K Number
K193380

Validate with FDA (Live)

Date Cleared
2020-07-20

(228 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CERENOVUS Large Bore Catheter, with the CERENOVUS® Aspiration Tubing Set and NOUVAG Vacuson 60 aspiration pump (or equivalent aspiration pump), is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA are candidates for treatment.

The CERENOVUS® Aspiration Tubing Set is intended to connect the CERENOVUS Large Bore Catheter to the canister of the NOUVAG Vacuson 60 Aspiration Pump (or equivalent vacuum pump) and to allow the user to control the fluid flow.

Device Description

The CERENOVUS Large Bore Catheter is a variable stiffness, single lumen catheter designed to be introduced over a steerable guide wire or microcatheter into the neuro vasculature. The catheter shaft is composed of a stainless steel variable pitch braid with a PTFE inner liner to facilitate movement of guide wires and other devices. The exterior of the catheter shaft is covered with polymer materials, which encapsulate the stainless steel braid construction. The catheter has a stiff proximal shaft which transitions into the flexible distal shaft to facilitate the advancement of the catheter in the anatomy. The distal end of the catheter has a radiopaque marker band to facilitate fluoroscopic visualization and has a hydrophilic coating to provide lubricity for navigation of vessels. The proximal end of the catheter has a luer fitting located on the end of the catheter hub which can be used to attach accessories for flushing and aspiration. An ID band is placed at the distal end of the hub over a strain relief. The catheter is packaged with a hemostasis valve with a side port and two peel-away introducers as accessories. The hemostasis valve with side port is used for flushing, insertion of catheters, and connection to an external aspiration system. The peel away introducer sheaths are designed to protect the distal tip of the catheter during insertion into the hemostasis valve.

The CERENOVUS Large Bore Catheter can be connected to the NOUVAG Vacuson 60 aspiration pump (or equivalent aspiration pump) using the CERENOVUS® Aspiration Tubing Set.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device (CERENOVUS Large Bore Catheter and Tubing Set), which focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance, animal, sterilization, shelf-life, and biocompatibility testing. It does not include information about AI/ML device performance, human reader studies, or the establishment of ground truth by expert consensus. Therefore, I cannot generate a response that fulfills the requirements of the prompt regarding acceptance criteria for an AI/ML device or studies proving its performance.

The document is purely about a physical medical device and its mechanical/physical/biological properties, not an AI/ML diagnostic or assistive technology.

{0}------------------------------------------------

July 20, 2020

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Codman & Shurtleff, Inc. Kirsten Franco, MS, RAC Associate Director of Regulatory Affairs 325 Paramount Drive Raynham, MA 02767

Re: K193380

Trade/Device Name: CERENOVUS Large Bore Catheter; CERENOVUS Aspiration Tubing Set Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NRY Dated: April 21, 2020 Received: April 22, 2020

Dear Kirsten Franco:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Naira Muradyan, Ph.D. Assistant Director (Acting) DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K193380

Device Name

CERENOVUS Large Bore Catheter; CERENOVUS® Aspiration Tubing Set

Indications for Use (Describe)

The CERENOVUS Large Bore Catheter, with the CERENOVUS® Aspiration Tubing Set and NOUVAG Vacuson 60 aspiration pump (or equivalent aspiration pump), is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA are candidates for treatment.

The CERENOVUS® Aspiration Tubing Set is intended to connect the CERENOVUS Large Bore Catheter to the canister of the NOUVAG Vacuson 60 Aspiration Pump (or equivalent vacuum pump) and to allow the user to control the fluid flow.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K193380

Pursuant to the requirements of 21 CFR Section 807.92(c), this 510(k) summary is provided as part of this Premarket Notification containing sufficient details to understand the basis for a determination of substantial equivalence.

SubmitterCodman & Shurtleff, Inc.325 Paramount DriveRaynham, MA 02767
ContactKirsten FrancoPhone: (484) 868-7991Email: kfranco5@its.jnj.com
Date PreparedJuly 15, 2020
Device Tradeor ProprietaryNameCERENOVUS Large Bore CatheterCERENOVUS® Aspiration Tubing Set
DeviceClassificationRegulatory Classification: IICommon or Usual Name: Catheter, Thrombus RetrieverClassification Name: Percutaneous CatheterRegulation Number: 21 CFR 870.1250Product Code: NRYClassification Panel: Neurology
Predicate DevicePrimary Predicate Device
--------------------------------------------
510(k)NumberDate ClearedDevice NameManufacturer
K161064June 12, 2016Penumbra System ACE 68Reperfusion CatheterPenumbra, Inc.
Reference Predicate Device
------------------------------
510(k)NumberDate ClearedDevice NameManufacturer
K191237Nov 8, 2019CERENOVUS Large BoreCatheterMedos InternationalSARL

{4}------------------------------------------------

DeviceDescriptionThe CERENOVUS Large Bore Catheter is a variable stiffness, single lumen catheterdesigned to be introduced over a steerable guide wire or microcatheter into the neurovasculature. The catheter shaft is composed of a stainless steel variable pitch braidwith a PTFE inner liner to facilitate movement of guide wires and other devices. Theexterior of the catheter shaft is covered with polymer materials, which encapsulatethe stainless steel braid construction. The catheter has a stiff proximal shaft whichtransitions into the flexible distal shaft to facilitate the advancement of the catheter inthe anatomy. The distal end of the catheter has a radiopaque marker band to facilitatefluoroscopic visualization and has a hydrophilic coating to provide lubricity fornavigation of vessels. The proximal end of the catheter has a luer fitting located onthe end of the catheter hub which can be used to attach accessories for flushing andaspiration. An ID band is placed at the distal end of the hub over a strain relief. Thecatheter is packaged with a hemostasis valve with a side port and two peel-awayintroducers as accessories. The hemostasis valve with side port is used for flushing,insertion of catheters, and connection to an external aspiration system. The peelaway introducer sheaths are designed to protect the distal tip of the catheter duringinsertion into the hemostasis valve.The CERENOVUS Large Bore Catheter can be connected to the NOUVAG Vacuson60 aspiration pump (or equivalent aspiration pump) using the CERENOVUS®Aspiration Tubing Set.
Indications forUseThe CERENOVUS Large Bore Catheter, with the CERENOVUS® Aspiration TubingSet and NOUVAG Vacuson 60 aspiration pump (or equivalent aspiration pump), isindicated for use in the revascularization of patients with acute ischemic strokesecondary to intracranial large vessel occlusive disease (within the internal carotid,middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hoursof symptom onset. Patients who are ineligible for intravenous tissue plasminogenactivator (IV t-PA) or who failed IV t-PA are candidates for treatment.The CERENOVUS® Aspiration Tubing Set is intended to connect the CERENOVUSLarge Bore Catheter to the canister of the NOUVAG Vacuson 60 Aspiration Pump(or equivalent vacuum pump) and to allow the user to control the fluid flow.
PredicateComparisonA comparison of the similarities and differences of product features between theCERENOVUS Large Bore Catheter and the primary predicate device is presented in

Table 1.

DescriptionSubject Device:CERENOVUS Large Bore CatheterPrimary Predicate Device:ACE 68 Reperfusion Catheter(K161064)
Product CodeNRYSame
Regulatory NameCatheter, PercutaneousSame
ClassificationClass II - 21 CFR 870.1250Same
Basic DesignVariable stiffness single lumen catheterSame
DescriptionSubject Device:CERENOVUS Large Bore CatheterPrimary Predicate Device:ACE 68 Reperfusion Catheter(K161064)
Indications For UseThe CERENOVUS Large Bore Catheter, with theCERENOVUS® Aspiration Tubing Set and NOUVAGVacuson 60 aspiration pump (or equivalent aspirationpump), is indicated for use in the revascularization ofpatients with acute ischemic stroke secondary tointracranial large vessel occlusive disease (within theinternal carotid, middle cerebral - M1 and M2 segments,basilar, and vertebral arteries) within 8 hours ofsymptom onset. Patients who are ineligible forintravenous tissue plasminogen activator (IV t-PA) orwho failed IV t-PA are candidates for treatment.The CERENOVUS® Aspiration Tubing Set is intendedto connect the CERENOVUS Large Bore Catheter to thecanister of the NOUVAG Vacuson 60 Aspiration Pump(or equivalent vacuum pump) and to allow the user tocontrol the fluid flow.The Penumbra System is intended for use inthe revascularization of patients with acuteischemic stroke secondary to intracraniallarge vessel occlusive disease (within theinternal carotid, middle cerebral – M1 andM2 segments, basilar, and vertebralarteries) within 8 hours of symptom onset.
Dimensions:
Length125 - 135 cm115 – 132 cm
ID0.071"0.068"
Distal OD0.081"0.084"
Proximal OD0.0825"0.084"
Catheter CoatingHydrophilicHydrophilic
Coating Length30 cmSame
Materials:
Marker BandMetal Platinum (90%) / Iridium (10%)Same
BraidStainless SteelStainless Steel, Nitinol
LinerPTFE LinerSame
HubPolyamidePolyamide
Strain ReliefPolyamidePolyamide, Stainless Steel
Outer JacketPebax, Urethane, NylonSame
Tip ConfigurationNon-shapeable tipSteam shapeable by user
Accessories Included:
Hemostasis valveHemostasis Valve with Side Port ExtensionTubingRotating Hemostasis Valve
IntroducerSheathPeel-Away Sheath Introducer (2)Peelable Sheath
ShapingN/AShaping Mandrel 0.038" SST
SterilizationMethodEthylene OxideSame
Sterility AssuranceLevel (SAL)10-6Same
PackagingPolyethylene Hoop and Mounting Card, Pouch,CartonSame
Shelf Life1 year8 months
RequiredAdditionalAccessoriesCERENOVUS Aspiration TubingNOUVAG Vacuson 60 PumpPenumbra Hi-Flow TubingPenumbra Pump MAX
Aspiration Pump Requirements
MinimumAspiration Pressure-20 inHg (-68 kPa)Same
MaximumAspiration Pressure-29 inHg (-98 kPa)Same
Flowrate (Air)0 to 60LPM0 to 23 LPM
Aspiration Tubing Requirements
Tubing ID0.110 in minimumSame
Tubing Length112 inSame
Flow ControlMechanismFlow Control SwitchSame

Table 1. Subject and Predicate Device Comparison Summary

{5}------------------------------------------------

{6}------------------------------------------------

Non-Clinical Testing Summary

Performance Testing - Bench

Appropriate testing was identified based on design, risk analyses and the intended use of the CERENOVUS Large Bore Catheter to demonstrate that it is substantially equivalent to the legally marketed Predicate device. The following performance data has been provided in support of the substantial equivalence determination. All testing was conducted using sampling methods as required by Codman & Shurtleff, Inc. Design Control procedures. The bench testing included the following tests:

TestTest SummaryResult
Visual InspectionConfirm that the CERENOVUS Large Bore Catheter meets the visual requirement described in ISO 10555-1 Section 4.4PASS:Samples met the established acceptance criteria
Catheter IDVerify that the catheter internal diameters meet the requirementsPASS:Samples met the established acceptance criteria
Introducer IDVerify that the introducer internal diameters meet the requirementsPASS:Samples met the established acceptance criteria
Catheter ODVerify that the catheter outer diameters meet the requirementsPASS:Samples met the established acceptance criteria
Introducer ODVerify that the introducer outer diameters meet the requirementsPASS:Samples met the established acceptance criteria
Catheter Working LengthConfirm the working length of a catheter as defined in ISO10555-1 Section 3.6.PASS:Samples met the established acceptance criteria
Introducer Working LengthConfirm the working length of the introducerPASS:Samples met the established acceptance criteria
Distal Tip LengthVerify the distal tip length of the catheterPASS:Samples met the established acceptance criteria
Hub Luer TaperVerify that the catheter hub luer taper fit standard luer fittings using a taper devicePASS:Samples met the established acceptance criteria
Air Leak testingVerify that there is no air leak into the hub subassemblyPASS:Samples met the established acceptance criteria
System Liquid LeakageVerify that the catheter joint strength meets the freedom from leakage (liquid during pressurization) requirements of ISO 10555-1:2013, section 4.7PASS:Samples met the established acceptance criteria
Delamination of PTFE LinerVerify that the PTFE has appropriately adhered to the inner lumen of the catheter with braid reinforcementPASS:Samples met the established acceptance criteria
Kink (Distal & Proximal)Confirm that the CERENOVUS Large Bore Catheter meets the requirement for the catheter to remain stable and not kink during usePASS:Samples met the established acceptance criteria
Tip MovementConfirm that the CERENOVUS Large Bore Catheter meets the tip column stiffness requirementPASS:Samples met the established acceptance criteria
Tip Linear StiffnessTest the tip flexibility of the CERENOVUS Large Bore catheter, relative to other devices of similar designPASS:Samples met the established acceptance criteria
Coating Lubricity & DurabilityVerify the lubriciousness and durability of the catheter hydrophilic coatingPASS:Samples met the established acceptance criteria
Coating LengthVerify that the catheter hydrophilic coatinglength meets the design requirementsPASS:Samples met the establishedacceptance criteria
Catheter TensileStrengthVerify that the catheter joint strength meets therequirements of Section 4.5 of ISO 10555-1PASS:Samples met the establishedacceptance criteria
IntroducerSeparation ForceConfirm the force required to separate the peel-away introducer accessoryPASS:Samples met the establishedacceptance criteria
Particle CountVerify that the coating integrity of the catheter'souter surface meets the requirements for contentof Particle Matter in alignment with USP<788>.PASS:Samples met the establishedacceptance criteria
Burst Pressure(static)Confirm the maximum hydrostatic pressure acatheter can withstand using a CrescentHydraulic Burst-leak TesterPASS:Samples met the establishedacceptance criteria
Static Flow RateDetermine the flow rate through a catheterPASS:Samples met the establishedacceptance criteria
Aspiration FlowRateDetermine the aspiration flow rate through acatheter when the catheter is connected to aconstant vacuum source.PASS:Samples met the establishedacceptance criteria
In-vitro UsabilityStudiesEvaluate catheter trackability, tip stability andvisibility under fluoroscopy, aspirationintegrity, ability to aspirate emboli/clot torestore flow and the durability, Subject andPredicate devices were tracked to the target sitewith the provided accessories to performsimulated neurothrombectomy procedure in theneurovascular model that replicates thetortuosity, diameter and location of the arteriesin the neurovasculature.PASS:Samples met theestablished acceptancecriteria

{7}------------------------------------------------

Codman also confirmed that the CERENOVUS Aspiration Tubing Set meets all design and performance requirements through the following bench testing:

TestTest SummaryResult
Dimensional/Visual InspectionConfirm that the Cerenovus Aspiration Tubingmeets all dimensional and visual inspectionspecifications.PASS:Samples met the establishedacceptance criteria
Tensile StrengthConfirm that the Cerenovus Aspiration Tubingmeets the existing tensile strengthspecifications.PASS:Samples met the establishedacceptance criteria
Connection toVacuum PumpSuction Connector of Aspiration TubingAssembly securely attaches to Pump Canisterlid via press fit.PASS:Samples met the establishedacceptance criteria
Connection toCatheterRotating Luer of Aspiration Tubing Assemblysecurely connects to the female luer of thecatheter hemostasis valve.PASS:Samples met the establishedacceptance criteria
Resist ovalization /Lumen PatencyAspiration Tubing Assembly maintainsfunctionality and maintains an open lumen, withno signs of ovalization, at vacuum pressure perproduct specification.PASS:Samples met the establishedacceptance criteria
Resist LeakAspiration Tubing Assembly maintainsfunctionality with no leaks at vacuum pressureper product specification.PASS:Samples met the establishedacceptance criteria
ComponentConnectionsAspiration Tubing Assembly maintainsfunctionality with no detachments of anybonded components during use.PASS:Samples met the establishedacceptance criteria
Flow SwitchFunctionFlow Control Switch arrests fluid flow in theOFF position and restores fluid flow in the ONposition.PASS:Samples met the establishedacceptance criteria

{8}------------------------------------------------

Performance Testing - Animal

Non-clinical animal testing was conducted to evaluate the safety, efficacy, and usability of the CERENOVUS Large Bore Catheter in comparison to the Penumbra ACE 68 Reperfusion Catheter at acute and chronic time points in a porcine model, both in the presence and absence of simulated clot. Non-clinical animal testing was conducted in accordance with 21 CFR Part 58 for Good Laboratory Practice (GLP) for Non-Clinical Laboratory Studies.

Performance Testing - Clinical

No clinical studies were required as appropriate verification and validation of the catheter and packaging modifications were achieved based on the similarities of the proposed device to the predicate device, and from results of bench testing.

Sterilization

The CERENOVUS Large Bore Catheter, as packaged with included accessories, and the CERENOVUS Aspiration Tubing are sterilized using a validated 100% Ethylene Oxide sterilization process to ensure sterility assurance level (SAL) of 106 in accordance with ISO 11135-1. The CERENOVUS Large Bore Catheter and all accessories meet EO residuals per EN ISO 10993-7 for a limited contact delivery system - externally communicating. The CERENOVUS Large Bore Catheter and all accessories are for single use only.

Shelf-Life Testing

The CERENOVUS Large Bore Catheter will have a shelf life of one year based on the successful completion of stability testing. Shelf life testing was performed using standard test methods and acceptance criteria. Prior to aging, all samples were exposed to standard transportation conditioning. Results of testing on the subject device all met established acceptance criteria. The CERENOVUS Aspiration Tubing will have a shelf life of three (3) years based on the successful completion of stability testing conducted by the manufacturer.

Biocompatibility Testing

The CERENOVUS Large Bore Catheter was assessed for biocompatibility in accordance with International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation of Testing within a Risk Management Process." and FDA Guidance for Industry and FDA Staff: Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing' (Issued June 16, 2016), as previously presented in K191237. The Subject device is considered an externally communicating medical device with circulating blood contact for less than 24 hours. The following Biocompatibility Testing was completed as part of this evaluation:

Biocompatibility testing previously presented in K191237 is representative of the subject CERENOVUS Large Bore Catheter because the subject device is comprised of the same materials and manufacturing processes.

{9}------------------------------------------------

Codman also confirmed that the CERENOVUS Aspiration Tubing Set is biocompatible for its intended use. Biological evaluation of the aspiration tubing was conducted pursuant to the recommendations in FDA guidance document titled: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". As the aspiration tubing does not have any patient contact, it was evaluated for intact skin contact as a worstcase scenario. All test results passed, indicating that the aspiration tubing is biocompatible for the intended use.

Summary of Catheter Biocompatibility Testing
TestSummary of ResultsConclusion
Cytotoxicity(MEM Elution)The test article extract showed no evidence of causing celllysis or toxicity (grade = 0). The test article extract met therequirements of the test since the grade was less than a grade2 (mild reactivity).The test article is considerednon-cytotoxic.PASS
Sensitization(MaximizationStudy)The test article extracts showed no evidence of causingdelayed dermal contact sensitization (all erythema scores=0).The test article was notconsidered a sensitizer.PASS
Irritation(IntracutaneousReactivity)The test article met the requirements of the test since thedifference between each test article extract overall meanscore and corresponding control overall mean score was 0.0for both the saline and the sesame oil test article extracts.The test article is considereda negligible irritant.PASS

Conclusion

Based upon the intended use, design, materials, function, side-by-side in-vitro testing and animal testing, it is concluded that the subject device, CERENOVUS Large Bore Catheter is substantially equivalent to the primary predicate device, ACE 68 Catheter (K161064, cleared 12 June 2016). The differences in verbiage in the Indications for Use statement, materials, design, and packaging, do not raise any questions regarding the safety and effectiveness of the device. The device, as designed, manufactured, packaged and sterilized, is substantially equivalent to the primary and referenced predicate device(s) currently marketed under the Federal Food, Drug and Cosmetic Act.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).