Search Results
Found 268 results
510(k) Data Aggregation
(267 days)
FOZ
BD Nexiva™ Closed IV Catheter System - Single Port
BD Nexiva™ Closed IV Catheter System is intended to be inserted into a patient's peripheral vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa) when access ports not suitable for use with power injectors are removed.
BD Nexiva™ Closed IV Catheter System - Dual Port
BD Nexiva™ Closed IV Catheter System is intended to be inserted into a patient's peripheral vascular system for short term use to sample blood or administer fluids. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa) when access ports not suitable for use with power injectors are removed.
BD Nexiva™ Closed IV Catheter Systems are over-the-needle, intravascular catheters. These devices have a radiopaque BD Vialon™ catheter, needle, needle shield, septum, stabilization platform, integrated extension tubing, clamp, Luer adapter (single or dual port), vent plug, and pre-attached needle-free connector (BD Q-Syte™ or BD MaxZero™) (dual port configurations only). The needle and catheter are protected by a needle cover. A BD Q-Syte™ Needle-free Connector, BD MaxZero™ Needle-free Connector, or end cap with protective cover is provided in the unit package (not available with all configurations).
The closed system is designed to keep blood contained within the device throughout the insertion process. The septum is designed to wipe visible blood from the needle surface as the needle is withdrawn from the catheter, further reducing the risk of blood exposure. The needle tip is passively protected when the needle is removed, reducing the risk of accidental needlestick injury.
These devices have BD Instaflash™ needle technology, allowing for immediate visualization of blood along the catheter. Continuous blood return is seen in the extension tubing. The vent plug prevents blood leakage from the extension tubing during insertion. Both the stabilization platform and Luer connector are color coded to indicate catheter gauge size (24 GA (0.7 mm) = Yellow, 22 GA (0.9 mm) = Blue, 20 GA (1.1 mm) = Pink, 18 GA (1.3 mm) = Green).
The provided 510(k) clearance letter and summary discuss the BD Nexiva™ Closed IV Catheter System, not an AI/ML medical device. Therefore, information regarding acceptance criteria, study details, and ground truth establishment pertinent to AI/ML device performance (sample sizes, expert consensus, MRMC studies, etc.) is not available in the provided text.
The document primarily focuses on demonstrating substantial equivalence to a predicate device (K183399) through performance testing of the physical catheter system, rather than validating an algorithm's diagnostic or predictive capabilities.
The performance tests mentioned are:
- Frequency Response Testing
- Kink Resistance Testing
- Blood Fill Time Testing
- ISO 80369-7 Testing
- ISO 11607-1 Testing
However, the specific acceptance criteria and detailed reported performance results for these tests are not explicitly listed in a tabular format, nor are sample sizes, ground truth methodologies, or expert involvement for these physical device performance tests detailed as one might expect for an AI/ML device. The document only states that the device "met all predetermined acceptance criteria for the above-listed performance tests."
Therefore, I cannot fulfill the request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria in the context of an AI/ML device as the input document describes a physical medical device.
Ask a specific question about this device
(88 days)
FOZ
BD Cathena™ Safety IV Catheter is intended to be inserted into a patient's peripheral vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. This device may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. This device is suitable for use with power injectors set to a maximum pressure of 325 psi (2240 kPa).
BD Cathena™ Safety IV Catheter is an over-the-needle, intravascular (IV) catheter. This device includes a radiopaque BD Vialon™ catheter, a needle, a grip, a passive safety needle shield, and a flash chamber with removable vent plug. The needle and catheter are protected by a needle cover. BD Cathena™ Safety IV Catheter has BD Instaflash™ Needle Technology, allowing for immediate visualization of blood along the catheter. The flash chamber provides confirmation that the device has entered the vessel. The needle tip is passively protected when the needle is removed, reducing the risk of accidental needlestick injury.
BD Cathena™ Safety IV Catheter is available with or without multi-access BD Multiguard™ Technology, which is designed to stop the flow of blood from the catheter hub until a Luer connection is made. Once a connection is made, fluids or blood can flow through the catheter hub in either direction.
BD Cathena™ Safety IV Catheter is available with or without wings. The catheter hub and wings are color coded to indicate the catheter gauge size (24 GA (0.7 mm) = Yellow, 22 GA (0.9 mm) = Blue, 20 GA (1.1 mm) = Pink, 18 GA (1.3 mm) = Green, 16 GA (1.7 mm) = Grey).
The FDA 510(k) clearance letter for the BD Cathena™ Safety IV Catheter (K251155) indicates that the submission is for manufacturing changes (new colorant supplier, new needle lubricant supplier) and the creation of new performance specifications for existing indications for use (blood sampling and blood pressure monitoring) rather than a novel AI/ML device. Therefore, a traditional acceptance criteria and study proving device meets acceptance criteria related to AI/ML performance, human reader improvement with AI assistance, or standalone algorithm performance, as typically understood in AI/ML medical devices, is not applicable here.
However, based on the provided document, the acceptance criteria and supporting studies are related to demonstrating that the modified device maintains substantial equivalence to its predicate for its stated indications for use, including the new performance specifications.
Here's an interpretation of the relevant information provided:
I. Acceptance Criteria and Reported Device Performance
The acceptance criteria are generally understood as the "predetermined design requirements" that the device must meet. The document states that the subject device met all predetermined acceptance criteria for the listed performance tests, demonstrating substantial equivalence to the predicate devices. Specific quantitative acceptance criteria are not explicitly detailed in this summary for each test, but the successful completion of these tests serves as evidence of meeting those criteria.
Acceptance Criteria Category | Reported Device Performance (Implied) |
---|---|
Blood Fill Time | Met predetermined design requirements (new performance specifications created and tested). |
Frequency Response | Met predetermined design requirements (new performance specifications created and tested). |
Catheter Kink Resistance | Met predetermined design requirements (new performance specifications created and tested). |
Average System Drag | Met predetermined design requirements (tested). |
Air Vent Time | Met predetermined design requirements (tested). |
Procedural Leak Time | Met predetermined design requirements (tested). |
Instaflash Time | Met predetermined design requirements (tested). |
Biocompatibility | Assessed and adopted from a representative BD device (K201075) utilizing the same intended use, device materials, and manufacturing processing, implying acceptance criteria for biocompatibility were met. |
Sterilization (ISO 11135 & ISO 10993-7) | Met applicable standards requirements (tested). |
Luer Connections (ISO 80369-7) | Met applicable standards requirements (tested). |
Material Equivalence (Colorant & Lubricant) | Design verification testing performed to demonstrate equivalence after supplier changes, implying acceptance criteria for material performance were met. |
II. Sample Size and Data Provenance for Test Set
The document does not specify exact sample sizes for each performance test (e.g., number of catheters tested for Blood Fill Time, Kink Resistance, or Frequency Response). However, it implies that sufficient samples were tested to "ensure that the subject device meets pre-determined design requirements."
The "test set" in this context refers to the samples of the subject device used for physical and functional testing. The provenance of these samples would be prospective, as they are newly manufactured devices undergoing verification for the submission. No information is provided regarding the country of origin of this test data, but it would typically be generated at the manufacturer's testing facilities or authorized contract labs.
III. Number of Experts and Qualifications for Ground Truth
This question is not applicable in the context of this 510(k) submission. This is not a study assessing diagnostic or predictive performance requiring expert interpretation for ground truth. The "ground truth" for the performance tests (e.g., Blood Fill Time, Kink Resistance) is established through standardized laboratory measurement methods and engineering specifications, not expert consensus on medical images or clinical outcomes.
IV. Adjudication Method for Test Set
Not applicable. As described above, the tests are objective physical and functional performance assessments, not subjective interpretations requiring adjudication.
V. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. An MRMC study is typically conducted for diagnostic imaging devices or AI tools where human readers are interpreting cases. This submission is for an intravascular catheter, a physical medical device. The "new performance specifications" are related to the physical and functional characteristics of the catheter (e.g., how quickly blood fills, how well it resists kinking for pressure monitoring), not to aiding human interpretation of medical data. Therefore, a study comparing human readers with and without AI assistance is not relevant or reported here.
VI. Standalone (Algorithm Only) Performance
Not applicable. The BD Cathena™ Safety IV Catheter is a physical medical device and is not an AI algorithm. Its performance is assessed through physical and functional bench testing.
VII. Type of Ground Truth Used
The "ground truth" here is based on engineering specifications and standardized test methods. For example:
- Blood Fill Time: Measured against a specific time objective.
- Frequency Response: Measured against a defined physiological frequency range for accurate blood pressure monitoring.
- Kink Resistance: Measured against criteria for maintaining lumen patency under stress.
- Biocompatibility: Assessed against ISO 10993 standards.
- Sterilization: Assessed against ISO 11135 standards.
- Luer Connections: Assessed against ISO 80369-7 standards.
These are objective, quantitative measurements derived from established industry standards and internal design requirements, not from clinical outcomes, pathology, or expert consensus in a diagnostic sense.
VIII. Sample Size for the Training Set
Not applicable. This is not an AI/ML device that requires a training set.
IX. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this type of device submission.
Ask a specific question about this device
(28 days)
FOZ
BD Insyte™ Autoguard™ Shielded IV Catheter is intended to be inserted into a patient's peripheral vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. This device may also be used to facilitate the placement of guidewires and other vascular access devices such as catheters without pre-attached hubs. This device may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa).
BD Insyte™ Autoguard™ BC Shielded IV Catheter is intended to be inserted into a patient's peripheral vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. This device may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa).
BD Insyte™ Autoguard™ BC Pro Shielded IV Catheter is intended to be inserted into a patient's peripheral vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. This device may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa).
BD Insyte™ Autoguard™ Shielded IV Catheter is an over-the-needle, intravascular (IV) catheter. This device includes a radiopaque BD Vialon™ Catheter Material, a needle, a grip with button, a flash chamber with vent plug, and a barrel. The needle and catheter are protected by a needle cover.
The 24-20 GA (0.7-1.1 mm) devices have BD Instaflash™ Needle Technology, allowing for immediate visualization of blood along the catheter. The flash chamber provides confirmation that the device has entered the vessel.
BD Insyte™ Autoguard™ Shielded IV Catheter incorporates BD Autoguard™ Push-button Shielding Technology which is activated when the button is depressed. A spring mechanism retracts the needle and flash chamber into the barrel, fully encasing the needle and reducing the risk of accidental needlestick injury.
BD Insyte™ Autoguard™ Shielded IV Catheter is available with or without wings. The catheter hub and wings are color coded to indicate catheter gauge size (24 GA (0.7 mm) = Yellow, 22 GA (0.9 mm) = Blue, 20 GA (1.1 mm) = Pink, 18 GA (1.3 mm) = Green, 16 GA (1.7 mm) = Grey, 14 GA (2.1 mm) = Orange).
BD Insyte™ Autoguard™ Shielded IV Catheter facilitates the introduction of guidewires and other vascular access devices. The following labeled gauge sizes correspond to the following inside diameter (ID): 24 GA (0.40 mm ID), 22 GA (0.55 mm ID), 20 GA (0.7 mm ID), 18 GA (0.8 mm ID), 16 GA (1.2 mm ID), and 14 GA (1.6 mm ID).
The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa).
BD Insyte™ Autoguard™ BC Shielded IV Catheter is an over-the-needle, intravascular (IV) catheter with blood control technology. This device includes a radiopaque BD Vialon™ Catheter Material, a needle, a grip with button, a flash chamber with vent plug, and a barrel. The needle and catheter are protected by a needle cover.
The blood control technology is designed to stop the flow of blood from the catheter hub until the initial Luer connection is made. Once the connection is made, fluids or blood can flow through the catheter hub in either direction.
The 24-18 GA (0.7-1.1 mm) devices have BD Instaflash™ Needle Technology, allowing for immediate visualization of blood along the catheter. The flash chamber provides confirmation that the device has entered the vessel.
BD Insyte™ Autoguard™ BC Shielded IV Catheter incorporate BD Autoguard™ Push-button Shielding Technology which is activated when the button is depressed. A spring mechanism retracts the needle and flash chamber into the barrel, fully encasing the needle and reducing the risk of accidental needlestick injury.
BD Insyte™ Autoguard™ BC Shielded IV Catheter is available with or without wings. The catheter hub and wings are color coded to indicate the catheter gauge size (24 GA (0.7 mm) = Yellow, 22 GA (0.9 mm) = Blue, 20 GA (1.1 mm) = Pink, 18 GA (1.3 mm) = Green, 16 GA (1.7 mm) = Grey).
The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa).
BD Insyte™ Autoguard™ BC Pro Shielded IV Catheter is an over-the-needle, intravascular (IV) catheter with blood control technology. This device includes a radiopaque BD Vialon™ Catheter Material, a needle, a grip with button, a flash chamber with vent plug, and a barrel. The needle and catheter are protected by a needle cover.
The blood control technology is designed to stop the flow of blood from the catheter hub until the initial Luer connection is made. Once the connection is made, fluids or blood can flow through the catheter hub in either direction.
The 24-18 GA (0.7-1.1 mm) devices have BD Instaflash™ Needle Technology, allowing for immediate visualization of blood along the catheter. The flash chamber provides confirmation that the device has entered the vessel.
BD Insyte™ Autoguard™ BC Pro Shielded IV Catheter incorporate BD Autoguard™ Push-button Shielding Technology which is activated when the button is depressed. A spring mechanism retracts the needle and flash chamber into the barrel, fully encasing the needle and reducing the risk of accidental needlestick injury.
BD Insyte™ Autoguard™ BC Pro Shielded IV Catheter is available with or without wings. The catheter hub and wings are color coded to indicate the catheter gauge size (24 GA (0.7 mm) = Yellow, 22 GA (0.9 mm) = Blue, 20 GA (1.1 mm) = Pink, 18 GA (1.3 mm) = Green, 16 GA (1.7 mm) = Grey).
The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa).
This document is a 510(k) clearance letter for three IV catheter devices, indicating that they are substantially equivalent to a previously cleared predicate device (K201075). It describes the changes made (new performance specifications and revised IFU) and provides a summary of performance tests conducted.
However, it does not contain the detailed acceptance criteria and reported device performance information that you specifically asked for in a table, nor does it describe a study that "proves the device meets the acceptance criteria" in the format of a clinical trial or algorithm validation study. The document primarily focuses on demonstrating substantial equivalence to a predicate device, which often relies on bench testing and comparisons of technological characteristics, rather than extensive clinical studies or AI algorithm performance validation.
Therefore, I cannot provide a table of acceptance criteria and reported device performance from this document because it is not present. Similarly, information regarding sample sizes for test/training sets, data provenance, expert ground truth adjudication for AI, MRMC studies, standalone algorithm performance, or training set ground truth establishment is not applicable or not provided within this 510(k) clearance document, as it pertains to a physical medical device (IV catheter) and not an AI/ML-driven device.
The study described to demonstrate substantial equivalence consists of bench testing to verify new performance specifications and ensuring compliance with relevant ISO and ASTM standards.
Here's an overview of the information that is available in relation to performance:
1. Table of Acceptance Criteria and Reported Device Performance:
- Not provided in this document. The document states that "the subject device met all predetermined acceptance criteria for the above-listed performance tests," but it does not specify what those acceptance criteria were (e.g., specific thresholds or ranges for frequency response, kink resistance, or blood fill time). It also does not present the numerical results ("reported device performance") of these tests.
2. Sample size used for the test set and the data provenance:
- Not applicable / Not provided. The document describes bench testing of physical devices, not an AI/ML system or a clinical study with a "test set" of patients or data in the context of AI validation.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable / Not provided. This information is relevant for AI/ML device validation and is not present for this physical medical device.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable / Not provided. This information is relevant for AI/ML device validation and is not present for this physical medical device.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This is not an AI/ML device, so an MRMC study is not relevant.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- No. This is not an AI/ML device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Not applicable / Not provided. For the performance tests listed (Frequency Response, Kink Resistance, Blood Fill Time, Catheter Separation Force), the "ground truth" would be the direct measurements obtained from the physical properties of the devices themselves, compared against pre-defined engineering specifications.
8. The sample size for the training set:
- Not applicable / Not provided. This is not an AI/ML device.
9. How the ground truth for the training set was established:
- Not applicable / Not provided. This is not an AI/ML device.
Summary of Performance Tests and Compliance (as described in the document):
The document states that a risk analysis was conducted to assess the impact of modifications. When technological characteristics were identical to the predicate, results from the predicate were applied. For other aspects, the following tests were conducted:
BD Internal Requirements:
- Frequency Response Testing: To support blood pressure monitoring indications.
- Kink Resistance Testing: To support blood pressure monitoring indications.
- Blood Fill Time Testing: To support blood sampling indications.
- Catheter Separation Force: (No specific indication mentioned, but likely related to structural integrity during withdrawal)
Compliance with Standards Testing:
- Luer Testing: According to ISO 80369-7 (Small-bore connectors for liquids and gases in healthcare applications).
- Packaging Testing: According to ISO 11607-1, ASTM F2096-11 (for gross leaks), and ASTM F88/F88M-15 (for seal strength).
- Biocompatibility: According to ISO 10993-1 (general biological evaluation) and ISO 10993-5 (for in vitro cytotoxicity).
Conclusion: The document explicitly states: "Per the design control requirements specified in 21 CFR §820.30, the subject device met all predetermined acceptance criteria for the above-listed performance tests, demonstrating substantial equivalence to the predicate devices." This indicates that the tests were performed and the results aligned with the internal and regulatory specifications, affirming the device's substantial equivalence.
Ask a specific question about this device
(89 days)
FOZ
BD Nexiva™ Diffusics™ Closed IV Catheter System is intended to be inserted into a patient's peripheral vascular system for short term use to sample blood, monitor blood pressure or administer fluids. This device may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. This device is suitable for use with power injectors set to a maximum pressure of 325 psi (2240 kPa) when access ports not suitable for use with power injectors are removed.
BD Nexiva™ Diffusics™ Closed IV Catheter Systems (Nexiva Diffusics) and BD Nexiva™ Diffusics™ Closed IV Catheter System with BD MaxZero™ Needle-free Connector (Nexiva Diffusics with Max Zero) are over-the-needle, intravascular (IV) catheters. These devices have a radiopaque BD Vialon™ Catheter Material with three side holes located near the tip of the catheter which are designed to optimize power injection procedures. These devices also have a needle, needle shield, septum, stabilization platform, integrated extension tubing, clamp, Luer connector, and vent plug. The Luer connector displays gauge-specific maximum flow rate and the maximum power injector pressure limit setting. The needle and catheter are protected by a needle cover. An end cap with protective cover is provided in the unit package.
A BD MaxZero™ device with protective cover is provided in the unit package for BD Nexiva™ Diffusics™ Closed IV Catheter System with BD MaxZero™ Needle-free Connector kit. The closed system is designed to keep blood contained within the device throughout the insertion process. The septum is designed to wipe visible blood from the needle surface as the needle is withdrawn from the catheter, further reducing the risk of blood exposure. The needle tip is passively protected when the needle is removed, reducing the risk of accidental needle stick injury.
These devices have BD Instaflash™ Needle Technology, allowing for immediate visualization of blood along the catheter. Continuous blood return is seen in the extension tubing. The vent plug prevents blood leakage from the extension tubing during insertion. The stabilization platform and Luer connector are color coded to indicate catheter gauge size (24 GA (0.7 mm) = Yellow, 22 GA (0.9 mm) = Blue, 20 GA (1.1 mm) = Pink, 18 GA (1.3 mm) = Green).
The provided FDA 510(k) clearance letter and summary for the BD Nexiva™ Diffusics™ Closed IV Catheter System does not contain information related to software, artificial intelligence, or diagnostic imaging. Therefore, it is impossible to describe acceptance criteria and study details related to a device using AI, ground truth, expert readers, or sample sizes for training/test sets as requested in the prompt.
The document pertains to a physical medical device (an IV catheter system) and its substantial equivalence to predicate devices, focusing on:
- Indications for Use: Changes to include "blood pressure monitoring indication" for one of the devices.
- Physical Characteristics: Materials, dimensions, sterilization, and packaging components.
- Performance Testing: Mechanical and functional tests relevant to IV catheters (frequency response, kink resistance, blood fill time, luer lock performance, packaging integrity).
There is no mention of:
- A device that uses AI.
- Any form of image analysis, signal processing, or diagnostic output that would require a "test set," "ground truth," "expert consensus," "human-in-the-loop performance," or "MRMC comparative effectiveness study."
- Specific acceptance criteria percentages for sensitivity, specificity, or similar metrics typically associated with AI/diagnostic device performance.
- Training or test data sets, data provenance, or the number/qualifications of experts for labeling or adjudication.
The questions in the prompt are designed for a different type of medical device submission, specifically those involving software as a medical device (SaMD) or AI/machine learning components for diagnostic or prognostic purposes.
Therefore, I cannot fulfill the request as the provided text does not contain the necessary information.
Ask a specific question about this device
(228 days)
FOZ
The RUBY Intravascular Catheter is inserted into a patient's vascular system to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be used with consideration given to adequacy of vascular anatomy, appropriateness of the solution being infused, and duration of therapy. The RUBY Intravascular Catheter is suitable for use with power injectors.
The RUBY Intravascular Catheter is intended to be inserted in the patient's vascular system for short-term use to sample blood, monitor blood pressure, or administer fluids intravenously.
The RUBY Intravascular Catheter is indicated for use in pediatric patients (with body mass higher or equal to 10kg) and adults.
The RUBY™ Intravascular Catheter consists of a radiopaque catheter having a usable length of 1.30 inches in 20-gauge size. The device is a single use, sterile intravascular catheters designed to be inserted into a patient's vascular system to sample blood, monitor pressure, or administer fluids intravenously. The RUBY™ Intravascular Catheter's hub has a blood control valve mechanism that automatically activates to stop the blood flow in the catheter hub when the needle is removed from the catheter during initial insertion by the clinician. Blood flow from the catheter is restricted immediately after needle retraction until a secure luer connection has been made. The RUBY™ Intravascular Catheter is provided with a catheter guiding element with an atraumatic tip design positioned within the formed lumen between the catheter and the needle and connected to a slider used to deploy and retract the catheter guiding element; a flashback chamber to enhance flashback visualization, and a safety container that prevents sharp injuries.
This document is an FDA Clearance Letter for a medical device (Ruby Intravascular Catheter), not a study report or clinical trial. Therefore, it does not contain the detailed information required to answer your prompt about acceptance criteria and study findings related to AI/algorithm performance.
The provided text focuses on:
- Regulatory Clearance: Explaining that the device is substantially equivalent to a predicate device, allowing it to be marketed.
- Device Description: What the Ruby Intravascular Catheter is and how it functions.
- Indications for Use: For what medical purposes the device can be used (e.g., blood sampling, fluid administration).
- Comparison to Predicate: Highlighting similarities and differences between the Ruby Catheter and a previously cleared device (AccuCath).
- Safety and Performance Tests (General): Listing the types of tests performed (e.g., performance, biocompatibility, sterilization, shelf-life and the standards they followed (e.g., ISO, ASTM).
Crucially, there is no mention of any AI, algorithm, or software component within the Ruby Intravascular Catheter. The device is a physical medical instrument (a catheter), and the testing described pertains to its physical and biological properties.
Therefore, I cannot provide a table of acceptance criteria for algorithm performance, sample sizes for an AI test set, expert qualifications for AI ground truth, or details about MRMC studies, as these concepts are not applicable to the information given in the FDA clearance letter for this specific device.
To answer your prompt with the requested details, the input would need to be a clinical study report or a 510(k) submission summary specifically for a software as a medical device (SaMD) or a device with a significant AI/algorithm component.
If you have a document related to an AI-powered medical device, please provide that, and I will do my best to extract the relevant information.
Ask a specific question about this device
(91 days)
FOZ
OSPREY Midline Closed IV Catheter System (OspreyEDC) is an intravascular catheter intended to be inserted into the patient’s vascular system for short-term use to sample blood, monitor blood pressure, or administer fluids intravenously. Blood is contained within the device during the catheter insertion process aiding in the prevention of blood exposure. This catheter may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of procedure.
The OSPREY Midline Closed IV Catheter System (OspreyEDC) is suitable for use with power injectors rated for a maximum of 325 psi when connected to male luer lock.
The OSPREY Midline Closed IV Catheter System (OspreyEDC) has a usable (deployed) catheter length of 3.25 inches. The device is a single use, sterile intravascular catheter designed to be inserted into a patient's vascular system to sample blood, monitor blood pressure, or administer fluids intravenously for short-term (
The provided FDA 510(k) clearance letter and summary discuss a medical device, the OSPREY Midline Closed IV Catheter System (OspreyEDC-F20), and its substantial equivalence to a predicate device. This document primarily focuses on demonstrating that the new device is as safe and effective as a legally marketed device, rather than providing detailed acceptance criteria and a study proving those criteria are met for an AI/software-as-a-medical-device (SaMD).
Here's an analysis based on the provided text, highlighting what information is available and what is not available regarding the specific questions about acceptance criteria and a study proving device performance in the context of AI/SaMD:
Summary of Acceptance Criteria and Device Performance (Based on provided document):
The document does not present explicit "acceptance criteria" in the format one might expect for a software or AI product's performance metrics (e.g., sensitivity, specificity, AUC). Instead, it demonstrates substantial equivalence by comparing the new device's technological characteristics and performance outcomes against a predicate device and relevant industry standards. The implicit "acceptance criteria" are that the new device performs at least as well as the predicate and meets established safety and performance standards.
Acceptance Criteria Category (Derived) | Stated Device Performance (OSPREY Midline Closed IV Catheter System) | Substantial Equivalence to Predicate |
---|---|---|
Indications for Use | Short-term IV catheter for blood sampling, pressure monitoring, fluid administration; aids in blood exposure prevention; suitable for all patient populations (with considerations); Power Injector compatible (max 325 psi). | Identical |
Classification | 21 CFR §880.5200, Class II, FOZ - Intravascular Catheter | Identical |
Critical Procedural Steps | Expanded IFU to include removal of pink wire slider, wire slider track, and guidewire components. | Different, but does not raise new safety/effectiveness questions. |
Materials of Construction | Polycarbonate, Stainless Steel, Polyurethane w/ radiopaque barium sulfate, Loctite, Polypropylene, Tygon®, Porous Polyethylene w/ CMC, Silicone. Additional components: Wire Slider Track, Wire Slider Cap, Guidewire (Stainless Steel). | Different, but materials are identical for common components and new materials passed biocompatibility. Does not raise new safety/effectiveness questions. |
Design Characteristics | Catheter-Needle Interface: Catheter through needle; Beveled Needle Tip; Rounded Catheter Tip; Spring-loaded needle retraction; Female Locking Luer Hub; Flashback Visualization. Catheter OD: 0.0435 – 0.0445 in. Catheter ID: 0.0275 – 0.0325 in. Catheter Length: 3.25 in. Integrated extension tube, pinch clamp, female luer with porous plug. Guidewire slider/track interface. Power Injector compatible (325 psi). | Different (catheter OD/ID/length slightly different, new guidewire mechanism). Does not alter or raise different questions of safety and effectiveness. |
Performance (Functional) | Flashback Chamber / Technology: Yes; Sharps Prevention Feature: Yes; Radiopaque: Yes; Flow Rate: 30 mL/min | Identical |
Biocompatibility | Passed ISO 10993-1 testing | Identical (PASSED) |
Sterilization | EtO Sterilized | Identical |
Packaging | Sterile Barrier: Individual Tyvek and PET Pouches | Identical |
Shelf Life | 6 Months | Identical |
Since this is a filing for a physical medical device (an IV catheter system) and not an AI/SaMD, many of the requested items related to AI model validation, such as test sets, ground truth establishment, expert adjudication, and training set details, are not applicable and therefore not present in the document.
Here's a breakdown of the specific points from your request:
-
A table of acceptance criteria and the reported device performance
- Available in document: Yes, a comparison table (Table VII - I) is provided on pages 7-11, outlining attributes of the subject device (OSPREY Midline Closed IV Catheter System) against the predicate device. This indirectly serves as acceptance criteria in the context of substantial equivalence for a physical medical device. The "Performance" section on page 12 also lists standards adhered to via bench testing.
- Not explicitly stated as AI/SaMD performance metrics (e.g., sensitivity, specificity): This is a hardware device, so these metrics are not relevant or provided.
-
Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not applicable/Not available in document: This information is typically for AI/SaMD. For a physical device, testing involves bench tests and biocompatibility, not data sets in the AI sense. No specific "sample sizes" for clinical data or data provenance are mentioned, as clinical studies in the AI context were not conducted or required for this type of device and submission. The document mentions "bench tests," but doesn't specify sample sizes for those tests.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable/Not available in document: "Ground truth" in this context refers to clinical data labeling for AI. For a physical device like a catheter, "ground truth" is established through engineering and biocompatibility standards, and performance against those standards is evaluated by engineers and lab personnel, not necessarily "experts" in the clinical decision-making sense with specific experience levels for data labeling.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable/Not available in document: This is relevant for clinical data labeling for AI performance evaluation.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable/Not available in document: This is entirely for AI-assisted diagnostic devices. This is a standalone physical medical device.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable/Not available in document: This pertains to AI algorithm performance.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Not applicable/Not available in document: As mentioned, for this physical device, "ground truth" is based on adherence to recognized industry standards (ISO, ASTM, USP) and established engineering principles and test methods (e.g., flow rate measurement, physical dimensions, material properties, biocompatibility assays).
-
The sample size for the training set
- Not applicable/Not available in document: This applies to AI models.
-
How the ground truth for the training set was established
- Not applicable/Not available in document: This applies to AI models.
In conclusion, the provided document is a standard 510(k) submission for a physical medical device. It successfully demonstrates substantial equivalence to a predicate device by comparing design, materials, indications for use, and performance based on engineering bench tests and adherence to recognized international standards for medical devices. The specific questions you asked are geared towards the evaluation of Artificial Intelligence/Software as a Medical Device (AI/SaMD), which is not the subject of this 510(k) clearance letter.
Ask a specific question about this device
(86 days)
FOZ
The HydroMID 5F Dual Lumen Midline Catheter is indicated for short term access (
The HydroMID catheter is a 5 French, dual lumen, midline catheter comprised of a radiopaque, hydrophilic catheter with a suture wing, Luer lock hub, and extension tubing made from materials commonly used in the manufacture of catheters. Catheters are provided packaged in kit configurations with the appropriate accessories for placement in the respective clinical environments.
HydroMID has been shown to be effective in reducing thrombus accumulation and thrombotic occlusions. Reduction of thrombus accumulation and thrombotic occlusions were evaluated using in vitro and in vivo models. Pre-clinical in vivo evaluations do not necessarily predict clinical performance with respect to thrombus formation.
Catheter and accessories are supplied sterile. Product is sterilized by ethylene oxide (EO) except where labeled otherwise. Product should be stored in a cool, dry, dark place. The HydroMID Catheter is supplied in a dry state and must be hydrated before use.
The provided text describes a 510(k) summary for the HydroMID 5F Dual Lumen Midline Catheter. It focuses on demonstrating substantial equivalence to a predicate device through bench testing. The document does not contain information about a study proving the device meets acceptance criteria in the context of an AI/human reader study or clinical performance with ground truth established by experts/pathology/outcomes data.
The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this document refer to engineering and performance bench testing of a physical medical device (a catheter), not a medical device software or AI algorithm. Therefore, many of the requested categories in the prompt (e.g., number of experts for ground truth, adjudication methods, MRMC study, standalone algorithm performance, training set details) are not applicable to this type of device submission.
However, I can extract the information relevant to the device's performance and the "acceptance criteria" as described in the 510(k) summary for this physical medical device.
1. Table of Acceptance Criteria and Reported Device Performance (Bench Testing):
The document states: "All tests passed the acceptance criteria specified for both the subject and predicate devices." It does not provide specific numerical acceptance criteria or numerical reported performance values for each test. Instead, it lists the types of tests conducted.
Test Type | Acceptance Criteria (General Statement) | Reported Device Performance (General Statement) |
---|---|---|
Power injection flow rate | Not explicitly detailed, but implied to meet an acceptable standard. | Passed; supports a 6ml/s rating (higher than predicate's 3.5ml/s). |
Static burst strength | Not explicitly detailed, but implied to meet an acceptable standard. | Passed. |
Catheter length | Not explicitly detailed, but implied to meet an acceptable standard. | Passed. |
Dimensional verification | Not explicitly detailed, but implied to meet an acceptable standard. | Passed. |
Catheter kink/resistance | Not explicitly detailed, but implied to meet an acceptable standard. | Passed. |
Tensile (tubing and joint) | Not explicitly detailed, but implied to meet an acceptable standard. | Passed. |
Particulate | Not explicitly detailed, but implied to meet an acceptable standard. | Passed. |
Packaging distribution | Not explicitly detailed, but implied to meet an acceptable standard. | Passed. |
Sterility | Not explicitly detailed, but implied to meet an acceptable standard. | Passed. (Sterilized by ethylene oxide, as stated in Device Description) |
MR compatibility | Not explicitly detailed, but implied to meet an acceptable standard (MRI Conditional). | Passed. (MRI Conditional) |
Shelf-life | Not explicitly detailed, but implied to meet an acceptable standard. | Passed. (Performance demonstrated at T=0 and over shelf life). |
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size: The document states: "Testing was identical to the predicate in standard conformance, sample size, methods, and acceptance criteria." However, the specific numerical sample size for each test is not provided.
- Data Provenance: The tests are described as "Bench Data" conducted internally by the manufacturer, Access Vascular Inc. The country of origin for the data is not explicitly stated, but the company address is Billerica, MA, which implies the testing was conducted in the United States. The data is prospective in the sense that it was generated for this 510(k) submission to demonstrate the performance of the new device design.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Experts:
- Not Applicable. This is a physical medical device. Ground truth for its performance is established by engineering and material science testing against predefined specifications, not by human expert review of images or clinical outcomes in the typical sense of AI/image analysis.
4. Adjudication Method for the Test Set:
- Not Applicable. See point 3. Testing results are objective measurements against established engineering standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs. without AI assistance:
- Not Applicable. This is a physical medical device (catheter), not a software or AI device intended to assist human readers.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not Applicable. This is a physical medical device, not an algorithm.
7. The Type of Ground Truth Used:
- Engineering Specifications and Standardized Test Methods: The "ground truth" for this device's performance revolves around its physical and functional properties meeting predetermined engineering specifications and validated test methods (e.g., pressure resistance, flow rates, dimensional accuracy, material properties like tensile strength and kink resistance, sterility, and MRI compatibility). These are objective criteria, not based on expert consensus, pathology, or outcomes data in a clinical trial sense.
8. The Sample Size for the Training Set:
- Not Applicable. This is a physical medical device; there is no "training set" in the context of machine learning. The device design and materials are based on established engineering principles and prior device iterations.
9. How the Ground Truth for the Training Set Was Established:
- Not Applicable. See point 8.
Ask a specific question about this device
(113 days)
FOZ
The Arrow™ Endurance™ Extended Dwell Peripheral Catheter System permits access to the patient's peripheral vascular system for short-term venous use (less than 30 days) to sample blood, administer fluids, and perform high pressure contrast injections at a maximum of 325 psi.
The Arrow™ Endurance™ Extended Dwell Peripheral Catheter System permits access to the patient's peripheral vascular system for short-term use (less than 30 days) to facilitate arterial blood pressure measurement and blood sampling.
The safety feature is intended to minimize the risk of sharps injuries.
The Arrow™ Endurance™ Extended Dwell Peripheral Catheter may be used for any patient population with consideration given to adequacy of anatomy and appropriateness of the procedure.
The Arrow™ Endurance™ Extended Dwell Peripheral Catheter may be used in hospitals, clinics, and other advanced clinical facilities.
The subject device, Arrow™ Endurance™ Extended Dwell Peripheral Catheter System, is a sterile, single use peripheral intravascular device designed to permit access to the peripheral vascular system. The Endurance catheter is intended to permit access to the patient's peripheral vascular system for short-term venous use to sample blood, administer fluids and for high pressure contrast injections at a maximum of 325 psi. The Arrow™ Endurance™ Extended Dwell Peripheral Catheter System permits access to the patient's peripheral vascular system for short-term arterial use to monitor arterial blood pressure and blood sampling. The safety feature is intended to minimize the risk of sharps injuries.
The Arrow™ Endurance™ Extended Dwell Peripheral Catheter System is available in a single lumen, 18 gauge configuration with usable length of 8 cm (3.15").
This document is a 510(k) summary for the Arrow™ Endurance™ Extended Dwell Peripheral Catheter System, seeking substantial equivalence to a predicate device. As such, it focuses on demonstrating equivalence through comparison of technological characteristics and nonclinical bench testing. It does not describe acceptance criteria and a study that proves the device meets those criteria in the context of an AI-powered medical device.
The information provided confirms that the device is a physical intravascular catheter, not an AI/software device. Therefore, the specific requirements listed in the prompt (e.g., sample size for test/training sets, expert ground truth establishment, MRMC studies, standalone algorithm performance) are not applicable to the content of this document.
The document highlights:
- Acceptance Criteria (Implied by equivalence): The "acceptance criteria" here are implicitly that the new device performs equivalently to the predicate device in relevant physical and functional characteristics, and that any differences do not raise new questions of safety or effectiveness.
- Reported Device Performance: This is demonstrated through a detailed comparison table (Section 7) and a list of "Nonclinical Testing" (Section 8).
Here's a breakdown of what is available and what is not relevant from your prompt:
1. Table of Acceptance Criteria and Reported Device Performance:
The document doesn't provide a typical "acceptance criteria" table with numerical targets as would be seen for an AI device. Instead, it presents a comparison to a predicate device. The "acceptance" is based on demonstrating that the subject device is "substantially equivalent" to the predicate, meaning it performs similarly and does not introduce new safety or effectiveness concerns.
Feature / Characteristic | Subject Device Performance (Reference to K243599) | Predicate Device Performance (Reference to K163513) | Assessment of Device Differences / Meeting "Acceptance" |
---|---|---|---|
Classification Name | Catheter, intravascular, therapeutic, short-term less than 30 days | Catheter, intravascular, therapeutic, short-term less than 30 days | Same |
Product Code | FOZ | FOZ | Same |
Regulation Number | 880.5200 | 880.5200 | Same |
Class | II | II | Same |
Intended Use | Access to patient's peripheral vascular system for short-term venous or short-term arterial use. | Access to patient's peripheral vascular system for short-term venous or short-term arterial use. | Same |
Indications for Use | Detailed venous (blood sample, fluids, high pressure contrast @ 325 psi) and arterial (BP measurement, blood sampling) use. Patient population/usage environment specified. | Less granular: sample blood, monitor blood pressure, administer fluids, high pressure injection. | Different granularity; deemed similar, no new safety/effectiveness questions. |
Single Use | Yes | Yes | Same |
Duration of Use | Less than 30 days | Less than 30 days | Same |
Insertion Technique | Guidewire, catheter advance, remove needle/guidewire, deploy safety. | Guidewire, catheter advance, remove needle/guidewire, deploy safety. | Same |
Principle of Operation | Closed fluid path system catheter. | Closed fluid path system catheter. | Same |
Shelf Life | 6 months | 2 years | Different, justified by business needs and bench testing. |
MR Safety | MR Safe (catheter) | MR Safe (catheter) | Same |
Device Components | Guard, Handle, Advancer, Slider, Needle Supports, Needle Safety, Extension Line Clamp, Needle, Guide Wire, Juncture Hub, Catheter with Extension Line. | Same components. | Same |
Device Materials (Key Differences Highlighted) | Catheter Body: Polyurethane with Silicone Coating (Quadraflex Polyurethane). | Catheter Body: Polyurethane (Tecoflex Polyurethane). | Different; biocompatibility and bench testing support no new safety/effectiveness concerns. |
Catheter Design | Ergonomically designed handle, echogenic needle, passive needle safety, needle support, guidewire with slider. | Same description. | Same |
Catheter Body OD | 18 Ga | 18 Ga, 20 Ga, 22 Ga | Same (for 18 Ga), excluding 20 & 22 Ga. |
Catheter Body ID | 0.039" (18 Ga) | 0.039" (18 Ga), 0.032" (20 Ga), 0.027" (22 Ga) | Same (for 18 Ga), excluding 20 & 22 Ga. |
Catheter Usable Length | 8 cm (3.15") | 6 cm (2.36") (20 Ga, 22 Ga), 8 cm (3.15") (18 Ga, 20 Ga, 22Ga) | Same (for 8cm), excluding 6cm. |
Needle Safety Feature | Yes | Yes | Same |
Blood Safety Feature | Bloodless (seal and extension lines) | Bloodless (seal and extension lines) | Same |
Pressure Injection Limits | 325 psi | 325 psi | Same |
Sidearm Clamp | Pinch | Slide | Different; bench testing supports no new safety/effectiveness concerns. |
Juncture Hub Advancer | Half Circle Suture Wing Posts, Removed Hub Nose Clips. | Round Suture Wing Posts, Hub Nose Clips. | Different; bench testing supports no new safety/effectiveness concerns. |
Handle | Lower needle support, No Catheter Release Tab. | No Lower Needle Support, Catheter Release Tab. | Different; bench testing supports no new safety/effectiveness concerns. |
Sterile | Yes | Yes | Same |
Sterilization Method | Ethylene Oxide | Ethylene Oxide | Same |
Biocompatibility | Biocompatible (per ISO 10993-1), Prolonged Contact: Circulating Blood. | Biocompatible (per ISO 10993-1), Prolonged Contact: Circulating Blood. | Same |
2. Sample sized used for the test set and the data provenance:
- This is not an AI/software device, so there isn't a "test set" in the sense of a dataset for algorithm evaluation.
- The performance is evaluated through bench testing (listed in Section 8). Details on the specific sample sizes for these physical and material tests are not provided in this summary.
- Data provenance: Not applicable as it's not clinical data. The tests are laboratory-based, performed on manufactured devices.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. This is not an AI/software device requiring expert labeling of data.
- The "ground truth" for a physical device is established through engineering specifications, material science, and performance standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable. This concept is for resolving disagreements in expert annotations for AI datasets, not for physical device testing.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is a catheter, not an AI software intended to assist human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is not an AI algorithm. Its performance is inherent in its physical properties and design.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- For physical devices like this, the "ground truth" for validating performance is typically derived from:
- Engineering specifications and design requirements: Does it meet the intended physical dimensions, material properties, and functional outputs (e.g., flow rate, pressure limits)?
- Recognized consensus standards: Compliance with ISO standards related to biocompatibility, sterilization, and general medical device safety.
- Bench testing results: Data from rigorous laboratory tests simulating various use conditions.
8. The sample size for the training set:
- Not applicable. This is not a machine learning model.
9. How the ground truth for the training set was established:
- Not applicable. This is not a machine learning model.
In summary, this 510(k) pertains to a physical medical device (catheter), not an AI/software product. Therefore, most of the requested information regarding AI device evaluation (test/training sets, expert ground truth, MRMC studies) is not relevant to this document. The "study" proving the device meets acceptance criteria is the comprehensive nonclinical bench testing outlined in Section 8, which demonstrates that the device performs functionally and safely, and is substantially equivalent to its predicate.
Ask a specific question about this device
(140 days)
FOZ
The Arrow™ Endurance™ Extended Dwell Peripheral Catheter System permits access to the patient's peripheral vascular system for short-term venous use (less than 30 days) to sample blood, administer fluids, and perform high pressure contrast injections at a maximum of 325 psi.
The Arrow™ Endurance™ Extended Dwell Peripheral Catheter System permits access to the patient's peripheral vascular system for short-term use (less than 30 days) to facilitate arterial blood pressure measurement and blood sampling.
The safety feature is intended to minimize the risk of sharps injuries.
The Arrow™ Endurance™ Extended Dwell Peripheral Catheter may be used for any patient population with consideration given to adequacy of anatomy and appropriateness of the procedure.
The Arrow™ Endurance™ Extended Dwell Peripheral Catheter may be used in hospitals, clinics, and other advanced clinical facilities.
The subject device, Arrow™ Endurance™ Extended Dwell Peripheral Catheter System, is a sterile, single use peripheral intravascular device designed to permit access to the peripheral vascular system. The Endurance catheter is intended to permit access to the patient's peripheral vascular system for short-term venous use to sample blood, administer fluids and for high pressure contrast injections at a maximum of 325 psi. The Arrow™ Endurance" Extended Dwell Peripheral Catheter System permits access to the patient's peripheral vascular system for short-term arterial use to monitor arterial blood pressure and blood sampling. The safety feature is intended to minimize the risk of sharps injuries.
The Arrow™ Endurance™ Extended Dwell Peripheral Catheter System is available in single lumen, 20 and 22 gauge configurations with usable lengths of 6 cm (2.36'') and 8 cm (3.15").
The provided text is a 510(k) Summary for the Arrow™ Endurance™ Extended Dwell Peripheral Catheter System, detailing its substantial equivalence to a predicate device. This document focuses on the physical and functional aspects of a medical device (intravascular catheter) and its comparison to a similar existing device. It does not contain information related to an AI/ML powered medical device, a multi-reader multi-case (MRMC) comparative effectiveness study, or details about ground truth establishment for a diagnostic algorithm.
Therefore, I cannot extract the specific information requested in your prompt regarding acceptance criteria and studies for an AI/ML device performance, as this information is not present in the provided document. The nonclinical testing listed relates to the physical performance and safety of the catheter itself, not to the diagnostic accuracy of an AI.
To elaborate on why the requested information cannot be found:
- Acceptance Criteria & Reported Device Performance (Table): The document provides a comparison of features, materials, and functional aspects of the subject device against a predicate device. It indicates "Same" or "Different" for many features and then lists various nonclinical bench tests in Section 8. However, it does not present quantitative acceptance criteria or reported performance for an AI/ML output. For example, it doesn't state "AI sensitivity > X%" or "AI accuracy > Y%" with corresponding actual performance.
- Sample Size (Test Set) & Data Provenance: This would be relevant for an AI study involving patient data. The document describes bench testing of a physical catheter, which does not typically involve a "test set" of patient data in the context of AI.
- Number of Experts, Qualifications & Adjudication Method: These points are critical for establishing ground truth in AI studies. Since this device is a physical catheter, there is no AI algorithm generating diagnostic outputs that would require expert adjudication for a ground truth.
- MRMC Comparative Effectiveness Study: This type of study assesses how AI affects human reader performance. As the device is not an AI, this study type is not applicable.
- Standalone Performance: This refers to the AI algorithm's performance without human intervention. Again, not applicable to a physical catheter.
- Type of Ground Truth: For an AI device, ground truth might be pathology, long-term outcomes, or expert consensus. For this physical catheter, "ground truth" is established through engineering specifications, material properties, and physical performance benchmarks.
- Sample Size for Training Set & How Ground Truth for Training Set was Established: These are fundamental elements of AI model development. Since this is not an AI device, there is no "training set" or corresponding ground truth establishment process in the context of machine learning.
In summary, the provided document describes the regulatory clearance for a physical medical device (intravascular catheter) through a substantial equivalence pathway, not an AI/ML-powered diagnostic or assistive device. Therefore, the information requested in your prompt is not available within this text.
Ask a specific question about this device
(218 days)
FOZ
The SV Spectrum MRC Central Venous Catheter is indicated to provide short-term (
The SV Spectrum MRC Central Venous Catheter (CVC) is a single-use, antibiotic impregnated, antimicrobial coated, power injectable catheter that is percutaneously inserted into the vasculature using the Seldinger technique, and advanced over a wire guide until its tip is positioned above the superior vena cava-right atrium (SVC-RA) junction within the lower third of the SVC. The SV Spectrum MRC Central Venous Catheter, 7 Fr triple lumen CVC, is made from an aliphatic polyether-based polyurethane catheter shaft with a radiopaque constituent and a bonded soft distal tip. The catheter shaft is impregnated with Spectrum® (minocycline, rifampin) antimicrobial agents and coated with chlorhexidine antiseptic. The SV Spectrum MRC Central Venous Catheter is not intended for treatment of existing infections in patients; the presence of the three antimicrobial agents has been shown to reduce microbial colonization of the catheter.
The SV Spectrum MRC CVC has demonstrated efficacy against the following organisms:
- Staphylococcus aureus
- Staphylococcus epidermidis
- Enterococcus faecalis
- Escherichia coli
- Klebsiella pneumoniae
- Acinetobacter baumanni
- Candida albicans
- Candida glabrata
The effectiveness was evaluated using in vitro methods. The antimicrobial activity of these agents is localized to the internal and external catheter surfaces.
The working length (15, 20, or 25 cm) of the catheter is measured from distal tip to the double bar line approximately 1.5 cm distal to the manifold. The catheter shaft features centimeter length markings. The proximal portion of the SV Spectrum MRC Central Venous Catheter, like the predicate catheter, has three extension tubes with injection-molded Luer hubs connected to a manifold. The injection-molded manifold with suture wings serves as the transition point for lumens of the extension tubes into each lumen of the catheter shaft. All three lumens are power injectable; the maximum flow rate is labeled on the hub of each extension tube.
The SV Spectrum MR Central Venous Catheter is identical to the SV Spectrum MRC Central Venous Catheter except that it is not coated with chlorohexidine antiseptic.
The SV Central Venous Catheter is identical to the SV Spectrum MRC Central Venous Catheter and SV Spectrum MR Central Venous Catheter except it is not impregnated with minocycline or rifampin, nor is it coated with chlorhexidine antiseptic.
This FDA 510(k) clearance letter pertains to a central venous catheter and, as such, does not contain information about the acceptance criteria and study that proves the device meets acceptance criteria for an AI/ML-driven medical device.
The document states:
- "Software: Not applicable. The device contains no software." (Page 18)
- The performance testing for this device focuses on mechanical and shelf-life properties, biocompatibility, and antimicrobial efficacy (for certain versions), all of which are standard for physical medical devices and unrelated to AI/ML performance.
Therefore, I cannot provide the requested information based on the provided text. The prompt asks for details specifically relevant to AI/ML device performance and testing (e.g., sample size for training/test sets, expert adjudication methods, MRMC studies, standalone algorithm performance, ground truth establishment) which are not present in this document as it's for a non-software, non-AI medical device.
Ask a specific question about this device
Page 1 of 27