K Number
K242104

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-02-13

(210 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Penumbra Reperfusion Catheters and Separators

As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.

Penumbra 3D Revascularization Device

As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.

Penumbra Aspiration Tubing

As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump.

Penumbra Aspiration Pump

The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

Device Description

The Penumbra System is comprised of the following devices:

  • Penumbra Reperfusion Catheter
  • . Penumbra 3D Revascularization Device
  • Penumbra Aspiration Pump
  • . Penumbra Aspiration Pump Canister/Tubing
  • Penumbra Aspiration Tubing
  • Penumbra Separator .

The Penumbra System is designed to remove thrombus and restore blood flow in the neurovasculature using aspiration. The Reperfusion Catheter delivers aspiration from the pump directly to the site of occlusion to remove the clot. The 3D Revascularization Device is used with Reperfusion Catheters to facilitate aspiration and removal of the thrombus when needed. Alternatively, a Separator may be used to clear the lumen of the Reperfusion Catheter should it become blocked with thrombus. The Reperfusion Catheter is introduced through a guide catheter or long femoral sheath and into the intracranial vasculature and guided over a neurovascular guidewire to the site of the primary occlusion. The Penumbra Reperfusion Catheter is used with the Penumbra Aspiration Pump to aspirate thrombus from an occluded vessel. The Penumbra Reperfusion Catheter is connected to the Penumbra Aspiration Pump using the Penumbra Aspiration Tubing and the Penumbra Aspiration Pump Canister. The Penumbra Reperfusion Catheter is provided with a steam shaping mandrel, rotating hemostasis valve (RHV), peelable sheath, and optionally, SENDit Technology (cleared under K191946). The Penumbra 3D Revascularization Device is provided with an introducer sheath. The Penumbra Separator is provided with an introducer and torque device. The devices are visible under fluoroscopy.

AI/ML Overview

The provided text describes specific acceptance criteria and performance data for the Penumbra System (Reperfusion Catheter RED 72). However, it notes that no animal or clinical studies were conducted for this particular device submission (K242104). The evidence for meeting acceptance criteria comes solely from bench and biocompatibility testing. Therefore, information regarding sample sizes for test sets, data provenance, number and qualifications of experts, adjudication methods, MRMC studies, standalone performance, and training set details, which are typically associated with clinical or AI/algorithm performance studies, are not applicable or available in this document.

Here's a summary of the available information:

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance CriteriaReported Device Performance
Dimensional/Visual TestUnits meet all dimensional and visual product specifications.Acceptance Criteria Met
Friction TestUnits meet product specification related to friction.Acceptance Criteria Met
Radiopacity TestMarkerband is fluoroscopically visible.Acceptance Criteria Met
Simulated Use TestFunctionality of units using clinically relevant benchtop model.Acceptance Criteria Met
Particulate and Coating Integrity TestParticulates generated during simulated use and coating integrity after simulated use (including multiple deployment cycles) were evaluated.Acceptance Criteria Met
Hub/Air TestUnits have no leaks when tested.Acceptance Criteria Met
Tensile TestUnits meet product specification related to tensile strength.Acceptance Criteria Met
Pressure TestUnits meet product specification related to pressure.Acceptance Criteria Met
Elongation TestUnits meet product specification related to elongation.Acceptance Criteria Met
Corrosion Resistance TestNo visible corrosion on the units when tested.Acceptance Criteria Met
Torque Strength TestUnits have sufficient torque strength.Acceptance Criteria Met
Burst Pressure TestUnits can withstand sufficient pressure.Acceptance Criteria Met
Distal Tip Stiffness TestUnits have distal tip stiffness comparable to the reference device.Acceptance Criteria Met
Kink Resistance TestUnits meet product specification related to kink resistance.Acceptance Criteria Met
Simulated Use Physician Usability TestUsability of units in clinically relevant benchtop model.Acceptance Criteria Met
Shelf LifeExpiration date based on accelerated aging test studies.Acceptance Criteria Met
Packaging ValidationPackaging of the units meets all product specifications.Acceptance Criteria Met
Sterilization TestUnits are sterilized in accordance with ISO 11135+A1 and ISO 10993-7.Acceptance Criteria Met
Biocompatibility Tests
Cytotoxicity: MEM ElutionSample extracts must have a cytotoxic reactivity score of Grade 2 or lower.Pass (Non-cytotoxic)
Sensitization: Magnusson-Kligman MethodTest group shall yield Grade < 1 score on Magnusson and Kligman scale (provided control group yields Grade < 1).Pass (Non-sensitizing)
Irritation: Intracutaneous ReactivityThe difference between the average scores for the extract of the test article and the control is ≤ 1.0.Pass (Non-irritating)
Systemic Toxicity: Acute Systemic InjectionSample extracts must not cause significant biological reaction greater than control. That is: Death in 2 or more animals. Signs of toxicity in 2 or more animals (i.e. convulsions, prostration). Weight loss > 10% in 3 or more animals.Pass (Non-toxic)
Systemic Toxicity: Material-Mediated PyrogenSample extracts must not cause a total rise in body temperature of ≥ 0.5 °C.Pass (Non-pyrogenic)
Hemocompatibility: In-vitro ThrombogenicityDevice must not be thrombogenic in vitro when compared to a predicate device.Pass (Comparable to the predicate)
Hemocompatibility: Partial Thromboplastin Time (PTT)Clotting times of test article must be similar to predicate values.Pass (Comparable to the predicate)
Hemocompatibility: Complement ActivationThe concentration of SC5b-9 of test article must be similar to predicate values.Pass (Non-activator of the complement system)
Hemocompatibility: Hemolysis (indirect contact)Sample extracts must be non-hemolytic (≤ 2% hemolytic index).Pass (Non-hemolytic)
Hemocompatibility: Hemolysis (direct contact)Sample must be non-hemolytic (≤ 2% hemolytic index).Pass (Non-hemolytic)
Hemocompatibility: Platelet and Leukocyte CountSample must meet the requirements of ASTM F2888-19, Standard Practice for Platelet Leukocyte Count.Pass (Similar to the predicate)

2. Sample size used for the test set and the data provenance: Not applicable. The document states that "No animal or clinical study was conducted as bench and biocompatibility testing were determined sufficient for verification and validation purposes." The tests listed are primarily bench (in-vitro) tests on manufactured units or material samples. The specific number of units or samples tested for each bench test is not provided in this summary.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. The "ground truth" for these engineering and biocompatibility tests is based on established scientific principles, industry standards, and specified material properties, not expert consensus on medical images or diagnoses.

4. Adjudication method for the test set: Not applicable. Adjudication methods are typically employed in clinical studies or studies involving human assessment of data, which were not performed in this case.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No MRMC study was performed as this device is a physical medical device (catheter system) and not an AI-powered diagnostic or assistive tool for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used:
* Bench Testing: Engineering specifications, material properties, and functionality performance standards.
* Biocompatibility Testing: Established biological safety standards (ISO 10993 series, USP standards, 21 CFR Part 58 GLP).

8. The sample size for the training set: Not applicable. This is a physical medical device, not an algorithm that requires a training set.

9. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

February 13, 2025

Penumbra, Inc. Sindokht (Sisi) Soltanzadeh Sr. Regulatory Affairs Specialist One Penumbra Place Alameda, California 94502

Re: K242104

Trade/Device Name: Penumbra System (Reperfusion Catheter RED 72) Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NRY Dated: January 8, 2025 Received: January 10, 2025

Dear Sindokht (Sisi) Soltanzadeh:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Naira Muradyan -S

Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K242104

Device Name

Penumbra System (Reperfusion Catheter RED 72)

Indications for Use (Describe)

Penumbra Reperfusion Catheters and Separators

As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.

Penumbra 3D Revascularization Device

As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.

Penumbra Aspiration Tubing

As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump.

Penumbra Aspiration Pump

The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "Penumbra" in a red font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern. The font is sans-serif and easy to read.

K242104

1 510(k) Summary

(as required by 21 CFR 807.92)

Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Penumbra, Inc. is providing the summary of Substantial Equivalence for the subject Penumbra System (Reperfusion Catheter RED 72).

1.1 Submitter

Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA

Contact Person: Sindokht (Sisi) Soltanzadeh Sr. Regulatory Affairs Specialist Tel: (510) 995-9792 E-mail: ssoltanzadeh(@penumbrainc.com

Date of Preparation:

February 10, 2025

1.2 Subject Device

Penumbra System (Reperfusion Catheter RED 72)

Regulatory Class: II Classification Panel: Neurology Classification Name: Percutaneous Catheter Regulation Number: 21 CFR 870.1250 Product Code: NRY

Predicate and Reference Devices 1.3

510(k) NumberName of DeviceName of Manufacturer
Predicate Device
K211654Penumbra System (Reperfusion Catheter RED 72)Penumbra, Inc.
Reference Device
K173761Penumbra System (Reperfusion Catheter JET 7)Penumbra, Inc.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Penumbra logo. The word "Penumbra" is written in red, with a stylized red circle and a white line through it to the right of the word. The logo is simple and modern.

1.4 Device Description

The Penumbra System is comprised of the following devices:

  • Penumbra Reperfusion Catheter ●
  • . Penumbra 3D Revascularization Device
  • Penumbra Aspiration Pump ●
  • . Penumbra Aspiration Pump Canister/Tubing
  • Penumbra Aspiration Tubing ●
  • Penumbra Separator .

The Penumbra System is designed to remove thrombus and restore blood flow in the neurovasculature using aspiration. The Reperfusion Catheter delivers aspiration from the pump directly to the site of occlusion to remove the clot. The 3D Revascularization Device is used with Reperfusion Catheters to facilitate aspiration and removal of the thrombus when needed. Alternatively, a Separator may be used to clear the lumen of the Reperfusion Catheter should it become blocked with thrombus. The Reperfusion Catheter is introduced through a guide catheter or long femoral sheath and into the intracranial vasculature and guided over a neurovascular guidewire to the site of the primary occlusion. The Penumbra Reperfusion Catheter is used with the Penumbra Aspiration Pump to aspirate thrombus from an occluded vessel. The Penumbra Reperfusion Catheter is connected to the Penumbra Aspiration Pump using the Penumbra Aspiration Tubing and the Penumbra Aspiration Pump Canister. The Penumbra Reperfusion Catheter is provided with a steam shaping mandrel, rotating hemostasis valve (RHV), peelable sheath, and optionally, SENDit Technology (cleared under K191946). The Penumbra 3D Revascularization Device is provided with an introducer sheath. The Penumbra Separator is provided with an introducer and torque device. The devices are visible under fluoroscopy.

Indications For Use 1.5

Penumbra Reperfusion Catheters and Separators

As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral -M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.

Penumbra 3D Revascularization Device

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the word "Penumbra" in a red sans-serif font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern.

As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral -M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.

Penumbra Aspiration Tubing

As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump.

Penumbra Aspiration Pump

The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

DeviceAttributePredicate DeviceSubject Device
Trade NamePenumbra System(Reperfusion Catheter RED 72)Penumbra System(Reperfusion Catheter RED 72)
FDA ProductClassificationClass II, NRY, 21 CFR 870.1250SAME
510(k)NumberK211654K242104
Indicationsfor UsePenumbra Reperfusion Catheters andSeparatorsAs part of the Penumbra System, theReperfusion Catheters and Separators areindicated for use in the revascularization ofpatients with acute ischemic strokesecondary to intracranial large vesselocclusive disease (within the internalcarotid, middle cerebral – M1 and M2segments, basilar, and vertebral arteries)within 8 hours of symptom onset. Patientswho are ineligible for intravenous tissueplasminogen activator (IV t-PA) or who failIV t-PA therapy are candidates fortreatment.Penumbra 3D Revascularization DeviceAs part of the Penumbra System, thePenumbra 3D Revascularization Device isindicated for use in the revascularization ofPenumbra ReperfusionCatheters and SeparatorsAs part of the PenumbraSystem, the ReperfusionCatheters and Separators areindicated for use in therevascularization of patientswith acute ischemic strokesecondary to intracranial largevessel occlusive disease (withinthe internal carotid, middlecerebral – M1 and M2segments, basilar, and vertebralarteries) within 8 hours ofsymptom onset. Patients whoare ineligible for thrombolyticdrug therapy or who failedthrombolytic drug therapy arecandidates for treatment.
DeviceAttributePredicate DeviceSubject Device
Trade NamePenumbra System(Reperfusion Catheter RED 72)Penumbra System(Reperfusion Catheter RED 72)
patients with acute ischemic strokesecondary to intracranial large vesselocclusive disease (within the internalcarotid, middle cerebral – M1 and M2segments) within 8 hours of symptomonset. Patients who are ineligible forintravenous tissue plasminogen activator(IV t-PA) or who fail IV t-PA therapy arecandidates for treatment.Penumbra Aspiration TubingAs part of the Penumbra System, thePenumbra Sterile Aspiration Tubing isindicated to connect the PenumbraReperfusion Catheters to the PenumbraAspiration Pump.Penumbra Aspiration PumpThe Penumbra Aspiration Pump isindicated as a vacuum source for PenumbraAspiration Systems.Penumbra 3DRevascularization DeviceAs part of the PenumbraSystem, the Penumbra 3DRevascularization Device isindicated for use in therevascularization of patientswith acute ischemic strokesecondary to intracranial largevessel occlusive disease (withinthe internal carotid, middlecerebral - M1 and M2segments) within 8 hours ofsymptom onset. Patients whoare ineligible for thrombolyticdrug therapy or who failedthrombolytic drug therapy arecandidates for treatment.Penumbra Aspiration TubingAs part of the PenumbraSystem, the Penumbra SterileAspiration Tubing is indicatedto connect the PenumbraReperfusion Catheters to thePenumbra Aspiration Pump.Penumbra Aspiration PumpThe Penumbra AspirationPump is indicated as a vacuumsource for Penumbra AspirationSystems.
Principles ofOperationSee Section 1.4SAME
DeviceMaterialsStainless Steel, PTFE, Polyurethane,Polyether Block Amide, Nylon 12, Nitinol,Platinum/IridiumStainless Steel, PTFE,Polyurethane, differing PolyetherBlock Amide, Nylon 12, Nitinol,Platinum/Iridium
ID BandColorPolyolefin, PET black [white foil]SAME
CoatingHydrophilic coating (proprietary)SAME
Min. ID0.072 in. (1.83 mm)SAME
Max. OD0.085 in. (2.16 mm)SAME
DeviceAttributePredicate DeviceSubject Device
Trade NamePenumbra System(Reperfusion Catheter RED 72)Penumbra System(Reperfusion Catheter RED 72)
Distal FlexLength30 cmSAME
CoatingLength30 cmSAME
EffectiveLengths115, 120, 125, 127, 132 cmSAME
AccessoriesPeelable Sheath, Shaping Mandrel, RHVPeelable Sheath, ShapingMandrel, RHV, SENDitTechnology (Optional, K191946)
PackagingMaterialsPolyester/Polyethylene/Tyvek, Polystyrene,SBS PaperboardSAME
ConditionSuppliedSterile and Single UseSAME
SterilizationMethodEthylene Oxide (EO)SAME

Comparison of Technological Characteristics with the Predicate Device 1.6

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "Penumbra" in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The "P" is stylized with a horizontal line running through it.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "Penumbra" in red font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern. The font is sans-serif.

Performance Data 1.7

The following performance data were provided in support of the substantial equivalence determination:

  • Design Verification .
  • Biocompatibility ●
  • Shelf Life .
  • Packaging Validation o
  • Sterilization

The subject device met all established requirements.

Design Verification Testing 1.7.1

The following design verification tests were performed on the subject device:

TestTest Method SummaryConclusion
Dimensional/Visual TestConfirms the units meet all dimensional and visual productspecifications.Acceptance Criteria Met
Friction TestConfirms units meet product specification related tofriction.Acceptance Criteria Met
Radiopacity TestConfirms the markerband is fluoroscopically visible.Acceptance Criteria Met
Simulated Use TestConfirms the functionality of units using clinicallyrelevant benchtop model.Acceptance Criteria Met
Particulate and CoatingIntegrity TestParticulates generated during simulated use and coatingintegrity after simulated use (including multiple deploymentcycles) were evaluated.Acceptance Criteria Met
Hub/Air TestConfirms units have no leaks when tested.Acceptance Criteria Met

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Penumbra, Inc., a global healthcare company focused on innovative therapies. The word "Penumbra" is written in a red sans-serif font. To the right of the word is a red circle with a white stylized "P" inside, which is the company's symbol.

TestTest Method SummaryConclusion
Tensile TestConfirms units meet product specification related to tensilestrength.Acceptance Criteria Met
Pressure TestConfirms units meet product specification related topressure.Acceptance Criteria Met
Elongation TestConfirms units meet product specification related toelongation.Acceptance Criteria Met
Corrosion ResistanceTestConfirms there is no visible corrosion on the units whentested.Acceptance Criteria Met
Torque Strength TestConfirms units have sufficient torque strength.Acceptance Criteria Met
Burst Pressure TestConfirms units can withstand sufficient pressure.Acceptance Criteria Met
Distal Tip Stiffness TestConfirms units have distal tip stiffness comparable to thereference device.Acceptance Criteria Met
Kink Resistance TestConfirms units meet product specification related to kinkresistance.Acceptance Criteria Met
Simulated Use PhysicianUsability TestConfirms the usability of units in clinically relevant benchtopmodel.Acceptance Criteria Met
Shelf LifeConfirms expiration date based on accelerated aging teststudies.Acceptance Criteria Met
Packaging ValidationConfirms the packaging of the units meets all productspecifications.Acceptance Criteria Met
Sterilization TestConfirms the units are sterilized in accordance with ISO11135+A1 and ISO 10993-7.Acceptance Criteria Met

1.7.2 Biocompatibility

The biocompatibility evaluation for the subject device was conducted in accordance with ISO 10993-1, USP standards, and 21 CFR, Part 58, Good Laboratory Practices (GLP). The battery of testing included the following tests:

TestsAcceptance CriteriaConclusion
Cytotoxicity:MEM ElutionSample extracts must have a cytotoxic reactivityscore of Grade 2 or lower.PassNon-cytotoxic
Sensitization:Magnusson-Kligman MethodTest group shall yield Grade $< 1$ score onMagnusson and Kligman scale (provided controlgroup yields Grade $< 1$ ).PassNon-sensitizing
Irritation:Intracutaneous ReactivityThe difference between the average scores forthe extract of the test article and the control is $\leq$1.0.PassNon-irritating
Systemic Toxicity:Acute Systemic InjectionSample extracts must not cause significantbiological reaction greater than control. That is:Death in 2 or more animals. Signs of toxicity in 2 or more animals (i.e.convulsions, prostration). Weight loss $> 10%$ in 3 or more animals.PassNon-toxic
Systemic Toxicity: Material-Mediated PyrogenSample extracts must not cause a total rise inbody temperature of $\geq 0.5$ °C.PassNon-pyrogenic

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Penumbra, Inc., a global healthcare company focused on innovative therapies. The logo consists of the word "Penumbra" in a bold, red font, followed by a stylized "P" symbol enclosed in a red circle. The "P" symbol is formed by a white line that creates the shape of the letter within the circle. The logo is clean and modern, reflecting the company's focus on innovation and technology in the healthcare industry.

TestsAcceptance CriteriaConclusion
Hemocompatibility:In-vitro ThrombogenicityDevice must not be thrombogenic in vitro whencompared to a predicate device.PassComparable to thepredicate
Hemocompatibility: PartialThromboplastin Time (PTT)Clotting times of test article must be similar topredicate values.PassComparable to thepredicate
Hemocompatibility:Complement ActivationThe concentration of SC5b-9 of test article mustbe similar to predicate values.PassNon-activator of thecomplement system
Hemocompatibility:Hemolysis (indirect contact)Sample extracts must be non-hemolytic (≤ 2%hemolytic index).PassNon-hemolytic
Hemocompatibility:Hemolysis (direct contact)Sample must be non-hemolytic (≤ 2% hemolyticindex).PassNon-hemolytic
Hemocompatibility: Plateletand Leukocyte CountSample must meet the requirements of ASTMF2888-19, Standard Practice for PlateletLeukocyte Count.PassSimilar to thepredicate

Biocompatibility test results demonstrate biological safety per ISO 10993 and USP requirements.

Performance Data - Animal, Clinical 1.7.3

No animal or clinical study was conducted as bench and biocompatibility testing were determined sufficient for verification and validation purposes.

1.8 Conclusions

The subject Penumbra System (Reperfusion Catheter RED 72) is substantially equivalent to the predicate Penumbra System (Reperfusion Catheter RED 72). The subject device has the identical intended use and similar operating principle, design concept, fundamental technology as the predicate. The differences between the subject and predicate devices do not raise new or different questions of safety or effectiveness. The device testing described in this 510(k) Summary demonstrates the subject device performs as intended.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).