K Number
K223186

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-02-13

(124 days)

Product Code
Regulation Number
870.5150
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

INDIGO Aspiration Catheters and Separators: As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems, and for the treatment of pulmonary embolism. INDIGO Aspiration Tubing: As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump. Penumbra Aspiration Pump: The Penumbra Aspiration Pump is indicated as a vacuum source for the Penumbra Aspiration Systems.

Device Description

The INDIGO® Aspiration System is comprised of the several devices: - INDIGO Aspiration Catheter - Penumbra Aspiration Pump - · INDIGO Aspiration Pump Canister - INDIGO Aspiration Tubing - INDIGO Separator The INDIGO Aspiration System is designed to remove thrombus from the vasculature using mechanical aspiration. The INDIGO Aspiration Catheter targets aspiration from the pump directly to the thrombus. The INDIGO Separator may be used to clear the lumen of the INDIGO Aspiration Catheter should it become blocked with thrombus. The INDIGO Aspiration Catheter is introduced through a guide catheter or vascular sheath into the peripheral vasculature and guided over a guidewire to the site of the primary occlusion. The INDIGO Aspiration Catheter is used with the Penumbra Aspiration Pump to aspirate thrombus from an occluded vessel. As needed, an INDIGO Separator may be deployed from the INDIGO Aspiration Catheter to assist with thrombus removal. The INDIGO Separator is advanced and retracted through the INDIGO Aspiration Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the INDIGO Aspiration Catheter tip. The INDIGO Aspiration Catheter may be provided with a steam shaping mandrel, rotating hemostasis valve, and introducer. The INDIGO Separator may be provided with an introducer and torque devices are visible under fluoroscopy. For the aspiration source, the INDIGO Aspiration Catheter is used in conjunction with the Penumbra Aspiration Pump, which is connected using the INDIGO Aspiration Tubing and the INDIGO Aspiration Pump Canister.

AI/ML Overview

This document describes the Indigo® Aspiration System - Lightning Bolt Aspiration Tubing (K223186). However, it does not describe a study that proves the device meets specific acceptance criteria in the context of image analysis, AI performance, or clinical effectiveness as one might expect for an AI/ML medical device. Instead, this document presents a 510(k) summary for a medical device (aspiration tubing) and focuses on demonstrating substantial equivalence to a predicate device through non-clinical (bench-top) testing.

Therefore, many of the requested elements regarding acceptance criteria, performance metrics, sample sizes for test/training sets, expert involvement, and ground truth establishment (which are standard for AI/ML device evaluations) are not applicable to the information provided in this document.

However, I can extract the relevant information on the non-clinical testing performed and the general acceptance of satisfactory performance.

Here's the breakdown based on the provided text, indicating where information is present and where it is not applicable:

1. Table of acceptance criteria and the reported device performance:

The document broadly states that the device met acceptance criteria, but it does not specify quantitative acceptance criteria for each test or provide detailed performance measurements.

Test PerformedReported Device Performance
BiocompatibilityLightning Bolt Aspiration Tubing shown to be biocompatible.
Dimensional/Visual InspectionMet all acceptance criteria.
Pressurization TestingMet all acceptance criteria.
Indigo Aspiration System Compatibility/Simulated Use Testing (Thrombus Removal Testing)Performed similarly to the predicate devices. Met all acceptance criteria.
Valve TestingMet all acceptance criteria.
Tensile TestingMet all acceptance criteria.
Post Destructive Testing Dimensional InspectionMet all acceptance criteria.
Electrical Safety/EMC TestingComplies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, AIM 7351731, and ISO 10079-1.
Software Verification & ValidationTesting and documentation provided as recommended by FDA guidance.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the document. The testing described is non-clinical (bench-top).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable as the document describes non-clinical engineering and biocompatibility testing, not evaluation by human experts, or clinical studies for AI/ML device.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable as this is not an AI/ML device for diagnostic imaging or similar applications that would involve human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. The device is an aspiration tubing, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the non-clinical tests, the "ground truth" would be established by engineering specifications, accepted ISO/IEC standards, and observation of physical performance (e.g., successful thrombus removal in a simulated environment, material integrity). It is not based on expert consensus, pathology, or outcomes data in the traditional sense for medical imaging or AI/ML evaluations.

8. The sample size for the training set

This information is not applicable; there is no AI/ML component with a training set described in this document.

9. How the ground truth for the training set was established

This information is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.

February 13, 2023

Penumbra, Inc. Deanna Kimlinger Senior Regulatory Specialist One Penumbra Place Alameda, California 95132

Re: K223186

Trade/Device Name: Indigo® Aspiration System - Lightning Bolt Aspiration Tubing Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: QEW Dated: January 13, 2023 Received: January 17, 2023

Dear Deanna Kimlinger:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Gregory W. Digitally signed by
Gregory W. Gregory W. O'connell -S O'connell -S Date: 2023.02.13
12-05'00'

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223186

Device Name

Indigo® Aspiration System - Lightning Bolt Aspiration Tubing

Indications for Use (Describe)

INDIGO Aspiration Catheters and Separators:

As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems, and for the treatment of pulmonary embolism.

INDIGO Aspiration Tubing:

As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump.

Penumbra Aspiration Pump:

The Penumbra Aspiration Pump is indicated as a vacuum source for the Penumbra Aspiration Systems.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

ਹ 510(k) Summary

(as required by 21 CFR 807.92)

Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Penumbra, Inc. is providing the summary of Substantial Equivalence for the Indigo "Aspiration System - Lightning Bolt Aspiration Tubing.

1.1 Sponsor/Applicant Name and Address

Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA

1.2 Sponsor Contact Information

Deanna Kimlinger Senior Regulatory Affairs Specialist Phone: (925) 212-9088 FAX: (510) 217-6414 Email: dkimlinger@penumbrainc.com

1.3 Date of Preparation of 510(k) Summary

February 08, 2023

1.4 Device Trade or Proprietary Name

Indigo Aspiration System - Lightning Bolt Aspiration Tubing

1.5 Device Classification

Regulatory Class:II
Classification Panel:Cardiovascular
Classification Name:Catheter, Embolectomy
Regulation Number:21 CFR §870.5150
Product Code:QEW

1.6 Predicate and Reference Devices – Lightning Bolt Aspiration Tubing

510(k) NumberName of Device
Lightning Bolt Aspiration Tubing
PredicateK210323 Indigo Aspiration System – Lightning Aspiration Tubing
Catheter 7 Aspiration Catheter
ReferenceK210083 Indigo Aspiration System Catheter 7, Indigo Aspiration System Separator 7,Lightning 7
System NameIndigo® Aspiration System
Lightning Aspiration Tubing[Predicate]Lightning Bolt Aspiration Tubing¹[Subject]
ClassificationClass II, QEWSAME
510(k) no.K210323K223186
IndicationINDIGO Aspiration Catheters andSeparators:As part of the INDIGO AspirationSystem, the INDIGO AspirationCatheters and Separators are indicatedfor the removal of fresh, soft emboli andthrombi from vessels of the peripheralarterial and venous systems, and for thetreatment of pulmonary embolism.INDIGO Aspiration Tubing:As part of the INDIGO AspirationSystem, the INDIGO Sterile AspirationTubing is indicated to connect theINDIGO Aspiration Catheters to thePenumbra Aspiration Pump.Penumbra Aspiration Pump:The Penumbra Aspiration Pump isindicated as a vacuum source for thePenumbra Aspiration Systems.SAME AS PREDICATE K210323
Aspiration TubingLightning Aspiration Tubing[Predicate]Lightning Bolt Aspiration Tubing[Subject]
510(k) No.K210323K223186
Materials
MaterialsBiocompatible, commonly utilized forinterventional devicesSAME
Dimensions
Tubing Inner Diameter (ID)0.125 in MINSAME
Distal Tubing Outer Diameter(OD)0.220 in MIN0.225 ± 0.007 in
Paratubing Outer Diameter(OD)0.223 in MIN0.230 + 0.009/- 0.007 in
Saline Tubing Outer Diameter(OD)N/A - Not present0.225 ± 0.007 in
Overall Length100 + 7 in120 + 7 in
System NameIndigo® Aspiration System
Lightning Aspiration Tubing[Predicate]Lightning Bolt Aspiration Tubing1[Subject]
Indigo System Attributes
Packaging MaterialsCommonly used materials forinterventional devicesSAME
Aspiration SourcePenumbra Aspiration PumpSAME
SterilizationEOSAME
Shelf-Life36 Months12 Months
UseSingle use, disposableSAME

{4}------------------------------------------------

1.7 Predicate Comparison

{5}------------------------------------------------

4 The Indigo System Aspiration Catheters, Separators, and Penumbra Aspiration Pumps are unchanged and remain identical to those currently cleared in Section 1.6.

1.8 Device Description

The INDIGO® Aspiration System is comprised of the several devices:

  • INDIGO Aspiration Catheter
  • Penumbra Aspiration Pump
  • · INDIGO Aspiration Pump Canister
  • INDIGO Aspiration Tubing
  • INDIGO Separator

The INDIGO Aspiration System is designed to remove thrombus from the vasculature using mechanical aspiration. The INDIGO Aspiration Catheter targets aspiration from the pump directly to the thrombus. The INDIGO Separator may be used to clear the lumen of the INDIGO Aspiration Catheter should it become blocked with thrombus. The INDIGO Aspiration Catheter is introduced through a guide catheter or vascular sheath into the peripheral vasculature and guided over a guidewire to the site of the primary occlusion. The INDIGO Aspiration Catheter is used with the Penumbra Aspiration Pump to aspirate thrombus from an occluded vessel. As needed, an INDIGO Separator may be deployed from the INDIGO Aspiration Catheter to assist with thrombus removal. The INDIGO Separator is advanced and retracted through the INDIGO Aspiration Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the INDIGO Aspiration Catheter tip. The INDIGO Aspiration Catheter may be provided with a steam shaping mandrel, rotating hemostasis valve, and introducer. The INDIGO Separator may be provided with an introducer and torque devices are visible under fluoroscopy. For the aspiration source, the INDIGO Aspiration Catheter is used in conjunction with the Penumbra Aspiration Pump, which is connected using the INDIGO Aspiration Tubing and the INDIGO Aspiration Pump Canister.

{6}------------------------------------------------

Lightning Bolt Aspiration Tubing

The Lightning Bolt Aspiration Tubing (INDIGO Aspiration Tubing) is designed to serve as a conduit to assist in thrombus removal, facilitating transfer of vacuum between the Penumbra Aspiration Pump and the INDIGO Aspiration Catheter while providing intermittent, continuous, or modulated aspiration. Modulated aspiration is provided when the Lightning Bolt Aspiration Tubing alternates between connecting the INDIGO Aspiration Catheter to the Penumbra Aspiration Pump and a sterile saline intravenous (IV) bag at ambient pressure.

1.9 Indications for Use

INDIGO Aspiration Catheters and Separators

As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems, and for the treatment of pulmonary embolism.

INDIGO Aspiration Tubing

As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump.

Penumbra Aspiration Pump

The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

1.10 Summary of Non-Clinical Data/ Performance Data

As required under Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, a summary of any information regarding substantial equivalence of the devices follows. Included in this section are summary descriptions of the testing which substantiates the performance of the subject Lightning Bolt Aspiration Tubing.

1.10.1 Biocompatibility

Biocompatibility testing was conducted on the subject Lightning Bolt Aspiration Tubing. The following tests were performed:

  • o Cytotoxicity
  • Sensitization o
  • o Irritation
  • O Systemic Toxicity

{7}------------------------------------------------

  • Acute Systemic Injection
  • I Material Mediated Pyrogenicity
  • Hemocompatibility o
    • . Indirect contact

The results from the testing performed showed the Lightning Bolt Aspiration Tubing to be biocompatible.

1.10.2 Design Verification (Bench-Top) Testing

Non-clinical laboratory testing was performed on the subject Lightning Bolt Aspiration Tubing device to determine substantial equivalence. The following tests were performed:

  • O Dimensional/Visual Inspection
  • O Pressurization Testing
  • Indigo Aspiration System Compatibility/Simulated Use Testing O
    • . Thrombus Removal Testing
  • O Valve testing
  • Tensile Testing O
  • Post Destructive Testing Dimensional Inspection o

The in vitro bench-top tests demonstrated that the subject Lightning Bolt Aspiration Tubing met all acceptance criteria and performed similarly to the predicate devices. Performance data demonstrate that the Lightning Bolt Aspiration Tubing device functions as intended and has a safety and effectiveness profile that is similar to the predicate devices.

1.10.3 Performance Data - Clinical

No clinical study was conducted as bench and previously performed animal testing was determined sufficient for verification and validation purposes.

1.10.4 Electrical Safety/EMC Testing

Electrical Safety and EMC testing were conducted on the subject Lightning Bolt Aspiration Tubing. The subject device complies with the requirements of IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, AIM 7351731, and ISO 10079-1.

1.10.5 Software

Software verification and validation testing and documentation for the subject Lightning Bolt Aspiration Tubing was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (issued May 11, 2005).

{8}------------------------------------------------

1.11 Summary of Substantial Equivalence

The subject Lightning Bolt Aspiration Tubing is substantially equivalent to the predicate devices, provided in Section 1.6 with regards to intended use, operating principle, design concept, materials, sterilization processes and packaging processes.

§ 870.5150 Embolectomy catheter.

(a)
Identification. An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.(b)
Classification. Class II (performance standards).