspiration. The Reperfusion Catheter targets aspiration from the pump directly to the thrombus. The 3D Revascularization Device is used with Reperfusion Catheters to facilitate aspiration and removal of the thrombus when needed. The Separator may be used to clear the lumen of the Reperfusion Catheter should it become blocked with thrombus. The use of the Separator may not be necessary when using a Reperfusion Catheter with an ID of 0.054 in. or larger. The Reperfusion Catheter is introduced through a guide catheter or long femoral sheath and into the intracranial vasculature and guided over a neurovascular guidewire to the site of the primary occlusion. The Penumbra Reperfusion Catheter is used with the Aspiration Pump to aspirate thrombus from an occluded vessel. As needed, a Penumbra Separator may be deployed from the Reperfusion Catheter to assist with thrombus removal. The Penumbra Separator is advanced and retracted through the Penumbra Reperfusion Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the Reperfusion Catheter tip. For the aspiration source, the Penumbra Reperfusion Catheter is used in conjunction with the Aspiration Pump, which is connected using the Penumbra Aspiration Tubing and the Penumbra Pump/Canister Tubing. The Penumbra Reperfusion Catheter is provided with a steam shaping mandrel and rotating hemostasis valve, and a peelable sheath. The Penumbra 3D Revascularization Device is provided with an introducer sheath. The Penumbra Separator is provided with an introducer and torque device. The Penumbra Reperfusion Catheters, 3D Revascularization Device and Separators are visible under fluoroscopy.

AI/ML Overview

This document is a 510(k) Summary for the Penumbra System (Reperfusion Catheter RED 62). It asserts substantial equivalence to predicate devices based on bench testing. The information provided is characteristic of a Class II medical device submission, which generally focuses on non-clinical performance data unless clinical data is deemed necessary to establish substantial equivalence.

Based on the provided text, the device in question is a physical medical device (catheter) and not an AI/ML powered device. Therefore, many of the requested criteria related to AI/ML device performance, such as ground truth establishment with experts, MRMC studies, standalone algorithm performance, and training set details, are not applicable to this submission.

Here's an analysis of the provided document regarding acceptance criteria and study proof:

1. A table of acceptance criteria and the reported device performance:

The document provides a table of "Design Verification Tests" and their conclusions. The "Acceptance Criteria" themselves are not explicitly detailed in numerical or categorical form within this table, but the "Conclusion" column consistently states "Acceptance Criteria Met," implying that internal, predefined criteria were satisfied.

TestTest Method SummaryReported Performance / Conclusion
Dimensional/Visual TestConfirms the units meet all dimensional and visual product specifications.Acceptance Criteria Met
Friction TestConfirms units meet product specification related to friction.Acceptance Criteria Met
Fluoroscopy TestConfirms the marker band is fluoroscopically visible.Acceptance Criteria Met
Simulated Use TestConfirms the functionality of units using clinically relevant benchtop model.Acceptance Criteria Met
Particulate TestParticulates generated during simulated use (including multiple deployment cycling) were evaluated.Acceptance Criteria Met
Hub Air TestConfirms units have no leaks when tested.Acceptance Criteria Met
Tensile TestConfirms units meet product specification related to tensile strength.Acceptance Criteria Met
Pressure TestConfirms units meet product specification related to pressure.Acceptance Criteria Met
Elongation TestConfirms units meet product specification related to elongation.Acceptance Criteria Met
Corrosion Resistance TestConfirms there is no visible corrosion on the units when tested.Acceptance Criteria Met
Torque Strength TestConfirms units have sufficient torque strength.Acceptance Criteria Met
Burst Pressure TestConfirms units can withstand sufficient pressure.Acceptance Criteria Met
Distal Tip Stiffness TestConfirms units have appropriate distal tip stiffness.Acceptance Criteria Met
Shelf-LifeConfirms expiration date based on accelerated aging test studies.Acceptance Criteria Met
Packaging Validation TestConfirms the packaging of the units meet all product specificationsAcceptance Criteria Met
Sterilization TestConfirms the units are sterilized in accordance with ISO 11135 and ISO 10993-7.Acceptance Criteria Met
Biocompatibility Tests
Cytotoxicity: MEM Elution (ISO 10993-5)No evidence of cell lysis or toxicity (Grade = 0, Reactivity None).Pass
Sensitization: Magnusson-Kligman Method (ISO 10993-10)Both Test Group & Control Group Grade = 0. None of the treated or negative control animals exhibited any reaction at the challenge. The positive control article elicited discrete reactions in all animals.Pass
Irritation: Intracutaneous Reactivity (ISO 10993-10)None of the animals exhibited overt signs of toxicity at any of the observation points. The test article sites did not show a significantly greater biological reaction than the sites injected with the control article.Pass
Systemic Toxicity: Acute Systemic Injection (ISO 10993-11)No evidence of systemic toxicity from sample extracts. (No deaths; No signs consistent with toxicity; No weight loss > 10%).Pass
Systemic Toxicity: Material Mediated Pyrogen (ISO 10993-11)Non-pyrogenic: no single animal had an individual rise in body temperature ≥ 0.5 °C.Pass
Hemocompatibility: In-vitro Thrombogenicity (ISO 10993-4)Device non-thrombogenic in vitro when compared to a predicate device.Pass
Hemocompatibility: Partial Thromboplastin Time (PTT) (ISO 10993-4)Test article coagulation times are statistically similar to predicate.Pass
Hemocompatibility: Complement Activation (ISO 10993-4)Test article concentrations of Sc5b-9 are statistically similar to predicate.Pass
Hemocompatibility: Hemolysis, indirect contact (ISO 10993-4)Hemolytic Index = 0.00%.Pass
Hemocompatibility: Hemolysis, direct contact (ISO 10993-4)Hemolytic Index = 0.00%.Pass

2. Sample size used for the test set and the data provenance:

  • Sample Size for Test Set: The document does not specify the exact sample size (number of units tested) for each of the design verification and biocompatibility tests. It broadly states "the units" were tested. This is common for bench testing summaries where detailed test reports would contain specific sample sizes.
  • Data Provenance: The data is from "Design Verification Testing" and "Biocompatibility" studies, which are typically performed in a laboratory setting by the manufacturer (Penumbra, Inc., Alameda, CA, USA). The studies are retrospective from the perspective of this 510(k) submission, meaning they were completed prior to the submission. The document does not specify the country of origin of where the testing itself was performed, but it references ISO standards and FDA GLP, implying internationally recognized practices. This is not clinical data from patients.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

This question is Not Applicable to this submission. This is a submission for a physical medical device (catheter) based on bench testing (performance and biocompatibility), not an AI/ML device that requires human expert review for establishing ground truth on image or patient data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

This question is Not Applicable. As stated above, this is bench testing of a physical device. There is no "ground truth" to be adjudicated by multiple readers or experts in the context of diagnostic agreement. The "acceptance criteria met" indicates that the device met pre-defined engineering and safety specifications.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This question is Not Applicable. This is not an AI/ML device, and no MRMC study was conducted or required. The document explicitly states: "No animal or clinical study was conducted as bench testing was determined sufficient for verification and validation purposes."

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

This question is Not Applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

This question is Not Applicable for the type of device and studies presented. The "ground truth" for this device is its adherence to engineering specifications, material properties, and biological safety standards, verified through objective bench tests (e.g., measurements for dimensions, leak tests, tensile strength tests, chemical analysis for biocompatibility). There is no "expert consensus" on imaging or clinical outcomes data required to validate the device's fundamental performance.

8. The sample size for the training set:

This question is Not Applicable. This is not an AI/ML device, so there is no training set mentioned or used.

9. How the ground truth for the training set was established:

This question is Not Applicable. As there is no training set (being a physical medical device and not an AI/ML device), no ground truth was established for it.

{0}------------------------------------------------

May 20, 2021

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Penumbra, Inc. Aditi Kolla Regulatory Affairs Program Manager One Penumbra Place Alameda, California 94502

Re: K203440

Trade/Device Name: Penumbra System (Reperfusion Catheter RED 62) Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NRY Dated: April 14, 2021 Received: April 15, 2021

Dear Aditi Kolla:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203440

Device Name Penumbra System (Reperfusion Catheter RED 62)

Indications for Use (Describe)

Penumbra Reperfusion Catheters and Separators

As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Penumbra 3D Revascularization Device

As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Penumbra Aspiration Tubing

As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump.

Penumbra Aspiration Pump The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "Penumbra" in red font, followed by a red circle with a white "P" inside. The "P" is stylized with a line through the middle. The logo is simple and modern. The text is clear and easy to read.

510(k) Summary

(as required by 21 CFR 807.92)

Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Penumbra, Inc. is providing the summary of Substantial Equivalence for the subject Penumbra System® (Reperfusion Catheter RED™ 62).

1.1 Submitter

Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA

Contact Person: Aditi Kolla Regulatory Affairs Program Manager Tel: (510) 995-2010 Fax: (510) 217-6414 E-mail: akolla@penumbrainc.com

Date of Preparation: May 19, 2021

Subject Device 1.2

Penumbra System® (Reperfusion Catheter RED™ 62)

Regulatory Class: II Classification Panel: Neurology Classification Name: Percutaneous Catheter Regulation Number: 21 CFR 870.1250 Product Code: NRY

1.3 Predicate Devices

510(k) NumberName of DeviceName ofManufacturer
K161640Penumbra System ACE 68 Reperfusion CatheterPenumbra,Inc.
K162901Penumbra 3D Revascularization DevicePenumbra,Inc.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Penumbra, a healthcare company. The word "Penumbra" is written in red, with a stylized "P" in a circle to the right. The logo is simple and modern, with a focus on the company's name.

Device Description 1.4

The Penumbra System® is comprised of the following devices:

  • Penumbra Reperfusion Catheter .
  • . Penumbra Aspiration Pump
  • Penumbra Aspiration Pump/Canister Tubing .
  • . Penumbra Aspiration Tubing
  • Penumbra Separator .
  • Penumbra 3D Revascularization Device .

The Penumbra System is designed to remove thrombus from the vasculature using continuous aspiration. The Reperfusion Catheter targets aspiration from the pump directly to the thrombus. The 3D Revascularization Device is used with Reperfusion Catheters to facilitate aspiration and removal of the thrombus when needed. The Separator may be used to clear the lumen of the Reperfusion Catheter should it become blocked with thrombus. The use of the Separator may not be necessary when using a Reperfusion Catheter with an ID of 0.054 in. or larger. The Reperfusion Catheter is introduced through a guide catheter or long femoral sheath and into the intracranial vasculature and guided over a neurovascular guidewire to the site of the primary occlusion. The Penumbra Reperfusion Catheter is used with the Aspiration Pump to aspirate thrombus from an occluded vessel. As needed, a Penumbra Separator may be deployed from the Reperfusion Catheter to assist with thrombus removal. The Penumbra Separator is advanced and retracted through the Penumbra Reperfusion Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the Reperfusion Catheter tip. For the aspiration source, the Penumbra Reperfusion Catheter is used in conjunction with the Aspiration Pump, which is connected using the Penumbra Aspiration Tubing and the Penumbra Pump/Canister Tubing. The Penumbra Reperfusion Catheter is provided with a steam shaping mandrel and rotating hemostasis valve, and a peelable sheath. The Penumbra 3D Revascularization Device is provided with an introducer sheath. The Penumbra Separator is provided with an introducer and torque device. The Penumbra Reperfusion Catheters, 3D Revascularization Device and Separators are visible under fluoroscopy.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "Penumbra" in red font. To the right of the word is a red circle with a white line going through the middle. The logo is simple and modern.

Indications For Use ાં રે

Penumbra Reperfusion Catheters and Separators

As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral -M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Penumbra 3D Revascularization Device

As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral -M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Penumbra Aspiration Tubing

As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump.

Penumbra Aspiration Pump

The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

1.6 Comparison of Indications for Use and Technological Characteristics with the Predicate Devices

Device AttributePredicate DeviceSubject Device
Trade NamePenumbra System ACE 68Reperfusion CatheterPenumbra System (ReperfusionCatheter RED 62)
FDA ProductClassificationClass II, NRY, 21 CFR 870.1250SAME
510(k) NumberK161640K203440
Indications for UsePenumbra Reperfusion Catheters andSeparatorsAs part of the Penumbra System, theReperfusion Catheters and SeparatorsSame as Predicate Device withK162901
Device AttributePredicate DeviceSubject Device
are indicated for use in the
revascularization of patients with
acute ischemic stroke secondary to
intracranial large vessel occlusive
disease (within the internal carotid,
middle cerebral - M1 and M2
segments, basilar, and vertebral
arteries) within 8 hours of symptom
onset.
Penumbra Aspiration Tubing
As part of the Penumbra System, the
Penumbra Sterile Aspiration Tubing
is indicated to connect the Penumbra
Reperfusion Catheters to the
Penumbra Pump MAX.
Penumbra Pump MAX
The Penumbra Pump MAX is
indicated as a vacuum source forPenumbra Aspiration Systems.
Principles ofOperationSee Section 1.4SAME
Device MaterialsStainless Steel, PTFE, Polyurethane,Polyether Block Amide, Nylon 12,Nitinol, Platinum/IridiumSAME
ID Band ColorYellow [black text]Black [white text]
CoatingHydrophilic coating (proprietary)Equivalent
Min. ID0.068 in. (1.73 mm)0.062 in. (1.57 mm)
Max. OD0.084 in. (2.13 mm)0.076 in. (1.93 mm)
Distal Flex Length30 cmSAME
Coating Length30 cmSAME
Effective Lengths115, 120, 125, 127, 132 cm115, 120, 125, 127, 132, 138,160 cm
AccessoriesPeelable Sheath, Shaping Mandrel,RHVSAME
PackagingMaterialsPolyester/Polyethylene/Tyvek,Polystyrene, SBS PaperboardSAME
Condition SuppliedSterile and Single UseSAME
SterilizationMethodEOSAME
DeviceAttributePredicate DeviceSubject Device
Trade NamePenumbra 3D Revascularization DevicePenumbra System(ReperfusionCatheter RED 62)
FDA ProductClassificationClass II, NRY, 21 CFR 870.1250SAME
510(k)NumberK162901K203440
IndicationsFor UsePenumbra Reperfusion Catheters and SeparatorsAs part of the Penumbra System, the ReperfusionCatheters and Separators are indicated for use in therevascularization of patients with acute ischemic strokesecondary to intracranial large vessel occlusive disease(within the internal carotid, middle cerebral - M1 andM2 segments, basilar, and vertebral arteries) within 8hours of symptom onset. Patients who are ineligible forintravenous tissue plasminogen activator (IV t-PA) orwho fail IV t-PA therapy are candidates for treatment.Penumbra 3D Revascularization DeviceAs part of the Penumbra System, the Penumbra 3DRevascularization Device is indicated for use in therevascularization of patients with acute ischemic strokesecondary to intracranial large vessel occlusive disease(within the internal carotid, middle cerebral - M1 andM2 segments) within 8 hours of symptom onset. Patientswho are ineligible for intravenous tissue plasminogenactivator (IV t-PA) or who fail IV t-PA therapy arecandidates for treatment.Penumbra Aspiration TubingAs part of the Penumbra System, the Penumbra SterileAspiration Tubing is indicated to connect the PenumbraReperfusion Catheters to the Penumbra AspirationPump.SAME
DeviceAttributePredicate DeviceSubject Device
Penumbra Aspiration PumpThe Penumbra Aspiration Pump is indicated as avacuum source for Penumbra Aspiration Systems.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for Penumbra, Inc., a healthcare company. The word "Penumbra" is written in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern, and the red color gives it a sense of energy and urgency.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "Penumbra" in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The logo appears to be for a company or organization called Penumbra.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "Penumbra" in a red, sans-serif font. To the right of the word is a red circle with a white outline. Inside the circle is a white "P" with a horizontal line through the middle.

Performance Data 1.7

The following performance data were provided in support of the substantial equivalence determination:

  • Design Verification ●
  • Biocompatibility ●
  • Shelf Life .
  • . Sterilization
  • Packaging Validation .

The subject device met all established requirements.

Design Verification Testing 1.7.1

The following design verification tests were performed on the subject device:

TestTest Method SummaryConclusion
Dimensional/VisualTestConfirms the units meet all dimensional and visualproduct specifications.Acceptance Criteria Met
Friction TestConfirms units meet product specification related tofriction.Acceptance Criteria Met
Fluoroscopy TestConfirms the marker band is fluoroscopically visible.Acceptance Criteria Met
Simulated Use TestConfirms the functionality of units using clinicallyrelevant benchtop model.Acceptance Criteria Met
Particulate TestParticulates generated during simulated use (includingmultiple deployment cycling) were evaluated.Acceptance Criteria Met
Hub Air TestConfirms units have no leaks when tested.Acceptance Criteria Met
Tensile TestConfirms units meet product specification related totensile strength.Acceptance Criteria Met
Pressure TestConfirms units meet product specification related topressure.Acceptance Criteria Met
Elongation TestConfirms units meet product specification related toelongation.Acceptance Criteria Met
Corrosion ResistanceTestConfirms there is no visible corrosion on the unitswhen tested.Acceptance Criteria Met
Torque Strength TestConfirms units have sufficient torque strength.Acceptance Criteria Met
Burst Pressure TestConfirms units can withstand sufficient pressure.Acceptance Criteria Met
Distal Tip StiffnessTestConfirms units have appropriate distal tip stiffness.Acceptance Criteria Met
Shelf-LifeConfirms expiration date based on accelerated agingtest studies.Acceptance Criteria Met
Packaging ValidationTestConfirms the packaging of the units meet all productspecificationsAcceptance Criteria Met

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image contains the word "Penumbra" in red font. To the right of the word is a red circle with a white "P" inside of it. The "P" is stylized with a white line going through the middle of it. The image is a logo for the company Penumbra.

K203440

TestTest Method SummaryConclusion
Sterilization TestConfirms the units are sterilized in accordance withISO 11135 and ISO 10993-7.Acceptance Criteria Met

1.7.2 Biocompatibility

The biocompatibility evaluation for the subject device was conducted in accordance with ISO 10993-1, USP standards, and FDA Good Laboratory Practices (GLP) as recognized by FDA. The battery of testing included the following tests:

TestsResultsConclusion
Cytotoxicity:MEM Elution(ISO 10993-5)No evidence of cell lysis or toxicity (Grade = 0,Reactivity None).Pass
Sensitization:Magnusson-KligmanMethod(ISO 10993-10)Both Test Group & Control Group Grade = 0None of the treated or negative control animals exhibitedany reaction at the challenge. The positive control articleelicited discrete reactions in all animals.Pass
Irritation:IntracutaneousReactivity(ISO 10993-10)None of the animals exhibited overt signs of toxicity atany of the observation points. The test article sites did notshow a significantly greater biological reaction than thesites injected with the control article.Pass
Systemic Toxicity:Acute Systemic Injection(ISO 10993-11)No evidence of systemic toxicity from sample extracts.That is:●No deaths;No signs consistent with toxicity;●No weight loss > 10%.●Pass
Systemic Toxicity:Material MediatedPyrogen(ISO 10993-11)Non-pyrogenic: no single animal had an individual rise inbody temperature ≥ 0.5 °C.Pass
Hemocompatibility:In-vitroThrombogenicity(ISO 10993-4)Device non-thrombogenic in vitro when compared to apredicate device.Pass
Hemocompatibility:Partial ThromboplastinTime (PTT)(ISO 10993-4)Test article coagulation times are statistically similar topredicate.Pass
Hemocompatibility:Complement Activation(ISO 10993-4)Test article concentrations of Sc5b-9 are statisticallysimilar to predicate.Pass
Hemocompatibility:Hemolysis, indirectcontact (ISO 10993-4)Hemolytic Index = 0.00%.Pass

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the word "Penumbra" in red font, followed by a red circle with a white "P" inside. The word "Penumbra" is written in a bold, sans-serif font. The "P" in the circle is also in a sans-serif font and is slightly offset to the right. The overall design is simple and modern.

TestsResultsConclusion
Hemocompatibility:Hemolysis, directcontact (ISO 10993-4)Hemolytic Index = 0.00%.Pass

Biocompatibility test results demonstrate biological safety per ISO 10993 and USP requirements.

1.7.3 Performance Data – Animal, Clinical

No animal or clinical study was conducted as bench testing was determined sufficient for verification and validation purposes.

Conclusions 1.8

The subject Penumbra System® (Reperfusion Catheter RED™ 62) is substantially equivalent to the predicate devices Penumbra System ACE 68 Reperfusion Catheter and Penumbra 3D Revascularization Device. The subject device has the same intended use as the predicate devices. The device testing described in the 510(k) Summary demonstrates the subject device is substantially equivalent to the predicate devices in regard to intended use, operating principle, design concept, fundamental technology and device performance.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).