(114 days)
TINGS FINDERS 2 Color Doppler Diagnostic Ultrasound System is intended for intracardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult patients.
Modes of operation include: B,M, Color Doppler,PW Doppler, CW Doppler,Power Doppler; Combined modes: B/M,B/Color,B/M/Color, B/PWD, B/CWD, B/Color/PWD, B/Color/CWD, B/Power.
The device is intended for use in Professional healthcare facility environment including echo lab, other hospital settings, operating room, Cath lab and EP lab.
The intended users of the Device is to be used exclusively by physicians who are appropriately trained personnel in a fully equipped electrophysiology or catheterization room or under their direct supervision.
The proposed TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound System is a Track 3 diagnostic device designed for imaging intracardiac anatomical structures and blood flow dynamics in adult patients. It comprises two primary components: the ultrasound console and the TINGSN Sonic Eyes 10 Eco Disposable Intracardiac Echocardiography Catheter. The console features a 17.3-inch touch-enabled LCD screen, a control panel, and interfaces for connectivity, powered by a rechargeable battery or AC adapter, with optional peripherals including a cart and foot switch. The catheter, a sterile, single-use device with a 95 cm insertable length and 10Fr diameter, is equipped with a 64-element phased array transducer at its tip for high-resolution imaging. It connects to the console via the Finders 2 Sonic Eyes Connector, featuring a locking mechanism to ensure secure attachment during procedures. The system supports multiple imaging modes—B-Mode, M-Mode, Pulsed Wave (PW) Doppler, Continuous Wave (CW) Doppler, Color Doppler, Power Doppler, and combined modes —along with advanced functionalities, displayed with real-time thermal (TI) and mechanical (MI) indices.
The provided text is a 510(k) Clearance Letter and 510(k) Summary for the TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound System. This document describes the device, its intended use, and a comparison to predicate devices to establish substantial equivalence.
Crucially, the document explicitly states under "8. CLINICAL DATA": "No clinical data was included in this submission."
Therefore, based on the provided text, a study that proves the device meets acceptance criteria through clinical data (either standalone or MRMC comparative effectiveness) was not performed, or at least, not included in this submission. The entire submission relies on non-clinical testing and comparison to predicate devices to establish substantial equivalence.
As the request specifically asks for information about a "study that proves the device meets the acceptance criteria" and mentions clinical outcomes, it is impossible to provide the requested details from the given text.
However, I can extract the information related to the non-clinical tests that were performed to verify the device met design specifications and was substantially equivalent, which serve as the basis for clearance in lieu of clinical data.
Here's an interpretation based on the non-clinical data presented:
Device Acceptance Criteria and Performance (Based on Non-Clinical Data)
The TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound System established its substantial equivalence (SE) based primarily on non-clinical testing and comparison to predicate devices, as no clinical data was included in the submission. The acceptance criteria and "performance" are therefore framed in terms of compliance with relevant standards and demonstration of functional equivalence.
1. Table of Acceptance Criteria and Reported Device Performance (Non-Clinical)
| Acceptance Criteria (Demonstrated via Compliance with Standards / Equivalence) | Reported Device Performance (Demonstrated via Non-Clinical Testing) |
|---|---|
| Safety & Essential Performance: Compliance with general medical electrical equipment safety standards. (IEC 60601-1 Ed 3.2) | Met: Non-clinical tests demonstrated compliance with IEC 60601-1 Edition 3.2. |
| Electromagnetic Compatibility (EMC): Compliance with EMC requirements for medical devices. (IEC 60601-1-2) | Met: Non-clinical tests demonstrated compliance with IEC 60601-1-2. |
| Specific Safety/Performance for Ultrasonic Equipment: Compliance with specific requirements for diagnostic ultrasound. (IEC 60601-2-37) | Met: Non-clinical tests demonstrated compliance with IEC 60601-2-37. |
| Acoustic Output Characterization: Compliance with methods for determining thermal and mechanical indices. (IEC 62359, NEMA UD2) | Met: Non-clinical tests demonstrated compliance with IEC 62359 and NEMA UD2. The device is a "Track 3" device, implying its acoustic output is within regulatory limits (global maximum derated ISPTA ≤720 mW/cm², MI ≤1.9 or ISPPA ≤ 190 W/cm²). |
| Intravascular Catheter Requirements: Compliance with standards for sterile, single-use intravascular catheters. (ISO 10555-1) | Met: Non-clinical tests demonstrated compliance with ISO 10555-1. |
| Biocompatibility: Evaluation and testing within a risk management process for medical devices. (ISO 10993-1) | Met: Non-clinical tests demonstrated compliance with ISO 10993-1 (specifically, Patient contact materials of the catheter shaft are "Pebax series Comply with ISO 10993-1"). |
| Packaging & Shipping Performance: Performance testing of shipping containers and systems. (ASTM D4169-2022) | Met: Non-clinical tests demonstrated compliance with ASTM D4169-2022. |
| Risk Management: Application of risk management to medical devices. (ISO 14971:2019) | Met: A thorough risk analysis was conducted, and risk controls were implemented to mitigate identified hazards. |
| Software Functionality & Safety: Compliance with FDA guidance for device software functions, meeting specifications and mitigating risks. (FDA Guidance for Device Software Functions) | Met: Software documentation included. Non-clinical tests (integration and functional) conducted. Test results support that all software specifications met acceptance criteria. Verification and validation testing found acceptable. |
| Functional Equivalence to Predicate Device (GE Vivid iq console): Similar intended use, imaging modes, principle of operation, and overall performance. | Demonstrated: The proposed device's indications for use are encompassed by the predicate. Technical differences in imaging modes, software, and compatible probes do not impact safety or effectiveness. Risk analysis assessed potential software-related risks. |
| Functional Equivalence to Predicate Device (SOUNDSTAR 3D Ultrasound Catheter): Similar indications for use, principle of operation, patient population, sterility, diameter, and transducer configuration. | Demonstrated: Slight differences in insertable length, imaging frequency range, and shelf life were determined not to impact safety or effectiveness. The proposed device's indications are encompassed by the predicate. |
| Shelf Life: Verification of product stability and packaging integrity for claimed shelf life. | Met: A 2-year shelf-life study was performed, verifying the claimed 2-year shelf life is scientifically reliable. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not explicitly stated for specific non-clinical tests. General statements indicate "non-clinical tests were conducted."
- Data Provenance: The tests were conducted internally by Jiangsu Tingsn Technology Co., Ltd. The document refers to "non-clinical tests (integration and functional) were conducted on the TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound System during product development."
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- Not applicable for non-clinical testing. The "ground truth" for non-clinical tests typically refers to engineering specifications, regulatory standards, and established physical principles, not expert interpretations of clinical data.
4. Adjudication Method for the Test Set
- Not applicable for non-clinical testing. Adjudication methods like 2+1 or 3+1 are used for human expert review of clinical cases to establish ground truth or compare diagnostic performance, which was not part of this submission.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No MRMC study was done. The submission explicitly states: "No clinical data was included in this submission." MRMC studies inherently involve human readers evaluating clinical cases.
6. Standalone (i.e., algorithm only without human-in-the-loop performance)
- Not applicable as an AI/algorithm-only device. The TINGSN FINDERS 2 is a diagnostic ultrasound system that produces images for human interpretation, not an AI algorithm performing diagnosis independently. The software features listed are for image optimization, measurements, and display, assisting the human user, not replacing them in a "standalone" diagnostic role.
7. The Type of Ground Truth Used
- For non-clinical testing: The "ground truth" was established by performance specifications, engineering requirements, and compliance with the international and national standards listed (e.g., IEC, ISO, NEMA, ASTM, FDA guidance). This includes electrical safety, mechanical robustness, image quality parameters (resolution, penetration, consistency), acoustic output levels, and software functionality validation against design requirements.
8. The Sample Size for the Training Set
- Not applicable. This device is a traditional ultrasound system, not an AI/ML product that undergoes "training."
9. How the Ground Truth for the Training Set Was Established
- Not applicable. See point 8.
U.S. Food & Drug Administration 510(k) Clearance Letter
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.08.00
July 19, 2025
Jiangsu Tingsn Technology Co., Ltd.
Jing Wan
Regulatory Affairs
21st Floor, Building B, No. 3-1 Xinjinhu Road
Jiangbei New District
Nanjing, Jiangsu 210032
China
Re: K250913
Trade/Device Name: TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound System
Regulation Number: 21 CFR 870.1200
Regulation Name: Diagnostic Intravascular Catheter
Regulatory Class: Class II
Product Code: OBJ, ITX, IYN, IYO
Dated: June 20, 2025
Received: June 20, 2025
Dear Jing Wan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Page 2
K250913 - Jing Wan Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
Page 3
K250913 - Jing Wan Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Aneesh S. Deoras -S
Aneesh Deoras
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
Indications for Use
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. K250913
Please provide the device trade name(s).
TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound System
Please provide your Indications for Use below.
TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound System is intended for intracardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult patients.
Modes of operation include: B,M, Color Doppler,PW Doppler, CW Doppler,Power Doppler; Combined modes: B/M,B/Color,B/M/Color, B/PWD, B/CWD, B/Color/PWD, B/Color/CWD, B/Power.
The device is intended for use in Professional healthcare facility environment including echo lab, other hospital settings, operating room, Cath lab and EP lab.
The intended users of the Device is to be used exclusively by physicians who are appropriately trained personnel in a fully equipped electrophysiology or catheterization room or under their direct supervision.
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
Please select the types of uses (select one or both, as applicable).
Color Doppler Diagnostic Ultrasound System
Page 5
TINGSN FINDERS 2 510(k) Premarket Notification Submission
510(k) Summary
Submission: K250913
In accordance with 21 CFR 807.92 the following summary of information is provided:
1. Sponsor Identification
Jiangsu Tingsn Technology Co., Ltd.
21st Floor, Building B,No.3-1 Xinjinhu Road,
Jiangbei New District, Nanjing, 210032, P.R. China
Contact Person: Jing Wan
Regulatory Affairs
Jiangsu Tingsn Technology Co., Ltd.
Tel: (+86) -18921028523
Email: wanjing@tingsn.com
Date prepared: March 18, 2025
2. DEVICE
Device Name: TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound System
Common name: Ultrasonic Pulsed Doppler Imaging System
Regulation number: 21 CFR 870.1200
Regulation Class: Class II
Product Code: OBJ,IYN,IYO,ITX
3. PREDICATE DEVICE
TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound System(The ultrasound console):
Predicate Device: GE Vivid iq (K221148)
Reference Device: Clivia series Diagnostic Ultrasound System(K211886)
Common name: Ultrasonic pulsed doppler imaging system
Regulation number: 21 CFR 892.1550
Regulation Class: Class II
Product Code: IYN, IYO
Page 6
TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound System (The ultrasound catheter):
Predicate Device: SOUNDSTAR 3D Ultrasound Catheter(K092064)
Common name: Catheter, ultrasound, intravascular
Regulation number: 21 CFR 870.1200
Regulation Class: Class II
Product Code: OBJ
4. DEVICE DESCRIPTION
The proposed TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound System is a Track 3 diagnostic device designed for imaging intracardiac anatomical structures and blood flow dynamics in adult patients. It comprises two primary components: the ultrasound console and the TINGSN Sonic Eyes 10 Eco Disposable Intracardiac Echocardiography Catheter. The console features a 17.3-inch touch-enabled LCD screen, a control panel, and interfaces for connectivity, powered by a rechargeable battery or AC adapter, with optional peripherals including a cart and foot switch. The catheter, a sterile, single-use device with a 95 cm insertable length and 10Fr diameter, is equipped with a 64-element phased array transducer at its tip for high-resolution imaging. It connects to the console via the Finders 2 Sonic Eyes Connector, featuring a locking mechanism to ensure secure attachment during procedures. The system supports multiple imaging modes—B-Mode, M-Mode, Pulsed Wave (PW) Doppler, Continuous Wave (CW) Doppler, Color Doppler, Power Doppler, and combined modes —along with advanced functionalities, displayed with real-time thermal (TI) and mechanical (MI) indices.
5. INDICATIONS FOR USE
TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound System is intended for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult patients.
Modes of operation include: B,M, Color Doppler,PW Doppler, CW Doppler,Power Doppler; Combined modes: B/M,B/Color,B/M/Color,B/PWD,B/CWD, B/Color/PWD, B/Color/CWD, B/Power.
The device is intended for use in Professional healthcare facility environment including echo lab, other hospital settings, operating room, Cath lab and EP lab.
The intended users of the Device is to be used exclusively by physicians who are appropriately trained personnel in a fully equipped electrophysiology or catheterization room or under their
Page 7
direct supervision.
6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound System, including its console and TINGSN Sonic Eyes 10 Eco Catheter, is compared to predicate devices (GE Vivid iq and SOUNDSTAR 3D Catheter) and a reference device (Clivia Series). The proposed device shares the same intended use, biocompatibility as the predicate devices and is similar in technological characteristics. The TINGSN FINDERS 2 system, designed for intracardiac and intra-luminal imaging in adults, differs from the broader-ranging Vivid iq and SOUNDSTAR 3D Catheter predicates but is substantially equivalent, as its indications are encompassed by these devices. Technical differences in imaging modes, software, and catheter length exist but do not impact safety or effectiveness. The system aligns with the predicates in core functions and principles, demonstrating substantial equivalence with no new risks.
Table 1: Comparison of the TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound System to the predicate device.
| Description | Proposed Device TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound System | Predicate Device vivid iq (K221148) | Reference Device Clivia series Diagnostic Ultrasound System (K211886) | Comments |
|---|---|---|---|---|
| FDA Classification | Class II | Class II | Class II | Same |
| Product Code | IYN, IYO | IYN, IYO | IYN, IYO | Same |
| Regulation | 21 CFR 892.1550 | 21 CFR 892.1550 | 21 CFR 892.1550 | Same |
| Indications for use | TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound System is intended for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as | The Vivid iq is high-performance compact diagnostic ultrasound system designed for cardiovascular and shared services.It is intended for use by Healtheare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid, Vivid iq clinical applications include: Fetal/Obstetrics, Abdominal (includes | The Clivia series Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It intended for use in | SE Analysis 1 |
Page 8
| visualization of other devices in the heart of adult patients. Modes of operation include: B,M, Color Doppler,PW Doppler, CW Doppler,Power Doppler; Combined modes: B/M,B/Color,B/M/Color, B/PWD, B/CWD, B/Color/PWD, B/Color/CWD, B/Power. The device is intended for use in Professional healthcare facility environment including echo lab, other hospital settings, operating room, Cath lab and EP lab. The intended users of the Device is to be used exclusively by physicians who are appropriately trained personnel in a fully equipped electrophysiology or catheterization room or under their direct supervision. | GYN), Pediatric, Small Organ(includes breast, testes, thyroid), Neonatal Cephalic, Adult Cephalic, Cardiac (includes Adult and Pediatric), Peripheral Vascular, Musculoskeletal Conventional, Musculoskeletal Superficial, Urology (Including prostate), Transeranial, Transrectal, Transvaginal, Transesophageal, Interventional Guidance (including Biopsy), Thoracic/Pleural,Intraoperative(Vascular), Intracardiac and Intraluminal. Modes of operation include: B, M, PW Doppler, CW Doppler, Color Doppler, Color M, Power Doppler, Harmonic lmaging, Real-Time (RT) 3D Mode (4D), Coded Pulse and Combined modes: B/M, B/Color M, B/PWD, B/CWD, B/Color/PWD, B/Color/CWD, B/Power/PWD. The device is intended for use in an indoor hospital environment including echo lab, other hospital settings, operating room, Cath lab and EP lab, in private medical offices, and in limited settings outside of professional Healthcare facilities. | Fetal/Obstetrics, Abdominal/GYN, Pediatrics, Small Organ(breast, thyroid, testes), Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Musculoskeletal (Conventional), Musculoskeletal (Superficial), Thoracic/Pleural, Cardiac Adult, Cardiac Pediatric), peripheral vessel and Urology exams. The operator for Clivia series Diagnostic Ultrasound System is intended for professional clinical staff or the qualified and trained personnel with experience in the use of ultrasound diagnostic equipment. The device is intended to be used in hospital or clinics. Modes of operation include: B-Mode, M-Mode, Color | |
Page 9
| | | Mode, Power (Dirpower)-Mode, PW Doppler Mode, CW Doppler Mode, 3D/4D, Holo PW, Tissue Harmonic Imaging, Tissue Doppler Imaging(TDI), Anatomic M(AMM) and combined mode. | |
| Acoustic Output Track | Track 3 | Track 3 | Track 3 | Same |
| Principle of operation | The TINGSN FINDERS 2 system employs Doppler ultrasound technology and ultrasound echo principles to simultaneously acquire information on blood flow dynamics, tissue motion, and organ structure imaging. When the ultrasound system is in operation, the transmission circuit converts electrical current through high-frequency oscillation into pulsed current, which is transmitted to the transducer at the tip of the ultrasound catheter. Utilizing the piezoelectric effect of the transducer crystal, | System console generates electrical current and sends to a connected, compatible transducer to stimulate its piezoelectric elements at the distal end. Stimulation causes the elements to expand and contract which creates a high-pressured wave . A series of soundwaves then propagate toward tissue medium (e.g., mucus membrane, bone). The transducer then receives echoed soundwaves that are reflected from the tissue medium and transmits it to the system. The system processes the echoed soundwaves into an image which is displayed on the display monitor screen of the system for user interpretation. | System console generates electrical current and sends to a connected, compatible transducer to stimulate its piezoelectric elements at the distal end. Stimulation causes the elements to expand and contract which creates a high-pressured wave . A series of soundwaves then propagate toward tissue medium (e.g., mucus membrane, bone). The transducer then receives | Same |
Page 10
| | the pulsed current is transformed into ultrasound waves that are emitted. The same transducer crystal receives the reflected ultrasound waves and converts them back into electrical signals. The PC system processes the received echoes and displays a ultrasound image on the monitor, using varying brightness to represent echo strength. | | echoed soundwaves that are reflected from the tissue medium and transmits it to the system. The system processes the echoed soundwaves into an image which is displayed on the display monitor screen of the system for user interpretation. | |
| Power Source | Ultrasound | Ultrasound | Ultrasound | Same |
| Configuration | Mobile cart with braking system | Mobile cart with braking system | Mobile cart with braking system | Same |
| Imaging modes | B, ,M, Color Doppler,PW Doppler, CW Doppler Power Doppler; Combined modes: B/M,B/Color,B/M/Color, B/PWD, B/CWD, B/Color/PWD, B/Color/CWD, B/Power. | B, M, PW Doppler, CW Doppler.Color Doppler, Color M, Power Doppler, Harmonic lmaging, Real-Time (RT) 3D Mode (4D), Coded Pulse and Combined modes: B/M, B/Color M, B/PWD, B/CWD, B/Color/PWD, B/Color/CWD, B/Power/PWD. | B-Mode,M-Mode, Color Mode, Power (Dirpower)-Mode, PW Doppler Mode, CW DoppleMode, 3D/4D, Holo Pw, Tissue Harmonic lmaging, Tissue Doppler lmaging(TDl), Anatomic M(AMM) and combined mode. | SE Analysis 2 |
Page 11
| Type of compatible probes | ICE Probes(TINGSN Sonic Eyes 10 Eco Disposable Intracardiac Echocardiography Catheter) | Vivid iq utilizes a variety of electronic array transducers operating in linear, curved, sector/phased array, matrix array, or dual array format, including dedicated CW transducers and real time 3D transducer. The system can also be used with compatible ICE transducers. | Convex Array Intracavitary Probe (used transvaginally); Convex Array Probe (abdominal); Convex Volume Probe (abdominal); Linear Array Transducers (superficial peripheral vessels, small organs, peripheral vessels); Phased Array (cardiac). | SE Analysis 3 |
| Primary Software features | One-click optimization (Auto) Line density Image enhancement Pseudo-color Smoothing Multi-focus synthesis* Grayscale mapping Frame correlation Wall filtering Color priority Color dual real-time* Edge enhancement Time-frequency resolution Tissue Doppler frequency spectrum imaging (TVD) Tissue Doppler imaging (TDI) Tissue velocity imaging (TVI) | Automatic EF (ejection fraction) Advanced cardiac measurements and analysis 4D/Multiplane LV measurement and analysis 4D Auto AVQ Composite imaging Advanced vascular measurements and analysis Pediatric calculations Obstetric measurements and calculations Quantitative analysis Measurements: depth, angle, distance, perimeter, area, time, slope, Doppler SV position, Doppler speed, Doppler angle correction | Tissue Doppler imaging (TDI), Tissue velocity imaging (TVI), Tissue Doppler frequency spectrum imaging (TVD), Anatomical M, 3D/4D imaging, Wide-angle imaging, Holo (plane wave PW) imaging, Image optimization features including gain, depth, TGC, image quality (frequency), sound power, focus, image display adjustment, line density, dynamic | SE Analysis 4 |
Page 12
| Measurements: depth, distance, perimeter, area, slope, angle, heart rate, acceleration, time, velocity | | range, image enhancement, frame correlation, rotation and mirror, spatial compound, grayscale mapping, pseudo-color, one-click optimization, wiNeedle, puncture-enhanced dual-window, time scale, display formats, edge enhancement, color gain, sample box adjustment, dual real-time, deflection, sensitivity, smoothing, scale, baseline, flip, waveform, wall filtering, color priority, trace lines, velocity, time/frequency resolution, display formats, duplex/triplex, angle correction, fast angle, volume, PW deflection, sample volume Measurements include: distance, perimeter, area, volume, angle, | |
Page 13
| | | depth, four-point three-line, hip joint angle, time, heart rate, slope, velocity, acceleration, spectrum, trace. | |
Substantial Equivalence (SE) Analysis
SE analysis 1: Indications for Use
The Proposed Device is compatible only with the ICE catheter, which limits its application to intracardiac and intra-luminal ultrasound imaging. In contrast, the Predicate Device (Vivid iq) supports multiple probes, offering a broader range of applications. However, the Predicate Device also supports ICE catheters (e.g., the Soundstar 3D Ultrasound Catheter, AcuNav Diagnostic Ultrasound Catheter) for intracardiac and intra-luminal ultrasound imaging. Therefore, the Predicate Device's indications for use encompass those of the Proposed Device. They can be considered Substantially Equivalent (SE) in terms of safety and effectiveness, as no new risks are introduced, and the SE is not affected.
SE analysis 2: Imaging Modes
Both the Proposed Device and the Predicate Device share several key imaging modes, including B-mode, M-mode, Color Doppler, PW Doppler, CW Doppler, and Power Doppler. However, the Predicate Device supports additional imaging modes, including Real-Time 3D Mode (4D) , which are not present in the Proposed Device. Despite this difference, the essential imaging functions for cardiac and vascular diagnostics are supported in both devices. The inclusion of Real-Time 3D Mode (4D) in the Predicate Device does not introduce new risks or affect the safety and effectiveness of the Proposed Device. Therefore, this difference does not impact substantial equivalence, and both devices are substantially equivalent in terms of imaging modes.
SE analysis 3:Type of Compatible Probes
The Proposed Device is compatible only with ICE catheters, while the Predicate Device (Vivid iq) supports a wider variety of probes, offering more flexibility. However, the Predicate Device also supports ICE catheters, so the SE is not affected by this difference.
SE Analysis 4:Primary Software features
While the specific features differ, the Predicate Device's software functions largely encompass the core functions of the Proposed Device, especially in terms of cardiac and vascular analysis,
Page 14
Doppler imaging, and basic measurement tools. Therefore, the Predicate Device's software is sufficient to cover the functionality of the Proposed Device without leading to substantial differences in safety or effectiveness.A thorough risk analysis was conducted to assess potential risks related to software differences, particularly in the area of measurement accuracy and imaging performance. The risk mitigation strategies implemented during development ensure that the software functions of the Proposed Device meet the necessary safety and effectiveness standards, and any deviations from the Predicate Device's software features were carefully evaluated for potential impact on diagnostic performance.
Additionally, a Reference Device, the Clivia series Diagnostic Ultrasound System (K211886), has been introduced. Its software features cover the software features of the Proposed Device. Therefore, we conclude that the differences between the software features of the Proposed Device and the Predicate Device do not introduce new risks that would compromise the safety or effectiveness of the device.
Table 2 Comparison of the proposed TINGSN Sonic Eyes 10 Eco Catheter to the currently marketed predicate device, the Soundstar 3d Ultrasound Catheter.
| Description | Proposed Device TINGSN Sonic Eyes 10 Eco Catheters | Predicate Device The Soundstar 3d Ultrasound Catheter (K092064) | Comments |
|---|---|---|---|
| Classification | Class II | Class II | Same |
| Regulation | 21 CFR 870.1200 | 21 CFR 870.1200 | Same |
| Product code | OBJ | OBJ | Same |
| Indications for use | TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound System is intended for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult patients. Modes of operation include: B,M, Color Doppler,PW Doppler, CW Doppler,Power Doppler; Combined modes: B/M,B/Color,B/M/Color, B/PWD, B/CWD, B/Color/PWD, B/Color/CWD, B/Power. | The Biosense Webster SOUNDSTAR 3D Diagnostic Ultrasound Catheter and related accessory devices are indicated for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. When used with the CARTO XP EP Navigation System, Version 9.7 or greater, the SOUNDSTAR 3D Catheter provides location information. | SE Analysis 1 |
Page 15
| The device is intended for use in Professional healthcare facility environment including echo lab, other hospital settings, operating room, Cath lab and EP lab. The intended users of the Device is to be used exclusively by physicians who are appropriately trained personnel in a fully equipped electrophysiology or catheterization room or under their direct supervision. | | | |
| Principle of operation | Catheter with ultrasound sensors that is used with an imaging system to display to provide high-resolution real-time visualization of cardiac structures, continuous monitoring of catheter location within the heart. | Catheter with ultrasound sensors that is used with an imaging system to display to provide high-resolution real-time visualization of cardiac structures, continuous monitoring of catheter location within the heart. | Same |
| Patient Population | Adult, not for fetal or pediatric use or for use in coronary vessels. | Adult, not for fetal or pediatric useor for use coronary vessels. | Same |
| Single Use | Yes | Yes | Same |
| Duration of Use | Limited (≤ 24 hours) | Limited (≤ 24 hours) | Same |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide | Same |
| Outside Diameter | 10F | 10F | Same |
| Insertable length | 95cm | 90cm | SE Analysis2 |
| Deflection mechanism | Integrates a tip deflection mechanism using a deflection handle and pull-line to provide integrated 4-way deflection of catheter. | Integrates a tip deflection mechanism using a deflection handle and pull-line to provide integrated 4-way deflection of catheter. | Same |
| Transducer configuration | Multi-element linear phased array ultrasound transducer at distal tip | Multi-element linear phased array ultrasound transducer at distal tip | Same |
| Number of Ultrasound Transducer elements | 64-channel acoustic phased array | 64-channel acoustic phased array | Same |
Page 16
| Ultrasound Imaging Frequency | 4.5-9.5MHz | 4-10 MHz | SE Analysis 3 |
| Acoustic Output Track | Track 3 recommendations: the global maximum derated ISPTA should be ≤720 mW/cm2, and either the global maximum MI should be ≤1.9 or the global maximum derated ISPPA should be ≤ 190 W/cm2. | Track 3 recommendations: the global maximum derated ISPTA should be ≤720 mW/cm2, and either the global maximum MI should be ≤1.9 or the global maximum derated ISPPA should be ≤ 190 W/cm2. | Same |
| Imaging modes | B-Mode (2D only) M-Mode Doppler Pulsed Wave Doppler Continuous Wave Doppler Color Doppler (2D only) Power Doppler | B-Mode (2D only) M-Mode Doppler Pulsed Wave Doppler Continuous Wave Doppler Color Doppler (2D only) Power Doppler | Same |
| Patient contact materials(Catheter Shaft) | Pebax series Comply with ISO 10993-1 | Pebax series Comply with ISO 10993-1 | Same |
| Shelf life | 2 years | 1 Year | SE Analysis 4 |
Substantial Equivalence (SE) Analysis
SE analysis 1:Indications for Use
Both the Proposed and Predicate Devices are intended for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology, as well as visualization of other devices in the heart of adult patients. However, the Predicate Device's catheter tip is equipped with a position sensor that works with the CARTO system to provide location information, making its indications for use broader. This broader indication encompasses the use of the Proposed Device. Therefore, the Proposed Device does not introduce a new intended use, and its indications for use fall within those of the Predicate Device.
SE analysis 2:Insertable Length
The insertable length of the Proposed Device, like that of the Predicate Device, meets clinical requirements for a wide range of patients. This slight difference does not raise any concerns regarding safety or effectiveness.
SE analysis 3:Ultrasound Imaging Frequency
The Proposed Device has a narrower ultrasound frequency spectrum compared to the Predicate Device. However, both devices operate within the low-frequency ultrasound range, and the
Page 17
Proposed Device's frequency range is adequate to meet the clinical requirements for intracardiac imaging. Therefore, this difference does not impact the safety or effectiveness of the device.
SE analysis 4: Shelf Life
The Proposed Device has undergone a 2-year shelf-life study, verifying product stability and packaging integrity. The claimed 2-year shelf life is scientifically reliable. These differences will not affect the safety and effectiveness of the Proposed Device.
7. PERFORMANCE DATA
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound System conforms to the following standards:
- IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)
- IEC 60601-1-2 2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-2-37 2015 Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
- IEC 62359 2017 Ultrasonics - Field characterization - Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields
- NEMA UD2 2004(2009) Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment Revision 3
- IS0 10555-1 2013 Intravascular catheters - Sterile and single-use intravascular catheters - Part 1: General requirements [Including AMENDMENT 1 (2017)]
- ISO 10993-1 2018, Biological evaluation of medical devices –Part 1: Evaluation and testing within a risk management process
- ASTM D4169-2022 Standard Practice for Performance Testing of Shipping Containers and Systems
Page 18
- ISO 14971: Medical devices- Applications of risk management to medical devices, 2019
- FDA Ultrasound Guidance document, "Marketing Clearance of Diagnostic Ultrasound Systems and Transducers," issued in February 2023
Software documentation is included in accordance with the FDA guidance "Guidance for the Content of Premarket Submissions for Device Software Functions," issued on June 14, 2023. Non-clinical tests (integration and functional) were conducted on the TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound System during product development.
The risk analysis was completed and risk controls were implemented to mitigate identified hazards. The test results support that all the software specifications have met the acceptance criteria. Verification and validation testing were found acceptable to support the claim of substantial equivalence.
8. CLINICAL DATA
No clinical data was included in this submission.
9. CONCLUSION
The indications for use statement for the proposed device is similar to that of the predicate. While its indications are more focused on intracardiac and intra-luminal imaging, the Vivid iq and SOUNDSTAR 3D Catheter encompass these uses,the differences between the proposed device and its predicate device do not raise new issues of safety and effectiveness. The non-clinical data support the safety of the device and the performance testing report demonstrate that the TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound System should perform as intended in the specified use conditions.
From the results of non-clinical data including the performance testing described, Jiangsu Tingsn Technology Co., Ltd. concludes that the TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound System is as safe, as effective, and performs as well as the predicate devices (K221148 and K092064), and is therefore considered substantially equivalent to the predicate devices.
§ 892.1550 Ultrasonic pulsed doppler imaging system.
(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.