Search Filters

Search Results

Found 187 results

510(k) Data Aggregation

    K Number
    K251358
    Date Cleared
    2025-06-24

    (55 days)

    Product Code
    Regulation Number
    870.4450
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    80921

    Re: K251358
    Trade/Device Name: Bridge Plus Occlusion Balloon
    Regulation Number: 21 CFR 870.4450
    , Intravascular Occluding, Temporary
    Device Class: II
    Classification Regulation: 21 CFR 870.4450

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Bridge Plus Occlusion Balloon catheter is is indicated for temporary vessel occlusion of the Superior Vena Cava (SVC) and Inferior Vena Cava (SVC) in applications including perioperative occlusion and emergency control of hemorrhage.

    Device Description

    The Bridge Plus Occlusion Balloon is designed to be delivered percutaneously to the superior vena cava (SVC) or the inferior vena cava (IVC) for the purpose of providing occlusion, emergency control of hemorrhage and perioperative occlusion in the event of an SVC or IVC tear or perforation during a vascular procedure.

    The Bridge Plus Occlusion Balloon catheter is constructed of a compliant polyurethane balloon mounted on a 12Fr dual lumen nylon shaft. This guidewire lumen is used to pass the catheter over a guidewire. The balloon has a length of 80mm and inflation diameter ranging from 20-29mm. Three platinum-iridium radiopaque markers are placed within the balloon segment of the catheter to provide visual reference points for balloon positioning within the SVC or IVC prior to inflation.

    AI/ML Overview

    This FDA 510(k) clearance letter pertains to a medical device, the Bridge Plus Occlusion Balloon, and therefore the "acceptance criteria" and "study that proves the device meets the acceptance criteria" refer to the performance specifications and verification/validation testing described in the submission, rather than clinical study endpoints for AI/software devices. The provided document does not describe a study involving human readers or AI algorithms.

    Here's a breakdown of the requested information based on the provided text:


    1. Table of Acceptance Criteria and Reported Device Performance

    The document lists various Design Verification and Validation Testing categories. For these types of medical devices, the "acceptance criteria" would be the pre-defined performance specifications for each test, which the device must meet to be considered safe and effective and substantially equivalent to a predicate device. The "reported device performance" indicates that the Bridge Plus Occlusion Balloon met these specifications.

    Note: The document does not provide the specific quantitative acceptance criteria or detailed numerical performance results for each test. Instead, it states that "Performance testing data have demonstrated that the Bridge Plus Occlusion Balloon meets performance specifications to support substantial equivalence to the predicate devices." and that the conducted tests "validate and verify that the subject device met all specifications."

    Acceptance Criteria CategoryReported Device Performance
    Design Verification & Validation TestingMet specifications for substantial equivalence.
    Dimensional and Visual TestsMet specified dimensions and visual appearance requirements (e.g., Crossing Profile, Balloon Compliance, Balloon Working Length, Catheter Effective Length, Surface Appearance, Distal Tip Configuration, Luer Compatibility, Guidewire Compatibility).
    Performance TestingMet specified performance requirements (e.g., Deployment and Retraction, Withdrawal Force, Inflation and Deflation Time, Inflated Balloon Size Stability, Balloon Bond Strength, Tip Bond Strength, Hub Bond Strength, Flexibility and Kink, Torque Strength, Vessel Occlusion, Leak Testing, Balloon Fragmentation & Burst Volume, Device Fatigue).
    SterilizationDemonstrated compliance with ISO 11135:2014 Amd 1, confirming the appropriateness of existing process challenge devices and validating bioburden methods.
    BiocompatibilityMet biocompatibility requirements for various tests (Cytotoxicity, Sensitization, Irritation or Intracutaneous Reactivity, Material Mediated Pyrogenicity, Acute Systemic Toxicity, Direct and Extract Hemolysis, Partial Thromboplastin Time, Platelet & Leukocyte Count, Comparative Surface Analysis, Complement Activation).

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes laboratory and bench testing rather than clinical studies with "test sets" of patient data. Therefore, the concept of sample size for a test set of data and data provenance (country of origin, retrospective/prospective) is not applicable in the context of this device's submission. The testing was conducted on the physical device itself.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Since this is a physical medical device (occlusion balloon) and not an AI/software device, the concept of "ground truth established by experts" for a test set is not applicable. Device performance was evaluated against engineering specifications and industry standards.


    4. Adjudication Method for the Test Set

    As this is not an AI/software device or a clinical study with subjective assessments, an "adjudication method" for a test set is not applicable.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of Human Reader Improvement with AI vs. Without AI Assistance

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for imaging or diagnostic AI products where human reader performance is being evaluated with and without AI assistance. The Bridge Plus Occlusion Balloon is a physical medical device for temporary vessel occlusion.


    6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

    No, a standalone (algorithm only) performance study was not done. This device is a physical medical instrument, not an algorithm.


    7. The Type of Ground Truth Used

    For this physical medical device, the "ground truth" is defined by engineering specifications, recognized industry standards (e.g., ISO for sterilization and biocompatibility), and performance benchmarks established by predicate devices. The device's physical and mechanical properties were tested against these pre-defined quantifiable criteria.


    8. The Sample Size for the Training Set

    The concept of a "training set" is not applicable to this physical medical device. Training sets are used for machine learning algorithms.


    9. How the Ground Truth for the Training Set Was Established

    Since there is no "training set" for this physical device, the method for establishing its "ground truth" is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232577
    Date Cleared
    2024-01-18

    (146 days)

    Product Code
    Regulation Number
    870.4450
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    China. 315336

    Re: K232577

    Trade/Device Name: Radial Compression Device Regulation Number: 21 CFR 870.4450
    device Classification Name: Vascular Clamp Classification Panel: Cardiovascular Regulation: 21 CFR 870.4450

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The radial compression device is designed to apply compression in order to achieve temporary hemostasis of the radial artery after a percutaneous transradial procedure.

    Device Description

    The Radial Compression Device (RC-001) is designed to promote hemostasis of the radial artery post percutaneous puncture. The radial compression device is fixed on the arm and the square pad of the device is pressed against the puncture site. The square cushion is adjusted to apply pressure and temporarily compress the puncture site of the radial artery to achieve rapid hemostasis. The product is primarily composed of a plate, turn cap, screw rod, hook and loop band, tapered cushion, and square cushion. The Radial Compression Device is available in one model, RC-001.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the Ningbo DIZENGENS Medical Science Co., Ltd. Radial Compression Device (K232577). The document presents a summary of non-clinical performance testing to support the claim of substantial equivalence to a predicate device. However, it does not contain the detailed acceptance criteria and study results in the format requested.

    Specifically, the document does not provide:

    • A table of acceptance criteria and reported device performance with specific metrics.
    • Sample sizes used for a "test set" in the context of an algorithm's performance.
    • Data provenance, number of experts, adjudication method, or ground truth details for an AI/algorithm study.
    • Information on Multi-Reader Multi-Case (MRMC) comparative effectiveness studies or standalone algorithm performance.
    • Sample size or ground truth establishment for a "training set."

    The document primarily focuses on non-clinical performance testing for a physical medical device (a radial compression device), not an AI/software algorithm. The tests conducted are typical for a physical medical device.

    Here's a breakdown of what is provided and what elements are missing or not applicable based on the provided text, in the context of your request for AI/algorithm study details:

    1. Table of Acceptance Criteria and Reported Device Performance:

    • Provided: A list of performance tests conducted. For each test, it states "All testing met the requirements and passed."
    • Missing from provided text: Specific numerical acceptance criteria for each test and the corresponding numerical results obtained by the Radial Compression Device. For example, for "Bond strength testing," it doesn't state "Acceptance criteria: > X N, Result: Y N."

    2. Sample size used for the test set and the data provenance:

    • Not Applicable in the provided text: This question is relevant for AI/algorithm performance evaluation. The document describes tests for a physical device (e.g., visual inspections, bond strength, biocompatibility), not an AI model. Therefore, there's no "test set" of data in the context of AI.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable in the provided text: This question is relevant for AI/algorithm performance. Ground truth for a physical device's non-clinical performance tests is typically established through established engineering standards, laboratory procedures, and physicochemical measurements, not through expert consensus on images or clinical data.

    4. Adjudication method for the test set:

    • Not Applicable in the provided text: This is relevant for AI/algorithm performance.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

    • No: The document does not mention any MRMC study, as it's for a physical device and not an AI assistant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • No: The device is a physical Radial Compression Device, not an algorithm.

    7. The type of ground truth used:

    • Implicitly: For the physical device tests, the "ground truth" is based on engineering specifications, standardized test methods (e.g., ISO standards for biocompatibility and sterilization), and validated laboratory measurements. For instance, biocompatibility is evaluated per ISO 10993, and sterilization per ISO 11135.
    • Missing from provided text: No explicit mention of ground truth in the context of expert consensus, pathology, or outcomes data, as these are typically for diagnostic AI/software.

    8. The sample size for the training set:

    • Not Applicable: There is no AI/algorithm training set.

    9. How the ground truth for the training set was established:

    • Not Applicable: There is no AI/algorithm training set.

    Summary of Provided Non-Clinical Performance Testing:

    The provided document lists the following non-clinical performance tests conducted for the Radial Compression Device:

    • Visual inspections
    • Dimensional inspections
    • Bond strength testing
    • Turn Cap and Screw Function
    • Distribution and Packaging Tests
    • Device and Packaging Aging Evaluation
    • Pressure applied at puncture site (to demonstrate adequate compression)
    • Biocompatibility evaluation per ISO 10993 (Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Pyrogenicity, Hemolysis)
    • Sterilization (EO gas, SAL of 10-6 per ISO 11135, EO/ECH residuals assessment per 10993-7)

    For all of these tests, the document states: "All testing met the requirements and passed."

    Conclusion:

    The provided text serves as a 510(k) summary for a physical medical device. It details non-clinical performance testing for that physical device to establish substantial equivalence. It does not provide the information requested regarding acceptance criteria and studies for an AI/algorithm device, as the product is not an AI/algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230281
    Device Name
    PICOCLAMP
    Date Cleared
    2023-10-23

    (264 days)

    Product Code
    Regulation Number
    870.4450
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Ichikawa city, Chiba 272-0837 Japan

    Re: K230281 Trade/Device Name: Picoclamp Regulation Number: 21 CFR 870.4450
    Vascular |
    | Classification Name: | Vascular clamp |
    | Regulation: | 21 CFR §870.4450

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A surgical instrument used to occlude a blood vessel temporarily.

    Device Description

    PICOCLAMP is Ethylene Oxide sterile single-use hemostasis clamp consisting of Polyetheretherketone (PEEK) and stainless steel (SUS series). PICOCLAMP is designed for use during surgery to temporarily occlude blood vessels during the anastomosis process. PICOCLAMP consists of three main direct patient-contacting parts, all of them are streilized; main body, axis, and spring. The body is consisting of two parts; the jaw which is used to occlude the blood vessel temporarily and the grip which functions to open and close jaws using the surgeon hand or the clamp forceps. The axis integrates the body and keep its parts intact. The spring functions to give clamping force to the jaws. Two types of clamps are provided: green color clamps used for arteries temporary occlusion and blue color clamps used for veins temporary occlusion. Both of the artery and vein clamp are having two types: single and double. PICOCLAMP is available in three different sizes; S (Small), M (Medium), and L (Large). PICOCLAMP is used for temporary occlusion of blood vessels with diameters ranging from 0.2mm to 2.0mm. The clipping power applied at the blood vessel differs depending on the size of the clamp and the pressure of the spring selected. Once the anastomosis process is complete, PICOCLAMP is removed from the patient.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding the PICOCLAMP vascular clamp. It outlines the device's characteristics, indications for use, comparison with a predicate device, and the performance data submitted to demonstrate substantial equivalence.

    Based on the provided text, there is no information about acceptance criteria or a study proving the device meets those criteria in the context of AI/ML performance. The document is a regulatory approval for a physical medical device (vascular clamp), not a software or AI/ML-driven device.

    Therefore, I cannot provide the requested information regarding acceptance criteria and a study proving device performance using AI/ML, as the document does not pertain to such a device or study.

    The performance data mentioned in the document are for:

    • Sterilization & Shelf-life & Packaging Testing: This ensures the device remains sterile and functional over its shelf life and that its packaging is adequate.
    • Biocompatibility Testing: This ensures the materials used in the device are safe for contact with biological tissues.
    • Bench Performance Testing: This evaluates the physical performance of the clamp, such as clipping power, ability for repeated opening/closing, and resistance to dislodgement. These tests compare the proposed device (PICOCLAMP) to the predicate device to demonstrate substantial equivalence of their physical performance.

    The document explicitly states:

    • "Animal Performance Testing was not required to demonstrate safety and effectiveness of the device."
    • "Clinical Performance Testing was not required to demonstrate the safety and effectiveness of the device."

    In summary, none of the requested information regarding AI/ML acceptance criteria, test set sizes, expert ground truth, MRMC studies, standalone algorithm performance, or training set details can be extracted from this document, as it is completely unrelated to AI/ML device performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221294
    Manufacturer
    Date Cleared
    2023-06-30

    (422 days)

    Product Code
    Regulation Number
    870.4450
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Massachusetts 01923

    Re: K221294

    Trade/Device Name: preCARDIA Occlusion System Regulation Number: 21 CFR 870.4450
    |
    |----------------------------|----------------------------|
    | Regulation Number: | 21 CFR 870.4450

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The preCARDIA device is intended for use in selectively stopping or controlling blood flow of the inferior and superior vena cava in applications including perioperative procedures in patients requiring emergency control of hemorrhage, and for blood pressure monitoring.

    Device Description

    The preCARDIA System is for use in selectively stopping or controlling blood flow of the inferior and superior vena cava, in applications including perioperative procedures in patients requiring emergency control of hemorrhage, and for blood pressure monitoring. The System includes a Balloon Occlusion Catheter (Catheter), Controller and Console. The Catheter has an atraumatic endovascular balloon that has been attached to the proximal portion of a commercially available Thermodilution (TD) Catheter. The single use, 110 cm disposable catheter comes in two balloon locations, 34 and 36 cm measured from the distal end of the catheter and is provided sterile. The catheter placement procedure is performed under fluoroscopy by clinicians trained in endovascular techniques. The balloon includes proximal and distal radiopaque markers to allow for visualization and confirmation of proper placement in the IVC or SVC. Following placement and securement of the catheter, the patient can be transferred to the operating room or other appropriate care location where selectively stopping or controlling blood flow in the SVC or IVC will continue alongside other standard of care treatments. The primary operating principle of the preCARDIA System is the selective stopping or controlling of the blood flow in the IVC or SVC through either sustained inflation or successive inflation and deflation cycles of the balloon as regulated by the Controller.

    AI/ML Overview

    The provided text is a 510(k) summary for the preCARDIA Occlusion System. It details the device, its intended use, and a comparison with predicate devices. While it lists various performance tests conducted, it does not contain specific acceptance criteria or the reported device performance for an AI/ML-based device.

    The document outlines bench testing and an animal study to evaluate the performance of the preCARDIA Catheter. However, it does not provide:

    • A table of acceptance criteria and reported device performance.
    • Sample sizes for a test set or data provenance for an AI/ML context.
    • Information on expert ground truth establishment (number of experts, qualifications, adjudication method).
    • Details on MRMC studies or a standalone algorithm performance.
    • The type of ground truth used relevant to AI/ML (e.g., pathology, outcomes data).
    • Sample size for a training set or how ground truth was established for AI/ML training data.

    The text is for a physical medical device (vascular clamp/occlusion system), not an AI/ML algorithm. Therefore, the concepts requested in the prompt, such as "acceptance criteria for an AI/ML model," "training set," "test set," "ground truth," "expert consensus," and "MRMC study," are not applicable to the content provided.

    Based on the provided document, I cannot fulfill the request as it pertains to AI/ML device validation. The document describes the regulatory submission for a physical medical device and its associated non-AI/ML performance testing.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221661
    Date Cleared
    2023-03-03

    (268 days)

    Product Code
    Regulation Number
    870.4450
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Device Name: Abdominal Aortic and Junctional Tourniquet - Stabilized (AAJT-S) Regulation Number: 21 CFR 870.4450
    -|-----------------|
    | Product Code: | DXC |

    FDA Regulation Number:21 CFR 870.4450
    Product Code:DXC
    FDA Regulation Number:21 CFR 870.4450
    ------------------
    Product Code:DXC
    FDA Regulation Number:21 CFR 870.4450
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Abdominal Aortic and Junctional Tourniquet - Stabilized (AAJT-S) is indicated for use to control bleeding in the pelvis, inguinal area, axilla and for pelvic fracture stabilization.

    Device Description

    The AAJT-S device is a wedge-shaped bladder that when inflated pushes in on the body Over the target area of application. One of the target areas is the lower abdomen, over the umbilicus, compressing all the structures within, including the descending aorta, to achieve the physiological effects of aortic compression. The other two areas are over the inguinal region or the axilla. In these two areas the region is compressed with the wedge- shaped bladder including the major vasculature that runs through that area.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Abdominal Aortic and Junctional Tourniquet - Stabilized (AAJT-S):

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device PerformanceStudy to Prove
    Pressure Maintenance: The AAJT-S must stay inflated and maintain the same pressure over the target area for at least a 4-hour period.Confirmed successful inflation and pressure maintenance for at least a 4-hour period.Pressure Testing
    Component Bond Strength: The bonds of components to the hard cover must withstand normal tensile forces during clinical use.Demonstrated ability to withstand normal tensile forces.Force Testing
    Biocompatibility (Cytotoxicity): No cytotoxic effects.Passed ISO 10993-5 (in vitro cytotoxicity).Biocompatibility Testing
    Biocompatibility (Irritation & Skin Sensitization): No irritation or skin sensitization.Passed ISO 10993-10 (irritation and skin sensitization).Biocompatibility Testing
    Wear Time (Stated Intended Use): Abdominal use: up to 1 hour; Junctional use: up to 4 hours.Not explicitly stated as a test result, but the pressure test confirms the device maintains pressure for the maximum stated junctional use duration (4 hours). The increased wear time of 1 hour is also noted as a difference from the predicate device.Pressure Testing implicitly supports junctional wear time.
    Pelvic Fracture Stabilization: Indication for use.Not explicitly stated what testing was conducted to prove this particular indication. The document refers to it as an "additional pelvic stability indication" compared to the predicate.Not explicitly detailed in the provided text.

    Important Note: The provided text does not include specific numeric acceptance values (e.g., "maintain pressure within X mmHg of initial pressure"). It states the qualitative criteria and that the tests were performed to confirm or evaluate these criteria.

    Missing Information:

    • There are no explicit quantitative acceptance criteria or specific performance values reported for the pressure testing (e.g., pressure drop over time).
    • Similarly, for force testing, no specific force values or deformation limits are given.
    • The document doesn't detail how the "pelvic fracture stabilization" indication was specifically proven or what criteria were used for that.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text does not specify sample sizes for the performance tests (Pressure Testing, Force Testing, Biocompatibility Testing). It mentions these tests were "performed" but not the number of units or replicates used.

    Data Provenance: Not applicable in the context of these non-clinical performance tests. These are laboratory tests conducted on the device itself, not patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. These are non-clinical performance tests conducted on the device, not clinical studies involving expert interpretation of medical images or patient outcomes.

    4. Adjudication Method for the Test Set

    Not applicable. This concept applies to clinical studies where experts might disagree on ground truth, requiring an adjudication process. The tests described are objective, non-clinical measurements.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. The provided text describes non-clinical performance tests of the device itself (pressure, force, biocompatibility), not a clinical study involving human readers or comparative effectiveness with or without AI assistance. The device is a physical medical device (tourniquet), not an AI-powered diagnostic tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, in the sense that the described tests are standalone device tests. The device's performance was evaluated independently without human-in-the-loop interaction as part of the test (beyond operating the test equipment). However, it's important to clarify that this refers to mechanical device performance, not AI algorithm performance.

    7. The Type of Ground Truth Used

    The "ground truth" for these non-clinical tests is based on objective physical measurements and established international standards:

    • Pressure Testing: The ground truth is the measured pressure (e.g., using a pressure sensor) and its change over time.
    • Force Testing: The ground truth is the applied force and the device's structural integrity under that force, measured by mechanical testing equipment.
    • Biocompatibility Testing: The ground truth is determined by the results of standardized assays (e.g., cytotoxicity assays, irritation/sensitization tests) as per ISO 10993 standards.

    8. The Sample Size for the Training Set

    Not applicable. This device is a physical medical device, not an AI model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As above, this is a physical device, not an AI model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230248
    Device Name
    VIOLA
    Date Cleared
    2023-02-28

    (29 days)

    Product Code
    Regulation Number
    870.4450
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Floor Philadelphia, Pennsylvania 19103

    Re: K230248

    Trade/Device Name: Viola Regulation Number: 21 CFR 870.4450

    Name of Device

    VIOLA

    Common or Usual Name

    Vascular Clamp

    Classification

    21 CFR 870.4450

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    VIOLA clampless proximal anastomosis sealing system is indicated for use by cardiac surgeons during coronary artery bypass grafting procedures to maintain hemostasis and facilitate the completion of proximal anastomoses to the aorta without application of an aortic clamp.

    Device Description

    The VIOLA is a sterile, single-use device designed to maintain hemostasis and facilitate the completion of multiple proximal anastomoses during a coronary artery bypass grafting procedure, without application of an aortic clamp. The VIOLA system can be used to create up to 4 anastomoses within the same patient.

    This modified version of VIOLA includes an ergonomic handle with operating button and a straight distal end.

    The VIOLA is comprised of:

    1. A concentric 4 mm aortic punch with a detachable handle.
    2. A sealing assembly comprising of a sealing element (which is available in two diameters), a catheter, fixation strap and a handle.
    3. A silicone boundary marker stencil for marking the maximal suture line around the aortic incision.

    The steps to create a sealed anastomosis hole that enables the surgeon to perform a clampless anastomosis include (1) the creation of a small "needle hole" (performed in the center of the boundary marker), (2) insertion and deployment of the sealing element, (3) creation of an anastomosis hole using the VIOLA's punch, and (4) performing the anastomosis.

    AI/ML Overview

    I am sorry, but the provided text does not contain the information required to answer your request. The document is an FDA 510(k) summary for a medical device called "VIOLA" and it focuses on demonstrating substantial equivalence to a predicate device. It briefly mentions performance data in the form of "risk analysis," "bond strength testing," and "simulated use testing" that "passed successfully all acceptance criteria," but it does not provide:

    • A table of acceptance criteria and reported device performance.
    • Sample sizes for test sets or their data provenance.
    • Details about experts for ground truth or adjudication methods.
    • Information on multi-reader multi-case studies or effect sizes.
    • Details on standalone algorithm performance.
    • The type of ground truth used.
    • Sample size for the training set or its ground truth establishment.

    The document primarily states that the modified device's technological characteristics are similar to the predicate and that performance data demonstrates it is as safe and effective.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222182
    Date Cleared
    2023-01-04

    (166 days)

    Product Code
    Regulation Number
    870.4450
    Reference & Predicate Devices
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    China

    Re: K222182

    Trade/Device Name: Radial Artery Compression Tourniquets Regulation Number: 21 CFR 870.4450

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Radial Artery Compression Tourniquet is designed for temporary hemostasis of radial artery post percutaneous puncture.

    Device Description

    Radial Artery Compression Tourniquets

    AI/ML Overview

    The provided document is a 510(k) clearance letter for the "Radial Artery Compression Tourniquets" and does not contain the specific information about acceptance criteria, device performance, study details, or AI-related data as requested. This letter confirms that the device is substantially equivalent to a predicate device and can be marketed, but does not include the detailed technical study information you're looking for.

    Therefore, I cannot provide the requested information based on the input text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K214060
    Date Cleared
    2022-09-19

    (266 days)

    Product Code
    Regulation Number
    870.4450
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    City, Utah 84115

    Re: K214060

    Trade/Device Name: LANDMARK REBOA Catheter Regulation Number: 21 CFR 870.4450
    |
    | Classification Name: | • 21 CFR §870.4450

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LANDMARK REBOA Catheter is intended for temporary occlusion of large vessels including patients requiring emergency control of hemorrhage.

    Device Description

    The LANDMARK REBOA Catheter is a 67cm, 59cm working length, 6.5 Fr (2.2mm), single lumen balloon catheter for temporary intravascular vessel occlusion in patients requiring emergency control of hemorrhage. Radio-opaque markers are provided on the catheter shaft at the proximal and distal ends of the balloon for radiological location of the device is additionally provided with depth markings referenced from the balloon to provide positioning information similar to the predicate device prior to confirmation of placement through radiological means. The device is intended to be deployed through a 7F introducer sheath placed into the femoral artery through normal vascular access procedures.

    The LANDMARK REBOA Catheter consists of five elements:

    • . Occluder Balloon Inflation Lumen with female luer (DEHP free PVC)
    • Catheter Handle (PA 12)
    • Catheter Shaft (polyimide and stainless steel) with Marker Bands (stainless steel)
    • Occluder Balloon (polyurethane)
    • Guide J-tip (stainless steel, nitinol, acrylic) ●
    AI/ML Overview

    The provided text is a 510(k) summary for the LANDMARK REBOA Catheter. While it outlines non-clinical testing performed to establish substantial equivalence to a predicate device, it does not contain information about studies involving human readers, AI assistance, or the use of expert consensus for ground truth. This device, a physical medical catheter, is evaluated based on its physical performance characteristics and biocompatibility, not on diagnostic accuracy derived from image analysis or human interpretation.

    Therefore, many of the requested points regarding acceptance criteria and studies would not be applicable in this context. The document focuses on demonstrating that the new device performs comparably to a legally marketed predicate device through bench testing and an animal model, rather than validating a diagnostic algorithm's performance.

    Here's an attempt to answer the applicable points based on the provided text, with an explanation for the missing information:

    1. A table of acceptance criteria and the reported device performance

    The document doesn't present a specific table of "acceptance criteria" with quantitative performance metrics for each test in the way one might for a diagnostic AI. Instead, it states that "Performance bench testing was conducted to ensure that the LANDMARK REBOA Catheter met the applicable design and performance requirements throughout its shelf life, verify conformity to applicable standards, and demonstrate substantial equivalence to the predicate system."

    The tests performed and implicitly, the criteria for passing them, are:

    Acceptance Criteria (Implied)Reported Device Performance (Implied)
    Balloon Burst Testing: Withstand pressure without bursting.Met applicable design and performance requirements.
    Balloon Inflation/Deflation Testing: Inflate and deflate correctly.Met applicable design and performance requirements.
    Balloon Diameter to Inflation Volume Testing: Consistent diameter for given inflation volume.Met applicable design and performance requirements.
    Occlusion Time Testing: Achieve and maintain occlusion for specified time.Met applicable design and performance requirements.
    Torque Testing: Withstand torsional forces.Met applicable design and performance requirements.
    Kink Diameter Testing: Resist kinking at specific diameters.Met applicable design and performance requirements.
    Fatigue Testing: Withstand repeated stress cycles.Met applicable design and performance requirements.
    Freedom From Leakage Testing: No leaks.Met applicable design and performance requirements.
    Tensile Strength Testing: Withstand pulling forces.Met applicable design and performance requirements.
    Dimensional Testing: Adhere to specified dimensions.Met applicable design and performance requirements.
    Maximum Inflation Volume Testing: Achieve specified max inflation volume.Met applicable design and performance requirements.
    Shelf-Life Testing: Maintain performance over specified shelf life (12 months).Demonstrated to meet requirements throughout its shelf life.
    Simulated Use: Clinically safe for prescription use.Supported conclusion of clinical safety through simulated use results and risk analysis (ISO 14971:2019).
    Biocompatibility (ISO 10993): Safe for patient contact.Fulfilled requirements for External Communicating Device, Circulating Blood, A-Limited (
    Ask a Question

    Ask a specific question about this device

    K Number
    K221843
    Date Cleared
    2022-09-14

    (82 days)

    Product Code
    Regulation Number
    870.4450
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Alabama 35211

    Re: K221843

    Trade/Device Name: GEM™ Biover Microvascular Clamps Regulation Number: 21 CFR 870.4450
    |
    | Classification Name | Clamp, Vascular
    Regulation Number: 21 CFR 870.4450
    |
    | Classification Name | Clamp, Vascular
    Regulation Number: 21 CFR 870.4450

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    GEM™ BIOVER disposable microvascular clamps are instruments which are used for all microsurgical procedures. They are used to occlude vessels during anastomosis, which is necessary as a result of vessel damage or thrombosis.

    Device Description

    The GEM™ Biover Microvascular Clamps (Clamps) are arterial and venous clamps in both single and double clamp configurations, available in varying sizes. The clamps are sterile, disposable clamps for single use. The clamps are used in microsurgery, plastic surgery, and hand surgery for the temporary occlusion of blood vessels during an anastomosis.

    The clamps are available for arteries and veins. The clamping force is higher for the arterial clamps than for the veinous clamps. The clamps are identified by a product reference code and by the color of the clamp. The clamps for veins are green, and the clamps for arteries are yellow.

    The clamps are packaged individually in two Tyvek pouches (double pouched). The devices in the Tyvek pouches are gamma sterilized and are placed in cartons of 10. The clamps are used once and are disposed of after use.

    AI/ML Overview

    This document is a 510(k) summary for the GEM™ Biover Microvascular Clamps, asserting their substantial equivalence to a previously cleared predicate device. This type of submission generally focuses on demonstrating that a new device is as safe and effective as a legally marketed predicate device, rather than proving performance against specific quantitative acceptance criteria in a robust statistical study.

    Based on the provided text, the device is not an AI/ML powered device, nor is it a diagnostic device that requires high-level performance metrics, such as sensitivity, specificity, or AUC. The submission is for a physical medical device (microvascular clamps). Therefore, the information requested in the prompt regarding AI/ML performance metrics, sample sizes for test/training sets, expert adjudication, MRMC studies, and ground truth establishment is not applicable to this document.

    The document states:

    • "The GEM Biover microvascular clamps remain unchanged from the predicate device."
    • "There are no engineering or performance changes to the device or its packaging."

    The "performance data" section mentions:

    • "The label design change and the outer carton sales unit design change were assessed and validated in a Packaging Summative Human Factors/Usability Study."
    • "Historical complaint data were reviewed and indicate no use-related concerns, thus, there are no use-related risks or complaints that trigger a need for further Human Factors validation."

    This indicates that the "performance" considered for this submission relates to packaging and user interface (labeling/carton design), and the study conducted was a human factors/usability study, not a clinical performance study measuring accuracy or efficacy of a medical algorithm.

    Therefore, I cannot populate the table or answer the specific questions related to acceptance criteria, AI/ML performance, and ground truth for a diagnostic or AI-powered device, as these details are not present and are not relevant to this type of device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211610
    Date Cleared
    2022-02-14

    (265 days)

    Product Code
    Regulation Number
    870.4450
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    California 92614

    Re: K211610

    Trade/Device Name: Fogarty Occlusion Catheter Regulation Number: 21 CFR 870.4450
    |
    | Regulation
    Class/Product
    Code | 21 CFR 870.4450

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Fogarty Occlusion Catheters are indicated for temporary vessel occlusion.
    The Large Occlusion Catheters are intended to be used for temporary occlusion in the aorta, vena cava and internal jugular vein.
    The Small Occlusion Catheters are intended to be used for temporary occlusion in the peripheral vascular system.

    Device Description

    Models: 620403F, 620404F, and 620405F
    The Fogarty Small Occlusion Catheters (4F, 5F) consist of a single-lumen polyvinylchloride catheter body with a latex balloon at the distal end and a gate valve at the proximal end. The 3F Fogarty Occlusion Catheter is made of Nylon. The catheter lumen is used for inflation of the balloon via a syringe connected to the gate valve. A removable stainless-steel stylet is provided with each catheter to maintain the straight catheter shape and to ensure that the lumen remains opened during storage of the product.
    Models: 62080814F and 62080822F
    The Fogarty 8F Large Occlusion Catheters consist of a single-lumen polyvinylchloride catheter body with a latex balloon at the distal end and a gate valve at the proximal end. The catheter lumen is used for inflation of the balloon via a syringe connected to the gate valve. A removable stainless-steel stylet is provided with each catheter.

    AI/ML Overview

    This document is a 510(k) premarket notification for the Fogarty Occlusion Catheter, indicating substantial equivalence to a predicate device. It's important to note that a 510(k) clearance does not involve clinical studies with human participants where AI performance is measured against human readers or a ground truth established by experts. Instead, it focuses on demonstrating that a new device is as safe and effective as a legally marketed predicate device.

    In this specific case, the submission emphasizes that the new device is identical to the predicate device in terms of intended use, indications for use, and technology. The changes proposed are related to new model numbers for existing devices, a new alternative packaging configuration, new labels, and updates to the Instructions For Use (IFU). Therefore, the "acceptance criteria" and "study" described are focused on proving that these packaging and labeling changes do not alter the device's performance or safety from the established predicate.

    Given this context, many of the typical questions for AI/CADe devices regarding acceptance criteria, ground truth, expert involvement, and MRMC studies are not applicable.

    Here's an interpretation based on the provided document:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance (Summary)
    Packaging IntegrityCompliance with ISO 11607-1 (for terminally sterilized medical devices)Packaging testing was performed and successfully passed, demonstrating the integrity of the new alternative packaging.
    Sterilization EfficacyCompliance with ISO 11135:2014 (ethylene oxide sterilization)Sterilization testing was successfully passed, ensuring sterility assurance.
    Ethylene Oxide (EO)/Esterification (ECH) ResidualsCompliance with ISO 10993-7:2008 (biological evaluation of medical devices)EO/ECH residuals evaluation was successfully passed.
    Product PerformanceMaintain original functional characteristics despite packaging changeProduct performance testing and balloon inflation testing were successfully passed.
    BiocompatibilityEnsure no new biocompatibility concerns due to packaging materialsBiocompatibility testing evaluations were conducted and ensured no alteration in performance.

    2. Sample size used for the test set and the data provenance:

    • The document does not specify sample sizes for each test mentioned (e.g., how many packages were tested for integrity, how many catheters were tested for performance).
    • The data provenance is not explicitly stated in terms of country of origin or whether it was retrospective or prospective. These tests are typically conducted in a laboratory setting by the manufacturer.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This question is not applicable in the context of this 510(k) submission. The "ground truth" here is compliance with established engineering and manufacturing standards (e.g., ISO standards) and performance specifications of the predicate device. There is no expert review of images or clinical data to establish a diagnostic ground truth.

    4. Adjudication method for the test set:

    • This question is not applicable. As there is no "ground truth" based on expert consensus for clinical interpretation, there is no need for an adjudication method. The tests are objective measurements against predefined standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This question is not applicable. This submission is for a physical medical device (an occlusion catheter) and does not involve AI or any form of image interpretation where human readers would be "assisted" by AI.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • This question is not applicable. This device does not have an algorithm or standalone performance in the context of AI.

    7. The type of ground truth used:

    • The "ground truth" in this context refers to compliance with recognized industry standards (e.g., ISO 11607-1, ISO 11135:2014, ISO 10993-7:2008) and engineering performance specifications that are identical to the predicate device.

    8. The sample size for the training set:

    • This question is not applicable. There is no "training set" in the context of an AI/machine learning model for this physical medical device submission.

    9. How the ground truth for the training set was established:

    • This question is not applicable. As there is no training set, there is no ground truth establishment for it.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 19