(198 days)
The Stryker Global Modular Replacement System components are sterile, single-use devices intended for use in situations where there is a need for replacement of bone due to radical bone loss. This loss can be related to oncology, trauma or failed previous prosthesis.
Specific Indications for Use are listed below.
Indications for the MRS Stems and Intercalary Stems presented in K952970:
- This device is intended for use in patients requiring extensive reconstruction of the femur and/or proximal tibia, including the hip or knee joint, resulting from extensive bone loss. Tumor resection for skeletal lesions (Oncology patients where radical bone resection and replacement may be required), revision surgery for failed arthroplasty, and acute trauma are the primary causes of extensive bone loss. These prostheses are intended for use with bone cement.
- . The Intercalary System is intended for use in situations arising from femoral mid-shaft turnor resection, or for prosthetic knee fusion.
Indications for the Global Modular Replacement System presented in K023087:
- . Femoral and/or proximal tibial replacement and total femoral replacement in Oncology cases where radical resection and replacement of bone is required, and in limb salvage procedures where radical resection and replacement of the bone is required. Limb salvage procedures would include surgical intervention for severe trauma, failed previous prosthesis, and/or Oncology indications.
Indications for the Global Modular Replacement System Press Fit Stems with HA Coating (presented in K022403, and K031217):
- Proximal femoral reconstruction secondary to: 0
- Trauma o
- Failed previous prosthesis о
- Tumor resection O
Indications for the GMRS Anteverted Proximal Femoral Component (presented in K032581)
-
Femoral replacement in Oncology cases where radical resection and replacement of bone is . required, and in limb salvage procedures where radical resection and replacement of bone is required. Limb salvage procedures would include surgical intervention for severe trauma, failed previous prosthesis, and/or Oncology indications. This component may also be used with the Distal Femoral segment components of the GMRS in total femoral replacement.
Indications for the Modular Replacement System Cemented Stems (cleared in K040749): -
Femoral and/or proximal tibial replacement due to: ●
- Trauma O
- Failed previous prosthesis O
- Tumor resection o
Indications for the Modular Rotating Hinge Knee System (cleared in K002552)
The Modular Rotating Hinge Knee System is intended for use with bone cement in cases where there is destruction of the joint surfaces, with or without significant bone deformity; the cruciate and/or collateral ligaments do not stabilize the knee joint; the ligaments are inadequate and/or the musculature is weak; and revision of a failed previous prosthesis where there is instability, with or without bone loss or inadequate soft tissue.
Expanded indications include for use with bone cement in situations where there is extensive bone loss, including limb salvage procedures for severe trauma, failed previous prosthesis, and/or Oncology indications.
The Global Modular Replacement System is comprised of a number of components that are intended to be used in conjunction with each other, or in conjunction with components of the Modular Replacement System or the Modular Rotating Hinge Knee System. These devices are intended to be used in clinical situations where there is radical bone loss of the femur and/or proximal tibia. This radical bone loss can be related to oncology (tumor), trauma or failed previous prosthesis.
This document is a 510(k) premarket notification from Howmedica Osteonics Corp., dba Stryker Orthopaedics, for their Global Modular Replacement System, Modular Replacement System, and Modular Rotating Hinge Knee. The submission, K222056, aims to modify the labeling of these devices to include MR Conditional language and to add an additional contraindication to the GMRS Press Fit Stems with Hydroxyapatite.
Here's an analysis of the provided text regarding acceptance criteria and supporting studies:
-
Table of Acceptance Criteria and Reported Device Performance:
The document does not explicitly present a table of acceptance criteria with corresponding device performance metrics for clinical outcomes. Instead, it focuses on non-clinical testing for MR compatibility and states that these tests were performed according to recognized ASTM standards. The "acceptance criteria" can be inferred as meeting the requirements of these standards.
Acceptance Criteria (Inferred) Reported Device Performance Compliance with ASTM F2052-15 for Magnetically Induced Displacement Force Testing performed as outlined in FDA guidance. Compliance with ASTM F2213-17 for Magnetically Induced Torque Testing performed as outlined in FDA guidance. Compliance with ASTM F2119-07 (2013) for Image Artifacts Testing performed as outlined in FDA guidance. Compliance with ASTM F2182-19e1 for Heating by RF Fields (including additional analyses for tissue heating) Testing performed as outlined in FDA guidance. -
Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
The document describes non-clinical testing performed on the devices themselves, not on human subjects. Therefore, the concept of a "test set" in the context of clinical data (e.g., patient cases) does not apply here. The "samples" would be the physical devices or components tested. The document does not specify the number of devices tested for each non-clinical evaluation (e.g., how many different GMRS components were subjected to MR compatibility testing). The data provenance is non-clinical laboratory testing, not human-derived data, and thus country of origin or retrospective/prospective status is not applicable in this context.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
This information is not applicable. Ground truth, in the sense of clinical diagnoses or outcomes, is not established for this type of non-clinical device testing. The "ground truth" for MR compatibility is determined by the physical properties of the materials and design, evaluated against established engineering standards like ASTM.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This information is not applicable. Adjudication methods are used in clinical studies or expert reviews to resolve discrepancies in diagnoses or assessments from multiple reviewers. For non-clinical engineering tests, the outcome is determined by measurements against predefined tolerances within the ASTM standards.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No MRMC comparative effectiveness study was done. The submission is for a medical device (orthopedic implants), not an AI-powered diagnostic or assistive tool.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. The device is a physical orthopedic implant, not an algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The "ground truth" for the non-clinical testing is based on established engineering principles and the requirements defined in the ASTM standards (e.g., limits for magnetically induced displacement, torque, image artifact levels, and localized heating).
-
The sample size for the training set:
Not applicable. This is not an AI/machine learning device; therefore, there is no "training set."
-
How the ground truth for the training set was established:
Not applicable. There is no training set for this type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 26, 2023
Howmedica Osteonics Corp., dba Stryker Orthopaedics Margaret Klippel Chief Regulatory Specialist 325 Corporate Drive Mahwah, New Jersey 07430
Re: K222056
Trade/Device Name: Global Modular Replacement System, Modular Replacement System, Modular Rotating Hinge Knee Regulation Number: 21 CFR 888.3350 Regulation Name: Hip Joint Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JDI, KRO, LZO, LPH Dated: January 13, 2023 Received: January 17, 2023
Dear Margaret Klippel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Limin Sun -S
Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
K222056
Device Name MRS Stems and Intercalary Stems
Indications for Use (Describe)
Indications for the MRS Stems and Intercalary Stems (presented in K952970):
· This device is intended for use in patients requiring extensive reconstruction of the femur and/or proximal tibia, including the hip or knee joint, resulting from extensive bone loss. Tumor resection for skeletal lesions (Oncology patients where radical bone resection and replacement may be required), revision surgery for failed arthroplasty, and acute trauma are the primary causes of extensive bone loss. These prostheses are intended for use with bone cement.
· The Intercalary System is intended for use in situations arising from in femoral mid-shaft tumor resection, or for prosthetic knee fusion.
Type of Use (Select one or both, as applicable)x Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{3}------------------------------------------------
510(k) Number (if known)
Device Name Global Modular Replacement System
Indications for Use (Describe) Indications for the Global Modular Replacement System (presented in K023087):
· Femoral and/or proximal tibial replacement and total femoral replacement in Oncology cases where radical resection and replacement of bone is required, and in limb salvage procedures where radical resection and replacement of the bone is required. Limb salvage procedures would include surgical intervention for severe trauma, failed previous prosthesis, and/or Oncology indications.
| Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 201 Subpart D) Over-The-Counter Use (21 CFR 201 Subpart C) |
|---|---|
| ------------------------------------------------- | -------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
_x Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{4}------------------------------------------------
510(k) Number (if known)
Device Name GMRS Press Fit Stems with HA Coating
Indications for Use (Describe)
Indications for the Global Modular Replacement System Press Fit Stems with HA Coating (presented in K022403, and K031217):
Proximal femoral reconstruction secondary to:
- o Trauma
o Failed previous prosthesis
o Tumor resection
Type of Use (Select one or both, as applicable)x Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{5}------------------------------------------------
510(k) Number (if known)
Device Name
GMRS Anteverted Proximal Femoral Component
Indications for Use (Describe) Indications for the GMRS Anteverted Proximal Femoral Component (presented in K032581)
· Femoral replacement in Oncology cases where radical resection and replacement of bone is required, and in limb salvage procedures where radical resection and replacement of bone is required. Limb salvage procedures would include surgical intervention for severe trauma, failed previous prosthesis, and/or Oncology indications. This component may also be used with the Distal Femoral segment components of the GMRS in total femoral replacement.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | |
|---|---|
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{6}------------------------------------------------
510(k) Number (if known)
Device Name MRS System Cemented Stems
Indications for Use (Describe)
Indications for the Modular Replacement System Cemented Stems (cleared in K040749):
Femoral and/or proximal tibial replacement due to:
o Trauma
o Failed previous prosthesis
o Tumor resection
| Type of Use (Select one or both, as applicable) | Residential Use (Dwelling, RCF, Adult Foster Care, etc.) Other Use (Commercial, RCF, Adult Foster Care, etc.) | Residential Use (Dwelling, RCF, Adult Foster Care, etc.) | Other Use (Commercial, RCF, Adult Foster Care, etc.) |
|---|---|---|---|
| Residential Use (Dwelling, RCF, Adult Foster Care, etc.) | Other Use (Commercial, RCF, Adult Foster Care, etc.) |
| * Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{7}------------------------------------------------
510(k) Number (if known)
Device Name Modular Rotating Hinge Knee System
Indications for Use (Describe)
Indications for the Modular Rotating Hinge Knee System (cleared in K002552)
The Modular Rotating Hinge Knee System is intended for use with bone cement in cases where there is destruction of the joint surfaces, with or without significant bone deformity; the cruciate and/or collateral ligaments do not stabilize the knee joint; the ligaments are inadequate and/or the musculature is weak; and revision of a failed previous prosthesis where there is instability, with or without bone loss or inadequate soft tissue.
Expanded indications include for use with bone cement in situations where there is extensive bone loss, including limb salvage procedures for severe trauma, failed previous prosthesis, and/or Oncology indications.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
× Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{8}------------------------------------------------
510(k) Summary
| Sponsor | Howmedica Osteonics Corp. dba Stryker Orthopaedics325 Corporate DriveMahwah, NJ 07430 |
|---|---|
| Contact Person | Margaret KlippelChief Specialist, Regulatory AffairsHowmedica Osteonics Corp325 Corporate DriveMahwah, NJ 07430(201) 831-5559Email: Margaret.klippel@stryker.com |
| Date Prepared: | January 4, 2023 |
| Proprietary Name: | Global Modular Replacement System |
| Common Name: | Hip Replacement, Rotating Hinge Knee Replacement |
| Regulatory Class: | Class II |
| Classification Panel: | 87, Orthopedic |
| Regulation: | 888.3350, 888.3510, 888.3353, 888.3358 |
Legally Marketed Devices to which Substantial Equivalence is Claimed:
Predicate Howmedica Osteonics (aka Stryker Orthopaedic) Global Modular Replacement System devices were cleared per the following primary 510(k) submission:
| Submission Number | Device Name | Product Codes |
|---|---|---|
| K023087 | Global Modular Replacement System | JDI, KRO |
Legally Marketed Reference Devices Used to Support Substantial Equivalence:
| Submission Number | Device Name | Product Code(s) |
|---|---|---|
| K952970 | Modular Replacement System | JDI, KRO |
| K022403 | GMRS Press Fit Stems with HA | LZO |
| K031217 | 11mm GMRS Press Fit Stems with HA | LZO |
| K032581 | GMRS Anteverted Proximal FemoralComponent | LPH |
{9}------------------------------------------------
| Submission Number | Device Name | Product Code(s) |
|---|---|---|
| K040749 | Modular Replacement System CementedStems | KRO |
| K002552 | Modular Rotating Hinge Knee | KRO |
Reason for 510(k) Submission:
The purpose of this submission is to modify the labeling of the Global Modular Replacement System components to add MR Conditional language. An additional contraindication is being added to the GMRS Press Fit Stems with Hydroxyapatite.
Device Description:
The Global Modular Replacement System is comprised of a number of components that are intended to be used in conjunction with each other, or in conjunction with components of the Modular Replacement System or the Modular Rotating Hinge Knee System. These devices are intended to be used in clinical situations where there is radical bone loss of the femur and/or proximal tibia. This radical bone loss can be related to oncology (tumor), trauma or failed previous prosthesis.
Indications for Use
The Stryker Global Modular Replacement System components are sterile, single-use devices intended for use in situations where there is a need for replacement of bone due to radical bone loss. This loss can be related to oncology, trauma or failed previous prosthesis.
Specific Indications for Use are listed below.
Indications for the MRS Stems and Intercalary Stems presented in K952970:
- This device is intended for use in patients requiring extensive reconstruction of the femur and/or proximal tibia, including the hip or knee joint, resulting from extensive bone loss. Tumor resection for skeletal lesions (Oncology patients where radical bone resection and replacement may be required), revision surgery for failed arthroplasty, and acute trauma are the primary causes of extensive bone loss. These prostheses are intended for use with bone cement.
- . The Intercalary System is intended for use in situations arising from femoral mid-shaft turnor resection, or for prosthetic knee fusion.
Indications for the Global Modular Replacement System presented in K023087:
- . Femoral and/or proximal tibial replacement and total femoral replacement in Oncology
{10}------------------------------------------------
cases where radical resection and replacement of bone is required, and in limb salvage procedures where radical resection and replacement of the bone is required. Limb salvage procedures would include surgical intervention for severe trauma, failed previous prosthesis, and/or Oncology indications.
Indications for the Global Modular Replacement System Press Fit Stems with HA Coating (presented in K022403, and K031217):
- Proximal femoral reconstruction secondary to: 0
- Trauma o
- Failed previous prosthesis о
- Tumor resection O
Indications for the GMRS Anteverted Proximal Femoral Component (presented in K032581)
-
Femoral replacement in Oncology cases where radical resection and replacement of bone is . required, and in limb salvage procedures where radical resection and replacement of bone is required. Limb salvage procedures would include surgical intervention for severe trauma, failed previous prosthesis, and/or Oncology indications. This component may also be used with the Distal Femoral segment components of the GMRS in total femoral replacement.
Indications for the Modular Replacement System Cemented Stems (cleared in K040749): -
Femoral and/or proximal tibial replacement due to: ●
- Trauma O
- Failed previous prosthesis O
- Tumor resection o
Indications for the Modular Rotating Hinge Knee System (cleared in K002552)
The Modular Rotating Hinge Knee System is intended for use with bone cement in cases where there is destruction of the joint surfaces, with or without significant bone deformity; the cruciate and/or collateral ligaments do not stabilize the knee joint; the ligaments are inadequate and/or the musculature is weak; and revision of a failed previous prosthesis where there is instability, with or without bone loss or inadequate soft tissue.
{11}------------------------------------------------
Expanded indications include for use with bone cement in situations where there is extensive bone loss, including limb salvage procedures for severe trauma, failed previous prosthesis, and/or Oncology indications.
Summary of Technological Characteristics:
There have been no changes to the technological characteristics of the subject Global Modular Replacement System devices as a result of the labeling to add MR Conditional language. The subject GMRS components have the same design and are manufactured from the same materials as the corresponding proximal femoral bodies and distal stems of the predicate devices.
Non-Clinical Testing:
The Global Modular Replacement System has been evaluated for use in a Magnetic Resonance Environment through non-clinical testing as outlined in the FDA guidance document "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment -Guidance for Industry and FDA Staff'', dated May 20, 2021. This testing was conducted to characterize the compatibility of Stryker Orthopaedics GMRS passive implants in the MR environment. FDA guidance "Assessment of Radiofrequency-Induced Heating in the Magnetic Resonance (MR) Environment for Multi-Configuration Passive Medical Devices -Guidance for Industry and FDA Staff", dated March 22, 2016, was also consulted for the heating evaluations performed. Testing was performed according to the standards listed below:
- Magnetically Induced Displacement Force Test Performed per ASTM F2052-15, Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment
- Magnetically Induced Torque Test Performed per ASTM F2213-17, Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment
- Image Artifact Test Performed per ASTM F2119-07 (2013), Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants
- Heating by RF Fields Test Performed per ASTM F2182-19e1, Standard Test Method for Measurement of Radio Frequency Induced Heating on or Near Passive Implants During Magnetic Resonance Imaging
Additional analyses have been performed to address heating in the tissue of interest as indicated in the May 2021 guidance document.
{12}------------------------------------------------
The labeling of the GMRS system has been modified to include the MR conditional symbol and to provide the parameters under which a patient who has the device can be safely scanned. The labeling of the GMRS Press Fit Stems with Hydroxyapatite has also been revised to add an additional contraindication.
Clinical Testing:
Clinical testing was not required as a basis for substantial equivalence.
Conclusion:
Based upon a comparison of the intended use, materials, summary of technological characteristics, and preclinical evaluation, the subject Global Modular Replacement System components are considered substantially equivalent to the predicate devices identified in this premarket notification.
N/A