K Number
K190011
Date Cleared
2019-07-10

(188 days)

Product Code
Regulation Number
870.1025
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Passport 12m and 17m Patient Monitors:

The Passport 17m and Passport 12m patient monitors are intended for monitoring, displaying, reviewing, and transferring of multiple physiological parameters including ECG (3-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, OT analysis, and heart rate (HR)), respiration rate (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure(IBP), pulmonary artery wedge pressure (PAWP), cardiac output (C.O.), continuous cardiac output (CCO), mixed/central venous oxygen saturation (SvO2/ScvO2), carbon dioxide (CO2), Oxygen (O2), anesthetic gas (AC), impedance cardiograph (CCG), bispectral index (BIS), respiration mechanics (RM), and neuromuscular transmission monitoring (NMT). The equipment also provides an interpretation of resting 12-lead ECG.

All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:
• The arrhythmia detection, ST Segment analysis of Mortara algorithm, BIS, RM, CCO, SvO2/ScvO2, PAWP monitoring and NMT monitoring are intended for adult and pediatric patients only;

· ST Segment analysis of Mindray algorithm is intended for adult patients only;

· C.O. monitoring is restricted to adult patients only;

• ICG monitoring is only for use on adult patients who meet the following requirements: height: 122 to 229cm, weight: 30 to 155kg.

The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use.

T1 Patient Monitor:

The T1 Patient Monitor is intended for monitoring, displaying, reviewing, and transferring of multiple physiological parameters including ECG (3-lead, or 12-lead selectable, arrhythmia detection, ST Segment analysis, QT analysis, and heart rate (HR)), respiration (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), pulmonary artery wedge pressure (PAWP), and carbon dioxide (CO2). The equipment also provides an interpretation of resting 12-lead ECG.

All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:
• The arthythmia detection, ST Segment analysis of Mortara algorithm, and PAWP monitoring are intended for adult and pediatric patients;

· ST Segment analysis of Mindray ECG algorithm is intended for adult patients only.

The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use.

Device Description

The subject Passport Series Patient Monitors includes three monitors:

  • o Passport 12m Patient Monitor
  • Passport 17m Patient Monitor ●
  • o T1 Patient Monitor

All of the devices in the family are multiparameter monitors indicated for monitoring, displaying, reviewing, alarming, and transferring multiple physiological parameters. The Passport 12m and 17m Patient monitors provide patient monitoring capabilities based on the user selected modules that are plugged into the main monitor.

The T1 patient monitor is one of the available modules that can be plugged into the Passport 17m or 12m monitor. The T1 can also be used as a standalone monitor and when used as a standalone monitor, it provides a subset of the functions that are provided by Passport 17m and 12m.

AI/ML Overview

This document is a 510(k) summary for the Mindray Passport Series Patient Monitors (Passport 12m, Passport 17m, and T1). It focuses on establishing substantial equivalence to previously cleared predicate devices and does not contain a full study report with detailed acceptance criteria and performance data for a new AI/ML device.

Therefore, the following information, relevant to AI/ML device studies, cannot be extracted from the provided text:

  • A table of acceptance criteria and the reported device performance: The document states that "functional and system level testing" was conducted and that the devices "continue to meet specifications and the performance of the device is equivalent to the predicate." However, it does not provide specific acceptance criteria values or detailed performance metrics.
  • Sample sized used for the test set and the data provenance: Not mentioned.
  • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not mentioned.
  • Adjudication method for the test set: Not mentioned.
  • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not relevant as this is not a new AI/ML device being submitted for comparative effectiveness.
  • If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not relevant as this is not a new AI/ML device.
  • The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not mentioned.
  • The sample size for the training set: Not mentioned.
  • How the ground truth for the training set was established: Not mentioned.

Summary of available relevant information from the document:

This 510(k) submission is for modifications to existing patient monitors, not for a new AI/ML device. The changes include:

  • Wireless Module: Addition of a Laird (2.4GHz/5GHz) wireless module.
  • MPM Module: Addition of an MPM 3.0 module with similar functionality to the MPM 2.0 module.
  • Masimo SpO2 accessories: Support for new Masimo SpO2 accessories.
  • ECG accessories: Support for new ECG accessories (12-lead, defibrillation-proof).
  • CO2 accessories: Support for new DRYLINE PRIME and DRYLINE PRIME+ gas sampling lines and water traps.
  • Signal Cord Resistance: Improvements to SpO2/CO2/BIS signal cord resistance to damage.
  • NIBP Issue Correction: Correction of an issue where NIBP misreported "NIBP Excessive Motion" after 49 days.
  • Anesthetic Agent (AA) Numerics: Support for End tidal (Et) and Fraction of Inspired (Fi) numerics for anesthetic agent (AA).
  • Physiological Alarms: Support for specific physiological alarm behaviors (triggered only when parameter value is higher/lower than limits, re-alarming after reset, non-overlapping relationships of alarm limit setting).
  • Alarm Setup Menu: Support for password protection for accessing the Alarm Setup menu.
  • User Authorization: Support for user authorization via LDAP.
  • Patient Monitor Information: Support for Facility information, Room No. information, and Visit Number in patient demographics.
  • Printing: Support for printing arrhythmia settings in reports, changing factory default printer paper size, and QT report printing.
  • MAC Address Display: Support to display the MAC address of the patient monitor.
  • ST Values Display: Support for displaying ST values in the parameter tile.
  • Patient Private Information: Support for encryption of patient private information sent to the network.
  • ST Point Adjustment: Support for adjustment of the ST point on multiple displayed leads.
  • NIBP Measurement on Clock: Support for NIBP measurement on the clock.
  • Network Application Transmission Priority Setting: Support for network application transmission priority setting.
  • ECG Beat Annotation: Support for ECG beat annotation.
  • DNS for ADT and LDAP server address: Support for DNS for ADT and LDAP server address.
  • Bed No. Issue Correction: Correction of an issue where the Bed No. of the Passport 12m/17m is changed after T1 is pulled out of the monitors.

Performance Data:
Mindray conducted functional and system level testing to evaluate the performance of these modifications. The testing demonstrated that the devices "continue to meet specifications and the performance of the device is equivalent to the predicate." The devices also meet relevant consensus standards (listed in the document).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 10, 2019

Shenzhen Mindray Bio-Medical Electronics Co., Ltd. Yanhong Bai Manager Regulatory Affairs Mindray Building, Keji 12th Road South Hi-tech Industrial Park, Nanshan Shenzhen, 518057, P.R. China

Re: K190011

Trade/Device Name: Passport Series Patient Monitors (including Passport 17m, Passport 12m and T1) Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: MHX, DQA, DRT, DSI, DSK, DXN, FLL, MLD, DPZ, CCK, CBQ, CBS, CBR, CCL, DSB, DXG, OLW, DSJ, KOI, GXY Dated: June 6, 2019 Received: June 10, 2019

Dear Yanhong Bai:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for Nicole Goodsell Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190011

Device Name

Passport Series Patient Monitors (including Passport 12m, Passport 17m and T1)

Indications for Use (Describe)

Passport 12m and 17m Patient Monitors:

The Passport 17m and Passport 12m patient monitors are intended for monitoring, displaying, reviewing, and transferring of multiple physiological parameters including ECG (3-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, OT analysis, and heart rate (HR)), respiration rate (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure(IBP), pulmonary artery wedge pressure (PAWP), cardiac output (C.O.), continuous cardiac output (CCO), mixed/central venous oxygen saturation (SvO2/ScvO2), carbon dioxide (CO2), Oxygen (O2), anesthetic gas (AC), impedance cardiograph (CCG), bispectral index (BIS), respiration mechanics (RM), and neuromuscular transmission monitoring (NMT). The equipment also provides an interpretation of resting 12-lead ECG.

All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following: • The arrhythmia detection, ST Segment analysis of Mortara algorithm, BIS, RM, CCO, SvO2/ScvO2, PAWP monitoring and NMT monitoring are intended for adult and pediatric patients only;

· ST Segment analysis of Mindray algorithm is intended for adult patients only;

· C.O. monitoring is restricted to adult patients only;

• ICG monitoring is only for use on adult patients who meet the following requirements: height: 122 to 229cm, weight: 30 to 155kg.

The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use.

T1 Patient Monitor:

The T1 Patient Monitor is intended for monitoring, displaying, reviewing, and transferring of multiple physiological parameters including ECG (3-lead, or 12-lead selectable, arrhythmia detection, ST Segment analysis, QT analysis, and heart rate (HR)), respiration (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), pulmonary artery wedge pressure (PAWP), and carbon dioxide (CO2). The equipment also provides an interpretation of resting 12-lead ECG.

All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following: • The arthythmia detection, ST Segment analysis of Mortara algorithm, and PAWP monitoring are intended for adult and pediatric patients;

· ST Segment analysis of Mindray ECG algorithm is intended for adult patients only.

The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(K) SUMMARY

In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the Passport Series Patient Monitors is provided below.

Device Common Name:Patient Monitor
Device Proprietary Name:Passport Series Patient Monitors (Passport 12m,Passport 17m and T1)
SubmitterShenzhen Mindray Bio-medical Electronics Co., LTD.Mindray Building, Keji 12th Road South,Hi-tech Industrial Park, Nanshan,Shenzhen, 518057, P. R. ChinaTel: +86 755 8188 8998Fax: +86 755 2658 2680
ContactMs. Yanhong BaiManager Regulatory AffairsShenzhen Mindray Bio-medical Electronics Co., LTD.Mindray Building, Keji 12th Road South,High-tech Industrial Park, Nanshan, Shenzhen518057, P.R. ChinaTel: +86 755 81885635Fax: +86 755 26582680E-mail: baiyanhong@mindray.com
Date PreparedDecember 31, 2018
Classification Regulation21 CFR 870.1025, Class II, Arrhythmia detector and alarm(including ST-segment measurement and alarm)
PanelCardiovascular

{5}------------------------------------------------

ProductCodeRegulationNumberPanelRegulation descriptionDevice Common Name
Primary
MHX21 CFR870.1025CardiovascularArrhythmia detector andalarm (includingST-segment measurementand alarm)monitor, physiological,patient(with arrhythmiadetection or alarms)
Secondary
ProductCodeRegulationNumberPanelRegulation descriptionDevice Common Name
DSI21 CFR870.1025CardiovascularArrhythmia detector andalarm (includingST-segment measurementand alarm)detector and alarm,arrhythmia
MLD21 CFR870.1025CardiovascularArrhythmia detector andalarm (includingST-segment measurementand alarm)monitor, st segment withalarm
DRT21 CFR870.2300CardiovascularCardiac Monitor(includingcardiotachometer and ratealarm)monitor, cardiac (incl.cardiotachometer & ratealarm)
DXN21 CFR870.1130CardiovascularNoninvasive bloodpressure measurementsystemsystem, measurement,blood-pressure,non-invasive
DSK21 CFR870.1110CardiovascularBlood pressure computercomputer, blood-pressure
FLL21 CFR880.2910CardiovascularClinical electronicthermometerthermometer, electronic,clinical
DQA21 CFR870.2700CardiovascularOximeteroximeter
DPZ21 CFR870.2710CardiovascularEar oximeteroximeter, ear
CCK21 CFR868.1400AnesthesiologyCarbon dioxide gasanalyzeranalyzer, gas,carbon-dioxide,gaseous-phase
CBQ21 CFR868.1400AnesthesiologyEnflurane gas analyzeranalyzer, gas, enflurane,
CBS868.1500868.162021 CFRAnesthesiologyHalothane gas analyzergaseous-phase (anesthetic concentration)analyzer, gas, halothane, gaseous-phase (anesthetic conc.)
CBR868.170021 CFRAnesthesiologyNitrous oxide gas analyzeranalyzer, gas, nitrous-oxide, gaseous phase (anesthetic conc.)
CCL868.172021 CFRAnesthesiologyOxygen gas analyzeranalyzer, gas, oxygen, gaseous-phase
DSB870.277021 CFRCardiovascularImpedance plethysmographplethysmograph, impedance
DXG870.143521 CFRCardiovascularSingle-function, preprogrammed diagnostic computercomputer, diagnostic, pre-programmed, single-function
OLW882.140021 CFRNeurologyElectroencephalographindex-generating electroencephalograph software
DSJ870.110021 CFRCardiovascularBlood pressure alarmalarm, blood-pressure
KOI868.277521 CFRAnesthesiologyElectrical peripheral nerve stimulatorstimulator, nerve, peripheral, electric
GXY870.132021 CFRNeurologyCutaneous electrode.electrode, cutaneous

Classification Regulation, Classification Name and Product Codes:

{6}------------------------------------------------

Primary Predicate Device:

  • K 152902 Passport Series Patient Monitors (Passport 12m, Passport 17m, and T1) Shenzhen Mindray Bio-Medical Electronics Co., Ltd
  • K170876 Passport Series Patient Monitors (Passport 12m and Passport 17m) Shenzhen Mindray Bio-Medical Electronics Co., Ltd

Secondary Predicate Device:

  • K182075 -BeneVision N Series Patient Monitors Shenzhen Mindray Bio-Medical Electronics Co., Ltd K171292 - A7 Anesthesia System Shenzhen Mindray Bio-Medical Electronics Co., Ltd

Indications for Use:

{7}------------------------------------------------

Passport 12m and 17m Patient Monitors:

The Passport 17m and Passport 12m patient monitors are intended for monitoring, displaying, reviewing, alarming, and transferring of multiple physiological parameters including ECG (3-lead , 5-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, QT analysis, and heart rate (HR)), respiration rate (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure(IBP), pulmonary artery wedge pressure (PAWP), cardiac output (C.O.), continuous cardiac output (CCO), mixed/central venous oxygen saturation (Sv02/Scv02), carbon dioxide (CO2), Oxygen (O2), anesthetic gas (AG), impedance cardiograph (ICG), bispectral index (BIS), respiration mechanics (RM), and neuromuscular transmission monitoring (NMT). The equipment also provides an interpretation of resting 12-lead ECG.

All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:

  • The arrhythmia detection, ST Segment analysis of Mortara algorithm, BIS, RM, ● CCO, SvO2/ScvO2, PAWP monitoring and NMT monitoring are intended for adult and pediatric patients only;
  • ST Segment analysis of Mindray algorithm is intended for adult patients only; ●
  • C.O. monitoring is restricted to adult patients only; ●
  • ICG monitoring is only for use on adult patients who meet the following requirements: height: 122 to 229cm, weight: 30 to 155kg.

The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use.

T1 Patient Monitor:

The T1 Patient Monitor is intended for monitoring, displaying, reviewing, storing , alarming, and transferring of multiple physiological parameters including ECG (3-lead, or 5-lead, or 12-lead selectable, arrhythmia detection, ST Segment analysis, QT analysis, and heart rate (HR)), respiration (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), pulmonary artery wedge pressure (PAWP), and carbon dioxide (CO2). The equipment also provides an interpretation of resting 12-lead ECG.

All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:

  • The arrhythmia detection, ST Segment analysis of Mortara algorithm, and ● PAWP monitoring are intended for adult and pediatric patients;
  • ST Segment analysis of Mindray ECG algorithm is intended for adult patients ● only.

{8}------------------------------------------------

The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use.

Device Description:

The subject Passport Series Patient Monitors includes three monitors:

  • o Passport 12m Patient Monitor
  • Passport 17m Patient Monitor ●
  • o T1 Patient Monitor

All of the devices in the family are multiparameter monitors indicated for monitoring, displaying, reviewing, alarming, and transferring multiple physiological parameters. The Passport 12m and 17m Patient monitors provide patient monitoring capabilities based on the user selected modules that are plugged into the main monitor.

The T1 patient monitor is one of the available modules that can be plugged into the Passport 17m or 12m monitor. The T1 can also be used as a standalone monitor and when used as a standalone monitor, it provides a subset of the functions that are provided by Passport 17m and 12m.

Substantial Equivalence:

Comparison of Indications

The indications for use of the subject devices have not changed in this 510(K). Indications of the predicate device (Passport 12m and 17m Patient Monitors K170876,T1 Patient Monitor K152902) and the subject device (Passport 12m,Passport 17m and T1)) are the same.

Comparison of Technological Characteristics

The table below compares the key technological feature of the subject devices to the primary predicate device (Passport 12m and 17m Patient Monitors K170876, T1 Patient Monitor K152902). The features in grey are the features that have been modified since their previous clearances and that are the subject of this 510(k).

Device Comparison Table (Compare with Passport 12m and 17m Patient Monitors K170876,T1 Patient Monitor K152902)

{9}------------------------------------------------

FeaturePassport 17mPassport 12mT1Subject Devices
Predicate Device (Passport 12m and 17m Patient Monitors K170876, T1 Patient Monitor K152902)
Integrated display and touchscreen17", 1280*1024 pixels12", 800*600 pixelsSameSame
Secondary displayIndependent display and displayDisplay is linked to integrated displayIndependent control via a VGA port in the docking T1 stationSame
Additional display featuresThe minitrends diagram, OxyCRG diagram, other monitor view, and calculation can be viewed when using an external LCD screenSameSameSame
Wireless moduleThe ASUS module is used for connecting to a network wirelessly, constructing a monitoring network with a central monitoring system (CMS).The ASUS, Silex and Laird modules are used for connecting to a network wirelessly, constructing a monitoring network with a central monitoring system (CMS).The Cyberlink modules is used for connecting to a network wirelessly, constructing a monitoring network with a central monitoring system (CMS).Laird (2.4GHz/5GHz) wireless module is added

Page 6 of 26

{10}------------------------------------------------

FeaturePredicate Device (Passport 12m and 17m Patient Monitors K170876, T1 Patient Monitor K152902)Subject Devices
Passport 17mPassport 12mT1Passport 17mPassport 12mT1
Module rackIndependent of the patient monitor, provides 8 standard module slots to extend the measurement capabilities of the systemNot supportedSameSameSame
Power supplyTwo rechargeable Lithium-ion battery or AC power supplyOne rechargeable Lithium-ion battery or AC power supplyOne rechargeable Lithium-ion battery or DC power supply or AC power supplySameSameSame
BatteryChargeable Lithium-Ion, 11.1 VDC, 4500 mAh, 350 gChargeable Lithium-Ion, 7.4 VDC, 2500 mAhSameSameSame
External memory cardCompact FlashSecure DigitalSameSameSame
Data RecorderThe thermal recorder records patient information, measurement numerics, up to three waveforms, etc.Not supportedSameSameSame
SpeakerGive alarm tones (45 to 85 dB), key tones, QRS tones; support PITCH TONE and multi-level tone modulationSameSameSame
Supports T1 as a moduleSupported/SameSameSame
Supports N1 as a moduleNot supportedNot supportedSupportedSame
Predicate Device (Passport 12m and 17m Patient Monitors K170876, T1 Patient Monitor K152902)Subject Devices
FeaturePassport 17mPassport 12mT1Passport 17mPassport 12mT1
MPM moduleMPM 2.0 module, it support:3/5/12 lead ECG, NIBP, dual channel Temp, SpO2, dual channel IBP measurement//MPM 3.0 module is added, it support:3/5/12 lead ECG, NIBP, dual channel Temp, SpO2, dual channel IBP measurementWhen MPM 3.0 module is used with Passport 12m and 17m, the functionality and performance is same with MPM 2.0 module./
ECG3-lead, 5-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, QT analysis, heart rate (HR), an interpretation of resting 12-lead ECG, J-point Auto detection, Dual Channel Pace detection and Adjustable QRS Detection ThresholdSame
Arrhythmia Analysis (Mindray Algorithm)Asystole, VFib/VTac, Vtac, Vent. Brady, Extreme Tachy, Extreme Brady, PVC, Couplet, Bigeminy, Trigeminy, R on T, Run PVCs, PVCs, Tachy, Brady, Missed Beats, Vent. Rhythm, PNP, PNC, Multif. PVC, Nonsus. Vtac, Pause, Irr. Rhythm, AFibSame
Arrhythmia Analysis (Mortara Algorithm)Asystole, Vfib, Vtac, Vent. Rhythm, Couplet, Run PVCs, Bigeminy, Trigeminy, R on T, Multif. PVC, Irr. Rhythm, Tachy, Brady, Missed Beats, PNP, PNC, PVCsSame
Predicate Device (Passport 12m and 17m Patient MonitorsK170876, T1 Patient Monitor K152902)Subject Devices
FeaturePassport 17mPassport 12mT1Passport 17mPassport 12mT1
Respiration rate(Resp)Uses the MPM (Multi Parameter Module) to measure the change in impedance measured across the thorax to determine the respiration rate.Same
Temperature(Temp)Uses the MPM (Multi Parameter Module) to measure core and skin temperature. The patient monitor can monitor two temperatures simultaneously using thermally sensitive resistors (thermistors).Same
Pulse oxygen saturation (SpO2)Uses the MPM (Multi Parameter Module) or the SpO2 module to measure the amount of oxygenated haemoglobin and pulse rate by measuring the absorption of selected wavelengths of light.Is compatible with the following 3 modules to measure oxygen saturation:Mindray SpO2 ModuleMasimo SpO2 ModuleNellcor SpO2 ModuleSame
FeaturePredicate Device (Passport 12m and 17m Patient Monitors K170876, T1 Patient Monitor K152902)Subject Devices
Passport 17mPassport 12mT1Passport 17mPassport 12mT1
PR from Mindray SpO2Measurement range: 20 to 254 bpmAccuracy: ±3 bpmPR from Mindray SpO2Same
PR from Masimo SpO2Measurement range: 25 to 240 bpmAccuracy: ±3 bpm (measured without motion)±5 bpm (measured with motion)PR from Masimo SpO2Same
Pulse rate (PR)PR from Nellcor SpO2Measurement range: 20 to 300 bpmAccuracy: 20 to 250 bpm: ±3 bpm251 to 300 bpm, not specifiedPR from Nellcor SpO2Same
PR from IBPMeasurement range: 25 to 350 bpmAccuracy:±1 bpm or ±1%, whichever is greaterPR from IBPSame
PR from NIBPMeasurement range: 40 to 240 bpmAccuracy: ±3 bpm or ±3%, whichever is greaterPR from NIBPMeasurement range: 30 to 300 bpmAccuracy: Same
FeaturePredicate Device (Passport 12m and 17m Patient Monitors K170876, T1 Patient Monitor K152902)Subject Devices
Passport 17mPassport 12mT1Passport 17mPassport 12mT1
Non-invasive blood pressure (NIBP)Measurement range:Adult:Systolic: 40-270 mmHgDiastolic: 10-210 mmHgMean:20-230 mmHgPediatric:Systolic:40-200mmHgDiastolic:10-150mmHgMean:20-165mmHgNeonate:Systolic:40-135 mmHgDiastolic:10-100mmHgMean:20-110 mmHgAccuracy:Max mean error: ±5 mmHgMax standard deviation: 8 mmHgMeasurement range:Adult:Systolic: 25-290 mmHgDiastolic: 10-250 mmHgMean:15-260 mmHgPediatric:Systolic:25-240mmHgDiastolic:10-200mmHgMean:15-215mmHgNeonate:Systolic:25-140 mmHgDiastolic:10-115 mmHgMean:15-125 mmHgAccuracy:Same
Invasive blood pressure (IBP)Uses the MPM (Multi Parameter Module) or the IBP Module to measure invasive blood pressure. The monitor can monitor up to 8 invasive blood pressures and displays systolic, diastolic and mean pressures and a waveform for each pressure. Supports the PPV function.Same
FeaturePredicate Device (Passport 12m and 17m Patient Monitors K170876, T1 Patient Monitor K152902)Subject Devices
Passport 17mPassport 12mT1Passport 17mPassport 12mT1
Pulse Pressure Variation (PPV)Supported feature of IBPSame
Cardiac output (C.O.)The temperature change is displayed as a curve in the C.O. split screen, and the monitor calculates the C.O. value from this curve. The monitor is capable of storing 6 measurements.Not supportedSameSame
Continuous cardiac output (CCO)CCO/SvO2 interface module is used to interface with Edwards Vigilance II monitor / Vigileo Monitor which measures continuous cardiac output (CCO), mixed venous oxygen saturation (SvO2) and central venous oxygen saturation (ScvO2).Not supportedSameSame
Central venous oxygen saturation (ScvO2)ScvO2 module is used to measure central venous oxygen saturation (ScvO2).Not supportedSameSame
FeaturePredicate Device (Passport 12m and 17m Patient Monitors K170876, T1 Patient Monitor K152902)Subject Devices
Passport 17mPassport 12mT1Passport 17mPassport 12mT1
Carbon dioxide (CO2)Is compatible with the following 3 modules to measure carbon dioxide:Sidestream CO2 Module Microstream CO2 Module Mainstream CO2 Module CO2 monitoring is based on calculations from measuring the absorption of infrared (IR) light of specific wavelengths using a photodetector.Sidestream CO2 2.0 module is added, the functionality and performance is same with Sidestream CO2 1.0 module.The other modules and specifications remains the same.Same
Anesthetic gas (AG)Is compatible with the following 2 modules to measure Anesthetic gas:3-slot AG Module 2-slot AG Module The AG module analyzes gas samples from the patient and calculates CO2, O2, N2O and AA waves and related numerics, airway respiratory rate, and MAC (minimum alveolar concentration).Supports the 3-slot AG module.Support Sample Gas Recycling feature for AG moduleNot supportedSame
FeaturePredicate Device (Passport 12m and 17m Patient MonitorsK170876, T1 Patient Monitor K152902)Subject Devices
Passport 17mPassport 12mT1Passport 17mPassport 12mT1
Impedancecardiograph (ICG)Uses the ICG Module tomeasure a patient'shemodynamic statususing a non-invasivemethod based onthoracic electricalbioimpedance(TEB) technology.Not supportedSameSame
Bispectral index(BIS)Measured parameters: EEG, BIS, BIS L,BIS RNot supportedSameSame
Respirationmechanics (RM)Measures respirationmechanics foradults, pediatricsand infants.Not supportedSameSame
Predicate Device (Passport 12m and 17m Patient MonitorsK170876, T1 Patient Monitor K152902)Subject Devices
FeaturePassport 17mPassport 12mT1Passport 17mPassport 12mT1
NMTUses the NMT module to monitor objective neuromuscular transmission.It evaluates muscle relaxation of patients under neuromuscular block by measuring the strength of muscle reaction after electrically stimulating the dedicated motor nerve.Not supportedSameSameSame
FeaturePassport 17mPassport 12mT1Subject DevicesPassport 17mPassport 12mT1
cleaninganddisinfectingagentsList of agents:Sodium hypochlorite bleachHydrogen peroxideIsopropanolRely+On TMVirkon®1-PropanolPerform®More agents are added:Metrex CaviCide1™Virex® II 256Virex® TBAlpet® D2 Surface Sanitizing WipesClorox Dispatch® Hospital Cleaner DisinfectantTowels with BleachClorox Healthcare® Bleach Germicidal WipesClorox HealthcareⓇ Hydrogen Peroxide CleanerDisinfectant WipesDiversey Oxivir® TB WipesMetrex CaviWipes™PDI Sani-Cloth® AF3 Germicidal Disposable WipePDI Sani-Cloth® Bleach Germicidal Disposable WipePDI Sani-Cloth® HB Germicidal Disposable WipePDI Sani-Cloth® Plus Germicidal Disposable ClothPDI Super Sani-Cloth® Germicidal Disposable WipeVIRAGUARD Hospital Surface Disinfectants
Predicate Device (Passport 12m and 17m Patient Monitors K170876, T1 Patient Monitor K152902)Subject Devices
FeaturePassport 17mPassport 12mT1Passport 17mPassport 12mT1
Support new Masimo SpO2 accessoriesAdded:040-003378-00RD SET MD 14-05, PC 5 ft
040-003379-00RD SET MD 14-12, Patient Cable 12 ft
040-003310-008pin masimo Cable (RD SET)
040-003426-00LNCS to RD Adapter
040-003381-00RD to LNC Adapter Cable
040-003376-00RD SET DCI, Adult Reusable Sensor
Not supported040-003377-00RD SET DCI, Pediatric Reusable Sensor
040-003380-00RD Set TC-I SpO2 Reusable Tip-Clip Ear Sensor, 3ft
040-003382-00RD SET Adhesive Sensor
040-003383-00RD SET PDT Adhesive Sensor
040-003384-00RD Set Infant Adhesive Sensor
040-003385-00RD Set Neo Adhesive Sensor
040-003386-00RD Set NeoPt Adhesive Sensor
040-003387-00RD Set NeoPt-500 Non-adhesive sensor
Support new ECG accessoriesNot supportedAdded:040-003528-00ECG cable, defibrillation-proof, AHA12-lead,
FeaturePassport 17mPassport 12mT1Predicate Device (Passport 12m and 17m Patient Monitors K170876, T1 Patient Monitor K152902) Subject Devices
Support new CO2 accessoriesNot supportedPassport 17mPassport 12mT1
Added:
100-000138-00DRYLINE PRIME Gas Sampling Line With Airway Adapter, Adult/Pediatric
100-000139-00DRYLINE PRIME Gas Sampling Line With Airway Adapter, Neonate
100-000140-00RYLINE PRIME+ Gas Sampling Line With Airway Adapter, Adult/Pediatric
100-000141-00DRYLINE PRIME+ Gas Sampling Line With Airway Adapter, Neonate
100-000142-00DRYLINE PRIME Nasal Gas Sampling Line, Adult
100-000143-00DRYLINE PRIME Nasal Gas Sampling Line, Pediatric
100-000144-00DRYLINE PRIME Nasal Gas Sampling Line, Neonate
100-000145-00DRYLINE PRIME+ Nasal Gas Sampling Line, Adult
100-000146-00DRYLINE PRIME+ Nasal Gas Sampling Line, Pediatric
100-000147-00DRYLINE PRIME+ Nasal Gas Sampling Line, Neonate
100-000151-00DRYLINE PRIME Water Trap Mini
Predicate Device (Passport 12m and 17m Patient Monitors K170876, T1 Patient Monitor K152902)Subject Devices
FeaturePassport 17mPassport 12mT1Passport 17mPassport 12mT1
Improve the SpO2/CO2/BIS signal cord resistance to damageNot supportedSupportedNote: T1 doesn't support BIS module
Corrected an issues where NIBP misreported "NIBP Excessive Motion" after 49 days of runningNot supportedSupported
FeaturePredicate Device (Passport 12m and 17m Patient MonitorsK170876, T1 Patient Monitor K152902)Subject Devices
Passport 17mPassport 12mT1Passport 17mPassport 12mT1
The End tidal (Et) and Fraction ofInspired (Fi) numerics foranesthetic agent (AA)Not supportedSupportedSame
Physiological alarms are trigged only when the monitored parameter value is higher than the high alarm limit or is lower than the low alarm limitNot supportedSupportedSame
Extend the setting range of low alarm limit of EtO2Not supportedSupportedSame

Page 7 of 26

{11}------------------------------------------------

Page 8 of 26

{12}------------------------------------------------

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

{16}------------------------------------------------

Page 13 of 26

{17}------------------------------------------------

{18}------------------------------------------------

{19}------------------------------------------------

{20}------------------------------------------------

Page 17 of 26

{21}------------------------------------------------

Page 18 of 26

{22}------------------------------------------------

{23}------------------------------------------------

Page 20 of 26

{24}------------------------------------------------

Predicate Device (Passport 12m and 17m Patient Monitors K170876, T1 Patient Monitor K152902)Subject Devices
FeaturePassport 17mPassport 12mPassport 17mPassport 12mT1
Support the non-overlapping relationships of alarm limit setting of physiological alarmsNot supportedSupported
Support re-alarming after a physiological alarm is resetNot supportedSupported
Support password protection for accessing the Alarm Setup menuNot supportedSupported
Support authorization of the user via the LDAPNot supportedSupported
Support the Facility information of patient monitorNot supportedSupported
Predicate Device (Passport 12m and 17m Patient Monitors K170876, T1 Patient Monitor K152902)Subject Devices
FeaturePassport 17mPassport 12mT1Passport 17mPassport 12mT1
Support the Room No. information of patient monitorNot supportedSupported
Support the Visit Number in the patient demographicsNot supportedSupported
Support printing of the arrhythmia settings in reportsNot supportedSupported
Change the factory default setting of printer paper sizeNot supportedSupported
Support to display the MAC address of patient monitorNot supportedSupported
Support the display mode of ST values in the parameter tileNot supportedSupported
Predicate Device (Passport 12m and 17m Patient Monitors K170876, T1 Patient Monitor K152902)Subject Devices
FeaturePassport 17mPassport 12mT1Passport 17mPassport 12mT1
Support encryption of patient private information sent to the networkNot supported
Support Laird 2.4GHz/5GHz Wifi moduleSupportedNot supportedSameSupported
Support adjustment of the ST point on multiple displayed leadsSupportedNot supportedSameSupported
Support QT report printingSupportedNot supportedSameSupported
Support NIBP measurement on clockSupportedNot supportedSameSupported
Support network application transmission priority settingSupportedNot supportedSameSupported
Support ECG beat annotationSupportedNot supportedSameSupported
Predicate Device (Passport 12m and 17m Patient Monitors K170876, T1 Patient Monitor K152902)Subject Devices
FeaturePassport 17mPassport 12mT1Passport 17mPassport 12mT1
Support DNS for ADT and LDAP server addressNot supportedSameSupported
Corrected an issues where the Bed No. of the Passport 12m/17m is changed after T1 is pulled out of the monitorsNot supportedSameSupported

Page 21 of 26

{25}------------------------------------------------

{26}------------------------------------------------

{27}------------------------------------------------

{28}------------------------------------------------

Performance Data

  • To establish the substantial equivalence of the Passport Series Patient Monitors O (Passport 12m, Passport 17m and T1), Mindray conducted functional and system level testing on the subject devices. The testing provided an evaluation of the performance of the device relevant to each of the modifications to the subject devices since their previous clearance. The functional and system level testing showed that the devices continue to meet specifications and the performance of the device is equivalent to the predicate.
    O In addition, Mindray has conducted testing to ensure the subject devices meet relevant consensus standards.

ANSI/AAMI ES60601-1:2005/(R)2012 and C1:2009/(R)2012 and A2:2010/(R)2012 Medical electrical equipment—Part 1: General requirements for basic safety and essential performance

IEC 60601-1-2 :2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests

IEC 60601-2-25: 2011 (2nd edition) Particular requirements for the basic safety and essential performance of electrocardiographs

IEC 60601-2-27:2011 (3rd edition) Particular requirements for the safety, including essential performance of electrocardiographic monitoring equipment ISO 80601-2-61:2011 Particular requirements for the basic safety and essential performance of pulse oximeter equipment

IEC 80601-2-30:2009/AMD1:2013 Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers ISO 80601-2-56:2009 (First edition) Particular requirements for the basic safety and essential performance of clinical thermometers for body temperature measurement

IEC 60601-2-34 (Third Edition): 2011 Particular requirements for the basic safety and essential performance of invasive blood pressure monitoring equipment

IEC 60601-2-49:2011 Medical electrical equipment - Part 2-49: Particular requirements for the basic safety and essential performance of multifunction patient monitoring equipment

ISO 80601-2-55 First edition 2011-12-15 Medical electrical equipment - Part

{29}------------------------------------------------

2-55: Particular requirements for the basic safety and essential performance of respiratory gas monitors

Substantial Equivalence Conclusion:

Based on the detailed comparison of specifications for each of the modifications to the previously cleared Passport Series devices, Passport 12m and Passport 17m (K170876) and relevant reference predicates, T1 (K152902) and relevant reference predicates, the performance testing results and conformance with applicable standards show that the Passport Series Patient Monitors (including Passport 12m, Passport 17m and T1) can be found substantially equivalent to the predicate devices.

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.