Search Filters

Search Results

Found 418 results

510(k) Data Aggregation

    K Number
    K251455
    Date Cleared
    2025-07-24

    (73 days)

    Product Code
    Regulation Number
    892.1550
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Automated Radiological Image Processing Software | 892.2050 | QIH |
    | Diagnostic Intravascular Catheter | 870.1200
    N/A | No change |
    | Secondary Regulation Number | 21 CFR 892.1570
    21 CFR 892.1560
    21 CFR 870.1200

    21 CFR 892.2050 | 21 CFR 892.1570
    21 CFR 892.1560
    21 CFR 870.1200
    21 CFR 892.2050 | N/A
    N/A | No change |
    | Secondary Regulation Number | 21 CFR 892.1570
    21 CFR 892.1560
    21 CFR 870.1200

    21 CFR 892.2050 | 21 CFR 892.1570
    21 CFR 892.1560
    21 CFR 870.1200
    21 CFR 892.2050 | N/A

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EPIQ Series Diagnostic Ultrasound System: The intended use of EPIQ Ultrasound Diagnostic System is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use: Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), intra-luminal, intra-cardiac echo, Musculoskeletal (Conventional), Musculoskeletal (Superficial), Ophthalmic, Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.
    Modes of operation include: B Mode, M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode, Power Doppler, and Harmonic Imaging.
    The clinical environments where EPIQ Series Diagnostic ultrasound Systems can be used include clinics, hospitals, and clinical point-of-care for diagnosis of patients.
    When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live echo guidance.
    The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best clinical procedure.

    Affiniti Series Diagnostic Ultrasound Systems: The intended use of Affiniti Series Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use: Abdominal, Cardiac Adult, Cardiac Other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.
    Modes of operation include: B Mode, M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode, Power Doppler, and Harmonic Imaging.
    The clinical environments where the Affiniti diagnostic ultrasound systems can be used include clinics, hospitals, and clinical point-of-care for diagnosis of patients. The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best clinical procedure.

    Device Description

    The purpose of this 510(k) Pre-market Notification is to introduce the SVS v2 Contrast software application with the EPIQ and Affiniti Series Diagnostic Ultrasound Systems as part of the VM13.0 program.
    SVS v2 Contrast is an automated software feature that assists in the selection of images for analysis with the new Philips license options 2D Auto EF Advanced (Adv) developed by DiA Imaging Analysis (K243235), now part of Philips Ultrasound LLC, and existing Philips license options AutoStrain LV; or 2D Auto LV (both K240850) license application in Adult Echo Transthoracic (TTE) examinations.
    As described in 510(k) K240850, this feature automatically classifies each acquired image by view and selects an appropriate set of images for Left ventricle (LV) analysis. The classification is based on a Deep Learning AI inference engine; the selection is a non-AI algorithm that considers the view classification and image depth to select the optimal set of images. The difference in SVS v2 Contrast from the predicate SVS v1 (K240850) is that the algorithm is revised to include the selection of optimal images for analysis when contrast is used in routine TTE exams.
    Users can launch existing 2D Auto EF; 2D Auto EF Adv; AutoStrain LV; or 2D Auto LV with the set of images that have been automatically selected without the need to review the acquired images and manually select the views. Users may select 2D Auto EF Adv, to process the selected views by SVS v2 Contrast. SVS v2 Contrast prioritizes the contrast image pair; however, if an appropriate pair of contrast images is not found, then Auto EF Adv may select non-contrast images. The manual approach to select views is still available and the user can override automatically selected images from SVS v2 Contrast.
    No hardware changes to the EPIQ or Affiniti Series Diagnostic Ultrasound Systems are required when using SVS v2 Contrast, and existing, cleared Philips TTE transducers.
    The SVS v2 Contrast feature is supported by all EPIQ models running software version 13.0 or higher including EPIQ CVx/CVxi, EPIQ Elite Advanced, EPIQ Elite, EPIQ 7, EPIQ 5. The SVS v2 Contrast feature is supported by the Affiniti models running software version VM13.0 or higher, including Affiniti CVx, Affiniti 70, Affiniti 50, and Affiniti 30. The SVS v2 Contrast feature is associated with the cardiac adult indication.

    AI/ML Overview

    Acceptance Criteria and Study Details for SVS v2 Contrast Software

    This document outlines the acceptance criteria and details of the study conducted to demonstrate that the Philips Ultrasound systems (EPIQ and Affiniti Series Diagnostic Ultrasound System) with the SVS v2 Contrast software application meet the established performance benchmarks.

    1. Table of Acceptance Criteria and Reported Device Performance

    CriteriaAcceptance CriteriaReported Device Performance (SVS Selected vs Manually Selected Ground Truth)
    Primary Endpoint: Pearson's correlation coefficient (r) for Biplane EF (Automated Selection vs. Manual Selection)Lower Confidence Bound for the Pearson's correlation coefficient (r) to be > 0.8. (Informed by published literature (Maret, E. B. (2008)) and previous regulatory submissions (K130779, K232500)). Note: The manual selection here refers to "manual contrast clip selection," which was automatically processed within the Contrast EF application, with the automated EF results serving as ground truth.N = 46: r = 0.953 (95% CI: 0.917, 0.974), p-value 0.8, meeting the acceptance criteria.
    Agreement Assessment: Bland-Altman Analysis for Biplane EF (Automated Selection vs. Manual Selection)Acceptable performance for automatic selection of appropriate clips by SVS software and subsequent output of Contrast Enhanced Biplane EF feature compared to the results of Contrast Enhanced Biplane EF from manually selected contrast clips. (No specific numerical thresholds provided for Bland-Altman, but indicated as demonstrating "acceptable performance" if the primary endpoint is met).N = 46: Mean Difference ± SD = 0.71 ± 4.36 (95% CI for Mean Difference: -0.58, 2.01). Lower LoA = -7.83 (95% CI: -10.06, -5.60), Upper LoA = 9.26 (95% CI: 7.03, 11.49). Reported as demonstrating acceptable performance.
    Agreement Assessment: Pearson's correlation coefficient (r) for Biplane EF (Automated Selection vs. Manual Tracing Ground Truth)No explicit independent acceptance criterion for this comparison; serves as an additional supporting analysis.N = 46: r = 0.938 (95% CI: 0.890, 0.965), p-value
    Ask a Question

    Ask a specific question about this device

    K Number
    K250913
    Date Cleared
    2025-07-19

    (114 days)

    Product Code
    Regulation Number
    892.1550
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Name:** TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound System
    Regulation Number: 21 CFR 870.1200
    System
    Common name: Ultrasonic Pulsed Doppler Imaging System
    Regulation number: 21 CFR 870.1200
    Catheter(K092064)
    Common name: Catheter, ultrasound, intravascular
    Regulation number: 21 CFR 870.1200
    ---|-------------|----------|
    | Classification | Class II | Class II | Same |
    | Regulation | 21 CFR 870.1200
    | 21 CFR 870.1200 | Same |
    | Product code | OBJ | OBJ | Same |
    | Indications for use | TINGSN FINDERS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TINGS FINDERS 2 Color Doppler Diagnostic Ultrasound System is intended for intracardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult patients.

    Modes of operation include: B,M, Color Doppler,PW Doppler, CW Doppler,Power Doppler; Combined modes: B/M,B/Color,B/M/Color, B/PWD, B/CWD, B/Color/PWD, B/Color/CWD, B/Power.

    The device is intended for use in Professional healthcare facility environment including echo lab, other hospital settings, operating room, Cath lab and EP lab.

    The intended users of the Device is to be used exclusively by physicians who are appropriately trained personnel in a fully equipped electrophysiology or catheterization room or under their direct supervision.

    Device Description

    The proposed TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound System is a Track 3 diagnostic device designed for imaging intracardiac anatomical structures and blood flow dynamics in adult patients. It comprises two primary components: the ultrasound console and the TINGSN Sonic Eyes 10 Eco Disposable Intracardiac Echocardiography Catheter. The console features a 17.3-inch touch-enabled LCD screen, a control panel, and interfaces for connectivity, powered by a rechargeable battery or AC adapter, with optional peripherals including a cart and foot switch. The catheter, a sterile, single-use device with a 95 cm insertable length and 10Fr diameter, is equipped with a 64-element phased array transducer at its tip for high-resolution imaging. It connects to the console via the Finders 2 Sonic Eyes Connector, featuring a locking mechanism to ensure secure attachment during procedures. The system supports multiple imaging modes—B-Mode, M-Mode, Pulsed Wave (PW) Doppler, Continuous Wave (CW) Doppler, Color Doppler, Power Doppler, and combined modes —along with advanced functionalities, displayed with real-time thermal (TI) and mechanical (MI) indices.

    AI/ML Overview

    The provided text is a 510(k) Clearance Letter and 510(k) Summary for the TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound System. This document describes the device, its intended use, and a comparison to predicate devices to establish substantial equivalence.

    Crucially, the document explicitly states under "8. CLINICAL DATA": "No clinical data was included in this submission."

    Therefore, based on the provided text, a study that proves the device meets acceptance criteria through clinical data (either standalone or MRMC comparative effectiveness) was not performed, or at least, not included in this submission. The entire submission relies on non-clinical testing and comparison to predicate devices to establish substantial equivalence.

    As the request specifically asks for information about a "study that proves the device meets the acceptance criteria" and mentions clinical outcomes, it is impossible to provide the requested details from the given text.

    However, I can extract the information related to the non-clinical tests that were performed to verify the device met design specifications and was substantially equivalent, which serve as the basis for clearance in lieu of clinical data.

    Here's an interpretation based on the non-clinical data presented:


    Device Acceptance Criteria and Performance (Based on Non-Clinical Data)

    The TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound System established its substantial equivalence (SE) based primarily on non-clinical testing and comparison to predicate devices, as no clinical data was included in the submission. The acceptance criteria and "performance" are therefore framed in terms of compliance with relevant standards and demonstration of functional equivalence.

    1. Table of Acceptance Criteria and Reported Device Performance (Non-Clinical)

    Acceptance Criteria (Demonstrated via Compliance with Standards / Equivalence)Reported Device Performance (Demonstrated via Non-Clinical Testing)
    Safety & Essential Performance: Compliance with general medical electrical equipment safety standards. (IEC 60601-1 Ed 3.2)Met: Non-clinical tests demonstrated compliance with IEC 60601-1 Edition 3.2.
    Electromagnetic Compatibility (EMC): Compliance with EMC requirements for medical devices. (IEC 60601-1-2)Met: Non-clinical tests demonstrated compliance with IEC 60601-1-2.
    Specific Safety/Performance for Ultrasonic Equipment: Compliance with specific requirements for diagnostic ultrasound. (IEC 60601-2-37)Met: Non-clinical tests demonstrated compliance with IEC 60601-2-37.
    Acoustic Output Characterization: Compliance with methods for determining thermal and mechanical indices. (IEC 62359, NEMA UD2)Met: Non-clinical tests demonstrated compliance with IEC 62359 and NEMA UD2. The device is a "Track 3" device, implying its acoustic output is within regulatory limits (global maximum derated ISPTA ≤720 mW/cm², MI ≤1.9 or ISPPA ≤ 190 W/cm²).
    Intravascular Catheter Requirements: Compliance with standards for sterile, single-use intravascular catheters. (ISO 10555-1)Met: Non-clinical tests demonstrated compliance with ISO 10555-1.
    Biocompatibility: Evaluation and testing within a risk management process for medical devices. (ISO 10993-1)Met: Non-clinical tests demonstrated compliance with ISO 10993-1 (specifically, Patient contact materials of the catheter shaft are "Pebax series Comply with ISO 10993-1").
    Packaging & Shipping Performance: Performance testing of shipping containers and systems. (ASTM D4169-2022)Met: Non-clinical tests demonstrated compliance with ASTM D4169-2022.
    Risk Management: Application of risk management to medical devices. (ISO 14971:2019)Met: A thorough risk analysis was conducted, and risk controls were implemented to mitigate identified hazards.
    Software Functionality & Safety: Compliance with FDA guidance for device software functions, meeting specifications and mitigating risks. (FDA Guidance for Device Software Functions)Met: Software documentation included. Non-clinical tests (integration and functional) conducted. Test results support that all software specifications met acceptance criteria. Verification and validation testing found acceptable.
    Functional Equivalence to Predicate Device (GE Vivid iq console): Similar intended use, imaging modes, principle of operation, and overall performance.Demonstrated: The proposed device's indications for use are encompassed by the predicate. Technical differences in imaging modes, software, and compatible probes do not impact safety or effectiveness. Risk analysis assessed potential software-related risks.
    Functional Equivalence to Predicate Device (SOUNDSTAR 3D Ultrasound Catheter): Similar indications for use, principle of operation, patient population, sterility, diameter, and transducer configuration.Demonstrated: Slight differences in insertable length, imaging frequency range, and shelf life were determined not to impact safety or effectiveness. The proposed device's indications are encompassed by the predicate.
    Shelf Life: Verification of product stability and packaging integrity for claimed shelf life.Met: A 2-year shelf-life study was performed, verifying the claimed 2-year shelf life is scientifically reliable.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated for specific non-clinical tests. General statements indicate "non-clinical tests were conducted."
    • Data Provenance: The tests were conducted internally by Jiangsu Tingsn Technology Co., Ltd. The document refers to "non-clinical tests (integration and functional) were conducted on the TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound System during product development."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Not applicable for non-clinical testing. The "ground truth" for non-clinical tests typically refers to engineering specifications, regulatory standards, and established physical principles, not expert interpretations of clinical data.

    4. Adjudication Method for the Test Set

    • Not applicable for non-clinical testing. Adjudication methods like 2+1 or 3+1 are used for human expert review of clinical cases to establish ground truth or compare diagnostic performance, which was not part of this submission.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC study was done. The submission explicitly states: "No clinical data was included in this submission." MRMC studies inherently involve human readers evaluating clinical cases.

    6. Standalone (i.e., algorithm only without human-in-the-loop performance)

    • Not applicable as an AI/algorithm-only device. The TINGSN FINDERS 2 is a diagnostic ultrasound system that produces images for human interpretation, not an AI algorithm performing diagnosis independently. The software features listed are for image optimization, measurements, and display, assisting the human user, not replacing them in a "standalone" diagnostic role.

    7. The Type of Ground Truth Used

    • For non-clinical testing: The "ground truth" was established by performance specifications, engineering requirements, and compliance with the international and national standards listed (e.g., IEC, ISO, NEMA, ASTM, FDA guidance). This includes electrical safety, mechanical robustness, image quality parameters (resolution, penetration, consistency), acoustic output levels, and software functionality validation against design requirements.

    8. The Sample Size for the Training Set

    • Not applicable. This device is a traditional ultrasound system, not an AI/ML product that undergoes "training."

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K250751
    Manufacturer
    Date Cleared
    2025-07-17

    (127 days)

    Product Code
    Regulation Number
    870.1200
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Jersey 08873

    Re: K250751
    Trade/Device Name: DualView Catheter
    Regulation Number: 21 CFR 870.1200
    Diagnostic Intravascular Catheter
    Classification Panel: Cardiovascular
    Regulation: 21 CFR 870.1200

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DualView Catheter is intended for the intravascular imaging of coronary arteries and is indicated in patients who are candidates for transluminal interventional procedures.

    Device Description

    The DualView Catheter is a catheter consisting of two assemblies: the catheter sheath and the imaging core (consisting of lens and transducer). During imaging, the imaging core rotates inside the catheter sheath to obtain a 360°image of the surface layer of the vessel wall by irradiating with near-infrared light and ultrasound. By pulling back the imaging core inside the catheter sheath, an image in the long axis direction can be obtained.

    This is a rapid exchange (RX) design (short monorail) catheter, which is used with a 0.014" (0.36 mm) guidewire. The catheter is 2.6 Fr (0.86 mm) in the imaging window section and 3.0 Fr (1.01 mm) in the shaft section with an effective length of 137 cm. The catheter has a 100 cm hydrophilic coating starting from the distal end, which becomes highly lubricious when wet. The catheter has a telescoping section, and the telescoping length is 155 mm. When connected to the OPUSWAVE, the imaging core can be pulled back 150 mm in the catheter sheath. There are two radiopaque markers. The distal radiopaque marker is located 7 mm from the distal end of the catheter sheath, and the sensor radiopaque marker is located where the near-infrared light and ultrasound are emitted. Those markers allow a user to confirm the positional relationship between the distal end of the catheter and the sensor position (imaging point). There are two depth markers, one at 90 cm and the other at 100 cm from the distal end of the catheter sheath, which serves as a guide for insertion.

    The transducer has an IPX7 ingress rating in accordance with IEC 60529.

    The catheter is stored in the holder tube and is secured to the catheter holder. The catheter comes with the Motor Drive Unit (MDU) Cover and accessories. The MDU Cover consists of an adapter and a plastic cover sheet to maintain the sterility of the catheter and clean field. The catheter accessories consist of a connection tube with a three-way stopcock, a priming syringe, and a reservoir syringe for priming the catheter lumen with heparinized saline solution.

    AI/ML Overview

    The provided FDA 510(k) Clearance Letter concerns the DualView Catheter, a diagnostic intravascular catheter. This document is a Summary of a Traditional 510(k) submission, which primarily focuses on demonstrating substantial equivalence to existing predicate devices based on non-clinical performance testing.

    Therefore, the submission does not include a Multi-Reader Multi-Case (MRMC) comparative effectiveness study, standalone algorithm performance, or extensive details on ground truth establishment involving human expert consensus for a clinical test set as would be typical for an AI/ML-based device. The clearance is based on direct device performance and safety, primarily through bench testing and animal studies, not a human reader study.

    Here's an analysis of the provided information, specifically addressing the questions as much as possible given the nature of this particular 510(k) (a medical device clearance, not an AI/ML algorithm clearance):

    Acceptance Criteria and Device Performance (Based on Non-Clinical Testing):

    Since this is a non-clinical device clearance, the "acceptance criteria" are the successful completion of the listed performance and safety tests, demonstrating the device meets its design specifications and is suitable for its intended use. The "reported device performance" is that it successfully met these criteria.

    Acceptance Criteria (Test Item)Reported Device Performance
    Radio-detectabilityMet requirements
    Appearance of CatheterMet requirements
    Corrosion ResistanceMet requirements
    Tensile StrengthMet requirements
    Freedom from leakageMet requirements
    Small bore connectorMet requirements
    Sliding resistance (Imaging window)Met requirements
    Particle countMet requirements
    Slidability of GuidewireMet requirements
    Bending strengthMet requirements
    DimensionsMet requirements
    Imaging qualityMet requirements
    Pullback/forward durabilityMet requirements
    Coating IntegrityMet requirements
    Stent crossabilityMet requirements
    Simulated use – Human factor usability engineeringMet requirements
    Software Verification and ValidationConducted per IEC 62304 and FDA guidance ("Enhanced" level) - Met requirements
    Electrical Safety & EMCCompliant with IEC 60601-1, -2-18, -1-2, -2-37, and IEC 60825 - Met requirements
    BiocompatibilityCompliant with ISO 10993-1, tested on accelerated aged devices - Met requirements
    SterilizationValidated per ISO 11135:2014/Amd 1:2018 (SAL of 10-6) - Met requirements
    Animal Study (Safety)No issues related to tissue damage and thrombus formation in coronary arteries
    Animal Study (Performance)System performed as intended compared to predicate devices

    Detailed Study Information (Where Applicable for this Device Type):

    1. Sample sizes used for the test set and the data provenance:

      • Test Set (Non-Clinical): The document refers to various non-clinical tests (e.g., performance testing, software V&V, electrical safety, biocompatibility, sterilization). The "sample size" for these tests would correspond to the number of catheters or test articles subjected to each specific test. This specific number is not provided in the summary but is assumed to be sufficient for each test type according to relevant standards.
      • Animal Study: A "swine model" was used for both safety and performance animal studies. The specific number of animals is not provided in this summary.
      • Data Provenance: The document does not explicitly state the country of origin for the data (e.g., test labs, animal facilities). This is a traditional 510(k) submission, and the manufacturer is Terumo Corporation (Japan), with manufacturing in Japan. Animal studies were likely prospective.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not Applicable in the context of this 510(k). This clearance is for a medical device (catheter), not an AI/ML algorithm that requires human expert consensus for image interpretation ground truth. The "ground truth" for this device's performance is established by direct physical measurements, engineering validations, and physiological outcomes in animal models against predefined specifications and safety standards.
    3. Adjudication method for the test set:

      • Not Applicable. Since there's no human interpretation component needing adjudication for ground truth establishment. Test outcomes are determined by objective measurements against acceptance criteria.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC study was NOT done. The document explicitly states: "This 510(k) does not include data from clinical tests." MRMC studies are typically used to assess the impact of AI algorithms on human reader performance, which is not relevant for this device's non-clinical clearance pathway.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • No, not in the sense of an AI algorithm. This device does not appear to involve an AI algorithm whose performance would be assessed in a standalone manner for image interpretation or diagnosis. It is a diagnostic imaging catheter that generates images for human interpretation.
    6. The type of ground truth used:

      • Engineering Specifications, Physical Measurements, and Physiological Outcomes (Animal Model):
        • For non-clinical performance tests (e.g., tensile strength, dimensions, imaging quality, etc.), the ground truth is the device's adherence to pre-defined engineering specifications and measurable performance characteristics.
        • For biocompatibility and sterilization, the ground truth is established by adherence to relevant ISO standards and successful completion of validated tests.
        • For animal studies, the "ground truth" relates to the physiological effects observed (e.g., absence of tissue injury, thrombus formation) and the successful performance of the device in a living system as intended by design, compared to predicate devices.
    7. The sample size for the training set:

      • Not Applicable. This is not an AI/ML device that requires a training set of data.
    8. How the ground truth for the training set was established:

      • Not Applicable. As no training set for an AI/ML algorithm is involved.

    Summary:

    The DualView Catheter received 510(k) clearance based on demonstrating substantial equivalence to predicate devices primarily through rigorous non-clinical performance testing and animal studies. This type of submission relies on showing that the new device meets established safety and performance standards equivalent to existing legally marketed devices, rather than a clinical study evaluating an AI algorithm's interpretive accuracy and its impact on human readers. Therefore, many of the questions pertinent to AI/ML device clearances (e.g., human expert ground truth, MRMC studies) are not applicable to this traditional medical device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250886
    Date Cleared
    2025-06-18

    (85 days)

    Product Code
    Regulation Number
    892.1550
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Automated Radiological Image Processing Software | 892.2050 | QIH |
    | Diagnostic Intravascular Catheter | 870.1200
    Automated Radiological Image Processing Software | 892.2050 | QIH |
    | Diagnostic Intravascular Catheter | 870.1200
    Identical to predicate |
    | Secondary Regulation Number | 21 CFR 892.1570
    21 CFR 892.1560
    21 CFR 870.1200

    21 CFR 892.2050 | 21 CFR 892.1570
    21 CFR 892.1560
    21 CFR 870.1200
    21 CFR 892.2050 | Identical

    21 CFR 892.2050 | 21 CFR 892.1570
    21 CFR 892.1560
    21 CFR 870.1200
    21 CFR 892.2050 | Identical

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended Use / Indications for Use

    EPIQ:

    The intended use of Philips EPIQ series diagnostic ultrasound systems is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use:

    Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intra-cardiac Echo, Intra-luminal, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Ophthalmic, Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.

    Modes of operation include: B Mode, M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode, Power Doppler, and Harmonic Imaging.

    The clinical environments where Philips EPIQ diagnostic ultrasound systems can be used include clinics, hospitals, and clinical point-of-care for diagnosis of patients.

    When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live echo guidance.

    The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best clinical procedure.

    Affiniti:

    The intended use of the Affiniti Series Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow analysis of the human body with the following Indications for Use:

    Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.

    Modes of operation include: B Mode, M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode, Power Doppler, and Harmonic Imaging.

    The clinical environments where the Affiniti Diagnostic Ultrasound Systems can be used include Clinics, Hospitals, and clinical point-of-care for diagnosis of patients.

    The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgment and best clinical procedure.

    Device Description

    The purpose of this Special 510(k) Pre-Market Notification is to expand the Smart View Select (SVS) software application onto the Affiniti Series Diagnostic Ultrasound Systems and to modify the Smart View Select software application onto both the EPIQ Series Diagnostic Ultrasound Systems.

    Smart View Select is an automated software feature that assists the user in selection of images for analysis with the existing Philips AutoStrain LV or 2D Auto LV application in Adult Echo Transthoracic (TTE) examination. This feature automatically classifies each acquired image by view and selects an appropriate set of images for left ventricle (LV) analysis. The classification is based on a Deep Learning AI interface engine; the selection is a non-AI algorithm that considers the view classification and image depth to select the optimal set of images.

    No hardware changes to the EPIQ or Affiniti systems are required when using SVS, and existing, cleared Philips TTE transducers are used with these software applications.

    The SVS v2 feature is supported by all EPIQ models running software version 13.0 or higher including EPIQ CVx/CVxi, EPIQ Elite Advanced, EPIQ Elite, EPIQ 7, EPIQ 5. The SVS v2 feature is supported by the Affiniti models running software version VM13.0 or higher, including Affiniti CVx, Affiniti 70, Affiniti 50, and Affiniti 30. The SVS and SWM software features are associated with the cardiac adult indication.

    AI/ML Overview

    The provided document details the 510(k) clearance for Philips Ultrasound's EPIQ and Affiniti Series Diagnostic Ultrasound Systems with the Smart View Select (SVS) v2 software. The SVS feature is an automated software that assists users in selecting images for Left Ventricle (LV) analysis using existing Philips AutoStrain LV or 2D Auto LV applications in Adult Echo Transthoracic (TTE) examinations. It classifies acquired images by view using a Deep Learning AI interface and then uses a non-AI algorithm to select an optimal set of images.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The study aimed to demonstrate agreement between LV analysis outputs (Ejection Fraction - EF, and Global Longitudinal Strain - GLS) derived from manually selected clips (ground truth) and automatically selected clips by the SVS software.

    Acceptance CriteriaReported Device Performance
    Co-primary Endpoint: Lower Confidence Bound for Pearson's correlation coefficient to be > 0.8 for both GLS and EF measurements.Pearson's Correlation Coefficient for EF: 0.891 (95% CI: 0.851, 0.920)
    Lower Confidence Bound: 0.851 (Met criteria)
    Pearson's Correlation Coefficient for GLS: 0.906 (95% CI: 0.871, 0.931)
    Lower Confidence Bound: 0.871 (Met criteria)

    2. Sample Size and Data Provenance

    • Test Set Sample Size: The exact number of patients or cases in the test set is not explicitly stated as a single number. However, the demographic characteristics section indicates n=71 for various demographic variables like Sex, Age, Height, Weight, BMI, Race, LV systolic function, RWMA, Known CAD, Previous reported MI Location, and LV Hypertrophy, suggesting the test dataset comprised data from at least 71 unique patients.
    • Data Provenance: The data was collected from a US-based medical center. The study was a retrospective data analysis.

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: Three (3) reviewers (clinical experts) participated in establishing the ground truth.
    • Qualifications of Experts: They are referred to as "clinical experts." No further specific qualifications (e.g., years of experience, specific board certifications like radiologist/cardiologist) are provided in the document.

    4. Adjudication Method for the Test Set

    The adjudication method used to establish ground truth was consensus/averaging. For each output (GLS and EF), the average across the three reviewers was used as the ground truth.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was described. The study focused on comparing the AI-selected clips' performance (GLS and EF) against the ground truth established by human experts using manually selected clips. There is no information provided regarding how much human readers improve with AI vs. without AI assistance. The study evaluates the AI's ability to select optimal clips, a step that precedes human analysis with AutoStrain LV.

    6. Standalone Performance Study (Algorithm Only)

    Yes, a standalone performance study was done. The performance of the SVS software was evaluated solely based on its ability to automatically select appropriate clips, and then the subsequent, un-edited evaluation of GLS and EF through AutoStrain LV using these SVS-selected clips was compared against the human-selected clips. This represents the algorithm's performance in its specific task (clip selection) without direct human intervention in the selection process.

    7. Type of Ground Truth Used

    The ground truth used was expert consensus. Specifically, it was the average of GLS and EF measurements obtained from clips manually selected and subsequently semi-automatically processed by three clinical experts using the AutoStrain LV software.

    8. Sample Size for the Training Set

    The document does not provide the sample size used for the training set of the deep learning AI model for image classification. It only states that "the neural network of the subject device was revised as described in attachment 002," implying updates to an existing model rather than a completely new one, but no training data details are given.

    9. How Ground Truth for Training Set was Established

    The document does not explicitly state how the ground truth for the training set was established. It mentions that the "classification is based on a Deep Learning AI interface engine," but details regarding the training data annotation and ground truth establishment are not provided within the excerpt.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    510k Summary Text (Full-text Search) :

    SOUNDSTAR eco Diagnostic Ultrasound Catheters (8F, 8FG, 10F, 10FG)

    Regulation Number: 21 CFR 870.1200
    Regulatory Name/Reference
    -------------------
    Regulatory Name/Reference
    Ultrasound catheter
    Regulation No.
    21 CFR § 870.1200
    Regulatory Class
    **Product
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medline ReNewal Reprocessed Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheters 8F, 8FG, 10F 10FG are indicated for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. When used with compatible CARTO 3 EP Navigation Systems, the SOUNDSTAR eco Catheter provides location information.

    Device Description

    The Medline ReNewal Reprocessed Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheters 8F, 8FG, 10F, 10FG are licensed for single use only. Models differ in shaft diameter (8F or 10F) and ultrasound system compatibility (Siemens [8F and 10F] or GE [8FG and 10FG]). The catheter consists of a CARTO connector, SwiftLink (ultrasound) connector, handle and shaft. In the distal tip of each catheter's shaft are a 64-element ultrasound transducer providing 2D imaging and a 3D location sensor providing location information to the compatible CARTO 3 Electrophysiology (EP) Navigation System. The deflection and tension knobs on the handle are used to articulate the distal tip and control the image plane orientation with four-directional (anterior-posterior, left-right) deflection.

    AI/ML Overview

    The provided FDA 510(k) clearance letter pertains to the Medline ReNewal Reprocessed Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheters. This document outlines the regulatory review and approval process for a reprocessed medical device, asserting its substantial equivalence to original predicate devices.

    However, the provided text does not contain any information regarding specific acceptance criteria for AI/ML performance, nor does it describe a study involving AI assistance or standalone AI performance evaluation. The "Non-clinical Testing Summary" focuses on the physical, electrical, and reprocessing aspects of the reprocessed catheter itself (e.g., functional performance, electrical safety, cleaning validation, biocompatibility, sterilization validation, packaging and shelf life). It explicitly states: "The current submission does not include consoles or any other system components as part of this respective submissions." The device subject to clearance is an ultrasound catheter, not an AI-powered diagnostic system.

    Therefore, I cannot fulfill the request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria for AI/ML device performance based on the provided document. The document pertains to a physical medical device (an ultrasound catheter) and its reprocessing, not an AI/ML-driven diagnostic or assistive technology.

    If you have a document describing an AI/ML device and its performance study, I would be happy to analyze it according to your requested criteria.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    510k Summary Text (Full-text Search) :

    Steerable Introducer; Agilis™ NxT Steerable Introducer Dual-Reach™

    Regulation Number: 21 CFR 870.1200
    Common Name
    Class
    Classification Name
    Common Name
    Class
    Classification Name
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ViewFlex™ Xtra ICE Catheter
    The ViewFlex™ Xtra ICE Catheter is indicated for use in adult and adolescent pediatric patients to visualize cardiac, structures, blood flow and other devices within the heart.

    ViewFlex™ Eco Reprocessed ICE Catheter
    The ViewFlex™ Eco Reprocessed Catheter is indicated for use in adult and adolescent pediatric patients to visualize cardiac structures, blood flow and other devices within the heart.

    Advisor™ HD Grid Mapping Catheter, Sensor Enabled™
    The Advisor™ HD Grid Mapping Catheter, Sensor Enabled™, is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart.

    Advisor™ HD Grid X Mapping Catheter, Sensor Enabled™
    The Advisor™ HD Grid X Mapping Catheter, Sensor Enabled™, is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart.

    Agilis™ NxT Steerable Introducer
    The Agilis™ NxT Steerable Introducer is indicated for the introduction of various cardiovascular catheters into the heart, including the left side of the heart, during the treatment of cardiac arrhythmias.

    Agilis™ NxT Steerable Introducer Dual-Reach™
    The Agilis™ NxT Steerable Introducer Dual-Reach™ is indicated for the introduction of various cardiovascular catheters into the heart, including the left side of the heart, during the treatment of cardiac arrhythmias.

    Device Description

    The Agilis™ NxT Steerable Introducer Dual-Reach™ is a sterile, single-use device that con-sists of a dilator and steerable introducer, which is designed to provide flexible catheter positioning in the cardiac anatomy. The inner diameter of the steerable introducer is 13F. The steerable introducer includes a hemostasis valve to minimize blood loss during catheter intro-duction and/or exchange. It has a sideport with three-way stopcock for air or blood aspiration, fluid infusion, blood sampling, and pressure monitoring. The handle is equipped with a rotating collar to deflect the tip clockwise ≥180° and counterclockwise ≥90°. The steerable introducer features distal vent holes to facilitate aspiration and minimize cavitation and a radiopaque tip marker to improve fluoroscopic visualization.

    AI/ML Overview

    This FDA 510(k) clearance letter (K251211) and its accompanying 510(k) summary pertain to a change in workflow for several existing cardiovascular catheters, specifically allowing for a "Zero/Low Fluoroscopy Workflow."

    The key phrase here is "Special 510(k) – Zero/Low Fluoroscopy Workflow". This type of submission is for modifications to a previously cleared device that do not significantly alter its fundamental technology or intended use, but rather introduce a change in how it's used or processed.

    Crucially, this submission does NOT describe a new AI/software device that requires extensive performance testing against acceptance criteria in the manner you've outlined for AI/ML devices. Instead, it's about demonstrating that existing devices, when used with a new, less-fluoroscopy-dependent workflow, remain as safe and effective as before.

    Therefore, many of the questions you've asked regarding acceptance criteria, study details, ground truth, and expert adjudication are not applicable to the information provided in this 510(k) document. The document explicitly states:

    • "Bench-testing was not necessary to validate the Clinical Workflow modifications."
    • "Substantial Equivalence of the subject devices to the predicate devices using the zero/low fluoroscopy workflow has been supported through a summary of clinical data across multiple studies in which investigators used alternative visualization methods."

    This indicates that the "study" proving the device (or rather, the new workflow) meets acceptance criteria is a summary of existing clinical data where alternative visualization methods were already employed, rather than a prospective, controlled study of a new AI algorithm.

    Based on the provided document, here's what can be answered:

    1. A table of acceptance criteria and the reported device performance:

    • Acceptance Criteria: The implicit acceptance criterion is that the devices, when used with "zero/low fluoroscopy workflow," maintain substantial equivalence to their predicate devices in terms of safety and effectiveness. This means they must continue to perform as intended for visualizing cardiac structures, blood flow, mapping, or introducing catheters.
    • Reported Device Performance: The document states that "Substantial Equivalence... has been supported through a summary of clinical data across multiple studies in which investigators used alternative visualization methods." This implies that the performance (e.g., adequate visualization, successful mapping, successful catheter introduction) was maintained. Specific quantitative metrics of performance (e.g., accuracy, sensitivity, specificity, or inter-reader agreement for a diagnostic AI) are not provided or applicable here as this is not an AI/ML diagnostic clearance.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified. The document refers to "a summary of clinical data across multiple studies." This suggests an aggregation of results from existing (likely retrospective) patient data where alternative visualization techniques (allowing for "zero/low fluoroscopy") were already utilized clinically. It's not a new, single, prospectively designed test set for an AI algorithm.
    • Data Provenance: Not specified regarding country of origin or specific patient demographics. It is implied to be clinical data collected from studies where these types of procedures were performed using alternative visualization. The data would be retrospective as it's a "summary of clinical data" that already exists.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable in the context of this 510(k). Ground truth in an AI/ML context typically refers to adjudicated labels for images or signals. Here, the "ground truth" is inferred from standard clinical practice and outcomes in the historical data summarized. There's no mention of a specific expert panel for new ground truth establishment for a diagnostic AI.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. This is not a study requiring adjudication of diagnostic outputs by multiple readers.

    5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is not an AI-assisted diagnostic device. The workflow change is about using alternative non-fluoroscopic imaging modalities (e.g., intracardiac echocardiography, electro-anatomical mapping systems), not about AI improving human reader performance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No. This is not an AI algorithm. The predicate devices are physical catheters.

    7. The type of ground truth used:

    • The "ground truth" is inferred from clinical outcomes and established clinical practice using the devices with alternative visualization methods in real-world scenarios. It's not a specific, adjudicated dataset for an AI algorithm. The performance of the devices (such as successful navigation, visualization, and mapping) under the "zero/low fluoroscopy" workflow is assumed to be equivalent to their performance under full fluoroscopy, as demonstrated by prior clinical use where such methods were employed.

    8. The sample size for the training set:

    • Not applicable. There is no AI model being trained discussed in this document.

    9. How the ground truth for the training set was established:

    • Not applicable. No training set for an AI model.

    In summary:

    This 510(k) is for a workflow modification for existing medical devices (catheters), not for an AI/ML diagnostic or assistive software. Therefore, the detailed data performance evaluation typically required for AI models against specific acceptance criteria (as requested in your template) is not presented or relevant in this clearance letter. The "proof" relies on the concept of substantial equivalence to previously cleared predicate devices, supported by a summary of existing clinical data that used alternative visualization methods, implying that the devices function safely and effectively even with reduced fluoroscopy.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243793
    Date Cleared
    2025-05-21

    (162 days)

    Product Code
    Regulation Number
    892.1550
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Medical image management and processing system | 892.2050 | QIH |
    | | Diagnostic Intravascular Catheter | 870.1200
    Identical to predicate |
    | Secondary Regulation Number | 21 CFR 892.1570
    21 CFR 892.1560
    21 CFR 870.1200

    21 CFR 892.2050 | 21 CFR 892.1570
    21 CFR 892.1560
    21 CFR 870.1200
    21 CFR 892.2050 | Identical

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EPIQ: The intended use of EPIQ Ultrasound Diagnostic System is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use: Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), intra-luminal, intra-cardiac echo, Musculoskeletal (Conventional), Musculoskeletal (Superficial), Ophthalmic, Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.
    Affiniti: The intended use of Affiniti Series Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use: Abdominal, Cardiac Adult, Cardiac Other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.

    Device Description

    The R-Trigger algorithm software feature on Philips EPIQ and Affiniti Ultrasound System is intended to support detection of R-wave peak (R-trigger) as an input to certain TTE clinical applications, initially including AutoStrain LV, AutoEF, 2D Auto LV (collectively referred to as "AutoStrain"), and AutoMeasure applications. The R-trigger algorithm is planned to be implemented as workflow enhancement for transthoracic clinical applications on EPIQ and Affiniti Ultrasound Systems in the VM13 software release. The Auto-Measure and AutoStrain features support users during B-mode (2D), CW-, PW- and TDI-Doppler measurements by automating some of the measurements needed to complete a routine transthoracic echo (TTE) exam for adult patients. The R-trigger feature (non-ECG-based) has been developed to enable clinical users to use AutoMeasure and AutoStrain application without the R-trigger (ECG based) input, which is currently required. There are no hardware changes to the EPIQ and Affiniti systems due to change to the introduction of the R-Trigger software application. The software application is supported by all EPIQ and Affiniti models running software version 13.0 or higher.

    AI/ML Overview

    The provided FDA 510(k) clearance letter describes the R-Trigger software application on Philips EPIQ and Affiniti Ultrasound Systems, which aims to provide an alternative method for detecting R-wave peaks (R-triggers) for cardiac clinical applications like AutoStrain and AutoMeasure, especially when the ECG signal is unavailable or unusable.

    Here's an analysis of the acceptance criteria and the study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The primary acceptance criteria for the R-Trigger algorithm are related to the agreement of its R-wave time stamp detection with the ground truth (ECG-based R-trigger) and the subsequent impact on the clinical outputs of AutoMeasure and AutoStrain. These are evaluated using Bland-Altman analysis (for agreement, specifically the Upper and Lower Limits of Agreement, LoA) and Pearson's correlation (for correlation, specifically the Lower Confidence Bound, LCB).

    Endpoint / Outcome ComparisonMeasurement TypeAcceptance Criteria (Upper/Lower LoA or LCB)Reported Device Performance (Upper/Lower LoA or Pearson's r with 95% CI)Met Criteria?
    Endpoint 1: R-trigger
    R-wave peak time stampTime Stamp[-99.5ms, 99.5ms]-58.06ms (-59.34, -56.78) to 69.69ms (68.41, 70.97)Yes (-58.06 > -99.5, 69.69 0.8
    GLSGLS (Correlation)LCB > 0.80.992 (0.990, 0.994)Yes (0.990 > 0.8)
    Endpoint 2: AutoMeasure
    MV E VelPw/cw Doppler velocity[-25%, 25%]-12.00 % (-13.17%, -10.84 %) to 12.98 % (11.81 %, 14.14%)Yes (-12.00 > -25, 12.98 -30, 9.33 -30, 5.35 -29, 17.44 -29, 17.91 -30, 9.39 -25, 14.69 -30, 9.30 -30, 9.55 -29, 14.41 -30, 13.00 -28, 17.97 -28, 17.53 -25, 13.83 -29, 14.77 -30, 10.64 -34, 18.23 -30, 8.83 -46, 32.89 -46, 20.47 -46, 21.51 -28, 21.44 -28, 20.89 -28, 21.62 -40, 13.59 -35, 23.62 -40, 14.21 -30, 14.74 -30, 12.98 -30, 12.68 -30, 16.86 -29, 17.04
    Ask a Question

    Ask a specific question about this device

    K Number
    K251231
    Manufacturer
    Date Cleared
    2025-05-20

    (28 days)

    Product Code
    Regulation Number
    870.1200
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Trade/Device Name: ViewFlex™ X ICE Catheter, Sensor Enabled™
    Regulation Number: 21 CFR 870.1200
    Name** | Catheter, Ultrasound, Intravascular |
    | Class | 2 |
    | Classification Name | 21 CFR 870.1200

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ViewFlex™ X ICE Catheter Sensor Enabled™ is indicated for use in adult and adolescent pediatric patients for intra-cardiac and intra-luminal visualization of cardiac and great vessels anatomy and physiology, as well as visualization of other devices in the heart. When used with a compatible three-dimensional mapping system, the catheter provides location information.

    Device Description

    The ViewFlex™ X ICE Catheter Sensor Enabled™ (SE) is a sterile, single use, temporary, radiopaque, intracardiac ultrasound catheter. The catheter shaft is a 9 French (F) catheter constructed with flexible tubing with a useable length of 90 cm. The shaft is compatible with a 10 French or larger introducer for insertion into the femoral or jugular veins. The catheter tip is a 64‑element linear phased array transducer. The distal portion of the shaft is deflectable utilizing two handle mechanisms which create four deflection directions including left, right, anterior and posterior. The distal tip contains an ultrasound transducer and 3-D location sensor providing 2-D imaging and 3-D location and orientation information when used with a compatible ultrasound system and the EnSite X Cardiac Mapping System.

    AI/ML Overview

    This document is a 510(k) clearance letter for the ViewFlex™ X ICE Catheter, Sensor Enabled™. The provided text does not contain any information regarding acceptance criteria or the study that proves the device meets those criteria, especially in the context of an AI/ML-enabled device as implied by the prompt's request for information about human readers, AI assistance, ground truth, and training sets.

    The device described is an intracardiac ultrasound catheter that provides 2-D imaging and 3-D location/orientation information. The mention of "Sensor Enabled™" and "3-D location sensor" suggests a technological upgrade, but there is no indication that this involves the use of artificial intelligence or machine learning for diagnostic interpretation.

    Therefore, I cannot provide the requested information based on the provided text. The prompt's questions (acceptance criteria, study details, sample sizes, expert qualifications, adjudication, MRMC studies, standalone performance, ground truth types, training set details) are all relevant to the evaluation of AI/ML-enabled medical devices, which is not what this 510(k) summary describes.

    In summary, the provided document explains the ViewFlex™ X ICE Catheter, Sensor Enabled™ as a traditional medical device (an intravascular ultrasound catheter) with an added 3-D location sensor for mapping systems. It details the substantial equivalence to a predicate device based on non-clinical testing (bench design verification, biocompatibility, mechanical integrity, etc.). It does not mention any AI or ML components, nor does it describe studies with human readers, AI assistance, or data sets for machine learning model evaluation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250592
    Date Cleared
    2025-05-12

    (74 days)

    Product Code
    Regulation Number
    870.1200
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Device Name:** NES Reprocessed Visions PV .018 Digital IVUS Catheter
    Regulation Number: 21 CFR 870.1200
    Name:** Reprocessed Intravascular Ultrasound Catheter
    Classification Number: Class II, 21 CFR 870.1200

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NES Reprocessed Visions PV .018 Digital IVUS Catheter is designed for use in the evaluation of vascular morphology in blood vessels of the coronary and peripheral vasculature by providing a cross-sectional image of such vessels. This device is not currently indicated for use in cerebral vessels.

    The NES Reprocessed Visions PV .018 Digital IVUS Catheter is designed for use as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures.

    Device Description

    The NES Reprocessed Visions PV .018 Digital IVUS Catheter utilizes a cylindrical ultrasound transducer array close to the distal tip to radiate acoustic energy into the surrounding tissue. By using the information gathered from the ultrasonic echoes, the device can generate real-time images of the coronary and peripheral vessels.

    The NES Reprocessed Visions PV .018 Digital IVUS Catheter tracks over a 0.018" (0.46mm) guide wire by through an internal lumen. Approximately 31 cm from the catheter tip is a guidewire port for the exiting guidewire. The NES Reprocessed Visions PV .018 Digital IVUS Catheter is introduced into the vascular system percutaneously or via surgical cutdown.

    The NES Reprocessed Visions PV .018 Digital IVUS Catheter may only be used with the Volcano s5 Series or CORE Series of Systems Operator's Manual. If connect to another system, the catheter will not operate.

    The NES Reprocessed Visions PV .018 Digital IVUS Catheter is intended to be reprocessed one (1x) time.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for K250592 describes a reprocessed medical device, the NES Reprocessed Visions PV .018 Digital IVUS Catheter. This document does not detail acceptance criteria or the study that proves the device meets AI/Machine Learning performance standards, as it is a traditional medical device (an Intravascular Ultrasound Catheter) and not an AI-enabled device.

    Therefore, many of the requested categories regarding AI model performance, ground truth establishment, expert adjudication, MRMC studies, and training/test set sizes are not applicable to this submission.

    However, I can extract the relevant information regarding the device's functional and safety testing as described in the 510(k) summary, which serves as the "study" proving the device meets its acceptance criteria for a reprocessed medical device.


    Summary of Device Acceptance Criteria and Performance (for a Reprocessed Medical Device, not AI)

    The NES Reprocessed Visions PV .018 Digital IVUS Catheter is a reprocessed version of a predicate device. The acceptance criteria and "study" proving the device meets these criteria focus on demonstrating that reprocessing does not compromise the device's safety and effectiveness compared to the original device. The "study" involves various bench and laboratory tests.

    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) summary states that "The test methods, requirements and acceptance criteria used for the .018 IVUS are the same used in cleared K200195 (reprocessed .014 IVUS)." It also indicates that there are "no changes to the claims, clinical applications, patient populations, performance specifications, or method of operation." This implies that the acceptance criteria are met if the reprocessed device performs comparably to a new device (or the predicate reprocessed device K200195) across all specified tests.

    Acceptance Criteria CategorySpecific Tests/RequirementsReported Device Performance (as implied by clearance)
    SafetyCleaning ValidationMeets validation requirements
    Sterilization ValidationMeets validation requirements
    Biocompatibility TestingMeets biocompatibility standards
    Transducer Thermal TestingMeets thermal output specifications
    Drying ValidationMeets validation requirements
    Packaging ValidationMeets validation requirements
    Functional/PerformanceVisual InspectionPasses visual inspection
    Dimensional VerificationWithin specified dimensional tolerances
    Simulated UsePerforms as expected in simulated use conditions
    Mechanical CharacteristicsMeets mechanical strength and integrity specifications
    System CompatibilityCompatible with Volcano s5 Series or CORE Series Systems
    Image AcuityProvides images with sufficient clarity and resolution
    Transducer Acoustic OutputWithin safe and effective acoustic output limits
    Flushing Adapter CompatibilityCompatible with relevant flushing adapters
    Surface Property TestingMeets specified surface property requirements

    Note: The document states that the "NES Reprocessed Visions PV .018 Digital IVUS Catheter is as safe and effective as the predicate devices described herein," which is the overarching conclusion of meeting all acceptance criteria.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated for each test, but standard for device validations. For a reprocessed device, this would typically involve a statistically significant number of reprocessed units across multiple reprocessing cycles.
    • Data Provenance: The tests are described as "Bench and laboratory testing," implying controlled, experimental data generated specifically for this submission. There is no mention of patient data.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • N/A: As this is a traditional medical device (not an AI/ML device), expert establishment of ground truth in the context of image interpretation or diagnostic accuracy is not relevant. The "ground truth" for the device's performance would be established by validated test methods and engineering specifications.

    4. Adjudication Method for the Test Set

    • N/A: Adjudication is typically relevant for subjective assessments (e.g., in AI or clinical studies). For bench and lab testing of a physical device, performance is measured against objective, predetermined specifications rather than requiring expert consensus or adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • N/A: An MRMC study is relevant for evaluating the impact of AI on human reader performance. This device is an IVUS catheter and does not involve AI assistance for image interpretation.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • N/A: This question pertains to AI algorithm performance. This device is a physical catheter, not an algorithm.

    7. The Type of Ground Truth Used

    • Engineering Specifications and Performance Standards: The "ground truth" for this device's performance is compliance with established engineering specifications, safety standards (e.g., sterilization, biocompatibility), and functional performance benchmarks (e.g., image acuity, acoustic output) that are deemed equivalent to the original predicate device. This is determined through objective bench and laboratory testing.

    8. The Sample Size for the Training Set

    • N/A: No training set is applicable as this is not an AI/ML device.

    9. How the Ground Truth for the Training Set was Established

    • N/A: Not applicable.
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    510k Summary Text (Full-text Search) :

    VSICE2D); VeriSight Pro Intracardiac Echocardiography Catheter (VSICE3D)
    Regulation Number: 21 CFR 870.1200
    Catheter |
    | Regulation Description | Diagnostic intravascular catheter |
    | Regulation Number | 870.1200

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VeriSight/VeriSight Pro ICE catheter is intended for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients. The catheter is intended for imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures.

    Device Description

    The VeriSight/VeriSight Pro Intracardiac Echocardiography (ICE) catheters are sterile, disposable, and for single use only. The catheter's distal end has an ultrasound transducer providing 2D and/or 3D imaging capabilities. A handle, located at the proximal end of the catheter, has two steering wheels that can be manually operated to control four-way articulation of the distal segment in anterior, posterior, left and right directions. The catheter has a 9 French (F) shaft and a usable length of 90 cm.
    The VeriSight and VeriSight Pro ICE Catheters are identical in all regards (material, processing, assembly and packaging). The VeriSight ICE catheter provides 2D ultrasound imaging capabilities. The VeriSight Pro ICE catheter provides 2D and/or 3D ultrasound imaging capabilities, depending on the model and configuration of the EPIQ ultrasound system it connects to. The catheters are compatible with ancillary equipment such as sheaths and introducers. The catheters are sterilized via Ethylene Oxide.
    The catheters connect to the Philips EPIQ Diagnostic Ultrasound System via a patient interface module (PIM); the connection between the catheter and the PIM is located outside the sterile field. The catheters will not operate if connected to any other imaging system. These catheters are for exclusive use with Philips EPIQ 7C, CVx, and CVxi series of ultrasound systems, cleared under K202216.

    AI/ML Overview

    Based on the provided FDA 510(k) clearance letter for the Philips VeriSight Intracardiac Echocardiography Catheters, the device is not an AI/ML-based device. The submission focuses on adding pediatric indications to an existing, cleared device, and thus, the information requested about acceptance criteria and a study proving an AI device meets those criteria is not applicable to this specific submission.

    The document explicitly states:

    • "The purpose of this submission is solely to add pediatric indications to the currently cleared VeriSight/VeriSight Pro ICE catheters (cleared under K200812)." (Page 6)
    • "No additional non-clinical performance testing was executed for this change." (Page 6)
    • "The VeriSight/VeriSight Pro ICE catheters use the same scientific technology, operating principles and shares similar indications for use as the predicate device. Therefore, clinical data is not required to establish substantial equivalence." (Page 6)

    Therefore, the document does not contain details regarding acceptance criteria, performance data, test sets, ground truth establishment, or any other aspects related to the validation of an AI/ML device. The clearance is based on the substantial equivalence to a predicate device for the expanded indications, rather than a new technological performance validation.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 42