Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    Why did this record match?
    Reference Devices :

    K182075, K171292

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Passport 12m and 17m Patient Monitors:

    The Passport 17m and Passport 12m patient monitors are intended for monitoring, displaying, reviewing, and transferring of multiple physiological parameters including ECG (3-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, OT analysis, and heart rate (HR)), respiration rate (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure(IBP), pulmonary artery wedge pressure (PAWP), cardiac output (C.O.), continuous cardiac output (CCO), mixed/central venous oxygen saturation (SvO2/ScvO2), carbon dioxide (CO2), Oxygen (O2), anesthetic gas (AC), impedance cardiograph (CCG), bispectral index (BIS), respiration mechanics (RM), and neuromuscular transmission monitoring (NMT). The equipment also provides an interpretation of resting 12-lead ECG.

    All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:
    • The arrhythmia detection, ST Segment analysis of Mortara algorithm, BIS, RM, CCO, SvO2/ScvO2, PAWP monitoring and NMT monitoring are intended for adult and pediatric patients only;

    · ST Segment analysis of Mindray algorithm is intended for adult patients only;

    · C.O. monitoring is restricted to adult patients only;

    • ICG monitoring is only for use on adult patients who meet the following requirements: height: 122 to 229cm, weight: 30 to 155kg.

    The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use.

    T1 Patient Monitor:

    The T1 Patient Monitor is intended for monitoring, displaying, reviewing, and transferring of multiple physiological parameters including ECG (3-lead, or 12-lead selectable, arrhythmia detection, ST Segment analysis, QT analysis, and heart rate (HR)), respiration (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), pulmonary artery wedge pressure (PAWP), and carbon dioxide (CO2). The equipment also provides an interpretation of resting 12-lead ECG.

    All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:
    • The arthythmia detection, ST Segment analysis of Mortara algorithm, and PAWP monitoring are intended for adult and pediatric patients;

    · ST Segment analysis of Mindray ECG algorithm is intended for adult patients only.

    The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use.

    Device Description

    The subject Passport Series Patient Monitors includes three monitors:

    • o Passport 12m Patient Monitor
    • Passport 17m Patient Monitor ●
    • o T1 Patient Monitor

    All of the devices in the family are multiparameter monitors indicated for monitoring, displaying, reviewing, alarming, and transferring multiple physiological parameters. The Passport 12m and 17m Patient monitors provide patient monitoring capabilities based on the user selected modules that are plugged into the main monitor.

    The T1 patient monitor is one of the available modules that can be plugged into the Passport 17m or 12m monitor. The T1 can also be used as a standalone monitor and when used as a standalone monitor, it provides a subset of the functions that are provided by Passport 17m and 12m.

    AI/ML Overview

    This document is a 510(k) summary for the Mindray Passport Series Patient Monitors (Passport 12m, Passport 17m, and T1). It focuses on establishing substantial equivalence to previously cleared predicate devices and does not contain a full study report with detailed acceptance criteria and performance data for a new AI/ML device.

    Therefore, the following information, relevant to AI/ML device studies, cannot be extracted from the provided text:

    • A table of acceptance criteria and the reported device performance: The document states that "functional and system level testing" was conducted and that the devices "continue to meet specifications and the performance of the device is equivalent to the predicate." However, it does not provide specific acceptance criteria values or detailed performance metrics.
    • Sample sized used for the test set and the data provenance: Not mentioned.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not mentioned.
    • Adjudication method for the test set: Not mentioned.
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not relevant as this is not a new AI/ML device being submitted for comparative effectiveness.
    • If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not relevant as this is not a new AI/ML device.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not mentioned.
    • The sample size for the training set: Not mentioned.
    • How the ground truth for the training set was established: Not mentioned.

    Summary of available relevant information from the document:

    This 510(k) submission is for modifications to existing patient monitors, not for a new AI/ML device. The changes include:

    • Wireless Module: Addition of a Laird (2.4GHz/5GHz) wireless module.
    • MPM Module: Addition of an MPM 3.0 module with similar functionality to the MPM 2.0 module.
    • Masimo SpO2 accessories: Support for new Masimo SpO2 accessories.
    • ECG accessories: Support for new ECG accessories (12-lead, defibrillation-proof).
    • CO2 accessories: Support for new DRYLINE PRIME and DRYLINE PRIME+ gas sampling lines and water traps.
    • Signal Cord Resistance: Improvements to SpO2/CO2/BIS signal cord resistance to damage.
    • NIBP Issue Correction: Correction of an issue where NIBP misreported "NIBP Excessive Motion" after 49 days.
    • Anesthetic Agent (AA) Numerics: Support for End tidal (Et) and Fraction of Inspired (Fi) numerics for anesthetic agent (AA).
    • Physiological Alarms: Support for specific physiological alarm behaviors (triggered only when parameter value is higher/lower than limits, re-alarming after reset, non-overlapping relationships of alarm limit setting).
    • Alarm Setup Menu: Support for password protection for accessing the Alarm Setup menu.
    • User Authorization: Support for user authorization via LDAP.
    • Patient Monitor Information: Support for Facility information, Room No. information, and Visit Number in patient demographics.
    • Printing: Support for printing arrhythmia settings in reports, changing factory default printer paper size, and QT report printing.
    • MAC Address Display: Support to display the MAC address of the patient monitor.
    • ST Values Display: Support for displaying ST values in the parameter tile.
    • Patient Private Information: Support for encryption of patient private information sent to the network.
    • ST Point Adjustment: Support for adjustment of the ST point on multiple displayed leads.
    • NIBP Measurement on Clock: Support for NIBP measurement on the clock.
    • Network Application Transmission Priority Setting: Support for network application transmission priority setting.
    • ECG Beat Annotation: Support for ECG beat annotation.
    • DNS for ADT and LDAP server address: Support for DNS for ADT and LDAP server address.
    • Bed No. Issue Correction: Correction of an issue where the Bed No. of the Passport 12m/17m is changed after T1 is pulled out of the monitors.

    Performance Data:
    Mindray conducted functional and system level testing to evaluate the performance of these modifications. The testing demonstrated that the devices "continue to meet specifications and the performance of the device is equivalent to the predicate." The devices also meet relevant consensus standards (listed in the document).

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BeneVision N12/N15/N17/N19/N22 patient monitors are intended for monitoring, displaying, reviewing, storing, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead, 6-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, QT Analysis, and Heart Rate (HR)), Respiration Rate (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure(IBP), Pulmonary Artery Wedge Pressure (PAWP), Cardiac Output (C.O.), Continuous Cardiac Output (CCO), Mixed/Central Venous Oxygen Saturation (SvO2/ScvO2), Carbon Dioxide (CO2), Oxygen (O2), Anesthetic Gas (AG), Impedance Cardiograph (ICG), Respiration Mechanics (RM), Neuromuscular Transmission Monitoring (NMT), Electroencephalograph (EEG), and Regional Oxygen Saturation (rSO2). The system also provides an interpretation of resting 12-lead ECG.

    All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following:

    • The arrhythmia detection, RM, CCO, SvO2/ScvO2, PAWP, and NMT monitoring are intended for adult and pediatric patients only;
    • C.O. monitoring is intended for adult patients only;
    • ICG monitoring is intended for only adult patients who meet the following requirements: height: 122 to 229cm, weight: 30 to 155kg.
    • rSO2 monitoring is intended for use in individuals greater than 2.5kg.

    The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They should only be used by persons who have received adequate training in their use. The BeneVision N12/N15/N17/N19/N22 monitors are not intended for helicopter transport, hospital ambulance, or home use.

    The BeneVision N1 Patient Monitor is intended for monitoring, displaying, storing, alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead, 6-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, QT Analysis, and Heart Rate (HR)), Respiration (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP) , Pulmonary Artery Wedge Pressure (PAWP), Carbon Dioxide (CO2) and Oxygen (O2). The system also provides an interpretation of resting 12-lead ECG.

    All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following:

    • The arrhythmia detection and PAWP is intended for adult and pediatric patients only
      The BeneVision N1 monitor is to be used in healthcare facilities. It can also be used during patient transport inside and outside of the hospital environment. It should be used by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for home use.
    Device Description

    The subject BeneVision N Series Patient Monitors includes six monitors:

    • BeneVision N12 Patient Monitor
    • BeneVision N15 Patient Monitor
    • BeneVision N17 Patient Monitor
    • BeneVision N19 Patient Monitor
    • BeneVision N22 Patient Monitor
    • BeneVision N1 Patient Monitor

    The BeneVision N Series Patient Monitors are Mindray's new generation monitoring product family with ergonomic and flexible design in platform of both software and hardware to meet the clinical needs of monitoring.

    AI/ML Overview

    The provided document is a 510(k) Summary for the Mindray BeneVision N Series Patient Monitors. It focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets specific acceptance criteria through a dedicated study with statistical endpoints.

    Therefore, many of the requested elements for a detailed study description (e.g., sample size for test/training sets, data provenance, number/qualifications of experts, adjudication methods, MRMC studies, standalone performance with specific metrics, and ground truth establishment for training data) are not present in the provided text.

    The document primarily highlights changes from predicate devices and states that functional and system-level testing, along with compliance with consensus standards, demonstrate equivalence.

    Here's a summary of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with corresponding device performance metrics in the format typically seen for a new device's efficacy study. Instead, it compares the specifications of the subject device with those of predicate devices, implicitly indicating that the subject device's performance meets or exceeds the predicate's established performance or relevant cleared standards.

    Feature Area / ParameterAcceptance Criteria (Implicitly, equivalent to/better than Predicate or standard)Reported Device Performance (Subject BeneVision Devices)
    Display-N22: 22" 1680*1050 pixels
    N19: 19" 1680*1050 pixels
    N17: 18.5" 1920*1080 pixels
    N15: 15.6" 1920*1080 pixels
    N12: 12.1" 1280*800 pixels
    N1: 5.5" 720*1280 pixels
    Wireless2.4GHz/5GHz dual band module (Passport 12m)2.4GHz/5GHz dual band module (All BeneVision N Series)
    Data StorageCompact Flash (Passport 17m)
    SD card (T1)Solid State Hard Drive (SSD) (N22, N19)
    Embedded Multi Media Card (eMMC) (N17, N15, N12, N1)
    Alarm SystemYellow/red alarm lamp (Predicate)Cyan, yellow, or red alarm lamp; Supports Alarm Volume Escalation (Feature cleared in K161531)
    ECG - 6-lead ECGNot supported by predicateSupported (Feature cleared in K162607)
    ECG - Intelligent Arrhythmia AlarmNot supported by predicateSupported (Feature cleared in K161531)
    ECG - ST Segment Analysis (Pediatric/Neonate)Only adult (Predicate)Pediatric and neonate supported (Feature cleared in K131414)
    SpO2 - Masimo SpO2 module in MPM 3.0Not supported by predicateSupported (Feature cleared in K053269)
    CO2 - Sidestream CO2 2.0 ModuleNot supported by predicate (Type 1.0 supported)Supported (Feature cleared in K170712)
    CO2 measurement range: 0-150mmHg (wider than predicate)
    AwRR measurement range: 0-150rpm (wider than predicate)
    AwRR accuracy improved
    NMT ModuleNot applicable (Predicate)Supported (Feature cleared in K170876)
    EEG ModuleNot applicable (Predicate)Supported (Feature cleared in K161531)
    rSO2 ModuleNot applicable (Predicate)Supported (Feature cleared in K082327)
    Gas Recycling (AG module)Not supported by predicateSupported (Feature cleared in K171292)
    Early Warning Score (EWS)Not applicable (Predicate)Supported (Feature cleared in K170712)
    Helicopter/ambulance transport (N1)Not applicable (Predicate)Supported for ECG, RESP, Temp, SpO2, PR, NIBP, IBP (Feature cleared in K161531)
    NIBP Measurement RangeAdult: 40-270 (Systolic), 10-210 (Diastolic), 20-230 (Mean)
    Pediatric: 40-200 (Systolic), 10-150 (Diastolic), 20-165 (Mean)
    Neonate: 40-135 (Systolic), 10-100 (Diastolic), 20-110 (Mean)Adult: 25-290 (Systolic), 10-250 (Diastolic), 15-260 (Mean)
    Pediatric: 25-240 (Systolic), 10-200 (Diastolic), 15-215 (Mean)
    Neonate: 25-140 (Systolic), 10-115 (Diastolic), 15-125 (Mean)
    NIBP AccuracyMax mean error: ±5 mmHg; Max standard deviation: 8 mmHg (Predicate)Max mean error: ±5 mmHg; Max standard deviation: 8 mmHg (Same as Predicate)
    IBP Measurement Range-50 to 300 mmHg (Predicate)-50 to 300 mmHg (Same as Predicate)
    IBP Accuracy±2% or ±1 mmHg, whichever is greater (without sensor) (Predicate)±2% or ±1 mmHg, whichever is greater (without sensor) (Same as Predicate)
    Cardiac Output Measurement Range0.1 to 20 L/min (C.O.); 23 to 43 °C (TB); 0 to 27 °C (TI) (Predicate)Same as Predicate
    Cardiac Output Accuracy±5% or ±0.1 L/min (C.O.); ±0.1 °C (TB, TI) (Predicate)Same as Predicate

    2. Sample size used for the test set and the data provenance

    The document does not specify sample sizes for test sets. The testing mentioned is referred to as "functional and system level testing" and "bench testing." It also states Mindray conducted "clinical testing to demonstrate that the Mindray and Nellcor SpO2 modules meet relevant consensus standards."
    There is no mention of data provenance (e.g., country of origin of data, retrospective or prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The document does not describe the use of experts to establish ground truth for testing. The evaluation focused on meeting specifications and consensus standards, and demonstrating equivalence to predicate devices.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. There is no mention of adjudication methods.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a patient monitor, not an AI-assisted diagnostic device, and no MRMC studies are mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The document describes performance in terms of functionality and adherence to technical specifications and consensus standards, not in terms of "algorithm-only" performance as would be relevant for an AI device. The tests performed are for the integrated device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The term "ground truth" is not used. The performance evaluation is based on meeting engineering specifications, comparing against predicate device performance, and compliance with recognized consensus standards (e.g., IEC, ISO, AAMI standards for physiological measurement accuracy).

    8. The sample size for the training set

    Not applicable. A "training set" is relevant for machine learning algorithms. This document describes a patient monitor, and no machine learning model training is discussed.

    9. How the ground truth for the training set was established

    Not applicable, as no training set for a machine learning model is mentioned.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1