Search Results
Found 460 results
510(k) Data Aggregation
(31 days)
K252364**
Trade/Device Name: Edwards eSheath Optima introducer set
Regulation Number: 21 CFR 870.1340
Introducer Set
Common Name: Catheter, Introducer
Classification: Catheter, Introducer
21 CFR 870.1340
The Edwards eSheath Optima introducer set is indicated for the introduction and removal of compatible devices used with Edwards transcatheter heart valves.
The Edwards eSheath Optima introducer set (herein referred to as Optima set), model 14000ES16, consists of a sheath, vessel dilator, introducer, and in-sheath dilator. The Optima set is available with inner sheath diameter of 16 French and is used to facilitate introduction and removal of compatible devices used with the Edwards transcatheter heart valve (THV) systems into/from the vasculature.
The sheath shaft is composed of two layers; as devices are passed through the sheath, the inner member expands by sliding against itself while the outer jacket expands by stretching radially, temporarily expanding the shaft diameter. A radiopaque marker on the distal end indicates the location of the sheath tip in the body and a hydrophilic coating on the sheath tubing exterior facilitates introduction into the vessel. The sheath tubing mates with a housing, which holds three seals to provide hemostasis, and an extension tube with stopcock for flushing.
The vessel dilator is used to dilate the vessel prior to sheath insertion. The introducer is inserted into the sheath hub and locked prior to insertion into the body over a guidewire. The in-sheath dilator is used to expand the sheath during device use at the physician's discretion. The introducer, vessel dilator, and in-sheath dilator are radiopaque to improve fluoroscopic visibility intra-procedure.
The 29mm loader (included with the Edwards delivery system) features a disc valve within the loader cap assembly to help maintain hemostasis, and a scored perforation on the loader tube allowing the loader tubing to be "peeled away" and removed to utilize the full working length of the inserted device.
The provided text is a 510(k) clearance letter for a medical device called the Edwards eSheath Optima Introducer Set. It outlines the regulatory process and asserts the device's substantial equivalence to a predicate device. However, it does not contain any information about acceptance criteria or a study proving that an AI/software device meets those criteria, as typically seen in AI/ML medical device submissions.
The document pertains to a physical medical device (catheter introducer) and its non-clinical performance testing. The "summary of non-clinical testing" section lists various engineering and biocompatibility tests performed on the device itself, not on an AI algorithm.
Therefore, I cannot fulfill your request for:
- A table of acceptance criteria and reported device performance for an AI/software device.
- Sample sizes, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, ground truth types, training set details, or how training ground truth was established, because this information is not present in the provided document regarding an AI/software device.
The document is about a physical medical device and its non-clinical (engineering and biocompatibility) testing, not an AI/ML-based device.
Ask a specific question about this device
(28 days)
Intravascular or IV Catheter/ Catheter Introducer |
| Regulation Number: | 21 CFR 880.5200
21 CFR 870.1340
Intravascular or IV Catheter/Catheter Introducer |
| Regulation Number: | 21 CFR 880.5200
21 CFR 870.1340
----------|------------|------------|------------|
| Classification | 21 CFR 880.5200
21 CFR 870.1340
– Catheter Introducer (BD Insyte™ Autoguard™ Shielded IV Catheter ONLY) | 21 CFR 880.5200
21 CFR 870.1340
BD Insyte™ Autoguard™ Shielded IV Catheter is intended to be inserted into a patient's peripheral vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. This device may also be used to facilitate the placement of guidewires and other vascular access devices such as catheters without pre-attached hubs. This device may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa).
BD Insyte™ Autoguard™ BC Shielded IV Catheter is intended to be inserted into a patient's peripheral vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. This device may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa).
BD Insyte™ Autoguard™ BC Pro Shielded IV Catheter is intended to be inserted into a patient's peripheral vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. This device may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa).
BD Insyte™ Autoguard™ Shielded IV Catheter is an over-the-needle, intravascular (IV) catheter. This device includes a radiopaque BD Vialon™ Catheter Material, a needle, a grip with button, a flash chamber with vent plug, and a barrel. The needle and catheter are protected by a needle cover.
The 24-20 GA (0.7-1.1 mm) devices have BD Instaflash™ Needle Technology, allowing for immediate visualization of blood along the catheter. The flash chamber provides confirmation that the device has entered the vessel.
BD Insyte™ Autoguard™ Shielded IV Catheter incorporates BD Autoguard™ Push-button Shielding Technology which is activated when the button is depressed. A spring mechanism retracts the needle and flash chamber into the barrel, fully encasing the needle and reducing the risk of accidental needlestick injury.
BD Insyte™ Autoguard™ Shielded IV Catheter is available with or without wings. The catheter hub and wings are color coded to indicate catheter gauge size (24 GA (0.7 mm) = Yellow, 22 GA (0.9 mm) = Blue, 20 GA (1.1 mm) = Pink, 18 GA (1.3 mm) = Green, 16 GA (1.7 mm) = Grey, 14 GA (2.1 mm) = Orange).
BD Insyte™ Autoguard™ Shielded IV Catheter facilitates the introduction of guidewires and other vascular access devices. The following labeled gauge sizes correspond to the following inside diameter (ID): 24 GA (0.40 mm ID), 22 GA (0.55 mm ID), 20 GA (0.7 mm ID), 18 GA (0.8 mm ID), 16 GA (1.2 mm ID), and 14 GA (1.6 mm ID).
The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa).
BD Insyte™ Autoguard™ BC Shielded IV Catheter is an over-the-needle, intravascular (IV) catheter with blood control technology. This device includes a radiopaque BD Vialon™ Catheter Material, a needle, a grip with button, a flash chamber with vent plug, and a barrel. The needle and catheter are protected by a needle cover.
The blood control technology is designed to stop the flow of blood from the catheter hub until the initial Luer connection is made. Once the connection is made, fluids or blood can flow through the catheter hub in either direction.
The 24-18 GA (0.7-1.1 mm) devices have BD Instaflash™ Needle Technology, allowing for immediate visualization of blood along the catheter. The flash chamber provides confirmation that the device has entered the vessel.
BD Insyte™ Autoguard™ BC Shielded IV Catheter incorporate BD Autoguard™ Push-button Shielding Technology which is activated when the button is depressed. A spring mechanism retracts the needle and flash chamber into the barrel, fully encasing the needle and reducing the risk of accidental needlestick injury.
BD Insyte™ Autoguard™ BC Shielded IV Catheter is available with or without wings. The catheter hub and wings are color coded to indicate the catheter gauge size (24 GA (0.7 mm) = Yellow, 22 GA (0.9 mm) = Blue, 20 GA (1.1 mm) = Pink, 18 GA (1.3 mm) = Green, 16 GA (1.7 mm) = Grey).
The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa).
BD Insyte™ Autoguard™ BC Pro Shielded IV Catheter is an over-the-needle, intravascular (IV) catheter with blood control technology. This device includes a radiopaque BD Vialon™ Catheter Material, a needle, a grip with button, a flash chamber with vent plug, and a barrel. The needle and catheter are protected by a needle cover.
The blood control technology is designed to stop the flow of blood from the catheter hub until the initial Luer connection is made. Once the connection is made, fluids or blood can flow through the catheter hub in either direction.
The 24-18 GA (0.7-1.1 mm) devices have BD Instaflash™ Needle Technology, allowing for immediate visualization of blood along the catheter. The flash chamber provides confirmation that the device has entered the vessel.
BD Insyte™ Autoguard™ BC Pro Shielded IV Catheter incorporate BD Autoguard™ Push-button Shielding Technology which is activated when the button is depressed. A spring mechanism retracts the needle and flash chamber into the barrel, fully encasing the needle and reducing the risk of accidental needlestick injury.
BD Insyte™ Autoguard™ BC Pro Shielded IV Catheter is available with or without wings. The catheter hub and wings are color coded to indicate the catheter gauge size (24 GA (0.7 mm) = Yellow, 22 GA (0.9 mm) = Blue, 20 GA (1.1 mm) = Pink, 18 GA (1.3 mm) = Green, 16 GA (1.7 mm) = Grey).
The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa).
This document is a 510(k) clearance letter for three IV catheter devices, indicating that they are substantially equivalent to a previously cleared predicate device (K201075). It describes the changes made (new performance specifications and revised IFU) and provides a summary of performance tests conducted.
However, it does not contain the detailed acceptance criteria and reported device performance information that you specifically asked for in a table, nor does it describe a study that "proves the device meets the acceptance criteria" in the format of a clinical trial or algorithm validation study. The document primarily focuses on demonstrating substantial equivalence to a predicate device, which often relies on bench testing and comparisons of technological characteristics, rather than extensive clinical studies or AI algorithm performance validation.
Therefore, I cannot provide a table of acceptance criteria and reported device performance from this document because it is not present. Similarly, information regarding sample sizes for test/training sets, data provenance, expert ground truth adjudication for AI, MRMC studies, standalone algorithm performance, or training set ground truth establishment is not applicable or not provided within this 510(k) clearance document, as it pertains to a physical medical device (IV catheter) and not an AI/ML-driven device.
The study described to demonstrate substantial equivalence consists of bench testing to verify new performance specifications and ensuring compliance with relevant ISO and ASTM standards.
Here's an overview of the information that is available in relation to performance:
1. Table of Acceptance Criteria and Reported Device Performance:
- Not provided in this document. The document states that "the subject device met all predetermined acceptance criteria for the above-listed performance tests," but it does not specify what those acceptance criteria were (e.g., specific thresholds or ranges for frequency response, kink resistance, or blood fill time). It also does not present the numerical results ("reported device performance") of these tests.
2. Sample size used for the test set and the data provenance:
- Not applicable / Not provided. The document describes bench testing of physical devices, not an AI/ML system or a clinical study with a "test set" of patients or data in the context of AI validation.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable / Not provided. This information is relevant for AI/ML device validation and is not present for this physical medical device.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable / Not provided. This information is relevant for AI/ML device validation and is not present for this physical medical device.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This is not an AI/ML device, so an MRMC study is not relevant.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- No. This is not an AI/ML device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Not applicable / Not provided. For the performance tests listed (Frequency Response, Kink Resistance, Blood Fill Time, Catheter Separation Force), the "ground truth" would be the direct measurements obtained from the physical properties of the devices themselves, compared against pre-defined engineering specifications.
8. The sample size for the training set:
- Not applicable / Not provided. This is not an AI/ML device.
9. How the ground truth for the training set was established:
- Not applicable / Not provided. This is not an AI/ML device.
Summary of Performance Tests and Compliance (as described in the document):
The document states that a risk analysis was conducted to assess the impact of modifications. When technological characteristics were identical to the predicate, results from the predicate were applied. For other aspects, the following tests were conducted:
BD Internal Requirements:
- Frequency Response Testing: To support blood pressure monitoring indications.
- Kink Resistance Testing: To support blood pressure monitoring indications.
- Blood Fill Time Testing: To support blood sampling indications.
- Catheter Separation Force: (No specific indication mentioned, but likely related to structural integrity during withdrawal)
Compliance with Standards Testing:
- Luer Testing: According to ISO 80369-7 (Small-bore connectors for liquids and gases in healthcare applications).
- Packaging Testing: According to ISO 11607-1, ASTM F2096-11 (for gross leaks), and ASTM F88/F88M-15 (for seal strength).
- Biocompatibility: According to ISO 10993-1 (general biological evaluation) and ISO 10993-5 (for in vitro cytotoxicity).
Conclusion: The document explicitly states: "Per the design control requirements specified in 21 CFR §820.30, the subject device met all predetermined acceptance criteria for the above-listed performance tests, demonstrating substantial equivalence to the predicate devices." This indicates that the tests were performed and the results aligned with the internal and regulatory specifications, affirming the device's substantial equivalence.
Ask a specific question about this device
(35 days)
-|
| Common Name | Introducer, Catheter |
| Class | II |
| Classification Name | 21 CFR 870.1340
-|
| Common Name | Introducer, Catheter |
| Class | II |
| Classification Name | 21 CFR 870.1340
ViewFlex™ Xtra ICE Catheter
The ViewFlex™ Xtra ICE Catheter is indicated for use in adult and adolescent pediatric patients to visualize cardiac, structures, blood flow and other devices within the heart.
ViewFlex™ Eco Reprocessed ICE Catheter
The ViewFlex™ Eco Reprocessed Catheter is indicated for use in adult and adolescent pediatric patients to visualize cardiac structures, blood flow and other devices within the heart.
Advisor™ HD Grid Mapping Catheter, Sensor Enabled™
The Advisor™ HD Grid Mapping Catheter, Sensor Enabled™, is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart.
Advisor™ HD Grid X Mapping Catheter, Sensor Enabled™
The Advisor™ HD Grid X Mapping Catheter, Sensor Enabled™, is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart.
Agilis™ NxT Steerable Introducer
The Agilis™ NxT Steerable Introducer is indicated for the introduction of various cardiovascular catheters into the heart, including the left side of the heart, during the treatment of cardiac arrhythmias.
Agilis™ NxT Steerable Introducer Dual-Reach™
The Agilis™ NxT Steerable Introducer Dual-Reach™ is indicated for the introduction of various cardiovascular catheters into the heart, including the left side of the heart, during the treatment of cardiac arrhythmias.
The Agilis™ NxT Steerable Introducer Dual-Reach™ is a sterile, single-use device that con-sists of a dilator and steerable introducer, which is designed to provide flexible catheter positioning in the cardiac anatomy. The inner diameter of the steerable introducer is 13F. The steerable introducer includes a hemostasis valve to minimize blood loss during catheter intro-duction and/or exchange. It has a sideport with three-way stopcock for air or blood aspiration, fluid infusion, blood sampling, and pressure monitoring. The handle is equipped with a rotating collar to deflect the tip clockwise ≥180° and counterclockwise ≥90°. The steerable introducer features distal vent holes to facilitate aspiration and minimize cavitation and a radiopaque tip marker to improve fluoroscopic visualization.
This FDA 510(k) clearance letter (K251211) and its accompanying 510(k) summary pertain to a change in workflow for several existing cardiovascular catheters, specifically allowing for a "Zero/Low Fluoroscopy Workflow."
The key phrase here is "Special 510(k) – Zero/Low Fluoroscopy Workflow". This type of submission is for modifications to a previously cleared device that do not significantly alter its fundamental technology or intended use, but rather introduce a change in how it's used or processed.
Crucially, this submission does NOT describe a new AI/software device that requires extensive performance testing against acceptance criteria in the manner you've outlined for AI/ML devices. Instead, it's about demonstrating that existing devices, when used with a new, less-fluoroscopy-dependent workflow, remain as safe and effective as before.
Therefore, many of the questions you've asked regarding acceptance criteria, study details, ground truth, and expert adjudication are not applicable to the information provided in this 510(k) document. The document explicitly states:
- "Bench-testing was not necessary to validate the Clinical Workflow modifications."
- "Substantial Equivalence of the subject devices to the predicate devices using the zero/low fluoroscopy workflow has been supported through a summary of clinical data across multiple studies in which investigators used alternative visualization methods."
This indicates that the "study" proving the device (or rather, the new workflow) meets acceptance criteria is a summary of existing clinical data where alternative visualization methods were already employed, rather than a prospective, controlled study of a new AI algorithm.
Based on the provided document, here's what can be answered:
1. A table of acceptance criteria and the reported device performance:
- Acceptance Criteria: The implicit acceptance criterion is that the devices, when used with "zero/low fluoroscopy workflow," maintain substantial equivalence to their predicate devices in terms of safety and effectiveness. This means they must continue to perform as intended for visualizing cardiac structures, blood flow, mapping, or introducing catheters.
- Reported Device Performance: The document states that "Substantial Equivalence... has been supported through a summary of clinical data across multiple studies in which investigators used alternative visualization methods." This implies that the performance (e.g., adequate visualization, successful mapping, successful catheter introduction) was maintained. Specific quantitative metrics of performance (e.g., accuracy, sensitivity, specificity, or inter-reader agreement for a diagnostic AI) are not provided or applicable here as this is not an AI/ML diagnostic clearance.
2. Sample size used for the test set and the data provenance:
- Sample Size: Not specified. The document refers to "a summary of clinical data across multiple studies." This suggests an aggregation of results from existing (likely retrospective) patient data where alternative visualization techniques (allowing for "zero/low fluoroscopy") were already utilized clinically. It's not a new, single, prospectively designed test set for an AI algorithm.
- Data Provenance: Not specified regarding country of origin or specific patient demographics. It is implied to be clinical data collected from studies where these types of procedures were performed using alternative visualization. The data would be retrospective as it's a "summary of clinical data" that already exists.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable in the context of this 510(k). Ground truth in an AI/ML context typically refers to adjudicated labels for images or signals. Here, the "ground truth" is inferred from standard clinical practice and outcomes in the historical data summarized. There's no mention of a specific expert panel for new ground truth establishment for a diagnostic AI.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable. This is not a study requiring adjudication of diagnostic outputs by multiple readers.
5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This is not an AI-assisted diagnostic device. The workflow change is about using alternative non-fluoroscopic imaging modalities (e.g., intracardiac echocardiography, electro-anatomical mapping systems), not about AI improving human reader performance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- No. This is not an AI algorithm. The predicate devices are physical catheters.
7. The type of ground truth used:
- The "ground truth" is inferred from clinical outcomes and established clinical practice using the devices with alternative visualization methods in real-world scenarios. It's not a specific, adjudicated dataset for an AI algorithm. The performance of the devices (such as successful navigation, visualization, and mapping) under the "zero/low fluoroscopy" workflow is assumed to be equivalent to their performance under full fluoroscopy, as demonstrated by prior clinical use where such methods were employed.
8. The sample size for the training set:
- Not applicable. There is no AI model being trained discussed in this document.
9. How the ground truth for the training set was established:
- Not applicable. No training set for an AI model.
In summary:
This 510(k) is for a workflow modification for existing medical devices (catheters), not for an AI/ML diagnostic or assistive software. Therefore, the detailed data performance evaluation typically required for AI models against specific acceptance criteria (as requested in your template) is not presented or relevant in this clearance letter. The "proof" relies on the concept of substantial equivalence to previously cleared predicate devices, supported by a summary of existing clinical data that used alternative visualization methods, implying that the devices function safely and effectively even with reduced fluoroscopy.
Ask a specific question about this device
(29 days)
Common / Usual Name:** Catheter introducer
Classification Name: Catheter introducer, 21 CFR §870.1340
Regulation** | 21 CFR § 870.3680 – Cardiovascular permanent or temporary pacemaker electrode
21 CFR § 870.1340
introducer | 21 CFR § 870.3680 – Cardiovascular permanent or temporary pacemaker electrode
21 CFR § 870.1340
Bard® Temporary Pacing Catheters are designed to transmit an electrical signal from an external pulse generator to the heart or from the heart to a monitoring device.
The Needle / Cannula (Introducer) is intended for the introduction of Bard® Temporary Pacing Electrode Catheters into the venous vasculature.
Bard® Temporary Pacing Electrode Catheters:
Bard® Temporary Pacing Catheters are designed to transmit an electrical signal from an external pulse generator to the heart or from the heart to a monitoring device.
Bard® Temporary Pacing Electrode Catheters are constructed of insulated electrical wires encased within a woven shaft, which is then coated with various blends of radiopaque polyurethane-based materials. All Temporary Pacing Electrode Catheters are bipolar and include two stainless steel electrodes - one located along the shaft and one at the catheter distal tip. The proximal end of the devices includes the two electrical leads which have shrouded jacks and are used to establish electrical connections with an external pulse generator or monitoring device. Certain catheters may incorporate a lumen for balloon inflation.
Some products may be packaged with accessories such as a needle / cannula, an ECG adapter, or a balloon inflation syringe. All Bard® Temporary Catheter devices are packaged with safety adapters, are intended for prescription use only, and are for single use only.
Needle / Cannula (Introducer):
The Needle / Cannula (Introducer) is intended for the introduction of Bard® Temporary Pacing Electrode Catheters into the venous vasculature. The Needle / Cannula (Introducer) is not sold separately by C. R. Bard, Inc. and this accessory is only included with certain Bard® Temporary Pacing Electrode Catheter kits.
The devices are provided in two different French size / length variants, dependent on size / type of the accompanying Bard® Temporary Pacing Electrode Catheter.
The Needle / Cannula (Introducer) includes a needle and a cannula. The two components of the Introducer are inserted into the vein simultaneously. The needle is then withdrawn, leaving in place the cannula, through which the Bard® Temporary Pacing Electrode Catheter can be advanced through the vessel and into the desired placement location. The device is intended for prescription use only and is for single use only.
The provided FDA 510(k) clearance letter and summary describe the Bard® Temporary Pacing Electrode Catheter Needle / Cannula (Introducer). The submission is a "Special 510(k)" primarily focused on removing a warning label related to unknown electromagnetic compatibility (EMC) with low-frequency emissions.
Here's an analysis of the acceptance criteria and the study that proves the device meets those criteria, based on the provided text:
Acceptance Criteria and Device Performance
1. Table of Acceptance Criteria and Reported Device Performance
The core of this Special 510(k) is the demonstration of electromagnetic compatibility, specifically concerning immunity to proximity magnetic fields, to justify the removal of a previous warning label.
Acceptance Criteria Category | Specific Acceptance Criterion | Reported Device Performance |
---|---|---|
Electromagnetic Compatibility (EMC) | IEC 60601-1-2 (Edition 4.1): 2020 Compliance | Met EMC requirements as defined in IEC 60601-1-2 Clause 8.11 for Immunity to Proximity Magnetic Fields. |
- IEC 60601-1-2 Clause 8.11 (Immunity to Proximity Magnetic Fields): Tested using methods specified in IEC 61000-4-39 (2017) per applicable specifications in Table 11 of IEC 60601-1-2. | - Demonstrated electromagnetic compatibility with electronics capable of generating low-frequency emissions. | |
Functional Performance (Implicit) | The device, when used as a system with representative external pulse generator equipment, should maintain its essential performance (transmitting electrical signals for pacing/monitoring) without degradation or disruption in the presence of low-frequency electromagnetic interference. The acceptance criteria were the same as the acceptance criteria utilized for other IEC 60601-1-2 immunity testing performed as part of the predicate (K241334) clearance, implying that the device's basic safety and essential performance were maintained. | Testing confirmed that the device met EMC requirements, indicating its functional performance was not negatively impacted by the tested electromagnetic fields. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not explicitly state a numerical sample size (e.g., "N=X devices were tested"). It refers to "representative external pulse generator equipment and the sample Temporary Pacing Electrode devices." This typically implies a limited number of devices/systems sufficient to represent the product line for EMC testing as per regulatory guidelines for in-vitro/bench testing.
- Data Provenance: The study was a bench (in-vitro) test performed to a recognized international standard (IEC 60601-1-2). There is no patient data involved, so there is no country of origin or retrospective/prospective distinction.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
Not applicable. This was a bench-top engineering test evaluating electromagnetic compatibility against a defined international standard (IEC 60601-1-2). "Ground truth" in the clinical sense (e.g., diagnosis by experts) is not relevant here. The "ground truth" is the established test methodology and performance limits defined by the IEC standard.
4. Adjudication Method for the Test Set
Not applicable. As a bench test against objective standards, there is no expert adjudication process like those used in clinical studies for diagnostic accuracy. The results are objectively measured against the specified limits of the IEC standard.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a passive pacing electrode catheter and introducer, not an AI-powered diagnostic or therapeutic tool with a human-in-the-loop component.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. The device is a medical hardware component, not an algorithm. The testing focused on the physical and electrical compatibility of the hardware with relevant external equipment.
7. The Type of Ground Truth Used
The "ground truth" for this study is compliance with international consensus standards for electromagnetic compatibility, specifically IEC 60601-1-2 (Edition 4.1): 2020, with particular focus on Clause 8.11 and the methods specified in IEC 61000-4-39 (2017). This standard defines objective tests and acceptance limits for device performance under various electromagnetic conditions.
8. The Sample Size for the Training Set
Not applicable. There is no "training set" as this is not a machine learning or AI-driven device.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set.
Ask a specific question about this device
(27 days)
84119
Re: K251051
Trade/Device Name: AuST Steerable Sheath
Regulation Number: 21 CFR 870.1340
Steerable Sheath
Classification Name: Catheter Introducer
Regulatory Class: Class II per 21 CFR 870.1340
| Device Class | II | II | Same |
| Product Code | DYB | DYB | Same |
| Regulation number | 21 CFR 870.1340
| 21 CFR 870.1340 | Same |
| Duration of use | Single-use, Transient | Single-use, Transient | Same
The AuST Steerable Sheath is intended for the introduction of diagnostic and therapeutic devices into the human vasculature, including but not limited to intracardiac, renal or other peripheral placements. Do not use this device for neural placements.
The modified AuST Steerable Sheath is a single-use percutaneous catheter intended to provide a pathway through which diagnostic and therapeutic devices are introduced in the human vasculature. The catheter is not intended for neural placements.
The modified AuST Steerable Sheath product family consists of a variety of configurations to accommodate different anatomies and/or devices being introduced into the human vasculature. Each of the sheaths in the product family is comprised of a braid-reinforced catheter shaft, bi-directional deflectable segment, radiopaque distal tip, a handle with a 3-way stopcock for flushing and aspiration, and a hemostasis valve to prevent air and fluid leakage.
The provided FDA 510(k) Clearance Letter for the AuST Steerable Sheath (K251051) does not describe an AI/Software as a Medical Device (SaMD). Instead, it describes a physical medical device – a catheter introducer sheath.
Therefore, many of the requested criteria related to AI/SaMD performance studies, such as sample sizes for test/training sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, or standalone algorithm performance, are not applicable to this type of device submission.
The clearance letter focuses on demonstrating substantial equivalence to a predicate physical device through engineering and performance testing rather than clinical effectiveness studies for an AI algorithm.
Here's an analysis based on the information provided for a physical medical device:
Acceptance Criteria and Device Performance for AuST Steerable Sheath (K251051)
Since this is a submission for a physical medical device (catheter introducer) and not an AI/SaMD, the "acceptance criteria" discussed are primarily related to general device safety and performance standards for such devices, demonstrating substantial equivalence to a predicate device. The study design described is engineering and bench testing, not a clinical trial involving patient data or expert interpretation of AI output.
1. Table of Acceptance Criteria and Reported Device Performance
The submission focuses on ensuring the modified device meets established safety and performance benchmarks for catheter introducers and is substantially equivalent to its predicate. The "acceptance criteria" are implied by the successful completion of the listed tests.
Acceptance Criteria (Implied by Test Type) | Reported Device Performance |
---|---|
Sterilization Validation (per ANSI/AAMI/ISO 11135) | Device passed sterilization validaiton. |
Visual Inspection | Device passed visual inspection. |
Simulated Use Testing (including compatibility with ancillary devices) | Device passed simulated use testing. |
Valve Liquid Leak Test | Device passed valve liquid leak test. |
Tensile Tests | Device passed tensile tests. |
Sheath and Dilator Dimensional Verification | Device passed dimensional verification. |
Dilator Removal Force | Device passed dilator removal force test. |
Biocompatibility (per ISO 10993-1) | All test results met the requirements of ISO 10993-1, demonstrating no adverse biocompatibility risks. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: Not explicitly stated. The document mentions "test units representative of finished devices" for engineering tests. For physical device performance testing, sample sizes are typically determined by statistical methods for engineering verification, but specific numbers are not provided in this summary.
- Data Provenance: The data is generated from in-vitro (bench) testing of the manufactured medical device components and finished product, adhering to recognized standards (e.g., ISO, ANSI/AAMI). Since it's a physical device, there's no "country of origin of data" in the sense of patient data, nor is it retrospective or prospective clinical data.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This criterion is not applicable for this physical device. Ground truth, in the context of an AI/SaMD, refers to accurate diagnostic labels or measurements. For a physical device, "ground truth" is typically defined by engineering specifications, validated test methods, and industry standards. No human experts are described as establishing "ground truth" in this context.
4. Adjudication Method for the Test Set
This criterion is not applicable for this physical device. Adjudication methods (e.g., 2+1, 3+1) are common in clinical trials or AI performance evaluations where expert disagreement on ground truth or interpretation needs resolution. For engineering testing, adherence to a test protocol and meeting predefined specifications are the primary "adjudication."
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC study was not done. MRMC studies are specific to evaluating the diagnostic performance of imaging systems or AI algorithms, often comparing human readers with and without AI assistance to measure reader improvement (effect size). This is a physical device, and such a study is not relevant.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
No, this is not applicable. This is a physical medical device. "Standalone performance" refers to the performance of an AI algorithm without human input, which is not relevant to a catheter introducer sheath.
7. The Type of Ground Truth Used
The "ground truth" for this physical device is defined by:
- Engineering Specifications: The design parameters and performance characteristics that the device is designed to meet.
- Standard Test Methods: Validated methodologies (e.g., per ISO or AAMI standards) used to assess physical properties and performance (e.g., strength, leak resistance, dimensions, biocompatibility).
- Predicate Device Performance: The modified device is compared to a legally marketed predicate device (AuST Steerable Sheath, K242106), implying that the predicate's established performance serves as a benchmark for substantial equivalence.
8. The Sample Size for the Training Set
This criterion is not applicable. "Training set" refers to data used to train machine learning models. This device is a physical medical device, not an AI or software product.
9. How the Ground Truth for the Training Set Was Established
This criterion is not applicable. As there is no AI component, there is no "training set" or ground truth establishment relevant to machine learning.
Ask a specific question about this device
(30 days)
Trade/Device Name: Prelude Wave Hydrophilic Sheath Introducer
Regulation Number: 21 CFR 870.1340
| Catheter Introducer |
| Regulatory Class: | 2 |
| Product Code: | DYB |
| 21 CFR §: | 870.1340
The Merit Prelude Wave Hydrophilic Sheath Introducer is intended to provide access and facilitate the percutaneous introduction of various devices into veins and/or arteries, including but not limited to the radial artery, while maintaining hemostasis for a variety of diagnostic and therapeutic procedures.
The access needle with inner metal needle and outer plastic cannula is used to gain access to the vein or artery for placement of guidewires.
The Prelude Wave™ Hydrophilic Sheath Introducer consists of a sheath introducer with compatible vessel dilator that snaps securely into the sheath introducer hub. The sheath hub contains an integral hemostasis valve. The sheath tubing is coated with a hydrophilic coating. A sidearm is affixed to the sheath hub and has a 3-way stopcock at its proximal end. The Prelude Wave™ Hydrophilic Sheath Introducer is available in 11cm and 23cm lengths, (French sizes 4F, 5F, 6F and 7F) and is designed to accept 0.018", 0.021" and 0.025" diameter guidewires. The Prelude Wave™ Hydrophilic Sheath Introducer is marketed with any of the following components, depending on the product configuration: guidewire, metal access needle, access needle with inner metal needle and outer plastic cannula and SnapFix™ securement device.
The provided FDA 510(k) clearance letter and summary for the Prelude Wave Hydrophilic Sheath Introducer (K250909) do not contain the specific information requested in the prompt regarding acceptance criteria, device performance, sample sizes for training and testing, expert qualifications, or detailed descriptions of ground truth establishment for an AI/ML medical device.
This document describes a traditional medical device (a catheter introducer), not an AI/ML-based device. Therefore, the types of studies and acceptance criteria typically associated with AI/ML device clearance (as outlined in your prompt questions) are not applicable here.
The document primarily focuses on:
- Substantial Equivalence: Demonstrating that the new device is as safe and effective as a legally marketed predicate device (Prelude IDEAL Hydrophilic Sheath Introducer, K173750). This is a common pathway for traditional medical devices.
- Performance Testing: Verification and validation studies to ensure the device meets its specifications and performs as intended, based on established international standards (ISO 10555-1, EN ISO 11070, ISO 10993-1, ISO 14971). These tests are standard for physical medical devices and include mechanical, functional, and biocompatibility assessments.
- Design Changes: Highlighting minor design modifications from the predicate, such as sheath material, sheath/dilator tip fit, and inclusion of an adhesive fixation device.
Therefore, I cannot populate the table or answer the specific questions about AI/ML device study parameters (like MRMC, standalone AI performance, expert labeling, training data) based on the provided text. The provided text simply does not contain this information because it's a submission for a physical medical device, not an AI software.
Ask a specific question about this device
(122 days)
K243954**
Trade/Device Name: PeriCross™ Epicardial Access Kit
Regulation Number: 21 CFR 870.1340
Introducer, Catheter
Regulatory Class: II
Product Code(s): DYB
Regulation Number: 21 CFR 870.1340 |
---|
Class |
Product Code |
Regulation (FDA) |
870.1340 |
Intended Use / Indications for Use |
The PeriCross™ Epicardial Access Kit is intended to access the epicardial surface of the heart via a subxiphoid approach to facilitate electrophysiology studies in adult patients.
The PeriCross™ Epicardial Access Kit is designed to provide safe and reliable access to the epicardium. The kit is supplied sterile and is designed to deliver a .018" guidewire to the pericardial space.
The PeriCross™ Epicardial Access Kit consists of the following components:
- One (1) PeriCross™ Tunneler Assembly
- One (1) PeriCross™ Access Device
- One (1) Guidewire – 0.018" OD × 80cm Long
- One (1) Coaxial Introducer – 5F × 18cm
The PeriCross™ Tunneler assembly consists of a cannula and a removable blunt obturator that, together, create an atraumatic assembly, which is utilized to gain access to the pericardium via the subxiphoid approach. When the tip of the PeriCross™ Tunneler Assembly reaches the pericardial surface, the obturator is removed, leaving a path to the pericardial surface through the open cannula. The PeriCross™ Access Device is inserted into the cannula and secured via a snap connection. The PeriCross™ Access Device incorporates a thumb slide that deploys tines to engage and retract the pericardium away from the surface of the epicardium. The space created allows for the deployment of an integrated 21ga needle to puncture the pericardium in a location away from the heart. The PeriCross™ Access Device prevents inadvertent deployment of the needle and retraction of the tines via an interlock mechanism. A .018" guidewire can then be deployed through the PeriCross™ Access Device. A 5F x 18cm coaxial introducer is provided to allow for subsequent dilation and guidewire exchange to a .032" guidewire.
The provided FCC 510(k) clearance letter and summary for the PeriCross™ Epicardial Access Kit discuss its design, comparison to predicate devices, and general non-clinical and clinical testing performed to demonstrate substantial equivalence. However, it does not contain the specific details required to fully address all aspects of acceptance criteria and the study that proves the device meets them as requested in the prompt.
Specifically, the document states:
- "The primary efficacy endpoint of achieving guidewire access to the pericardial space using the PeriCross Epicardial Access Kit, confirmed by standard X-ray technique, was achieved in all subjects."
- "There were two non-serious cardiac perforations (≤ 80 mL of bleeding) and one serious cardiac perforation (≥ 80 mL of bleeding), with no surgery required for closure."
This information is related to the outcome of the study, but it does not provide a table of acceptance criteria (e.g., minimum success rate, maximum complication rate) or explicitly state how these specific results meet predefined acceptance criteria. It also lacks details about the test set, ground truth experts, adjudication, MRMC studies, or training set specifics.
Therefore, I will provide a response based on the inference of typical medical device study reporting and what might be implied by the passing results, acknowledging the limitations of the provided text.
Acceptance Criteria and Study Proving Device Performance: PeriCross™ Epicardial Access Kit
Based on the provided 510(k) summary, the primary efficacy endpoint for the PeriCross™ Epicardial Access Kit was the successful achievement of guidewire access to the pericardial space. Safety endpoints related to cardiac perforations were also assessed. While explicit numerical acceptance criteria are not detailed in the provided document, the successful clearance implies that the study results met the FDA's requirements for demonstrating substantial equivalence and safety/efficacy.
Inferring from the "Pass" results and the successful clearance, the implied acceptance criteria were met.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Inferred) | Reported Device Performance |
---|---|
Efficacy: | |
Guidewire access to the pericardial space achieved (primary efficacy endpoint) | Achieved in all subjects (39/39) |
Safety: | |
Acceptable rate of non-serious cardiac perforations ( |
Ask a specific question about this device
(87 days)
92614
Re: K244046
Trade/Device Name: Edwards eSheath Optima introducer set Regulation Number: 21 CFR 870.1340
|
| Classification: | Catheter, Introducer
21 CFR 870.1340
The Edwards eSheath Optima introducer set is indicated for the introduction and removal of compatible devices used with Edwards transcatheter heart valves.
The Edwards eSheath Optima introducer set consists of a sheath, vessel dilator, introducer, and insheath dilator. The Edwards Optima introducer set is available with inner sheath diameter of 14 French (model 14000ES14). The Edwards eSheath Optima introducer set is used to facilitate introduction of the Edwards transcatheter heart valve systems into the vasculature.
The sheath shaft is composed of two layers; as devices are passed through the sheath, the inner member expands by sliding against itself while the outer iacket expands by stretching radially. temporarily expanding the shaft diameter. A radiopaque marker on the distal end indicates the location of the sheath tip in the body and a hydrophilic coating on the sheath tubing exterior facilitates introduction into the vessel. The sheath tubing mates with a housing, which holds three seals to provide hemostasis, and an extension tube with stopcock for flushing.
The vessel dilator is used to dilate the vessel prior to sheath insertion. The introducer is inserted into the sheath hub and locked prior to insertion into the body over a guidewire. The in-sheath dilator is used to expand the sheath during device use at the physician's discretion. The introducer, vessel dilator, and in-sheath dilator are radiopaque to improve fluoroscopic visibility intra-procedure.
The loader (included in the Edwards delivery system) features a disc valve within the loader cap assembly to help maintain hemostasis, and a scored perforation on the loader tube allowing the loader tubing to be "peeled away" and removed to utilize the full working length of the inserted device.
The provided document is an FDA 510(k) clearance letter and summary for the Edwards eSheath Optima introducer set. It demonstrates substantial equivalence to a predicate device based on non-clinical testing. However, it does not contain a detailed study meeting the specific criteria you've outlined, particularly for AI/algorithm performance. The information provided relates to the physical and functional performance of a medical device, not an AI or software component.
Therefore, I cannot extract the following information from the provided text:
- 1. A table of acceptance criteria and the reported device performance: While the document mentions various bench tests were "successfully completed" and "all design requirements were met," it does not provide specific quantitative acceptance criteria or detailed numerical results for each test.
- 2. Sample sized used for the test set and the data provenance: Not applicable, as this is non-clinical bench testing of a physical device, not an AI or software study with a test set of data.
- 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
- 4. Adjudication method for the test set: Not applicable.
- 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this device does not involve AI or human readers.
- 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
- 7. The type of ground truth used: Not applicable in the context of AI/software performance. For a physical device, the "ground truth" would be the engineering specifications and established test methods.
- 8. The sample size for the training set: Not applicable.
- 9. How the ground truth for the training set was established: Not applicable.
Summary of available information from the document:
The document describes the Edwards eSheath Optima introducer set, a physical medical device.
Device Description:
The introducer set comprises a sheath, vessel dilator, introducer, and in-sheath dilator. It's designed to facilitate the introduction and removal of compatible devices used with Edwards transcatheter heart valves. Key features include an expandable inner member and outer jacket, a radiopaque marker, a hydrophilic coating, hemostatic seals, and an optional in-sheath dilator.
Non-Clinical Testing:
A list of bench tests was performed to demonstrate substantial equivalence to the predicate device and that all design requirements were met. These tests cover various physical and functional aspects of the device, including:
- Recovered Outer Diameter (OD)
- In-Sheath Dilator (ISD) Max Distal OD
- Tip OD
- ISD Insertion & Retrieval
- Tip Inner Diameter (ID)
- Sheath Insertion & Retrieval
- Sheath Working Length
- ISD Working Length
- Hemostasis
- Kink Radius
- Fishmouth
- Lubricity and Durability
- Bond Tensile Strengths (Sheath Housing to Shaft, Sheath Shaft to Tip, Flush Tube to Housing, Stopcock to Flush Tub, ISD Hub to Shaft)
- Transcatheter Heart Valve (THV)/Sheath Interaction
- Device Interaction
- Guidewire Compatibility
- Delivery System Insertion & Retrieval
- Crimped THV Retrieval
- Radiopacity
- Particulate Testing
- Sterilization Validation
- Biocompatibility (Cytotoxicity, Sensitization, Irritation, Material Mediated Pyrogenicity, Acute Systemic Toxicity, Hemocompatibility, Thrombogenicity)
The document states that these tests were "successfully completed" and that "all design requirements were met," leading to the conclusion that the device is substantially equivalent to the predicate. However, it does not provide specific acceptance criteria values or the quantitative results from these tests.
Ask a specific question about this device
(90 days)
17111 Israel
Re: K243928
Trade/Device Name: ViaOne Epicardial Access System Regulation Number: 21 CFR 870.1340
Usual Name: ViaOne Classification Name: Introducer, Catheter Regulatory Class: Class II, per 21 CFR 870.1340
EPi-Ease™ Epicardial Access System (EAS) cleared under K233959 (product . code DYB Regulation No. 21 CFR 870.1340
|
| CFR | 21 CFR 870.1340
| 21 CFR 870.1340
The ViaOne Epicardial Access System is intended to access the epicardial surface of the heart via a subxiphoid approach to deliver diagnostic and therapeutic interventions for ventricular tachycardia.
ViaOne Epicardial Access System is a sterile, single-use access tool consisting of a concealed needle within a leading channel to facilitate guidewire access into the pericardium through a subxiphoid incision. The device consists of a radiopaque distal tip designed to retract a portion of the pericardial tissue into the device, and away from the heart muscle, where it is punctured by the concealed needle and a guidewire is introduced into the pericardial space.
The provided text describes the 510(k) premarket notification for the ViaOne Epicardial Access System. However, it does not contain the detailed acceptance criteria and the specific study results in a format that would allow for a complete table of acceptance criteria and reported device performance.
The text generally states that the device "met the predetermined acceptance criteria" and "All tests were successfully completed and passed, and the device met the defined acceptance criteria." but does not quantify these criteria or results.
Here's a breakdown of what can be extracted and what information is missing:
1. Table of Acceptance Criteria and Reported Device Performance
This information is not provided in the document. The document lists performance tests but does not state the specific acceptance criteria (e.g., "tensile strength > X N") nor the quantitative results from those tests (e.g., "tensile strength Y N").
2. Sample size used for the test set and the data provenance
- Biocompatibility: Not specified.
- Sterilization, Packaging and Shelf Life Testing: Not specified.
- Performance Testing: Not specified for individual tests.
- Animal Study: Not specified (e.g., number and type of animals).
- Clinical Information: "multi-central clinical trial which included VT subjects." The exact number of subjects is not specified. Data provenance (country of origin, retrospective/prospective) explicitly is not specified, though "multi-central" implies a prospective clinical trial.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not provided for any of the studies mentioned.
4. Adjudication method for the test set
This information is not provided for any of the studies mentioned.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
An MRMC study is not applicable to this device. The ViaOne Epicardial Access System is an access tool, not an AI-powered diagnostic device.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
A standalone performance study of an algorithm is not applicable to this device as it is a physical medical device, not an AI algorithm.
7. The type of ground truth used
- Biocompatibility: Established by ISO 10993-1 standards.
- Sterilization, Packaging and Shelf Life Testing: Established by ISO 11135-1 standards and internal performance criteria.
- Performance Testing: Established by internal design specifications and ISO 11070:2014 standards.
- Animal Study: Ground truth was based on observation of device performance, malfunction, and Histopathology evaluation and pathological examination.
- Clinical Information: Ground truth for adverse events would be clinical observation and diagnosis by medical professionals. Ground truth for pericardial space access success would be confirmed during the procedure.
8. The sample size for the training set
This information is not applicable as this is a physical medical device, not an AI model requiring a training set.
9. How the ground truth for the training set was established
This information is not applicable as this is a physical medical device.
Summary of available and missing information:
Information Category | Status in Document |
---|---|
1. Table of Acceptance Criteria and Reported Device Performance | Missing (Qualitative statements only) |
2. Sample size for test set | Missing (Not specified for any study type) |
2. Data Provenance (country, retrospective/prospective) | "multi-central clinical trial" implies prospective; country not specified for clinical or animal studies |
3. Number of experts & qualifications for ground truth | Missing |
4. Adjudication method | Missing |
5. MRMC comparative effectiveness study? Effect size? | Not Applicable (Not an AI device) |
6. Standalone (algorithm only) performance? | Not Applicable (Not an AI device) |
7. Type of Ground Truth Used | ISO standards, pathological examination (animal), clinical observation (human trials) |
8. Sample size for training set | Not Applicable (Not an AI device) |
9. How ground truth for training set was established | Not Applicable (Not an AI device) |
Ask a specific question about this device
(30 days)
, Minnesota 55311
Re: K250468
Trade/Device Name: iSLEEVE Introducer Set Regulation Number: 21 CFR 870.1340
| |
| Classification | Class II, 21 CFR Part 870.1340
The iSLEEVE™ Introducer Set is intended to facilitate femoral access to the vascular system for introduction and removal of cardiovascular devices. The iSLEEVE™ Introducer Set is suitable for use in vessels > 5.5 mm diameter.
The iSLEEVE Expandable Introducer Set (iSLEEVE Introducer Set) is a sterile, single-use introducer catheter that provides percutaneous access to the femoral artery for introduction and removal of cardiovascular devices into the vascular system. The iSLEEVE Introducer Set is composed of a dilator and an introducer sheath with a three-way stopcock.
The distal end of the introducer sheath is passively expandable which allows for transient sheath expansion when the compatible valve system is introduced through it. Once expanded, this region of the introducer sheath is radially compliant, which allows it to expand and collapse as needed during device delivery and removal. This reduces the time the access vessel is expanded. A hydrophilic coating is applied to the working end of the iSLEEVE introducer sheath which increases the lubricity to aid in delivery when activated.
A product mandrel and flushing tube are also supplied with the iSLEEVE Introducer Set.
The provided text is related to a 510(k) premarket notification for a medical device called the "iSLEEVE Introducer Set." It describes the device, its intended use, and the regulatory process. However, it does not contain a detailed study with acceptance criteria and reported device performance in the format requested. The document focuses on regulatory approval based on substantial equivalence to a predicate device and mentions general categories of non-clinical performance data.
Therefore, I cannot provide the specific information requested in the format of a table of acceptance criteria and reported device performance, nor details about sample sizes for test sets, data provenance, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, or training set details as these are not present in the provided text.
The text does state:
"Testing demonstrated that the modified iSLEEVE Introducer Set met the previously established acceptance criteria."
And lists the types of non-clinical testing performed:
- Luer Connections
- Simulated Use Testing
- Dimensional Requirements
- Tensile Testing
- Packaging Testing
- Radiopacity
- Sterilization
- Usability
However, it does not provide the specific acceptance criteria values or the reported performance data against those criteria. It also doesn't contain the detailed study design elements you asked for, such as the number of experts, adjudication methods, or specific study types.
Ask a specific question about this device
Page 1 of 46