Search Filters

Search Results

Found 7 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Triathlon® Hinge Knee System is intended to be implanted with bone cement for the following condition(s):

    • There is destruction of the joint surfaces, with or without significant bone deformity.
    • The cruciate and/or collateral ligaments do not stabilize the knee joint.
    • The ligaments are inadequate and/or the musculature is weak. And/or
    • Revision is required of a failed prosthesis where there has been gross instability, with or without bone loss or inadequate soft tissue.

    When used with MRH femur and/or MRH tibial baseplate replacement indicated in revision of an existing prosthesis:

    • Revision is required of a failed prosthesis where there has been gross instability, with or without bone loss or inadequate soft tissue.

    When used with compatible GMRS components:

    • Where segmental resection and/or replacement of femur and/or proximal tibia is required
    Device Description

    The subject THK components are a line extension of the previously cleared Triathlon® Hinge Knee System (K223528, K230416). The THK System (K223528, K230416) is a tricompartmental knee system consisting of a stemmed femoral component and a stemmed tibial bearing component connected by a set of previously cleared MRH or THK Bushings and an MRH or THK Axle (K994207, K002552, K223528, K230416). A bumper locks this assembly. This assembly provides motion through the MRH or THK Axle/Bushings combination in the flexion/extension plane. The articulation between the bearing surfaces on the underside of a tibial bearing component and a tibial insert provides motion in the rotating plane. A hinge tibial insert is assembled to a Triathlon® Hinge Revision Tibial Baseplate that incorporates a longitudinal bore to accept a previously cleared MRH Tibial Sleeve (K994207, K002552) or Triathlon Tibial Sleeve (K223528, K230416). Optional distal femoral and tibial augments are available to fill bone defects.

    The subject Triathlon® Hinge Small Proximal Tibial Bearing Component and Triathlon® End Cap are sterile, single use devices intended for cemented use only and are being added to the previously cleared THK system (K223528, K230416) as an extension. They can be used with previously cleared MRH Knee components (K994207, K002552), GMRS (K023087, K222056), and Triathlon® Knee System components (K172634, K172326, K190991, K143393, K141056, K132624, K070095, K061521, K053514, K052917, K051948, K051146, K040267).

    AI/ML Overview

    This FDA 510(k) clearance letter is for a mechanical medical device (a knee implant system), not an AI/software as a medical device. Therefore, the request for "acceptance criteria and the study that proves the device meets the acceptance criteria" using metrics relevant to AI/software (like sensitivity, specificity, MRMC studies, ground truth for training/test sets, etc.) is not applicable here.

    The document provided details about the substantial equivalence of the "Triathlon® Hinge Knee System" to previously cleared predicate devices. The "study" proving the device meets acceptance criteria for a mechanical implant typically involves non-clinical (benchtop) testing and engineering analyses to demonstrate performance, material compatibility, and safety, rather than clinical trials or AI performance evaluations with ground truth.

    Here's how the provided information relates to the typical evaluation of such a device, addressing the closest analogues to your requested points:


    Acceptance Criteria and Device Performance for a Mechanical Implant:

    For mechanical implants like the Triathlon® Hinge Knee System, "acceptance criteria" are generally tied to the successful completion of various non-clinical (benchtop) tests and analyses. These tests simulate the physical stresses and conditions the implant will experience in the human body. The "study" proving the device meets these criteria is the execution and successful outcome of these non-clinical tests.

    No direct table of "acceptance criteria" vs. "reported device performance" in the AI sense is provided in this 510(k) summary. Instead, the summary lists the types of non-clinical tests performed to demonstrate performance and substantial equivalence to predicate devices. The implied acceptance is that the device passed these tests, meaning its performance in these simulated scenarios was deemed acceptable and comparable to existing, legally marketed devices.

    Table of Non-Clinical Tests (Analogous to Performance Metrics):

    Acceptance Criteria (Test Type)Reported Device Performance (Implied "Pass")Notes
    Triathlon Hinge Bearing Component Varus/Valgus FatigueMet performance standardsTests the component's ability to withstand repeated bending stresses in side-to-side directions, simulating loading during gait.
    Triathlon Hinge Bearing Component Chair Rise TestingMet performance standardsSimulates stresses experienced during a common activity (standing up from a chair).
    Triathlon Hinge Full Construct FatigueMet performance standardsEvaluates the fatigue life of the entire assembled knee system under cyclic loading.
    Wear Test Rationale for New ConstructsAcceptable wear characteristics demonstratedRationale provided to show that wear performance is acceptable, likely through comparison to predicate devices or established standards. Direct wear testing is implied by "wear test rationale".
    Analysis of Contact Area/Contact Stress & Constraint AnalysisAcceptable contact mechanics & constraintComputer modeling and/or benchtop tests to evaluate how the components interact under load, ensuring appropriate stress distribution and constrained movement.
    Range of Motion and Rotational Freedom Analysis (ASTM F1223-20)Met specified ranges and freedomsTested for flexion/extension, internal/external rotation, varus/valgus, and translation (medial/lateral, proximal/distal, anterior/posterior) to ensure physiological movement.
    Total Femur Construct Compatibility & Triathlon Hinge Bushing/Axle CompatibilityCompatible with legacy componentsEnsured new components integrate properly with existing cleared femoral and tibial components.
    Triathlon Revision Baseplate-End Cap Locking Strength AnalysisAdequate locking strengthEvaluated the mechanical strength of the connection points.
    Triathlon End Cap Tightening AnalysisOptimal tightening confirmedEnsured secure fixation of the end cap.
    Triathlon End Cap Load Carrying Capacity RationaleAcceptable load capacityAnalysis provided to demonstrate the end cap can withstand expected loads.
    Triathlon End Cap Stability AnalysisStable under normal useEvaluated the end cap's ability to remain in place and function without loosening.
    MRI Testing (Safety for imaging) - Displacement, Torque, Artifacts, RF HeatingAll MRI safety criteria metPerformed to confirm the device is safe for patients undergoing MRI procedures (e.g., no excessive magnetic pull, acceptable image distortion, no dangerous heating). Standards: ASTM F2052-15, ASTM F2213-17, ASTM F2119-07 (reapproved 2013), ASTM F2182-191ae.
    Biocompatibility (ISO 10993-1:2020)Biocompatible with human tissueTesting to ensure the materials used in the implant do not cause adverse biological reactions in the body.
    Shelf-life Validation (ISO 11607-1:2019, ISO 11607-2:2019, ASTM F1980-21)Validated shelf-lifeEnsures the sterility and integrity of the packaging system and the device itself are maintained over the specified shelf-life. Methods: ASTM F1886/F1886M-16, ASTM F88/88M-21, ASTM F2096-11(2019).
    Bacterial Endotoxin Testing (ANSI/AAMI ST72:2019)Endotoxin limit
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Triathlon® Hinge Knee System:
    Rotating Hinge Knee System is intended to be implanted with bone cement for the following condition(s):

    • · There is destruction of the joint surfaces, with or without significant bone deformity.
    • · The cruciate and/or collateral ligaments do not stabilize the knee joint.
    • · The ligaments are inadequate and/or the musculature is weak. And/or
    • · Revision is required of a failed prosthesis where has been gross instability, with or without bone loss or inadequate soft tissue.
    • · And/or where segmental resection and replacement of the distal femur is required.

    Triathlon® Revision Insert X3®:
    General Total Knee Arthroplasty (TKR) Indications:
    · Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis, or avascular necrosis), rheumatoid arthritis or post-traumatic arthritis.
    · Post-traumatic loss of knee joint configuration and function.
    · Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
    · Revision of previous unsuccessful knee replacement or other procedure.
    · Fracture of the distal femur and/or proximal tibia that cannot be standard fracture-management techniques. Additional Indications for Total Stabilizer (TS) Components:

    • · Ligamentous instability requiring implant bearing surface geometries with increased constraint.
    • · Absent or non-functioning posterior cruciate ligament.
    • · Severe anteroposterior instability of the knee joint.
    • · Severe instability of the knee secondary to compromised collateral ligament integrity or function.
    Device Description

    The previously cleared Triathlon® Hinge Knee (THK) System (K223528) is a tricompartmental knee system consisting of a stemmed femoral component and a stemmed tibial bearing component connected by a set of Modular Rotating Hinge (MRH) bushings and MRH axle (K222056, K002552, K994207). A bumper locks this assembly. This assembly provides motion through the MRH axle/bushing combination in the flexion/extension plane. The articulation between bearing surfaces on the underside of a tibial bearing component and a hinge tibial insert provide motion in the rotating plane. A hinge tibial insert is assembled to a revision tibial baseplate which incorporates a longitudinal bore to accept a Triathlon® tibial sleeve or an MRH tibial sleeve. Optional distal femoral and tibial augments are available to fill bone defect. The Instructions for Use and package labels for the THK components are being updated to bear the MR Conditional symbol and MR Conditional parameters.

    The subject Triathlon® Bushing and Axle (Standard Assembly Pack) contains sterile, singleuse devices that are being added to the previously cleared THK System (K223528) as an alternate option to MRH bushings and MRH axle to connect a stemmed femoral component and a stemmed tibial bearing component and provide motion through the flexion/extension plane.

    This premarket notification also introduces the subject Triathlon® Revision Insert X3®, which is a sterile, single-use device that is intended for use in a total knee arthroplasty with the previously cleared Triathlon® Revision Tibial Baseplate (K223528) and Triathlon® TS Femoral Component (K172326, K141056, K070095) as part of the Triathlon® Total Knee System. The subject insert is available in seven sizes, and each size is available in seven different thicknesses. The subject insert is packaged together with Cobalt-Chrome (CoCr) stabilizer pin and filler bushing subcomponents that are assembled intraoperatively. The subject insert is assembled to the previously cleared Triathlon® Revision Baseplate (K223528), which incorporates a longitudinal bore to accept the filler bushing subcomponent. The stabilizer pin is inserted through the subject Triathlon® Revision Insert X3® and extends into filler bushing assembled within the Triathlon® Revision Tibial Baseplate to provide additional stability in the insert post.

    AI/ML Overview

    This FDA K-number document (K230416) is for a medical device, specifically orthopaedic implants (knee systems), and thus does not involve acceptance criteria or studies related to AI/ML device performance. The document only lists pre-clinical (non-clinical) testing performed on the device components, such as materials characterization, wear analysis, fatigue testing, and biocompatibility, to demonstrate substantial equivalence to predicate devices. There is no mention of an algorithm or AI model, nor any associated acceptance criteria, study designs, or ground truth establishment relevant to AI/ML performance.

    Therefore, I cannot fulfill your request for information related to AI/ML device acceptance criteria and study details based on the provided document. The document explicitly states: "Clinical testing was not required as a basis for substantial equivalence."

    Ask a Question

    Ask a specific question about this device

    K Number
    K223528
    Date Cleared
    2023-01-20

    (58 days)

    Product Code
    Regulation Number
    888.3510
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This Rotating Hinge Knee System is intended to be implanted with bone cement for the following condition(s):

    • · There is destruction of the joint surfaces, with or without significant bone deformity.
    • · The cruciate and/or collateral ligaments do not stabilize the knee joint.
    • · The ligaments are inadequate and/or the musculature is weak. And/or
    • · Revision is required of a failed prosthesis where there has been gross instability, with or without bone loss or madequate soft tissue.
    • · And/or where segmental resection and replacement of the distal femur is required.
    Device Description

    The subject THK System is a line extension to the existing Triathlon® Knee System and will be a modified version of the predicate Modular Rotating Hinge (MRH) Knee System and Triathlon® Total Knee System. The THK System is a tricompartmental knee system consisting of a new hinge femoral component and a new tibial bearing component connected by a set of previously cleared MRH bushings and an MRH axle (K994207, K002552). A new bumper locks this assembly. This assembly provides motion through the MRH axle/bushings combination in the flexion/extension plane. The articulation between the bearing surfaces on the underside of the new tibial bearing component and a new hinge tibial insert provides motion in the rotating plane. The hinge tibial insert is assembled to a new revision tibial baseplate which incorporates a longitudinal bore to accept a new tibial sleeve or previously cleared MRH tibial sleeve (K994207, K002552). Optional new distal femoral and tibial augments are available to fill bone defects.

    The THK System is designed to provide varus/valgus stability throughout the range of motion, internal/external rotation about the tibial axis, constraint by the bearing surface radius on the tibial bearing component, and an extensive range of size, modularity, and resection options. The subject THK System consists of:

    • Hinge Femoral Components in six sizes and in left and right configurations o
    • Revision Tibial Baseplates in seven sizes O
    • Hinge Inserts in seven sizes and five thicknesses each hinge insert is packaged with a O standard sleeve subcomponent
    • Tibial Bearing Components in three sizes O
    • Bumper inserts in neutral and three degree flexion options O
    • Revision Tibial Augments in eight sizes, two thicknesses and in right medial/left lateral and O right lateral/left medial options
    • o Femoral Distal Augments in six sizes and two thicknesses.

    The components of the subject THK System are sterile, single-use devices intended for cemented use only. They can be used with previously cleared Modular Rotating Hinge (MRH) Knee components (K994207, K002552), the Global Modular Replacement System (GMRS) (K023087), and Triathlon® Knee System components (K172634, K172326, K190991, K143396, K141056. K132624. K070095. K061521. K053514. K052917. K051948. K051146. K040267).

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding the "Triathlon® Hinge Knee System". It describes a medical device, specifically a knee replacement system, and outlines the non-clinical testing performed to establish its substantial equivalence to previously cleared predicate devices.

    Crucially, this document states: "Clinical testing was not required as a basis for substantial equivalence."

    This means that a study proving the device meets acceptance criteria involving human performance or clinical outcomes was not conducted or submitted for this 510(k) clearance due to the nature of the device (a modification/line extension of existing, cleared devices) and the regulatory pathway followed.

    Therefore, I cannot provide the information requested in your prompt regarding:

    • A table of acceptance criteria and reported device performance (in a clinical context)
    • Sample sizes for a test set (clinical)
    • Number of experts for ground truth establishment (clinical)
    • Adjudication method (clinical)
    • MRMC comparative effectiveness study
    • Standalone (algorithm only) performance
    • Type of ground truth used (clinical)
    • Sample size for a training set (clinical)
    • How ground truth for a training set was established (clinical)

    All the testing described in the document under "Non-Clinical Testing" (pages 6-7) relates to engineering analysis, materials characterization, mechanical fatigue testing, wear analysis, biocompatibility, and shelf-life validation. These are physical and mechanical performance tests comparing the new device to the established performance characteristics of predicate devices, not studies involving human subjects or AI performance.

    The document's conclusion reinforces this: "Based upon a comparison of the intended use, indications for use, design, materials and sterilization method, performance characteristics, and operational principles, the components of the subject THK System are substantially equivalent to those of the predicate devices identified in this premarket notification." The substantial equivalence is based on non-clinical data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221048
    Date Cleared
    2022-12-23

    (259 days)

    Product Code
    Regulation Number
    888.3540
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K181280, K071413, K172326

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Episealer Patellofemoral System is intended to be used in patients with osteoarthritis limited to the distal patellofemoral joint, patients with a history of patellar dislocation or patellar fracture, and those patients with failed previous surgery (arthroscopy, tibial tubercle elevation, lateral release, etc.) where pain, deformity or dysfunction persists.

    The device is intended for cemented fixation.

    Device Description

    The Episealer Patellofemoral System is a patient-individualized arthroplasty device which replaces a damaged patellofemoral joint. The subject device consists of two components:

    • . Episealer PF
    • Patellar Component ●

    The Episealer PF component is implanted centrally in the trochlear area of the distal femur. The Patellar Component is implanted on the backside of the patella and articulates with the Episealer PF.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Episealer® Patellofemoral System, an orthopedic device. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving a new device's safety and effectiveness through extensive clinical trials with pre-defined acceptance criteria.

    As such, detailed information regarding acceptance criteria for a specific performance metric, and a study proving those criteria were met in the same way one would assess an AI/ML algorithm, is not available in this document.

    However, I can extract the information that is present regarding performance testing, even though it's not structured around explicit acceptance criteria and corresponding performance metrics in the way you've requested for an AI model.

    Here's an attempt to answer your request based on the provided text, highlighting what is included and what is explicitly not mentioned:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance metrics like "sensitivity", "specificity", "AUC", or "agreement" as would be seen for an AI/ML device. Instead, it describes general performance testing and analyses.

    Performance Aspect (Implied Criteria)Reported Device Performance Statement
    Wear (Criteria: Acceptable wear characteristics comparable to predicate/standard)"Wear testing of the Episealer Patellofemoral System was performed." (No specific wear rates or comparison values are provided in this summary.)
    Contact Area (Criteria: Appropriate load distribution/contact area)"In addition, analyses of contact area... were conducted." (No specific contact area values or comparison values are provided in this summary.)
    Cantilever Bending Strength (Criteria: Sufficient mechanical strength)"...and cantilever bending strength were conducted." (No specific strength values or comparison values are provided in this summary.)
    Component Placement Accuracy (Criteria: Accurate and proper component placement relative to pre-plan)"A cadaver study was conducted to demonstrate that the components of the patient-matched Episealer Patellofemoral System can be accurately placed relative to the pre-planned position, with proper recession and engagement of the Episealer PF and Patellar components." (The statement indicates successful demonstration, but no quantitative metrics for "accuracy" or "proper recession/engagement" are provided.)
    Overall Performance (Criteria: Performs as intended)"The testing, engineering analyses, and cadaver study demonstrate the ability of the Episealer Patellofemoral System to perform as intended in the target population." (General conclusion of successful performance, without specific metrics.)

    Summary of Performance Testing:

    The document states: "Performance Testing Summary: Wear testing of the Episealer Patellofemoral System was performed. In addition, analyses of contact area and cantilever bending strength were conducted. A cadaver study was conducted to demonstrate that the components of the patient-matched Episealer Patellofemoral System can be accurately placed relative to the pre-planned position, with proper recession and engagement of the Episealer PF and Patellar components. The testing, engineering analyses, and cadaver study demonstrate the ability of the Episealer Patellofemoral System to perform as intended in the target population."

    This section indicates that the device underwent standard engineering and mechanical testing relevant to orthopedic implants, as well as a cadaver study to verify surgical placement. However, specific acceptance criteria (e.g., "wear rate must be less than X," "contact area must be within Y range") and the quantitative results against those criteria are not provided in this 510(k) summary. The conclusion is a qualitative statement: "demonstrate the ability... to perform as intended."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not specified in the provided summary. The text only mentions "wear testing," "analyses of contact area and cantilever bending strength," and "A cadaver study." The number of cadavers or individual test repetitions is not given.
    • Data Provenance: Not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable in the context of this device. A medical device like the Episealer Patellofemoral System undergoes engineering, mechanical, and cadaveric testing, not evaluation against expert-defined "ground truth" labels in the way an AI/ML diagnostic tool would.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable for the type of testing described (mechanical testing, cadaver study).

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. This is an orthopedic implant, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. This is an orthopedic implant, not an algorithm. The device itself is a standalone implant that is surgically placed.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The concept of "ground truth" as used for AI/ML evaluation does not directly apply here. Instead, performance is assessed through:

    • Engineering measurements (wear, contact area, bending strength) against industry standards or internal specifications.
    • Cadaveric surgical validation, where the "truth" would be successful implantation according to surgical protocols and demonstration of intended function.

    8. The sample size for the training set

    This is not applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    This is not applicable. This is not an AI/ML device that requires a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220930
    Date Cleared
    2022-06-02

    (63 days)

    Product Code
    Regulation Number
    888.3520
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    RESTORIS™ Multicompartmental Knee (MCK) System is indicated for single or multi-compartmental knee replacement used in conjunction with RO®, the Robotic Arm Interactive Orthopedic System, in individuals with osteoarthritis or posttraumatic arthritis of the tibiofemoral and/or patellofemoral articular surfaces. The specific knee replacement configurations include:

    • · Medial unicondylar
    • · Lateral unicondylar
    • · Patellofemoral
    • · Medial bi-compartmental (medial unicondylar and patellofemoral)

    RESTORIS™ Multicompartmental Knee (MCK) System is for single use only and is intended for implantation with bone cement.

    Device Description

    The purpose of this Changes Being Effected premarket notification is to add a contraindication for the Restoris Multicompartmental Knee (MCK) System. Additionally, minor modifications are being made to the contraindications for clarity. There is no change to the intended use, indications, design, technological characteristics or operational principles for the devices. The subject implant components are identical to the predicate components. The Restoris MCK system was most recently cleared in K180612, and has also been cleared previously in K172326, K150307, K090763, K082172, K082088, K082081, and K080368.

    AI/ML Overview

    This FDA 510(k) submission [K220930] for the Restoris Multi-Compartmental Knee System does not contain any information regarding acceptance criteria or a study proving the device meets acceptance criteria.

    The document explicitly states:

    • "The purpose of this Changes Being Effected premarket notification is to add a contraindication for the Restoris Multicompartmental Knee (MCK) System. Additionally, minor modifications are being made to the contraindications for clarity." (Page 4)
    • "There is no change to the intended use, indications, design, technological characteristics or operational principles for the devices." (Page 4)
    • "No additional testing was performed as part of this submission, as the only changes being made are to the labeling for the devices. There is no impact to the device design or the physical characteristics of the devices. Testing submitted as part of previously cleared premarket notifications is applicable to this submission." (Page 5)

    Therefore, based on the provided text, I cannot fill in the requested information about acceptance criteria and study results because this particular 510(k) submission is for a labeling change and explicitly states that no new testing was performed. The performance data and acceptance criteria would have been part of previous 510(k) submissions (e.g., K180612, K172326, etc.) referenced in this document.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Reference Devices :

    K051380, K141056, K172326

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    General Total Knee Arthroplasty (TKR) Indications:

    • Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis, or avascular necrosis), rheumatoid arthritis or post-traumatic arthritis.
    • Post-traumatic loss of knee joint configuration and function.
    • Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
    • Revision of previous unsuccessful knee replacement or other procedure.
    • Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard fracture-management techniques.

    The Triathlon Tritanium Tibial Baseplate and Tritanium Metal-Backed Patella components are indicated for both uncemented and cemented use.

    The Triathlon Total Knee System beaded and beaded with Peri-Apatite components are intended for uncemented use only.

    The Triathlon All Polyethylene tibial components are indicated for cemented use only.

    Additional Indications for Posterior Stabilized (PS) and Total Stabilizer (TS) Components:

    • Ligamentous instability requiring implant bearing surface geometries with increased constraint.
    • Absent or non-functioning posterior cruciate ligament.
    • Severe anteroposterior instability of the knee joint.

    Additional Indications for Total Stabilizer (TS) Components:

    • Severe instability of the knee secondary to compromised collateral ligament integrity or function.

    Indications for Bone Augments:

    • Painful, disabling joint disease of the knee secondary to: degenerative arthritis, rheumatoid arthritis, or post-traumatic arthritis, complicated by the presence of bone loss.
    • Salvage of previous unsuccessful total knee replacement or other surgical procedure, accompanied bone loss.

    Additional Indications for Cone Augments

    • Severe degeneration or trauma requiring extensive resection and replacement
    • Femoral and tibial bone voids
    • Metaphyseal reconstruction

    The Triathlon Tritanium® Cone Augments are intended for cemented or cemented or cementless use.

    Device Description

    The Triathlon® Posterior Stabilized (PS) Femoral Component - beaded with Peri-Apatite (PA) and Triathlon® Cruciate Retaining (CR) Femoral Component - beaded with Peri-Apatite (PA) are components of the Triathlon® Total Knee System previously cleared via 510(k)s K051380, K141056 and K172326. The subject Triathlon® Femoral Components are components of the Triathlon® Total Knee System and are used in total knee arthroplasty procedures. The subject Triathlon® Femoral Components are available in 8 different sizes and in left and right configurations that are anatomically suited to the left or right knee. The subject Triathlon® Femoral Components are manufactured from cast cobalt-chrome-molybdenum alloy (ASTM F75) with a cobalt-chrome-molybdenum alloy (ASTM F75) double layer porous coating (beading) on the femoral fixation surface. Additionally, the subject Triathlon® Femoral Components are coated with Peri-Apatite (ASTM F1185).

    The design and materials of the subject Triathlon® Femoral Components are identical to the predicate Triathlon® Femoral Components cleared via 510(k)s K051380, K141056 and K172326. The proposed modification is to introduce automated equipment for the first and second porous coating (beading) application process steps for the subject Triathlon® Femoral Components. There are no modifications proposed to the other components of the Triathlon® Total Knee System previously cleared via 510(k)s K051380, K141056 and K172326. The subject Triathlon® Femoral Components will have the same performance specifications as the predicate Triathlon® Femoral Components cleared via 510(k)s K051380, K141056 and K172326.

    AI/ML Overview

    This document describes the Stryker Corporation's 510(k) submission for the Triathlon Posterior Stabilized Femoral Component - beaded with Peri-Apatite and Triathlon Cruciate Retaining Femoral Component - beaded with Peri-Apatite. The submission, K200782, aims to demonstrate substantial equivalence to previously cleared devices.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied by the comparison to predicate devices and the performance of non-clinical testing. The device is considered to meet these criteria if it demonstrates "substantial equivalence" and the proposed modification does not affect safety or effectiveness.

    Acceptance Criteria (Implied)Reported Device Performance
    Substantial equivalence in intended useThe subject Triathlon Femoral Components have the same intended use as the predicate Triathlon Femoral Components.
    Substantial equivalence in indications for useThe subject Triathlon Femoral Components have the same Indications for Use as the predicate Triathlon Femoral Components.
    Substantial equivalence in designThe design of the subject Triathlon Femoral Components is identical to the predicate Triathlon Femoral Components. The proposed modification is to introduce automated equipment for the porous coating application, not a design change of the component itself.
    Substantial equivalence in materialsThe materials of the subject Triathlon Femoral Components are identical to the predicate Triathlon Femoral Components (cast cobalt-chrome-molybdenum alloy (ASTM F75) with a cobalt-chrome-molybdenum alloy (ASTM F75) double layer porous coating, coated with Peri-Apatite).
    Substantial equivalence in performance characteristicsThe subject Triathlon Femoral Components will have the same performance specifications as the predicate Triathlon Femoral Components. Non-clinical testing for porous coating type (beaded, sintered cobalt-chrome coatings) showed substantial equivalence.
    Substantial equivalence in operational principlesNot explicitly detailed, but implied by the overall statement of substantial equivalence.
    Proposed modification does not affect safety or effectivenessNon-clinical testing for porous coating type (beaded, sintered cobalt-chrome coatings) demonstrated that the proposed modification does not affect safety or effectiveness.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a sample size in the context of a test set for clinical performance. The evaluation conducted here is primarily non-clinical, focusing on the comparison of an updated manufacturing process against existing devices.

    • Test Set: Not applicable in the traditional sense of a clinical test set with patient data. The "test set" consists of the subject Triathlon Femoral Components undergoing non-clinical testing.
    • Data Provenance: The testing was non-clinical laboratory testing. No geographical or temporal (retrospective/prospective) information is provided as it's not human subject data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This submission relies on non-clinical testing for performance characteristics, specifically for the porous coating process. The "ground truth" is established through physical and material science testing standards, not expert consensus on medical images or diagnoses.

    4. Adjudication Method for the Test Set

    Not applicable. As described above, the evaluation is based on non-clinical laboratory testing, not a human reader assessment requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a submission for an orthopedic implant (knee femoral component), not an AI-powered diagnostic or assistive device that would involve human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm or AI device. The "performance" being evaluated is the physical and material characteristics of a medical implant after a manufacturing process change.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The ground truth used for proving the device meets acceptance criteria is based on non-clinical laboratory testing standards and direct comparison of material properties and design to legally marketed predicate devices. Specifically, it mentions "Testing for porous coating type a. beaded, sintered cobalt-chrome coatings on a cobalt-chrome substrate as per FDA guidance document, 'Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,' dated April 28, 1994."

    8. The Sample Size for the Training Set

    Not applicable. Since there's no clinical study or AI algorithm involved, there is no "training set" in the context of this submission. The device is a physical implant.

    9. How the Ground Truth for the Training Set was Established

    Not applicable as there is no training set.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Reference Devices :

    K172326

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indications for Use:

    General Total Knee Arthroplasty (TKR) Indications:

    · Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis, or avascular necrosis), rheumatoid arthritis or post-traumatic arthritis.

    • · Post-traumatic loss of knee joint configuration and function.
    • · Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
    • · Revision of previous unsuccessful knee replacement or other procedure.
    • · Fracture of the distal femur and/or proximal tibia that cannot be standard fracture -management techniques.

    The Triathlon® Tritanium® Total Knee System components are indicated for both uncemented use.

    The Triathlon® Total Knee System beaded with Per-Apatite components are intended for uncemented use only.

    The Triathlon® All Polyethylene tibial components are indicated for cemented use only.

    Additional Indications for Posterior Stabilized (PS) and Total Stabilizer (TS) Components:

    • · Ligamentous instability requiring implant bearing surface geometries with increased constraint.
    • · Absent or non-functioning posterior cruciate ligament.
    • · Severe anteroposterior instability of the knee joint.

    Additional Indications for Total Stabilizer (TS) Components:

    · Severe instability of the knee secondary to compromised collateral ligament integrity or function.

    Indications for Bone Augments:

    · Painful, disabling joint disease of the knee secondary to: degenerative arthritis, rheumatoid arthritis, or post-traumatic arthritis, complicated by the presence of bone loss.

    · Salvage of previous unsuccessful total knee replacement or other surgical procedure, accompanied by bone loss.

    Additional Indications for Cone Augments:

    • Severe degeneration or trauma requiring extensive resection and replacement
    • · Femoral and Tibial bone voids
    • · Metaphyseal reconstruction

    The Triathlon TS Cone Augments are intended for cemented or cementless use.

    Device Description

    The additional components being added to the Triathlon Total Knee System are:

    • Size 0 Cruciate Retaining Femoral Component (cemented use) .
    • o Size 0 Primary Cemented Tibial Baseplate
    • Size 0 Tritanium Tibial Baseplate for cemented/cementless use ●
    • Size 0 CR, CS, and PS Tibial Inserts in X3 ETO and N2Vac ●
    • New PS Femoral Component with modified intercondylar PS box (cemented use) ●
    • o New PSR (Posterior Stabilizing Rotation ) tibial insert in sizes 0-8 in X3 ETO

    Additionally, engineering analyses and finite element analyses were presented to document the range of motion available with the Triathlon Total Knee System.

    AI/ML Overview

    This document is a 510(k) summary, which is a premarket notification to the FDA to demonstrate that a device is at least as safe and effective as a legally marketed predicate device. It is not a clinical study report that would typically include detailed acceptance criteria and performance data for an AI/ML device.

    Therefore, the provided text does not contain the information needed to describe acceptance criteria and a study proving a device meets them, especially for an AI-based medical device. The document concerns modifications to an existing total knee system (implants), not an AI/ML device.

    The questions you've asked are highly relevant for the regulatory approval of AI/ML medical devices. If this document were for an AI/ML device, here's how I would break down why the requested information is missing:


    Reasoning for Missing Information (Applicable if this were an AI/ML device submission):

    The provided text is a 510(k) summary for a "Triathlon Total Knee System - Additional Components." This is a premarket notification for an orthopedic implant (knee replacement components), not an AI/ML (Artificial Intelligence/Machine Learning) device.

    The questions posed are specifically for describing the regulatory submission and study design for an AI/ML medical device, which often requires:

    • Demonstrating performance metrics (e.g., sensitivity, specificity, AUC) against established acceptance criteria.
    • Detailing the ground truth establishment process (expert consensus, pathology, etc.).
    • Describing the size and provenance of training and test datasets.
    • Explaining human-in-the-loop studies (multi-reader multi-case studies).

    Since the device under review is a physical medical implant (knee components), these types of studies and acceptance criteria are not applicable. The sponsor instead provides:

    • Engineering analyses and finite element analyses: To document range of motion.
    • Non-clinical testing: Fatigue testing, static shear testing, contact area/stress analysis, range of constraint, sterilization validation, and MR Conditional status.
    • Bacterial endotoxin testing.

    The document explicitly states: "Clinical Testing: Clinical testing was not required as a basis for substantial equivalence." This further confirms that the type of performance evaluation relevant to AI/ML devices was not conducted or required for this orthopedic implant.


    In summary, none of the requested information regarding acceptance criteria, study design for AI/ML device performance, sample sizes for training/test sets, ground truth establishment, or human expert involvement is present in this 510(k) summary because it pertains to a mechanical orthopedic implant, not an AI/ML software device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1