Search Results
Found 3 results
510(k) Data Aggregation
(258 days)
The CEREGLIDE 92 Intermediate Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the neurovascular system.
The CEREGLIDE™ 92 Intermediate Catheter is a single lumen, variable stiffness catheter designed to be introduced over a steerable guidewire along with a microcatheter and/or compatible support device into the neuro vasculature. The catheter consists of a lubricious PTFE lined inner lumen to facilitate movement of the guidewires and other devices, variable pitch stainless steel and tungsten braid, and various durometer polymer jackets. These jackets provide distal flexibility and gradually transition to a stiffer proximal shaft to facilitate the advancement of the catheter in the anatomy. The outer surface of the catheter is hydrophilic coated in order to reduce friction during manipulation in the vessel. A radiopaque marker at the distal end of the catheter provides fluoroscopic visualization of the catheter tip. The proximal end of the catheter incorporates a standard Luer adapter to facilitate the attachment of accessories, a hub, and an ID band.
The CEREGLIDE™ 92 Intermediate Catheter is packaged with a Tuohy Borst hemostasis valve (RHV) with a side port and two slit introducer accessories. The RHV with side port is used for flushing and insertion of catheters. The slit introducers are designed to introduce the CEREGLIDE™ 92 Intermediate Catheter into the base catheter and protect the distal tip of the CEREGLIDE™ 92 Intermediate Catheter during insertion into the hemostasis valve of the base catheter.
The provided document is a 510(k) Premarket Notification from the FDA regarding the "CEREGLIDE 92 Intermediate Catheter." It details various performance tests conducted on the device to demonstrate its substantial equivalence to a legally marketed predicate device.
However, it's crucial to understand that this document does NOT describe the acceptance criteria and study proving device performance for an AI/ML-driven medical device.
Instead, this document pertains to a traditional, physical medical device (a catheter). The "performance testing" described are typical bench testing (physical and mechanical properties) and biocompatibility testing for hardware, not algorithmic performance.
Therefore, many of the specific questions asked in the prompt, such as "Sample sized used for the test set and the data provenance," "Number of experts used to establish the ground truth," "Adjudication method," "MRMC study," "Standalone performance," and "ground truth for the training set," are not applicable to the content of this document. These questions are relevant to the validation of AI/ML models in medical devices.
I will address the questions to the extent possible with the information provided, explicitly stating when information is not present or not applicable.
Acceptance Criteria and Device Performance for the CEREGLIDE 92 Intermediate Catheter
The CEREGLIDE 92 Intermediate Catheter is a physical medical device, not an AI/ML-driven one. Therefore, the "acceptance criteria" and "device performance" in this context refer to its physical, mechanical, and biological properties rather than algorithmic performance.
1. A table of acceptance criteria and the reported device performance
The document provides a summary of "Performance Testing - Bench" and "Biocompatibility Testing." For each test, the "Result" uniformly states "PASS: Samples met the established acceptance criteria" or simply "PASS." The specific numerical or quantitative acceptance criteria are generally not explicitly stated in the summary tables but are implied by the "PASS" result.
Test Category | Specific Test | Acceptance Criteria (Implicit) | Reported Device Performance |
---|---|---|---|
Design Verification | Visual Inspection | Meets visual requirements described in ISO 10555-1 Section 4.4 | PASS |
Catheter ID | Meets specified internal diameter requirements | PASS | |
Catheter OD | Meets specified outer diameter requirements | PASS | |
Catheter Working Length | Confirms length as defined in ISO 10555-1 Section 3.6 | PASS | |
Catheter Tip Length | Meets specified tip length requirements | PASS | |
Hub Luer Taper | Fits standard Luer fittings | PASS | |
System Air Leakage | No air leak into hub subassembly | PASS | |
System Liquid Leakage | Freedom from leakage during pressurization (ISO 10555-1:2013, section 4.7) | PASS | |
Burst Pressure | Withstands maximum hydrostatic pressure | PASS | |
Hub Pull Testing | Hub-to-joint strength meets acceptance criteria | PASS | |
Shaft Tensile Strength | Joint strength meets acceptance criteria | PASS | |
Particulate Count | Meets particulate size and counts under simulated use | PASS | |
Coating Lubricity and Durability | Lubriciousness and durability of hydrophilic coating | PASS | |
Coating Length | Hydrophilic coating length meets design requirements | PASS | |
Kink Resistance (Distal & Proximal) | Remains stable and does not kink during use | PASS | |
Delamination of PTFE Liner | PTFE appropriately adhered to inner lumen | PASS | |
Tip Movement | Meets tip column stiffness requirement | PASS | |
Tip Linear Stiffness | Flexibility relative to other similar devices | PASS | |
Trackability | Meets trackability requirement | PASS | |
Torque Strength | Meets torque strength requirement | PASS | |
Introducer Working Length | Confirms working length of introducer | PASS | |
Introducer Separation Force | Confirms force required to separate introducer | PASS | |
Design Validation | In Vitro Usability Studies | Evaluates usability parameters (trackability, tip stability, durability, device compatibility) | PASS |
Biocompatibility | Cytotoxicity | Grade ≤ 2 | PASS |
Sensitization - Guinea Pig Maximization | Grade 50% of the negative control | PASS | |
ASTM Heparinized Platelet and Leukocyte Count | % platelet and leukocyte comparison to control article | PASS | |
Comparative Surface Assessment | 150x microscopic comparative evaluation of surface and geometric features | PASS |
2. Sample sized used for the test set and the data provenance
The document states: "All testing was conducted using sampling methods as required by internal design control procedures." However, the specific sample sizes for each test are not provided.
Data provenance: Given that these are bench tests for a physical device, the "data provenance" would refer to the testing conditions and lab environment, rather than patient data. The tests were conducted internally by the manufacturer. No country of origin for specific "data" (in the sense of patient data) is applicable. This was a non-clinical study involving bench testing and in-vitro/in-vivo animal tests for biocompatibility, not a clinical trial involving human patients or patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This device is a physical catheter, not an AI/ML algorithm requiring expert ground truth for image interpretation or diagnosis. The "ground truth" for these tests refers to established engineering specifications, ISO standards, and biological safety limits.
4. Adjudication method for the test set
Not applicable. There is no "adjudication" in the sense of reconciling expert opinions on data. Adjudication methods like 2+1 or 3+1 are used for human-labeled datasets, typically for AI/ML models.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. MRMC studies are performed to evaluate the diagnostic performance of AI-assisted human readers compared to human readers alone, typically in medical imaging or diagnostics. This document is about a physical interventional device, not a diagnostic AI tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This is a physical device, not an algorithm.
7. The type of ground truth used
For the bench testing and biocompatibility assessments, the "ground truth" is based on:
- Established engineering specifications and design requirements for the catheter's physical dimensions, mechanical properties (tensile strength, burst pressure, kink resistance, trackability, etc.).
- International standards (e.g., ISO 10555-1) for catheter performance.
- Regulatory guidance and standards for biocompatibility (e.g., ISO 10993-1, ISO 11135 for sterilization, EN ISO 10993-7 for EO residuals).
- Comparison to predicate device characteristics to demonstrate substantial equivalence.
8. The sample size for the training set
Not applicable. This is not an AI/ML device that requires a "training set."
9. How the ground truth for the training set was established
Not applicable. As above, no training set for an AI/ML model.
Ask a specific question about this device
(156 days)
The PROWLER SELECT PLUS Microcatheter is intended for the introduction of interventional devices, delivery of therapeutic devices, and infusion of diagnostic agents into the peripheral and neuro vasculature.
The PROWLER EX Microcatheter is intended for the introduction of interventional devices, delivery of therapeutic devices, and infusion of diagnostic agents into the peripheral and neuro vasculature.
The PROWLER SELECT PLUS Microcatheter and PROWLER EX Microcatheter are variable stiffness, single lumen catheters designed to access small, tortuous vasculature. They are available in a variety of outer and inner diameters. Each configuration has a hydrophilic coating to provide lubricity for navigation of vessels. The inner lumen is lined with lubricious PTFE to facilitate movement of guidewires and other devices. The distal sections of the catheter bodies are radiopaque to aid visualization under fluoroscopy, and the distal tips are clearly distinguished by a radiopaque marker.
The provided FDA 510(k) summary (K210838) is for a medical device, specifically a microcatheter. It focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance testing. The document does not describe studies proving an AI/ML powered device meets acceptance criteria.
Therefore, I cannot extract the requested information regarding AI/ML acceptance criteria, performance tables, sample sizes, expert involvement, or MRMC studies from the provided text, as those concepts are not applicable to the device described (a physical microcatheter).
The document details performance testing for this physical device as follows:
Acceptance Criteria and Device Performance (Not applicable to AI/ML):
The document describes non-clinical performance testing for the PROWLER SELECT PLUS Microcatheter and PROWLER EX Microcatheter. The acceptance criteria are implicit in the "Test Method Summary" and the "Results" sections.
Test | Test Method Summary | Results (Device Performance) |
---|---|---|
Simulated Use Testing in a tortuous anatomical model | To provide evidence that the subject devices can safely and effectively deliver interventional devices to the peripheral and neurovasculature. The test method and anatomical model were the same established for reference devices K162563 and K191237. | Test sample microcatheters successfully delivered interventional devices, including vascular stents, stent retrievers, and aneurysm coils. Results demonstrated substantial equivalence for delivery of interventional devices. |
Particulate Testing | To evaluate and compare the quantity of particles generated by the subject device during simulated device delivery in a tortuous anatomical model versus particles generated by an applicable reference device (K131437). The test method and anatomical model were the same established for reference devices K162563 and K191237. | Particle generation from baseline and aged subject devices were comparable with the reference device (K131437). |
Static Burst Pressure Testing | This test verified that the subject devices meet the established static burst pressure specification, after simulating the delivery of an interventional device. Static burst pressure specifications were the same as those established for predicate K021591. The test method was the same established for reference devices K162563 and K191237. | All tested baseline and aged samples met the minimum static burst pressure acceptance criteria with a demonstrated 95% confidence and 99% reliability and substantial equivalence for delivery of interventional devices. |
Regarding the specific questions asked for an AI/ML device:
- A table of acceptance criteria and the reported device performance: Provided above for the physical device.
- Sample sized used for the test set and the data provenance: Not explicitly stated (e.g., number of microcatheters tested in "test sample microcatheters" or "all tested baseline and aged samples"). The data is from non-clinical bench testing, not patient data (so no country of origin, retrospective/prospective information).
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a physical microcatheter's performance is based on direct measurement and observation of physical properties and simulated use, not expert interpretation of AI output.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is a physical medical device, not an AI-powered diagnostic/analytic tool.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Ground truth for this device is based on established engineering specifications, physical testing standards, and direct observation of mechanical performance (e.g., successful delivery, particle generation counts, burst pressure measurements).
- The sample size for the training set: Not applicable (no AI model).
- How the ground truth for the training set was established: Not applicable (no AI model).
Ask a specific question about this device
(228 days)
The CERENOVUS Large Bore Catheter, with the CERENOVUS® Aspiration Tubing Set and NOUVAG Vacuson 60 aspiration pump (or equivalent aspiration pump), is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA are candidates for treatment.
The CERENOVUS® Aspiration Tubing Set is intended to connect the CERENOVUS Large Bore Catheter to the canister of the NOUVAG Vacuson 60 Aspiration Pump (or equivalent vacuum pump) and to allow the user to control the fluid flow.
The CERENOVUS Large Bore Catheter is a variable stiffness, single lumen catheter designed to be introduced over a steerable guide wire or microcatheter into the neuro vasculature. The catheter shaft is composed of a stainless steel variable pitch braid with a PTFE inner liner to facilitate movement of guide wires and other devices. The exterior of the catheter shaft is covered with polymer materials, which encapsulate the stainless steel braid construction. The catheter has a stiff proximal shaft which transitions into the flexible distal shaft to facilitate the advancement of the catheter in the anatomy. The distal end of the catheter has a radiopaque marker band to facilitate fluoroscopic visualization and has a hydrophilic coating to provide lubricity for navigation of vessels. The proximal end of the catheter has a luer fitting located on the end of the catheter hub which can be used to attach accessories for flushing and aspiration. An ID band is placed at the distal end of the hub over a strain relief. The catheter is packaged with a hemostasis valve with a side port and two peel-away introducers as accessories. The hemostasis valve with side port is used for flushing, insertion of catheters, and connection to an external aspiration system. The peel away introducer sheaths are designed to protect the distal tip of the catheter during insertion into the hemostasis valve.
The CERENOVUS Large Bore Catheter can be connected to the NOUVAG Vacuson 60 aspiration pump (or equivalent aspiration pump) using the CERENOVUS® Aspiration Tubing Set.
The provided text describes a 510(k) premarket notification for a medical device (CERENOVUS Large Bore Catheter and Tubing Set), which focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance, animal, sterilization, shelf-life, and biocompatibility testing. It does not include information about AI/ML device performance, human reader studies, or the establishment of ground truth by expert consensus. Therefore, I cannot generate a response that fulfills the requirements of the prompt regarding acceptance criteria for an AI/ML device or studies proving its performance.
The document is purely about a physical medical device and its mechanical/physical/biological properties, not an AI/ML diagnostic or assistive technology.
Ask a specific question about this device
Page 1 of 1