K Number
K242319

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-09-03

(29 days)

Product Code
Regulation Number
870.5150
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

INDIGO Aspiration Catheters and Separators: As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems, and for the treatment of pulmonary embolism.

INDIGO Aspiration Tubing: As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump.

Penumbra Aspiration Pump: The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

Device Description

The INDIGO® Aspiration System is comprised of the several devices: INDIGO Aspiration Catheter, Penumbra Aspiration Pump, INDIGO Aspiration Pump Canister, INDIGO Aspiration Tubing, INDIGO Separator. The INDIGO Aspiration System is designed to remove thrombus from the vasculature using mechanical aspiration. The INDIGO Aspiration Catheter targets aspiration from the pump directly to the thrombus. The INDIGO Separator may be used to clear the lumen of the INDIGO Aspiration Catheter should it become blocked with thrombus. The INDIGO Aspiration Catheter is introduced through a guide catheter or vascular sheath into the peripheral vasculature and guided over a guidewire to the site of the primary occlusion. TraX may be used with its prepackaged Aspiration Catheter to facilitate access to the primary occlusion. The INDIGO Aspiration Catheter is used with the Penumbra Aspiration Pump to aspirate thrombus from an occluded vessel. As needed, an INDIGO Separator may be deployed from the INDIGO Aspiration Catheter to assist with thrombus removal. The INDIGO Separator is advanced and retracted through the INDIGO Aspiration Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the INDIGO Aspiration Catheter tip. The devices are visible under fluoroscopy. For the aspiration source, the INDIGO Aspiration Catheter is used in conjunction with the Penumbra Aspiration Pump, which is connected using the INDIGO Aspiration Tubing and the INDIGO Aspiration Pump Canister. The INDIGO Aspiration Catheter may be provided with a steam shaping mandrel, rotating hemostasis valve, Select Catheter, introducer, and TraX. TraX compatibility is limited to its prepackaged Aspiration Catheter. The INDIGO Separator may be provided with an introducer and torque device.

AI/ML Overview

The provided document is a 510(k) Premarket Notification from the FDA for the Penumbra, Inc. Indigo® Aspiration System - Aspiration Catheter 6X. It details the device, its intended use, and its substantial equivalence to predicate devices based on non-clinical (bench-top) performance testing.

1. A table of acceptance criteria and the reported device performance

The document does not provide a specific table of acceptance criteria with corresponding performance metrics like sensitivity, specificity, or AUC, as it is related to a physical medical device (catheter) rather than a diagnostic algorithm or AI. Instead, it lists various bench-top performance tests conducted. For each test, the implicit acceptance criterion is that the device met all requirements.

Test CategoryReported Device Performance (Acceptance Criteria)
Particulate TestingMet all requirements
Dimensional/Visual InspectionMet all requirements
Friction TestingMet all requirements
Performance/Simulated Use TestingMet all requirements
Markerband VisibilityMet all requirements
Tensile TestingMet all requirements
Pressure TestingMet all requirements
Elongation TestingMet all requirements
Torsion TestingMet all requirements
Hub Air AspirationMet all requirements
Corrosion TestingMet all requirements
Biocompatibility (ISO 10993-1)Met all requirements
Shelf-LifeEstablished for 12 months based on accelerated aging data

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document primarily describes bench-top testing for a physical device. It does not mention a "test set" in the context of data used for an algorithm or AI. Therefore, concepts like sample size for a test set, country of origin, retrospective or prospective data are not applicable. The data provenance would be from internal laboratory testing conducted by Penumbra, Inc.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable as the document describes bench-top performance testing of a physical medical device (catheter), not a diagnostic algorithm or AI that would require expert-established ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable for the same reason as point 3.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

An MRMC comparative effectiveness study was not done. The study described is for a physical medical device, not an AI or diagnostic tool that would involve human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

A standalone performance study in the context of an algorithm or AI was not done. The performance evaluation was of the physical device's mechanical, material, and functional properties on a test bench.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the bench-top tests, the "ground truth" would be defined by engineering specifications, material science standards, and functional performance benchmarks for the device. For example, a tensile test would measure the force required to break the catheter, with the "ground truth" being the specified minimum tensile strength that the device must meet according to its design requirements. Similarly, biocompatibility is assessed against ISO 10993-1 standards.

8. The sample size for the training set

This information is not applicable as the document describes bench-top performance testing of a physical medical device, not an AI or algorithm that uses a training set.

9. How the ground truth for the training set was established

This information is not applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

September 3, 2024

Penumbra, Inc. Samyukta Rangachari Regulatory Affairs Specialist III One Penumbra Place Alameda, California 94502

Re: K242319

Trade/Device Name: Indigo® Aspiration System - Aspiration Catheter 6X Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: QEW Dated: August 2, 2024 Received: August 5, 2024

Dear Samyukta Rangachari:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ariel G. Ash-Digitally signed by Ariel G.Ash-shakoor -S
shakoor -SDate: 2024.09.03 14:20:05-04'00'

{2}------------------------------------------------

For

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K242319

Device Name

Indigo® Aspiration System - Aspiration Catheter 6X

Indications for Use (Describe)

INDIGO Aspiration Catheters and Separators

As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems, and for the treatment of pulmonary embolism.

INDIGO Aspiration Tubing

As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump.

Penumbra Aspiration Pump

The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours be response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "Penumbra" in red font, followed by a red circle with a white "P" inside. The word "Penumbra" is written in a sans-serif font and is the main focus of the image. The red circle with the white "P" is located to the right of the word "Penumbra" and serves as a logo for the company.

K242319 Page 1 of 7

510(k) Summary

1. Submitter

Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA

Contact Person: Samyukta Rangachari Regulatory Affairs Specialist III Phone: (763) 742-9306 Email: srangachari@penumbrainc.com

Date of Preparation: August 29, 2024

2. Subject Device

Indigo® Aspiration System – Aspiration Catheter 6X

Regulatory Class: ll Classification Panel: Cardiovascular Classification Name: Catheter, Embolectomy Regulation Number: 21 CFR §870.5150 Product Code: QEW

3. Predicate/Reference Devices

510(k) NumberName of DeviceSystem NameINDIGO Aspiration SystemPenumbra SystemINDIGO Aspiration System
PredicateDevice NameIndigo Aspiration Catheter 5[Predicate]Reperfusion Catheter RED 72 [Reference]Indigo Aspiration Catheter6X [Subject]
K192833Indigo Aspiration System – Aspiration Catheter 5510(k) No.K192833K211654K242319
ReferenceClassificationClass II, QEWClass II, NRYSAME as Predicate
K211654Penumbra System (Reperfusion Catheter RED 72)IndicationINDIGO Aspiration Catheters andSeparators:As part of the INDIGO AspirationSystem, the INDIGO AspirationCatheters and Separators are indicatedfor the removal of fresh, soft emboliand thrombi from vessels of theperipheral arterial and venoussystems, and for the treatment ofpulmonary embolism.INDIGO Aspiration Tubing:As part of the INDIGO AspirationSystem, the INDIGO Sterile AspirationTubing is indicated to connect theINDIGO Aspiration Catheters to thePenumbra Aspiration Pump.Penumbra Aspiration Pump:The Penumbra Aspiration Pump isindicated as a vacuum source for thePenumbra Aspiration Systems.Penumbra Reperfusion Catheters andSeparators:As part of the Penumbra System, theReperfusion Catheters and Separators areindicated for use in the revascularization ofpatients with acute ischemic stroke secondaryto intracranial large vessel occlusive disease(within the internal carotid, middle cerebral –M1 and M2 segments, basilar, and vertebralarteries) within 8 hours of symptom onset.Patients who are ineligible for intravenoustissue plasminogen activator (IV t-PA) or whofail IV t-PA therapy are candidates fortreatment.Penumbra 3D Revascularization Device:As part of the Penumbra System, the Penumbra3D Revascularization Device is indicated for usein the revascularization of patients with acuteischemic stroke secondary to intracranial largevessel occlusive disease (within the internalcarotid, middle cerebral – M1 and M2segments) within 8 hours of symptom onset.SAME as Predicate
K191946Penumbra System – MAX Delivery Device

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the word "Penumbra" in a bold, red font. To the right of the word is a circular logo, also in red, with a white "P" inside. The logo has a white line extending from the left side of the "P" to the edge of the circle, creating a visual effect of light or shadow. The logo is a registered trademark.

K242319

4. Predicate Comparison

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the Penumbra company logo. The word "Penumbra" is written in a red serif font. To the right of the word is a red circle with a white "P" inside of it.

K242319

Page 3 of 7
System NameINDIGO Aspiration SystemPenumbra SystemINDIGO Aspiration System
Device NameIndigo Aspiration Catheter 5[Predicate]Reperfusion Catheter RED 72 [Reference]Indigo Aspiration Catheter6X [Subject]
Patients who are ineligible for intravenoustissue plasminogen activator (IV t-PA) or whofail IV t-PA therapy are candidates fortreatment.Penumbra Aspiration Tubing:As part of the Penumbra System, the PenumbraSterile Aspiration Tubing is indicated to connectthe Penumbra Reperfusion Catheters to thePenumbra Aspiration Pump.Penumbra Aspiration Pump:The Penumbra Aspiration Pump is indicated as avacuum source for Penumbra AspirationSystems.
CatheterAspiration Catheter 5Reperfusion Catheter RED 72Aspiration Catheter 6X
MaterialsBiocompatible, commonly utilized forinterventional devicesBiocompatible, commonly utilized forinterventional devicesSAME
CoatingHydrophilic (proprietary)Hydrophilic (proprietary)SAME as Reference
Proximal andDistal OD
Proximal andDistal IDEffectiveLengthsAppropriately sized for the targetvesselSAME, with larger diameters and additionaleffective length offeringsSAME as Reference
System NameINDIGO Aspiration SystemPenumbra SystemINDIGO Aspiration System
Device NameIndigo Aspiration Catheter 5 [Predicate]Reperfusion Catheter RED 72 [Reference]Indigo Aspiration Catheter 6X [Subject]
AccessoriesIntroducer, Shaping Mandrel, RHVIntroducer, Shaping Mandrel, RHV, SENDit Technology (Optional)Introducer, Shaping Mandrel, RHV, TraX (Optional)
Aspiration TubingModified 110 Aspiration Tubing (K180939)Modified 110 Aspiration TubingSAME
Packaging MaterialsCommonly used materials for medical devicesCommonly used materials for medical devicesSAME
Aspiration SourceAspiration PumpAspiration PumpSAME
SterilizationEOEOSAME
Shelf-Life36 months12 monthsSAME as Reference
UseSingle use, disposableSingle use, disposableSAME

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "Penumbra" in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The "P" is stylized with a horizontal line extending from the right side of the "P" to the edge of the circle. The logo is simple and modern.

K242319

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "Penumbra" in red font, followed by a red circle with a white "P" inside. The "P" is stylized with a line extending from the top of the "P" to the edge of the circle. The logo is simple and modern.

5. Device Description

The INDIGO® Aspiration System is comprised of the several devices:

  • INDIGO Aspiration Catheter
  • Penumbra Aspiration Pump
  • · INDIGO Aspiration Pump Canister
  • INDIGO Aspiration Tubing
  • INDIGO Separator

The INDIGO Aspiration System is designed to remove thrombus from the vasculature using mechanical aspiration. The INDIGO Aspiration Catheter targets aspiration from the pump directly to the thrombus. The INDIGO Separator may be used to clear the lumen of the INDIGO Aspiration Catheter should it become blocked with thrombus. The INDIGO Aspiration Catheter is introduced through a guide catheter or vascular sheath into the peripheral vasculature and guided over a guidewire to the site of the primary occlusion. TraX may be used with its prepackaged Aspiration Catheter to facilitate access to the primary occlusion. The INDIGO Aspiration Catheter is used with the Penumbra Aspiration Pump to aspirate thrombus from an occluded vessel. As needed, an INDIGO Separator may be deployed from the INDIGO Aspiration Catheter to assist with thrombus removal. The INDIGO Separator is advanced and retracted through the INDIGO Aspiration Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the INDIGO Aspiration Catheter tip. The devices are visible under fluoroscopy. For the aspiration source, the INDIGO Aspiration Catheter is used in conjunction with the Penumbra Aspiration Pump, which is connected using the INDIGO Aspiration Tubing and the INDIGO Aspiration Pump Canister. The INDIGO Aspiration Catheter may be provided with a steam shaping mandrel, rotating hemostasis valve, Select Catheter, introducer, and TraX. TraX compatibility is limited to its prepackaged Aspiration Catheter. The INDIGO Separator may be provided with an introducer and torque device.

6. Indications for Use

INDIGO Aspiration Catheters and Separators

As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems, and for the treatment of pulmonary embolism.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the Penumbra company logo. The logo consists of the word "Penumbra" in a red, sans-serif font, followed by a red circle with a white "P" inside. A white line extends from the right side of the "a" in "Penumbra" to the right side of the circle, visually connecting the text and the symbol.

INDIGO Aspiration Tubing

As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump.

Penumbra Aspiration Pump

The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

7. Summary of Non-Clinical Data/Performance Data

Included in this section are summary descriptions of the testing which substantiates the performance of the subject Aspiration Catheter 6X.

7.1 Summary of Performance Testing - (Bench-Top) Testing

The following bench-top performance (design verification) tests were used in support of the subject device to determine substantial equivalence:

  • . Particulate Testing
  • Dimensional/Visual Inspection ●
  • Friction Testing ●
  • Performance/Simulated Use Testing
  • Markerband Visibility
  • Tensile Testing ●
  • Pressure Testing
  • Elongation Testing
  • Torsion Testing ●
  • Hub Air Aspiration
  • Corrosion Testing

7.2 Summary of Biocompatibility

Biocompatibility was evaluated for the subject device in accordance with ISO 10993-1 and the FDA Guidance for Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", for limited exposure (≤24 hours), externally communicating devices with circulating blood contact. The subject device met all requirements.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the Penumbra company logo. The logo consists of the word "Penumbra" in red font, followed by a red circle with a white "P" inside. The "P" is stylized with a horizontal line extending from the top of the "P" to the edge of the circle.

7.3 Summary of Shelf-Life

The shelf-life of the subject device is established for 12 months based on accelerated aging data.

7.4 Summary of Performance Data – Animal, Clinical

No animal or clinical study was conducted as bench testing was determined sufficient for verification and validation purposes.

8. Summary of Substantial Equivalence

The subject Aspiration Catheter 6X is substantially equivalent to the predicate device, Aspiration Catheter 5, with regards to intended use, operating principle, design concept, fundamental technology, and device performance.

§ 870.5150 Embolectomy catheter.

(a)
Identification. An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.(b)
Classification. Class II (performance standards).