Search Results
Found 2 results
510(k) Data Aggregation
(56 days)
The Radical the Dude 8F Guide Catheter is indicated for the introduction of intravascular catheters into the peripheral, coronary, and neuro vasculature.
The Radical the Dude 8F Guide Catheter (Dude 8F Catheter) is an 8 French (Fr) guide catheter designed to aid the physician in accessing the target vasculature during interventional procedures. The Dude 8F Catheter has a usable length between 80 cm and 110 cm, and an outer diameter (OD) size designation of 8 Fr. The Dude 8F Catheter has variable stiffness along its length, incorporating hybrid ribbon technologies to maintain stability and vary stiffness along the device length. The distal portion of the Dude 8F Catheter has a hydrophilic coating. The Dude 8F Catheter is packaged with a rotating hemostasis valve (RHV) and a peel-away sheath.
The provided FDA 510(k) summary (K243577) describes a medical device, the Radical the Dude 8F Guide Catheter, and its substantial equivalence to a predicate device. However, this document does not describe a study involving AI or software performance evaluation with human-in-the-loop scenarios.
Therefore, most of the questions regarding acceptance criteria, study details, sample sizes, expert involvement, and ground truth establishment, which are typically relevant for assessing the performance of AI/software medical devices, cannot be answered based on the provided text.
The 510(k) pertains to a physical catheter device and details traditional bench testing and biocompatibility assessments to establish substantial equivalence with a previously cleared device.
Based on the provided document, here's what can be extracted, and where limitations exist:
-
A table of acceptance criteria and the reported device performance
The document provides a summary of "Nonclinical Performance Testing" (Bench Testing and Biocompatibility). The acceptance "criteria" are implied by the "Pass" result for each test.
Table of Acceptance Criteria and Reported Device Performance
Test | Test Method/Applicable Standard | Acceptance Criteria (Implied by "Pass") | Reported Device Performance |
---|---|---|---|
Visual Inspection | Visual inspection completed for surface defects. | No surface defects | Pass |
Dimensional Inspection | Critical dimensions were verified. | Critical dimensions met specifications | Pass |
Simulated Use Test | Simulated use in a bench anatomical model with femoral artery access. | Successful use in simulated anatomical model | Pass |
PTFE Delamination | Assessed for PTFE delamination at distal tip following simulated use testing. | No PTFE delamination | Pass |
Tensile Testing | Tensile strength measured along entire catheter length. | Tensile strength met specifications | Pass |
Torque Strength | The distal end of the catheter was constrained from movement while the proximal end was turned until failure in a simulated anatomy model. | Torque strength met specifications | Pass |
Kink Resistance | Resistance to kink was tested at various locations along the catheter shaft using successively smaller radii to challenge the catheter. | Acceptable kink resistance | Pass |
Catheter Burst | Catheter burst tested per ISO 10555-1. | Met ISO 10555-1 burst requirements | Pass |
Liquid Leak Test | Liquid leak tested per ISO 10555-1. | Met ISO 10555-1 liquid leak requirements | Pass |
Air Leak Test | Air leak tested per ISO 10555-1. | Met ISO 10555-1 air leak requirements | Pass |
Hydrophilic Coating Integrity | The integrity of the hydrophilic coating was inspected before and after simulated use testing in an in vitro model. | Hydrophilic coating integrity maintained | Pass |
Particulate Testing | During simulated use testing in an in vitro model the particle size and count were analyzed using light obscuration method and compared to the reference device. | Particle size and count comparable to reference device | Pass |
Tip Stiffness | Compared the tip stiffness of the Dude 8F Catheter with the reference device. | Tip stiffness comparable to reference device | Pass |
Biocompatibility: | |||
Hemocompatibility: Partial Thromboplastin Time (PTT) | The test article average clotting time was higher and was not statistically significant when compared to the negative control. The test article average clotting time was lower and was not statistically significant when compared to the vehicle control. | PTT similar to the reference device | Pass |
Thrombogenicity in Canine Model | An in vivo canine test to evaluate the thrombogenic potential of the subject device compared to a comparator device. | Thrombogenicity similar to reference device | Pass |
-
Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- The document describes bench testing and biocompatibility testing. Specific sample sizes for each bench test are not specified in this summary.
- For the "Thrombogenicity in Canine Model", it used an in vivo canine test, implying an animal study. The exact number of animals or the test setup details are not provided in this summary.
- There is no mention of "test sets" in the context of clinical data or patient data, as this is a device for physical performance and biocompatibility.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable. The tests are physical or chemical evaluations (e.g., visual inspection, dimensional inspection, tensile strength, PTT). "Ground truth" in the context of expert review for diagnostic accuracy is not relevant here.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. This is not a study requiring adjudication of expert opinions for a test set.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. An MRMC study was not done, as this is neither an AI device nor does it involve human readers interpreting images.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- No. This is not an algorithm or AI device.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- For bench testing, the "ground truth" is established by adherence to engineering specifications, relevant ISO standards (e.g., ISO 10555-1), and comparison to the predicate/reference device's performance. For biocompatibility, it's based on standard biological evaluation methods and comparison to controls or reference devices.
-
The sample size for the training set
- Not applicable. This is not an AI/ML device employing a training set.
-
How the ground truth for the training set was established
- Not applicable. There is no training set for this device.
Ask a specific question about this device
(258 days)
The CEREGLIDE 92 Intermediate Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the neurovascular system.
The CEREGLIDE™ 92 Intermediate Catheter is a single lumen, variable stiffness catheter designed to be introduced over a steerable guidewire along with a microcatheter and/or compatible support device into the neuro vasculature. The catheter consists of a lubricious PTFE lined inner lumen to facilitate movement of the guidewires and other devices, variable pitch stainless steel and tungsten braid, and various durometer polymer jackets. These jackets provide distal flexibility and gradually transition to a stiffer proximal shaft to facilitate the advancement of the catheter in the anatomy. The outer surface of the catheter is hydrophilic coated in order to reduce friction during manipulation in the vessel. A radiopaque marker at the distal end of the catheter provides fluoroscopic visualization of the catheter tip. The proximal end of the catheter incorporates a standard Luer adapter to facilitate the attachment of accessories, a hub, and an ID band.
The CEREGLIDE™ 92 Intermediate Catheter is packaged with a Tuohy Borst hemostasis valve (RHV) with a side port and two slit introducer accessories. The RHV with side port is used for flushing and insertion of catheters. The slit introducers are designed to introduce the CEREGLIDE™ 92 Intermediate Catheter into the base catheter and protect the distal tip of the CEREGLIDE™ 92 Intermediate Catheter during insertion into the hemostasis valve of the base catheter.
The provided document is a 510(k) Premarket Notification from the FDA regarding the "CEREGLIDE 92 Intermediate Catheter." It details various performance tests conducted on the device to demonstrate its substantial equivalence to a legally marketed predicate device.
However, it's crucial to understand that this document does NOT describe the acceptance criteria and study proving device performance for an AI/ML-driven medical device.
Instead, this document pertains to a traditional, physical medical device (a catheter). The "performance testing" described are typical bench testing (physical and mechanical properties) and biocompatibility testing for hardware, not algorithmic performance.
Therefore, many of the specific questions asked in the prompt, such as "Sample sized used for the test set and the data provenance," "Number of experts used to establish the ground truth," "Adjudication method," "MRMC study," "Standalone performance," and "ground truth for the training set," are not applicable to the content of this document. These questions are relevant to the validation of AI/ML models in medical devices.
I will address the questions to the extent possible with the information provided, explicitly stating when information is not present or not applicable.
Acceptance Criteria and Device Performance for the CEREGLIDE 92 Intermediate Catheter
The CEREGLIDE 92 Intermediate Catheter is a physical medical device, not an AI/ML-driven one. Therefore, the "acceptance criteria" and "device performance" in this context refer to its physical, mechanical, and biological properties rather than algorithmic performance.
1. A table of acceptance criteria and the reported device performance
The document provides a summary of "Performance Testing - Bench" and "Biocompatibility Testing." For each test, the "Result" uniformly states "PASS: Samples met the established acceptance criteria" or simply "PASS." The specific numerical or quantitative acceptance criteria are generally not explicitly stated in the summary tables but are implied by the "PASS" result.
Test Category | Specific Test | Acceptance Criteria (Implicit) | Reported Device Performance |
---|---|---|---|
Design Verification | Visual Inspection | Meets visual requirements described in ISO 10555-1 Section 4.4 | PASS |
Catheter ID | Meets specified internal diameter requirements | PASS | |
Catheter OD | Meets specified outer diameter requirements | PASS | |
Catheter Working Length | Confirms length as defined in ISO 10555-1 Section 3.6 | PASS | |
Catheter Tip Length | Meets specified tip length requirements | PASS | |
Hub Luer Taper | Fits standard Luer fittings | PASS | |
System Air Leakage | No air leak into hub subassembly | PASS | |
System Liquid Leakage | Freedom from leakage during pressurization (ISO 10555-1:2013, section 4.7) | PASS | |
Burst Pressure | Withstands maximum hydrostatic pressure | PASS | |
Hub Pull Testing | Hub-to-joint strength meets acceptance criteria | PASS | |
Shaft Tensile Strength | Joint strength meets acceptance criteria | PASS | |
Particulate Count | Meets particulate size and counts under simulated use | PASS | |
Coating Lubricity and Durability | Lubriciousness and durability of hydrophilic coating | PASS | |
Coating Length | Hydrophilic coating length meets design requirements | PASS | |
Kink Resistance (Distal & Proximal) | Remains stable and does not kink during use | PASS | |
Delamination of PTFE Liner | PTFE appropriately adhered to inner lumen | PASS | |
Tip Movement | Meets tip column stiffness requirement | PASS | |
Tip Linear Stiffness | Flexibility relative to other similar devices | PASS | |
Trackability | Meets trackability requirement | PASS | |
Torque Strength | Meets torque strength requirement | PASS | |
Introducer Working Length | Confirms working length of introducer | PASS | |
Introducer Separation Force | Confirms force required to separate introducer | PASS | |
Design Validation | In Vitro Usability Studies | Evaluates usability parameters (trackability, tip stability, durability, device compatibility) | PASS |
Biocompatibility | Cytotoxicity | Grade ≤ 2 | PASS |
Sensitization - Guinea Pig Maximization | Grade 50% of the negative control | PASS | |
ASTM Heparinized Platelet and Leukocyte Count | % platelet and leukocyte comparison to control article | PASS | |
Comparative Surface Assessment | 150x microscopic comparative evaluation of surface and geometric features | PASS |
2. Sample sized used for the test set and the data provenance
The document states: "All testing was conducted using sampling methods as required by internal design control procedures." However, the specific sample sizes for each test are not provided.
Data provenance: Given that these are bench tests for a physical device, the "data provenance" would refer to the testing conditions and lab environment, rather than patient data. The tests were conducted internally by the manufacturer. No country of origin for specific "data" (in the sense of patient data) is applicable. This was a non-clinical study involving bench testing and in-vitro/in-vivo animal tests for biocompatibility, not a clinical trial involving human patients or patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This device is a physical catheter, not an AI/ML algorithm requiring expert ground truth for image interpretation or diagnosis. The "ground truth" for these tests refers to established engineering specifications, ISO standards, and biological safety limits.
4. Adjudication method for the test set
Not applicable. There is no "adjudication" in the sense of reconciling expert opinions on data. Adjudication methods like 2+1 or 3+1 are used for human-labeled datasets, typically for AI/ML models.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. MRMC studies are performed to evaluate the diagnostic performance of AI-assisted human readers compared to human readers alone, typically in medical imaging or diagnostics. This document is about a physical interventional device, not a diagnostic AI tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This is a physical device, not an algorithm.
7. The type of ground truth used
For the bench testing and biocompatibility assessments, the "ground truth" is based on:
- Established engineering specifications and design requirements for the catheter's physical dimensions, mechanical properties (tensile strength, burst pressure, kink resistance, trackability, etc.).
- International standards (e.g., ISO 10555-1) for catheter performance.
- Regulatory guidance and standards for biocompatibility (e.g., ISO 10993-1, ISO 11135 for sterilization, EN ISO 10993-7 for EO residuals).
- Comparison to predicate device characteristics to demonstrate substantial equivalence.
8. The sample size for the training set
Not applicable. This is not an AI/ML device that requires a "training set."
9. How the ground truth for the training set was established
Not applicable. As above, no training set for an AI/ML model.
Ask a specific question about this device
Page 1 of 1